Model MB HTHP Filter Press

For temperatures up to 350°F (177°C)

Instruction Manual

11302 Steeplecrest Dr. · Houston, Texas · 77065 · U.S.A.

Tele: 832.320.7300 · Fax: 713.880.9886 · www.ofite.com

171-50 (115V)

171-51 (230V)

Updated 5/6/2011

Ver. 2.1

OFI Testing Equipment, Inc.

Copyright OFITE 2011

©

loading

Related Manuals for OfiTE MB HTHP

Summary of Contents for OfiTE MB HTHP

  • Page 1
    171-50 (115V) 171-51 (230V) For temperatures up to 350°F (177°C) Instruction Manual Updated 5/6/2011 Ver. 2.1 OFI Testing Equipment, Inc. 11302 Steeplecrest Dr. · Houston, Texas · 77065 · U.S.A. Tele: 832.320.7300 · Fax: 713.880.9886 · www.ofite.com Copyright OFITE 2011 ©…
  • Page 2: Table Of Contents

    Table of Intro…………………..2 Specifications ………………2 Contents Components………………3 Safety ………………..6 Procedure …………………7 OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com…

  • Page 3: Intro

    Maximum Pressure: 1,500 PSI (10,342.5 kPa) Pressure Source: 2 CO Cartridges Test Cell Capacity: 250 mL Voltage: #171-50: 115V; #171-51-C: 230V Heater: 200 Watt OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com…

  • Page 4: Components

    #170-07 O-ring; Qty: 2 #170-32 1/8″ x 1/8″ NPT Male Needle Valve #171-66 Stainless Steel Receiver Tube #171-75 Receiver Body #171-79 ¼» Hex Wrench OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com…

  • Page 5
    #143-05 EZ Puncture CO Bulbs; 8-Gram; UN #1013; Package of 10 #143-20 Repair Kit for High-Pressure Victor Regulator (#171-53) #171-81 Stainless Steel Carrying Case OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com…
  • Page 6
    Red Pilot Light #171-69 White Pilot Light #171-71 Thermostat #171-73-1 1000-PSI Gauge; 1/8″ Bottom; 1½» Face #171-78 Cell Screw; Qty: 4 #171-79 ¼» Hex Wrench OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com…
  • Page 7: Safety

    All safety precautions must be met, especially in the cell breakdown procedure after the filtration procedure has been completed. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com…

  • Page 8: Procedure

    Evenly tighten the set screws with the allen wrench. O-ring Groove Cell Cap (#171-62) Locking Screws (#171-78) OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com…

  • Page 9
    Back Pressure Receiver (#171-58) T-screw T-screw Cartridge Holder (#143-03) Needle Valve Needle Valve (#170-32) (#170-34) Male Coupling (#171-64) Locking Ring O-ring (#142-58) Back Pressure Receiver OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com…
  • Page 10
    500 PSI (3,447.5 kPa) above the back pressure. Open the bottom valve to initiate filtration and begin timing. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com…
  • Page 11
    Measure and report the thickness of the filter cake to the nearest 1/32″ (0.8 mm). 16. Clean and dry the apparatus thoroughly after each use. Inspect all o- rings and replace if necessary. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com…
  • Page 12
    Safety Bleeder Valve (#143-06) (#171-75) Cartridge Barrel (#143-03) Male Needle Valve (#170-32) Graduated Cylinder (#153-16) Filter Paper (#170-19) Stopcock Grease (#153-55) Stand and Base (#171-59) OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com…

Предложите, как улучшить StudyLib

(Для жалоб на нарушения авторских прав, используйте

другую форму
)

Ваш е-мэйл

Заполните, если хотите получить ответ

Оцените наш проект

1

2

3

4

5

OFITE 170-00-4S 4-Unit HTHP Filter Press Instruction Manual
OFITE 170-00-4S 4-Unit HTHP Filter Press

Intro

The OFITE 4-Unit HTHP Filter Press Heat Jacket is designed to hold four, 175-mL HTHP test cells for simultaneous testing. Each cell has a separate temperature controller and two pressure regulators (drive pressure and back pressure) for independent testing.

Specifications

Maximum Temperature: 425°F (218.3°C)
Maximum Inlet Pressure: 2,000 PSI (13.8 MPa)
Maximum Drive Pressure: 1,350 PSI (9.3 MPa)
Maximum Back Pressure: 750 PSI (5.2 MPa)
Power Requirements: 115 VAC or 230 VAC configurations available (power cord included)
Cell Size: 175 mL (Cells not included)

Components

Included:

  • #120-05-005: Fuse Holder (230 Volt)
  • #120-05-005-1: Fuse Holder (115 Volt)
  • #120-70-1-052: Hose, Stainless Steel, 18″
  • #130-81-015: Relief Valve
  • #130-81-028: Regulator
  • #152-37: AC Power Cord (115 Volt)
  • #152-38: AC Power Cord (230 Volt)
  • #153-14: Graduated Cylinder, 50 mL × 1 mL, Glass
  • #170-06-1: Back Pressure Receiver
  • #170-07: O-ring for Receiver Body
  • #170-19: Filter Paper, Specially Hardened for Filter Presses, 2.5″
  • (6.35 cm): Diameter, Box of 100
  • #170-20: Manifold Block
  • #170-93: Wrench for Valve Stem
  • #171-45-7: Thermocouple
  • #171-23-1: Safety Pin with Lanyard
  • #174-03: Temperature Controller

Test Cells:

Test cells are not included with the instrument. The following 175 mL cells are available.

  • #170-12-1: Test Cell for Filter Paper, 175 mL, Single Cap
  • #170-45: Test Cell with Cement Screens, 175 mL, Double Cap
  • #170-48: Test Cell for Ceramic Disks, 175 mL, Single Cap
  • #170-46: Test Cell for Ceramic Disks, 175 mL, Double Cap

Accessories:

  • #170-13-3: O-ring for Test Cell, Viton®/Fluorocarbon (FKM)
  • #170-16-06: Valve Stem, 4.5″ Long
  • #170-17: O-ring for Valve Stem, Viton®/Fluorocarbon (FKM)
  • #170-26-1: Locking Screw, Hardened Alloy Steel

Safety

Optional:
Load Cell Accessory Kit:
#170-00-4S-LC: Load Cell Assembly; Qty: 4
#170-00-4S-013: 4-Port USB Hub

Nitrogen must be supplied in an approved Nitrogen Gas Cylinder and secured to meet safety standards.

Due to the high temperatures and pressures involved in this test, extreme care must be exercised at all times. All safety precautions must be met, especially in the cell breakdown procedure after the filtration procedure has been completed.

Cell Corrosion
Test fluids under high tempreature and pressure can corrode the cell body and caps. Carefully inspect the cell body and calls for corrosion before and after each test.

Some materials are more susceptible to corrosion than other. Also, some fluids and additives are more corrosive than others. OFITE offers a variety ofcell materials for different levels of corrosion resistance and cost.

Setup

  1. Remove the instrument from the crate and place it on a flat, stable surface.
  2. Turn off the Nitrogen valve (horizontal). Turn off both Main Power switches and all four Heat switches. Make sure all regulators are backed off completely. Turn them completely counterclockwise.
  3. Connect a Nitrogen source to the back of the unit. The inlet connector is ¼” NPT.
  4. On the back of the unit is a power inlet configured for 115 VAC or 230 VAC. Plug the supplied power cord into an appropriate power source.
  • #170-00-4S: 115 VAC, 50/60 Hz
  • #170-00-4S-230: 230 VAC, 50/60 Hz

Operation

Although all four test cells share a single Nitrogen source, they each have two regulators for adjusting drive pressure and back pressure. Each cell also has a separate temperature controller.

There are two main power switches on the front of the instrument. The switch on the left controls power to the two cells on the left. The switch on the right controls power to the two cells on the right.

The valve on the front controls the flow of Nitrogen pressure to the regulators. When the valve is horizontal, the pressure is turned on. When it is vertical, the pressure is turned off.
Operation

IconNote

IconTip

  1. Before starting a test, make sure the Nitrogen valve is off (vertical). Turn off both Main Power switches and all four Heat switches. Make sure all regulators are backed off completely. Turn them completely counterclockwise.
  2. Turn the Main Power switches on.
  3. Press the up and down arrows on the temperature controller to set the desired test temperature.
  4. Turn the Heat switch on. Each cell has a separate Heat switch. The light next to the cell will turn on when the heaters are engaged.
  5. Be sure all of the o-rings are in good working condition (pliable with no nicks or cuts, etc.), and are not damaged during the assembly procedures. Place a thin film of silicone grease on all o-rings.
    Nitril-NBR (Buna N) o-rings (#170-13) are recommended for tests up to 250°F (121°C). For tests up to 400°F (204°C), Fluorocarbon/FKM (Viton®) o-rings (#170-13-3) are recommended. For tests up to 500°F (260°C), Perfluorocarbon (FFKM) o-rings are recommended. Also, for tests with water-based fluids containing no hydrocarbons, Ethylene propylene (EPM/EPDM) o-rings may be used up to 400°F (204°C).
  6. Screw a valve stem into the test cell on the side opposite the cell cap. Tighten the valve stem completely.
  7. Prepare the sample according to API specifications.
  8. Carefully pour the sample into the cell. Do not fill the cell closer than 0.5″ 13 mm) from the o-ring groove to allow for heat expansion of the fluid. Be careful not to spill fluid in the o-ring groove.
  9. Place an o-ring in the o-ring groove inside the cell. Place another o-ring in the recess in the cell cap.
  10. Place a circle of filter paper on top of the cell o-ring and slowly push the cell cap into the test cell. Make sure the arrow on the cell cap lines up with the arrow on the cell body. Tighten the cap locking screws.
    Important
    If the cap locking screw seats are oval shaped and no longer round, there is a possibility of stress failure and the cap should be replaced.
  11. Screw another valve stem into the cell cap and tighten it completely. The cell is now completely sealed.
    Operation Instruction
  12. Place the cell in the heating jacket with the outlet or filter side of the cell pointed down. Rotate the cell in the heating jacket so that the pin in the bottom of the heating well seats into the hole in the bottom of the cell. This will anchor the cell inside the well and prevent the cell from rotating as the valve stems are opened and closed.
  13. Transfer the thermocouple from the heating jacket to the thermometer well within the cell.
  14. Connect the manifold block to the top valve stem and lock it in place with the retainer pin. Place the back pressure receiver on the bottom valve stem and lock it in place with the other retainer pin. Make sure both relief valves and both valve stems are closed.
    Operation Instruction
  15. Turn the Nitrogen valve on.
  16. Set the top and bottom regulators to the recommended back pressure for your test.
    This instrument is equipped with check valves between the regulators and the pressure outlet to protect the regulators from backflow. If you increase the regulator pressure and then decrease it, there may be extra pressure remaining in the lines.
    Recommended Minimum Back Pressure
    Test Temperature Vapor Pressure Minimum Back Pressure
    °F °C psi kPa psi kPa
    200-299 95-149 12.1-67 84-462 100 700
    300-374 150-189 67-184 462-1269 200 1400
    375-399 190-199 184-247 1269-1704 275 1900
    400-424 200-219 247-326 1704-2245 350 2500
    425-450 220-230 326-422 2245-2912 450 3100
  17. Open the top valve stem by turning it one quarter turn. This will add pressure to the test cell. Maintain this pressure until the sample temperature stabilizes. The heating time of the sample should never exceed one hour.
  18. When the fluid sample reaches the desired test temperature, increase the pressure on the top pressure unit to 500 PSI (3,448 kPa) more than the back pressure. Open the bottom valve stem to initiate filtration.
  19. Collect the filtrate for 30 minutes. If the back pressure rises above the initial setting during the test, cautiously reduce the pressure by opening the valve on the receiver and drawing off some of the filtrate into the graduated cylinder.
  20. At the end of the test, close (tighten) the top and bottom valve stems. This will seal the cell. Failure to close the valve stems before releasing pressure (step 22) will damage a regulator by drawing fluid into it.
  21. Turn off the heat switch and set the temperature controller to the lowest setting.
  22. Open the outlet valve on the back pressure receiver to collect all of the filtrate in the graduated cylinder. Open the relief valve on the manifold block. This will flush the pressure lines and ensure no test fluid has flowed back into them.
  23. Slowly turn the regulators counterclockwise to close off the flow of pressurized gas.
  24. Remove the top and bottom retaining pins and remove the manifold block and the back pressure receiver. Drain any residual filtrate collected in the receiver into the graduated cylinder.
  25. Remove the cell from the heating jacket after once again checking that the cell valve stems are tightly closed. Allow it to cool to room temperature or quick cool the cell by immersion in cool water. Pressure inside the sample cell will still be approximately 500 PSI (3,450 kPa). Keep the cell upright and cool it to room temperature before disassembling. The cell must be cool for at least one hour at room temperature or at least 10 minutes in cool water before loosening the cap locking screws and removing the cell cap.
  26. Correct the total filtrate volume collected to a standard filtration test area of 7.1 in2 (45.8 cm2 ) by doubling the filtrate volume collected in 30 minutes. Record this total filtrate volume (doubled), temperature, pressure, and time.
  27. Using extreme care to save the filter paper and deposited cake, place the cooled cell upright with the outlet (cap side) or filter side down. Open (loosen) the inlet valve stem to bleed off pressure from the cell body. Pressure cannot be relieved from the cell by opening the outlet valve stem as the filter cake will seal off the cell. It is a good idea to open the valve stem with the cell inside a sink or with a rag over the valve stem in order to catch any liquid that might be ejected.
  28. Loosen, but do not remove, the six cap locking screws, and separate the cap from the cell with a slight rocking motion. Discard the fluid inside the cell and retrieve the filter cake.
  29. Wash the filter cake on the paper with a gentle stream of water. Measure and report the thickness of the filter cake to the nearest 1/32″ (0.8 mm).
  30. Clean and dry the apparatus thoroughly after each use. Inspect and, if necessary, replace all of the o-rings.
  31. After all testing is complete, make sure all four heat switches are off, the temperature controllers are set to 0, the regulators are backed off completely (counterclockwise), and both Main Power switches are off.

Load Cell Option

The Load Cell option (#170-00-4S-LC) consists of four low profile precision scales designed to take more accurate measurements of fluid dispensed from filtration tests. Information from each load cell is sent to the user’s computer via USB connections. The included software records the filtrate volume in each cell and graphs it with respect to time.

  1. Set up Load Cells in their designated positions on the base of the 4-unit filter press making sure that the corners of the base plates meet with their correlated posts.
    IconNote
    Place load cells 1 and 2 to the left of their designated support rods and load cells 3 and 4 to the right of their designated support rods.
    Operation Instruction
  2. Run the USB cables through the grommets on the front panel and into the unit cabinet.
  3. Plug the USB cables into the 4-Port Hub (#170-00-4S-013).
    Actual component may vary.
    Operation Instruction
  4. Plug the Filter Press into the computer with the supplied USB cable.

Software Setup

Before running the software, install the driver as described on page 15.

  1. Select Setup on the Utilities menu.
    Software Setup
  2. Select a COM port for each cell.
    Each load cell is assigned a COM port the first time it is connected to the computer. This COM port will remain the same unless the computer or the internal electronics of the load cell are changed.
    Repeat the following procedure for each of the four cells:
    a. Under Cell X Com, choose a COM port from the drop-down list.
    b. Click OK.
    c. On the Main Screen, locate the Filtrate Volume field under Cell X. Place a weight on the load cell and watch the field for changes.
    Software Setup
    • If the value changes when weight is applied, you have found the correct COM port for Cell X. Continue to step 3.
    • If the value does not change, return to step 1 and try a different COM port.
  3. Choose an Archive Path. This is where the software will save data files.
  4. Choose a Logo Path. Select a logo (.JPG) to print on the graph at the end of the test.
  5. “Print to Printer” – When this option is on, a graph of the test results will automatically print to the default printer when a test is complete.

Operation

  1. On the Test Setup menu, select a Cell from the list.
  2. Enter a test name and a specific gravity (g/cm3 ). In the Sample Composition field, enter any notes about the test or the components of the sample.
    Software Setup
  3. Click OK.
  4. Place a clean, dry graduated cylinder onto the load cell. Click the Zero button to zero the Filtrate Volume.
  5. Click the Start Test button before initiating filtration on the cell.
  6. When the test is complete, click the Stop Test button to stop recording.
    Software Setup

Calibrate each load cell (one at a time) prior to testing.

  1. Navigate to the calibration window. Utilities → Select a Cell Number to calibrate.
  2. Place a desired amount of weights onto a load cell.
  3. Enter the value of the weight in the 1st set weight (g) field and press the green arrow to display the Raw value.
  4. Add desired weight to the load cell.
  5. Enter the value of the weight in the 2nd set weight (g) field and press the green arrow to display the Raw value.
  6. Add desired weight to the load cell.
  7. Enter the value of the weight in the 3rd set weight (g) field and press the green arrow to display the Raw value.
  8. Press the “Store Cal” button to save the calibrations and exit.
  9. Repeat steps 2 – 8 for the remaining three load cells
    Operation

Appendix

Driver Install

A driver is software that a device (the HTHP Filter Press) uses to work with a computer. Every HTHP Filter Press requires a driver to work with your PC before the software can be operated.

  1. Insert the software/driver CD to the computer and select the file labeled: CDM20828.exe
  2. A prompt will open, asking if you want to run this file. Press Enter or click “Run”.
    Driver Install
  3. The FTDI CDM Driver will open to allow you to extract the driver and launch the installer. Click “Extract”.
    Driver Install
  4. A welcome window will open. Click “Next”.
    Driver Install
  5. Accept the license agreement and click “Next”.
    Driver Install
  6. Once the driver installation is complete, Click “Finish”.
    Driver Install
  7. Plug he USB portion of the USB/RS232 cable (#130-76-19) to the computer.
    The computer will continue installing the software. Wait for it to finish installing.
  8. Once the installation is complete, navigate to the Device Manager: Click the Start Menu. Open the “Control Panel”. Click the “Hardware and Sound” link. Click on the Device Manager link in the “Devices and Printers” options.
  9. Confirm which communication port (COM Port) the USB Serial Port is located in. The window below is indicating that this sample port was connected to “COM3” which can be a different COM# depending on the computer and the port it was connected to. This is the COM Port that the driver has been assigned to.
    Driver Install
  10. Open the Options screen in the software. Choose the COM port listed in Device Manager.

Warranty

OFI Testing Equipment, Inc. (OFITE) warrants that the products shall be free from liens and defects in title, and shall conform in all respects to the terms of the sales order and the specifications applicable to the products. All products shall be furnished subject to OFITE’s standard manufacturing variations and practices. Unless the warranty period is otherwise extended in writing, the following warranty shall apply: if, at any time prior to twelve (12) months from the date of invoice, the products, or any part thereof, do not conform to these warranties or to the specifications applicable thereto, and OFITE is so notified in writing upon discovery, OFITE shall promptly repair or replace the defective products. Notwithstanding the foregoing, OFITE’s warranty obligations shall not extend to any use by the buyer of the products in conditions more severe than OFITE’s recommendations, nor to any defects which were visually observable by the buyer but which are not promptly brought to OFITE’s attention.

In the event that the buyer has purchased installation and commissioning services on applicable products, the above warranty shall extend for an additional period of twelve (12) months from the date of the original warranty expiration for such products.

In the event that OFITE is requested to provide customized research and development for the buyer, OFITE shall use its best efforts but makes no guarantees to the buyer that any products will be provided.

OFITE makes no other warranties or guarantees to the buyer, either express or implied, and the warranties provided in this clause shall be exclusive of any other warranties including ANY IMPLIED OR STATUTORY WARRANTIES OF FITNESS FOR PURPOSE, MERCHANTABILITY, AND OTHER STATUTORY REMEDIES WHICH ARE WAIVED.

This limited warranty does not cover any losses or damages that occur as a result of:

  • Improper installation or maintenance of the products
  • Misuse
  • Neglect
  • Adjustment by non-authorized sources
  • Improper environment
  • Excessive or inadequate heating or air conditioning or electrical power failures, surges, or other irregularities
  • Equipment, products, or material not manufactured by OFITE
  • Firmware or hardware that have been modified or altered by a third party
  • Consumable parts (bearings, accessories, etc.)

Returns and Repairs:
Items being returned must be carefully packaged to prevent damage in shipment and insured against possible damage or loss. OFITE will not be responsible for equipment damaged due to insufficient packaging.

Any non-defective items returned to OFITE within ninety (90) days of invoice are subject to a 15% restocking fee. Items returned must be received by OFITE in original condition for it to be accepted. Reagents and special order items will not be accepted for return or refund.

OFITE employs experienced personnel to service and repair equipment manufactured by us, as well as other companies. To help expedite the repair process, please include a repair form with all equipment sent to OFITE for repair. Be sure to include your name, company name, phone number, email address, detailed description of work to be done, purchase order number, and a shipping address for returning the equipment. All repairs performed as “repair as needed” are subject to the ninety (90) day limited warranty. All “Certified Repairs” are subject to the twelve (12) month limited warranty

Returns and potential warranty repairs require a Return Material Authorization (RMA) number. An RMA form is available from your sales or service representative.

Please ship all equipment (with the RMA number for returns or warranty repairs) to the following address:

OFI Testing Equipment, Inc.
Attn: Repair Department 11302 Steeplecrest Dr. Houston, TX 77065 USA

OFITE also offers competitive service contracts for repairing and/or maintaining your lab equipment, including equipment from other manufacturers. For more information about our technical support and repair services, please contact techservice@ofite.com.

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com

OFI Testing Equipment, Inc.
11302 Steeplecrest Dr. · Houston, Texas · 77065 · U.S.A.
Tele: 832.320.7300 ·
Fax: 713.880.9886 · www.ofite.com
©Copyright OFITE 2015

OFITE Logo

Documents / Resources

OFITE 170-00-4S 4-Unit HTHP Filter Press [pdf] Instruction Manual
170-00-4S, 170-00-4S-230, 170-00-4S-LC, 170-00-4S 4-Unit HTHP Filter Press, 4-Unit HTHP Filter Press, HTHP Filter Press, Filter Press, Press

Download or browse on-line these Instruction Manual for OfiTE HTHP Measuring Instruments.

Summary of Contents:

[Page 1] OfiTE HTHP

OFITE HTHP Viscometer Part No. 130-77 Instruction Manual Updated 8/22/2019 Ver. 7 OFI Testing Equipment, Inc. 11302 Steeplecrest Dr. · Houston, Texas · 77065 · U.S.A. Tele: 832.320.7300 · Fax: 713.880.9886 · www.ote.com © Copyright OFIT…

[Page 2] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 1 Intro ……………………………………………………………………………………..2 Components ……………………

[Page 3] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 2 When extremely high-temperature and/or high-pressure viscosity measurements are required, the OFITE HTHP viscometer is the solution. This…

[Page 4] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 3 Components #120-00-016-1 Pressure TDR, 50,000 PSI (344.8 MPa) #120-001 Mineral Oil, 2 Gallons #120-106 High Pressure Filter #130-75-71 …

[Page 5] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 4 Optional: #130-77-SP Spare Parts Kit #130-77-002 O-ring for Cell Assembly, Teon, For tests above 400°F, Qty: 6 #130-77-080-1 Torsi…

[Page 6] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 5 Specications Instrument Geometry True Couette Coaxial Cylinder Motor Technology Stepper Motor Speeds Variable: .01 – 1,000 RPM Speed …

[Page 7] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 6 Setup Computer 1. Connect the monitor, mouse, and keyboard to the computer and connect the monitor and computer to an AC power outlet. T…

[Page 8] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 7 Setup Viscometer 1. Place the unit on a solid, level surface away from any large motors or strong magnetic elds. 2. The HTHP Viscome…

[Page 9] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 8 Control Panel During a test, the computer will be continuously adjusting the pressure inside the cell. As the pressure increases due to t…

[Page 10] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 9 Operation 1. Remove the entire test cell from the heating well and place it in the preparation stand. 2. Remove the cell and the cell c…

[Page 11] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 10 7. Check the movement of the rotor. To test for resistance, move a magnet or magnetized metal object around the outside of the cell at …

[Page 12] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 11 12. Place an o-ring (#130-77-3) into the designated groove on top of the coupling. Place a metal seal ring (#130-77-22) on top of the o-…

[Page 13] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 12 20. For tests above 200°F (93.3°C), screw the two clamps into the holes on the bulkhead to secure the cell in place. 21. Inspect the o…

[Page 14] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 13 25. Fill a syringe with test uid. Inject 15 mL of the test uid into the “Front” port. This step is only necessary when testing…

[Page 15] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 14 Software Start 1. Double-click the “ORCADA ® ” icon on the desktop. 2. If this is a fresh install, you will be asked to select a h…

[Page 16] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 15 The “Mud” and “Cement” tests ignore the “Dead Time” and “DAQ Time” settings on the Options screen. Temperature and press…

[Page 17] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 16 For example, assume a test has a conditioning time of 10 minutes and the test is paused at the 5 minute mark for 1 minute. If the test i…

[Page 18] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 17 “PreHeat” — This eld is not used on the HTHP Viscometer. “Cool” — Place a check in this box to activate the cooling solenoid a…

[Page 19] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 18 Calibration The HTHP Viscometer is designed to be calibrated with certied calibration uids available from OFITE. These uids are…

[Page 20] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 19 To zero the unit, loosen the bolts on the transducer housing and rotate the hood slightly while observing the “Shear Stress Raw” val…

[Page 21] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 20 “S.S. Reference” — the shear stress as calculated by the software “S.S. Raw” — the actual signal coming from the unit “RPM” -…

[Page 22] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 21 “Temperature” — the temperature of the sample “Shear Rate” — rotational speed ( s) “RPM” — rotational speed (RPM) “Sh…

[Page 23] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 22 Calibration Fluid Manager Before calibrating the unit, you must input the specications of the calibration uid that you are using….

[Page 24] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 23 Software Options To access the Options screen, choose “Options” from the “Edit” menu. “Pressure Minimum Deadband” — During te…

[Page 25] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 24 “Enable Auto Cool” — Check this box to have the viscometer automatically cool the cell after a test. “Deactivate Cooling when Temp…

[Page 26] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 25 “Wait for Temp Mode” — If a step in the test is set to wait for the temperature setpoint, this option will control how the software …

[Page 27] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 26 Software Save Rate Settings By default, the ORCADA ® Software saves data to le based on a set time interval. This time interval ca…

[Page 28] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 27 Software Test Builder The Test Builder is designed to help you build custom tests for use in Auto mode. To access the test builder, go t…

[Page 29] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 28 d. “Pressure” — the pressure setpoint e. “Cond. Time” — the time period to condition the sample f. “Cond. RPM” — the rotat…

[Page 30] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 29 “Stir Prior to Gel” — Check this box to stir the sample before calculating gel strength. “Gel Stir Rate” — This is the rate the …

[Page 31] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 30 Software Select Analysis Outputs After building a test in the Test Builder, the chart at the bottom of the Main Screen will show the va…

[Page 32] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 31 Software Saved Test Data To review test data from past tests: 1. Select “Saved Test Data” from the “File” menu. 2. The saved t…

[Page 33] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 32 “Export” — Click this button if you wish to export the test data to a le. You will be asked to select a lename and destination…

[Page 34] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 33 Software Calibration History On the Calibration History Screen, you can retrieve the data from all past calibration tests. 1. Begin on…

[Page 35] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 34 Software Import/Export The Import/Export feature of the ORCADA ® Software makes it possible to transfer a test from one PC to another…

[Page 36] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 35 Software Starting a Test The ORCADA ® software has two modes for running tests: Auto and Manual. By default, the software is in Manua…

[Page 37] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 36 Maintenance In order to keep the HTHP Viscometer in good working condition, proper cleaning and maintenance are essential. After each te…

[Page 38] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 37 Disassembly Cell Before releasing the pressure in the test cell, make sure it has cooled to 150°F (65.5°C) or less. The cell should b…

[Page 39] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 38 Disassembly Rotor 1. Using a at-head screwdriver, remove the two screws on the bottom of the rotor and remove the magnet holder. 2….

[Page 40] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 39 5. The rotor should be further disassembled by removing the rotor bearing. a. Using a small punch, press the rotor bearing out of the m…

[Page 41] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 40 Disassembly Cell Top 1. Unscrew the cell cap and remove it. The strap wrench may be required. 2. Lift the pivot cap off of torsion spr…

[Page 42] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 41 Disassembly Coupling 1. Remove the bafe. It may be necessary to put a small screwdriver or allen wrench through the holes in the ba…

[Page 43] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 42 Remove/Install Angular Contact Bearing 1. Remove the rotor from the test cell. 2. Inspect the thermowell for damage or wear. Replace it…

[Page 44] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 43 Remove/Install Pivot Bushing Removal: Once the angular contact bearing inner race is removed, the pivot bushing can also be removed with…

[Page 45] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 44 Remove/Install Rotor Bushing Removal: Refer to page 38 for instructions. Installation: 1. Set the rotor cup down with the open end down….

[Page 46] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 45 Remove/Install Torsion Assembly Removal: 1. Slide the pivot cap off of the torsion assembly. 2. Unscrew the two screws that secure the …

[Page 47] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 46 Installation: 1. Slide the zeroing sleeve into the spring housing and tighten the set screw to secure it in place. 2. Slide the spring…

[Page 48] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 47 Appendix Zeroing the Compass The compass, inside the transducer housing, tracks the movement of the spring assembly by measuring the ma…

[Page 49] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 48 9. Place the pivot cap back onto the torsion assembly and place the cell cap back onto the test cell. 10. Unplug the serial cable from …

[Page 50] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 49 13. Click “Start 8-Point Cal”. 14. Follow the on-screen instructions. The software will prompt you to turn the cap 45° and then clic…

[Page 51] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 50 Appendix Reinstalling Software Reinstallation of the ORCADA ® and data acquisition software packages will only be necessary in the cas…

[Page 52] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 51 Appendix Minimum Required Pressures Temperature Pressure °F °C PSI kPa Ambient – 200 Ambient – 93.3 0 0 201 – 295 93.9 – 146….

[Page 53] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 52 Appendix Conversion Charts Viscosity Conversions To convert from units on left side to units on top, multiply by factor @ intercept. Cen…

[Page 54] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 53 Vee Jewel (#130-77-1) Torsion Spring Module (#130-77-080) Appendix Test Cell Assembly Port Adapter (#130-77-36) Cell Body (#130-77…

[Page 55] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 54 Appendix Test Cell Assembly — Exploded Cell Cap (#130-77-31) Vee Jewel (#130-77-1) Pivot Cap (#130-77-46) Pivot (#130-77-12) Torsion…

[Page 56] OfiTE HTHP

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 55 Warranty and Return Policy Warranty: OFI Testing Equipment, Inc. (OFITE) warrants that the products shall be free from liens and defects…

Фильтр пресс низкого давления OFITE имеет конструкционные особенности ячейки бурового раствора, узла создания давления и нижней крышки с ситом и фильтровальной бумагой. Разработанные для полевого и лабораторного использования эти установки стали промышленным стандартом для исследования фильтрации низкого давления / низкой температуры.

OFITE также производит другие типы фильтр-пресса и полный ассортимент запасных частей и расходных материалов. Возможно изготовление большинства установок из нержавеющей стали. В случае необходимости выполнения нескольких испытаний одновременно, доступны установки с двумя, четырьмя или шестью ячейками. Эти комбинированные установки комплектуются всей необходимой трубопроводной арматурой. Полевые портативные приборы обеспечены блоками для создания давления, трубопроводами. Американский Нефтяной Институт (API) рекомендует стандартное давление в ячейке 100 ± 1.0 пси (690 ± 6.9 kPa), в пределах времени выполнения теста от 30 секунд до 30 минут .
Когда требуются измерения фильтрации при высокой температуре и высоком давлении, рекомендуются фильтр-прессы HPHT , предназначенные специально для того, чтобы моделировать скважинные условия. Температуры испытания для всех моделей могут превышать 200°F (93°C), для сбора фильтрата установлен приемник с системой обеспечения противодавления. Компоненты HPHT фильтр-прессов OFITE полностью взаимозаменяемы с компонентами поставляемыми другими изготовители. Доступны три различных размера , чтобы соответствовать различным требованиям. Серия 170-00 — полевой портативный аппарат с ячейкой для образца на 175 мл и малым потреблением энергии, состоящим из двух нагревателей на 200 ватт. Испытательная ячейка имеет рубашку для обогрева, чтобы гарантировать полный и однородный нагрев образца. Для создания давления в ячейке и противодавления в приемнике обычно используются баллончики с CO2 , однако другие системы для создания давления также доступны. Особенности фильтр-пресса HPHT на 250 мл (№ 171-50) быстрое подключение для создания давления. Он также портативен и может быть заказан из нержавеющей стали (Номер 171-81). Фильтр-пресс типа 175 мл и 250 мл предназначен для температур испытания до 350°F (176°C). № 171-00 ряд оборудован ячейкой для образцов на 500 мл и четырьмя нагревателями на 200 ватт для температур выше 350°F (176°C). Этот ряд наиболее полно соответствует для лабораторного использования. Также включена система регулятора для подключения к баллоном с азотом при более высоких давлениях. Все приборы HPHT можно заказать или для использования в сетях с напряжением на 230 Вольт или на 115 Вольт. Доступна многоблочная модель, которая может быть скомпонована в соответствии с вашими спецификациями. Также доступен широкий выбор дополнительных принадлежностей и запчастей.

Понравилась статья? Поделить с друзьями:
  • Ht 1890 инструкция на русском
  • Hsp savagery 1 8 инструкция
  • Hsp brontosaurus инструкция на русском
  • Hp deskjet plus ink advantage 6075 инструкция
  • Hp deskjet plus 4100 инструкция