ADVIA® 60 Hematology System
Operator’s Guide
Closed Tube Model
075D0002-01 Rev. A, 2003-09
Related Manuals for Bayer HealthCare ADVIA 60
Summary of Contents for Bayer HealthCare ADVIA 60
-
Page 1
ADVIA® 60 Hematology System Operator’s Guide Closed Tube Model 075D0002-01 Rev. A, 2003-09… -
Page 2
If the system is used in a manner differently than specified by Bayer HealthCare LLC, the protection provided by the equipment may be impaired. See warning and hazard statements. -
Page 3
The LED stops blinking and the tube holder carriage moves down when the sampling is completed. Shutdown procedure 1. Perform a general cleaning/rinse cycle of the ADVIA 60 Hematology System using the STAND BY key on the front panel. This cycle lasts for approximately one (1) minute. The instrument will cycle into the STAND BY mode. -
Page 5: Table Of Contents
Table of Contents CHAPTER 1. INTRODUCTION ………………. 1-1 1.1. INTENDED USE ……………………..1-1 1.2. PRESENTATION ……………………..1-2 1.3. OPEN TUBE AND CLOSED TUBE MODELS ………………1-2 1.4. NOTES ……………………….. 1-2 CHAPTER 2. INSTALLATION ………………2-1 2.1. INSPECTION ………………………. 2-1 2.2. UNPACKING ……………………….. 2-1 2.3.
-
Page 6
4.3. STUDY OF GENERAL FLAGS ………………….4-6 4.3.1. Platelet flags ……………………..4-7 4.3.2. HGB flag ……………………….. 4-7 4.3.3. WBC Flags ……………………..4-7 4.3.4. Comments on the flagging capabilities ………………. 4-8 CHAPTER 5. DESCRIPTION ………………5-1 5.1. INSTRUMENT ……………………… 5-1 5.1.1. Front panel and command keys ………………..5-1 A — Using the keyboard …………………… -
Page 7
7.5.4. QC print targets ……………………. 7-16 7.5.5. QC statistics ……………………..7-17 7.5.5.1. Select statistics ……………………7-17 7.5.5.2. Select level ……………………..7-17 7.5.6. QC Graphs ……………………..7-18 7.5.6.1. Select graphs ……………………7-18 7.5.6.2. Select level ……………………..7-18 CHAPTER 8. INSTRUMENT CONFIGURATION …………8-1 8.1. RESULTS OPTIONS ……………………8-1 8.1.1. -
Page 8
9.1.3.1. Backflush ……………………..9-2 9.1.3.2. Drain chambers ……………………9-2 9.1.3.3. Prime reagents ……………………9-2 9.1.3.4. Concentrated cleaning ………………….9-3 9.1.3.5. Mechanical checks ………………….. 9-4 9.1.3.6. Cycle functions ……………………9-6 9.1.3.7. Technician functions ………………….9-6 9.1.3.8. Automatic cleaning ………………….9-6 9.1.4 Disposal of System Waste and Supplies ……………… 9-6 9.2. -
Page 9: Chapter 1. Introduction
(Closed Tube) system is a fully automated (Microprocessor controlled) hematology analyzer used for the in vitro diagnostic testing of whole blood specimens. ADVIA 60- CT16 — WBC, LYM%, LYM#, MON%, MON#, GRA%, GRA#, — RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, and 3 distribution curves: WBC, RBC, PLT.
-
Page 10: Presentation
1.4. NOTES Bayer HealthCare retains the right to make changes to the instrument and this document. Your local technical support provider or distributor will provide the latest revision of this document. No part of this document may be copied or reproduced in any form or by any means without prior written consent of Bayer HealthCare.
-
Page 11: Chapter 2. Installation
CHAPTER 2. INSTALLATION 2.1. INSPECTION A thorough inspection is performed prior to the release of an ADVIA 60- Hematology System. It is important to verify receipt of all parts. Notify any descrepancies with the carrier. As instructed, the installation procedures must be followed in the order listed below.
-
Page 12: Package Contents
O ring 6 mm x 1.5 mm FAA 036 A O ring 0.74 mm x 1.02 mm FAA 054 A ADVIA 60 common installation kit XEA 312 A Table 2.1 — The ADVIA 60- common installation kit (XEA 312 A) includes :…
-
Page 13: Working Conditions
2.4.2. Location ADVIA 60- should be placed on a clean and level table or work station. Please note that ADVIA 60- printer and reagents weigh approximately 30 kilograms (66 lbs). Avoid exposure to sunlight. Proper ventilation requires that a space of at least 20 cm (8 inches) must be left behind the apparatus.
-
Page 14: Connection Check
2. INSTALLATION Push the black plastic carriage locking clip as far as possible to the left and place the sample needle carriage as far forward as possible to the right-hand side. Check that the aspiration needle is not bent and make sure it is in its upper position. 1 — Carriage locking clip 2 — Sample needle carriage 3 — Needle…
-
Page 15: Reagent Pack Connections
: they should be 1 Amperes, 220 Volts Slow-Blow. Diag.2.8 2.6. REAGENT PACK CONNECTIONS The ADVIA 60 Hematology System TIMEPAC ® Reagents includes the entire set of reagents in a «Pack» and is able to receive the waste liquids. Three soft pockets contain the 3 reagents sysDIL ®…
-
Page 16: Reagent And Waste Handling Precautions
The ADVIA 60- responds to the UL 3101 norm. Refer to the Declaration of Conformity Statement for details. System performance is guaranteed by Bayer HealthCare under the following condi- tions only: — services and repairs are provided by your local technical support provider or distributor — the electrical supply of the laboratory follows the national or international regulations — the system is operated under the following instructions.
-
Page 17: Connection
Lock the connector in place by tightening the 2 screws on each end of the connector to the ADVIA 60- . Attach the other end of the cable to the printer and lock the printer connector in place by the means of the 2 clips located on the connector itself.
-
Page 18
2. INSTALLATION From the MAIN MENU, move the cursor to the function SERVICE and press ENTER. The service menu is displayed : SERVICE > 2 DRAIN CHAMBER HH : MM 3 PRIME REAGENTS Move the cursor to PRIME REAGENTS and press the ENTER key. Select the function CHANGE PACK and follow the instructions given by the LCD in order to install the pack. -
Page 19: Chapter 3. Specifications
31°C decreasing linearly to 50% relative humidity at 40°C. — Smart Card option : * Reader : GCI 400 — ADVIA 60 SETpoint Calibrator memory card : — GFM 2K — Patient memory card : — MCOS 24K (capacity : 60 CBCs)
-
Page 20
3. SPECIFICATIONS — Storage capacity : Number of QC available # of QC Aspirations/day QC8P card QC16P card Table 3.1 — Hemoglobin : * HGB/WBC chamber, LED 550 nm. — Size of apertures : * WBC = 80 µm * RBC/PLT = 50 µm — Final dilutions : * WBC = approximately 1/250 * RBC / PLT = approximately 1/15000… -
Page 21
3. SPECIFICATIONS — Reagent pack * Total capacity : 4.2 liters — Measuring principles: * WBC.RBC.PLT = Impedance change * Hematocrit = Numeric integration * HGB = Cyanmethemoglobin method (550 nm) — Barcode reader option: * EAN 8, EAN 13, C 39, C 128, ITF (2/5), CODABAR, STF, C 93 with or without checksum. -
Page 22: Expected Values
3. SPECIFICATIONS — Carryover : Carry-over was tested by analyzing samples with high concentrations of WBC’s, RBC’s, HGB and PLT’s. Each sample was run in triplicate, followed by three background cycles. The % carryover is calculated using the following formula : Background 1 — Background 3 Carryover = ——————————————— X 100…
-
Page 23
3. SPECIFICATIONS — Accuracy : Approximately 200 patient specimens were analyzed on the ADVIA 60- and the ADVIA ® 120 Hematology Systems at three different locations. The following table summarizes the data: SITE 1 SITE 2 SITE 3 PARAMETER WBC (x 10 0.996… -
Page 24: Reagent Specifications
Physio-chemical properties : Boiling point : About 100°C, pH : neutral. 3.2.2. Lyse : sysLYSE (1 x 200 mL) For in vitro diagnostic use as an eythrocyte-lysing agent for leukocyte counting, differentiation, and hemoglobin determination on the ADVIA 60 Hematology system. Composition : · Potassium cyanide, 0.03%.
-
Page 25: Detergent : Sysklen
3. SPECIFICATIONS 3.2.3. Detergent : sysKLEN Function : For in vitro diagnostic use in hte cleaning of the Advia 60 Hematology system. Composition: Organic buffer, < 0.2%. Physio-chemical properties: Boiling point : around 100°C, pH = 9.7 ± 0.2 at 20°C.
-
Page 26: Caps Of The Sampling Tubes
3. SPECIFICATIONS 3.3.2. Caps of the sampling tubes Some caps of the sampling tubes are more adapted to «cap piercing» sampling systems. Plastic caps cannot be used. Rubber caps can be of different quality materials. Use the best quality materials in order to avoid any rubber particles entering the sample tube when the cap is pierced. It is also recommended to use caps specially designed to avoid any blood retention in the upper part of the cap.
-
Page 27
3. SPECIFICATIONS Following the manual differential white blood cell count, the WBC assay value must be corrected for the presence of nucleated red blood cells. The formula utilized for correcting the WBC parameter, when nucleated red blood cells are present, is: counted WBC X 100 CORRECT WBC: ————————————-… -
Page 28
3. SPECIFICATIONS HGB (Hemoglobin): Turbidity of the blood sample — Due to any number of physiologic and/or therapeutic factors may produce falsely elevated HGB results. To obtain accurate hemoglobin results when increased turbidity of the blood sample occurs, determine the cause of the turbidity and follow the appropriate method below : 1. -
Page 29
3. SPECIFICATIONS RDW (Red Blood Cell Distribution Width): The red blood cell distribution width is a function of the RBC count. The red blood cell dilution contains all of the formed elements in the blood : erythrocytes, leukocytes and platelets. During the counting of the erythrocytes (Red Blood Cells) platelets are not included in RBC count since their size falls below the minimum threshold. -
Page 30
3. SPECIFICATIONS Agglutinated erythrocytes — May trap Platelets, causing an erroneous MPV result. The presence of agglutinated erythrocytes may be detected by observation of the abnormal MCH and MCHC values and by careful examination of the stained blood film. Chemotherapy — May also affect the sizing of PLTs. CAUTION: Blood samples collected in EDTA will not maintain a stable Mean Platelet Volume. -
Page 31: Chapter 4. Technology
CHAPTER 4. TECHNOLOGY 4.1. MEASUREMENT PRINCIPLES 4.1.1. RBC / WBC / PLT detection principles The counting principle used for red blood cells (RBC), white blood cells (WBC), and platelets (PLT) is based on the variation in impedance generated by the passage of cells through a calibrated microaperture.
-
Page 32
7 — Each type of cell (WBC, RBC, PLT) is analyzed by the microprocessor which also handles the cell distribution (histograms). 8 — To count the platelets, the ADVIA 60- uses high performing electronics, which avoids the use of complex hydraulic systems for the elimination of faulty impulses generated at the rear side of the aperture. -
Page 33: Hemoglobin Measurement Principle
3 — The result is given with the units set up in Section 8 «Instrument Configuration». 4.2. CELL DISTRIBUTION STUDY ADVIA 60- carries out volumetric distributions (histograms) for WBC and RBC on 256 analysis channels, and 128 channels for PLT with the following measuring range: * WBC = approximately 30 to 460 fL.
-
Page 34: Diluent And Lysing Action
4.2.1.3. Volumetric study After the differential lysing action, ADVIA 60- analyzes the heights of each impulse, coming from the aperture and ranked in the 256 counting channels. A curve is obtained with number of cells on the y-axis and cell volume on the x-axis.
-
Page 35: Distribution Of Red Blood Cells
4. TECHNOLOGY 4.2.2. Distribution of Red Blood Cells The study of the distribution of RBC detects erythrocyte anomalies linked to anisocytosis. A Red Cell Distribution Width (RDW) will enable you to follow the evolution of the width of the curve in relation to the cell number and average volume. K SD With System constant…
-
Page 36: Study Of General Flags
4. TECHNOLOGY 4.3. STUDY OF GENERAL FLAGS With the print-out of results the following general flags may occur : following either WBC, RBC, HCT or PLT indicates that the system, analyzed the sample three times but that all three counts differed and were outside the systems precision limits. The result should be verified by repeating the sample or checked according to one of the laboratory reference methods.
-
Page 37: Platelet Flags
If this suspicious flag occurs more than three consecutive times run the checkup procedure. 4.3.3. WBC Flags The ADVIA 60- has a system of WBC differential flags alerting the operator to the possible presence of pathological cells, abnormal volume distribution histograms or abnormal elevated populations such as in the case of the eosinophils and basophils.
-
Page 38: Comments On The Flagging Capabilities
The lab operator can adjust the limit which triggers this flag. 4.3.4. Comments on the flagging capabilities All anomalies and/or abnormal distributions signaled by the ADVIA 60- should be verified manually for the presence of pathological elements. As a result of the differential resistance of cytoplasmic membranes in the different cell types, pathological elements can be found in a number of different zones.
-
Page 39: Chapter 5. Description
Door to pneumatic part Piercing carriage door Front panel Diag.5.1 1 — ADVIA 60- Cover The instrument cover is fixed by the means of 5 screws. Before any attempt to remove the cover, open the pneumatic access door on the front of the system.
-
Page 40: B — Command Keys
5. DESCRIPTION Diag.5.2 STARTUP key DELETE key STAND BY key ENTER key Identification Key (ID/SEQ) ESCAPE key Start analysis cycle key (START) Display scroll keys Numerical keyboard Cycle light indicators B — Command keys 1 — STARTUP key: When this key is pressed, a startup cycle including a cleaning and rinsing procedure is carried out.
-
Page 41
5. DESCRIPTION 5 — Numerical keyboard : keys from 0 to 9 allow the operator to enter the following figures : — Date — Calibration values — Laboratory limits — Patient number for analysis — Leukocytes differential flag values. These 2 keys pressed simultaneously will allow an automatic adjustment of the LCD. -
Page 42: Rear Panel / Main Fuses
5. DESCRIPTION 5.1.2. Rear panel / Main fuses Rear panel RS 232 computer connection Printer output ON/OFF switch F1,F2 Main fuses Main power socket Diag.5.3 Diag.5.4 5.1.3. Left side internal view Vacuum/pressure syringe Liquid electrovalve block 1 Dilution block Liquid electrovalve block 2 Diag.5.5 Page 5/4 075D0002-01…
-
Page 43: Front Internal View
5. DESCRIPTION 5.1.4. Front internal view LCD display Keyboard Cycle light indicators Sampling needle carriage Start cycle key RBC chamber WBC chamber Spectrophotometer Piercing mechanism Tube holder switches Piercing carriage door switch Diag.5.6 075D0002-01 Page 5/5…
-
Page 44
5. DESCRIPTION Page 5/6 075D0002-01… -
Page 45: Chapter 6. Startup And Sample Run
CHAPTER 6. STARTUP AND SAMPLE RUN 6.1. STARTUP CHECKS 6.1.1. INSTRUMENT STARTUP Turn ON instrument by pressing the ON/OFF switch located on the rear panel. The display shows the following : PLEASE WAIT FOR 3 MIN ESCAPE : ESC Then the front panel LED turns from red to green to indicate that the initialization phase is completed. Wait for the end of the automatic startup cycle or press the STARTUP key, the instrument will perform a startup cycle and perform a blank cycle for a background count (an analysis cycle on reagent without any blood sample).
-
Page 46: Reagent Pack
Sample collection should be done on venous blood by the means of vacuum or atmospheric sample collection tubes. It is possible to collect capillary blood into a microtainer with a minimum volume of 100 µL (for example in a pediatric lab.) and analyze this on the ADVIA 60- . EDTA is the recommended anticoagulent for analysis on the ADVIA 60- Samples should be processed as soon as possible, and within 6-8 hours of collection.
-
Page 47: Daily Quality Control/Calibration Verification
6. STARTUP AND SAMPLE RUN 6.3. DAILY QUALITY CONTROL/CALIBRATION VERIFICATION Refer to your laboratory quality assurance program to ensure quality throughout the entire testing process. It is recommended that the system be controlled using Bayer TESTpoint Hematology Controls Low, Normal, and High (Prod. Nos. B03-4200-54, B03-4201-54, and B03-4202-54, respectively). Refer to the package insert supplied with the control for instructions for use.
-
Page 48: Us Mode Identification With Barcode Reader
6. STARTUP AND SAMPLE RUN 6.3.1. US mode identification without barcode reader Press the ID/SEQ key of the front panel to enter the control ID. The identification menu is displayed : PAT. ID. ? : EXIT : ESC CURRENT: SAVE : ENTER The identification can be entered using 13 characters, letters or numbers.
-
Page 49: Standard Identification Mode
6. STARTUP AND SAMPLE RUN 6.3.3. Standard identification mode Press the ID/SEQ key to enter the control Run #. The following menu is displayed: RUN #? : EXIT : ESC NEXT : SAVE : ENTER Enter the run number (from 1 to 9999) using the numeric keys of the front panel. Press the ENTER key to record the number (or the ESC key to save the displayed one) : A message «PLEASE CLOSE HOLDER DOOR»…
-
Page 50: Analysis
6. STARTUP AND SAMPLE RUN 6.3.5. Analysis Install the sample tube in the tube holder and close the sample door to start the analysis if this mode has been selected or close the sample door and press the START key. Diag.
-
Page 51: Running Samples
6. STARTUP AND SAMPLE RUN 6.4. RUNNING SAMPLES 6.4.1. Sample identification After completing the start up procedure, run a non pathological blood from the previous day, followed by the quality control analyses and then the sample analyses. 1 — Enter either the patient identification (US mode) or the patient run # (Standard mode). 2 — Install the sample tube in the tube holder and close the sample door to start the analysis if this mode has been selected or close the sample door and press the START key.
-
Page 52: Results
6. STARTUP AND SAMPLE RUN 6.5. RESULTS When the analysis cycle is completed, results are displayed and printed out according to the setup of the instrument: 6.5.1. Displayed results The first group of parameters is displayed: MCHC 5.22 15.6 46.5 29.9 33.5 The second group of parameters can be accessed when moving the cursor to the top:…
-
Page 53: Result Printout
6. STARTUP AND SAMPLE RUN 6.5.2. Result printout 1. US mode Diag. 6.4 On the result printout can be found: 1 — The sample identification that was entered by the operator. 2 — The sequence number. 3 — The STARTUP status. 4 — The PLT flags.
-
Page 54
6. STARTUP AND SAMPLE RUN 2. Standard mode Diag. 6.5 On the result printout can be found: 1 — The sample run # that was entered by the operator. 2 — The sequence number. 3 — The PLT flags. 4 — The CBC results. 5 — The WBC flags. -
Page 55: Chapter 7. Calibration And Quality Control
Calibration can be achieved in two different ways. 1 — Bayer SETpoint Calibrator (B03-4203-51) is used to calibrate the ADVIA 60- Hematology System.
-
Page 56: Smart Card Option
7. CALIBRATION AND QUALITY CONTROL 7.2. SMART CARD OPTION When the instrument is equipped with a smart card reader proceed as follows. Install the card in the smart card reader. Make sure that the cursor is positioned on function AUTOCALIBRATION and press ENTER. Diag.
-
Page 57: Change Lot Number
7. CALIBRATION AND QUALITY CONTROL 7.3.2. Change lot number CHANGE LOT # ? NO : ESC CURRENT : SETPOINT YES : ENTER The current lot number is displayed. If the lot number of the calibration blood has to be changed, press ENTER.
-
Page 58: Change Expiration Date
7. CALIBRATION AND QUALITY CONTROL 7.3.3. Change expiration date The current expiration date is displayed. CHANGE EXP. DATE ? (DD.MM.YY) NO : ESC CURRENT : YES : ENTER If it is correct, press the escape key, the menu turns to the target values. If the expiration date needs to be changed, press ENTER, the following menu is displayed : EXP.
-
Page 59: Change Number Of Calibrator Samples
NO : ESC CURRENT : YES : ENTER ADVIA 60- calibration may be performed using 3 to 11 sample aspirations. The autocalculation module performs statistics on these results in order to obtain the best calibration coefficients. In order to obtain the best calibration possible, it is recommended to run the calibration NOTE: blood a minimum of 5 times.
-
Page 60: Run Calibration
7. CALIBRATION AND QUALITY CONTROL 7.3.6. Run calibration Prepare the calibrator according to the specific instructions (temperature, mixing, etc…). RUN CAL. ? NO : ESC YES : ENTER Press the ENTER key. The start calibration menu is displayed : START CALIBRATION # 1/X ESC TO EXIT CALIBRATION ENTER TO ASPIRATE…
-
Page 61
7. CALIBRATION AND QUALITY CONTROL A — Calibration passed : If the statistical figures are within the acceptable limits: — Coefficient of variation is within the limits given in Table 7.1 and, — The percentage difference between the target and the mean value is less than 20. The calibration passed and the results are printed out. -
Page 62
7. CALIBRATION AND QUALITY CONTROL B — Calibration failed: If the statistical figures are not within the acceptable limits, the following menu is displayed: CALIBRATION FAILED !!! PRESS A KEY TO CONTINUE … — Coefficient of variation is not within the limits given in Table 7.1 or, — The percentage difference between the target and the mean value is greater than 20. -
Page 63: Calibration Coefficients
7. CALIBRATION AND QUALITY CONTROL 7.4. CALIBRATION COEFFICIENTS Calibration can be achieved directly by changing the calibration coefficients when they are known. Move the cursor to function COEFFICIENTS and press ENTER. The following menu is displayed: COEFFICIENTS 1 CALIB. COEFF. HH:MM 2 PRINT COEFF.
-
Page 64: Print Coefficients
7. CALIBRATION AND QUALITY CONTROL 7.4.2. Print coefficients From the COEFFICIENTS menu, it is possible to print the coefficient values. Move the cursor to function PRINT COEFF. and press ENTER. The printout of the coefficient values starts automatically. COEFFICIENTS DATE : 01/11/1997 TIME : HH : MM CURRENT : 0.97…
-
Page 65: Quality Control Program
7. CALIBRATION AND QUALITY CONTROL 7.5. QUALITY CONTROL PROGRAM 7.5.1. Introduction The ADVIA 60- quality control program contains five different functions : 1 — AUTOMATIC — The function of this submenu is to allow the operator to analyze the commercial control bloods and store the results on the memory smart card.
-
Page 66: Qc — Automatic
7. CALIBRATION AND QUALITY CONTROL 7.5.2. QC — Automatic From the QC menu, move the cursor to function AUTOMATIC and press ENTER. The AUTOMATIC menu will unfold step by step through the smart card insertion, the operator selection, the commercial control identification and lot, etc … 7.5.2.1.
-
Page 67: Select Operator
7. CALIBRATION AND QUALITY CONTROL List of the different messages : NEW QC : The smart card has not been used, it is the beginning of a new QC. XX QC RUN : xx QC are stored on this smart card and the index in the smart card and in the analyzer are identical.
-
Page 68: Run Commercial Control
7. CALIBRATION AND QUALITY CONTROL 7.5.2.4. Run commercial control Prepare the commercial control according to the specific instructions (temperature, mixing, etc … ). Install the calibrator tube in the correct holder position. Close the sample door to start the analysis if this mode has been selected or close the sample door and press the cycle button located on the front panel.
-
Page 69: Exiting Qc Automatic
7. CALIBRATION AND QUALITY CONTROL 1 — ACCEPT: If the results are accepted they will be stored on the QC smart card and the display will return to the SELECT LEVEL. SELECT LEVEL > * 1 LOW BLOOD HH:MM 2 HIGH BLOOD It is possible to exit the QC or run another control level.
-
Page 70: Analysis
7. CALIBRATION AND QUALITY CONTROL 7.5.3. Analysis This function ANALYSIS allows the operator to run a control as a normal analysis cycle but with specific LMG thresholds for QC blood (independant from the temperature). From the QC main menu, move the cursor to the function and press ENTER.
-
Page 71: Qc Statistics
7. CALIBRATION AND QUALITY CONTROL 7.5.5. QC statistics The commercial control files can be printed for permanent records at any time. It is recommended to print the files at the end of each month. Each file printout includes the following information: File Name (blood level), Lot number of control, expiration date of control, date and time of print request, date and time of the run, operator and the data points for each stored quality control run, the reference means and upper and lower limits, the actual mean results of the quality controls runs, the 2 standard deviation value, and the…
-
Page 72: Qc Graphs
ETC ..Diag. 7.9 7.5.6. QC Graphs The ADVIA 60- plots Levey-Jennings charts for each parameter of the quality control files stored. The Levey-Jennings chart will plot one point for every control data point stored. Beneath each chart, the Reference Mean, two SD value and the actual mean, two SD value and % CV are provided.
-
Page 73
7. CALIBRATION AND QUALITY CONTROL Using the UP or DOWN arrows select the commercial control level to print one set of graphs or all graphs automatically. Press ENTER and the following message appears: PRINTING QC RESULTS PLEASE WAIT… All graphs will be printed as shown in the example below. NOTE: QC graphs will be printed out even when values are equal to zero. -
Page 74
7. CALIBRATION AND QUALITY CONTROL Page 7/20 075D0002-01… -
Page 75: Chapter 8. Instrument Configuration
CHAPTER 8. INSTRUMENT CONFIGURATION The ADVIA 60- has several operator options : — Specific laboratory limits. — Date and time formats. — Result format. — RS 232 options. — Special functions. These options can be configurated according to the operator needs through the SETUP function of the MAIN MENU.
-
Page 76: Reprint Results
8. INSTRUMENT CONFIGURATION 8.1.1. Reprint results To reprint the results of the last sample in memory, move the cursor to function of the RESULTS menu and press ENTER. The results of the last sample is automatically reprinted with the date and time, the associated identification, sample run and sequence numbers, the possible flags and the histograms if their printout is selected.
-
Page 77: Printer Selection
8. INSTRUMENT CONFIGURATION 8.1.4. Printer selection Four different printers and a no printout option can be selected from the RESULTS menu. Move the cursor to the function PRINTER and press ENTER. The PRINTER menu is displayed: PRINTER > * 1 RESERVED 1 HH:MM 2 RESERVED 2 Move the cursor to the required printer :…
-
Page 78: Print Limits
8. INSTRUMENT CONFIGURATION 8.1.6. Print limits The laboratory limits can be printed out underneath each result when this option is selected. From the RESULTS menu, move the cursor to function and press ENTER. The PRINT LIMITS menu is displayed: PRT LIMITS >…
-
Page 79: Result High Limits
8. INSTRUMENT CONFIGURATION Move the cursor next to the required parameter and press ENTER, the following menu (RBC for example) is displayed: RBC LOW ?: EXIT : ESC CURRENT : SAVE : ENTER Enter the required low value and press ENTER or press the ESC key to keep the current value. Move the cursor to the next required parameter and follow the same procedure.
-
Page 80: Flag Limits
8. INSTRUMENT CONFIGURATION 8.2.4. Flag limits A — Platelet flag adjustment From the menu CHG LAB LIMITS, move the cursor to the function FLAGS and press ENTER. The flag menu is displayed: FLAGS LIMITS > 1 SCL <8.0> HH:MM 2 SCH <8.0>…
-
Page 81: Special Functions
8. INSTRUMENT CONFIGURATION 8.3. SPECIAL FUNCTIONS These special functions accessible through a password allow the user to : 1 — Identify 4 users 2 — Change the password 3 — Choose the startup mode 4 — Adjust the cleaning frequency 5 — Print the internal setup of the instrument 6 — Set ON/OFF the cycle end audible signal 7 — Choose the identification mode…
-
Page 82: Change Password
8. INSTRUMENT CONFIGURATION 8.3.2. Change password The use of a password is mandatory to access some important functions such as: — Changing the calibration factors. — Accessing the technician functions. — Changing the password. The original password is <123>. If the change is requested, move the cursor from the SPECIAL menu to function and press ENTER.
-
Page 83: Autocleaning Frequency
8. INSTRUMENT CONFIGURATION 8.3.4. Autocleaning frequency An automatic cleaning involving a cleaner solution is normally carried out every 50 samples. The user has the possibilty to setup the automatic cleaning frequency according to the number of samples usually run in his laboratory. From the SPECIAL menu, move the cursor to the function CLEAN.
-
Page 84: Cycle End Audible Signal
8. INSTRUMENT CONFIGURATION 8.3.6. Cycle end audible signal A cycle end audible signal (BEEP) can be set up with the SPECIAL menu. Move the cursor to the function BUZZER and press ENTER. The audible signal menu is displayed: BUZZER > * 1 ON HH:MM 2 OFF Move the cursor on the required option and press ENTER.
-
Page 85: Start Mode
8. INSTRUMENT CONFIGURATION 8.3.8. Start mode The START MODE function allows the operator to choose between 2 different analysis starting modes: — In the automatic mode, the analysis can start directly when the operator closes the sample door, — In the manual mode, the analysis can start after closing the sample door, when the operator presses the START cycle key.
-
Page 86: Change Time
8. INSTRUMENT CONFIGURATION 8.4.1. Change time Move the cursor to the function and press ENTER. The CHANGE TIME menu is displayed: NEW TIME (HH.MM) ? : EXIT : ESC CURRENT : SAVE : ENTER Enter the required time in the format HH.MM and press ENTER. The new time is recorded. 8.4.2.
-
Page 87: Host Options
The ADVIA 60-CT has to be setup according to the external computer specifications. The following functions have to be used by your laboratory computer specialist only.
-
Page 88: Barcode Setup
8.7. Patient Memory Card 8.7.1. Introduction This menu is available only if the ADVIA 60- is equipped with a smart card reader. The MEMORY program contains the seven following functions: 1 — MEMO : allows the operator to enable/disable the memory function.
-
Page 89: Memo Operation Mode
8. INSTRUMENT CONFIGURATION 8.7.2. Memo operation mode 8.7.2.1. Memo on/off Insert the Patient Memory Card into the slot in the upper left corner of the analyzer. From the MEMOCARD menu, move the cursor to function MEMO and press ENTER. The following menu is displayed: MEMO >…
-
Page 90: Running The Samples
8. INSTRUMENT CONFIGURATION 8.7.3. Running the samples Place the sample tube in the sample holder and close the door to start the analysis cycle. The following menu is displayed: PAT ID ? : NO SAVE : ESC CURRENT SAVE : ENTER Enter the patient identification and press ENTER.
-
Page 91: Print List
8. INSTRUMENT CONFIGURATION 8.7.4. Print list The PRINT LIST option allows the operator to print out all the patient identifications recorded on the Patient Memory Card, the time and the date of the analysis as well as the backup number (column MEMO).
-
Page 92: Trans. All
8. INSTRUMENT CONFIGURATION 8.7.6. Trans. all This function allows the operator to reprint or to send to the host all the results stored on the smart card. Check the quantity of paper before running a REPRINT ALL. From the MEMOCARD (Patient Card) menu, move the cursor to TRANS.
-
Page 93: Chapter 9. Maintenance And Troubleshooting
Appendix B for recommended precautions when working with biohazardous materials. 9.1.2. Daily maintenance These cleaning procedures are required daily to maintain optimum performance of your ADVIA 60- CAUTION: Failure to perform any of these recommended cleaning procedures may result in decreased reliability of your system.
-
Page 94: Service Functions
9. MAINTENANCE AND TROUBLESHOOTING 9.1.3. Service functions Several service functions are available for the user to clean and check his instrument. These functions are accessible from the Main menu. Move the cursor to the Function SERVICE and and press ENTER. The SERVICE menu is displayed: SERVICE >…
-
Page 95: Concentrated Cleaning
9. MAINTENANCE AND TROUBLESHOOTING 9.1.3.4. Concentrated cleaning This procedure provides a strong cleaning of the RBC and WBC apertures. This cycle lasts approximately 3 minutes and involves the operator intervention to fill the chambers with sysCLEAR or diluted bleach solution. The diluted bleach solution is obtained by a 1:5 dilution of a commercial bleach solu- NOTE: tion containing from 10 to 15% of sodium hypochlorite.
-
Page 96: Mechanical Checks
F — Valves : Move the cursor to the function VALVES and press ENTER. Closely observe the valve operations, the movement has to be straight and regular. NOTE: On the ADVIA 60- , the valve #3 is not installed. Page 9/4 075D0002-01…
-
Page 97
9. MAINTENANCE AND TROUBLESHOOTING G — Change contrast : The display contrast can be adjusted for a better readability. Move the cursor to the function CHG CONTRAST and press ENTER. The CHG CONTRAST menu is displayed: CONTRAST PRESS ENTER TO VALID HH:MM Press the UP key of the numeric keyboard to increase the contrast, or press the DOWN key to reduce the contrast. -
Page 98: Cycle Functions
The principal wastes associated with the use of the ADVIA 60- Hematology System are the TIMEPAC reagent containers, and the sysCLEAR reagent.
-
Page 99: Troubleshooting
9. MAINTENANCE AND TROUBLESHOOTING 9.2. TROUBLESHOOTING 9.2.1. Overview Your ADVIA 60- may occasionally require troubleshooting if : — System operations are faulty. — The background count is unacceptably high. — Your control values are out-of-ranges, or patient results are suspicious (e.g., consistently high RBC counts or inability to verify results by manual methods).
-
Page 100: Calibration
9.2.2.4. Calibration If the system appears to be operating properly, fresh uncontaminated reagents are being used, and the precision is within the specifications, the ADVIA 60- may need calibration. Calibrate your instrument as described in section 7. 9.2.3. Troubleshooting parameters The procedures described below should be performed whenever the precision of a parameter is not within the specifications noted above, or a parameter result is incorrect or suspicious.
-
Page 101
9. MAINTENANCE AND TROUBLESHOOTING If it is, continue. If sample probe is not in its correct position in the mixing chamber, call your local technical support provider or distributor. Diag. 9.1 Diluent dispenser : Do you see bubbles in the diluent dispenser? Is the plunger of the diluent dispenser moving up and down smoothly during sample analysis? -
Page 102: Wbc
9. MAINTENANCE AND TROUBLESHOOTING 9.2.3.2. WBC Check the following if only the WBC count is incorrect or suspicious. Concentrated cleaning procedure : Was the concentrated cleaning procedure performed earlier as part of the identification procedure? If not, perform the concentrated cleaning procedure. Calibration : Was the system calibrated earlier as part of the identification procedure.
-
Page 103: Rbc, Hct And Plt
9. MAINTENANCE AND TROUBLESHOOTING HGB LED : Is the HGB LED illuminated when the system power is on? If it is and therefore all system operations are acceptable, call your local technical support provider or distributor. If the HGB LED is not illuminated when the system power is on, call your local technical support provider or distributor.
-
Page 104: Rbc
9. MAINTENANCE AND TROUBLESHOOTING RBC chamber : Is the sample probe between the edge of the chamber and the center of the chamber, close to the bottom? If the sample probe is in its proper position in the RBC chamber (between the edge of the chamber and the center of the chamber, close to the bottom), and therefore system operations appear to be acceptable, call your local technical support provider or distributor.
-
Page 105: Hct
The procedures described on the next pages should be performed whenever system operations are faulty. Identify the source of the malfunction and initiate the appropriate troubleshooting procedure. A «SERVICE» menu is available in the ADVIA 60- program to help the user in the troubleshooting of the hydraulic transfers and mechanical operations.
-
Page 106: Motors
9. MAINTENANCE AND TROUBLESHOOTING Diag. 9.6 9.2.4.3. Motors After having switched on the instrument or whenever a problem occurs, the display may show some error messages concerning a failure in one of the motor initialization. These messages appear for a few seconds on the display and are as follow : — «ERROR : NEEDLE MOTOR»…
-
Page 107: Error Messages
9. MAINTENANCE AND TROUBLESHOOTING 9.3. ERROR MESSAGES Some error messages can be displayed and require the operator intervention. Follow the instruc- tions given for each message and refer to the specified section. If the problem cannot be solved, call your local technical support provider or distributor. A — «ERROR : PRESSURE SYRINGE MOTOR»: This message can be displayed at the instru- ment startup.
-
Page 108
9. MAINTENANCE AND TROUBLESHOOTING K — «ERROR : BAD DATE ! TRY AGAIN» : This message is displayed in the date change function when the new date entered is incompatible with the date format previously entered. Re-enter a correct date or change the date format. L — «ERROR : BAD TIME ! TRY AGAIN»… -
Page 109
9. MAINTENANCE AND TROUBLESHOOTING Q — «HGB REFERENCE FAILED»: this message is displayed when the instrument gives out of range HGB blank values after 3 consecutive HGB reference blank cycles. Check the expiration dates, replace the reagents if necessary or perform a concentrated cleaning. R — «4 HOURS ELAPSED BETWEEN LAST OPERATION»: this message is displayed when the instrument has not been used for at least 4 hours. -
Page 110: Purpose Of The Valves
9. MAINTENANCE AND TROUBLESHOOTING 9.4. PURPOSE OF THE VALVES Valve #1 : Controls the lyse distribution Valve #2 : Cancels the pressure/vacuum in the pressure/vacuum syringe Valve #4 : Controls the cleaner input in the WBC counting head during the rinsing Valve #5 : Controls the drain of the pressure/vacuum syringe Valve #6 : Activates the vacuum needed in the WBC &…
-
Page 111: Menu Overview
9. MAINTENANCE AND TROUBLESHOOTING 9.5. MENU OVERVIEW ADVIA 60-CT Parameters 2 SELECT LEVEL 1 — Low Select OP Parameters 1 2 — Normal 1 — Automatic 1 — RESULTS 3 — High Pat .ID / Run # 2 — Analysis…
-
Page 114
9. MAINTENANCE AND TROUBLESHOOTING Page 9/22 075D0002-01… -
Page 115: Chapter 10. Service And Supplies
Bleach Solution B01-4198-01 02488831 Bayer TESTpoint Hematology Controls B03-4200-54 04575197 Normal B03-4201-54 00133238 High B03-4202-54 00056756 ADVIA 60 SETpoint Calibrator B03-4203-51 09345602 Calibrator Memory Card 075-1100-02 08366266 Control Memory Card 075-1000-02 08564998 Patient Memory Card 075-1200-02 05963395 Bar Code Reader…
-
Page 116: For Service
10.2.1 Bayer Authorized Representative Bayer Diagnostics Europe Limited Chapel Lane, Swords, Co. Dublin, Ireland 10.2.2 Bayer Offices Worldwide Manufactured by: Bayer S.A. Bayer HealthCare LLC Produtos Diagnósticos Subsidiary of Bayer Corporation Rua Domingos Jorge 1100 Diagnostics Division 04779-900 — São Paulo — SP…
-
Page 117
Centro de Atención Telefónica (CAT) Athens 151 25, Greece +55 57 28 33 12 +30 (0) 1-6883648 +55 57 28 33 06 Bayer HealthCare Co. Ltd. Bayer B.V. 20/F Gee Chang Hong Centre Health Care 65 Wong Chuk Hang Road… -
Page 118
10. SERVICE AND SUPPLIES Bayer (South East Asia) Pte Ltd. No. 9 Benoi Sector Singapore 629844 +65-261-3389 Bayer (Pty) Ltd. Healthcare Division 27 Wrench Road Isando 1600, South Africa +27 (0) 11-921-5048 Química Farmacéutica Bayer, S.A. División Diagnósticos Calabria, 268 08029 Barcelona, España +34-93-4956500 Bayer (Schweiz) AG… -
Page 119: Appendix A. System And Reagent Rolls
APPENDIX A. SYSTEM AND REAGENT ROLLS A.1 System and Reagent Symbols The following symbols appear on the Advia 60 Hematology System, and the Advia 60 TIMEPAC reagent package. Warning. Biohazard Warning. Biological Risk WARNING: Indicates the risk of personal injury or loss of life if operating procedures and practices are not strictly observed.
-
Page 120: System And Reagent Symbols
APPENDIX A. SYSTEM AND REAGENT SYMBOLS Up Navigation Down Navigation ON/OFF Ground Fuse Serial Port Printer Port The input electricity is alternating current. Caution. Risk of static discharge. Consult the instructions for use. The analyzer meets the safety requirements of Underwriters Laboratories. The analyzer meets the safety requirements of the Canadians Standards Association.
-
Page 121
APPENDIX A. SYSTEM AND REAGENT SYMBOLS Serial Number Catalog Number In Vitro Diagnostic Device Manufactured Location Authorized Representative Date of Manufacture CE Mark. Product meets the requirements of applicable European Directives. Temperature limitation (store between x°C – y°C) Batch code Use by Store upright Recycle… -
Page 122
APPENDIX A. SYSTEM AND REAGENT SYMBOLS Page A/4 075D0002-01… -
Page 123: Appendix B. Biohazard Protection
APPENDIX B. BIOHAZARD PROTECTION B.1 Protecting Yourself from Biohazards BIOHAZARD All products or objects that come in contact with human or animal body fluids should be handled, before and after cleaning, as if capable of transmitting infectious diseases. Wear facial protection, gloves, and protective clothing.
-
Page 124: References
APPENDIX B. BIOHAZARD PROTECTION • Keep your hands away from your face. • Cover superficial cuts and wounds. • Dispose of contaminated materials according to your laboratory’s biohazard control procedures. • Disinfect your work area with a 15% bleach solution. •…
-
Page 125: Appendix C. Bibliography
APPENDIX C. BIBLIOGRAPHY C.1 Bibliography The following information was used to create the reagent information in this manual. Bessman, JD; (1986) Automated Blood Counts and Differentials, Johns Hopkins University Press Groner W, Simson E; (1995) Practical Guide to Modern Hematology Analyzers: Wiley 1-241 Corash L;…
-
Page 126: Bibliography
APPENDIX C. BIBLIOGRAPHY Page C/2 075D0002-01…
-
Page 127
INDEX Index Autocleaning Frequency 8-9 Printer Select 8-3 Calibration 6-3, 7-2, 7-6 Change Date and Time 8-11 Change Laboratory Limits 8-4 Quality Control 6-3 Change Operator 8-7 Messages 7-13 Checks Visual 2-3 Concentrated Cleaning 9-3 Reagent Pack Connnections 2-5 Installation 2-5 Date Priming 2-7 Change 8-11… -
Page 128: Index
INDEX Index/2 075D0002-01…
Электроника и Медтехника
|
|
|
© 2003 — 2019 Электроника и Медтехника |
Назначение
Описание
Программное обеспечение
Знак утверждения типа
Комплектность
Сведения о методах измерений
Рекомендации к применению
Назначение
Система гематологическая аналитическая Advia 60 (далее по тексту — система) предназначена для измерения следующих параметров крови:
o WBC — Концентрация лейкоцитов
o RBC — Концентрация эритроцитов
0 HGB — Концентрация гемоглобина
Система также может рассчитывать следующие параметры крови: гематокрит; средняя концентрация гемоглобина в эритроците; среднее содержание гемоглобина в эритроците; ширина распределения эритроцитов; содержание лимфоцитов; содержание гранулоцитов; содержание средних лейкоцитов; средний объем тромбоцитов; тромбокрит; ширина распределения тромбоцитов; регистрировать крупные тромбоциты.
Описание
Принцип действия системы основан на двух методах измерения:
1 Кондуктометрический метод для подсчета количества эритроцитов, основанный на изменении сопротивления раствора между электродами при прохождении клетки крови через апертуру фиксированного диаметра. Поток клеток создает последовательность импульсов, которые усиливаются, измеряются и затем математически пересчитываются в числовое значение.
2 Спектрофотометрический метод для определения концентрации гемоглобина, основанный на поглощении измеряемым раствором излучения с длиной волны 550 нм.
Конструкция системы включает в себя следующие блоки и узлы:
— Основной блок, предназначенный для проведения анализа образцов и включающий в себя блок измерительных камер для выполнения измерений.
— Блок подачи образцов, предназначенный для автоматизации ввода образцов в основной блок.
— Блок обработки информации, предназначенный для обработки данных и управления основным блоком.
— Узел гидравлической автоматики, предназначенный для дозирования и разведения образцов цельной крови необходимыми реагентами с целью обеспечения работы измерительных камер основного блока.
Рисунок 1 — Общий вид системы Advia 60
► ADVIA 60
наименование
HEMATOLOC
SYSTEM
| |yp | For b Vitro DiagnostK Um
REF ADVIA 60 ОТ SN
Pour UtfcMtW A Para изо tfagnoitico «1 «иге Para Use am Diagndstice in f*n
esfl С
07/06
6070T81194
серийный номер
100 240 V — 150 V.A 50 60 Hz
■ фирма-изготовител!
©
LISTED
Mrabr) Imiwi «0»
(€
«VIAmlrKlWMrt /ТЧ ,, jire
gl Bay» HMfthCare UC ИЧЯГПИГОЫ^,
WaiiutKlurwJ r f>««* *» в***
■ (I*,» ■ m I_ Щ >и11мШ1мИМ _
АЛАЯ К ^ мшМаЬМ»»*
т,.Лиж «»1вв’ *в,“м I I
RS23Z IOIOI
A
1 |
° |
I § НЕ В |
СКРЫВАТЬ!!! |
* лК |
Программное обеспечение
В системе используется встроенное программное обеспечение, которое устанавливается заводом-изготовителем непосредственно в ПЗУ системы.
Программное обеспечение предназначено для управления системой, контроллером внутренних исполнительных механизмов и измерительных устройств и его настроек, а также для обеспечения функционирования интерфейса, обработки информации, полученной от измерительных устройств в процессе проведения измерений.
Идентификационные данные (признаки) метрологически значимой части программного обеспечения системы указаны в таблице 1.
Таблица 1
Наименование программного обеспечения |
Идентифика ционное наименование программного обеспечения |
Номер версии (идентификаци онный номер) программного обеспечения |
Цифровой идентификатор программного обеспечения (контрольная сумма исполняемого кода) |
Алгоритм вычисления цифрового идентификатора программного обеспечения |
ПО анализа ADVIA 60 |
Advia 60 |
1.0 |
Данные являются собственностью производителя и являются защищенными для доступа дилера и пользователя |
Защита программного обеспечения от непреднамеренных и преднамеренных изменений соответствует уровню «А» по МИ 3286-2010.
Таблица 2
Количество определяемых параметров |
18 |
Диапазон измерений |
|
WBC, х109 1/л |
0,5-80 |
RBC, х1012 1/л |
0,2-7,5 |
HGB, г/дл |
2,5-23 |
Предел относительного среднего квадратичного |
|
отклонения измерения, %, не более |
|
WBC |
3 |
RBC |
2 |
HGB |
2 |
Объем пробы |
|
Режим работы с цельной кровью, мкл |
10 |
Напряжение питания, В |
100/240 |
При частоте, Гц |
50/60 |
Память, измерений |
60 |
Г абаритные размеры, мм |
440х360х330 |
Масса, кг |
14 |
Условия эксплуатации |
|
Температура воздуха, оС |
18-32 |
Относительная влажность, %, не более |
80 |
Знак утверждения типа
Знак утверждения типа наносится на титульный лист руководства по эксплуатации типографским способом.
Комплектность
1 Система
2 Руководство по эксплуатации
3 Комплект расходных материалов и запасных частей Поверка
Поверка прибора осуществляется в соответствии с Методикой поверки МП 59.Д4-12 «Система гематологическая аналитическая Advia 60», утвержденной ГЦИ СИ ФГУП «ВНИИОФИ» 12 июля 2012 г.
Основное средство поверки — ГСО 9624-2010 Состава форменных элементов крови.
Сведения о методах измерений
Руководство пользователя на Систему гематологическую аналитическую Advia 60.
Нормативные и технические документы, устанавливающие требования к Системе гематологической аналитической Advia 60
1 ГОСТ Р 50444-92. Приборы, аппараты и оборудование медицинское. Общие технические условия.
2 Техническая документация фирм «Bayer HealthCare LLC», «Bayer Diagnostics», США, Франция.
Рекомендации к применению
Осуществление деятельности в области здравоохранения
© 2003 Bayer HealthCare LLC
No part of this manual or the products it describes may be reproduced by any means or in any form without
prior consent in writing from Bayer HealthCare LLC.
ADVIA is a trademark of Bayer HealthCare LLC.
Vacutainer is a trademark of Bectin Dickson and Company.
Manufactured in France for Bayer HealthCare LLC.
The information in this manual was correct at the time of printing. However, Bayer HealthCare LLC continues
to improve products and reserves the right to change specifications, equipment, and maintenance procedures
at any time without notice.
If the system is used in a manner differently than specified by Bayer HealthCare LLC, the protection provided
by the equipment may be impaired. See warning and hazard statements.
ADVIA® 60 Hematology System
Operator’s Guide
Closed Tube Model
075D0002-01 Rev. A, 2003-09
© 2003 Bayer HealthCare LLC
No part of this manual or the products it describes may be reproduced by any means or in any form without prior consent in writing from Bayer HealthCare LLC.
ADVIA is a trademark of Bayer HealthCare LLC.
Vacutainer is a trademark of Bectin Dickson and Company.
Manufactured in France for Bayer HealthCare LLC.
The information in this manual was correct at the time of printing. However, Bayer HealthCare LLC continues to improve products and reserves the right to change specifications, equipment, and maintenance procedures at any time without notice.
If the system is used in a manner differently than specified by Bayer HealthCare LLC, the protection provided by the equipment may be impaired. See warning and hazard statements.
User Manual Closed Tube Model
Daily Startup and Shutdown Procedures
Start-up procedure
ON/OFF switch located on the ADVIA
® three (3) minutes.
60 Hematology System rear panel. Wait for approximately
2. Verify that the printer is connected and turned to ON.
3. Wait for the end of the STARTUP CYCLE or press the STARTUP key and validate.
4. The background values must be less than:
WBC
RBC
HGB
PLT
0.3 x 10
3
WBC/mm
3
0.02 x 10
6
RBC/mm
3
0.0 g/dL
10 x 10
3
PLT/mm
3
If the background values are not within these limits, the ADVIA 60 Hematology System will automatically perform another startup cycle. If the STARTUP fails, refer to Section 9 for troubleshooting procedures.
5. Run a blood sample to prime the system: a. Enter the patient identification or run # (according to the identification mode chosen) using the ID/SEQ key.
b. Place the sample tube into the tube holder.
c. Close the door of the tube holder in its sampling position to start the analysis if this starting mode has been setup, or press the START key after closing the door. The tube holder carriage moves up in the piercing position, and the sample aspiration begins. The green/red LED will blink during sampling.
d. The LED stops blinking and the tube holder carriage moves down when the sampling is completed.
6. Perform a QC procedure using ADVIA 60 TESTpoint
™
Hematology Controls (Prod. No. B03-4200-54, B03-
4201-54, and B03-4202-54).
7. Perform a calibration procedure only if necessary.
8. Run patient samples : a. Enter patient identification:
• Enter the identification # (up to 13 characters) or Run # (ranging from 1 to 10000).
b. Place the sample tube into the tube holder.
c. Close the door of the tube holder in its sampling position to start the analysis if this starting mode has been setup, or press the START key after closing the door. The tube holder carriage moves up in the piercing position, and the sample aspiration begins. The green/red LED will blink during sampling.
d. The LED stops blinking and the tube holder carriage moves down when the sampling is completed.
Shutdown procedure
1. Perform a general cleaning/rinse cycle of the ADVIA 60 Hematology System using the STAND BY key on the front panel. This cycle lasts for approximately one (1) minute. The instrument will cycle into the STAND BY mode.
2. Press the ON/OFF switch to OFF.
NOTE: When the instrument is left in STAND BY mode, a STARTUP CYCLE must be performed prior to any analysis cycle
© 2003 Bayer HealthCare LLC. All rights reserved.
Table of Contents
CHAPTER 1. INTRODUCTION ……………………………………………………………………….. 1-1
1.1. INTENDED USE ……………………………………………………………………………………………………………….. 1-1
1.2. PRESENTATION ………………………………………………………………………………………………………………. 1-2
1.3. OPEN TUBE AND CLOSED TUBE MODELS ………………………………………………………………………… 1-2
1.4. NOTES …………………………………………………………………………………………………………………………… 1-2
CHAPTER 2. INSTALLATION …………………………………………………………………………. 2-1
2.1. INSPECTION …………………………………………………………………………………………………………………… 2-1
2.2. UNPACKING ……………………………………………………………………………………………………………………. 2-1
2.3. PACKAGE CONTENTS ……………………………………………………………………………………………………… 2-2
2.4. WORKING CONDITIONS ………………………………………………………………………………………………….. 2-3
2.4.1. Environment ………………………………………………………………………………………………………………. 2-3
2.4.2. Location ……………………………………………………………………………………………………………………. 2-3
2.4.3. Grounding …………………………………………………………………………………………………………………. 2-3
2.4.4. Humidity and temperature conditions …………………………………………………………………………. 2-3
2.5. VISUAL CHECKS …………………………………………………………………………………………………………….. 2-3
2.5.1. Mechanical check ………………………………………………………………………………………………………. 2-3
2.5.2. Connection check ………………………………………………………………………………………………………. 2-4
2.6. REAGENT PACK CONNECTIONS ………………………………………………………………………………………. 2-5
2.7. REAGENT AND WASTE HANDLING PRECAUTIONS ……………………………………………………………. 2-6
2.8. ELECTRICAL CONNECTIONS …………………………………………………………………………………………… 2-6
2.9. GENERAL POINTS …………………………………………………………………………………………………………… 2-6
2.10. PRINTER (optional) ……………………………………………………………………………………………………….. 2-6
2.10.1. Connection ………………………………………………………………………………………………………………. 2-7
2.11. REAGENT PRIMING ……………………………………………………………………………………………………….. 2-7
CHAPTER 3. SPECIFICATIONS ……………………………………………………………………… 3-1
3.1. PERFORMANCE SPECIFICATIONS ……………………………………………………………………………………. 3-1
3.1.1. Expected Values ………………………………………………………………………………………………………… 3-4
3.2. REAGENT SPECIFICATIONS …………………………………………………………………………………………….. 3-6
3.2.1.Diluent : sysDIL …………………………………………………………………………………………………………… 3-6
3.2.2. Lyse : sysLYSE ……………………………………………………………………………………………………………. 3-6
3.2.3. Detergent : sysKLEN …………………………………………………………………………………………………… 3-7
3.2.4. sysCLEAR (bleach solution) ………………………………………………………………………………………… 3-7
3.2.5. TIMEPAC ……………………………………………………………………………………………………………………. 3-7
3.3. LIMITS …………………………………………………………………………………………………………………………… 3-7
3.3.1. Cleaning ……………………………………………………………………………………………………………………. 3-7
3.3.2. Caps of the sampling tubes ………………………………………………………………………………………… 3-8
3.3.3. Blood specimens ……………………………………………………………………………………………………….. 3-8
3.3.4. Known interfering substances …………………………………………………………………………………….. 3-8
CHAPTER 4. TECHNOLOGY………………………………………………………………………….. 4-1
4.1. MEASUREMENT PRINCIPLES …………………………………………………………………………………………… 4-1
4.1.1. RBC / WBC / PLT detection principles …………………………………………………………………………. 4-1
4.1.2. Hemoglobin measurement principle …………………………………………………………………………… 4-3
4.1.3. Hematocrit measurement principle …………………………………………………………………………….. 4-3
4.2. CELL DISTRIBUTION STUDY ……………………………………………………………………………………………. 4-3
4.2.1. White blood cell distribution ………………………………………………………………………………………. 4-3
4.2.1.1. Analysis principle ………………………………………………………………………………………………….. 4-3
4.2.1.2. Diluent and lysing action ……………………………………………………………………………………….. 4-4
4.2.1.3. Volumetric study ……………………………………………………………………………………………………. 4-4
4.2.1.4. Results ………………………………………………………………………………………………………………….. 4-4
4.2.2. Distribution of Red Blood Cells ……………………………………………………………………………………. 4-5
4.2.3. Platelet distribution ……………………………………………………………………………………………………. 4-5
i
4.3. STUDY OF GENERAL FLAGS …………………………………………………………………………………………… 4-6
4.3.1. Platelet flags ……………………………………………………………………………………………………………… 4-7
4.3.2. HGB flag ……………………………………………………………………………………………………………………. 4-7
4.3.3. WBC Flags …………………………………………………………………………………………………………………. 4-7
4.3.4. Comments on the flagging capabilities ……………………………………………………………………….. 4-8
CHAPTER 5. DESCRIPTION ………………………………………………………………………….. 5-1
5.1. INSTRUMENT ………………………………………………………………………………………………………………….. 5-1
5.1.1. Front panel and command keys ………………………………………………………………………………….. 5-1
A — Using the keyboard ………………………………………………………………………………………………………. 5-1
B — Command keys ……………………………………………………………………………………………………………. 5-2
5.1.2. Rear panel / Main fuses ……………………………………………………………………………………………… 5-4
5.1.3. Left side internal view ……………………………………………………………………………………………….. 5-4
5.1.4. Front internal view …………………………………………………………………………………………………….. 5-5
CHAPTER 6. STARTUP AND SAMPLE RUN …………………………………………………… 6-1
6.1. STARTUP CHECKS ………………………………………………………………………………………………………….. 6-1
6.1.1. INSTRUMENT STARTUP ………………………………………………………………………………………………… 6-1
6.1.2. Reagent pack …………………………………………………………………………………………………………….. 6-2
6.2. SAMPLE COLLECTION AND MIXING ………………………………………………………………………………… 6-2
6.2.1. Sample collection ……………………………………………………………………………………………………… 6-2
6.2.2. Mixing ……………………………………………………………………………………………………………………….. 6-2
6.3. DAILY QUALITY CONTROL/CALIBRATION VERIFICATION …………………………………………………… 6-3
6.3.2. US mode identification with barcode reader ……………………………………………………………….. 6-4
6.3.1. US mode identification without barcode reader ………………………………………………………….. 6-4
6.3.3. Standard identification mode ……………………………………………………………………………………… 6-5
6.3.4. Sample tube holder selection …………………………………………………………………………………….. 6-5
6.3.5. Analysis …………………………………………………………………………………………………………………….. 6-6
6.4. RUNNING SAMPLES ……………………………………………………………………………………………………….. 6-7
6.4.1. Sample identification …………………………………………………………………………………………………. 6-7
6.4.2. Automatic cleaning ……………………………………………………………………………………………………. 6-7
6.4.3. End of the day rinsing ………………………………………………………………………………………………… 6-7
6.5. RESULTS ……………………………………………………………………………………………………………………….. 6-8
6.5.1. Displayed results………………………………………………………………………………………………………… 6-8
6.5.2. Result printout ……………………………………………………………………………………………………………. 6-9
CHAPTER 7. CALIBRATION AND QUALITY CONTROL ……………………………………. 7-1
7.1. INTRODUCTION ………………………………………………………………………………………………………………. 7-1
7.2. SMART CARD OPTION …………………………………………………………………………………………………….. 7-2
7.3. AUTOCALIBRATION ………………………………………………………………………………………………………… 7-2
7.3.1. Select operator ………………………………………………………………………………………………………….. 7-2
7.3.2. Change lot number…………………………………………………………………………………………………….. 7-3
7.3.3. Change expiration date ……………………………………………………………………………………………… 7-4
7.3.4. Change target values …………………………………………………………………………………………………. 7-4
7.3.5. Change number of calibrator samples ………………………………………………………………………… 7-5
7.3.6. Run calibration ………………………………………………………………………………………………………….. 7-6
7.4. CALIBRATION COEFFICIENTS ………………………………………………………………………………………….. 7-9
7.4.1. Changing calibration coefficients ……………………………………………………………………………….. 7-9
7.4.2. Print coefficients ………………………………………………………………………………………………………..7-10
7.4.3. Coefficient limits ………………………………………………………………………………………………………..7-10
7.5. QUALITY CONTROL PROGRAM ………………………………………………………………………………………. 7-11
7.5.1. Introduction ………………………………………………………………………………………………………………. 7-11
7.5.2. QC — Automatic …………………………………………………………………………………………………………..7-12
7.5.2.1. Insert QC smart card ………………………………………………………………………………………………7-12
7.5.2.2. Select operator ……………………………………………………………………………………………………..7-13
7.5.2.3. Select commercial control level …………………………………………………………………………….7-13
7.5.2.4. Run commercial control…………………………………………………………………………………………7-14
7.5.2.5. Accepting/rejecting results …………………………………………………………………………………….7-14
7.5.2.6. Exiting QC automatic …………………………………………………………………………………………….7-15
7.5.3. Analysis …………………………………………………………………………………………………………………….7-16
ii
7.5.4. QC print targets ………………………………………………………………………………………………………….7-16
7.5.5. QC statistics ……………………………………………………………………………………………………………….7-17
7.5.5.1. Select statistics ……………………………………………………………………………………………………..7-17
7.5.5.2. Select level …………………………………………………………………………………………………………..7-17
7.5.6. QC Graphs …………………………………………………………………………………………………………………7-18
7.5.6.1. Select graphs ………………………………………………………………………………………………………..7-18
7.5.6.2. Select level …………………………………………………………………………………………………………..7-18
CHAPTER 8. INSTRUMENT CONFIGURATION ……………………………………………….. 8-1
8.1. RESULTS OPTIONS ………………………………………………………………………………………………………… 8-1
8.1.1. Reprint results ……………………………………………………………………………………………………………. 8-2
8.1.2. Printout ……………………………………………………………………………………………………………………… 8-2
8.1.3. Units ………………………………………………………………………………………………………………………….. 8-2
8.1.4. Printer selection ………………………………………………………………………………………………………… 8-3
8.1.5. Temperature printout …………………………………………………………………………………………………. 8-3
8.1.6. Print limits …………………………………………………………………………………………………………………. 8-4
8.1.7. Print differential results ………………………………………………………………………………………………. 8-4
8.2. CHANGE LABORATORY LIMITS ……………………………………………………………………………………….. 8-4
8.2.1. Result low limits ………………………………………………………………………………………………………… 8-4
8.2.2. Result high limits ……………………………………………………………………………………………………….. 8-5
8.2.3. Print limits and flag values …………………………………………………………………………………………. 8-5
8.2.4. Flag limits………………………………………………………………………………………………………………….. 8-6
8.3. SPECIAL FUNCTIONS ……………………………………………………………………………………………………… 8-7
8.3.1. Change operator ………………………………………………………………………………………………………… 8-7
8.3.2. Change password ………………………………………………………………………………………………………. 8-8
8.3.3. Startup cycle ……………………………………………………………………………………………………………… 8-8
8.3.4. Autocleaning frequency ……………………………………………………………………………………………… 8-9
8.3.5. Internal setup printout ………………………………………………………………………………………………… 8-9
8.3.6. Cycle end audible signal ……………………………………………………………………………………………8-10
8.3.7. Identification mode ……………………………………………………………………………………………………8-10
8.3.8. Start mode ……………………………………………………………………………………………………………….. 8-11
8.4. DATE AND TIME ……………………………………………………………………………………………………………… 8-11
8.4.1. Change time ………………………………………………………………………………………………………………8-12
8.4.2. Date format ……………………………………………………………………………………………………………….8-12
8.4.3. Change date ………………………………………………………………………………………………………………8-12
8.5. HOST OPTIONS ………………………………………………………………………………………………………………8-13
8.5.1. Host communication …………………………………………………………………………………………………..8-13
8.5.2. Baud rate …………………………………………………………………………………………………………………..8-13
8.5.3. Transmission ……………………………………………………………………………………………………………..8-13
8.6. BARCODE SETUP ……………………………………………………………………………………………………………8-14
8.7. Patient Memory Card ……………………………………………………………………………………………………..8-14
8.7.1. Introduction ……………………………………………………………………………………………………………….8-14
8.7.2. Memo operation mode ……………………………………………………………………………………………….8-15
8.7.2.1. Memo on/off ………………………………………………………………………………………………………….8-15
8.7.2.2. Transmission mode………………………………………………………………………………………………..8-15
8.7.3. Running the samples ………………………………………………………………………………………………….8-16
8.7.4. Print list …………………………………………………………………………………………………………………….8-17
8.7.5. Trans. one………………………………………………………………………………………………………………….8-17
8.7.6. Trans. all ……………………………………………………………………………………………………………………8-18
8.7.7. Trans. from to …………………………………………………………………………………………………………….8-18
8.7.8. Clear smart card ………………………………………………………………………………………………………..8-18
CHAPTER 9. MAINTENANCE AND TROUBLESHOOTING ……………………………….. 9-1
9.1. MAINTENANCE AND SERVICE ………………………………………………………………………………………….. 9-1
9.1.1. Overview …………………………………………………………………………………………………………………… 9-1
9.1.2. Daily maintenance …………………………………………………………………………………………………….. 9-1
9.1.2.1. Startup and standby cycles …………………………………………………………………………………….. 9-1
9.1.2.2. Automatic cleaning ……………………………………………………………………………………………….. 9-1
9.1.2.3. Instrument general cleaning ………………………………………………………………………………….. 9-1
9.1.3. Service functions ……………………………………………………………………………………………………….. 9-2
iii
9.1.3.1. Backflush ………………………………………………………………………………………………………………. 9-2
9.1.3.2. Drain chambers …………………………………………………………………………………………………….. 9-2
9.1.3.3. Prime reagents ……………………………………………………………………………………………………… 9-2
9.1.3.4. Concentrated cleaning …………………………………………………………………………………………… 9-3
9.1.3.5. Mechanical checks ………………………………………………………………………………………………… 9-4
9.1.3.6. Cycle functions………………………………………………………………………………………………………. 9-6
9.1.3.7. Technician functions ……………………………………………………………………………………………… 9-6
9.1.3.8. Automatic cleaning ……………………………………………………………………………………………….. 9-6
9.1.4 Disposal of System Waste and Supplies………………………………………………………………………. 9-6
9.2. TROUBLESHOOTING ………………………………………………………………………………………………………. 9-7
9.2.1. Overview …………………………………………………………………………………………………………………… 9-7
9.2.2. Identification procedure …………………………………………………………………………………………….. 9-7
9.2.2.1. System operations …………………………………………………………………………………………………. 9-7
9.2.2.2. Reagents ………………………………………………………………………………………………………………. 9-7
9.2.2.3. Precision ………………………………………………………………………………………………………………. 9-7
9.2.2.4. Calibration ……………………………………………………………………………………………………………. 9-8
9.2.3. Troubleshooting parameters ……………………………………………………………………………………….. 9-8
9.2.3.1. WBC and HGB ……………………………………………………………………………………………………….. 9-8
9.2.3.2. WBC ……………………………………………………………………………………………………………………..9-10
9.2.3.3. HGB ………………………………………………………………………………………………………………………9-10
9.2.3.4. RBC, HCT and PLT ………………………………………………………………………………………………… 9-11
9.2.3.5. RBC ………………………………………………………………………………………………………………………9-12
9.2.3.6. HCT ………………………………………………………………………………………………………………………9-13
9.2.3.7. PLT ……………………………………………………………………………………………………………………….9-13
9.2.4. Troubleshooting system operations …………………………………………………………………………….9-13
9.2.4.1. Power …………………………………………………………………………………………………………………..9-13
9.2.4.2. Display ………………………………………………………………………………………………………………….9-13
9.2.4.3. Motors …………………………………………………………………………………………………………………..9-14
9.3. ERROR MESSAGES ………………………………………………………………………………………………………..9-15
9.4. PURPOSE OF THE VALVES ……………………………………………………………………………………………..9-18
9.5. MENU OVERVIEW …………………………………………………………………………………………………………..9-19
9.6. PNEUMATIC DIAGRAM …………………………………………………………………………………………………….9-20
CHAPTER 10. SERVICE AND SUPPLIES ……………………………………………………… 10-1
10.1 Supplies ………………………………………………………………………………………………………………………..10-1
10.2 For Service ……………………………………………………………………………………………………………………10-2
10.2.1 Bayer Authorized Representative ………………………………………………………………………………10-2
10.2.2 Bayer Offices Worldwide …………………………………………………………………………………………..10-2
APPENDIX A. SYSTEM AND REAGENT ROLLS ……………………………………………… A-1
A.1 System and Reagent Symbols …………………………………………………………………………………………. A-1
APPENDIX B. BIOHAZARD PROTECTION …………………………………………………….. B-1
B.1 Protecting Yourself from Biohazards ………………………………………………………………………………… B-1
B.2 References ……………………………………………………………………………………………………………………… B-2
APPENDIX C. BIBLIOGRAPHY ……………………………………………………………………… C-1
C.1 Bibliography …………………………………………………………………………………………………………………… C-1
Index …………………………………………………………………………………………………….. Index-1
iv
CHAPTER 1. INTRODUCTION
1.1. INTENDED USE
The ADVIA
60-
CT
(Closed Tube) system is a fully automated (Microprocessor controlled) hematology analyzer used for the in vitro diagnostic testing of whole blood specimens.
ADVIA 60-
CT16
:
— WBC, LYM%, LYM#, MON%, MON#, GRA%, GRA#,
— RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, and 3 distribution curves: WBC, RBC, PLT.
ADVIA 60-
CT18
:
— WBC, LYM%, LYM#, MON%, MON#, GRA%, GRA#,
— RBC, HGB, HCT, MCV, MCH, MCHC, RDW,
— PLT, MPV, PDW**, PCT**, and 3 distribution curves: WBC, RBC, PLT.
from 10 µL of whole blood (taken from EDTA).
WBC :
RBC :
HGB :
HCT :
White blood cell count
Red blood cell count
Hemoglobin
Hematocrit
MCV :
MCH :
Mean cell volume
Mean corpuscular hemoglobin
MCHC : Mean corpuscular hemoglobin concentration
RDW : Red cell distribution width
PLT :
MPV :
Platelet count
Mean platelet volume
LYM% : Lymphocyte percent
LYM# : Lymphocyte number
MON% : Monocyte percent*
MON# : Monocyte number*
GRA% : Granulocyte percent
GRA# : Granulocyte number
PDW :
PCT :
Platelet distribution width **
Plateletcrit **
* The MON count (both # and %) is a composite count that includes monocytes, eosinophils and basophils.
** PCT and PDW are not available in the UNITED STATES
The rate of determinations is 55 samples per hour in the optimum configuration. The system is totally automated including the cap piercing of the sample tube, with an internal dilution system and a graphic printer (optional) for recording all test results including flags and graphic printouts.
075D0002-01 Page 1/1
1. INTRODUCTION
1.2. PRESENTATION
The instrument, which is small in size, has 9 main parts:
1 — The electrical supply.
2 — The electronic board.
3 — The dilution pneumatics.
4 — A control panel including a keyboard and a LCD screen.
5 — A cap piercing mechanism.
6 — A reagent compartment.
7 — A printer (optional) that prints out the results and the plotting of the distribution curves.
8 — A smart card reader (optional) for quality control, result records and calibration direct entries.
9 — A barcode reader (optional) for a direct entry of the alphanumerical identifications.
All the controls are grouped together on one panel, at the front of the system.
Diag.1.1
1.3. OPEN TUBE AND CLOSED TUBE MODELS
The ADVIA 60 is also available in two mechanical models. The ADVIA
60-
OT
is an Open Tube model whereby the operator needs to remove the stopper from the blood tube before introducing the sample via the sampling probe.
The ADVIA 60-
CT
is a Closed Tube model permitting sampling of the blood specimen without removing the stopper from the blood collection tube.
1.4. NOTES
Bayer HealthCare retains the right to make changes to the instrument and this document. Your local technical support provider or distributor will provide the latest revision of this document.
No part of this document may be copied or reproduced in any form or by any means without prior written consent of Bayer HealthCare.
Page 1/2 075D0002-01
CHAPTER 2. INSTALLATION
2.1. INSPECTION
A thorough inspection is performed prior to the release of an ADVIA 60-
CT
Hematology System.
It is important to verify receipt of all parts. Notify any descrepancies with the carrier. As instructed, the installation procedures must be followed in the order listed below.
2.2. UNPACKING
The instrument is enveloped in a special, protective foam before being placed in a cardboard box. Cut the four angles of the box to unpack the system. Remove the cardboard box containing the ADVIA 60-
CT
installation kit from its location.
Diag.2.1
— The ADVIA 60-
CT
«pack» installation kit : XEA 335 A
075D0002-01 Page 2/1
2. INSTALLATION
2.3. PACKAGE CONTENTS
— The ADVIA 60-
CT
boxes contain the following parts :
— ADVIA 60-
CT
— Printer (optional)
— User’s daily startup and shutdown procedure
— User’s manual
— ADVIA 60-
CT
power cable (European) : or
— ADVIA 60-
CT
power cable (US) :
DAC 011 A
DAC 012 A
— The ADVIA 60-
CT
«pack» installation kit (XEA 335 A) includes :
Table 2.1
DESIGNATION
Sampling needle ADVIA 60-CT
O ring 6 mm x 1.5 mm
O ring 0.74 mm x 1.02 mm
ADVIA 60 common installation kit
PART NUMBER
GBC 052 AS
FAA 036 A
FAA 054 A
XEA 312 A
— The ADVIA 60-
CT
common installation kit (XEA 312 A) includes :
Table 2.2
DESIGNATION
Fuse 1A, 220 V 5 x 20 mm
Fitting 1.6mm
ADVIA 60 cover
ADVIA 60 cover key
Allen key 1.5 mm
Allen key 2.0 mm
Allen key 2.5 mm
Bent wrench 2.5 mm
TORX key
Tygon tube 1.52 mm
Tygon tube 2.29 mm
O ring 30.8 x 3.6 mm
O ring 15 x 1.5 mm
Grease KM 1011
PART NUMBER
DAR 040 A
EAB 032 A
FBH 015 A
FAJ 004 A
MAB 003 A
MAB 001 A
MAB 069 A
MAB 002 A
MAB 090 A
EAE 007 A
EAE 009 A
FAA 017 A
FAA 029 A
XEA 019 A
2
2
1
1
1
1
1
1
1
1
QTY
2
1
1
1
QTY
1
2
2
1
Page 2/2 075D0002-01
2. INSTALLATION
2.4. WORKING CONDITIONS
2.4.1. Environment
ADVIA 60-
CT
should be operated in an indoor location only. Operation at an altitude over 2000 meters (6562 feet) is not recommended. The instrument is designed to be safe for transient voltages according to INSTALLATION CATEGORY II and POLLUTION DEGREE 2. Please ask your local technical support provider or distributor for any information about operating location when it does not comply with the specifications.
2.4.2. Location
ADVIA 60-
CT
should be placed on a clean and level table or work station. Please note that ADVIA 60-
CT
, printer and reagents weigh approximately 30 kilograms (66 lbs). Avoid exposure to sunlight.
Proper ventilation requires that a space of at least 20 cm (8 inches) must be left behind the apparatus.
2.4.3. Grounding
Proper grounding is required. Check that the wall ground (earth) plug is correctly connected to the laboratory grounding electricity installation. If there is no ground then use a ground stake. Current electricity norms must be applied.
2.4.4. Humidity and temperature conditions
ADVIA 60-
CT
can function between 18 to 32°C (65 to 90°F). Maximum relative humidity is 80% for temperatures up to 31°C decreasing linearly to 50% relative humidity at 40°C. If it is kept at a temperature less than 10°C (50°F), the instrument should be allowed to sit for an hour at the correct room temperature before use.
2.5. VISUAL CHECKS
2.5.1. Mechanical check
Using the key from the installation kit, unloosen the locker. Open the pneumatic protection door.
WARNING: Do not open or close the instrument front door when the door of the piercing mechanism is open.
Diag.2.2
Unscrew the 5 cover fixation screws and remove the cover: pull it backward and lift it up to the rear of the instrument.
Diag.2.3
075D0002-01 Page 2/3
2. INSTALLATION
Diag.2.4
Diag.2.5
Push the black plastic carriage locking clip as far as possible to the left and place the sample needle carriage as far forward as possible to the right-hand side. Check that the aspiration needle is not bent and make sure it is in its upper position.
1 — Carriage locking clip
2 — Sample needle carriage
3 — Needle
Check the position of the chambers. Each chamber should be in its proper position with its clips and the electrode block is attached firmly to the RBC chamber.
1 — Clip
2 — RBC chamber
Unscrew the 2 screws of the WBC/HGB chamber protection cover slightly. Remove the cover and check that the chamber is fixed properly in its clips and the electrode block is attached firmly to the chamber.
Re-install the HGB/WBC chamber cover.
1 — Chamber protection cover
Diag.2.6
2.5.2. Connection check
Diag.2.7
Page 2/4
Check that the connectors on the printed circuit board are securely in place.
Re-install the instrument cover.
075D0002-01
2. INSTALLATION
Remove the fuse holder from its location on the rear panel pressing on the holder lock and check the fuse characteristics : they should be 1 Amperes, 220 Volts
Slow-Blow.
Diag.2.8
2.6. REAGENT PACK CONNECTIONS
The ADVIA 60 Hematology System TIMEPAC
®
Reagents includes the entire set of reagents in a
«Pack» and is able to receive the waste liquids. Three soft pockets contain the 3 reagents sysDIL
®
,
sysKLEN
®
and sysLYSE
®
and are closed by means of the valve connectors located at the bottom of the pack. The fourth pocket is empty and is intended for receiving waste liquids. To order additional reagents, contact your local technical support provider or distributor.
Remove the reagent output protections, as well as the waste input protection.
Diag. 2.9
Install the pack directly into the compartment of the instrument. Push the pack down in order to plug correctly the pack on the male connectors.
Diag. 2.10
The free male connector must be plugged on the pack upper valve in order to receive the waste liquids.
CAUTION: To prevent leaks in the reagent pack avoid unplugging reagent pack more than once.
075D0002-01 Page 2/5
2. INSTALLATION
2.7. REAGENT AND WASTE HANDLING PRECAUTIONS
Reagents have to be stored at room temperature (18°C to 25°C). Lyse reagent contains cyanides and has to be handled according to the local or national regulations. Always follow the recommended precautions. Collect all the waste generated during testing to facilitate compliance with
the local environmental regulations.
BIOHAZARD: Wear personal protective equipment. Use universal precautions. Refer to Appendix B for recommended precautions when working with biohazardous materials.
2.8. ELECTRICAL CONNECTIONS
The ADVIA 60-
CT
is connected to the laboratory electrical supply using the power cable included in the installation kit. Connect the power cable to the plug located on the rear left-hand side of the device.
Two 1A fuses are located, under the power plug. The instrument can be operated at any other voltage (from 100 to 240V) or frequency (from 50 to 60Hz) without modification.
1 — ON/OFF SWITCH
2 — FUSE HOLDER
3 — MAIN SUPPLY PLUG
4 — RS 232 OUTPUT
5 — PRINTER OUTPUT
Diag.2.11
If the instrument has to be connected to a laboratory computer, use the plug (4) RS 232.
2.9. GENERAL POINTS
The ADVIA 60-
CT
responds to the UL 3101 norm. Refer to the Declaration of Conformity Statement for details. System performance is guaranteed by Bayer HealthCare under the following conditions only:
— services and repairs are provided by your local technical support provider or distributor
— the electrical supply of the laboratory follows the national or international regulations
— the system is operated under the following instructions.
2.10. PRINTER (optional)
It is mandatory to connect a printer which fulfills the following conditions :
— the printer is recommended by your local technical support provider or distributor,
— the printer is approved by the CE norms (EEC only).
— the printer is certified CSA.
Page 2/6 075D0002-01
2. INSTALLATION
2.10.1. Connection
The printer is connected to ADVIA 60-
CT
with the cable delivered with the instrument. Lock the connector in place by tightening the 2 screws on each end of the connector to the ADVIA 60-
CT
. Attach the other end of the cable to the printer and lock the printer connector in place by the means of the 2 clips located on the connector itself.
Diag.2.12
NOTE:
On the ADVIA 60-
CT
, it is necessary to select the printer format RESERVED 1 of the
«PRINTER» menu (function 4) accessible through the «OPTIONS» menu (function 5 of the main menu) then «RESULTS» (function 1).
2.11. REAGENT PRIMING
When the ADVIA 60-
CT
is first installed, it contains no reagents. All the reagents have to be primed now. Turn ON instrument by pressing the ON/OFF switch located on the rear panel. When the instrument turns on, the display shows:
PLEASE WAIT FOR 3 MIN
ESCAPE : ESC
This time is required at the startup for the instrument initialization and stabilization, specifically for the HGB diode to reach its operationnal temperature. After three minutes, the LED of the front panel turns from red to green and the display shows the following:
STARTUP NOT INITIATED
PRESS A KEY TO CONTINUE…
This message appears when the instrument is setup with the manual startup cycle to prevent any analysis cycle before running a startup cycle. Press any key, the main menu is displayed:
MAIN MENU
HH : MM
1 RESULTS
2 QC
075D0002-01 Page 2/7
2. INSTALLATION
From the MAIN MENU, move the cursor to the function
4 service menu is displayed :
SERVICE and press ENTER. The
SERVICE
HH : MM
> 2 DRAIN CHAMBER
3 PRIME REAGENTS
Move the cursor to
3
PRIME REAGENTS and press the ENTER key. Select the function
1
CHANGE PACK and follow the instructions given by the LCD in order to install the pack.
REAGENT PACK
HH : MM
> 1 CHANGE PACK
2 CBC LEFT < 150>
Once the new PACK is installed a priming cycle will be automatically carried out and the following menu displayed.
PRIME
* * * * * * * * * *
WAIT FOR 2 MIN 3 S
NOTE:
Before analyzing samples, visually inspect reagent lines and pumps for air bubbles.
Repeat priming if air bubbles are still present. Call the local technical support provider or distributor if priming does not eliminate air bubbles.
From the REAGENT PACK menu, the function
2 cycles left to run with the same pack.
CBC LEFT displays the number of analysis
It is also possible to run a priming cycle at any time using the selection
3
REAGENT PACK menu.
PRIME of the
CAUTION: To prevent leaks in the reagent pack avoid unplugging reagent pack more than once.
Page 2/8 075D0002-01
CHAPTER 3. SPECIFICATIONS
3.1. PERFORMANCE SPECIFICATIONS
ADVIA 60-
CT
performs automated blood counts, and requires no manual operations for aspirating blood, dilution, measuring, calculations, print-out, and computer transfer. The parameters given according to the internal setup :
ADVIA 60-
CT16
: — WBC, LYM%, LYM#, MON%, MON#, GRA%, GRA#,
— RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, and 3 distribution curves: WBC, RBC, PLT.
ADVIA 60-
CT18
: — WBC, LYM%, LYM#, MON%, MON#, GRA%, GRA#,
— RBC, HGB, HCT, MCV, MCH, MCHC, RDW,
— PLT, MPV, PDW*, PCT*, and 3 distribution curves: WBC, RBC, PLT.
* : PCT and PDW are not available in the United States.
— Dimensions:
— Height : approximately 440 mm (16.5 inches)
— Width : approximately 360 mm (14.2 inches)
— Depth : approximately 330 mm (12.6 inches)
Diag.3.1
— Weight :
— Power supply :
— LCD screen :
— Power Consumption :
* approximately 14 Kgs (31 lbs).
* 100 Vac to 240 Vac + 10%, 50 Hz to 60 Hz
* 2 lines of 40 characters, backlighted
* Maximum : 150 VA (-30%, + 10%)
* In use : 110 VA (-30%, + 10%)
* Stand-by mode : 35 VA (-30%, +10%)
— Conditions for use :
— Smart Card option :
* ADVIA 60-
CT
can function between 18 to 32°C
(65 to 90°F). Maximum relative humidity 80% for
temperatures up to 31°C decreasing linearly to 50%
relative humidity at 40°C.
* Reader : GCI 400
— ADVIA 60 SETpoint
TM
Calibrator memory card : — GFM 2K
— Patient memory card :
— ADVIA 60 TESTpoint
TM
Control memory card :
— MCOS 24K (capacity : 60 CBCs)
— MCOS 24K (16 parameters)
— MCOS 16K (8 parameters)
075D0002-01 Page 3/1
3. SPECIFICATIONS
— Storage capacity :
Table 3.1
# of QC Aspirations/day
1
2
3
— Hemoglobin :
— Size of apertures :
— Final dilutions :
— Throughput :
— Capacity of internal memory :
— Anticoagulent:
— Volume of whole blood sample :
— Reagent consumption :
QC8P card
80
44
31
Number of QC available
80
60
41
* HGB/WBC chamber, LED 550 nm.
* WBC = 80 µm
* RBC/PLT = 50 µm
* WBC = approximately 1/250
* RBC / PLT = approximately 1/15000
* 55 samples / hour approximately
* Last sample only
* 60 samples with the memory Smart Card
EDTA is recommended
* 10 µL
* Software version V 1.6
QC16P card
Table 3.2
CYCLE
Analysis cycle
Prime all reagents
Prime diluent
Prime lyse
Prime cleaner
Startup cycle
Standby cycle
Auto clean cycle
Concentrated cleaning
HGB blank cycle
Backflush cycle
sysDIL
17.5 mL
40.0 mL
27.0 mL x x
21.0 mL x
16.4 mL
16.4 mL
6.0 mL x
sysLYSE
0.60 mL
11.6 mL x
11.6 mL x
0.6 mL x
0.6 mL
0.6 mL
1.3 mL x
sysKLEN
0.85 mL
5.2 mL x x
6.3 mL
1.4 mL
13.6 mL
15.2 mL
1.5 mL x x
sysCLEAR
x x x x x x x x
6.0 mL x x
Page 3/2 075D0002-01
Table 3.3
3. SPECIFICATIONS
— Reagent pack
— Measuring principles:
* Total capacity : 4.2 liters
* WBC.RBC.PLT = Impedance change
* Hematocrit = Numeric integration
* HGB = Cyanmethemoglobin method (550 nm)
— Barcode reader option: * EAN 8, EAN 13, C 39, C 128, ITF (2/5),
CODABAR, STF, C 93 with or without checksum.
— Reproducibility: (based on 20 consecutive samplings from one fresh, normal, whole blood sample)
PARAMETERS
— WBC :
— RBC :
— HGB :
— HCT :
— MCV :
— PLT :
— LYM :
— MON :
— GRA :
% CV
< 2.5%
< 2%
< 1.5%
< 2%
< 1%
< 5%
< 5%
< 10%
< 5% at 10
3
/mm
3
TEST LEVEL
at 5 x 10
6
/mm
3 at 15 g/dL at 45% at 90 µm 3 at 300 x 10
3
/mm
3 at 40% at 10% at 50%
— Linearity: Linearity was tested using commercially available low range and full range linearity test kits. The kits were analyzed and data was computed according to the manufacturer’s instructions.
Each kit included six levels and one level was used as the reference value. Each level was run four times. The results of this study are as follows :
Table 3.4
PARAMETERS
WBC (x 10
3
cells/mm
3
)
RBC (x 10
6
cells/mm
3
)
HGB (g/dL)
HCT (%)
PLT (x 10 3 cells/mm 3 ) for HGB concentrations > 2 g/dL
PLT (x 10 3 cells/mm 3 ) for HGB concentrations < 2 g/dL
LINEARITY
RANGE
LIMITS
0 to 100 +/- 0.5 or +/- 5 % (whichever is greater)
0 to 8.0
+/- 0.3 or +/- 3 % (whichever is greater)
0 to 26
0 to 80
+/- 0.3 or +/- 3 % (whichever is greater)
+/- 2 or +/- 3% (whichever is greater)
0 to 2200
0 to 4000
+/- 10 or +/- 10 % (whichever is greater)
+/- 10 or +/- 10 % (whichever is greater)
NOTE: Platelet linearity depends on hemoglobin concentration.
075D0002-01 Page 3/3
3. SPECIFICATIONS
— Carryover : Carry-over was tested by analyzing samples with high concentrations of WBC’s,
RBC’s, HGB and PLT’s. Each sample was run in triplicate, followed by three background cycles.
The % carryover is calculated using the following formula :
Background 1 — Background 3
Carryover = ——————————————— X 100
Sample 3 — Background 3
Table 3.5
Level
%Carryover (actual)
%Carryover (claim)
WBC
63.0
0.3
<0.5%
RBC
7.58
0.00
<0.5%
HGB
23.4
0.0
<0.5%
PLT
988
0.0
<0.5%
3.1.1. Expected Values
Normal ranges were established at a study performed in Tarrytown (NY, USA). The results were derived from the central 95% of the values in the distribution of 50 apparently healthy individuals.
Table 3.6
WBC (x 10
3
/mm
3
)
LYM (%)
MON (%)
GRA (%)
RBC (x 10 6 /mm 3 )
HGB (g/dL)
HCT (%)
MCV (µm 3 )
MCH (pg)
MCHC (g/dL)
RDW (%)
PLT (x 10
3
/mm
3
)
MPV (µm
3
)
MALE (N=25)
4.5 — 10.8
20 — 47
3 — 9
46 — 72
4.7 — 6.1
13.8 — 17.0
42 — 50
80 — 94
27 — 32
31 — 34
13 — 16
185 — 402
7.1 — 9.5
FEMALE (N=25)
4.3 — 10.4
19 — 48
3 — 9
48 — 71
4.2 — 5.4
11.3 — 15.5
36 — 46
81 — 99
27 — 32
31 — 34
13 — 15
132 — 440
7.8 — 9.3
PCT and PDW have not been established as indications in the United States for this product.
Their use should be restricted to research or investigational use only.
Expected values vary with sample population and/or geographic location. It is recommended that each laboratory establish its own normal ranges based on the local population.
Page 3/4 075D0002-01
3. SPECIFICATIONS
— Accuracy : Approximately 200 patient specimens were analyzed on the ADVIA 60-
CT
and the
ADVIA
®
120 Hematology Systems at three different locations. The following table summarizes the data:
Table 3.7
PARAMETER
WBC (x 10
3
/mm
3
)
RBC (x 10
6
/mm
3
)
HGB (g/dL)
HCT (%)
MCV (µm
3
)
MCH (pg)
MCHC (g/dL)
RDW (%)
PLT (x 10
3
/mm
3
)
MPV (µm
3
)
LYM (%)
MON (%)
GRA (%)
205
205
205
204
205
205
205
205
205
204
204
N
205
205
SITE 1
R
0.996
0.993
0.996
0.985
0.947
0.974
0.360
0.656
0.990
0.877
0.980
0.809
0.980
180
180
N/A
177
180
180
180
180
180
177
177
N
180
180
SITE 2
R
0.992
0.983
0.986
0.972
0.940
0.945
0.314
0.400
0.972
N/A
0.994
0.940
0.982
208
208
208
200
208
208
208
208
208
200
200
N
208
208
SITE 3
R
0.986
0.992
0.994
0.984
0.902
0.916
0.330
0.607
0.985
0.801
0.983
0.819
0.979
075D0002-01 Page 3/5
3. SPECIFICATIONS
3.2. REAGENT SPECIFICATIONS
In order for the instrument to operate correctly, high-quality reagents must be used. Bayer
HealthCare provides all the necessary reagents. To order additional reagents, contact your local technical support provider or distributor. Do not use beyond the expiration date and store the reagents at a temperature between 18 to 25°C. The reagents are ready to use and require no preparation.
3.2.1.Diluent : sysDIL
(1 x 3.2 L)
For in vitro diagnostic use as an isotonic solution for the determination and differentiation of blood cells.
Composition :
· Stabilized buffer solution
· Sodium hydroxide
· Sodium azide
0.04 %
0.09 %
CAUTION: Sodium azide can react with copper and lead plumbing to form explosive metal azides. On disposal, if disposal into a drain is in compliance with federal, state, and local requirements, flush reagents with a large volume of water to prevent the buildup of azides.
Physio-chemical properties : Boiling point : About 100°C, pH : neutral.
3.2.2. Lyse : sysLYSE
(1 x 200 mL)
For in vitro diagnostic use as an eythrocyte-lysing agent for leukocyte counting, differentiation, and hemoglobin determination on the ADVIA 60 Hematology system.
Composition :
· Potassium cyanide, 0.03%.
Physio-chemical properties : Boiling point : approximately 100°C, pH : basic.
Page 3/6 075D0002-01
3. SPECIFICATIONS
3.2.3. Detergent : sysKLEN
Function : For in vitro diagnostic use in hte cleaning of the Advia 60 Hematology system.
Composition: Organic buffer, < 0.2%.
Physio-chemical properties: Boiling point : around 100°C, pH = 9.7 ± 0.2 at 20°C.
3.2.4. sysCLEAR (bleach solution)
sysCLEAR is used in the concentrated cleaning procedure. Diluted bleach solution can also be used. Follow the recommended precautions for safe use. See the Material Safety Data Sheet for first aid.
REF: 02488831
Part number: B01-4198-01 (4 x 500 mL)
For in vitro diagnostic use as a cleaning and bleaching solution for the ADVIA 60 Hematology system.
Composition:
· Sodium Hypochlorite (4%)
· Sodium Hydroxide (0.26%)
Physio-chemical properties: Boiling point : 100°C, pH : 12.9 ± 0.5 at 20°C.
3.2.5. TIMEPAC
See the reagent specifications above.
REF: 07622536
Part number: B01- 4199-54 TIMEPAC (4 x 145 aspirations)
The TIMEPAC reagent pack contains the following reagents:
— sysDIL
— sysLYSE
— sysKLEN
3.3. LIMITS
As with any hematological analysis, users must be alert to the possible effect on results of unknown interferences from medications or endogenous substances. All patient results must be evaluated by the laboratory and the physician in light of the total clinical status of the patient.
3.3.1. Cleaning
In section 9, specific start-up, shutdown, and maintenance procedures are listed. The maintenance procedures identified are mandatory for the proper use and operation of the ADVIA 60-
CT
. FAILURE
TO EXECUTE ANY OF THESE RECOMMENDED PROCEDURES MAY RESULT IN DECREASED
RELIABILITY OF THE SYSTEM.
075D0002-01 Page 3/7
3. SPECIFICATIONS
3.3.2. Caps of the sampling tubes
Some caps of the sampling tubes are more adapted to «cap piercing» sampling systems. Plastic caps cannot be used. Rubber caps can be of different quality materials. Use the best quality materials in order to avoid any rubber particles entering the sample tube when the cap is pierced.
It is also recommended to use caps specially designed to avoid any blood retention in the upper part of the cap.
Diag.3.3
1 Type 1 — Wrong
2 Type 2 — Good
Avoid using Type 1 caps as they may block the atmosphere connection because of the blood trapped inside the cap. This type of cap also can drop some blood when opening the tube, and introduce pressure when closing the tube. Type 2 caps are recommended as they are specially designed to prevent these problems.
NOTE:
It is recommended not to pierce more than three times through the caps.
3.3.3. Blood specimens
Verification of any abnormal test result (including flagged results or results outside of the normal range) should be performed using reference methods or other standard laboratory procedures for the conclusive verification of results.The sections below list known limitations of automated blood cell counters which use the principle of impedance.
BIOHAZARD: Wear personal protective equipment. Use universal precautions. Refer to Appendix B for recommended precautions when working with biohazardous materials.
3.3.4. Known interfering substances
WBC White Blood Cells (Leukocytes):
WBC results that exceed the linearity limits of the system will require dilution of the blood sample.
Re-assaying the diluted sample will help to obtain the correct assay value.
NOTE: Dilute the sample using autologous plasma or Saline.
NRBC — Immature nucleated red blood cells will be counted in the WBC (White Blood Cell) parameter. If the number of nucleated red blood cells is sufficient to activate an L1 alarm, such interference will be detected. However, the manual differential white blood cell count — performed on the stained blood film — will reveal the presence of NRBC’s.
Page 3/8 075D0002-01
075D0002-01
3. SPECIFICATIONS
Following the manual differential white blood cell count, the WBC assay value must be corrected for the presence of nucleated red blood cells. The formula utilized for correcting the WBC parameter, when nucleated red blood cells are present, is:
CORRECT WBC:
counted WBC X 100
————————————-
100 + (# of NRBC/100 WBC)
Unlysed Red Cells — In particularly rare instances, the erythrocytes in the blood sample may not completely lyse. These non-lysed red blood cells may be detected on the WBC histogram with an L1 alarm or as an elevated baseline on the side (leading edge) of the lymphocytes population.
Non-lysed erythrocytes will cause a falsely elevated WBC count.
Multiple myeloma — The precipitation of proteins in multiple myeloma patients may give elevated
WBC counts.
Hemolysis — Hemolyzed specimens contain red cell stroma which may elevate white cell counts.
Leukemia — A spurious low WBC count may result in this disease state because of the possible increased fragility of the leukocytes leading to some destruction of these cells during counting.
These white cell fragments will also interfere with the white cell partial differential parameters;
LYM% + #, MON% + #, GRAN% + #. A spurious low WBC count may also be seen in patients with lymphocytic leukemias due to the presence of abnormally small lymphocytes which may not be counted by the instrument.
Chemotherapy — Cytotoxic and immunosuppressive drugs may increase the fragility of the leukocytes which may cause low WBC counts.
Cryoglobulins — Increased levels of cryoglobulin that may be associated with myeloma, carcinoma, leukemia, macroglobulinemia, lymphoproliferative disorders, metastic tumors, autoimmune disorders, infections, idiopathic disease, aneurism, pregnancy, thromboembolic phenomena, diabetes, etc can elevate the WBC, RBC or PLT counts and the HGB value. The specimen can be warmed up to 37°C and reanalyzed immediately or a manual WBC, RBC, or PLT count can be performed.
RBC Red Blood Cells (Erythrocytes):
The red blood cell dilution contains all the formed elements in the blood : erythrocytes, leukocytes, and platelets. During the counting of the erythrocytes (red blood cells), platelets are not counted since their size falls below the minimum threshold. Leukocytes (White blood cells), on the other hand, are included in the RBC count. However, since the normal ratio between red blood cells and white blood cells is so extreme, the influence of the WBC on the RBC is negligible.
High WBCs — In rare cases where the WBC is extremely high, the RBC count may be corrected, especially if the RBC count is extremely low.
Agglutinated red blood cells — May cause a falsely decreased RBC count. Blood samples containing the agglutinated red blood cells may be identified by observing abnormal MCH and
MCHC values, as well as by examination of the stained blood film.
Cold agglutinins — IgM immunoglobulins which are elevated in cold agglutinin disease may lower
RBC and PLT counts and increase MCV.
Page 3/9
3. SPECIFICATIONS
HGB (Hemoglobin):
Turbidity of the blood sample — Due to any number of physiologic and/or therapeutic factors may produce falsely elevated HGB results. To obtain accurate hemoglobin results when increased turbidity of the blood sample occurs, determine the cause of the turbidity and follow the appropriate method below :
1. Elevated WBC : An extremely elevated WBC will cause excessive light scatter. In these cases use reference (manual) methods.The diluted sample should be centrifuged, and the supernatant fluid measured on a spectrophotometer.
2. Elevated lipids : Elevated lipids in the blood sample will give the plasma a «milky» appearance.
This condition can occur with hyperlipidemia, hyperproteinemia (as in gammapathies) and hyperbilirubinemia. Accurate hemoglobin determinations can be achieved by using reference
(manual) methods and a plasma blank.
Increased turbidity may also be seen in case where the red blood cells are resistant to lysing.
This condition will cause a falsely elevated Hgb result, but may be detected by observing the abnormal MCH, MCHC values, and the increased baseline on the leading edge of the WBC histogram. Erroneous hemoglobin results will cause the results of the MCH and MCHC to be erroneous as well.
Fetal bloods — The mixing of fetal and maternal bloods may produce a falsely elevated HGB value.
HCT (Hematocrit):
Red blood cells agglutination — May produce erroneous HCT and MCV values. Red blood cells agglutination may be detected by observing the abnormal MCH and MCHC values, as well as by examination of the stained blood film In such cases, manual methods may be required to obtain an accurate HCT value.
MCV (Mean Corpuscular Volume):
Red blood cell agglutination — May produce an erroneous MCV value. Red blood cell agglutination may be detected by observing the abnormal MCH and MCHC values, as well as by examination of the stained blood film. In such cases, manual methods may be required to obtain an accurate
MCV value.
Excessive numbers of large platelets and/or the presence of an excessively high WBC
count — May interfere with the accurate determination of the MCV value. In such cases, careful examination of the stained blood film may reveal the error.
MCH (Mean Corpuscular Hemoglobin):
The MCH is a function of th HGB value and the RBC count. The limitations listed for the HGB and
RBC will have an effect on the MCH and may cause erroneous values.
MCHC (Mean Corpuscular Hemoglobin Concentration):
The MCHC is a function of the HGB and HCT values. The limitations listed for the HGB and HCT will have an effect on the MCHC and may cause erroneous values.
Page 3/10 075D0002-01
075D0002-01
3. SPECIFICATIONS
RDW (Red Blood Cell Distribution Width):
The red blood cell distribution width is a function of the RBC count. The red blood cell dilution contains all of the formed elements in the blood : erythrocytes, leukocytes and platelets. During the counting of the erythrocytes (Red Blood Cells) platelets are not included in RBC count since their size falls below the minimum threshold. However, leukocytes (White Blood Cells) are counted and included in the RBC count. Since the normal ratio between RBC and WBC is so extreme, the influence of the WBC is extremely low and the RBC count may need to be corrected, especially if the RBC count is extremely low.
Agglutinated RBC — May cause a falsely decreased RBC count and erroneous RDWs. Blood samples containing the agglutinated RBC may be detected by observing abnormal MCH and
MCHC values, as well as by examination of the stained blood film.
Nutritional deficiency or blood transfusion — May cause elevated RDW results due to iron, vitamin B12 or folate conditions. High RDWs may also be present from bi-modal RBC distribution in blood transfusion.
PLT (Platelets):
Very small erythrocytes (microcytes) , erythrocytes fragments (schizocytes), and WBC
fragments — May interfere with the proper counting of platelets and cause elevated PLT counts.
Agglutinated erythrocytes — May trap platelets, causing an erroneously low platelet count. The presence of agglutinated erythrocytes may be detected by observation of abnormal MCH and
MCHC values and by careful examination of the stained blood film.
Giant platelets in excessive numbers — May cause an erroneously low platelet count since these large platelets may exceed the upper threshold for the platelet parameter and are not counted.
Chemotherapy — Cytotoxic and immunosuppressive drugs may increase the fragility of these cells which may cause low PLT counts. Reference (manual) methods may be necessary to obtain an accurate platelet count.
Hemolysis — Hemolyzed specimens contain red cell stroma which may elevate platelet counts.
A.C.D. blood — Blood anticoagulated with Acid Citrate Dextrose may contain platelet aggregates which could depress the platelet count.
RBC inclusions — Erythrocyte inclusions, such as Howell-Jolly bodies, Heinz bodies, siderotic and basophilic granules, etc may produce a spuriously increased platelet count.
Platelet agglutination — Clumped platelets due to poor collection techniques or platelet satellitosis caused by EDTA activation of immunoglobulins may cause a decreased platelet count and/or an elevated WBC count. The specimen should be recollected in sodium citrate anticoagulant and reanalyzed for only the platelet count. The final PLT result must be corrected for the sodium citrate dilution effect.
MPV (Mean Platelet Volume):
Giant platelets that exceed the upper threshold of the Platelet parameter — May not be counted as platelets. Consequently, these larger platelets will not be included in the instrument’s calculation of Mean Platelet Volume.
Very small erythrocytes (microcytes), erythrocytic fragments (Schizocytes), and white
blood cell fragments — May interfere with the proper counting and sizing of Platelets.
Page 3/11
3. SPECIFICATIONS
Agglutinated erythrocytes — May trap Platelets, causing an erroneous MPV result. The presence of agglutinated erythrocytes may be detected by observation of the abnormal MCH and MCHC values and by careful examination of the stained blood film.
Chemotherapy — May also affect the sizing of PLTs.
CAUTION: Blood samples collected in EDTA will not maintain a stable Mean Platelet
Volume. Platelets collected in EDTA swell with time and temperature.
LYM# (Lymphocyte count absolute value):
The lymphocyte count is derived from the WBC count. The presence of nucleated red blood cells
(NRBC), certain parasites, and erythrocytes that are resistant to lysis may interfere with an accurate LYM count. Limitations listed for the WBC count pertain to the LYM # count as well.
LYM% (Lymphocyte percentage):
The lymphocyte percent is a function of the WBC count and the number of lymphocytes. The presence of nucleated RBC (NRBC), certain parasites, and erythrocytes that are resistant to lysis may interfere with an accurate LYM% count. Limitations listed for the WBC count pertain to the LYM% as well.
MON# (Mononcyte cell count absolute)*:
The monocyte cell count absolute is derived from the WBC count.The presence of large lymphocytes, atypical lymphocytes, blasts, and excessive number of basophils may interfere with an accurate number of monocytes
MON% (Monocyte percentage)*:
The monocyte percentage is a function of the WBC count and the number of monocytes. The presence of large lymphocytes, atypical lymphocytes, blasts, and excessive number of basophils may interfere with an accurate MON% count.
* : The MON count (both # and %) is a composite count that includes monocytes, eosinophils and basophils.
GRA# (Granulocyte cell count absolute):
The granulocyte cell count is derived from the WBC cell count. The excessive presence of eosinophils, metamyelocytes, myelocytes, promyelocytes, blasts, and plasma cells may interfere with an accurate granulocyte count.
GRA% (Granulocyte percentage):
The granulocyte percentage is a function of the WBC count and the number of the granulocytes.
The excessive presence of eosinophils, metamyelocytes, myelocytes, promyelocytes, blasts, and plasma cells may interfere with an accurate GRA% count.
Page 3/12 075D0002-01
CHAPTER 4. TECHNOLOGY
4.1. MEASUREMENT PRINCIPLES
4.1.1. RBC / WBC / PLT detection principles
The counting principle used for red blood cells (RBC), white blood cells (WBC), and platelets
(PLT) is based on the variation in impedance generated by the passage of cells through a calibrated microaperture.
1 — The sample is diluted in an electrolytic diluent (current conductor). The conductivity of the diluent differs considerably from the non conductivity of the blood cells.
2 — The dilution is pulled through the calibrated microaperture. Two electrodes are placed on each side of the aperture. Electric current passes through the electrodes continuously.
3 — When the cell passes through the aperture, electric resistance (or impedance) between the two electrodes increases proportionately with the cell volume.
Diag. 4.1
1 Volts
2 Pulse
3 Time
4 Electrodes
5 Analyzing Electronic Circuit
6 I = Constant
7 Vacuum = Constant
8 Solution to be Analyzed
075D0002-01 Page 4/1
4. TECHNOLOGY
We can derive from Ohm’s law :
V = Voltage I = Current
V = RI
R = Resistance
Since I is constant, R increases with each cell passage (through the aperture), thus V increases proportionately to the cell volume.
4 — The generated impulses have a very low voltage, which the amplification circuit increases, so that the electronic system can analyze them and eliminate the background noise.
5 — Two measuring chambers and detection circuits separately carry out the analysis of white blood cells, and that of platelets and red blood cells.
1
2
3
Volts
Threshold
Time
Diag. 4.2
6 — The dilutions used for these different measurements are the following:
* WBC = 10 µL of whole blood mixed with 2.50 mL of diluent, the final dilution is at 1/250, then
0.60 mL of lyse are added before the count, resulting in a final dilution of 1/300.
* RBC / PLT = 30 µL of the dilution at 1/250 are mixed with 2.50 mL of the diluent, resulting in a dilution of 1/21,000.
7 — Each type of cell (WBC, RBC, PLT) is analyzed by the microprocessor which also handles the cell distribution (histograms).
8 — To count the platelets, the ADVIA 60-
CT
uses high performing electronics, which avoids the use of complex hydraulic systems for the elimination of faulty impulses generated at the rear side of the aperture.
When red blood cells reenter the analysis zone, this creates impulses with a height comparable to platelet impulses, but with a different shape. The instrument uses a very sophisticated impulse sorting system. This system rejects any impulse which does not have the typical platelet shape.
This sorting system maintains a very reliable aperture and a traditional hydraulic system.
9 — ADVIA 60-
CT
provides distribution curves, by analysis on 256 counting channels for the WBC,
RBC and Platelets.
Page 4/2 075D0002-01
4. TECHNOLOGY
4.1.2. Hemoglobin measurement principle
1 — During the STARTUP cycle an HGB blank test sequence including two blank measures is run.
If the difference between these two measures is too large a third measurement is performed.
NOTE:
HGB reference blank measurement will occur if the operator:
— has required a CALIBRATION operation.
— has required a QC operation.
— has left the instrument more than 10 minutes after the STARTUP.
— has left the instrument more than 60 minutes after ANALYSIS.
— has not carried out the STARTUP cycle after switching on the instrument.
2 — On every cycle an Hgb blank is carried out on diluent and compared to the previous HGB blank analysis.
3 — 0.60 mL of lyse agent is added to the 2.5 mL of 1/250 dilution.
4 — The hemoglobin freed by the lyse of the red blood cells combines with potassium cyanide to form chromogenous cyanmethemoglobin compound.
5 — The compound is then measured by spectrophotometry, through the WBC cuvette at 550 nm.
6 — The result is given with the units set up in Section 8 «Instrument Configuration».
4.1.3. Hematocrit measurement principle
1 — The height of the impulse generated by the passage of a cell through the microaperture is
directly proportional to the volume of the analyzed cell.
2 — The hematocrit is measured as a function of the numeric integration of the MCV.
3 — The result is given with the units set up in Section 8 «Instrument Configuration».
4.2. CELL DISTRIBUTION STUDY
ADVIA 60-
CT
carries out volumetric distributions (histograms) for WBC and RBC on 256 analysis channels, and 128 channels for PLT with the following measuring range:
* WBC = approximately 30 to 460 fL.
* RBC = approximately 25 to 300 fL.
* Plt = approximately 2 to 33 fL.
4.2.1. White blood cell distribution
The WBC volumetric distribution study reveals the following three leukocyte subpopulations :
Lymphocytes, Monocytes, Granulocytes.
4.2.1.1. Analysis principle
The principle is based on the volumetric study of leukocytes after use of a (patented) diluent and a lysing reagent.
NOTE:
The lysing action varies according to the temperature of the dilution (ambient temperature). In order to correct these fluctuations within the operating temperature range (18°C to 32°C, 65°F to 90°F), a temperature sensor is placed on the diluent circuit and the position of the separation thresholds varies according to the temperature.
075D0002-01 Page 4/3
4. TECHNOLOGY
4.2.1.2. Diluent and lysing action
The diluent preserves and prepares the cell membrane for the differential reaction. The lyse has a differential mode of action on cytoplasmic membranes. The lymphocyte cytoplasmic membranes allow the release of water soluble cytoplasm and shrink the membrane around the nucleous, the monocytes undergo an intermediate reaction, while the granulocytes have a limited reaction due to a molecule and their cytoplasmic structure which protects them from the shrinking action of the lyse.
4.2.1.3. Volumetric study
After the differential lysing action, ADVIA 60-
CT
analyzes the heights of each impulse, coming from the aperture and ranked in the 256 counting channels. A curve is obtained with number of cells on the y-axis and cell volume on the x-axis. The cells will be broken down accordingly:
* the lymphocytes between 30 — 100 fL .
* the monocytes between 100 — 150 fL .
* the granulocytes between 150 fL and a maximum (unlimited volumetrically).
Pathologic cells will, of course, place themselves in different zones in the distribution curve.
Mobile and fixed flags will alert the lab operator of the presence of such pathologic elements.
NOTE: The WBC distribution curve disappears when the WBC result is rejected.
Diag. 4.3
1 Leukocyte Flags
2 LYM
3 MON
4 GRA
5 Number of Cells
4.2.1.4. Results
The lymphocytes, monocytes, and granulocytes are expressed in percents (%) and absolute numbers (#) : LYM%, LYM#, MON%, MON#, GRA%, GRA#.
Page 4/4 075D0002-01
4. TECHNOLOGY
4.2.2. Distribution of Red Blood Cells
The study of the distribution of RBC detects erythrocyte anomalies linked to anisocytosis.
A Red Cell Distribution Width (RDW) will enable you to follow the evolution of the width of the curve in relation to the cell number and average volume.
RDW =
K SD
——
MCV
With K
SD
MCV
=
=
=
System constant
Determined standard deviation according to statistical studies on cell distribution.
Mean Corpuscular Volume of erythrocytes
4.2.3. Platelet distribution
The platelet distribution study counts platelets, detects platelet anomalies and alerts the lab operator in the event of a non-platelet cell population (schistocytes, microcytes, etc…).
Diag. 4.4
1
2
3
1 — Platelets are counted, between a low threshold placed at 2 fL and a variable high threshold. The variable high threshold varies according to the microcyte population, which is present in the platelet analysis zone.
Small Cells (SCL)
Schistocytes (SCH)
Microcytes (MIC)
Diag. 4.5
2 — Measuring the MPV : The MPV (Mean Platelet Volume) is directly derived from the analysis of the platelet distribution curve. The MPV is expressed in µm
3
or fL.
3 — Calculating PCT :
PLT (10
3
/mm
3
) x MPV (µm
3
)
PCT% = ——————————————
10000
4 — Measuring the PDW (Platelet Distribution Width):
This count is derived from the platelet curve.
PDW = Width of the curve between 15% of the number of platelets starting from 2 fL (S1) and 15% of the number of platelets beginning with the variable top threshold (S2).
1 PDW
075D0002-01 Page 4/5
4. TECHNOLOGY
4.3. STUDY OF GENERAL FLAGS
With the print-out of results the following general flags may occur :
*
following either WBC, RBC, HCT or PLT indicates that the system, analyzed the sample three times but that all three counts differed and were outside the systems precision limits. The result should be verified by repeating the sample or checked according to one of the laboratory reference methods.
$
following a parameter : indicates that the instrument performed 3 counts, 2 of the 3 counts are within :
— 7% for WBC
— 5% for RBC
— 15% for PLT
If the maximum of the first two raw counts is lower than :
— 3000 for WBC, the limit becomes 9%
— 16000 for RBC, the limit becomes 8%
— 400 for PLT, the limit becomes 20%
Result for the concerned parameter can be accepted.
— D
If results exceed the linearity limits shown below, a <D> will be printed after the result. If the results greatly exceed the limits, a result will not be shown on the LCD or the printout. The flag
DIL will be shown on the LCD, and —D or —0 will be printed and transmitted to the LIS.
Repeat the sample using a 1:2 dilution and repeat each time the flag reappears.
Limits to the linearity ranges : — WBC >
— RBC >
— PLT >
— PLT >
— HGB >
— HCT >
100 x 10
8.0 x 10
2200 x 10
4000 x 10
26 g/dL
80%
6
3
/mm
/mm
3
3
3
3
/mm
/mm
3
3
(HGB > 2 g/dL)
(HGB < 2 g/dL)
NOTE: Dilute the sample using autologous plasma or Saline.
H : located next to a result of a parameter shows that the value is above the upper limit set up by the operator.
L : located next to a result of a parameter shows that the value is below the lower limit set up by the operator.
NOTE:
Temperature : Operating reagent temperature should be within the recommended limits (18°C to 32°C, 65°F to 90°F). The operating temperature can be printed out on the result form. When these limits are exceeded, the values obtained cannot be guaranteed with certainty and the result transmitted on the RS shows an error code.
Page 4/6 075D0002-01
4. TECHNOLOGY
4.3.1. Platelet flags
MIC
following the Platelet result indicates the excessive presence of microcytes in the Platelet measurement zone. Verify the result using a Platelet Rich Plasma (PRP) or a manual count. This flag can be adjusted by the user.
SCH
following the Platelet result indicates the presence of schistocytes or Platelet aggregates in the Platelet measurement zone. Review/scan slide before reporting result. This flag can be adjusted by the user.
SCL
following the Platelet result indicates the presence of small cells in the 2 and 3 fL zone. A second sample cycle should be carried out and the results verified. If this flag should persist, perform an automatic cleaning cycle and resample. If the flag persists verify using a Platelet Rich
Plasma (PRP) of the sample and make a manual slide count for the Platelets. This flag can be adjusted by the user.
4.3.2. HGB flag
!
: located next to the HGB result shows that the HGB blank carried out during the analysis and the previous analysis blank differed and were outside the system’s precision limits.
Nevertheless the instrument provides a result according to the previous HGB blank. This result can be reported.
If this suspicious flag occurs more than three consecutive times run the checkup procedure.
4.3.3. WBC Flags
The ADVIA 60-
CT
has a system of WBC differential flags alerting the operator to the possible presence of pathological cells, abnormal volume distribution histograms or abnormal elevated populations such as in the case of the eosinophils and basophils.
1 — Flag
L1
: This flag shows an abnormal number of cells in comparison with lymphocytes (in the
30 — 60 fL zone). The pathologic elements which may be found in this area include platelet agregates, nucleated red blood cells or atypical lymphocytes. The flag corresponds to the number of cells counted in the first five channels out of the total number of lymphocytes. The lab operator can adjust the limit which triggers this flag.
075D0002-01 Page 4/7
4. TECHNOLOGY
2 — Flag
M2
: Located in the 130 to 160 fL zone, this flag informs the lab operator of the presence of lymphoblasts, myelocytes, abnormal lymphocytes, or basophilia (too many basophils). This flag can be adjusted by the operator. It corresponds to the number of cells counted by the detection zone out of the total number of granulocytes. The limit which triggers this flag is adjusted by the lab operator.
3 — Flag
G1
: Situated in the 160 to 220 fL zone, this detects the presence of eosinophils, myelocytes, and sometimes, neutrophilia. This flag can be adjusted by the operator. It corresponds to the number of cells counted in the detection area over the total number of granulocytes.
The limit which triggers this flag is adjusted by the lab operator.
4 — Flag
G2
: This flag makes it possible to follow an abnormal granulocyte peak displacement, shows anomalies in the membrane of the granulocytes, and also possible lyse or hydraulic anomalies. This flag is also triggered, if the blood is too old (after 6 — 8 hours). The flag is also triggered if the granulocyte volume is less than 250 fL .
5 — Flag
G3
: This flag is situated in the zone which is greater than 400 fL. It detects the presence of metamyelocytes. The flag is set off when the percentage in number, compared to the number of granulocytes is higher than the level set. The lab operator can adjust the limit which triggers this flag.
4.3.4. Comments on the flagging capabilities
All anomalies and/or abnormal distributions signaled by the ADVIA 60-
CT
should be verified manually for the presence of pathological elements. As a result of the differential resistance of cytoplasmic membranes in the different cell types, pathological elements can be found in a number of different zones. This also applies to the presence of normal or non-pathological cells that have been subject to chemotherapy or some other form of treatment in alarm zones. This will result in a false alarm.
Page 4/8 075D0002-01
CHAPTER 5. DESCRIPTION
5.1. INSTRUMENT
3
4
1
2
Cover
Door to pneumatic part
Piercing carriage door
Front panel
Diag.5.1
1 — ADVIA 60-
CT
Cover
The instrument cover is fixed by the means of 5 screws. Before any attempt to remove the cover, open the pneumatic access door on the front of the system.
WARNING: Do not open or close the instrument front door when the piercing carriage door is open. Do not open the piercing carriage door when the aspiration needle is in its lower position.
2 — Pneumatic access door. This door gives an access to the pneumatic parts. It also allows the operator to check the hydraulic cycle operation. It is recommended to keep the door locked during the measuring cycles as it is equipped with an electrical interference shield.
3 — Piercing carriage door. This door gives access to the sample tube holder which moves up during the sampling cycle in order to pierce the tube cap. Then the aspiration needle moves down into the sample tube inside the piercing needle and aspirates the 10 µL whole blood sample. Its internal and external rinsing is patented and is fully automatic.
4 — Control panel. This panel allows the operator to communicate with the instrument :
— To access the different cycles
— To identify the patients
— To setup the instrument, etc…
5.1.1. Front panel and command keys
A — Using the keyboard
All functionalities of the instrument have been dispatched in «menus». Each menu has a certain number of functions. These menus are all accessible through the MAIN MENU displayed at the instrument startup. Navigation in the menus is done using the «>» cursor command keys UP and
DOWN . Entry in a function is done by moving the cursor in front of the function and pressing the
ENTER key.
NOTE:
When menus and functions are known, it is possible to enter directly the menu or function number to access it (and press ENTER when a function is requested).
075D0002-01 Page 5/1
5. DESCRIPTION
Diag.5.2
3
4
1
2
5
STARTUP key
STAND BY key
Identification Key (ID/SEQ)
Start analysis cycle key (START)
Numerical keyboard
B — Command keys
6
7
8
9
10
DELETE key
ENTER key
ESCAPE key
Display scroll keys
Cycle light indicators
1 — STARTUP key: When this key is pressed, a startup cycle including a cleaning and rinsing procedure is carried out. Detergent left in the chambers is rinsed with diluent and the system is ready for the analysis cycles. This cycle has a duration of approximately 130 seconds (this cycle can be run 2 or 3 times if blank values are not within the acceptable limits).
2 — STAND BY key: this key is used for shutdown at the end of the working day. When this key is pressed, it is mandatory to carry out a startup cycle before any analysis cycle. This cycle has a duration of approximately 65 seconds.
3 — Identification Key (ID/SEQ): This key is used to enter the patient identification (13 characters maximum, letters or numbers), and the run number.
NOTE:
The patient identification can be entered also using the barcode reader when the instrument is set up in the «US» identification mode.
4 — Start Analysis Cycle key (START): This key starts the analysis when the manual mode for starting the analysis has been set up.
Page 5/2 075D0002-01
5. DESCRIPTION
5 — Numerical keyboard : keys from 0 to 9 allow the operator to enter the following figures :
— Date
— Calibration values
— Laboratory limits
— Patient number for analysis
— Leukocytes differential flag values.
These 2 keys pressed simultaneously will allow an automatic adjustment of the LCD.
6 — DELETE Key: When it is pressed, this key deletes the information entered on the LCD display.
7 — ENTER Key: This key is pressed to validate the informations entered on the LCD display.
8 — ESCAPE Key (ESC): When it is pressed, this key allows the operator to exit a function and to come back to the previous menu. This key can be used to stop an hydraulic cyle (see NOTE below). It also opens the tube holder compartment door.
9 — UP/DOWN Keys: These keys allow the user to scroll up or down in the instrument menus to access the different functions and to choose alphabetic letters in the patient identification.
10 — When the START key is activated, the indicator light flashes during the sampling time. When the indicator stops flashing, the operator is allowed to remove the tube from the sampling position. This cycle has a duration of 65 seconds approximately.
NOTE:
When the ESC key is pressed during a hydraulic cycle, it is necessary to run a STARTUP cycle to rinse the instrument before any analysis.
C — Screen details
SERVICE
HH : MM
> 2 DRAIN CHAMBER
3 PRIME REAGENTS
The position of the cursor is given by the «>» as shown above. The and on the right side of the screen indicate that more menus are available up and down.
075D0002-01 Page 5/3
5. DESCRIPTION
5.1.2. Rear panel / Main fuses
3
4
1
2
5
6
Rear panel
RS 232 computer connection
Printer output
ON/OFF switch
F1,F2 Main fuses
Main power socket
Diag.5.3
Diag.5.4
5.1.3. Left side internal view
Diag.5.5
Page 5/4
3
4
1
2
Vacuum/pressure syringe
Liquid electrovalve block 1
Dilution block
Liquid electrovalve block 2
075D0002-01
5.1.4. Front internal view
5. DESCRIPTION
7
8
5
6
3
4
1
2
9
10
11
LCD display
Keyboard
Cycle light indicators
Sampling needle carriage
Start cycle key
RBC chamber
WBC chamber
Spectrophotometer
Piercing mechanism
Tube holder switches
Piercing carriage door switch
Diag.5.6
075D0002-01 Page 5/5
5. DESCRIPTION
Page 5/6 075D0002-01
CHAPTER 6. STARTUP AND SAMPLE RUN
6.1. STARTUP CHECKS
6.1.1. INSTRUMENT STARTUP
Turn ON instrument by pressing the ON/OFF switch located on the rear panel. The display shows the following :
PLEASE WAIT FOR 3 MIN
ESCAPE : ESC
Then the front panel LED turns from red to green to indicate that the initialization phase is completed.
Wait for the end of the automatic startup cycle or press the STARTUP key, the instrument will perform a startup cycle and perform a blank cycle for a background count (an analysis cycle on reagent without any blood sample). If the background count does not exceed the following :
— WBC :0.3x 10
3
/mm
3
(0.3 on the display)
— RBC : 0.02 x 10
6
/mm
3
(0.02 on the display)
— PLT : 10 x 10
3
/mm
3
(10 on the display) the Startup is completed and the message «STARTUP PASSED» is printed out with results (US mode).
If the results are above these limits, the instrument will automatically perform another background cycle. If the problem persists after 3 cycles, a message «STARTUP FAILED, CHECK REAGENTS» is displayed, run the checkup procedure.
NOTE:
During the startup cycle an HGB blank test is carried out. If the HGB blank is not within acceptable values after 3 cycles, a flag «STARTUP FAILED, CHECK REAGENTS» and «HGB REFE-
RENCE FAILED» is printed out with startup results.Run the checkup procedure.
If the startup fails the message «STARTUP FAILED» is printed out with all the results until a new startup is carried out.
During the startup, if the HGB blank test is unacceptable, the background count is not carried out and a new attempt is automatically performed (3 maximum).
— When the instrument has not been used for 4 hours, a startup cycle must be run before running sample analyses.
NOTE:
If analyses are performed when the startup cycle has not been carried out, the message
«STARTUP NOT INITIATED» Is displayed and printed out.
075D0002-01 Page 6/1
6. STARTUP AND SAMPLE RUN
6.1.2. Reagent pack
A low reagent level is indicated by the number of CBCs left (from 5 to 0) when the operator tries to run an analysis cycle :
WARNING : PACK LOW LEVEL : 5
ESC TO EXIT START TO CONFIRM
If this occurs operator has the choice either to run the cycle pressing the START key or to replace the Reagent pack with a new one.
6.2. SAMPLE COLLECTION AND MIXING
6.2.1. Sample collection
Sample collection should be done on venous blood by the means of vacuum or atmospheric sample collection tubes. It is possible to collect capillary blood into a microtainer with a minimum volume of 100 µL (for example in a pediatric lab.) and analyze this on the ADVIA 60-
CT
. EDTA is the recommended anticoagulent for analysis on the ADVIA 60-
CT
.
Samples should be processed as soon as possible, and within 6-8 hours of collection.
BIOHAZARD: Wear personal protective equipment. Use universal precautions. Refer to Appendix B for recommended precautions when working with biohazardous materials.
CAUTION: The sample collection tube should be filled with the exact quantity of blood indicated on the tube itself. Any incorrectly measured blood sample collection will show a possible variation in the results.
6.2.2. Mixing
Bayer HealthCare has a list of recommended sample collection tubes at your disposal.
NOTE:
It is critical to assure that all the blood samples are thoroughly and gently mixed (with a gentle up and down and rolling motion), prior to each sample aspiration.
Page 6/2 075D0002-01
6. STARTUP AND SAMPLE RUN
6.3. DAILY QUALITY CONTROL/CALIBRATION VERIFICATION
Refer to your laboratory quality assurance program to ensure quality throughout the entire testing process.
It is recommended that the system be controlled using Bayer TESTpoint Hematology Controls
Low, Normal, and High (Prod. Nos. B03-4200-54, B03-4201-54, and B03-4202-54, respectively).
Refer to the package insert supplied with the control for instructions for use.
Controls are intended to be integrated into a clinical laboratory’s own quality control program and procedures.
The following is a suggested protocol. The actual frequency of control in a laboratory will be based upon many factors such as workflow, system experience, government regulation, etc.
These controls should be assayed:
1 At the beginning of each shift.
2 Whenever a new reagent container (same or different lot) is used.
3 Following the performance of any system maintenance or cleaning.
A satisfactory level of performance is achieved when the control values obtained for each level are within the acceptable range for the system as published in the package insert provided with the
Bayer TESTpoint Hematology Controls.
Refer to Section 7 «Calibration and Quality Control» for instructions for use.
Before analyzing samples it is recommended that the operator analyze three levels of control material (low, normal and high) to verify that the system is within acceptable limits.
Two Identification modes are available :
1 — US mode : that requires the patient (or control) identification on each analysis. This mode allows the use of the barcode reader.
2 — Standard mode : increments a run number on each analysis.
BIOHAZARD: Wear personal protective equipment. Use universal precautions. Refer to Appendix B for recommended precautions when working with biohazardous materials.
075D0002-01 Page 6/3
6. STARTUP AND SAMPLE RUN
6.3.1. US mode identification without barcode reader
Press the ID/SEQ key of the front panel to enter the control ID. The identification menu is displayed :
PAT. ID. ? :
CURRENT:
EXIT : ESC
SAVE : ENTER
The identification can be entered using 13 characters, letters or numbers. Letters can be entered using the UP and DOWN keys of the front panel. Press ENTER for each letter to step to the next one. Press the ENTER key when the identification is entered (or ESC key to save the displayed one) : a message «PLEASE CLOSE HOLDER DOOR» is displayed.
NOTE: The identification displayed in «actual» stays in memory until a cycle is run.
6.3.2. US mode identification with barcode reader
When the blood control tube is equipped with a barcode label, press the ID/SEQ key of the front panel to enter the control ID. The identification menu is displayed :
PAT. ID. ? :
CURRENT:
EXIT : ESC
SAVE : ENTER
Read the control identification barcode label using the barcode reader. When the reading is completed, a beep occurs and the control lot identification is displayed on the screen. Press the ENTER key when the identification is entered (or ESC key to save the displayed one) : a message «PLEASE CLOSE
HOLDER DOOR» is displayed. Install the sample tube in the tube holder and close the sample door to start the analysis if this mode has been selected or close the sample door and press the START key.
Diag. 6.1
Page 6/4 075D0002-01
6. STARTUP AND SAMPLE RUN
6.3.3. Standard identification mode
Press the ID/SEQ key to enter the control Run #. The following menu is displayed:
RUN #? :
NEXT :
EXIT : ESC
SAVE : ENTER
Enter the run number (from 1 to 9999) using the numeric keys of the front panel. Press the ENTER key to record the number (or the ESC key to save the displayed one) : A message «PLEASE
CLOSE HOLDER DOOR» is displayed.
6.3.4. Sample tube holder selection
The sample tube holder has 4 positions according to the sample tube characteristics. The required position is selected when it is facing the inside of the sampling compartment. Turn the sample tube holder in either side to place the required position until the
‘click’ is heard.
The tube holder is associated with 3 switches which are able to detect the sampling position, the presence of the tube holder and a wrong position of the tube holder. The 4 positions can be used for the following sample tubes (list is not exhaustive):
Position 1 : Vacutainers
Position 2 : Mini Vacutainers
Position 3 : Control and calibration vial position
Position 4 : Micro sample collection devices.
Diag. 6.2
075D0002-01 Page 6/5
6. STARTUP AND SAMPLE RUN
6.3.5. Analysis
Install the sample tube in the tube holder and close the sample door to start the analysis if this mode has been selected or close the sample door and press the START key.
Diag. 6.3
NOTE:
If the instrument has not been used for 1 hour, when the first analysis is requested, the instrument will start an HGB reference cycle, the message «PLEASE WAIT» is displayed.
At the end of this cycle the first analysis will be automatically performed when the sample door mode is selected or press the START key again to start the analysis.
BIOHAZARD: Wear personal protective equipment. Use universal precautions. Refer to Appendix B for recommended precautions when working with biohazardous materials.
NOTE:
The current Identification stays in memory until the analysis cycle has begun and can be displayed on the screen by pressing the ID/SEQ key.
The analysis cycle lasts 65 seconds. At the end of the result printout, the LED turns to green and the instrument is ready for the next analysis. If any of the control results are outside the acceptable ranges perform the following : a — Rerun the control b — Clean the system and rerun the control c — Open a new vial d — Recalibrate the system
Page 6/6 075D0002-01
6. STARTUP AND SAMPLE RUN
6.4. RUNNING SAMPLES
6.4.1. Sample identification
After completing the start up procedure, run a non pathological blood from the previous day, followed by the quality control analyses and then the sample analyses.
1 — Enter either the patient identification (US mode) or the patient run # (Standard mode).
2 — Install the sample tube in the tube holder and close the sample door to start the analysis if this mode has been selected or close the sample door and press the START key.
6.4.2. Automatic cleaning
When the instrument has run 50 samples from the date changing, an automatic cleaning procedure is carried out. This cycle lasts 10 minutes approximately while the autocleaning menu is displayed:
AUTO CLEANING
***************************
PLEASE WAIT 2MNS 13S
NOTE:
The automatic cleaning frequency can be adjusted by the user from 1 to 50 as described in the instrument setup.
6.4.3. End of the day rinsing
It is necessary to run a standby/shutdown cycle at the end of the day. Press the STAND BY key, the instrument performs a complete cleaning with detergent, and puts the system into the stand by mode. The instrument can be switched off if the working day is completed or left in this standby mode overnight or until the next analysis.
NOTE:
When the instrument is left in standby mode, it is necessary to carry out a STARTUP cycle before any sample analysis.
075D0002-01 Page 6/7
6. STARTUP AND SAMPLE RUN
6.5. RESULTS
When the analysis cycle is completed, results are displayed and printed out according to the setup of the instrument:
6.5.1. Displayed results
The first group of parameters is displayed:
WBC RBC HGB HCT MCV MCH MCHC PLT
7.5 5.22 15.6 46.5 89 29.9 33.5 233
The second group of parameters can be accessed when moving the cursor to the top:
MPV RDW %LYM %MO %GRA #LYM #MO #GRA
8.8 13.0 39.3 7.3 53.4 2.21 0.41 3.00
* Identification
1. US mode
The patient identification can be reviewed when moving the cursor to the bottom:
MM/DD/YY
HH:MM
PAT. ID :
2. Standard mode
The patient run number can be reviewed when moving the cursor to the bottom:
MM/DD/YY
HH:MM
RUN#
* Flags:
The PLT and LMG flags can be reviewed moving the cursor to the bottom
PLT FLAGS :
LMG FLAGS :
NOTE:
The last sample run can be displayed again at any time when moving the cursor from the main menu to the function
1
RESULTS and pressing ENTER.
Page 6/8 075D0002-01
6.5.2. Result printout
1. US mode
6. STARTUP AND SAMPLE RUN
Diag. 6.4
On the result printout can be found:
1 — The sample identification that was entered by the operator.
2 — The sequence number.
3 — The STARTUP status.
4 — The PLT flags.
5 — The CBC results.
6 — The WBC flags.
7 — The Diff results.
8 — The histogram representations.
9 — The date sample was run.
10 — The time sample was run.
The sequence number is updated to 1 everyday and increases by 1 on each cycle. The sequence number cannot be modified by the operator.
075D0002-01 Page 6/9
6. STARTUP AND SAMPLE RUN
2. Standard mode
Diag. 6.5
On the result printout can be found:
1 — The sample run # that was entered by the operator.
2 — The sequence number.
3 — The PLT flags.
4 — The CBC results.
5 — The WBC flags.
6 — The Diff results.
7 — The histogram representations.
8 — The date sample was run.
9 — The time sample was run.
The sequence number is updated to 1 everyday and increases by 1 on each cycle. The sequence number cannot be modified by the operator.
NOTE:
The STARTUP status is printed out only when it has failed (message : «STARTUP
FAILED»).
Page 6/10 075D0002-01
CHAPTER 7. CALIBRATION AND QUALITY CONTROL
7.1. INTRODUCTION
Calibration should be performed upon installation of the ADVIA 60-
CT
Hematology system.
Subsequent recalibration is required if there is a significant shift in control values after replacement of instrument parts, a change in reagent lot number, or whenever indicated by quality control data.
Calibration can be achieved in two different ways.
1 — Bayer SETpoint Calibrator (B03-4203-51) is used to calibrate the ADVIA 60-
CT
Hematology
System.
Refer to the package insert supplied with the calibrator for instructions for use and assay values specific to the ADVIA 60 Hematology system used.
2 — Calibration coefficients are known and can be entered directly.
NOTE:
Calibration must be performed on a clean and reproducible instrument, blank values must be within the acceptable limits.
BIOHAZARD: Wear personal protective equipment. Use universal precautions. Refer to Appendix B for recommended precautions when working with biohazardous materials.
From the MAIN MENU, move the cursor to the function
3
The CALIBRATION menu is displayed :
CALIBRATION and press ENTER.
CALIBRATION
HH:MM
1 AUTO CALIBRATION
2 COEFFICIENTS
075D0002-01 Page 7/1
7. CALIBRATION AND QUALITY CONTROL
7.2. SMART CARD OPTION
When the instrument is equipped with a smart card reader proceed as follows.
Install the card in the smart card reader.
Make sure that the cursor is positioned on function
1
AUTOCALIBRATION and press ENTER.
Diag. 7.1
As the lot number, the expiration date and the target values are recorded into the memory card, only 2 operations remain :
— The operator selection,
— The selection of the number of the calibrator samples.
Once these two steps are done, run calibration.
NOTE: The card can be removed as soon as the lot number has been validated.
7.3. AUTOCALIBRATION
From the calibration menu, move the cursor to function
1
AUTOCALIBRATION and press
ENTER. The CALIBRATION menu will unfold step by step through the operator selection, the calibrator lot, identification, the target values and the number of samples to be analyzed before starting the calibration.
7.3.1. Select operator
Move the cursor to one of the 4 required operator identification and press ENTER. A star (*) is displayed next to the chosen identification and the menu turns to the calibrator identification.
SELECT OP.
HH:MM
1 OP_1
> *2 OP_2
Page 7/2 075D0002-01
7. CALIBRATION AND QUALITY CONTROL
7.3.2. Change lot number
CHANGE LOT # ?
CURRENT : SETPOINT
NO : ESC
YES : ENTER
The current lot number is displayed. If the lot number of the calibration blood has to be changed, press ENTER. The following menu is displayed:
LOT # :
CURRENT : SETPOINT
EXIT : ESC
SAVE : ENTER
Without barcode reader: enter the new lot number (10 characters maximum). Numbers can be entered directly using the numerical keyboard, letters can be entered using the DOWN and UP keys. For each letter, press the ENTER key to validate it and enter the next one, validate a second time to record the lot number. The menu turns to the expiration date. If the lot number has not changed, press the ESC key, the menu turns directly to the expiration date.
With barcode reader: When the calibration blood tube is equipped with a barcode label, read the calibration blood identification barcode label using the barcode reader. When the reading is completed, a beep occurs and the control lot identification is displayed on the screen. Press ENTER to validate.
The menu turns to the expiration date.
Diag. 7.2
075D0002-01 Page 7/3
7. CALIBRATION AND QUALITY CONTROL
7.3.3. Change expiration date
The current expiration date is displayed.
CHANGE EXP. DATE ? (DD.MM.YY)
CURRENT :
NO : ESC
YES : ENTER
If it is correct, press the escape key, the menu turns to the target values. If the expiration date needs to be changed, press ENTER, the following menu is displayed :
EXP. DATE : (DD.MM.YY)
CURRENT :
EXIT : ESC
SAVE : ENTER
Change the expiration date according to the format requested on the display and press ENTER.
The menu turns to the target value modifications.
7.3.4. Change target values
The following menu with the current WBC target value is displayed :
CHANGE TARGET WBC ?
CURRENT :
NO : ESC
YES : ENTER
If the WBC target value is correct, press ESC, the menu turns to RBC target value. If the value has to be changed, press ENTER, the following menu is displayed:
TARGET WBC :
CURRENT :
EXIT : ESC
SAVE : ENTER
Enter the new WBC target value if required and press ENTER or press ESC directly. The menu turns to RBC target value. Repeat the same procedure for RBC, HGB, HCT, PLT, and MPV. When this last value has been modified (or ESC pressed), the number of calibrator samples to be analyzed is displayed.
Page 7/4 075D0002-01
7. CALIBRATION AND QUALITY CONTROL
7.3.5. Change number of calibrator samples
CHANGE # SAMPLE ?
CURRENT :
NO : ESC
YES : ENTER
ADVIA 60-
CT
calibration may be performed using 3 to 11 sample aspirations. The autocalculation module performs statistics on these results in order to obtain the best calibration coefficients.
NOTE:
In order to obtain the best calibration possible, it is recommended to run the calibration blood a minimum of 5 times.
NOTE:
The first result is not taken into account in the statistical calculations used for the calibration. The first sample is used as a «primer».
Press ENTER to change the required number of calibrator samples (or press ESC). The following menu is displayed :
# SAMPLE
CURRENT :
EXIT : ESC
SAVE : ENTER
Enter the new number of calibrator samples and press ENTER if required or press ESC. The run calibration menu is displayed.
075D0002-01 Page 7/5
7. CALIBRATION AND QUALITY CONTROL
7.3.6. Run calibration
Prepare the calibrator according to the specific instructions (temperature, mixing, etc…).
RUN CAL. ?
NO : ESC
YES : ENTER
Press the ENTER key. The start calibration menu is displayed :
START CALIBRATION
ESC TO EXIT CALIBRATION
# 1/X
ENTER TO ASPIRATE
Instrument is requesting the first run of the calibrator. Install the calibrator tube in the correct holder position. Close the sample door to start the analysis if this mode has been selected or close the sample door and press the cycle button located on the front panel. When the analysis cycle ends, the first result menu is displayed:
WBC RBC HGB HCT PLT MPV PRESS ENTER
X.X X.XX XX.X XX XXX X.X
TO CONTINUE
Check that the results are within the limits given in the calibrator instructions and press ENTER.
The first validation menu is displayed:
VALID. CALIBRATION
ESC TO DISCARD
# 1/X
ENTER TO VALID
If the results are not within the acceptable limits, use the ESC key to reject the results and to restart the first run.
NOTE: Results having flags ($ , * or ! for HGB) are automatically rejected.
If the results are correct, press the ENTER key. The menu of the second calibration run is displayed:
START CALIBRATION
ESC TO EXIT CALIBRATION
# 2/X
ENTER TO ASPIRATE
Run the second calibrator sample and follow the same procedure until the number of samples set up is obtained. When the last result has been validated, the instrument calculates the statistical calibration factors for each parameter : Mean, target, coefficient of variation, percentage difference between the target value and the mean (% CHG), new calibration coefficient as well as a recall of the previous calibration coefficient and a pass/fail status appear on the result printout.
Page 7/6 075D0002-01
7. CALIBRATION AND QUALITY CONTROL
Table 7.1
A — Calibration passed :
If the statistical figures are within the acceptable limits:
— Coefficient of variation is within the limits given in Table 7.1 and,
— The percentage difference between the target and the mean value is less than 20.
The calibration passed and the results are printed out. The calculated coefficients become the new calibration coefficients and the calibration is completed.
PARAMETERS
CV
WBC
2.5
RBC
2.0
HGB
1.5
HCT
2.0
PLT
5.0
MPV
3.0
NOTE:
The first run of the calibrator is not taken into account in the statistical calculations (a
«P» is printed out next to the run number).
Diag. 7.3
CALIBRATION
DATE : 04/02/98
OPERATOR : JMG
LOT Nr : SETpoint
4
5
6
2
3
RUN
1 P
MEAN
TARGET
CV
% CHG
OLD CAL
CURRENT
STATUS
WBC
6.3
6.3
6.4
6.5
6.6
6.5
6.46
6.5
1.76
0.61
0.98
0.97
OK
RBC
5.30
5.35
5.40
5.31
5.28
5.30
5.32
5.34
0.89
0.37
0.90
0.88
OK
TIME : 09:21
The following message is displayed:
CALIBRATION ENDED WITH NEW COEFF.
PRESS A KEY TO CONTINUE …
Press any key to return to the MAIN MENU.
HGB
16.0
16.1
16.2
16.1
16.1
16.0
16.1
16.1
0.43
0
1.13
1.13
OK
HCT
47.2
47.1
47
46.8
47.5
47.1
47.1
47
0.54
0.21
0.98
1.03
OK
PLT
230
224
230
240
228
235
231.4
221
2.69
4.70
0.91
0.95
OK
075D0002-01 Page 7/7
7. CALIBRATION AND QUALITY CONTROL
B — Calibration failed:
If the statistical figures are not within the acceptable limits, the following menu is displayed:
CALIBRATION FAILED !!!
PRESS A KEY TO CONTINUE …
— Coefficient of variation is not within the limits given in Table 7.1 or,
— The percentage difference between the target and the mean value is greater than 20.
The calibration failed and the results are printed out. The calculated coefficients are rejected, the instrument keeps in memory the previous calibration coefficients and returns to the calibration menu.
NOTE: If the calibration has failed on one or more parameters no parameter is calibrated.
NOTE:
If the calibration fails, the user can restart the calibration, or enter directly the calibration coefficients through the coefficient function of the calibration menu, using the password.
Diag. 7.4
CALIBRATION
DATE : 04/02/98
CALIBRATION FAILED
OPERATOR : JMG
LOT Nr : SETpoint
4
5
6
2
3
RUN
1 P
MEAN
TARGET
CV
% CHG
6.78
6.5
11.7
4.30
REJ. COEFF 1.16
CURRENT 0.97
STATUS
WBC
6.3
6.3
6.4
6.5
8.2
6.5
5.32
5.34
0.89
0.37
0.90
0.88
RBC
5.30
5.35
5.40
5.31
5.28
5.30
FAILED OK
TIME : 09:21
16.1
16.1
0.43
0
1.13
1.13
HGB
16.0
16.1
16.2
16.1
16.1
16.0
OK
HCT
47.2
47.1
47
46.8
47.5
47.1
47.1
47
0.54
0.21
0.98
1.03
OK
If the printer is not used, when the calibration fails, the following menu is displayed:
SAVED COEFF. WBC 0.97
RBC 0.88
HGB 0.95
REJECT. COEFF.
1.16
0.90
0.90
Rejected and saved coefficients are displayed using the DOWN and UP keys in order to check and to record the faulty parameter(s). Press the ESC key to return to the main calibration menu.
PLT
230
224
230
240
228
235
231.4
221
2.69
4.70
0.90
0.95
OK
Page 7/8 075D0002-01
7. CALIBRATION AND QUALITY CONTROL
7.4. CALIBRATION COEFFICIENTS
Calibration can be achieved directly by changing the calibration coefficients when they are known.
Move the cursor to function
2
COEFFICIENTS and press ENTER. The following menu is displayed:
COEFFICIENTS
HH:MM
1 CALIB. COEFF.
2 PRINT COEFF.
7.4.1. Changing calibration coefficients
Move the cursor to function
1
CALIB.COEFF. and press ENTER. The following menu is displayed
PASSWORD ? :
HH:MM
A specific password is requested to enter the function. Enter the password «123» or the userdefined password and press ENTER. The following menu is displayed:
CALIB. COEFF.
HH:MM
1 WBC < 0.97 >
2 RBC < 0.98 >
Enter the new coefficients for WBC and RBC, then move the cursor down to the HGB, HCT, PLT,
MPV, RDW and PDW positions and enter the required new coefficients.
CALIB. COEFF.
HH:MM
3 HGB < 1.13 >
4 HCT < 1.03 >
CALIB. COEFF.
HH:MM
5 PLT < 0.95 >
6 MPV < 0.92 >
CALIB. COEFF.
HH:MM
7 RDW < 1.00 >
8 PDW < 1.00 >
When the required coefficients have been changed, press ENTER to record the setup. RDW and
PDW can be calibrated by means of calibration coefficients. These coefficients are incremented to 1.00 by default. The RDW and PDW are calculated according to the below formulas :
* RDW result = RDW coeff x RDW measured
* PDW result = PDW coeff x PDW measured
— Press ENTER to modify one of both coefficients. Type in a new value and press ENTER to validate.
NOTE: PDW is not available in the United States.
075D0002-01 Page 7/9
7. CALIBRATION AND QUALITY CONTROL
7.4.2. Print coefficients
From the COEFFICIENTS menu, it is possible to print the coefficient values. Move the cursor to function PRINT COEFF. and press ENTER. The printout of the coefficient values starts automatically.
Diag. 7.5
DATE : 01/11/1997
COEFFICIENTS
TIME : HH : MM
CURRENT
RDW COEFF : 1.00
PDW COEFF : 1.00
WBC RBC HGB HCT PLT MPV
: 0.97 0.88 1.13 1.03 0.95 0.92
NOTE: PDW is not available in the United States.
7.4.3. Coefficient limits
Check that the calibration coefficients corresponds to the following ranges :
LIMITS
Minimum
Mean
Maximum
WBC RBC HGB HCT PLT MPV RDW PDW
0.89 0.73 0.83 0.87 0.99 0.75 0.75 0.75
1.09 0.89 1.11 1.08 1.20 0.94 1.00 1.00
1.29 1.05 1.39 1.29 1.41 1.13 1.25 1.25
Diag. 7.6
If not, call your local technical support provider or distributor.
Page 7/10 075D0002-01
7. CALIBRATION AND QUALITY CONTROL
7.5. QUALITY CONTROL PROGRAM
7.5.1. Introduction
The ADVIA 60-
CT
quality control program contains five different functions :
1 — AUTOMATIC — The function of this submenu is to allow the operator to analyze the commercial control bloods and store the results on the memory smart card.
2 — ANALYSIS — The function of this submenu is to allow the operator to analyze a commercial control with fixed WBC thresholds whatever is the ambient temperature.
3 — PRINT TARGETS — The function of this submenu is to allow the operator to print the targets values of the commercial control bloods.
4 — STATISTICS — The function of this submenu is to allow the operator to print the cumulative statistics of the commercial control files.
5 — GRAPHS — The function of this submenu is to allow the operator to print the Levey-Jennings graphs of the commercial control files.
From the MAIN MENU, move the cursor to the function
2 is displayed:
QC and press ENTER. The QC menu
QC
HH:MM
1 AUTOMATIC
2 ANALYSIS
075D0002-01 Page 7/11
7. CALIBRATION AND QUALITY CONTROL
7.5.2. QC — Automatic
From the QC menu, move the cursor to function
1
AUTOMATIC and press ENTER. The
AUTOMATIC menu will unfold step by step through the smart card insertion, the operator selection, the commercial control identification and lot, etc …
7.5.2.1. Insert QC smart card
First, the analyzer checks if a CARD READER is present, if a card reader is not present, the QC
PROGRAM will be aborted and the following message will appear:
ERROR : NO SMART CARD READER
PRESS A KEY TO CONTINUE …
After a key is pressed, the analyzer comes back to the QC menu because it is impossible to run automatic QC without a smart card reader.
Then, the analyzer checks if a QC smart card is present. If the smart card has not been inserted, the following message appears:
ERROR : NO SMART CARD…
INSERT NEW CARD
NO : ESC
YES : ENTER
Insert the QC smart card into the slot in the upper left corner of the analyzer and press ENTER to
VALIDATE.
If you have no QC smart card, press ESC to come back to the QC menu because it is impossible to run automatic QC without a QC smart card.
If the smart card is present the previous message does not appear and automatically the lot # and the expiration date will be read and are processed and the following message appears:
LOT# TESTPOINT
EXP. DATE 06.03.98
NEW QC NO : ESC
YES :ENTER
The current lot number is displayed. Verify the lot number of the commercial control. The message
«NEW QC» is to explain that this smart card is used for the first time. With every BLOOD CONTROL
LOT a smart card is available and to avoid confusion, the number of runs stored on the QC smart card or the message NEW QC is displayed. If the analyzer index and the smart card index (index
= the number of QC stored on the smart card) are different, a warning message is also displayed.
Page 7/12 075D0002-01
7. CALIBRATION AND QUALITY CONTROL
NOTE:
List of the different messages :
NEW QC : The smart card has not been used, it is the beginning of a new QC.
XX QC RUN : xx QC are stored on this smart card and the index in the smart card and in the analyzer are identical.
QC DIFF.: The QC index in the analyzer and the index on the QC smart card are different, certainly it comes from the confusion between 2 QC smart cards.
If the user presses the ENTER key, the change is validated and index analyzer equal index QC smart card.
If the user presses the ESC key, the analyzer requests a new smart card, it reads the new smart card information and displays the information.
SMART CARD FULL : From 31 to 80 QC sample runs (varies according to type of smart card) have been stored, it is impossible to continue running the QC with this smart card, you must change your smart card.
Press the ENTER key to validate the smart card, or if it is not the proper smart card, press the
ESC key.
If the user presses the ESC key, the analyzer requests a new smart card, reads the new smart card information and displays the information.
When the smart card is validated, the program initializes the first day values to zero (as a protection against wrong results). This operation is automatic and a waiting message is displayed during this time (about 2 seconds for 8 parameters and 4 seconds for 16 parameters).
If the smart card is removed during the QC operation, the instrument goes back to the main menu.
7.5.2.2. Select operator
Move the cursor to one of the 4 required operator identification and press ENTER. A star (*) is displayed next to the chosen identification and the menu turns to the commercial control identification.
SELECT OP
HH:MM
> * 1 OP.1
2 OP.2
NOTE:
If the operator name is changed during the use of the QC card, it is memorized on the card. The QC card allows up to 5 operator name changes.
7.5.2.3. Select commercial control level
SELECT LEVEL
HH:MM
> * 1 LOW BLOOD
2 HIGH BLOOD
Using the upper or lower arrows select the commercial control level that will be analyzed. Press
ENTER .
A message «LOADING LEVEL PLEASE WAIT» appears during a half of a second, the information on the smart card is read during this time. After reading the smart card, the following message appears:
TESTPOINT
ESC TO EXIT
LOW START QC
PRESS START TO ASPIRATE
The current lot number is displayed. Verify the lot number of the commercial control blood.
075D0002-01 Page 7/13
7. CALIBRATION AND QUALITY CONTROL
7.5.2.4. Run commercial control
Prepare the commercial control according to the specific instructions (temperature, mixing, etc … ).
Install the calibrator tube in the correct holder position. Close the sample door to start the analysis if this mode has been selected or close the sample door and press the cycle button located on the front panel. When the analysis cycle ends, the first QC results menu is displayed :
WBC
XX.X H
RBC
X.XX
HGB
XX.X L XX.X
HCT
The complete listing of results can be rewieved when moving the cursor up or down. Press the
ESC key to exit this menu.
Diag. 7.7
7.5.2.5. Accepting/rejecting results
The results are compared to the control assay ranges stored on the memory card. If any result is out of range, an H (High) or L (Low) will be shown on the display and the printout.
If a third count must be processed, a dollar ($) or a star (*) will be shown on the display and the run is rejected.
If the HGB blank is not within acceptable limits a (!) is displayed and the run is rejected.
You must rerun the control. The following message appears when the ENTER key is pressed:
RUN REJECTED
PRESS A KEY TO CONTINUE…
If the results are correct (no star, dollar or ! will appear), the following menu appears:
VALID LOW NO : ESC
YES : ENTER
Page 7/14 075D0002-01
7. CALIBRATION AND QUALITY CONTROL
075D0002-01
1 — ACCEPT: If the results are accepted they will be stored on the QC smart card and the display will return to the SELECT LEVEL.
SELECT LEVEL
HH:MM
> * 1 LOW BLOOD
2 HIGH BLOOD
It is possible to exit the QC or run another control level.
The control level stored on the smart card disappears in the listing of the control levels remaining
(it is impossible to store 2 identical control levels) to avoid confusion.
After the second run (in order to save time), the analyzer loads automatically the last control blood level without going to the SELECT LEVEL (no choice is requested).
2 — REJECT AND RERUN: If any results are not within the acceptable limits, it is possible to reject the results and to repeat the control analysis.
7.5.2.6. Exiting QC automatic
EXIT BEFORE END OF QC: If at any time the operator wants to exit this QC AUTOMATIC MENU, the ESC key can be pressed and the following warning message appears:
EXIT QC ?
NO : ESC
YES : ENTER
After 1, 2 or 3 correct results are stored, it is possible to exit QC and save it. The analyzer checks the number of correct results before exiting the QC PROGRAM.
Then the following message appears:
VALID QC ?
NO : ESC
YES : ENTER
Then the following message appears:
QC STORED
PRESS A KEY TO CONTINUE…
1 — VALID QC: If the QC is validated, the index of the QC smart card is increased as well as the internal index of the analyzer.
2 — INVALID QC: If the QC is not validated, the following message appears:
QC NOT VALID
PRESS A KEY TO CONTINUE …
All the previous data stored this day are erased to avoid confusion in the processing of the printing results. This operation is automatic and a waiting message is displayed during this time (about 2 seconds for 8 parameters and 4 seconds for 16 parameters ).
Page 7/15
7. CALIBRATION AND QUALITY CONTROL
7.5.3. Analysis
This function
2
ANALYSIS allows the operator to run a control as a normal analysis cycle but with specific LMG thresholds for QC blood (independant from the temperature).
From the QC main menu, move the cursor to the function
2
and press ENTER.
LOT # :
CURRENT : MX223N
EXIT : ESC
SAVE : ENTER
Enter a lot identification when required and press ENTER. An HGB blank reference measurement is done prior to the analysis. When this cycle is completed, the following message is displayed:
ANALYSIS
PRESS THE SAMPLING BAR…
Install one of the 3 levels of control blood tubes in the correct holder position. Close the sample door to start the analysis if this mode has been selected or close the sample door and press the cycle button located on the front panel. The analysis is carried out. Results are displayed and printed out.
Diag. 7.8
7.5.4. QC print targets
The commercial control blood target values can be printed at any time. Normally these values are printed on the assay sheets of your control.
From the QC MENU, move the cursor to the function
3
The QC targets are printed out.
PRINT TARGETS and press ENTER.
QC
HH:MM
2 ANALYSIS
3 PRINT TARGETS
Page 7/16 075D0002-01
7. CALIBRATION AND QUALITY CONTROL
7.5.5. QC statistics
The commercial control files can be printed for permanent records at any time. It is recommended to print the files at the end of each month.
Each file printout includes the following information: File Name (blood level), Lot number of control, expiration date of control, date and time of print request, date and time of the run, operator and the data points for each stored quality control run, the reference means and upper and lower limits, the actual mean results of the quality controls runs, the 2 standard deviation value, and the percent coefficient of variation.
7.5.5.1. Select statistics
From the QC MENU, move the cursor to the function
4
STATISTICS and press ENTER.
QC
HH:MM
4 STATISTICS
5 GRAPHS
The QC card is read.
7.5.5.2. Select level
SELECT LEVEL
HH:MM
> * 1 ALL
2 LOW BLOOD
SELECT LEVEL
HH:MM
3 NORMAL BLOOD
4 HIGH BLOOD
Using the UP or DOWN arrows select the commercial control level to print or select ALL to print all three levels automatically. Press ENTER and the following message appears:
PRINTING QC RESULTS
PLEASE WAIT…
All statistics will be printed.
075D0002-01 Page 7/17
7. CALIBRATION AND QUALITY CONTROL
QC
LOW
LOT # : XXX
EXP DATE : 12/31/97
TIME : 12:10PM
DATE : 12/14/97
Diag. 7.9
No DATE TIME
2 12/01/97 10:12
3 ———- ——
———- ——
———- ——
ETC ….
REFERENCE :
MEAN
LOW
HIGH
ACTUAL :
:
:
:
OP
JOHN
———
———
WBC
7.4
6.8
7.9
WBC RBC
7.8 H 4.22
——-
——-
RBC
——
——
——
——
——
HGB
——
——
——
HGB
HCT
——
——
——
HCT
11.6 L 33.0
——- —— —— ——
——
——
——
——
——
MCV
—-
—-
—-
MCV
81
—-
—-
—-
MCH
——
——
——
MN
2SD
CV
ETC ……
:
:
:
WBC
7.8
0.3
2.0
RBC HGB
—— ——
—— ——
—— ——
HCT
——
——
——
MCV
—-
—-
—-
MCH
——
——
——
MCH
28.7
——
——
——
MCHC PLT
———
——
——
—-
—-
MCHC PLT
——
——
——
—-
—-
—-
MCHC PLT
35.6
——
257
—-
——
——
—-
—-
7.5.6. QC Graphs
The ADVIA 60-
CT
plots Levey-Jennings charts for each parameter of the quality control files stored. The Levey-Jennings chart will plot one point for every control data point stored. Beneath each chart, the Reference Mean, two SD value and the actual mean, two SD value and % CV are provided.
7.5.6.1. Select graphs
From the QC MENU, move the cursor to function
5
GRAPHS and press the ENTER key.
4 STATISTICS
> 5 GRAPHS
QC
HH:MM
The QC smart card is read.
7.5.6.2. Select level
SELECT LEVEL
HH:MM
> * 1 ALL
2 LOW BLOOD
SELECT LEVEL
HH:MM
3 NORMAL BLOOD
4 HIGH BLOOD
Page 7/18 075D0002-01
7. CALIBRATION AND QUALITY CONTROL
Using the UP or DOWN arrows select the commercial control level to print one set of graphs or all graphs automatically. Press ENTER and the following message appears:
PRINTING QC RESULTS
PLEASE WAIT…
All graphs will be printed as shown in the example below.
NOTE: QC graphs will be printed out even when values are equal to zero.
Diag. 7.10
075D0002-01
1 — High target value
2 — Mean
3 — Low target value
Page 7/19
7. CALIBRATION AND QUALITY CONTROL
Page 7/20 075D0002-01
CHAPTER 8. INSTRUMENT CONFIGURATION
The ADVIA 60-
CT
has several operator options :
— Specific laboratory limits.
— Date and time formats.
— Result format.
— RS 232 options.
— Special functions.
These options can be configurated according to the operator needs through the SETUP function of the MAIN MENU. From the MAIN MENU, move the cursor to the function
5
and press
ENTER. The SETUP menu is displayed:
SETUP
HH:MM
> 1 RESULTS
2 CHG LAB LIMITS
8.1. RESULTS OPTIONS
The RESULTS menu allows the operator to access the following options :
— To reprint the results of the last sample in memory.
— To select or not the histogram printouts.
— To select the unit type.
— To select the printer type.
— To select the temperature printout.
— To select or not the limit printouts.
— To select or not the Differential result printout.
From the setup menu, move the cursor to the function
1
RESULTS and press ENTER, the
RESULTS menu is displayed:
RESULTS
HH:MM
> 1 REPRINT LAST RESULT
2 PRINTOUT < WITH >
075D0002-01 Page 8/1
8. INSTRUMENT CONFIGURATION
8.1.1. Reprint results
To reprint the results of the last sample in memory, move the cursor to function
1
of the
RESULTS menu and press ENTER. The results of the last sample is automatically reprinted with the date and time, the associated identification, sample run and sequence numbers, the possible flags and the histograms if their printout is selected.
8.1.2. Printout
The distribution curves (histograms) for WBC, RBC, and PLT can be printed out when this option is selected. From the RESULTS menu, move the cursor to function
2
HISTOGRAM menu is displayed:
and press ENTER. The
PRINTOUT
HH:MM
> * 1 WITH HISTOGRAMS
2 WITHOUT HISTOGRAMS
Move the cursor to the required function and press ENTER : results will be printed and reprinted with or without the histograms according to the selection The third function «HISTO WITHOUT
RBC» allows to print out the results with the PLT and WBC histograms : This function avoids the lapse of time taken to print out all the histograms at the end of an analysis cycle (recommended with the CITIZEN printer).
8.1.3. Units
The operator has the choice between 4 different unit systems accessible when moving the cursor from the RESULTS menu to the function
3 displayed:
UNITS and pressing ENTER. The UNITS menu is
UNITS
HH:MM
> * 1 STANDARD
2 SI
Table 8.1
The 4 different systems are:
WBC
RBC
PLT
HCT
HGB
MCV
MCH
MCHC
MPV
STANDARD
10
3
/mm
3
10
6
/mm
3
10
3
/mm
3
% g/dL
µm
3 pg g/dL
µm
3
SI
10
9
/L
10
12
/L
10
9
/L
L/L mmol/L fL fmol mmol/L fL
INTER 1
10
3
/mm
3
10
6
/mm
3
10
3
/mm
3
% g/dL fL pg g/dL fL
Move the cursor in front of the required unit system and press ENTER.
INTER 2 g/L fL pg g/L fL
10
9
/L
10
12
/L
10
9
/L
L/L
Page 8/2 075D0002-01
8. INSTRUMENT CONFIGURATION
8.1.4. Printer selection
Four different printers and a no printout option can be selected from the RESULTS menu. Move the cursor to the function
4
PRINTER and press ENTER. The PRINTER menu is displayed:
PRINTER
HH:MM
> * 1 RESERVED 1
2 RESERVED 2
Move the cursor to the required printer :
— 1 RESERVED 1 : EPSON (80 columns) printer
— 2 RESERVED 2 : STAR printer
— 3 RESERVED 3 : SEIKO (thermal) printer
— 4 STANDARD : CITIZEN (dot matrix) printer
— 5 NONE or if no printout is required, move the cursor to the function
5
NONE and press ENTER. In this last case, results will have to be recorded by the operator from the display at the end of each analysis.
8.1.5. Temperature printout
The temperature of the diluent during the analysis has to remain in between the specified limits
(18 — 32°C, 65 — 90°F). Results obtained for temperature outside these limits cannot be certified.
When the temperature printout is requested, move the cursor to the function
5
PRINT TEMP and press ENTER. The TEMPERATURE menu is displayed:
PRT. TEMP.
HH:MM
> * 1 YES
2 NO
Move the cursor to the required option and validate. The temperature measured on the diluent circuit will be printed out on each result (sample and QC results).
075D0002-01 Page 8/3
8. INSTRUMENT CONFIGURATION
8.1.6. Print limits
The laboratory limits can be printed out underneath each result when this option is selected.
From the RESULTS menu, move the cursor to function
6 menu is displayed:
and press ENTER. The PRINT LIMITS
PRT LIMITS
HH:MM
> * 1 YES
2 NO
The * indicates the current state. Move the cursor to YES to print the results with the limits and press ENTER. Move the cursor to NO to print the results without the limits and press ENTER.
8.1.7. Print differential results
The operator has the choice to print out or not the differential results. From the RESULTS menu, move the cursor to function
7
and press ENTER. The PRT LMG menu is displayed:
PRT LMG
HH:MM
> * 1 YES
2 NO
The current state of the printout is indicated by the *. Move the cursor to YES to print out the differential and press ENTER. Move the cursor to NO to print out the results without the differential and press ENTER.
8.2. CHANGE LABORATORY LIMITS
Laboratory limits can be set by the operator according to its own specifications. Results that exceed the laboratory limits are identified with a flag : H for results above the upper limit, L for results below the lower limit. From the SETUP menu, move the cursor to the function
2
LAB LIMITS and press ENTER. The LAB LIMITS menu is displayed:
CHG
CHG LAB LIMITS
HH:MM
> 1 LOW LIMITS
2 HIGH LIMITS
8.2.1. Result low limits
Move the cursor to function
1
LOW LIMITS and press ENTER. The LOW LIMITS menu is displayed:
LOW LIMITS
HH:MM
> 1 WBC LOW < XX >
2 RBC LOW < XX >
Page 8/4 075D0002-01
8. INSTRUMENT CONFIGURATION
Move the cursor next to the required parameter and press ENTER, the following menu (RBC for example) is displayed:
RBC LOW ?:
CURRENT :
EXIT : ESC
SAVE : ENTER
Enter the required low value and press ENTER or press the ESC key to keep the current value.
Move the cursor to the next required parameter and follow the same procedure. Press the ESC key when all modifications have been carried out.
8.2.2. Result high limits
From the CHG LAB LIMITS menu, move the cursor to function
2
ENTER.
HIGH LIMITS and press
HIGH LIMITS
HH:MM
> 1 WBC HIGH < XX >
2 RBC HIGH < XX >
Follow the same procedure for the result low limits.
8.2.3. Print limits and flag values
Move the cursor to function
3
PRINT LIMITS and press
ENTER. The high and low limits as well as the flag values are printed out:
Diag. 8.1
NOTE: PCT and PDW are not available in the United States.
075D0002-01 Page 8/5
8. INSTRUMENT CONFIGURATION
8.2.4. Flag limits
A — Platelet flag adjustment
From the menu CHG LAB LIMITS, move the cursor to the function
4
The flag menu is displayed:
FLAGS and press ENTER.
FLAGS LIMITS
HH:MM
> 1 SCL
2 SCH
<8.0>
<8.0>
Move the cursor in front of the flag to be adjusted and press ENTER. The following menu (SCL for exemple) is displayed :
SCL : ?
ACTUEL : 8.00
EXIT : ESC
SAVE : ENTER
Enter the flag required value and press ENTER or press ESC to keep the current value. Move the cursor in front of the next flag and repeat the same procedure. Press the ESC key when all required values have been adjusted to return to the previous menu.
NOTE: The lower the flag value, the higher the triggering sensitivity.
The factory adjustment values are:
SCL : 8.00
SCH : 8.00
MIC : 8.00
B — WBC morphological flags
These flags have to be adjusted by the user according to the representative population of the samples to be analyzed. Specialized hospital laboratories may not have the same detection requirements as outpatient laboratories.
Move the cursor in front of the required flag and press ENTER. Enter the new flag value and press
ENTER. Press the ESC key when all required values have been adjusted to return to the previous menu.
The factory adjustment values are:
L1 : 8.00
M2 : 8.00
G1 : 15.00
G3 : 8.00
Page 8/6 075D0002-01
8. INSTRUMENT CONFIGURATION
8.3. SPECIAL FUNCTIONS
These special functions accessible through a password allow the user to :
1 — Identify 4 users
2 — Change the password
3 — Choose the startup mode
4 — Adjust the cleaning frequency
5 — Print the internal setup of the instrument
6 — Set ON/OFF the cycle end audible signal
7 — Choose the identification mode
8 — Choose the start analysis mode
From the SETUP menu, move the cursor to function
3
SPECIAL and press ENTER. The message : <PASSWORD ?> is displayed. Enter the password <123> or the user-defined password and press ENTER. If the password is correct, the SPECIAL menu is displayed:
SPECIAL
HH:MM
> 1 CHANGE OP
2 CHG PASS. <123>
8.3.1. Change operator
Some of the instrument functions (calibration, QC) requires the operator identification. Four different identifications can be entered and modified at any time by the user. From the SPECIAL menu, move the cursor to the function
1
and press ENTER. The CHANGE OP menu is displayed:
CHANGE OP
HH:MM
> * 1 OP1
2 OP2
Move the cursor in front of the required operator identification to be changed and press ENTER. The following menu is displayed(example for operator 1):
OP1 ? :
CURRENT :
EXIT : ESC
SAVE : ENTER
Enter the new operator identification (4 characters maximum). Alphabetic identification can be entered using the UP and DOWN keys. Press the ENTER key to memorize the identification or
ESC to keep the current figure. Repeat the same procedure for the 3 other identifications if necessary.
075D0002-01 Page 8/7
8. INSTRUMENT CONFIGURATION
8.3.2. Change password
The use of a password is mandatory to access some important functions such as:
— Changing the calibration factors.
— Accessing the technician functions.
— Changing the password.
The original password is <123>. If the change is requested, move the cursor from the SPECIAL menu to function
2
and press ENTER. The PASSWORD menu is displayed:
CHG PASS ? :
CURRENT :
EXIT : ESC
SAVE : ENTER
The current password is displayed. Enter the new password if necessary, 3 numerical characters maximum using the numeric keyboard or ESC to keep the current one.
8.3.3. Startup cycle
The STARTUP cycle is used every day at the instrument startup to rinse out any detergent in the system. The STARTUP cycle includes a background count which must be determined before any samples are analyzed. This is necessary to ensure that there are no extraneous interferences that may be detected as background noise and affect the cell count. If the results of the background count are not within those specified, the analyzer performs a second STARTUP cycle automatically.
This STARTUP cycle can be run automatically at each instrument startup, or manually accessed using the STARTUP key. To configure the instrument according to the operator’s needs, from the
SPECIAL menu, move the cursor to the function
3 menu is displayed:
STARTUP and press ENTER. The STARTUP
STARTUP
HH:MM
1 AUTO
2 MANUAL
Move the cursor to the required setup and press ENTER. When the MANUAL setup is selected, the following message will be displayed at the instrument startup:
STARTUP NOT INITIATED
PRESS A KEY TO CONTINUE…
Press a key to access the MAIN menu then the STARTUP key if an analysis is required.
NOTE:
A message «STARTUP NOT INITIATED» will be printed out with the analysis results when the STARTUP cycle is not carried out after the startup of the instrument.
Page 8/8 075D0002-01
8. INSTRUMENT CONFIGURATION
8.3.4. Autocleaning frequency
An automatic cleaning involving a cleaner solution is normally carried out every 50 samples. The user has the possibilty to setup the automatic cleaning frequency according to the number of samples usually run in his laboratory. From the SPECIAL menu, move the cursor to the function
4
CLEAN. FREQ. and press ENTER. The CLEAN. FREQ.menu is displayed:
CLEAN FREQ. ? :
CURRENT :
EXIT : ESC
SAVE : ENTER
Enter the required frequency and press ENTER. The new automatic cleaning frequency is recorded.
8.3.5. Internal setup printout
The setup of all user options can be printed out using this function. Move the cursor to function
5
PRINT CONFIG. and press ENTER. The instrument internal setup (limits and configuration) is printed out.
Diag. 8.2
075D0002-01 Page 8/9
8. INSTRUMENT CONFIGURATION
8.3.6. Cycle end audible signal
A cycle end audible signal (BEEP) can be set up with the SPECIAL menu. Move the cursor to the function
6
BUZZER and press ENTER. The audible signal menu is displayed:
BUZZER
HH:MM
> * 1 ON
2 OFF
Move the cursor on the required option and press ENTER. The new setup is recorded.
8.3.7. Identification mode
Two identification modes are available:
— US mode: the operator must type in the identification of each patient using letters or numbers.
The identification will be printed out with the results. The STARTUP results are printed out too.
or
— Standard mode: the operator can enter a run number before running an analysis series. This run number will be incremented on each cycle and printed out with the results.
Move the cursor to the function
7
ID MODE and press ENTER. The following menu is displayed:
ID MODE
HH:MM
> 1 STANDARD
* 2 US
Move the cursor to the required option and press ENTER. The new «ID MODE» is recorded.
NOTE:
The US Identification mode allows the use of the barcode reader for an alphanumerical identification.
Page 8/10 075D0002-01
8. INSTRUMENT CONFIGURATION
8.3.8. Start mode
The START MODE function allows the operator to choose between 2 different analysis starting modes:
— In the automatic mode, the analysis can start directly when the operator closes the sample door,
— In the manual mode, the analysis can start after closing the sample door, when the operator presses the START cycle key.
To setup the instrument according to the operator’s needs, from the SPECIAL menu, move the cursor to the function
8 displayed:
START MODE and press ENTER. The START MODE menu is
START MODE
HH:MM
> * 1 AUTO
2 MANUAL
NOTE: Pressing the ESC key will open the door of the tube holder compartment.
8.4. DATE AND TIME
Date and time can be changed according to the country specifications. From the SETUP menu, move the cursor to the function
4 displayed:
DATE TIME and press ENTER. The DATE TIME menu is
DATE TIME
HH:MM
> 1 CHG TIME
2 DATE FORMAT
075D0002-01 Page 8/11
8. INSTRUMENT CONFIGURATION
8.4.1. Change time
Move the cursor to the function
1
and press ENTER. The CHANGE TIME menu is displayed:
NEW TIME (HH.MM) ? :
CURRENT :
EXIT : ESC
SAVE : ENTER
Enter the required time in the format HH.MM and press ENTER. The new time is recorded.
8.4.2. Date format
From the DATE TIME menu, move the cursor to the function
2
DATE FORMAT (the current setting is displayed) and press ENTER. The DATE FORMAT menu is displayed:
DATE FMAT
HH:MM
> * 1 MM.DD.YY
2 DD.MM.YY
4 different date formats can be used : MM.DD.YY
DD.MM.YY
YY.MM.DD
YY.DD.MM
Move the cursor in front of the required selection and press ENTER. The new date format is recorded.
8.4.3. Change date
From the DATE TIME menu, move the cursor to the function
3
ENTER. The CHG DATE menu is displayed:
CHANGE DATE and press
NEW DATE (MM.DD.YY) ? :
CURRENT :
EXIT : ESC
SAVE : ENTER
Enter the new date according to the format recalled in the menu and press ENTER. The new date is recorded.
Page 8/12 075D0002-01
8. INSTRUMENT CONFIGURATION
8.5. HOST OPTIONS
The ADVIA 60-
CT
is capable of transmitting data to an external laboratory computer via the
RS 232 interface. If you have an external computer to be connected to the analyzer, plug one end of the computer cable (provided by your computer vendor) into the external computer. Plug the other end of the computer cable into the cable receptacle at the rear of the instrument.
NOTE:
The ADVIA 60-CT has to be setup according to the external computer specifications.
The following functions have to be used by your laboratory computer specialist only.
The default setup of the serial port are as follow :
1 — Byte number : 8
2 — Parity : none
3 — Stop byte : 1
4 — Xon / Xoff : none
From the SETUP menu, move the cursor to the function
5
HOST OPTIONS and press ENTER.
The HOST OPTIONS menu is displayed:
HOST OPTIONS
HH:MM
> 1 HOST COMM.
2 BAUD RATE
Move the cursor in front of the required option and press ENTER to access the different settings.
Move the cursor in front of the required setting and press ENTER to record the new setup. The different options and their settings are as follow:
8.5.1. Host communication
1 — FORMAT 1
2 — FORMAT 2
3 — STANDARD
4 — TR OFF (transmission OFF)
8.5.2. Baud rate
1 — 300
2 — 1200
3 — 2400
4 — 4800
5 — 9600
8.5.3. Transmission
From the HOST OPTIONS menu, move the cursor to the function
3
TRANSMISSION and press ENTER. The last results are transmitted to the external laboratory computer via the RS 232.
075D0002-01 Page 8/13
8. INSTRUMENT CONFIGURATION
8.6. BARCODE SETUP
The barcode reader can be set up according to the barcode label specifications (checksum).
From the SETUP menu, move the cursor to the function
6
BARCODE menu is displayed:
BARCODE and press ENTER. The
CHECKSUM
HH:MM
Move the cursor to the required option and validate.
> 1 YES
2 NO
NOTE:
Make sure that the «US» identification mode has been set up for the use of the barcode reader.
8.7. Patient Memory Card
8.7.1. Introduction
This menu is available only if the ADVIA 60-
CT
is equipped with a smart card reader.
The MEMORY program contains the seven following functions:
1 — MEMO : allows the operator to enable/disable the memory function.
2 — TRANSMISSION : allows the operator to select the transmission mode : printer or host.
3 — PRINT LIST : the purpose of this submenu is to print or to send to the host the list of the patient identification stored on the memory smart card.
4 — TRANS. ONE : this submenu allows the operator to print or to send to the host one result only.
5 — TRANS. ALL : this submenu allows the operator to print out or to send to the host all the results recorded on the memory smart card.
6 — TRANS. FROM TO : print out or send to the host results from XXX to YYY.
7 — CLEAR CARD : erase all the results recorded on the memory smart card.
From the SETUP menu move the cursor to the function
7
The memory menu is displayed:
MEMO. CARD and press ENTER.
MEMO. CARD
HH:MM
> 1 MEMO < ON >
2 TRANSMISSION
Page 8/14 075D0002-01
8. INSTRUMENT CONFIGURATION
8.7.2. Memo operation mode
8.7.2.1. Memo on/off
Insert the Patient Memory Card into the slot in the upper left corner of the analyzer. From the
MEMOCARD menu, move the cursor to function
1 menu is displayed:
MEMO and press ENTER. The following
MEMO
HH:MM
> * 1 ON
2 OFF
This function allows the user to enable or disable the result backup on the Patient Memory Card.
The * indicates the current state of the memory function. To activate the MEMO function, move the cursor to 1 — ON and press ENTER. To disable move the cursor to 2 — OFF and press ENTER.
NOTE:
As soon as the MEMO is ON, the identification mode turns to «US» mode. It is mandatory to enter the identification before running the analysis.
8.7.2.2. Transmission mode
The operator can chose the transmission mode of the results recorded on the Patient Memory
Card : printer or host computer. From the MEMOCARD menu, move the cursor to function
2
MEMO and press ENTER. The following menu is displayed:
TRANSMISSION
HH:MM
> * 1 PRINTER
2 HOST
Move the cursor to the required mode and validate. The * indicates the current state of the memory function.
075D0002-01 Page 8/15
8. INSTRUMENT CONFIGURATION
8.7.3. Running the samples
Place the sample tube in the sample holder and close the door to start the analysis cycle. The following menu is displayed:
PAT ID ? :
CURRENT
NO SAVE : ESC
SAVE : ENTER
Enter the patient identification and press ENTER. Then, the analyzer checks if a Patient Memory
Card is present.
— If the Patient Memory Card has not been inserted, the following message appears:
NOTE:
ERROR : NO SMART CARD…
INSERT A NEW CARD
Insert the Patient Memory Card and press ENTER.
NO : ESC
YES : ENTER
— If the smart card is not a Patient Memory Card, the following message warns the operator:
ERROR : BAD SMART CARD…
INSERT A NEW CARD
NO : ESC
YES : ENTER
Replace the smart card with a Patient Memory Card and press ENTER.
— If the smart card has been introduced in a wrong way
ERROR : BAD CARD INSERTION
INSERT A NEW CARD
NO : ESC
YES : ENTER
Remove the smart card and insert it with the arrow facing toward the system, and on the upper side and press ENTER.
When the smart card is full:
ERROR : MEMORY CARD FULL
INSERT A NEW CARD
Insert a new one and press ENTER.
NO : ESC
YES : ENTER
When the analysis is completed the results are stored on the smart card, as well as the patient identification.
The graphic results are not recorded on the smart card.
Sixty results can be stored on the smart card. When this one is full, the operator must either replace the smart card with a new one or clear the results from the old smart card.
The results are stored under a backup number which corresponds to the index of the results recorded on the smart card.
NOTE: Results cannot be recorded if the MEMO is OFF.
Page 8/16 075D0002-01
8. INSTRUMENT CONFIGURATION
8.7.4. Print list
The PRINT LIST option allows the operator to print out all the patient identifications recorded on the Patient Memory Card, the time and the date of the analysis as well as the backup number
(column MEMO).
From the MEMOCARD (Patient Card) menu, move the cursor to
3
ENTER.
PRINT LIST and press
MEMO CARD
HH:MM
> 3 PRINT LIST
4 TRANS ONE
Diag. 8.3
Example :
MEMO
3
4
1
2
5
DATE
10/23/1998
10/23/1998
10/23/1998
10/26/1998
10/26/1998
TIME
16:39
17:18
17:26
17:33
17:36
NAME
123
951
1235
784
895
8.7.5. Trans. one
Reprinting or sending to the host one result involves the backup number. This backup number is displayed with results underneath the patient identification.
MM/DD/YY
HH:MM
PAT. ID :
MEMO :
From the MEMOCARD (Patient Card) menu, move the cursor to
4
ENTER. the following menu is displayed :
TRANS. ONE and press
MEMO ? :
CURRENT
NO SAVE : ESC
SAVE : ENTER
Enter the backup number of the results to be printed out and press ENTER. The results are printed out with the backup number (on the right, below the time).
NOTE: The total number of results stored on the card is displayed in CURRENT [ ].
075D0002-01 Page 8/17
8. INSTRUMENT CONFIGURATION
8.7.6. Trans. all
This function allows the operator to reprint or to send to the host all the results stored on the smart card. Check the quantity of paper before running a REPRINT ALL. From the MEMOCARD
(Patient Card) menu, move the cursor to
5
TRANS. ALL and press ENTER. The results are printed out or sent to the host in the usual format without graphs.
If the printout has to be interrupted press ESC until it has been detected by the program (the program checks the keys only at the end of a result printout).
8.7.7. Trans. from to
This function allows the operator to printout or to send to the host only one part of the results recorded on the smart card. From the MEMOCARD (Patient Card) menu, move the cursor to
6
TRANS. FROM TO and press ENTER.
TRANS. FROM TO > 1 — BEGIN <
2 — END <
>
>
Move the cursor to
1
BEGIN and press ENTER : the BEGIN submenu is displayed:
BEGIN ? :
CURRENT : 1
NO SAVE : ESC
SAVE : ENTER
Type in the backup number of the first result to be transmitted and press ENTER.
Move the cursor to
2
END and press ENTER : the END submenu is displayed:
END ? :
CURRENT : 3
NO SAVE : ESC
SAVE : ENTER
Type in the backup number of the last result to be transmitted and press ENTER. Move the cursor to
3
SEND RESULTS and press ENTER. The results are printed out in the usual format without the graphs.
8.7.8. Clear smart card
A single result cannot be erased by itself. This function allows the operator to erase all the results recorded on the smart card. From the MEMOCARD (Patient Card) menu, move the cursor to
7
CLEAR CARD and press ENTER. A message is displayed to alert the operator before erasing any data. Press the ENTER key to erase any data or press the ESC key to cancell this function.
The results are erased and the backup number is updated to 0. The smart card can be reused.
Page 8/18 075D0002-01
CHAPTER 9. MAINTENANCE AND TROUBLESHOOTING
9.1. MAINTENANCE AND SERVICE
9.1.1. Overview
One of the principle factors contributing to accurate and reliable results is a well-maintained instrument. Routine maintenance is required to keep your instrument functioning properly. The
ADVIA 60-
CT
is designed to keep this maintenance automated and to a minimum. This section describes the daily and periodic maintenance procedures.
BIOHAZARD: Wear personal protective equipment. Use universal precautions. Refer to Appendix B for recommended precautions when working with biohazardous materials.
9.1.2. Daily maintenance
These cleaning procedures are required daily to maintain optimum performance of your ADVIA 60-
CT.
CAUTION: Failure to perform any of these recommended cleaning procedures may result in decreased reliability of your system.
9.1.2.1. Startup and standby cycles
At the beginning of each day, a startup cycle must be performed. This can be performed automatically without the operator’s involvement or manually by pressing the STARTUP key.
At the end of each day, press the STAND BY key. The instrument goes into «STANDBY» at the end of the cycle. This cycle takes 1 minute. Leave the instrument in this mode overnight (instrument has to be switched off in this mode) with sysKLEN in the chambers.
9.1.2.2. Automatic cleaning
A cleaning cycle is activated automatically after the number of samples programmed by the operator. The cycle frequency can be adjusted to the laboratory workload. This one can be run directly from the service menu.
9.1.2.3. Instrument general cleaning
In general, intrument has to be cleaned with a wet piece of cloth. Use water and a drop of liquid soap if necessary to clean the outside of the instrument. Never use solvant or abrasive materials.
Wipe off any trace of blood as soon as possible. Disconnect instrument from the main electrical supply before any cleaning intervention and make sure the instrument is clean and dry before reconnection.
075D0002-01 Page 9/1
9. MAINTENANCE AND TROUBLESHOOTING
9.1.3. Service functions
Several service functions are available for the user to clean and check his instrument. These functions are accessible from the Main menu. Move the cursor to the Function
4 and and press ENTER. The SERVICE menu is displayed:
SERVICE
SERVICE
HH:MM
> 1 BACKFLUSH
2 DRAIN CHAMBERS
9.1.3.1. Backflush
This cycle allows the user to clean the instrument aperture in case of blockages. Move the cursor to the function
1
BACKFLUSH and press ENTER. The backflush cycle is carried out and lasts approximately 22 seconds. Check that the liquids are aspirated from the WBC and RBC chambers through the apertures. Check that the liquids are also pushed back into the chambers. If it is not the case, the aperture may be blocked. Perform a concentrated cleaning procedure.
9.1.3.2. Drain chambers
This cycle allows the user to check that the chambers are drained properly and to maintain some parts of the hydraulic manifold as it flushes the liquid out of the instrument. This cycle lasts approximately 25 seconds. Move the cursor to the function
2
ENTER.
DRAIN CHAMBERS and press
9.1.3.3. Prime reagents
This cycle allows the user to prime the reagents when replacing one or all reagent containers (or the pack). Move the cursor to the function
3 required priming cycle.
PRIME REAGENTS and press ENTER. Run the
Select the function
1
: CHANGE PACK and follow the instructions given by the LCD in order to install the pack. Once the new PACK installed a priming cycle will be automatically carried out.
Visually inspect reagent lines and pump. Check that they are clear of air bubbles.
Page 9/2 075D0002-01
9. MAINTENANCE AND TROUBLESHOOTING
NOTE:
9.1.3.4. Concentrated cleaning
This procedure provides a strong cleaning of the RBC and WBC apertures. This cycle lasts approximately 3 minutes and involves the operator intervention to fill the chambers with sysCLEAR or diluted bleach solution.
The diluted bleach solution is obtained by a 1:5 dilution of a commercial bleach solution containing from 10 to 15% of sodium hypochlorite.
WARNING: Commercial strength bleach is 10 to 15% sodium hypochlorite. When handling bleach, which can be used as a cleaning and antiviral agent, wear protective clothing, gloves, and safety glasses. It is harmful if swallowed and may cause eye or skin irritation.
Use bleach that is free of heavy metals.
To prepare a 1:5 dilution of a commercial bleach solution, dilute one part of bleach with five parts of clean distilled water, or clean deionized water. The prepared solution is stable for one week when stored at room temperature.
Move the cursor to the function
4 menu is displayed:
CONCENTRATED CLEANING and press ENTER. The following
PLEASE OPEN COVER DOOR
PRESS A KEY TO CONTINUE …
Open the instrument cover door and press any key. The following menu is displayed:
POUR 3ML OF SYSCLEAR IN WBC CHAMBER
PRESS A KEY TO CONTINUE …
Using a 5 mL syringe, pour 3 mL of sysCLEAR into the WBC chamber and press any key. The next menu is displayed :
POUR 3ML OF SYSCLEAR IN RBC CHAMBER
PRESS A KEY TO CONTINUE …
Pour 3 mL of sysCLEAR inside the RBC chamber and press any key. The next menu is displayed:
PLEASE CLOSE COVER DOOR
PRESS A KEY TO CONTINUE …
Close the instrument cover door and press any key. The following menu is displayed:
CONCENTRATED CLEANING WAIT 2MNS 27S
************
075D0002-01
Stars are displayed at the beginning of the cycle. Every 10 seconds a star is cleared off. The procedure involves different cycles, backflush, aspiration, rinsing, which allow a perfect cleaning of the apertures. After this procedure is completed, run a startup cycle, then sample analysis can begin.
Page 9/3
9. MAINTENANCE AND TROUBLESHOOTING
9.1.3.5. Mechanical checks
Move the cursor to the function
5
MECHANIC and press ENTER. The MECHANIC menu is displayed :
MECHANIC
HH:MM
> 1 CHECK SENSORS
2 NEEDLE U/P
A — Sensor operations : This function allow the user to check the correct detection of the motor home positions. Move the cursor to the function
1
CHECK SENSORS and press ENTER. The
CHECK SENSORS menu is displayed :
NEEDLE SENSOR :
CARRIAGE SENSOR :
0
0
— NEEDLE SENSOR : open the instrument front door and push upward the sampling needle support to the top. Check the correct detection on the display : 0 turns to 1 followed by 10 stars.
— CARRIAGE SENSOR : with the sampling needle in its upper position, move the sampling carriage on the right hand side position. Check the correct detection on the display : 0 turns to 1 followed by 10 stars.
NEEDLE SENSOR :
CARRIAGE SENSOR :
1 **********
1 **********
Press any key to exit the function.
B — Needle up and down operation : Move the cursor to the function
2
NEEDLE U/D and press ENTER. Closely observe the needle translation, the movement has to be smooth and regular.
C — Carriage left/right operation : Move the cursor to the function
3
CARRIAGE L/R and press ENTER. Closely observe the needle translation, the movement has to be smooth and regular.
D — Liquid syringe : Move the cursor to the function
4
LIQUID SYRINGE and press ENTER.
Closely observe the syringe translation, the movement has to be smooth and regular.
E — Pressure/vacuum syringe : Move the cursor to the function
5
PRESSURE SYRINGE and press ENTER. Closely observe the syringe translation, the movement has to be smooth and regular.
F — Valves : Move the cursor to the function
6
VALVES and press ENTER. Closely observe the valve operations, the movement has to be straight and regular.
NOTE: On the ADVIA 60-
CT
, the valve #3 is not installed.
Page 9/4 075D0002-01
9. MAINTENANCE AND TROUBLESHOOTING
G — Change contrast : The display contrast can be adjusted for a better readability. Move the cursor to the function
7 displayed:
CHG CONTRAST and press ENTER. The CHG CONTRAST menu is
CONTRAST +
HH:MM :
: PRESS ENTER TO VALID
Press the UP key of the numeric keyboard to increase the contrast, or press the DOWN key to reduce the contrast. When the best contrast is obtained, press ENTER to validate the setup.
NOTE:
The contrast adjustment menu can be accessed from any other menu by pressing simultaneously the DEL and «.» keys.
H — Park : Raises the vacuum/waste syringe piston in the upper position and the dilution block syringes in the lower position.
I — Piercing : Allows the operator to check the needle height adjustment acccording to the sample tube holder position. Close the sample compartment door. Move the cursor to the function
9
PIERCING and press ENTER : A piercing test is carried out and the needle low position measured. The following menu is displayed:
NEEDLE X
CURRENT XXX STANDARD XXX
If the needle height has to be readjusted, call your local technical support provider or distributor.
075D0002-01 Page 9/5
9. MAINTENANCE AND TROUBLESHOOTING
9.1.3.6. Cycle functions
These functions allow the user to check the cycle numbers run on the instrument. From the
SERVICE menu, move the cursor to the function
6 menu is displayed :
CYCLE and press ENTER. The CYCLE
CYCLES
HH:MM
> 1 STARTUP <4097>
2 STANDBY <6234>
The following cycles can be reviewed : STARTUP, STANDBY, and CBCs.
9.1.3.7. Technician functions
This functions allows the field service engineer to check the instrument on a technical basis.
These functions can be accessed only with the use of a specific password.
9.1.3.8. Automatic cleaning
Move the cursor to function
8
and press ENTER.
AUTO CLEANING
***************************
PLEASE WAIT 3MNS 00S
The cycle frequency can be adjusted to the laboratory workload.
9.1.4 Disposal of System Waste and Supplies
Laws and regulations enacted to protect the environment and to encourage resource conservation require the disposal of hazardous and biohazardous wastes in a specified manner. Some of the wastes from the ADVIA 60-
CT
Hematology System can be classified as hazardous or biohazardous wastes. It is essential that the laboratory take appropriate steps to determine the laws and regulations applicable to their disposal and to effect compliance. If it is necessary to sample instrument wastes and effluent in order to evaluate compliance with applicable regulations, the laboratory should contact a local licensed biohazardous waste disposal firm for assistance.
The principal wastes associated with the use of the ADVIA 60-
CT
Hematology System are the
TIMEPAC reagent containers, and the sysCLEAR reagent.
Test tubes with human specimens and control materials should also be handled and disposed of in accordance with the prevailing regulations and guidelines of agencies with jurisdiction over the laboratory. Refer to the product label and to Material Safety Data Sheets for details concerning any special precautions related to the handling of ADVIA 60-
CT
TIMEPAC containers. Material
Safety Data Sheets are available from Bayer HealthCare.
Page 9/6 075D0002-01
9. MAINTENANCE AND TROUBLESHOOTING
9.2. TROUBLESHOOTING
9.2.1. Overview
Your ADVIA 60-
CT
may occasionally require troubleshooting if :
— System operations are faulty.
— The background count is unacceptably high.
— Your control values are out-of-ranges, or patient results are suspicious (e.g., consistently high
RBC counts or inability to verify results by manual methods).
— Precision is poor.
— Calibration is drifting.
NOTE:
To locate parts of the analyzer mentioned in the following discussions, see the diagrams in section 5.
9.2.2. Identification procedure
The first step in any troubleshooting session is to identify the source of the system malfunction: system operations, reagents, precision, or calibration. These steps should be carried out in sequence as described below :
9.2.2.1. System operations
Press the START key and observe the instrument operation as described in section 4. If the ADVIA
60-
CT
appears to be operating properly, continue with the identification procedure. If operations are faulty, identify the source of the malfunction and initiate appropriate troubleshooting procedures.
9.2.2.2. Reagents
If your background count is unacceptable, your control values are out-of-range, or your patient results are suspicious, reagent deterioration or contamination can be suspected. Replace your reagents and perform the concentrated cleaning procedure. Obtain a background count, and if appropriate, reassay controls or patient samples. If the background count is acceptable, but control values are still out-of-range (or patient results are still suspicious), continue with identification procedure. If replacing reagents and performing concentrated cleaning does not correct the background count, call your local technical support provider or distributor.
9.2.2.3. Precision
Analyze a fresh patient sample 5 to 10 times and calculate the coefficient of variation (%CV).
NOTE:
The %CV is calculated by dividing the standard deviation of the measurements by the mean and multiplying this result by 100.
075D0002-01 Page 9/7
9. MAINTENANCE AND TROUBLESHOOTING
SD =
Σ ( X — Xi ) 2 n — 1 where :
X
= mean
X
i
= individual value
SD
= standard deviation
The following CVs should be obtained : n
= # of observations
Parameters
WBC
RBC
HGB
HCT
PLT
%CV
<2.5%
<2.0%
<1.7%
<2.0%
<5.0%
Proceed with the identification procedure if precision is acceptable. If the precision of any parameter is not within these specifications, identify the out-of-range parameter(s) and initiate appropriate troubleshooting procedures.
9.2.2.4. Calibration
If the system appears to be operating properly, fresh uncontaminated reagents are being used, and the precision is within the specifications, the ADVIA 60-
CT
may need calibration. Calibrate your instrument as described in section 7.
9.2.3. Troubleshooting parameters
The procedures described below should be performed whenever the precision of a parameter is not within the specifications noted above, or a parameter result is incorrect or suspicious.
9.2.3.1. WBC and HGB
Perform the following if both your WBC and HGB results are incorrect or suspicious. Press the
START key and closely observe the specific operations of the analyzer listed below in the order specified. Identify the malfunction(s) and initiate appropriate troubleshooting procedure(s).
Sample probe : During the initial dilution cycle, is the sample probe between the edge of the
WBC chamber and the center of the chamber, close to the bottom?
Page 9/8 075D0002-01
9. MAINTENANCE AND TROUBLESHOOTING
If it is, continue. If sample probe is not in its correct position in the mixing chamber, call your local technical support provider or distributor.
Diag. 9.1
Diluent dispenser : Do you see bubbles in the diluent dispenser? Is the plunger of the diluent dispenser moving up and down smoothly during sample analysis?
If you do not see bubbles and the diluent dispenser is operating properly, continue. If you see bubbles, or if the diluent dispenser is not operating properly, refer to Mechanical Checks for appropriate actions.
Diag. 9.2
Transfer of lyse to WBC chamber : Is the lyse pump plunger moving up and down smoothly? Can you see air bubbles?
If the lyse pump is operating properly, and no bubbles are seen (thus system operations all appear to be acceptable), call your local technical support provider or distributor. If bubbles are seen, change the reagent pack and prime the reagent lines.
Diag. 9.3
Call your local technical support provider or distributor if this does not correct the WBC and HGB results.
075D0002-01 Page 9/9
9. MAINTENANCE AND TROUBLESHOOTING
9.2.3.2. WBC
Check the following if only the WBC count is incorrect or suspicious.
Concentrated cleaning procedure : Was the concentrated cleaning procedure performed earlier as part of the identification procedure? If not, perform the concentrated cleaning procedure.
Calibration : Was the system calibrated earlier as part of the identification procedure. If it was not, and the WBC count is still erroneous after the concentrated cleaning procedure was performed, calibrate the instrument as described in section 7. Continue troubleshooting if:
— This does not correct the WBC results.
— The instrument has already been calibrated as part of the identification procedure.
— Earlier attempts to calibrate the WBC during the identification procedure failed.
Analyze a sample and observe the operation of the liquid valve <6>.
Valve <6> : Is liquid valve <6> opening and closing during analysis cycle? If the liquid valve <6> is not opening and closing, replace the valve. If this does not correct the WBC count, call your local technical support provider or distributor.
If the valve is operating properly, call your local technical support provider or distributor.
9.2.3.3. HGB
Check the following if only the HGB results are incorrect or suspicious.
Concentrated cleaning : Was the concentrated cleaning performed? If it was not, perform the concentrated cleaning. Continue troubleshooting if this does not correct the HGB result.
Calibration : Was the system calibrated earlier as part of the identification procedure? If HGB was not calibrated as part of the identification procedure, calibrate it as described in section 7.
If all attempts to calibrate this parameter have failed, press the START key and closely observe the specific operations of the analyzer listed below in the order specified. Identify the malfunction(s) and initiate appropriate troubleshooting procedure(s).
If upon close inspection all operations still appear to be acceptable, call your local technical support provider or distributor.
Lyse pump : Is the lyse pump plunger moving up and down smoothly? Can you see air bubbles?
If the lyse pump is operating properly and bubbles are not observed, continue. If lyse pump is not operating properly or bubbles are observed, see Mechanical Checks for troubleshooting procedure.
Page 9/10 075D0002-01
9. MAINTENANCE AND TROUBLESHOOTING
HGB LED : Is the HGB LED illuminated when the system power is on? If it is and therefore all system operations are acceptable, call your local technical support provider or distributor. If the HGB LED is not illuminated when the system power is on, call your local technical support provider or distributor.
Diag. 9.4
9.2.3.4. RBC, HCT and PLT
Check the following if RBC, HCT, and PLT results all appear to be erroneous.
Concentrated cleaning procedure : Was the concentrated cleaning performed earlier as part of the identification procedure? If not, perform the concentrated cleaning procedure. Then press the START key and closely observe the specific operations of the analyzer listed below in the order specified. Identify the malfunction(s) and initiate appropriate troubleshooting procedure(s). If upon close inspection all operations still appear to be acceptable, call your local technical support provider or distributor.
Sampling syringe : Is the sample syringe moving up and down? If the 10 µL syringe is operating properly during sample analysis, continue. If the sampling syringe is not operating properly during sample analysis, call your local technical support provider or distributor.
Diluent dispenser : Do you see any bubbles in the diluent dispenser? Is the plunger of the diluent dispenser moving up and down smoothly during sample analysis? If no air bubbles are seen, and the diluent dispenser plunger is operating properly, continue. If bubbles are seen, or the plunger of the diluent dispenser is not operating properly, see Mechanical Checks for actions.
First dilution : Is there a stream of air bubbles in the WBC chamber during the initial dilution cycle? Is the sample probe between the edge of the chamber and the center of the chamber, close to the bottom?
If there is a stream of bubbles, and the sample probe is in its proper position in the WBC chamber
(between the edge of the chamber and the center of the chamber, close to the bottom) continue by observing the RBC chamber (see RBC chamber below). If there is no stream of bubbles in the
WBC chamber during the initial dilution cycle, check the operation of the liquid valve <12>.
If the liquid valve <12> is not opening and closing, replace the valve. If this does not correct the problem, call your local technical support provider or distributor. If the valve is operating properly, call your local technical support provider or distributor.
If the sample probe is not in its proper position in the WBC chamber (between the edge of the chamber and the center of the chamber, close to the bottom) call your local technical support provider or distributor.
075D0002-01 Page 9/11
9. MAINTENANCE AND TROUBLESHOOTING
RBC chamber : Is the sample probe between the edge of the chamber and the center of the chamber, close to the bottom?
If the sample probe is in its proper position in the
RBC chamber (between the edge of the chamber and the center of the chamber, close to the bottom), and therefore system operations appear to be acceptable, call your local technical support provider or distributor.
If the sample probe is not in its proper position in the
RBC chamber (between the edge of the chamber and the center of the chamber, close to the bottom) call your local technical support provider or distributor.
Diag. 9.5
9.2.3.5. RBC
Perform the following actions if only the RBC parameter is erroneous or suspicious. Analyze a normal control and observe the specific operation of the analyzer listed below in the order specified.
Identify the malfunction(s) and initiate appropriate troubleshooting procedure(s). If upon close inspection all operations still appear to be acceptable, call your local technical support provider or distributor.
Valve <6> : Is liquid valve <6> opening and closing during analysis cycle? If the liquid valve <6> is not operating properly, replace the valve. If the valve is still not operating properly, call your local technical support provider or distributor. If the valve is operating properly, continue the procedure.
Sampling syringe : Is the sample syringe moving up and down ? If the 10 µL syringe is not operating properly during sample analysis, see Mechanical Checks for troubleshooting actions. If the sampling syringe is operating properly during sample analysis, continue the procedure.
Diluent dispenser : Do you see any bubbles in the diluent dispenser? Is the plunger of the diluent dispenser moving up and down smoothly during sample analysis? If no air bubbles are seen, and the diluent dispenser plunger is operating properly, continue. If bubbles are seen, or the plunger of the diluent dispenser is not operating properly, see Mechanical Checks for troubleshooting actions.
Valves <8> and <11> : Are liquid valves <8> and <11> opening and closing during analysis cycle? If the liquid valves <8> and <11> are not opening and closing, replace the valves. If this does not correct the RBC count, call your local technical support provider or distributor.
If the valves are operating properly, call your local technical support provider or distributor.
Page 9/12 075D0002-01
9. MAINTENANCE AND TROUBLESHOOTING
9.2.3.6. HCT
Check the following if only the HCT result is incorrect or suspicious.
Calibration : Was the system calibrated earlier as part of the identification procedure? If HCT was not calibrated as part of the identification procedure, calibrate it as described in section 7.
Call your local technical support provider or distributor if this does not correct the HCT result.
9.2.3.7. PLT
Perform the following actions if only the PLT parameter is erroneous or suspicious. Review patient results.
Sample flags : Were there frequent $ flags next to the PLT results? If there were, replace your reagents and perform the concentrated cleaning procedure if it was not already done as part of the identification procedure. If you do not observe $ flags on patient samples, continue.
Calibration : Was the system calibrated earlier as part of the identification procedure? If PLT was not calibrated as part of the identification procedure, calibrate it as described in section 7.
Call your local technical support provider or distributor if this does not correct the PLT result.
Sample probe : During the initial dilution cycle, is the sample probe between the edge of the mixing chamber and the center of the chamber, close to the bottom? If it is, continue. If sample probe is not in its correct position in the mixing chamber, call your local technical support provider or distributor.
9.2.4. Troubleshooting system operations
The procedures described on the next pages should be performed whenever system operations are faulty. Identify the source of the malfunction and initiate the appropriate troubleshooting procedure. A «SERVICE» menu is available in the ADVIA 60-
CT
program to help the user in the troubleshooting of the hydraulic transfers and mechanical operations. This SERVICE menu can be accessed when the following checks have been passed.
9.2.4.1. Power
Check the following if you are unable to analyze a sample because of a lack of power to the
ADVIA 60-
CT
. Is the power switch on? If it is, continue. Is the power cord plugged into the wall outlet? If it is, continue.
Is there current in the wall outlet? If there is, continue. If there is no current, call your maintenance department.
Are the fuses still functioning? If they are, and you are to identify the source of the power failure, call your local technical support provider or distributor. If the fuses are defective or blown, replace them.
9.2.4.2. Display
When the instrument is switched on, if an audible alarm occurs and the display remains dark, check the following : Are both connectors properly fitted?
— Check the flat cable connection on the display.
— Check the flat cable connection on the electronic board.
If they are, and you have therefore been unable to identify the source of the display failure (and no audible alarm occured), press simultaneously on the «.» and DEL keys to access the display adjustment menu. Follow the procedure described in Mechanical Checks.
If the display remains dark, call your local technical support provider or distributor.
075D0002-01 Page 9/13
9. MAINTENANCE AND TROUBLESHOOTING
Diag. 9.6
9.2.4.3. Motors
After having switched on the instrument or whenever a problem occurs, the display may show some error messages concerning a failure in one of the motor initialization. These messages appear for a few seconds on the display and are as follow :
— «ERROR : NEEDLE MOTOR»
— «TRANSFER MOTOR ERROR»
— «ERROR : LIQUID SYRINGE MOTOR»
— «ERROR : PRESSURE SYRINGE MOTOR»
— «ERROR : PIERCING MOTOR»
— «ERROR : CARRIAGE MOTOR»
Are the motor connectors properly fitted on the electronic board? If they are, and you have therefore been unable to identify the source of the motor failure, call your local technical support provider or distributor.
Diag. 9.7
1 — Pressure syringe motor
2 — Liquid syringe motor
3 — Transfer motor
4 — Needle motor
5 — Fan motor
6 — Piercing motor
When the STARTUP checks have been passed, it is now possible to access the SERVICE menu in order to check the operation of the hydraulics and the mechanical parts.
Page 9/14 075D0002-01
9. MAINTENANCE AND TROUBLESHOOTING
9.3. ERROR MESSAGES
Some error messages can be displayed and require the operator intervention. Follow the instructions given for each message and refer to the specified section. If the problem cannot be solved, call your local technical support provider or distributor.
A — «ERROR : PRESSURE SYRINGE MOTOR»: This message can be displayed at the instrument startup. It shows a malfunction on the pressure syringe motor. Check that the motor is properly connected on the electronic board. If it is, call your local technical support provider or distributor.
B — «ERROR : LIQUID SYRINGE MOTOR»: This message can be displayed at the instrument startup. It shows a malfunction on the liquid syringe motor. Check that the motor is properly connected on the electronic board. If it is, call your local technical support provider or distributor.
C — «ERROR : TRANSFER MOTOR»: This message can be displayed at the instrument startup.
It shows a malfunction on the transfer motor. Check that the motor is properly connected on the electronic board. If it is, call your local technical support provider or distributor.
D — «ERROR : NEEDLE MOTOR» : This message can be displayed at the instrument startup. It shows a malfunction on the needle motor. Check that the motor is properly connected on the electronic board. If it is, call your local technical support provider or distributor.
E — «ERROR : PIERCING MOTOR»: This message can be displayed at the instrument startup. It shows a malfunction on the piercing motor. Check that the motor is properly connected on the electronic board. If it is, call your local technical support provider or distributor.
F — «PRINT IN PROGRESS»: This message appears when the operator attempts to print some data during a result printout.
G — «ERROR : OUT OF PAPER»: This message is displayed when the printer runs out of paper.
— Press ESC to clear off the message and to complete the cycle, the last result can be reprinted after installing a new roll of paper or,
— install a new roll of paper and press ENTER. the result will be automatically printed out.
H — «ERROR : PRINTER OFF LINE»: This message is displayed when the printer is off line.
Check the status of the online lamp (or select lamp, according to the type of the printer in use).
Press ENTER when the line is on to print automatically the result.
I — «ERROR : NO PRINTER» : This message is displayed when the printer is off or not connected.
Reconnect the printer or set up the instrument without printer.
J — «ERROR : PRINTER NOT SELECTED» : This message is displayed when a printout is requested and no printer has been selected. Select a printer.
075D0002-01 Page 9/15
9. MAINTENANCE AND TROUBLESHOOTING
NOTE:
K — «ERROR : BAD DATE ! TRY AGAIN» : This message is displayed in the date change function when the new date entered is incompatible with the date format previously entered. Re-enter a correct date or change the date format.
L — «ERROR : BAD TIME ! TRY AGAIN» : This message is displayed in the change time function when the new time has not the required format : HH.MM. Re-enter the correct time.
M — «CYCLE ABORTED BY USER» : This message is displayed whenever the ESC key is pressed during an hydraulic cycle. A confirmation message «CYCLE ABORTED?» is displayed. If the abortion is confirmed by pressing the ENTER key, an initialization cycle is carried out to reinitialize the motors in their home position.
After aborting a hydraulic cycle, it is necessary to run a STARTUP cycle to rinse the instrument before any further sample analysis.
N — «BAD VALUE…MINI : XXX, MAXI : XXX» : this message is displayed in the following occasions :
— When the sample run number is above 9999.
— When the target values entered in the AUTOCALIBRATION are out of the limits.
— When the number of sample selected in the AUTOCALIBRATION is out of the limits (3 to 11).
— When the calibration factors entered in the CALIBRATION function are out of the factor limits.
— When the laboratory limit values entered in the CHANGE LAB. LIMITS are out of the limits.
— When the flag values entered in the FLAG ADJUST function are out of the limits.
— When the automatic cleaning frequency value entered in the AUTO CLEANING function is out of the limits.
NOTE:
Maximum and minimum values available are displayed with the corresponding error message. Correct the values according to the specified indications.
O — «STARTUP NOT INITIATED» : This message is displayed when the instrument is turned on, and the startup cycle is not automatically carried out. This STARTUP cycle is mandatory after each STANDBY cycle in order to flush the instrument from the cleaning reagent. A blank cycle on reagents is performed to check the cleanliness of the instrument. When this message is displayed and a sample analysis is requested, press the STARTUP cycle key.
P — «STARTUP FAILED, CHECK REAGENTS» : This message is displayed when the instrument gives out of range blank values after 3 consecutive startup cycles. Check the expiration dates, replace the reagents if necessary or perform a concentrated cleaning.
Page 9/16 075D0002-01
9. MAINTENANCE AND TROUBLESHOOTING
Q — «HGB REFERENCE FAILED»: this message is displayed when the instrument gives out of range HGB blank values after 3 consecutive HGB reference blank cycles.
Check the expiration dates, replace the reagents if necessary or perform a concentrated cleaning.
R — «4 HOURS ELAPSED BETWEEN LAST OPERATION»: this message is displayed when the instrument has not been used for at least 4 hours. In order to prevent any drift in the results, a
STARTUP has to be performed before a sample is run. Press the ENTER key in order to run the
STARTUP. Run a STANDBY cycle if the instrument has to be stopped.
S — «ERROR : TUBE HOLDER POSITION» : This message is displayed when the tube holder is not in its proper position. Turn it slightly to the right or to the left until a click is heard.
T : «ERROR : NO SAMPLE TUBE HOLDER» : This message is displayed when an analysis cycle is requested but no sample tube holder is installed.
U : «PLEASE, CLOSE SAMPLE TUBE HOLDER DOOR» : This message is displayed when the instrument is set up with the manual start mode and the START key is pressed with the tube holder door open.
V — «TUBE HOLDER DOOR ERROR, PLEASE OPEN
THE DOOR MANUALLY» : This message is displayed together with an audible alarm when the tube holder door is blocked. Open the instrument cover and by means of a flat screwdriver or a finger nail, push slightly on the solenoid washer to open the door.
Diag. 9.8
W — «SENSOR ERROR OR DILUENT EMPTY» :This message is displayed when the instrument detects a problem on the drainage operation.The waste detection cell may be faulty, or the diluent may have run out.
X — «MAX. OP. SAVED» : The operator attempted to change a sixth operator’s name on the current QC card. It is possible to change up to 5 operator names, the sixth one will not be recorded.
075D0002-01 Page 9/17
9. MAINTENANCE AND TROUBLESHOOTING
9.4. PURPOSE OF THE VALVES
Valve #1 : Controls the lyse distribution
Valve #2 : Cancels the pressure/vacuum in the pressure/vacuum syringe
Valve #4 : Controls the cleaner input in the WBC counting head during the rinsing
Valve #5 : Controls the drain of the pressure/vacuum syringe
Valve #6 : Activates the vacuum needed in the WBC & RBC counting heads
Valve #7 : Controls the diluent input in the RBC counting head during the rinsing
Valve #8 : Controls the aspiration of the diluent/air output inside the needle rinse block
Valve #9 : Routes the diluent distribution to the inside or outside of the piercing needle
Valve #10 : Controls the diluent inside the aspiration needle
Valve #11 : Controls the diluent distribution
Valve #12 : Controls the drain of the WBC chamber
Valve #13 : Controls the drain of the RBC chamber
Diag. 9
1 Two Ways NC
2 3 Ways
Page 9/18 075D0002-01
9. MAINTENANCE AND TROUBLESHOOTING
9.5. MENU OVERVIEW
ADVIA 60-CT
4 — DATE TIME
5 — HOST OPTIONS
6 — BARCODE
7 — MEMO CARD
(patient card)
075D0002-01
1 — RESULTS
2 — QC
3 — CALIBRATION
1 — Autocalibration
2 — Coefficients
4 — SERVICE
5 — SETUP
1 — RESULTS
2 — CHG LAB LIMITS
3 — SPECIAL
Password
1 — Backflush
2 — Drain
3 — Prime
4 — Conc. clean
5 — Mechanic
6 — Cycles
7 — Technician
8 — Auto clean
Parameters 2
Parameters 1
Pat .ID / Run #
Flags
Chg lot #
Startup
Stand by
CBC
1 — Automatic
2 — Analysis
3 — Print targets
4 — Statistics
5 — Graphs
Chg target
Select OP
1 — Calibration
2 — Print coefficients
Chg exp date
Calibration card
Password
Select OP
Lot #
SELECT LEVEL
1 — Low
2 — Normal
3 — High
SELECT LEVEL
1 — All
2 — Low
3 — Normal
4 — High
Chg sample Nb
WBC
RBC
HGB
MCV
PLT
MPV
RDW
PDW
1 — Sensors
2 — Needle U/D
3 — Carriage L/R
4 — Liquid syringe
5 — Pres. syringe
6 — Valves
7 — Chg. Contrast
8 — Park
9 — Piercing
Smart card
Run calibration Valid. calibration
Start calibration with pack
Without pack
1 — All reagents
2 — Diluent
3 — Lyse
4 — Cleaner
1 — Chg. pack
2 — Capacity
3 — Prime
1 — Reprint result
2 — Printout < >
3 — Units
4 — Printer
5 — Print temp.Y/N
6 -Print Limits Y/N
7 — Flag LMG Y/N
1 — Reserved 1
2 — Reserved 2
3 — Reserved 3
4 — Standard
5 — None
1 — Standard
2 — SI
3 — Inter 1
4 — Inter2
1 — With histo
2 — Without histo
3 — Histo w/o RBC
1 — Host Comm.
2 — Baud rate
3 — Transmission
Checksum Y/N
1 — Chg OP
2 — Chg. pass.
3 — Startup cycle
4 — Auto clean #
5 — Print config
6 — Buzzer on/off
7 — ID mode
8 — Start mode
1 — Auto
2 — Manual
1 — STD
2 — US
1 — Auto
2 — Manual
1 — Format ABX
2 — Format 2
3 — Standard
4 — TR OFF
1 — 300
2 — 1200
3 — 2400
4 — 4800
5 — 9600
1 — Low limits
2 — High limits
3 — Print limits
4 — Flag limits
2 — SCH
1 — Chg Time
2 — Date format
3 — Chg date
1 — SCL
4 — L1
3 — MIC
1 — Memo
2 — Transmission
3 — Trans. list
4 — Trans. one
5 — Trans. all
6 — Trans. from to
7 — Clear card
6 — G1
5 — M2
1 — Begin
2 — End
7 — G3
3 — Send results
RAM 035 A Ind.C
Page 9/19
9. MAINTENANCE AND TROUBLESHOOTING
Page 9/22 075D0002-01
CHAPTER 10. SERVICE AND SUPPLIES
10.1 Supplies
The following supplies are available from your local distributor:
Product Name
TIMEPAC Reagent Pack
Part Number
B01-4199-54
REF Number
07622536 sysKLEN Detergent sysCLEAR Bleach Solution
B01-4197-01
B01-4198-01
Bayer TESTpoint Hematology Controls
03340897
02488831
Low
Normal
High
B03-4200-54 04575197
B03-4201-54 00133238
B03-4202-54 00056756
ADVIA 60 SETpoint Calibrator B03-4203-51 09345602
Calibrator Memory Card 075-1100-02 08366266
Control Memory Card 075-1000-02 08564998
Patient Memory Card
Bar Code Reader
075-1200-02 05963395
075-0142-01 02184638
075D0002-01 Page 10/1
10. SERVICE AND SUPPLIES
10.2 For Service
To contact the legal representative for Bayer within the European community, contact the Bayer
Authorized Representative. For service, contact your local technical support provider or distributor.
10.2.1 Bayer Authorized Representative
Bayer Diagnostics Europe Limited
Chapel Lane, Swords, Co. Dublin, Ireland
10.2.2 Bayer Offices Worldwide
Manufactured by:
Bayer HealthCare LLC
Subsidiary of Bayer Corporation
Diagnostics Division
511 Benedict Avenue
Tarrytown, NY 10591-5097 USA
914-631-8000
Bayer S.A.
División Diagnósticos
Ricardo Gutiérrez 3652
B1605EHD Munro — Buenos Aires
Argentina
+54 (0) 11-4762-7000
Bayer Australia Limited
ABN 22 000 138 714
Diagnostics Business Group
2 Keith Campbell Court
Scoresby Victoria 3179
Australia
+61 (0) 3-9212-8444
Bayer Austria GesmbH
Geschäftsbereich Diagnostika
Lerchenfelder Gürtel 9-11
A-1164 Wien, Austria
+43 (0) 1-711-46-2424
Bayer s.a.-n.v.
Division Diagnostics
Avenue Louise 143 Louizalaan
1050 Bruxelles-Brussel,
Belgium
+32 (0) 2-535-66-81
Bayer S.A.
Produtos Diagnósticos
Rua Domingos Jorge 1100
04779-900 — São Paulo — SP
Brazil
+55-11-5694-5166
Bayer Inc.
Bayer Diagnostics
Healthcare Division
77 Belfield Road
Toronto, Ontario
Canada M9W 1G6
416-248-0771
Bayer S.R.O.
Litvinovska 609/3
CZ- 190 21 Prague 9 –Prosek
+420 (0) 2-66101463
Bayer S.A.
División Diagnóstica
A.A. 80387
Av. De las Américas No 57-52
Santafé de Bogotá, D.C. – Colombia
+571 (09) 423-4199
Bayer A/S
Nørgaardsvej 32
DK-2800 Lyngby, Denmark
+45-45-23-50-00
Bayer OY
Suomalaistentie 7
FIN 02270 Espoo, Finland
+358-9-887-887
Page 10/2 075D0002-01
Bayer Diagnostics
Tour Horizon
52, quai de Dion Bouton
92807 Puteaux Cedex, France
+33 (0) 1-49-06-56-00
Bayer Vital GmbH
Geschäftsbereich Diagnostics
Siemensstrabe 3
D-35463 Fernwald, Germany
+49 (0) 641-40-03-170
Bayer Hellas S.A.
Akakion 54A
Marousi
Athens 151 25, Greece
+30 (0) 1-6883648
Bayer HealthCare Co. Ltd.
20/F Gee Chang Hong Centre
65 Wong Chuk Hang Road
Hong Kong
852-28147337
Bayer Hungaria Kft.
H-1012 Budapest, Hungary
Pálya u.4-6
+36 (06) 1-212-1540
Bayer Diagnostics India Limited
589, Sayajipura
Ajwa Road
Baroda – 390 019
Gujarat, India
+91 (0) 26-5562720
Bayer S.p.A.
Divisione Diagnostici
Via Grosio 10/4
20151 Milano, Italia
+39-2-3978-1
Bayer Medical Ltd.
1-19-15, Ebisu
Shibuya-Ku
Tokyo 150-0013, Japan
+81 (0) 3-440-2411
Bayer Diagnostics Korea Limited
Kye Myung Bldg. 4F
Myungil-Don, 48-7
Kangdong-Ku
Seoul 134070
Korea
+82 (0) 2-428-5987
075D0002-01
10. SERVICE AND SUPPLIES
Bayer (Malaysia) Sdn. Bhd
19th & 20th Floor Wisma MPSA
P.O. Box 7252
40708 Shah Alam
Selangor Darul Ehsan
Malaysia
+603 (0) 551-02818
Bayer de México, S.A. de C.V.
División Diagnósticos
Via Morelos 330-E
Santa Clara
55540 Ecatepec, Estado de México
México
Centro de Atención Telefónica (CAT)
+55 57 28 33 12
+55 57 28 33 06
Bayer B.V.
Health Care
Division Diagnostics
Energieweg 1
3641 RT Mijdrecht
The Netherlands
+31 (0) 297-280666
Bayer New Zealand Ltd
Diagnostics Business Group
3 Argus Place,
Glenfield,
Auckland, New Zealand
+64-800-724-269
Bayer As
Brennaveien 18
N-1483 Skytta, Norway
+47-67-06-86-00
Bayer Sp. Z o.o.
Al. Jerozolimskie 158
02-326 Warszawa, Polska
+48 (0) 22-572-3500
Bayer Diagnostics Europe Ltd.
(sucursal em Portugal)
Rua da Quinta do Pinheiro, 5
2795-653 Carnaxide, Portugal
+351-21-416-4227
Bayer Puerto Rico Inc.
Diagnostics Division
Victoria Industrial Park
Building 1
Carolina, Puerto Rico
787-752-8989
Page 10/3
10. SERVICE AND SUPPLIES
Bayer (South East Asia) Pte Ltd.
No. 9 Benoi Sector
Singapore 629844
+65-261-3389
Bayer (Pty) Ltd.
Healthcare Division
27 Wrench Road
Isando 1600, South Africa
+27 (0) 11-921-5048
Química Farmacéutica Bayer, S.A.
División Diagnósticos
Calabria, 268
08029 Barcelona, España
+34-93-4956500
Bayer (Schweiz) AG
GB DS
Grubenstrasse 6
CH-8045 Zürich
+41 (0) 1-465-81-11
Bayer AB
Drakegatan 1
S-402 24 Göteborg, Sweden
+46-31-83-98-00
Bayer Taiwan Co. Ltd.
Diagnostics Division
11/F 237 Sung Chiang Road
Taipei 104
Taiwan
+886-2-2-5039123
Bayer plc
Diagnostics Division
Bayer House
Strawberry Hill
Newbury, RG14 1JA
United Kingdom
+44 (0) 1635-563000
Page 10/4 075D0002-01
APPENDIX A. SYSTEM AND REAGENT ROLLS
A.1 System and Reagent Symbols
The following symbols appear on the Advia 60 Hematology System, and the Advia 60 TIMEPAC reagent package.
Warning. Biohazard
Warning. Biological Risk
WARNING: Indicates the risk of personal injury or loss of life if operating procedures and practices are not strictly observed.
CAUTION: Indicates the possibility of damage to or distruction of equipment if operating procedures and practices are not strictly observed. Also indicates the possibility of causing erroneous results and actions that could affect system performance.
Standby State
Start
Start up cleaning cycle
Keypad Input
Enter Function
Escape Function
Delete Function
075D0002-01 Page A/1
APPENDIX A. SYSTEM AND REAGENT SYMBOLS
Page A/2
Up Navigation
Down Navigation
ON/OFF
Ground
Fuse
Serial Port
Printer Port
The input electricity is alternating current.
Caution. Risk of static discharge.
Consult the instructions for use.
The analyzer meets the safety requirements of Underwriters Laboratories.
The analyzer meets the safety requirements of the Canadians Standards Association.
Revision Number
075D0002-01
APPENDIX A. SYSTEM AND REAGENT SYMBOLS
Serial Number
Catalog Number
In Vitro Diagnostic Device
Manufactured Location
Authorized Representative
Date of Manufacture
CE Mark. Product meets the requirements of applicable European Directives.
Temperature limitation (store between x°C – y°C)
Batch code
Use by
Store upright
Recycle
2003-06
Date Format (year-month)
075D0002-01 Page A/3
APPENDIX A. SYSTEM AND REAGENT SYMBOLS
Page A/4 075D0002-01
APPENDIX B. BIOHAZARD PROTECTION
B.1 Protecting Yourself from Biohazards
BIOHAZARD
All products or objects that come in contact with human or animal body fluids should be handled, before and after cleaning, as if capable of transmitting infectious diseases. Wear facial protection, gloves, and protective clothing.
The operator should follow the recommendations to prevent the transmission of infectious agents in healthcare settings as recommended for potentially infectious specimens in Protection of
Laboratory Workers from Infectious Disease Transmitted by Blood, Body Fluids, and Tissue, 2d edition; Approved Guideline (1997) Document M29-A, National Committee for Clinical Laboratory
Standards (NCCLS). This document contains complete information on user protection and it can be used as reference material for instructions on laboratory safety.
The following information summarizes the established guidelines for handling laboratory biohazards.
This summary is based on the guidelines developed by the National Institutes of Health (NIH), the
Centers for Disease Control (CDC), the NCCLS Document M29-A, Protection of Laboratory Workers
from Instrument Biohazards and Infectious Disease Transmitted by Blood, Body Fluids, and
Tissue, and the Occupational Safety and Health Administration’s Bloodborne Pathogens Standard.
1–3
Use this summary for general information only. It is not intended to replace or supplement your laboratory or hospital biohazard control procedures.
By definition, a biohazardous condition is a situation involving infectious agents biological in nature, such as the hepatitis B virus, the human immunodeficiency virus (HIV), and the tuberculosis bacterium. These infectious agents may be present in human blood and blood products and in other body fluids.
The following are the major sources of contamination when handling potentially infectious agents:
• needlesticks
• sharp objects, such as probe tips
• hand-to-mouth contact
• hand-to-eye contact
• direct contact with superficial cuts, open wounds, and other skin conditions that may permit absorption into subcutaneous skin layers
• splashes or aerosol contact with skin and eyes
To prevent accidental contamination in a clinical laboratory, strictly adhere to the following procedures:
• Wear gloves while handling parts of the instrument that have contact with body fluids such as serum, plasma, urine, or whole blood.
• Wash your hands before going from a contaminated area to a non-contaminated area, or when you remove or change gloves.
• Perform procedures carefully to minimize aerosol formation.
• Wear facial protection when splatter or aerosol formation are possible.
• Wear personal protective equipment such as safety glasses, gloves, lab coats or other protective clothing when working with possible biohazard contaminants.
075D0002-01 Page B/1
APPENDIX B. BIOHAZARD PROTECTION
• Keep your hands away from your face.
• Cover superficial cuts and wounds.
• Dispose of contaminated materials according to your laboratory’s biohazard control procedures.
• Disinfect your work area with a 15% bleach solution.
• Do not eat, drink, smoke, or apply cosmetics or contact lenses while in the laboratory.
• Do not pipette any liquid, including water, with your mouth.
• Do not place any tools or any other items in your mouth.
• Do not use the biohazard sink for any personal cleaning, such as rinsing cups or washing hands.
To prevent needlestick injuries, needles should not be recapped, purposely bent, cut, broken, removed from disposable syringes, or otherwise manipulated by hand.
B.2 References
1. Centers for Disease Control. 1988. Update: Universal precautions for prevention of transmission of human immunodeficiency virus, hepatitis B virus and other bloodborne pathogens in healthcare settings. MMWR, 37:377–382, 387, 388.
2. National Committee for Clinical Laboratory Standards. Protection of laboratory workers from instrument biohazards and infectious disease transmitted by blood, body fluids, and tissue; approved guideline. NCCLS Document M29-A. Villanova (PA): NCCLS; 1997 Dec. 90p.
3. Federal Occupational Safety and Health Administration. Bloodborne Pathogens Standard. 29
CFR 1910. 1030.
Page B/2 075D0002-01
APPENDIX C. BIBLIOGRAPHY
C.1 Bibliography
The following information was used to create the reagent information in this manual.
Bessman, JD; (1986) Automated Blood Counts and Differentials, Johns Hopkins University Press
Groner W, Simson E; (1995) Practical Guide to Modern Hematology Analyzers: Wiley 1-241
Corash L; (1983) Platelet Sizing: Techniques, Biological Significance, and Clinical Application; Current
Topics in Hematology, Alan R. Liss Inc. pp 99-122
Hadley GG and Weiss SP; (1955) Further notes on the use of EDTA as an anticoagulant. Am J Clin Path 25:
1090-1093
Koepke JA; (1976) The calibration of automated instruments for accuracy in hemoglobinometry. Lab World
July: 180-184
NCCLS; (1984) (H3-A2) Procedures for the Collection of Diagnostic Blood Specimens by Venipuncture.
NCCLS; (1984) (H4-A2) Procedures for the Collection of Diagnostic Blood Specimens by Skin Puncture.
Grover NB, Naaman J, Ben-asson S and Dojanski F; (1972) Electrical sizing of particles in suspension III.
Rigid spheroids and red blood cells. Biophys J 12:1099-1116.
Hughes-Jones; (1974) Differential Leucocyte counts by volume distribution analysis. Brit J Hem 128:148
International Committee for Standardization in Hematology; (1978) Recommendations for reference method for hemoglobinometry in human blood (ICSH Standard EP6/2: 1977) and specifications for international hemoglobincyanide reference preparation (ICSH Standard EP6/3: 1977) J Clin Path 31: 139-143
075D0002-01 Page C/1
APPENDIX C. BIBLIOGRAPHY
Page C/2 075D0002-01
Index
A
Autocleaning Frequency 8-9
C
Calibration 6-3, 7-2, 7-6
Change Date and Time 8-11
Change Laboratory Limits 8-4
Change Operator 8-7
Checks
Visual 2-3
Concentrated Cleaning 9-3
D
Date
Change 8-11
F
I
Flags 4-6
Print 8-5
Review 6-8
Set Limits 8-6
Troubleshooting 9-8
Identification Mode 9-7
Interfering Substances 3-8
L
Laboratory Limits
Change 4-6, 8-4, 8-6
Levy-Jennings Charts 7-18
O
Operating Conditions 2-3
Operator
Change 8-7
P
Printer
Select 8-3
Q
Quality Control 6-3
Messages 7-13
R
Reagent Pack
Connnections 2-5
Installation 2-5
Priming 2-7
Specifications 3-6
S
Select Printer 8-3
Smart Card
Calibration 7-2, 7-12
Patient Results 8-14
Quality Control 7-12
Specifications
Expected Values 3-4
Reagents 3-6
Start Up 6-1, 8-8
Startup Cycle 8-8
T
Time
Change 8-11
Troubleshooting 9-8
INDEX
075D0002-01 Index/1
INDEX
Index/2 075D0002-01