Glucocard x mini plus инструкция

Datasheet | Arkray Glucocard X-Meter Technical information | Manualzz

GLUCOCARD X-mini plus INT-IVD mmol
Operating Manual (83-04345)
page 1
FRONT
2.2 What You Need for Each Test
Chapter 10 Technical Information
10.1 Specifications
Product
Model
Test item
Sample
Sample size
Test strips
Unit of measure
Test range
Test time
Temperature
compensation
Battery
Battery life
Power consumption
Memory capacity
Clock accuracy
Operating
environment
Expected life
Dimensions
Weight
GLUCOCARDTM X-mini plus
GT-1960
Blood glucose level
Fresh capillary whole blood*1
0.3 μL
GLUCOCARDTM X premium Test Strips
mmol/L
0.6 to 33.3 mmol/L
5 seconds from blood detection
Automatic compensation using built-in thermo sensor
a. GLUCOCARDTM
X-mini plus (meter)
83-04341
Thank you for choosing the GLUCOCARDTM X-mini plus.
We have designed this compact blood glucose meter for faster and easier
testing of your blood glucose. We hope our product will help you manage your
diabetes. This manual explains how to use your new meter. Before testing,
carefully read through this manual and the package inserts that come with the
GLUCOCARDTM X premium Test Strips and GLUCOCARDTM X-METER
CONTROL. Pay particular attention to listed warnings and cautions. Please
keep this manual handy for future reference. If you have any questions, please
contact our authorized distributor(s) at the end of this manual.
3 V lithium battery (CR2032) × 2
Approx. 1,000 tests (Actual number of tests may be
less under some conditions of use.)
0.02 W (Max.)
500 test results
Within ±75 seconds/month
Temperature: 10 to 40°C (50 to 104°F)
Humidity: 20 to 80% RH (No condensation)
3 years (according to company data)
L102 × W33 × 18.5 mm (L4.0 × W1.3 × 0.7 inches)
Approx. 46 g (1.49 ounces)
Chapter 1 Introduction
*1: Although whole blood samples are used for measurement, displayed results
are equivalent to plasma glucose levels.
1.1 Intended Use
Your GLUCOCARDTM X-mini plus and GLUCOCARDTM X premium Test Strips
are for quantitatively measuring the glucose level in fresh capillary whole blood.
They are intended for use outside of the body (in vitro diagnostic use) at home
or in a clinical setting as an aid to monitor and control blood glucose levels. Do
not use them to diagnose diabetes. Also, do not alter treatment based on
the test results of this meter without instructions from your doctor or healthcare
professional.
 Store the meter at 0 to 50°C (32 to 122°F).
 The meter is shipped with the batteries loaded. So, you can start testing
immediately. However, the batteries may not satisfy the “Battery life”.
 Product specifications and appearance are subject to change without notice.
1.2 Measurement Principle
Glucose in the blood reacts with the reagent in the test strip and this produces
a small electric current. The strength of this current is proportional to the
concentration of glucose in the blood. The meter measures this current and
calculates your glucose level.
1.3 What’s Included
 GLUCOCARDTM X-mini plus
 Carrying case
 Operating Manual (this manual)
 Simplified Manual / Information for Users
 MultiletTM (lancets)*
 Multi-Lancet DeviceTM (lancing device)*
 GLUCOCARDTM X premium Test Strips*
Some kits do not include the above items with an asterisk (*).
Chapter 2 Before Tests
2.1 Important Health-Related Information
( 84-04345 )
WARNING
Do not use this meter for your blood glucose testing if you are being tested
for galactose tolerance or xylose absorption. You may use this meter if you
are taking PAM (1-Methylpyridine-6-carbaldehyde oxime).
These treatments may produce higher test results than your actual
blood glucose level. Taking unnecessary measures to lower your blood
glucose level may induce severe low glucose symptoms such as coma.
b. GLUCOCARDTM
X premium Test Strips
d. Lancet
WARNING
Keep the meter, test strips and other items away from young children. Small
items are choking hazards.
2.3 Cautions on Using the Meter
CAUTION
 For accurate test results, allow the meter to adjust to the surroundings:
Temperature: 10 to 40°C (50 to 104°F)
Humidity: 20 to 80% RH (Relative Humidity)
for 30 minutes before testing your blood glucose.
 Do not store or use the meter where:
• There are sharp temperature fluctuations.
• Humidity is high enough to cause condensation (bathrooms, drying
rooms, kitchen, etc.).
• There is a strong electromagnetic field (microwave oven, cell phone,
etc.).
 Do not use the meter after it has been dropped in liquid or liquids have
entered inside, even if dried afterwards.
 Keep hands away from the test strip slot on the meter. A thermo sensor is
housed inside the meter to minimize detection of errors.
 Do not connect the communication cable to the connection terminal
during testing. The meter may be damaged, leading to inaccurate test
results.
 Do not apply blood directly to the test strip slot on the meter.
 Do not share your meter with anyone to avoid the risk of infection.
2.4 Cautions on Using the Test Strips
CAUTION
 Use only the GLUCOCARDTM X premium Test Strips for testing with the
GLUCOCARDTM X-mini plus. Do not use other test strips as it causes
inaccurate test results.
 Do not use test strips beyond their expiration date. The expiration date is
written on the test strip bottle next to “Exp.”.
 For accurate test results, allow test strips to adjust to the surroundings:
Temperature: 10 to 40°C (50 to 104°F)
Humidity: 20 to 80% RH
for 20 minutes before testing your blood glucose.
 Do not use the test strips if 6 months or more have passed since
opening the bottle.
 The test strips are for single-use only. Do not use test strips that have
already absorbed blood or control solution.
NOTE
Store test strips in their original bottle to maintain quality. Do not transfer them
to other containers.
2.5 Meter Parts
Strap hole
Battery holder
Display
with backlight
Test strip
disposal lever
Connection
terminal
Test strip slot with light
Insert a test strip here before
testing.
REV. 2011.09
c. Lancing device
Button
GLUCOCARD X-mini plus INT-IVD mmol
Operating Manual (83-04345)
Button Operations
This manual uses the following graphic icons to
show you how to operate the button.
Left
Right
Press the button once.
Twice
page 1
REAR
Next ...
Go to chapter 3 “Blood Sampling”. Draw blood and apply it to the test strip no
more than 5 minutes after you insert the test strip into the meter. Double beeps
begin to sound 4 minutes and 30 seconds after inserting the test strip. Thirty
seconds later, the meter turns itself off.
Chapter 8 Control Test
Chapter 9 Meter Care
This meter checks itself every time you insert a test strip. You do not need to
run frequent control tests. Do a control test if:
9.1 Storing Your Meter
NOTE
In the ECO mode, the blood symbol flashes 2 minutes shorter than otherwise.
See chapter 7 “Meter Setup”.





Press the button the specified number of
times.
Slide the button to the right.
Chapter 3 Blood Sampling
Slide the button to the left.
This chapter briefly explains how to get a blood sample using the Multi-Lancet
DeviceTM and MultiletTM. If using some other lancing device to sample blood, see
the manual for that device.
Hold
Hold the button in the direction of the arrow
until the option you want appears on the
screen.
Hold for
5 seconds
Hold the button in the direction of the arrow
for the specified duration of time.
3.1 Lancing Device Parts
MultiletTM (lancet)
Multi-Lancet DeviceTM (lancing device)
Head cap
Holder
Puncture button
Adjustment dial
Protective cap
2.6 Display
Delete flag
Control test flag
HYPO flag
Special band
Band hooks
Disposal-and-set lever
Strap hole
Sound mute symbol
After-meal flag
Battery symbol
Displays the
blood symbol,
date and time,
etc.
Clock counter
Meal amount flag
Decimal point
Thermometer symbol
2.7 Inserting Test Strips into the Meter
3.2 Cautions on Using the Multi-Lancet DeviceTM
Follow these cautions to reduce the risk of infection by pathogenic
microorganisms.
CAUTION
 Wash the puncture site with soap and water. Dry the site thoroughly
before sampling blood.
 Do not share the same lancet or lancing device with anyone to avoid the
risk of infection.
 Always use a new lancet. Lancets are for single-use only. Do not reuse a
lancet that you have already used.
Start with the following steps to test your blood glucose level.
1. With clean, dry hands, remove 1 test strip from the bottle.
Do not bend the test strip.
NOTE
 Tightly close the bottle immediately after taking out a test strip.
 Do not apply blood to the test strip before inserting it into the meter.
2. Insert the test strip fully into the test strip slot.
The meter turns on. Check that a full display appears as
in chapter 2.6.
After the most recent test result appears, the
blood symbol will flash on the screen.
If nothing appears on the screen, remove the test strip and insert it back into
the test strip slot.
NOTE
 You can turn on the test strip light.
 The test result that appears right after power turns on is the most recent
result without a control test flag or delete flag.
 If the thermometer symbol appears on the screen, see “If a
Message Appears” in the “Information for Users”.
 If you insert a test strip beyond the expiration date you set, the
blood symbol and expiration symbol will alternately flash on the
screen. If this occurs, see chapter 6.3.
3.3 Selecting a Sampling Site
Alternative site testing (AST): This meter can test the glucose level of blood
from your fingertip, palm, forearm or upper arm. However, test results from
sites other than the fingertip may differ from a fingertip measurement. Consult
your doctor or healthcare professional before testing blood from the palm,
forearm or upper arm.
Use blood from:
If you are going to test:
Fingertip, palm,
forearm or
upper arm
Fingertip




Before meals
Two hours or more after meals
Two hours or more after exercise
When there is the possibility of blood glucose level
changing rapidly (e.g. after meals or exercise)
 When experiencing symptoms of hypoglycemia
such as perspiration, cold sweats, a floating
sensation or trembling
 When immediate testing is needed for suspected
hypoglycemia
 When in poor physical condition such as having a
cold, etc.
3.4 Sampling from Your Fingertip
1. Remove the head cap.
CAUTION
If you cannot test your blood glucose due to a problem with your meter or
test strips, promptly contact your healthcare professional and distributor.
2. Insert a new lancet into the holder until it stops.
3. To flag the test result:
See step 2 in chapter 5.1.
NOTE
Use only the control solution, GLUCOCARDTM X-METER CONTROL, to test
your meter and test strips
CAUTION
 Do not use control solutions beyond their expiration date. The expiration
date is written on the control solution bottles next to “Exp.”.
 Leave the meter and test strips at
10 to 40°C (50 to 104°F) and 20 to 80% RH,
and the control solution at
22 to 28°C (72 to 82°F)
for at least 30 minutes before running the control test.
 Do not drink control solution. It is not for human consumption.
 Keep control solution away from skin and eyes. Contact may cause
inflammation.
1. See steps 1 and 2 in chapter 2.7.
Before-meal flag
After-exercise flag
You suspect the meter or test strips are not working properly.
You dropped the meter.
You damaged the meter.
Your test results do not match how you feel.
You want to check performance of the meter and test strips before a blood
glucose test.
NOTE
Use a lancet before its expiration date.
3. Twist off the protective cap on the lancet. Keep the
protective cap for disposing of the lancet after the test.
2. Select the control test flag.
As an alternative, you can attach the control test flag after the
test result appears on the screen. Test results with the control test flag are not
part of your result averages.
3. Make sure that the blood symbol is flashing on the screen.
4. Squeeze a drop of control solution to the tip of the test strip.
After use, tightly close the caps of the test strip bottle and control solution bottle
to maintain their quality.
Store your meter, test strips, control solutions and manuals in your carrying
case and store it in a dry location. The correct storage temperature is 0 to 50°C
(32 to 122°F) for the meter and 1 to 30°C (34 to 86°F) for the test strips and
control solutions. Do not freeze. Avoid heat, humidity and direct sunlight.
CAUTION
To obtain accurate test results:
 Do not use test strips or control solutions if their bottles are broken or
have been left open.
 Do not use the test strips or control solutions beyond their expiration
dates.
9.2 Cleaning Your Meter and Lancing Device
Your meter and lancing device do not need special cleaning. If your meter and
lancing device get dirty, wipe them with a soft piece of cloth moistened with
water. To disinfect these items after cleaning, wipe with a soft cloth moistened
with 70% ethanol, 70% isopropanol or diluted household bleach (0.05% sodium
hypochlorite solution).
9.3 Changing the Batteries
When the battery symbol appears on the screen, the batteries
are getting low. Before using your meter, change the batteries.
Past results remain in the memory even when the batteries are
changed.
NOTE
Your meter uses two CR2032 3V lithium batteries. This type of batteries is
available in many stores. Keep spare batteries handy at all times.
The test will not start if you apply your control solution to
the check window directly.
You do not need to set the date and time if you insert new batteries within 1
minute after removing the old batteries. Even if you change the batteries within
1 minute, your meter’s clock may stop if:
The test starts when the meter detects the control
solution. The meter counts down from 5 to 1 during the
test.
 The batteries are not replaced soon after the battery symbol appears.
 The batteries are inserted upside down.
 You touch the metal parts inside the meter with hands or metal.
NOTE
 Tightly close the control solution bottle.
 Do not touch the test strip once the meter starts counting down.
1. Make sure the meter is off.
5. Check your test result is within the acceptable range.
The acceptable range is shown on the package insert that come with the
control solution. If within range, the meter and the test strip are working
properly. If out of range, test the control solution again.
CAUTION
If you still have test results that are out of the acceptable range, stop using
your meter, test strips and control solutions, and contact your distributor.
6. Slide the test strip disposal lever to eject the test strip.
The meter turns itself off.
2. Grip the strap hole with the tip of your finger and pull
open.
3. Remove the batteries.
4. Set the new batteries into the battery holder with the
“+” side facing upward.
NOTE
If the batteries are inserted upside down, the meter will
not operate.
5. Close the battery holder.
6. The “SET” screen appears when the time has been reset. If it appears, set
the correct time. See the setup mode flowchart in chapter 7. If you do not set
the time, all subsequent test results will be stored as 12:00 AM or 00:00.
Dispose of old batteries according to your local environmental
regulations.
GLUCOCARD X-mini plus INT-IVD mmol
Operating Manual (83-04345)
page 2
FRONT
4. Attach the head cap to the lancing device body.
Chapter 7 Meter Setup
When you hear a beep, remove the test strip from the blood. The test starts
when the meter detects enough amount of blood. The meter counts down from
5 to 1 during the test.
Your meter is set up as below at the time of purchase. To change the settings, see the setup mode flowchart.
 Alarm: Off
 Result average display: 7d
 ECO: Off
 Beeper volume: High (3)
 Meal amount: Off
 Test strip expiration: Off
 Date: Current local date
 Bar graph: Upper limit 16.7 mmol/L, Lower limit 2.8 mmol/L
 Time: Current local time
 HYPO: Off
6. Wash the puncture site with soap and water. Dry
thoroughly.
Setup mode
To go to “SET”:
to follow the black
arrows (
).
Alarm
1. Make sure
the meter is
off.
Current setting
2.
You can make 4
alarm settings.
Hold for
5 seconds
to follow the gray
arrows (
).
Beeper volume
Date
Time
Current setting
Current setting
Current setting
1. Select a
setting.
1. Set the year.
1. Select a time
format.
1. Select an
alarm number.
or
A1, A2, A3 or A4
2. Turn on the
alarm.
to change the settings or
numbers.
or
Hold
to scroll through the
settings or numbers:
on1: Low
on2: Normal
on3: High
oFF: Off
To the top of the
beeper volume
setting.
12h or 24h
2. Set the month.
3. Set the day.
2. Set the hour.
3. Set the minute.
4. Select a date
format.
To the top of the
time setting.
4. Set the minute.
To make another alarm
setting, repeat steps 1 to 4.
To the top of the
date setting.
If you accidentally press the puncture button before
reaching the puncture site, press the disposal-and-set
lever until you hear a click. The device is then reset.
HYPO
(hypoglycemic)
Bar graph
Meal amount
Result average
display
Current setting
Current setting
Current setting
Current setting
Current setting
Current setting
1. Turn on the
expiration
warning.
1. Turn on the
ECO mode.
1. Turn on the
HYPO flag.
1. Set the upper
limit of the
graph.
1. Turn on the
meal amount
flag.
1. Select a
setting.
You will hear a beep when the test result appears on the
screen. The meter stores this test result and the test date
in the memory.
8. Obtain as much blood as that shown in “Actual size”
on the right.
Actual
size:
Next ...
Go to chapter 4 “Testing Your Blood Glucose”.
3.5 Sampling from Your Palm, Forearm or Upper Arm
2. Select a soft, fleshy area on the palm, forearm or upper arm that is free of
visible veins, moles and hair and away from bone.
3. Massage the puncture site.
The date 6 months
from the current
date appears.
To the top of the
ECO mode
setting.
OK
To change
the date:
See steps 1 to
3 in “Date.”
To the top of the
expiration
setting.
7d, 14d, 30d or
90d
1.1 to 33.3 mmol/L
2. Set the
threshold value.
2. Set the lower
limit of the
graph.
0.6 to 4.4 mmol/L
A HYPO flag
appears when
your test results
are below the
threshold value.
To the top of the
HYPO setting.
Bar graph
Test results
The 6 most recent test results
appear as a bar graph, with the
newest result at the far right.
Upper limit
Lower limit
To switch the bar graph to the
current date and time:
If you obtained an insufficient amount of blood, rub or
warm a new puncture site to increase blood flow.
Then, make another puncture.
0.6 to 32.8
mmol/L
To the top of the
bar graph
setting.
To the top of the
meal amount
setting.
 If test results do not match how you feel:
Make sure you performed the test properly as explained in chapters 2 to
4. Then, do a control test to check there are no problems with the meter
or test strips. If you tested blood from the palm, forearm or upper arm,
repeat the test with blood sample from a fingertip. If test results still do
not match how you feel, contact your doctor or healthcare professional.
Actual size:
Next ...
Go to chapter 4 “Testing Your Blood Glucose”.
Chapter 4 Testing Your Blood Glucose
1. Make sure that the blood symbol is flashing on the screen.
To the top of the
result average
display setting.
2. Touch the tip of the test strip to the drop of blood.
Let the test strip draw up blood until the dark-colored
check window is filled with blood.
WARNING
Test results are shown in mmol/L. You cannot change the unit of measure. If
the results appear in mg/dL by any possibility, contact your distributor
immediately. Using mg/dL may cause you to misread test results and lead
to incorrect treatment.
CAUTION
 If “Lo” or “Hi” appears on the screen:
Repeat the test. If “Lo” or “Hi” still appears, contact your doctor or
healthcare professional. “Lo” appears if your test result is less than 0.6
mmol/L. “Hi” appears if your test result is more than 33.3 mmol/L.
If nothing appears on the screen, remove the test strip and insert it back into
the slot. Then, wait until the blood symbol starts flashing.
The ECO mode
extends battery
life.
NOTE
 For accurate test results, touch the test strip to your blood in less than 20
seconds after puncturing.
 Do not test blood that runs or spreads out from the puncture site.
 Do not smear blood on the test strip.
 Do not press the test strip into your puncture site with force.
 If the meter does not start counting down, touch the test strip to the drop of
blood again within 10 seconds. Use a new test strip if more than 10 seconds
have passed.
 Keep your hands away from the test strip once the meter starts counting
down.
3. Read your test result.
6. Hold the head cap against the site. Add and reduce
pressure to the site repeatedly, one cycle per second,
to obtain as much blood as that shown in “Actual size”
on the right.
Back to “Alarm”
NOTE
The meter turns
itself off if left
alone for 3
minutes (ECO
mode: 1 minute).
Your new settings
will be saved.
7. Press the head cap against your fingertip and press
the puncture button lightly.
If you accidentally press the puncture button before
reaching the puncture site, press the disposal-and-set
lever until you hear a click. The device is then reset.
To the top of the alarm setting.
The meter turns
off.
Deeper
5. Press the head cap hard against the puncture site for
about 10 seconds. Press the puncture button lightly.
5. Go to “End”.
ECO
1 2 3 4 5 6 7
4. Wash the puncture site with soap and water. Dry
thoroughly.
d-n: Day-month
n-d: Month-day
Test strip
expiration
Level indications
1. See steps 1 to 5 in chapter 3.4.
3. Set the hour.
Setup end
5. If needed, adjust the lancing depth by turning the
adjustment dial.
 Do not ignore test results. Do not alter your blood glucose management
or treatment without consulting your doctor or healthcare professional. It
is important to follow their instructions.
NOTE
In the ECO mode, the test result appears on the screen 2 minutes shorter than
otherwise. See chapter 7 “Meter Setup”.
4. Slide the test strip disposal lever to eject the test strip.
The meter turns itself off.
CAUTION
Point your meter down and away from others when ejecting the used test
strip.
5. Remove the used lancet from your lancing device.
Accurate test results will not be obtained if you apply
your blood to the check window directly.
Remove the head cap from the lancing device body.
Stab the protective cap with the tip of the lancet.
Push the disposal-and-set lever in the direction of the
arrow to remove the lancet.
Disposal of Biohazardous Waste
Test strips and lancets qualify as biohazardous waste once used to test blood
glucose. After use, dispose of them according to your local regulations on
biohazardous waste.
GLUCOCARD X-mini plus INT-IVD mmol
Operating Manual (83-04345)
page 2
REAR
3. View the test result you want.
Chapter 5 Managing Your Test Results
Chapter 6 Useful Functions for Your Test
To go to the next result:
5.1 Flagging Test Results
Flags help you categorize and identify results.
1. Leave the test strip in the meter after the test result appears on the screen.
To go back to the previous
result:
2. Select the flag you want.
To scroll through results:
1. Repeat steps 1 to 3 in “Before Your Meal: Starting the Clock Counter”.
6.1 Using the Alarm
The alarm reminds you when to test your blood glucose. You can set the alarm
to sound at 4 different times in a day.
See chapter 7 “Meter Setup.”
Hold
To stop the alarm:
6.3 Managing the Expiration Date of Test Strips
You can set the expiration date of test strips. The expiration warning then
appears if you insert a test strip into the meter after the set date.
5.3 Viewing Result Averages
You can view result averages for the set interval (7, 14, 30 or 90 days).
Before-meal flag
For results obtained before eating
Setting the Expiration Date of Your Test Strips
6.2 Using the Clock Counter
1. See steps 1 and 2 in chapter 5.2.
You can store after-meal results together with the time elapsed since the
beginning of your last meal (up to 180 minutes).
2. View the averages of your
glucose test results.
After-meal flag
For results obtained after eating
mmol/L
3. To view the averages with a
flag:
Checking the Set Expiration Date
1. Make sure the meter is off.
mmol/L
2. Go to the “DATE:” screen.
2. View the expiration date.
The number of tests
included in the calculation
of the result average.
Meal amount flag
When the meal amount setting is on, the meal
amount flag appears below the after-meal flag.
50%
Select the amount of food you ate from the 3
levels shown on the left.
3. From the “DATE:” screen, go
to the “Meal Time” screen.
4. To turn off the meter:
Hold for
2 seconds
4. Turn on the clock counter.
5.4 Deleting All Test Results from the Memory
You can delete all test results from the memory. Remember that results cannot
be retrieved once deleted.
After-exercise flag
For results obtained after exercise
1. See steps 1 and 2 in chapter 5.2.
3. Slide the test strip disposal lever to eject the test strip. The meter turns itself
off.
2. Go to the “AVG (average)”
screen.
3. Go to the “ALL CLEAR”
screen.
5.2 Reviewing Past Results
You can review past results stored in the memory. Your meter stores up to 500
test results. New results beyond the 500th overwrite previously stored results in
the order of oldest first.
2. Turn on the meter.
The most recent test result
appears on the screen.
Hold for
2 seconds
The meter then turns itself off.
mmol/L
The 3-digit number following “M:” is the order in which test results are stored,
with the most recent result first. The screen also shows the date and time
when the test was done.
If you forgot to start the clock
counter, select the time
elapsed since the beginning
of your last meal in
15-minute blocks.
After the current date and
time appear, the meter turns
itself off.
To view the elapsed time
while the meter is off:
4. Select “YES”.
5. Delete the test results from
the memory.
1. Make sure the meter is off.
5. When you are about to start your meal, go to step 6 with “0 min” flashing.
6. Start the clock counter.
Hold for
5 seconds
Hold for
2 seconds
Testing after Your Meal: Changing the Elapsed Time
1. After eating, test your blood glucose.
When the clock counter setting is on, the elapsed time
and after-meal flag appear on the right of the test result.
2. To change (or set) the
elapsed time:
Select the time.
The expiration date appears
after the current date and
time.
Expiration Date Reminder
80%
Full
Select the earlier date of the following:
• Six months from opening the test strip bottle
• Date next to “Exp.” on the test strip bottle
To set the expiration date, see chapter 7 “Meter Setup”.
Before Your Meal: Starting the Clock Counter
1. Make sure the meter is off.
Clock counter
If the clock counter setting is on, the elapsed
time appears on the right of the after-meal
flag. See chapter 6.2.
3. Confirm your selection.
A series of short beeps sound for 1 minute.
Twice
Control test flag
Control test results. Not part of your result averages.
Delete flag
For inaccurate results. Not part of your result averages. The
delete flag can be attached only after testing.
2. Select “no”.
Setting the Alarm
When the Alarm Reaches the Set Time
4. To turn off the meter:
Stopping the Clock Counter
Hold for
2 seconds
or
Confirm your setting.
3. Slide the test strip disposal lever to eject the test strip.
The meter turns itself off and the test result is stored together with the
elapsed time.
Do not use the inserted test strip and other test
strips in the same bottle, if the blood symbol and
expiration symbol alternately flash on the screen.

Catalog excerpts

GLUCOCARD X-mini plus - 1

Advanced Blood Glucose Test Meter GLUCOCARD™X-mini plus More advanced SMBG meter which successfully benefits the user with no coding,

Open the catalog to page 1

GLUCOCARD X-mini plus - 2

Advanced Blood Glucose Test Meter GLUCOCARD™X-mini plus Impressive & Stylish design Comprehensive display New user friendly functions Small Sample Maximum flexibility Short fill detection Strip discard button Test strip port light External output Test strip for GLUCOCARD X-mini plus Sample Volume Measuring Time Power source Glucose in whole blood Test Strip_GLUCOCARD X-SENSOR_ Operating Environment Humidity 20 to 80% RH (Without condensation) Design and specifications may be changed without prior notice arKray sLObai business.inc. Kyoto Miyuki Bldg. 10F, 689 Takanna-cho, ^P^K Nakagyo-ku,…

Open the catalog to page 2

  1. Manuals
  2. Brands
  3. Arkray Manuals
  4. Blood Glucose Meter
  5. GLUCOCARD 01-mini
  6. User instruction manual
  • Contents

  • Table of Contents

  • Troubleshooting

  • Bookmarks

Quick Links

loading

Related Manuals for Arkray GLUCOCARD 01-mini

Summary of Contents for Arkray GLUCOCARD 01-mini

  • Page 1
    User Instruction Manual…
  • Page 2
    CAUTION: Before using any product to test your blood sugar (blood glucose), read all instructions and practice the test. Do all quality control checks as directed and consult with a diabetes healthcare professional. These recommendations apply to all blood glucose monitoring systems and are supported by the American Association of Diabetes Educators, the American Diabetes Association, the U.S.
  • Page 3: Table Of Contents

    TABLE OF CONTENTS Introduction ……… . . 5 Intended Use .

  • Page 4
    Blood Glucose Testing ……..35 The Lancing Device .
  • Page 5: Introduction

    INTRODUCTION Thank you for choosing the GLUCOCARD™ 01-mini Blood Glucose Monitoring System. This system measures blood glucose in fresh capillary whole blood. Results are shown as plasma values. This makes it easy to compare the GLUCOCARD™ 01-mini Blood Glucose Meter and lab results.

  • Page 6: Intended Use

    Intended Use The GLUCOCARD™ 01-mini Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips or palm. Testing is done outside the body (In Vitro diagnostic use). It is indicated for use at home (over the counter [OTC]) by persons with diabetes, or in clinical settings by healthcare professionals, as an aid to monitor the effectiveness of diabetes control.

  • Page 7: Understanding The Glucocard ™ 01-Mini Blood Glucose Meter

    UNDERSTANDING THE GLUCOCARD 01-mini BLOOD GLUCOSE METER ™ Bottom Button(1) Button(2) Battery Cover Display Strap Hole Test Strip Port…

  • Page 8: Explanation Of The Full Display Screen

    Explanation of the full display screen: Memory symbol: Appears when you are Shows you a past result setting the date or time or reviewing in the memory. a past result. Sound mute symbol Delete flag Control test flag Battery symbol Pre-meal flag Thermometer symbol Post-meal flag…

  • Page 9
    GLUCOCARD™ 01-SENSOR These test strips need only 0.3 µL (microliter) of blood. Apply blood to the sample application tip. The strip draws in the blood by capillary action. The blood glucose reacts with the enzyme on the test strip. This produces a current proportional to the blood glucose level.
  • Page 10: Changing The Battery

    Changing the Battery -mini blood glucose The GLUCOCARD™ 01 meter comes with one lithium battery CR2032 (or DL2032). CAUTION Replacing the battery will not delete meter settings or test results stored in the meter unless you: • Touch the metal portion on main unit with your fingers or any metal. •…

  • Page 11
    To replace battery: 1. Turn meter off. 2. Remove the battery cover on the back of the meter. 3. Slide the battery to the right (a) and remove it (b). CAUTION: Do not touch the metal contacts inside the battery compartment with your finger or any metal.
  • Page 12
    5. Replace battery cover. If the clock setting has been deleted, SET will appear on the screen. If SET appears, set the time and date. If you do not set the time and date, all subsequent results will be stored with a test date and time of January 1, 2008, 12:00 a.m.
  • Page 13: Changing Colored Faceplates

    Changing the Colored Faceplates The GLUCOCARD™ 01-mini comes with multiple interchangeable colored faceplates. You can attach the color of your choice. Display cover 1. Pressing on the dotted area shown on the right with your thumb, slide the cover in the direction of the arrow.

  • Page 14
    4. Hook the end of the display cover into the hole Color of your preferred color faceplate. faceplate 5. Fit one of the tabs on the display cover into the corresponding slot on your meter. 6. Press forcefully on the dotted area shown on the right to snap the display cover back into your meter.
  • Page 15: Meter Set-Up

    METER SET-UP Meter set-up allows you to adjust the audible beep volume and to set time, date, and test averaging. Before using the GLUCOCARD™ 01-mini blood glucose meter for the first time, set the time and date. Do not perform a blood glucose test until you have set-up the meter.

  • Page 16
    Setup mode Date and time Test average display Audible Beep See page 23. See page 18. See page 20. To go to SET: To turn off the 1. Make sure meter, press To change the the meter To change the button.
  • Page 17: Basic Rules

    Basic rules for meter set-up: Step 1: Turn off meter. Step 2: buttons at the Press the same time for 5 seconds. A beep sounds and the meter turns on. Next, a second beep sounds and “SET” shows on the screen. Step 3: Use the button to change the set-up screen.

  • Page 18
    Setting the Audible Beep If the audible beep is turned “on”, the beep will sound when: • You insert a test strip into the meter. • You apply blood or control solution to a test strip. • The test has ended and the result is displayed. •…
  • Page 19
    Press the button to cycle through the volume levels: “OFF” (no sound) “on 1” (low) “on 2”(middle) “on 3” (high) “OFF” (no sound) …, etc. A beep sounds at the set volume at each step. If “OFF” is set, the “audible beep mute (Off)”…
  • Page 20: Year

    Setting the Year In the year set-up screen, the last two digits of the currently set year appear at the top of the screen. “ y Er” appears in the middle of the screen. The last two digits of the currently set year appear at the (audible beep) bottom of the screen and blink.

  • Page 21: Month/Day

    Setting the Month Press the button to change the month. After you set the month, press the button to confirm the setting. The screen changes to the day set-up screen. Setting the Day Press the button to change the day. After you set the day, press the button to confirm the setting.

  • Page 22: Hour/Minute

    Setting the Hour Press the button to change the hour. After you set the hour, press the button to confirm the setting. The screen changes to the minute set-up screen. Setting the Minute Press the button to change the minute. After you set the minute, press the button to confirm the setting.

  • Page 23: Test Average Display

    Setting Test Average Display NOTE: “Lo” results are included as 20 mg/dL in the averages. “Hi” results are included as 600 mg/dL in the averages. Test results with the following symbol or flag are not part of the result average: temperature error symbol, control solution flag, or delete flag.

  • Page 24: Ending Set-Up

    Ending Set-up To make further changes to set-up settings, press button while “End” is displayed. This will bring up the Audible Beep set-up screen. You can then move through the set-up mode and adjust any settings that you wish. If you are done with set-up, press the button while “End”…

  • Page 25: Perform A Control Solution Test

    • the meter and test strips are working correctly as a system. • you are testing correctly. Call your local supplier or pharmacy to order control solution. If they do not stock it, call ARKRAY Order Services at 800.818.8877 option 4 (USA and Canada only).

  • Page 26
    Perform a Control Solution Test: • Before testing with the GLUCOCARD™ 01-mini blood glucose meter for the first time. • When you open a new bottle of test strips. • Whenever you suspect the meter or test strips may not be functioning properly.
  • Page 27
    • GLUCOCARD™ 01 CONTROL is not intended for human consumption. Do not drink. • Do not apply to eye. Contact your healthcare professional immediately if solution is swallowed, injected, or applied to the eye. • Dye in control solution may stain skin, clothing, or surfaces. •…
  • Page 28
    Step 1: Insert a test strip as shown in diagram. Test strip should seat firmly into meter. The meter will turn on automatically. The full screen will come on for a moment. Make sure all symbols appear on the screen (see page 8).
  • Page 29
    Step 2: Press the button to enter the control mode. You will see “CONT” on the upper left of the screen. A “CONT” flag can also be attached when results are shown on screen.
  • Page 30
    Step 3: Remove the cap from the control solution bottle. While the Apply Blood symbol and Code No. (F-2, F-3 or F-4) are alternating, place a very small drop on the end of the test strip. The test strip will draw up the solution and the meter will beep.
  • Page 31
    Repeat Steps 1-5 with other Control Solution levels as per recommendations of your healthcare professional. Call your local supplier or pharmacy to order control solution. If they do not stock it, call ARKRAY Order Services at 800.818.8877 option 4 (USA and Canada only). NOTE:…
  • Page 32: Troubleshooting Control Solution Test

    Troubleshooting Control Solution Test If the control solution test result is out-of-range, check the following and repeat the test: Check Action Did you do the test in If not, do the test again. control solution mode? Insert a test strip; press the button Do you see “CONT”…

  • Page 33
    Check Action Did you insert test strip Make sure test strip is inserted firmly into meter? all the way into test strip port. See diagram on page 28. Did you follow the Reread “Performing a Control procedure correctly? Solution Test” (pages 25-31) and retest. Were test strips stored correctly If not, retest with a (between 34-86°F (1-30°C))?
  • Page 34
    Does your Control Solution result still read outside the range printed on the test strip bottle? If yes, the system may not be working correctly. DO NOT USE the system to test your blood glucose until the control solution result is within range. If you need help, please call Customer Service at 800.566.8558 (24 hours a day, 7 days a week, USA and Canada).
  • Page 35: Blood Glucose Testing

    BLOOD GLUCOSE TESTING The Multi-Lancet Device ™ Removable cap Lancing Button Adjustment dial Multilet (lancet) Multi-Lancet Device body ™ Band hooks Protective cap Disposal-and-set lever Step 1: Pull the removable Strap hole Special band cap to Groove of special band remove it.

  • Page 36
    Step 2: Insert a new, unused lancet into the Holder Holder until it stops. Step 3: Twist off the protective cap. Step 4: Attach the removable cap.
  • Page 37
    Step 5: The Multi-Lancet Device offers 7 different levels of skin ™ penetration. For soft or thin skin or first time use, select “1 or 2.” For alternate sites, select “7.” If the obtained amount of blood is excessive, turn the dial to a lower setting. If the obtained amount of blood is insufficient, turn the dial to a larger setting.
  • Page 38: Getting A Drop Of Blood

    Getting a Drop of Blood Step 1: Wash hands with soap and warm water. Dry hands thoroughly. If you use alcohol wipes to cleanse fingers, make sure the area is dry before lancing finger. Residual alcohol may lead to inaccurate readings. Step 2: Let the arm hang down for 10 to 15 seconds.

  • Page 39
    Step 5: Set aside the Multi-Lancet Device and wait a few seconds for ™ a blood drop to form. Keeping the hand warm, lowering the hand to waist level and gently massaging your wrist, palm or the base of the finger helps the flow of blood.
  • Page 40: Performing A Blood Glucose Test

    Performing a Blood Glucose Test NOTE: Four and a half minutes after test strip insertion, a beep sounds every 5 seconds. After 5 minutes, a beep sounds and the meter turns off. Take out unused test strip and insert it again to turn meter back on.

  • Page 41
    Step 2: The last test result will then appear. The Apply Blood symbol and the code number (F-2, F-3, or F-4) will alternate on the screen. Verify that the code number on the screen matches the code number printed on the bottle. If the code number does not match, the meter may give a false reading.
  • Page 42
    Step 3: Obtain a blood drop Apply (see pages 38-39). blood to the end of the test strip. Capillary action will pull the blood into the strip. The meter will beep when it detects that blood has been applied. The meter will begin to count down.
  • Page 43
    If it takes longer than 20 seconds to obtain a blood sample and to touch the test strip to the blood drop, repeat Step 3. Step 4: The meter will count down, starting from “7”. After 7 seconds, a beep sounds and the test result and date/time are shown.
  • Page 44: Flagging Test Results

    Flagging Test Results Flags help you categorize and identify results. They can be used to omit specific results from result average calculations. You can attach any of the below flags to test results. These flags are then stored in the memory with the test results.

  • Page 45
    Step 2: When you have selected the desired marker, remove the test strip. When you recall the result from memory, the selected marker will be displayed.
  • Page 46: Alternate Site Testing (Ast)

    ALTERNATE SITE TESTING (AST) Important Information About AST: Sites other than your fingertip may have fewer nerve endings so obtaining a blood sample from these sites may be less painful. The technique for alternate site testing is different from fingertip testing.

  • Page 47
    Use Fingertip Testing: • Within two hours after a meal. • Within two hours after insulin dosing. • Within two hours after physical activity. • If you have a history of hypoglycemia, are experiencing low blood glucose, or suffer from hypoglycemic unawareness (you cannot tell when you have low blood glucose).
  • Page 48
    Preparing to Test Your Blood Glucose From an Alternate Site Choose a Site: Select a soft, fleshy area on the palm that is free of visible veins and moles and away from bone. Wash the area with warm, soapy water. Rinse and dry completely.
  • Page 49
    Prepare Your Lancing Device: To order an AST Lancing Kit containing the lancing device, lancets and instructions, please call 800.566.8558. Performing a Blood Glucose Test From an Alternate Site Step 1: Insert lancet and cock lancing device.
  • Page 50
    Step 2: Remove test strip from the bottle. Immediately Insert test strip as replace the bottle cap tightly. shown in diagram. A beep will sound and the meter will turn on automatically. Verify that all symbols appear on the screen (see page 8). Verify that the code number displayed on the screen matches the code number on the bottle of test strips used.
  • Page 51
    Step 3: a) PRESS AND VIGOROUSLY RUB THE SELECTED AREA for 10 seconds until it starts to feel warm to the touch. b) Firmly hold the cocked lancing device against the clean skin for 5-10 seconds.
  • Page 52
    c) Press the release button d) Once a on the lancing device to large enough Continue to lance the skin. drop of blood hold the lancing device has formed, firmly against the skin until remove the a blood drop forms lancing device.
  • Page 53
    IMPORTANT IF NOT, • Was the sample free of clear or watery fluid? REPEAT TEST. • Was the drop large enough (at least 0.3 µL)? • Did the reaction site fill with enough blood all at once? Step 5: The meter will count down and display result in 7 seconds. CONSIDER YOUR RESULT.
  • Page 54
    If the repeated alternate site result is still not consistent with how you feel, confirm your blood glucose level with fingertip testing. Step 6: record your results in your self-testing logbook Always along with other information such as insulin dosage, diet, and exercise. The result is automatically stored in memory with the time and date.
  • Page 55: Understanding Test Results

    UNDERSTANDING TEST RESULTS The GLUCOCARD™ 01-mini blood glucose meter is referenced to a fresh plasma sample. Your meter results may differ from clinical lab results. This is due to normal variation. To compare your meter with lab results, follow the guidelines below. Before you compare Perform a control solution test.

  • Page 56
    ALWAYS wash hands with warm soapy water. Rinse and dry them ALWAYS before testing. use fresh capillary whole blood on meter tests. You may still see a difference from the lab result. This can be due to several causes. Blood glucose levels can change over short periods of time, for example due to food or stress.
  • Page 57
    Normal Blood Glucose Readings Expected blood glucose levels for adults without diabetes¹ , ²: Fasting 70 to 110 mg/dL 1 to 2 hours after meals <120 mg/dL High Blood Glucose Readings If your blood glucose is above 600 mg/dL, you will receive a “Hi.” Repeat the test with a new test strip.
  • Page 58: Diabetes Findings

    Low Blood Glucose Readings If your blood glucose is below 20 mg/dL, you will receive a “Lo.” Repeat the test with a new test strip. If this message shows again, contact your healthcare professional immediately! Contact your physician for advice if test results are very low (below 70 mg/dL) and/or you have symptoms of low blood glucose.

  • Page 59: What This Means For You

    What This Means For You Frequent blood glucose testing is the best means to track how well you are doing with your diabetes management. It helps you track the effects of medications, diet, exercise, and stress management. Blood glucose test results can also tell you if your diabetes is changing. This may alert you to adjust your treatment plan.

  • Page 60: Recalling Results — Entering Memory Mode

    RECALLING RESULTS — ENTERING MEMORY MODE The GLUCOCARD™ 01-mini blood glucose meter stores up to 50 test results. When more than 50 test results have been performed, the meter drops the oldest result each time you add a new result. Before using the memory, remove test strip from meter and make sure meter is turned off.

  • Page 61
    • If Test Averaging has been turned on, the average result will appear first. • If Test Averaging has been set to “Off”, the display will briefly flash the number “1.” Then, the most recent test result appears with the date and time of the test.
  • Page 62
    Step 3: After all test results have been reviewed, “End” appears. • Press the button once to go back to the first result in memory. You can then scroll through results in memory again. • Press the button for 2 seconds to exit Memory and turn off the meter.
  • Page 63
    Step 2: Hold down the button for 2 seconds to display the most recent test result. If you have turned the result average display on, the result average will appear instead of the most recent result. Step 3: Press the button to go to the “End”…
  • Page 64: Maintenance

    MAINTENANCE Cleaning To prevent malfunction of the meter, keep the test strip port free of blood, moisture, dirt, or dust. Use a lint-free cloth dampened with water to clean meter. Thoroughly wring out cloth before use. Do not use an abrasive cloth or antiseptic solution, as these may damage the display screen.

  • Page 65: Storage And Handling

    Storage and Handling • Store meter at 32°F-122°F (0°C-50°C). • Do not leave the meter in very hot or cold places. Do not leave it near a heat source (radiator) or in a car in hot or cold weather. • Do not store or use meter or test strips: •…

  • Page 66: Display Screen And Error Messages

    Service at 800.566.8558 (24 hours a day, 7 days a week, USA and Canada only). Contact your healthcare professional if you cannot reach Customer Service. Do not send your meter to ARKRAY or your local dealer without first calling to get approval to do so.

  • Page 67
    Your blood glucose level is less than 20 mg/dL or there was not enough blood on test strip. Repeat test with new test strip. Make sure you use enough blood to fill the strip. If this message shows again, contact your healthcare professional immediately! Your blood glucose level is more than 600 mg/ dL.
  • Page 68
    The battery power is too low for testing. Test results and changes are not stored in the memory. Change the battery. See “Changing the battery” page 10. The meter is faulty. Contact Customer Service at 800.566.8558. The meter temperature is outside acceptable range of 50˚F-104˚F (10˚C-40˚C).
  • Page 69
    Test strip problem. Contact bars on test strip may be dirty. Incorrect test strip used. Used test strip has been inserted. Repeat test with new test strip. Additional blood was applied to test strip after test started. Test strip was moved during test. Insufficient blood sample.
  • Page 70
    Apply Control Solution. The Apply Blood symbol flashes alternately with the strip code number (F-2, F-3, or F-4) and “CONT” appears at the top of the screen. This shows the meter is ready for you to apply control solution. Test Result. The meter shows results between 20 and 600 mg/dL.
  • Page 71: Product Information

    PRODUCT INFORMATION Product Specifications Test Strips: GLUCOCARD™ 01-SENSOR. Result Range: 20-600 mg/dL . Calibration: Plasma referenced. Sample Size: Minimum 0.3 µL. Blood Source: Fresh capillary whole blood. Hematocrit Range: 30% — 54%. Test Time: 7 seconds. Assay Method: Biosensor, Glucose Oxidase (Aspergillus niger sourced). Power Source: One 3-volt lithium battery (CR2032 or DL2032) Battery Life:…

  • Page 72: Warnings, Precautions & Limitations

    Warnings, Precautions & Limitations • You may get test results higher than your actual glucose level if taking PAM (1-Methylpyridine-6-carbaldehyde oxime). Taking unnecessary measures to lower your blood glucose level may induce severe low blood glucose symptoms such as coma. •…

  • Page 73
    • Extremes in hematocrit may affect test results. Hematocrit levels less than 30% may cause falsely high readings. Hematocrit levels greater than 54% may cause falsely low readings. • Inaccurate results may occur in severely hypotensive individuals or patients in shock. Inaccurate low results may occur for individuals experiencing a hyperglycemic-hyperosmolar state, with or without ketosis.
  • Page 74
    • The GLUCOCARD™ 01-mini Blood Glucose Monitoring System is not designed to be a substitute for pathology laboratory equipment and should not be used for the diagnosis of diabetes. • Use only fresh capillary blood. Do not use serum or plasma or venous whole blood.
  • Page 75: Warranty

    Warranty You are a valued customer of ARKRAY. It is important to us that you are completely satisfied with your GLUCOCARD™ 01-mini blood glucose meter. ARKRAY warrants your GLUCOCARD™ 01-mini blood glucose meter will be free from defects in materials and workmanship for a period of five years from the date of the original purchase.

  • Page 76
    Returns You must contact ARKRAY Customer Service at 800.566.8558 before returning your meter. You will be instructed how to return the meter to ARKRAY. Returned meters without this authorization will not be accepted. TEL 800.566.8558 FAX 952.646.3110 www.arkrayusa.com P/N 7344-00 ©2008 ARKRAY, Inc.

GLUCOCARD X-mini plus INT-IVD mmol Operating Manual (83-04345) page 1 FRONT

10.1 Specifications

*

1

: Although whole blood samples are used for measurement, displayed results

are equivalent to plasma glucose levels.

Store the meter at 0 to 50°C (32 to 122°F).

The meter is shipped with the batteries loaded. So, you can start testing

immediately. However, the batteries may not satisfy the “Battery life”.

Product specifications and appearance are subject to change without notice.

Chapter 10 Technical Information

Product

Model

Test item

Sample

Sample size

Test strips

Unit of measure

Test range

Test time

Temperature

compensation

Battery

Battery life

Power consumption

Memory capacity

Clock accuracy

Operating

environment

Expected life

Dimensions

Weight

GLUCOCARD

TM

X-mini plus

GT-1960

Blood glucose level

Fresh capillary whole blood*

1

0.3 μL

GLUCOCARD

TM

X premium Test Strips

mmol/L

0.6 to 33.3 mmol/L

5 seconds from blood detection

Automatic compensation using built-in thermo sensor

3 V lithium battery (CR2032) × 2

Approx. 1,000 tests (Actual number of tests may be

less under some conditions of use.)

0.02 W (Max.)

500 test results

Within ±75 seconds/month

Temperature: 10 to 40°C (50 to 104°F)

Humidity: 20 to 80% RH (No condensation)

3 years (according to company data)

L102 × W33 × 18.5 mm (L4.0 × W1.3 × 0.7 inches)

Approx. 46 g (1.49 ounces)

REV. 2011.09

( 84-04345 )

Thank you for choosing the GLUCOCARD

TM

X-mini plus.

We have designed this compact blood glucose meter for faster and easier

testing of your blood glucose. We hope our product will help you manage your

diabetes. This manual explains how to use your new meter. Before testing,

carefully read through this manual and the package inserts that come with the

GLUCOCARD

TM

X premium Test Strips and GLUCOCARD

TM

X-METER

CONTROL. Pay particular attention to listed warnings and cautions. Please

keep this manual handy for future reference. If you have any questions, please

contact our authorized distributor(s) at the end of this manual.

NOTE

Store test strips in their original bottle to maintain quality. Do not transfer them

to other containers.

1.1 Intended Use

Your GLUCOCARD

TM

X-mini plus and GLUCOCARD

TM

X premium Test Strips

are for quantitatively measuring the glucose level in fresh capillary whole blood.

They are intended for use outside of the body (in vitro diagnostic use) at home

or in a clinical setting as an aid to monitor and control blood glucose levels. Do

not use them to diagnose diabetes. Also, do not alter treatment based on

the test results of this meter without instructions from your doctor or healthcare

professional.

1.2 Measurement Principle

Glucose in the blood reacts with the reagent in the test strip and this produces

a small electric current. The strength of this current is proportional to the

concentration of glucose in the blood. The meter measures this current and

calculates your glucose level.

1.3 What’s Included

GLUCOCARD

TM

X-mini plus

Carrying case

Operating Manual (this manual)

Simplified Manual / Information for Users

Multilet

TM

(lancets)*

Multi-Lancet Device

TM

(lancing device)*

GLUCOCARD

TM

X premium Test Strips*

Some kits do not include the above items with an asterisk (*).

2.2 What You Need for Each Test

Do not use this meter for your blood glucose testing if you are being tested

for galactose tolerance or xylose absorption. You may use this meter if you

are taking PAM (1-Methylpyridine-6-carbaldehyde oxime).

These treatments may produce higher test results than your actual

blood glucose level. Taking unnecessary measures to lower your blood

glucose level may induce severe low glucose symptoms such as coma.

Chapter 1 Introduction

For accurate test results, allow the meter to adjust to the surroundings:

Temperature: 10 to 40°C (50 to 104°F)

Humidity: 20 to 80% RH (Relative Humidity)

for 30 minutes before testing your blood glucose.

Do not store or use the meter where:

There are sharp temperature fluctuations.

Humidity is high enough to cause condensation (bathrooms, drying

rooms, kitchen, etc.).

There is a strong electromagnetic field (microwave oven, cell phone,

etc.).

Do not use the meter after it has been dropped in liquid or liquids have

entered inside, even if dried afterwards.

Keep hands away from the test strip slot on the meter. A thermo sensor is

housed inside the meter to minimize detection of errors.

Do not connect the communication cable to the connection terminal

during testing. The meter may be damaged, leading to inaccurate test

results.

Do not apply blood directly to the test strip slot on the meter.

Do not share your meter with anyone to avoid the risk of infection.

2.1 Important Health-Related Information

2.3 Cautions on Using the Meter

Chapter 2 Before Tests

WARNING

CAUTION

2.4 Cautions on Using the Test Strips

2.5 Meter Parts

Use only the GLUCOCARD

TM

X premium Test Strips for testing with the

GLUCOCARD

TM

X-mini plus. Do not use other test strips as it causes

inaccurate test results.

Do not use test strips beyond their expiration date. The expiration date is

written on the test strip bottle next to “Exp.”.

For accurate test results, allow test strips to adjust to the surroundings:

Temperature: 10 to 40°C (50 to 104°F)

Humidity: 20 to 80% RH

for 20 minutes before testing your blood glucose.

Do not use the test strips if 6 months or more have passed since

opening the bottle.

The test strips are for single-use only. Do not use test strips that have

already absorbed blood or control solution.

CAUTION

Keep the meter, test strips and other items away from young children. Small

items are choking hazards.

WARNING

Connection

terminal

a. GLUCOCARD

TM

X-mini plus (meter)

b. GLUCOCARD

TM

X premium Test Strips

c. Lancing device d. Lancet

Strap hole

Display

with backlight

Test strip slot with light

Insert a test strip here before

testing.

Test strip

disposal lever

Battery holder

Button

83-04341

Glucocard Σ mini, Глюкокард Сигма-мини

глюкокард сигма мини

Glucocard Σ mini, Глюкокард Сигма-мини — это уменьшенная и немного упрощенная версия глюкометра Glucocard Σ. Глюкометр Glucocard Σ mini имеет более компактные размеры и более легкий все – весит глюкометр всего 25г.
Для измерения уровня сахара Glucocard Σ mini требуется небольшая капля крови.
Глюкометр можно соединить с компьютером для передачи обработки результатов.
Глюкометр Glucocard Σ mini протестирован в России, и свободно продается в аптеках и интернет магазинах. Одна из линий по производству тест-полосок находится в России, в городе Дубна.
(далее…)

Glucocard Σ, Глюкокард Сигма

Glucocard Σ Glucocard Σ, Глюкокард Сигма — современный и быстрый глюкометр – проводит анализ всего за 7 секунд, имеет большой диапазон измерений и маленький вес.
Для измерения уровня сахара в крови Glucocard Σ требуется небольшая капля крови.
Глюкометр можно соединить с компьютером для передачи обработки результатов.
Glucocard Σ — протестирован в России и свободно продается в аптеках и интернет магазинах. На 2018 год этот глюкометр выписывают по бесплатным рецептам, а так-же выписывают рецепты на тест полоски к этому глюкометру.
Одна из линий по производству тест-полосок находится в России в подмосковной Дубне.
(далее…)

Glucocard X mini plus

Glucocard X-mini plusGlucocard X mini plus — стильный, надежный глюкометр, небольшого размера.
Глюкометр Glucocard X mini plus имеет подсвечивающийся, цветной экран, что делает возможным измерение сахара в плохо освещенном помещении.
Есть возможность ставить метки перед измерениями и делать заметки.
Выпускается в разных цветовых решениях.
Glucocard X mini plus — выпускается для Европейского рынка, в Россию официально не поставляется, и сертификацию не проходил. Тест-послоски можно приобрести в интернет магазинах.
(далее…)

Glucocard X mini

Glucocard X miniGlucocard X mini — Простой, надежный глюкометр, компактный и стильный.
Есть функция ставить метки при измерении сахара.
Glucocard X mini — это упрощенная версия глюкометра Glucocard X mini plus, отличия в отсутствие подсветки экрана, и менее информативный экран.

Glucocard X mini — производится для Европейского рынка, на Российском рынке он не представлен, тест-полоски можно приобрести через интернет магазины.
(далее…)

Glucocard Expression

Glucocard ExpressionGlucocard Expression — это продвинутый глюкометр, со множеством функций, он умеет проговаривать результаты, что удобно для людей с ослабленным зрением. Глюкометр может подключаться к компьютеру через USB и передавать результаты измерений для дальнейшего анализа как самостоятельного, так и врачом.

Глюкометр имеет большой экран с четким изображением, удобные крупные кнопки, которые помогут в проведении анализов людям с ослабленным зрением.

Glucocard Expression — производится для Американского рынка, в Россию официально не поставляется, тест-полоски можно купить через интернет магазины, в свободной продаже тест-полосок для этого глюкометра нет.
(далее…)

Glucocard Vital — Arkray

Glucocard VitalGlucocard Vital — хороший глюкометр отличающийся своей простотой, точностью и быстротой выполнения измерений.
Корпуса глюкометра выполнены в разных цветах.
Glucocard Vital — не распространен в России, официально в Россию и на территорию Европы не поставляется, глюкометр производиться для рынка США и северной Америки. Тест-полоски для этого глюкометра можно приобрести в западных интернет магазинах.
(далее…)

Glucocard 01 mini plus

Glucocard 01 mini plusGlucocard 01 mini plus от компании Arkray сочетает в себе Маленький и удобный корпус и современный дизайн. Экран глюкометра имеет подсветку. Glucocard 01 mini plus может соединяться с компьютером для дальнейшей обработки результатов с помощью программы MEQNET SMBG Viewer
Glucocard 01 mini plus — производиться для Европейского рынка и не распространен в России, тест-полоски для этого глюкометра можно приобрести в популярных западных интернет магазинах
(далее…)

Glucocard 01 mini

Glucocard 01 miniGlucocard 01 mini — Маленький, компактный глюкометр.

Его корпус выполнен в различных цветовых решениях, яркий и веселый. Можно выбрать из более чем 20ти вариантов расцветок.

Glucocard 01 mini — официально не поставлялся в Россию, тест-полоски для этого глюкометра найти сложно.
(далее…)

Glucocard 01 — Arkray

Glucocard 01 Glucocard 01 — Глюкометр очень удобный и простой в использовании.
Имеет компактные размеры – практически плоский, легкий.
Небольшие тест-полоски и компактный бокс для них.
Glucocard 01 — производился для рынка США, в Россию и Европу не поставлялся, на данный момент купить расходные материалы для данного глюкометра еще можно, но уже сложно, стоит рассмотреть более современную замену.
(далее…)

Glucocard X-METER GT 1910

Glucocard X-METER GT 1910Glucocard X-METER GT 1910 — Надежный и точный глюкометр, большой и высококонтрастный дисплей, высокое японское качество.
Простой в применении, невычурный белый дизайн корпуса.
Glucocard X-METER GT 1910 — производиться для рынка Европы, в Россию не поставляется, тест-полоски можно купить через интернет магазины. На данный момент приобрести тест-полоски сложно, глюкометр снимается производства, и заменяется на глюкометры нового поколения.
(далее…)

Glucocard 2

Glucocard 2

Glucocard 2 — этот довольно распространённый глюкометр отличает знаменитое японское качество и точность в измерениях, а также широкий диапазон измерений.

Но для измерения глюкометру требуется пол минуты на проведение анализ и очень большая капля крови.

Небольшая память глюкометра и отсутствие выведения средних значений.

Glucocard 2 проходил тестирование в России и продавался в апетках, как сам глюкометр, так и тест-полоски к нему, но особой популярности он не приобрел, поэтому сейчас для него уже довольно трудно найти расходные материалы. Скорее всего глюкометр снят с производства, и расходные материалы тоже не производятся.
UPD 2019. Glucocard 2 — глюкометр снять с производства, тест-полоски не производятся.
Glucocard 2 — очень похож на глюкометр Ascensia Elite от фирмы Bayer, но тест полоски не взаимозаменяемы, хотя и очень похожи.
(далее…)