Jumo ecotrans ph03 инструкция на русском

JUMO ecoTRANS pH 03 Operating Manual

  • Contents

  • Table of Contents

  • Bookmarks

Quick Links

JUMO ecoTRANS pH 03

Microprocessor/transmitter / switching device

for pH value / redox voltage and temperature

Operating manual

20272300T90Z001K000

V2.00/EN/00506536

loading

Related Manuals for JUMO ecoTRANS pH 03

Summary of Contents for JUMO ecoTRANS pH 03

  • Page 1
    JUMO ecoTRANS pH 03 Microprocessor/transmitter / switching device for pH value / redox voltage and temperature Operating manual 20272300T90Z001K000 V2.00/EN/00506536…
  • Page 3: Table Of Contents

    Content Typographical conventions ……6 Warning signs ……… . 6 Note signs .

  • Page 4
    Content Calibration ……..40 General information ……..40 9.1.1 When is calibration required? .
  • Page 5
    Display and LED messages ….. . . 62 12.1 Operating states of JUMO ecoTrans pH 03 ….62 12.2 Underrange .
  • Page 6: Typographical Conventions

    1 Typographical conventions Warning signs Danger This symbol is used when there may be danger to personnel if the instructions are ignored or not followed correctly! Caution This symbol is used when there may be damage to equipment or data if the instructions are ignored or not followed correctly! Caution This symbol is used where special care is required when handling components…

  • Page 7: Notes

    Please assist us in improving these operating instructions where necessary. For technical questions Service hotline: Phone: 0661 6003-300 or 0661 6003-653 Fax: 0661 6003-881300 or 0661 6003-881653 E-mail: Service@jumo.net…

  • Page 8: Identifying The Instrument Version

    3.1 Type description (1) Basic type 202723 JUMO ecoTRANS pH 03, Microprocessor/transmitter / switching device for pH value / redox voltage and temperature (2) Output I (pH value / redox voltage) Analog actual value output, freely programmable…

  • Page 9: Mounting

    4 Mounting °C ≤ 75 5 Electrical connection The choice of cable, the installation, and the electrical connection must conform to the requirements of VDE 0100 “Regulations on the Installation of High-Voltage Systems with Nominal Voltages below 1000 V” or the appropriate local regulations. To protect the device from static electrical discharge, the user must be electrostatically discharged before touching the instrument!

  • Page 10: Connecting The Sensor

    5.1 Connecting the sensor To connect the pH redox electrode, special coaxial lines are recommended as specified by JUMO datasheet 202990. Standard commercial antenna or computer lines with coaxial construction are generally not suitable. They may result in faulty readings or could destroy the pH or redox electrode.

  • Page 11: Customizing The Special Coaxial Line

    5.2 Customizing the special coaxial line Example: ~40 mm Strip insulation from the line as shown in the drawing. Remove the black semiconducting layer (2) from the insulation of the inner conductor as indicated. Twist the shield (1) and insulate with a shrink tubing. Insulate the black semiconducting layer with shrink tubing (2) to prevent short circuits.

  • Page 12
    Terminal assignment EXIT…
  • Page 13
    Measurement Terminal assignment Symbol inputs pH electrode Reference system (braiding) Redox Glass electrode/metal combination electrode (inner conductor) electrode pH glass electrode 13 Glass electrode/metal electrode (inner conductor) Metal electrode (with isolated reference electrode) Reference 16 Reference system electrode (with isolated electrodes) Liquid potential (connect only with…
  • Page 14
    Binary input Outputs Terminal assignment Symbol Analog process value output pH/ redox (galvanically isolated) Analog process value output — temperature (galvanically isolated) Relay Break (SPST-NC) Make (SPST-NO) Ö Power Terminal assignment Symbol supply Power supply (with reverse polarity protection) DC 20 to 30 V…
  • Page 15: Commissioning

    6 Commissioning 6.1 Brief description of the instrument The instrument measures and, depending on the configuration, regulates the pH or redox voltage in aqueous solutions. The range applications includes general water wastewater management, measuring drinking and wastewater, process water, surface and ocean water, swimming pool and fountain water, and professional aquariums, etc.

  • Page 16: Block Diagram

    Note: After the instrument is started, the output signal is 0 V or 0 mA. The relay is in idle state (inactive). After about 2 seconds, the JUMO ecoTRANS pH 03 begins working according to the configuration.

  • Page 17: Adjusting / Changing Instrument Functions

    7 Adjusting / changing instrument functions Changes can be made in the setup program or with the JUMO ecoTRANS pH 03 keys. 7.1 Process value display The process value display can be either — in static mode or in — alternating mode…

  • Page 18: Operation

    7.2 Operation The instrument can be operated on different levels. Access to all levels (except the user level) is protected by different codes On the User level (USER) all parameters can be viewed and/or modified depending on user rights (see Enable level). On the Calibration level (CALIB) the electrode zero point (intercept point) and/or the electrode slope can be calibrated.

  • Page 19
    — To increase or reduce the value, press the UP or DOWN key. — The value can then be accepted with the P key. — The EXIT key is used to cancel the entry and switch to the next higher level.
  • Page 20: Selecting Levels

    7.4 Selecting levels Process value display > 2 seconds EXIT Selection level EXIT USER (Operator level) EXIT ADMIN (Administrator level) EXIT RIGHT (Enable level) EXIT CALIB (Calibration level) Timeout (automatically jumps back after 60 sec. with no activity). See Section 7.5 «User level (USER)», page 21. See Section 7.6 «Administrator level (ADMIN)», page 22.

  • Page 21: User Level (User)

    7.5 User level (USER)

  • Page 22: Administrator Level (Admin)

    7.6 Administrator level (ADMIN)

  • Page 23: Enable Level (Right)

    7.7 Enable level (RIGHT)

  • Page 24: Calibration Level (Calib)

    7.8 Calibration level (CALIB) CALIB CODE See Section 9.5.1 «One-point calibration (zero point)», page 44 NULL See Section 9.5.2 «Two-point calibration (zero point and slope)», page 45 SLOPE The timeout function is not active during calibration! See Section 9 «Calibration», page 40.

  • Page 25: Configurable Parameters

    8 Configurable parameters Parameters can be set with the setup program or on the instrument. Since some parameters are independent of each other, it may be necessary when one parameter is changed to adjust others as well. Example: When the measurement variable is changed (from pH to redox voltage or vice versa) the display format, setpoints, and other parameters are adjusted.

  • Page 40: Calibration

    9 Calibration 9.1 General information The zero point (intercept) and slope of pH sensors stray from copy to copy. During the service life of the sensor, up to the time it must be replaced by a new sensor, the zero point and slope change. To allow for precise measurements, the transmitter must be adjusted (calibrated) to the current sensor parameters.

  • Page 41: Starting Calibration From The Calibration Level «Calib

    Calibration can be cancelled at any time with the EXIT key. Old calibration data will not be lost. Calibration is performed using the instrument keys. The zero point and slope of a sensor can also be entered manually. This should only be done in exceptional cases.

  • Page 42: Canceling Calibration And Error Messages

    9.3 Canceling calibration and error messages Calibration can be cancelled at any time with the EXIT key. Old calibration data will not be lost. During the calibration process, the transmitter calculates the electrode parameters zero point and if applicable slope. If the calculated values fall outside the permissible parameter limits, an error message is generated.

  • Page 43: Acknowledging Errors

    The parameter limits for NULL and SLOPE have been violated and/ or the two calibration points Ref.1 and Ref.2 are too close to each other. For two-point calibration of a pH sensor, the minimum spacing is 2 pH. For two-point calibration of a redox sensor, the minimum spacing is 2 mV.

  • Page 44: Selecting Two-Point Calibration

    9.4.2 Selecting two-point calibration Press the key and select CAL 2-PT. For more details see Section 9.5.2 «Two-point calibration (zero point and slope)», page 45. 9.5 Calibrating a pH measurement chain 9.5.1 One-point calibration (zero point) Confirm the selection with Display or editing option of the buffer temperature.

  • Page 45: Two-Point Calibration (Zero Point And Slope)

    Flashing Enter the actual buffer value with and confirm with The calculated zero point of the measurement chain appears. Accept the value with or press to cancel the calibration. EXIT The instrument goes into measurement mode. If an error message appears, see Section 9.3 «Canceling calibration and error messages», page 42.

  • Page 46
    Confirm the selection with Display and editing option of the buffer temperature. Confirm the selection with The first reference value is then measured. Immerse the measurement chain in the first buffer (for example pH 7.00) Wait until the measurement value has stabilized. Confirm the measurement value with Flashing Enter the actual buffer value with…
  • Page 47
    Remove the measurement chain from the first buffer and rinse with distilled water. Immerse the measurement chain in the second buffer (for example pH 4.00) Wait until the measurement value has stabilized. Confirm the measurement value with Flashing Enter the actual buffer value with confirm with The calculated zero point (top line) and the calculated slope (bottom line) appear.
  • Page 48: Ph Antimony Measurement Chain

    9.6 pH antimony measurement chain Antimony measurement chains are calibrated similarly to «normal» pH measurement chains, see Section 9.5 «Calibrating a pH measurement chain», page 44. 9.7 Redox measurement chain 9.7.1 General information The instrument offers two calibrating options for adjusting it to the redox measurement chain.

  • Page 49
    The reference value is then measured. Wait until the measurement value has stabilized. Confirm the measurement value with Flashing Enter the actual buffer value with confirm with The calculated zero point of the measurement chain appears. Accept the value with or press to cancel the calibration.
  • Page 50: Two-Point Calibration

    9.7.3 Two-point calibration In this type of calibration the display range can be scaled freely from 0 to 100%. Example: A span of -10 mV … +1000 mV can be scaled to 0 … 100% werden. The zero point can fall within the range of -999 …

  • Page 51
    Enter the desired value with (for example 20) and confirm with Remove the measurement chain from the first solution and rinse with distilled water. Immerse the measurement chain in the second solution (for example 295 mV). Wait until the measurement value has stabilized. Confirm the measurement value with Flashing Enter the desired value with…
  • Page 52: Analog Output

    10 Analog output Analog outputs are configured on the user level (USER) or on the administrator level (ADMIN) in PH.OUT redox output) TE.OUT (temperature output); see Section 7.5 «User level (USER)», page 21. 10.1 Behavior of the output signal during calibration Two options are possible, «Concurrent»…

  • Page 53: Output Signal In Case Of Error

    10.3 Output signal in case of error Depending on the configuration, the output signal may assume the «LOW» or «HIGH» state in case of error. Output signal Output signal Output signal nominal HIGH 0…20 mA 22.0 mA 0 mA 4…20 mA 22.0 mA 3.4 mA 0…10 V…

  • Page 54: Relay Output

    11 Relay output 11.1 Relay behavior Depending on the setting, the JUMO ecoTRANS pH 03 monitors a limit value. 11.2 Binary output 1 (submenu «BIN.1») Setting options = No function (default setting) = Window contact main value (active inside a window)

  • Page 55
    Contact function for settings 1 and 5 Hysteresis HYS.1. Distance DST.1 Switching point ALAR.1 Contact function for settings 2 and 6 Hysteresis HYS.1. Distance DST.1 Switching point ALAR.1…
  • Page 56
    Contact function for settings 3 and 7 Hysteresis HYS.1. Switching point ALAR.1 Contact function for settings 4 and 8 Hysteresis HYS.1. Switching point ALAR.1…
  • Page 57: Manual Mode Of The Relay Output

    11.3 Manual mode of the relay output For test purposes or when starting up systems, a constant signal can be generated by the transmitter. Manual mode can be set with parameter: USER / BIN.1 / SIM.1 to = inactive = active => LED «K1» is lit = No manual mode.

  • Page 58: Behavior Of The Relay During Calibration

    After «Supply voltage On» manual mode is always deactivated. 11.4 Behavior of the relay during calibration The behavior of the relay with the parameter: USER / BIN.1 / CAL.1 set to Relay inactive Relay active Relay unchanged (during calibration the relay status remains set to the status that was valid before the calibration process) 11.5 Pulse function of the relay output The limit comparator is reset after an adjustable «pulse time.»…

  • Page 59
    Triggering condition is longer than pulse duration Triggering condition Time Pulse contact Time Pulse duration T.PUL1 Triggering condition is shorter than pulse duration Triggering condition Time Pulse contact Time Pulse duration T.PUL1…
  • Page 60: Behavior Of The Relay In Case Of Error

    11.6 Behavior of the relay in case of error The behavior of the relay can be adjusted with the following parameters: USER / BIN.1 and ERR.1 set to Relay inactive Relay active Relay unchanged («frozen»: the relay status remains set to the status that was valid before the error) Function pH value…

  • Page 61
    Limit comparator of pH value/redox voltage with active temperature compensation — Underrange pH value/redox voltage — Overrange pH value/redox voltage — Underrange temperature — Overrange temperature Limit comparator — temperature — Underrange temperature — Overrange temperature Calibration timer — Timeout…
  • Page 62: Display And Led Messages

    12 Display and LED messages 12.1 Operating states of JUMO ecoTrans pH 03 Two LED indicate operating states Instrument status Red LED (top) Yellow LED (bottom) Normal mode On if LC1 is active Error Flashing On if LC1 is active Initialization 12.2 Underrange…

  • Page 63: Short-Circuit

    If no automatic temperature compensation or temperature measurement is required parameter SENS.T must be set accordingly; see Section 8.1.2 «Measurement input — temperature (submenu «TEMP»)», page 28. 12.5 Short-circuit 12.6 Initialization of dependent parameters After a parameter has been changed, other dependent parameters are automatically changed.

  • Page 64: Operation Via Setup Interface

    13 Operation via setup interface EXIT (1) JUMO ecoTRANS pH 03 (2) PC interface line (optional accessory) (3) JUMO PC setup software, multilingual D / GB / F (optional accessory) (4) PC or Notebook with USB interface ® ® operating system: Windows 2000…

  • Page 65: Operation With Pc Setup Program

    13.1 Operation with PC setup program (1) Navigation tree The navigation tree provides fast access (double click) to individual setting options. (2) Diagnostics window As soon as a connection is established with an instrument, current data is displayed here. (3) Working area Clicking the arrow ( ) shows possible settings.

  • Page 66: Technical Data

    14 Technical data Inputs Analog input 1 (pH / redox) — Electrodes — Glass or metal electrodes with isolated reference electrode — Antimony electrode Measurement ranges for pH / redox -2 … 16 pH or -1500 … +1500 mV Accuracy for pH / redox ±…

  • Page 67
    ≤ 400 Ω or 0(4) … 20mA load ≤ 400 Ω 20 … (4)0mA load galvanically isolated from inputs: ΔU ≤ 30V AC or ΔU ≤ 50 V DC Scope of scaling at least 10% of the scope of the measurement range deviation from the characteristic curve of the output signal ≤…
  • Page 68
    Supply DC 20 … 30 V, residual ripple<5%, power consumption≤ 4 W, with reverse polarity protection. Operates only on SELV or PELV circuits. Electrical connection Screw terminals up to 2.5 mm Operating temperature range 0 … +50°C Functional temperature range -10 ……
  • Page 69: Environment/Disposal

    Mounting On DIN rail 35mm x 7.5mm to DIN EN 60 715 Installation position Weight Approx. 150g 15 Environment/disposal Defective instruments can be sent to JUMO for proper disposal.

  • Page 72
    JUMO GmbH & Co. KG Street address: Moritz-Juchheim-Straße 1 36039 Fulda, Germany Delivery address: Mackenrodtstraße 14 36039 Fulda, Germany Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 Email: mail@jumo.net Internet: www.jumo.net JUMO Instrument Co. Ltd.

39

9 Calibration

9.1 General information

The zero point (intercept) and slope of pH sensors stray from copy to copy. During the service life of the sensor, up to the time it must be replaced by a new sensor, the zero point and slope change. To allow for precise measurements, the transmitter must be adjusted

(calibrated) to the current sensor parameters. The calibration is performed using buffer solutions.

Only the zero point needs to be calibrated for redox sensors.

9.1.1 When is calibration required?

The electrochemical sensors should be cleaned and the transmitter should be calibrated at regular intervals (depending on the measuring medium)!

With the correct wiring of device and sensor, and correct configuration of the device, a measurement without calibration should be possible.

If this is not the case, influencing factors such as short circuit, open circuit, EMC and flow conditions should also be considered.

9.2 Activating and starting calibration mode

During the calibration process, the display flashes.

The analog outputs respond as they were calibrated on the USER LEVEL / ANALOG OUTPUT x /

FOR CALIBRATION.

The response of the relay depends on the configuration of the switching output!

40

Calibration can be cancelled at any time with the EXIT key. Old calibration data will not be lost.

Calibration is performed using the instrument keys.

The zero point and slope of a sensor can also be entered manually. This should only be done in exceptional cases. Some sensors come with a test report that specifies the zero point and slope. These values simply provided documentation that the sensor was in proper condition upon delivery. Since these values change during storage, they are not suitable for manual entry. We always advise performing calibration with buffer solutions.

The timeout function is not active during calibration!

9.2.1 Starting calibration from the calibration level «CALIB»

Use the menu to switch to the calibration level (see

Section 7.4 «Selecting levels», page 20 and Section

7.8 «Calibration level (CALIB)», page 24).

P

Press and hold (for more than 2 sec) / then select CALIB.

P

Press and hold (for less than 1 sec) / then enter code 110.

✱ Confirm with

P

,

continue with Section 9.4 «One-point or two-point calibration», page 43.

9.2.2 Starting calibration with the hot key (rapid access)

Activation of the calibration mode by rapid access must be previously enabled:

P

> Press and hold for 2 seconds / Set ADMIN / CAL /

CA.LVL to 1.

✱ Press the

P

and keys,

continue with Section 9.4 «One-point or two-point calibration», page 43.

41

9.3 Canceling calibration and error messages

Calibration can be cancelled at any time with the EXIT key. Old calibration data will not be lost.

During the calibration process, the transmitter calculates the electrode parameters zero point and if applicable slope. If the calculated values fall outside the permissible parameter limits, an error message is generated.

For the permitted value ranges of parameters NULL and SLOPE see

Section 8.1.1 «Measurement input — main value (submenu «PH»)», page 26.

9.3.1 Zero point error

Parameter limits for NULL have been violated.

9.3.2 Slope error

Parameter limits for SLOPE have been violated.

9.3.3 General error during calibration

42

The parameter limits for NULL and SLOPE have been violated and/ or the two calibration points Ref.1 and Ref.2 are too close to each other.

For two-point calibration of a pH sensor, the minimum spacing is

2 pH.

For two-point calibration of a redox sensor, the minimum spacing is

2 mV.

The greater the spacing, the more accurate the calibration.

9.3.4 Acknowledging errors

✱ Press the

P

keys or EXIT.

The error message is deleted.

The faulty parameters NULL and/or SLOPE are not saved.

The instrument continues working with old calibration data.

9.3.5 Additional measures

✱ Check the quality and condition (age) of the buffer solutions.

✱ Check whether the sensor is dirty or needs to be replaced.

✱ Check whether the sensor plug is moist or the wiring is faulty.

✱ Observe the minimum spacings for buffer or reference values.

9.4 One-point or two-point calibration

For one-point calibration, the transmitter is only adjusted to the zero point of the sensor.

In two-point calibration the transmitter is adjusted to both the zero point of the sensor and its slope. We expressly recommend twopoint calibration!

9.4.1 Selecting one-point calibration

Select CAL I-PT.

For more details see Section 9.5.1 «One-point calibration (zero point)», page 44.

43

9.4.2 Selecting two-point calibration

Press the key and select CAL 2-PT.

For more details see Section 9.5.2 «Two-point calibration (zero point and slope)», page 45.

9.5 Calibrating a pH measurement chain

9.5.1 One-point calibration (zero point)

✱ Confirm the selection with

P

.

Display or editing option of the buffer temperature.

✱ Confirm the selection with

P

.

The first reference value is then measured.

✱ Wait until the measurement value has stabilized.

Confirm the measurement value with

P

.

44

Flashing

✱ Enter the actual buffer value with

or and confirm with

P

.

The calculated zero point of the measurement chain appears.

✱ Accept the value with

P

or press

EXIT

to cancel the calibration.

The instrument goes into measurement mode.

If an error message appears, see Section 9.3

«Canceling calibration and error messages», page 42.

9.5.2 Two-point calibration (zero point and slope)

Calibrate zero point and slope.

The buffer solutions (reference solutions) used for calibration must differ by at least 2 pH!

During the calibration, the temperature of the two buffer solutions must be identical and remain constant!

45

✱ Confirm the selection with

P

.

Display and editing option of the buffer temperature.

✱ Confirm the selection with

P

.

The first reference value is then measured.

✱ Immerse the measurement chain in the first buffer (for example pH 7.00)

✱ Wait until the measurement value has stabilized.

Confirm the measurement value with

P

.

Flashing

✱ Enter the actual buffer value with

or and confirm with

P

.

46

✱ Remove the measurement chain from the first buffer and rinse with distilled water.

✱ Immerse the measurement chain in the second buffer (for example pH 4.00)

✱ Wait until the measurement value has stabilized.

Confirm the measurement value with

P

.

Flashing

✱ Enter the actual buffer value with

or and confirm with

P

.

The calculated zero point (top line) and the calculated slope (bottom line) appear.

✱ Accept the values with

P

or press

EXIT

to cancel the calibration.

The instrument goes into measurement mode.

If an error message appears, see Section 9.3

«Canceling calibration and error messages», page 42.

47

9.6 pH antimony measurement chain

Antimony measurement chains are calibrated similarly to «normal»

pH measurement chains, see Section 9.5 «Calibrating a pH measurement chain», page 44.

9.7 Redox measurement chain

9.7.1 General information

The instrument offers two calibrating options for adjusting it to the redox measurement chain.

— One-point calibration

If «mV» was configured as UNIT.

— Two-point calibration

If «%» was configured as UNIT.

The display flashes during calibration.

The analog outputs respond the same way as on

USER LEVEL / ANALOG OUTPUT x / FOR

CALIBRATION.

The response of the relay depends on the configuration of the switching output!

)

9.7.2 One-point calibration (recommended calibration)

Calibration of the zero point.

✱ Confirm the selection with

P

.

✱ Rinse off the measurement chain with distilled water or clean if necessary (see operating instructions for electrode).

✱ Immerse the measurement chain in the test solution (e.g. 468 mV).

48

The reference value is then measured.

✱ Wait until the measurement value has stabilized.

✱ Confirm the measurement value with

P

.

Flashing

✱ Enter the actual buffer value with

or and confirm with

P

.

The calculated zero point of the measurement chain appears.

Accept the value with

P

or press

EXIT

to cancel the calibration.

The instrument goes into measurement mode.

If an error message appears, see Section 9.3

«Canceling calibration and error messages», page 42.

49

)

9.7.3 Two-point calibration

In this type of calibration the display range can be scaled freely from

0 to 100%.

Example:

A span of -10 mV … +1000 mV can be scaled to 0 … 100% werden.

The zero point can fall within the range of

-999 … +999 mV.

Two different reference fluids must be used for the calibration. The redox voltage of a measurement solution is not temperature-dependent!

Confirm the selection with

P

.

The first reference value is then measured.

✱ Immerse the measurement chain in the first solution (for example

59 mV)

✱ Wait until the measurement value has stabilized.

Confirm the measurement value with

P

.

Flashing

50

✱ Enter the desired value with

or (for example 20) and confirm with

P

.

✱ Remove the measurement chain from the first solution and rinse with distilled water.

✱ Immerse the measurement chain in the second solution (for example 295 mV).

✱ Wait until the measurement value has stabilized.

Confirm the measurement value with

P

.

Flashing

✱ Enter the desired value with

or (for example 80) and confirm with

P

.

✱ The calculated zero point (top line in mV) and the calculated slope (bottom line) appear.

Accept the values with

P

or press

EXIT

to cancel the calibration.

The instrument goes into measurement mode.

If an error message appears, see Section 9.3

«Canceling calibration and error messages», page 42.

51

10 Analog output

Analog outputs are configured on the user level

(USER) or on the administrator level (ADMIN) in

PH.OUT (pH or redox output) and TE.OUT

(temperature output); see Section 7.5 «User level

(USER)», page 21.

10.1 Behavior of the output signal during calibration

Two options are possible, «Concurrent» and «Unchanged»

(constant).

10.2 Behavior of the output signal in case of error

If one of the following errors occurs, the output signal assumes the

defined status (see Section 10.3 «Output signal in case of error», page 53):

Analog output of pH value / redox voltage with non-active temperature compensation

— Underrange pH value / redox voltage

— Overrange pH value / redox voltage

Analog output of pH value / redox voltage with active temperature compensation

— Underrange pH value / redox voltage

— Overrange pH value / redox voltage

— Underrange temperature

— Overrange temperature

Analog output — temperature

— Underrange temperature

— Overrange temperature

52

10.3 Output signal in case of error

Depending on the configuration, the output signal may assume the

«LOW» or «HIGH» state in case of error.

Output signal nominal

0…20 mA

4…20 mA

0…10 V

2…10 V

Output signal

HIGH

22.0 mA

22.0 mA

10.7 V

10.7 V

Output signal

LOW

0 mA

3.4 mA

0 V

1.4 V

10.4 Output signal when leaving the scaling range

When leaving the scaling range, the output returns a proportional signal up to a defined limit (in compliance with NAMUR NE43). The limits are listed in the table below:

Value below scaling range

0.0 mA

3.8 mA

0.0 V

20.5 mA

20.5 mA

10.2 V

1.8 V

10.2 V

In the scaling range

0…20 mA

4…20 mA

0…10 V

20…0 mA

20…4 mA

10…0 V

2…10 V

10…2 V

Scaling range was exceeded

20.5 mA

20.5 mA

10.2 V

0.0 mA

3.8 mA

0.0 V

10.2 V

1.8 V

10.5 Manual mode of the analog output

For test purposes or startup, a constant analog signal can be generated by the instrument

(see also Section 11.3 «Manual mode of the relay output», page

57).

After a power supply failure, manual mode is deactivated.

53

11 Relay output

11.1 Relay behavior

Depending on the setting, the JUMO ecoTRANS pH 03 monitors a limit value.

11.2 Binary output 1 (submenu «BIN.1»)

Setting options

0

= No function (default setting)

1 = Window contact main value (active inside a window)

2 = Window contact main value (active outside a window)

3 = Max contact main value (normally open (NO), similar to LK7)

4 = Min contact main value (normally closed (NC), similar to LK8)

5 = Window contact temperature (active inside a window)

6 = Window contact temperature (active outside a window)

7 = Max contact temperature (normally open (NO), similar to

LK7)

8 = Min contact temperature (normally closed (NC), similar to

LK8)

9 = Error output (every device error causes the relay to switch)

10 = Calibration timer elapsed

54

Contact function for settings 1 and 5

Hysteresis

HYS.1.

On

Off

Distance

DST.1

Switching point

ALAR.1

Contact function for settings 2 and 6

Hysteresis

HYS.1.

On pH

Off

Distance

DST.1

Switching point

ALAR.1

pH

55

Contact function for settings 3 and 7

Hysteresis

HYS.1.

On

Off

Switching point

ALAR.1

pH

Contact function for settings 4 and 8

Hysteresis

HYS.1.

On

Off

Switching point

ALAR.1

pH

56

11.3 Manual mode of the relay output

For test purposes or when starting up systems, a constant signal can be generated by the transmitter.

Manual mode can be set with parameter: USER / BIN.1 / SIM.1 to

0

1

2

= inactive

= active => LED «K1» is lit

= No manual mode.

Indication of manual mode

About 3

If one of the outputs is in manual mode, this is indicated in the change to process value display by an «S» or «-«.

S The relevant output is in manual mode.

The relevant output is not in manual mode.

1 2 3

1st place Analog output — pH/redox voltage

2nd place Analog output — temperature

3rd place Binary output 1

In the example above, the analog temperature output is in manual mode. No other outputs are in manual mode.

When the user exits manual mode, the output signal immediately assumes the value proportional to the pH value/redox voltage or the temperature process value.

57

After «Supply voltage On» manual mode is always deactivated.

11.4 Behavior of the relay during calibration

The behavior of the relay with the parameter:

USER / BIN.1 / CAL.1 set to

0 = Relay inactive

1 = Relay active

2 = Relay unchanged

(during calibration the relay status remains set to the status that was valid before the calibration process)

11.5 Pulse function of the relay output

The limit comparator is reset after an adjustable «pulse time.» The parameter for this is: USER / BIN.1 / T.PUL1 .

It can be set from 0 = 0 seconds (no pulse function) to 999 = 999 seconds.

LED «K1» is lit red as long as the switching condition is met.

No OFF delay is possible in pulse mode.

58

Triggering condition is longer than pulse duration

Triggering condition

On

Off

Time

Pulse contact

On

Off

Time

Pulse duration

T.PUL1

Triggering condition is shorter than pulse duration

Triggering condition

On

Off

Time

Pulse contact

On

Off

Time

Pulse duration

T.PUL1

59

11.6 Behavior of the relay in case of error

The behavior of the relay can be adjusted with the following parameters:

USER / BIN.1 and ERR.1 set to

0 = Relay inactive

1 = Relay active

2 = Relay unchanged

(«frozen»: the relay status remains set to the status that was valid before the error)

Function pH value redox voltage

Underrange

Overrange

Temperature

Underrange

Overrange

Limit comparator pH/redox voltage without temperature compensation

Limit comparator pH/redox voltage with temperature compensation

Limit comparator temperature

Calibration timer elapsed

Error output x x x x x x x x x x x x x x x x

11.7 Error detection

The relay output is active for the following errors:

Limit comparator for pH value/redox voltage with non-active temperature compensation

— Underrange pH value/redox voltage

— Overrange pH value/redox voltage

60

Limit comparator of pH value/redox voltage with active temperature compensation

— Underrange pH value/redox voltage

— Overrange pH value/redox voltage

— Underrange temperature

— Overrange temperature

Limit comparator — temperature

— Underrange temperature

— Overrange temperature

Calibration timer

— Timeout

61

12 Display and LED messages

12.1 Operating states of JUMO ecoTrans pH 03

Two LED indicate operating states

Instrument status Red LED (top)

Normal mode Off

Error

Initialization

Flashing

Off

12.2 Underrange

Yellow LED (bottom)

On if LC1 is active

On if LC1 is active

Off

Value is below lower measurement range limit (ORP means redox).

12.3 Overrange

Measurement range exceeded (ORP means redox).

12.4 Broken sensor

Broken sensor, wrong or no temperature sensor connected;

see Section 8.1.2 «Measurement input — temperature

(submenu «TEMP»)», page 28.

62

If no automatic temperature compensation or temperature measurement is required parameter SENS.T must be set

accordingly; see Section 8.1.2 «Measurement input — temperature (submenu «TEMP»)», page 28.

12.5 Short-circuit

12.6 Initialization of dependent parameters

After a parameter has been changed, other dependent parameters are automatically changed.

Please check all dependent parameters!

12.7 Calibration timer elapsed

Depending on specifications (for example of the equipment manufacturer), calibration of the measurement chain should be performed.

After a correct calibration, the calibration timer is reset and automatically restarted.

63

13 Operation via setup interface

(1)

(4)

(2)

P

K1

EXIT

(3)

(1) JUMO ecoTRANS pH 03

(2) PC interface line (optional accessory)

(3) JUMO PC setup software, multilingual D / GB / F (optional accessory)

(4) PC or Notebook with USB interface operating system: Windows 2000

®

,

NT

®

4.0 and up, Windows Vista

®

Windows XP

® or Windows

64

13.1 Operation with PC setup program

(1)

(3)

(2)

(1) Navigation tree

The navigation tree provides fast access (double click) to individual setting options.

(2) Diagnostics window

As soon as a connection is established with an instrument, current data is displayed here.

(3) Working area

Clicking the arrow ( ) shows possible settings.

Double clicking the text opens the appropriate editing window.

65

14 Technical data

Inputs

Analog input 1 (pH / redox)

— Electrodes

— Glass or metal electrodes with isolated reference electrode

— Antimony electrode

Measurement ranges for pH / redox

-2 … 16 pH or

-1500 … +1500 mV

Accuracy for pH / redox

± 1% of the measurement range

Analog input 2 (temperature)

— Resistance thermometer Pt100 or Pt1000

The temperature sensor can be connected in a 2-wire circuit.

Measurement display can be switched between °C / °F

Temperature offset for analog input 2

A process value can be corrected by an offset in the range from —

20 … +20°C.

Temperature measurement range

-10 … +150°C or 14 … 302°F

Deviation from characteristic temperature curve

for Pt 100 / Pt 1000:

1.5 K

Outputs

Two analog outputs:

Freely configurable:

0(2) … 10V

10 … (2)0V

R load

R load

2 k

Ω

or

2 k

Ω

or else

66

0(4) … 20mA R load

20 … (4)0mA R load

400

Ω or

400

Ω galvanically isolated from inputs:

Δ

U

30V AC or

Δ

U

50 V DC

Scope of scaling at least 10% of the scope of the measurement range

deviation from the characteristic curve of the output signal

0.075% of the measurement range

Relay output:

Switching contact switching capacity: 8 A, 250 V AC or 8 A, 24 V DC with resistive load contact life: > 100,000 operations at normal load

Key general parameters

A/D converter

14-bit resolution

Sampling time

500ms = 2 measurements / second

Effect of ambient temperature

0.6% / 10 K

Measuring circuit monitoring

Input 1(main value): out-of-range

Input 2 (temperature): out-of-range, sensor short circuit, broken sensor.

In case of error, the outputs assume a defined

(configurable) state.

Data backup

EEPROM

67

Supply

DC 20 … 30 V, residual ripple<5%, power consumption

4 W, with reverse polarity protection.

Operates only on SELV or PELV circuits.

Electrical connection

Screw terminals up to 2.5 mm

2

Operating temperature range

0 … +50°C

Functional temperature range

-10 … +60°C

Storage temperature range

-20 … +75°C

Climatic rating

Rel. humidity

75% no condensation

Enclosure protection (complies with EN 60 529)

IP 20

Electrical safety

Complies with EN 61 010 air gaps and creep zones for

— Overvoltage category II

— Pollution degree 2

Electromagnetic compatibility

Complies with EN 61 326

Interference immunity:To industrial requirements

Interference emission:Class B

Enclosure

DIN rail enclosure made of PC (polycarbonate)

68

Mounting

On DIN rail 35mm x 7.5mm to DIN EN 60 715

Installation position

Any

Weight

Approx. 150g

15 Environment/disposal

Defective instruments can be sent to JUMO for proper disposal.

69

70

JUMO GmbH & Co. KG

Street address:

Moritz-Juchheim-Straße 1

36039 Fulda, Germany

Delivery address:

Mackenrodtstraße 14

36039 Fulda, Germany

Postal address:

36035 Fulda, Germany

Phone: +49 661 6003-0

Fax: +49 661 6003-607

Email: [email protected]

Internet: www.jumo.net

JUMO Instrument Co. Ltd.

JUMO House

Temple Bank, Riverway

Harlow, Essex CM 20 2DY, UK

Phone: +44 1279 63 55 33

Fax: +44 1279 62 50 29

Email: [email protected]

Internet: www.jumo.co.uk

JUMO Process Control, Inc.

6733 Myers Road

East Syracuse, NY 13057, USA

Phone: +1 315 437 5866

Fax: +1 315 437 5860

Email: [email protected]

Internet: www.jumousa.com

39

9 Calibration

9.1 General information

The zero point (intercept) and slope of pH sensors stray from copy to copy. During the service life of the sensor, up to the time it must be replaced by a new sensor, the zero point and slope change. To allow for precise measurements, the transmitter must be adjusted

(calibrated) to the current sensor parameters. The calibration is performed using buffer solutions.

Only the zero point needs to be calibrated for redox sensors.

9.1.1 When is calibration required?

The electrochemical sensors should be cleaned and the transmitter should be calibrated at regular intervals (depending on the measuring medium)!

With the correct wiring of device and sensor, and correct configuration of the device, a measurement without calibration should be possible.

If this is not the case, influencing factors such as short circuit, open circuit, EMC and flow conditions should also be considered.

9.2 Activating and starting calibration mode

During the calibration process, the display flashes.

The analog outputs respond as they were calibrated on the USER LEVEL / ANALOG OUTPUT x /

FOR CALIBRATION.

The response of the relay depends on the configuration of the switching output!

40

Calibration can be cancelled at any time with the EXIT key. Old calibration data will not be lost.

Calibration is performed using the instrument keys.

The zero point and slope of a sensor can also be entered manually. This should only be done in exceptional cases. Some sensors come with a test report that specifies the zero point and slope. These values simply provided documentation that the sensor was in proper condition upon delivery. Since these values change during storage, they are not suitable for manual entry. We always advise performing calibration with buffer solutions.

The timeout function is not active during calibration!

9.2.1 Starting calibration from the calibration level «CALIB»

Use the menu to switch to the calibration level (see

Section 7.4 «Selecting levels», page 20 and Section

7.8 «Calibration level (CALIB)», page 24).

P

Press and hold (for more than 2 sec) / then select CALIB.

P

Press and hold (for less than 1 sec) / then enter code 110.

✱ Confirm with

P

,

continue with Section 9.4 «One-point or two-point calibration», page 43.

9.2.2 Starting calibration with the hot key (rapid access)

Activation of the calibration mode by rapid access must be previously enabled:

P

> Press and hold for 2 seconds / Set ADMIN / CAL /

CA.LVL to 1.

✱ Press the

P

and keys,

continue with Section 9.4 «One-point or two-point calibration», page 43.

41

9.3 Canceling calibration and error messages

Calibration can be cancelled at any time with the EXIT key. Old calibration data will not be lost.

During the calibration process, the transmitter calculates the electrode parameters zero point and if applicable slope. If the calculated values fall outside the permissible parameter limits, an error message is generated.

For the permitted value ranges of parameters NULL and SLOPE see

Section 8.1.1 «Measurement input — main value (submenu «PH»)», page 26.

9.3.1 Zero point error

Parameter limits for NULL have been violated.

9.3.2 Slope error

Parameter limits for SLOPE have been violated.

9.3.3 General error during calibration

42

The parameter limits for NULL and SLOPE have been violated and/ or the two calibration points Ref.1 and Ref.2 are too close to each other.

For two-point calibration of a pH sensor, the minimum spacing is

2 pH.

For two-point calibration of a redox sensor, the minimum spacing is

2 mV.

The greater the spacing, the more accurate the calibration.

9.3.4 Acknowledging errors

✱ Press the

P

keys or EXIT.

The error message is deleted.

The faulty parameters NULL and/or SLOPE are not saved.

The instrument continues working with old calibration data.

9.3.5 Additional measures

✱ Check the quality and condition (age) of the buffer solutions.

✱ Check whether the sensor is dirty or needs to be replaced.

✱ Check whether the sensor plug is moist or the wiring is faulty.

✱ Observe the minimum spacings for buffer or reference values.

9.4 One-point or two-point calibration

For one-point calibration, the transmitter is only adjusted to the zero point of the sensor.

In two-point calibration the transmitter is adjusted to both the zero point of the sensor and its slope. We expressly recommend twopoint calibration!

9.4.1 Selecting one-point calibration

Select CAL I-PT.

For more details see Section 9.5.1 «One-point calibration (zero point)», page 44.

43

9.4.2 Selecting two-point calibration

Press the key and select CAL 2-PT.

For more details see Section 9.5.2 «Two-point calibration (zero point and slope)», page 45.

9.5 Calibrating a pH measurement chain

9.5.1 One-point calibration (zero point)

✱ Confirm the selection with

P

.

Display or editing option of the buffer temperature.

✱ Confirm the selection with

P

.

The first reference value is then measured.

✱ Wait until the measurement value has stabilized.

Confirm the measurement value with

P

.

44

Flashing

✱ Enter the actual buffer value with

or and confirm with

P

.

The calculated zero point of the measurement chain appears.

✱ Accept the value with

P

or press

EXIT

to cancel the calibration.

The instrument goes into measurement mode.

If an error message appears, see Section 9.3

«Canceling calibration and error messages», page 42.

9.5.2 Two-point calibration (zero point and slope)

Calibrate zero point and slope.

The buffer solutions (reference solutions) used for calibration must differ by at least 2 pH!

During the calibration, the temperature of the two buffer solutions must be identical and remain constant!

45

✱ Confirm the selection with

P

.

Display and editing option of the buffer temperature.

✱ Confirm the selection with

P

.

The first reference value is then measured.

✱ Immerse the measurement chain in the first buffer (for example pH 7.00)

✱ Wait until the measurement value has stabilized.

Confirm the measurement value with

P

.

Flashing

✱ Enter the actual buffer value with

or and confirm with

P

.

46

✱ Remove the measurement chain from the first buffer and rinse with distilled water.

✱ Immerse the measurement chain in the second buffer (for example pH 4.00)

✱ Wait until the measurement value has stabilized.

Confirm the measurement value with

P

.

Flashing

✱ Enter the actual buffer value with

or and confirm with

P

.

The calculated zero point (top line) and the calculated slope (bottom line) appear.

✱ Accept the values with

P

or press

EXIT

to cancel the calibration.

The instrument goes into measurement mode.

If an error message appears, see Section 9.3

«Canceling calibration and error messages», page 42.

47

9.6 pH antimony measurement chain

Antimony measurement chains are calibrated similarly to «normal»

pH measurement chains, see Section 9.5 «Calibrating a pH measurement chain», page 44.

9.7 Redox measurement chain

9.7.1 General information

The instrument offers two calibrating options for adjusting it to the redox measurement chain.

— One-point calibration

If «mV» was configured as UNIT.

— Two-point calibration

If «%» was configured as UNIT.

The display flashes during calibration.

The analog outputs respond the same way as on

USER LEVEL / ANALOG OUTPUT x / FOR

CALIBRATION.

The response of the relay depends on the configuration of the switching output!

)

9.7.2 One-point calibration (recommended calibration)

Calibration of the zero point.

✱ Confirm the selection with

P

.

✱ Rinse off the measurement chain with distilled water or clean if necessary (see operating instructions for electrode).

✱ Immerse the measurement chain in the test solution (e.g. 468 mV).

48

The reference value is then measured.

✱ Wait until the measurement value has stabilized.

✱ Confirm the measurement value with

P

.

Flashing

✱ Enter the actual buffer value with

or and confirm with

P

.

The calculated zero point of the measurement chain appears.

Accept the value with

P

or press

EXIT

to cancel the calibration.

The instrument goes into measurement mode.

If an error message appears, see Section 9.3

«Canceling calibration and error messages», page 42.

49

)

9.7.3 Two-point calibration

In this type of calibration the display range can be scaled freely from

0 to 100%.

Example:

A span of -10 mV … +1000 mV can be scaled to 0 … 100% werden.

The zero point can fall within the range of

-999 … +999 mV.

Two different reference fluids must be used for the calibration. The redox voltage of a measurement solution is not temperature-dependent!

Confirm the selection with

P

.

The first reference value is then measured.

✱ Immerse the measurement chain in the first solution (for example

59 mV)

✱ Wait until the measurement value has stabilized.

Confirm the measurement value with

P

.

Flashing

50

✱ Enter the desired value with

or (for example 20) and confirm with

P

.

✱ Remove the measurement chain from the first solution and rinse with distilled water.

✱ Immerse the measurement chain in the second solution (for example 295 mV).

✱ Wait until the measurement value has stabilized.

Confirm the measurement value with

P

.

Flashing

✱ Enter the desired value with

or (for example 80) and confirm with

P

.

✱ The calculated zero point (top line in mV) and the calculated slope (bottom line) appear.

Accept the values with

P

or press

EXIT

to cancel the calibration.

The instrument goes into measurement mode.

If an error message appears, see Section 9.3

«Canceling calibration and error messages», page 42.

51

10 Analog output

Analog outputs are configured on the user level

(USER) or on the administrator level (ADMIN) in

PH.OUT (pH or redox output) and TE.OUT

(temperature output); see Section 7.5 «User level

(USER)», page 21.

10.1 Behavior of the output signal during calibration

Two options are possible, «Concurrent» and «Unchanged»

(constant).

10.2 Behavior of the output signal in case of error

If one of the following errors occurs, the output signal assumes the

defined status (see Section 10.3 «Output signal in case of error», page 53):

Analog output of pH value / redox voltage with non-active temperature compensation

— Underrange pH value / redox voltage

— Overrange pH value / redox voltage

Analog output of pH value / redox voltage with active temperature compensation

— Underrange pH value / redox voltage

— Overrange pH value / redox voltage

— Underrange temperature

— Overrange temperature

Analog output — temperature

— Underrange temperature

— Overrange temperature

52

10.3 Output signal in case of error

Depending on the configuration, the output signal may assume the

«LOW» or «HIGH» state in case of error.

Output signal nominal

0…20 mA

4…20 mA

0…10 V

2…10 V

Output signal

HIGH

22.0 mA

22.0 mA

10.7 V

10.7 V

Output signal

LOW

0 mA

3.4 mA

0 V

1.4 V

10.4 Output signal when leaving the scaling range

When leaving the scaling range, the output returns a proportional signal up to a defined limit (in compliance with NAMUR NE43). The limits are listed in the table below:

Value below scaling range

0.0 mA

3.8 mA

0.0 V

20.5 mA

20.5 mA

10.2 V

1.8 V

10.2 V

In the scaling range

0…20 mA

4…20 mA

0…10 V

20…0 mA

20…4 mA

10…0 V

2…10 V

10…2 V

Scaling range was exceeded

20.5 mA

20.5 mA

10.2 V

0.0 mA

3.8 mA

0.0 V

10.2 V

1.8 V

10.5 Manual mode of the analog output

For test purposes or startup, a constant analog signal can be generated by the instrument

(see also Section 11.3 «Manual mode of the relay output», page

57).

After a power supply failure, manual mode is deactivated.

53

11 Relay output

11.1 Relay behavior

Depending on the setting, the JUMO ecoTRANS pH 03 monitors a limit value.

11.2 Binary output 1 (submenu «BIN.1»)

Setting options

0

= No function (default setting)

1 = Window contact main value (active inside a window)

2 = Window contact main value (active outside a window)

3 = Max contact main value (normally open (NO), similar to LK7)

4 = Min contact main value (normally closed (NC), similar to LK8)

5 = Window contact temperature (active inside a window)

6 = Window contact temperature (active outside a window)

7 = Max contact temperature (normally open (NO), similar to

LK7)

8 = Min contact temperature (normally closed (NC), similar to

LK8)

9 = Error output (every device error causes the relay to switch)

10 = Calibration timer elapsed

54

Contact function for settings 1 and 5

Hysteresis

HYS.1.

On

Off

Distance

DST.1

Switching point

ALAR.1

Contact function for settings 2 and 6

Hysteresis

HYS.1.

On pH

Off

Distance

DST.1

Switching point

ALAR.1

pH

55

Contact function for settings 3 and 7

Hysteresis

HYS.1.

On

Off

Switching point

ALAR.1

pH

Contact function for settings 4 and 8

Hysteresis

HYS.1.

On

Off

Switching point

ALAR.1

pH

56

11.3 Manual mode of the relay output

For test purposes or when starting up systems, a constant signal can be generated by the transmitter.

Manual mode can be set with parameter: USER / BIN.1 / SIM.1 to

0

1

2

= inactive

= active => LED «K1» is lit

= No manual mode.

Indication of manual mode

About 3

If one of the outputs is in manual mode, this is indicated in the change to process value display by an «S» or «-«.

S The relevant output is in manual mode.

The relevant output is not in manual mode.

1 2 3

1st place Analog output — pH/redox voltage

2nd place Analog output — temperature

3rd place Binary output 1

In the example above, the analog temperature output is in manual mode. No other outputs are in manual mode.

When the user exits manual mode, the output signal immediately assumes the value proportional to the pH value/redox voltage or the temperature process value.

57

After «Supply voltage On» manual mode is always deactivated.

11.4 Behavior of the relay during calibration

The behavior of the relay with the parameter:

USER / BIN.1 / CAL.1 set to

0 = Relay inactive

1 = Relay active

2 = Relay unchanged

(during calibration the relay status remains set to the status that was valid before the calibration process)

11.5 Pulse function of the relay output

The limit comparator is reset after an adjustable «pulse time.» The parameter for this is: USER / BIN.1 / T.PUL1 .

It can be set from 0 = 0 seconds (no pulse function) to 999 = 999 seconds.

LED «K1» is lit red as long as the switching condition is met.

No OFF delay is possible in pulse mode.

58

Triggering condition is longer than pulse duration

Triggering condition

On

Off

Time

Pulse contact

On

Off

Time

Pulse duration

T.PUL1

Triggering condition is shorter than pulse duration

Triggering condition

On

Off

Time

Pulse contact

On

Off

Time

Pulse duration

T.PUL1

59

11.6 Behavior of the relay in case of error

The behavior of the relay can be adjusted with the following parameters:

USER / BIN.1 and ERR.1 set to

0 = Relay inactive

1 = Relay active

2 = Relay unchanged

(«frozen»: the relay status remains set to the status that was valid before the error)

Function pH value redox voltage

Underrange

Overrange

Temperature

Underrange

Overrange

Limit comparator pH/redox voltage without temperature compensation

Limit comparator pH/redox voltage with temperature compensation

Limit comparator temperature

Calibration timer elapsed

Error output x x x x x x x x x x x x x x x x

11.7 Error detection

The relay output is active for the following errors:

Limit comparator for pH value/redox voltage with non-active temperature compensation

— Underrange pH value/redox voltage

— Overrange pH value/redox voltage

60

Limit comparator of pH value/redox voltage with active temperature compensation

— Underrange pH value/redox voltage

— Overrange pH value/redox voltage

— Underrange temperature

— Overrange temperature

Limit comparator — temperature

— Underrange temperature

— Overrange temperature

Calibration timer

— Timeout

61

12 Display and LED messages

12.1 Operating states of JUMO ecoTrans pH 03

Two LED indicate operating states

Instrument status Red LED (top)

Normal mode Off

Error

Initialization

Flashing

Off

12.2 Underrange

Yellow LED (bottom)

On if LC1 is active

On if LC1 is active

Off

Value is below lower measurement range limit (ORP means redox).

12.3 Overrange

Measurement range exceeded (ORP means redox).

12.4 Broken sensor

Broken sensor, wrong or no temperature sensor connected;

see Section 8.1.2 «Measurement input — temperature

(submenu «TEMP»)», page 28.

62

If no automatic temperature compensation or temperature measurement is required parameter SENS.T must be set

accordingly; see Section 8.1.2 «Measurement input — temperature (submenu «TEMP»)», page 28.

12.5 Short-circuit

12.6 Initialization of dependent parameters

After a parameter has been changed, other dependent parameters are automatically changed.

Please check all dependent parameters!

12.7 Calibration timer elapsed

Depending on specifications (for example of the equipment manufacturer), calibration of the measurement chain should be performed.

After a correct calibration, the calibration timer is reset and automatically restarted.

63

13 Operation via setup interface

(1)

(4)

(2)

P

K1

EXIT

(3)

(1) JUMO ecoTRANS pH 03

(2) PC interface line (optional accessory)

(3) JUMO PC setup software, multilingual D / GB / F (optional accessory)

(4) PC or Notebook with USB interface operating system: Windows 2000

®

,

NT

®

4.0 and up, Windows Vista

®

Windows XP

® or Windows

64

13.1 Operation with PC setup program

(1)

(3)

(2)

(1) Navigation tree

The navigation tree provides fast access (double click) to individual setting options.

(2) Diagnostics window

As soon as a connection is established with an instrument, current data is displayed here.

(3) Working area

Clicking the arrow ( ) shows possible settings.

Double clicking the text opens the appropriate editing window.

65

14 Technical data

Inputs

Analog input 1 (pH / redox)

— Electrodes

— Glass or metal electrodes with isolated reference electrode

— Antimony electrode

Measurement ranges for pH / redox

-2 … 16 pH or

-1500 … +1500 mV

Accuracy for pH / redox

± 1% of the measurement range

Analog input 2 (temperature)

— Resistance thermometer Pt100 or Pt1000

The temperature sensor can be connected in a 2-wire circuit.

Measurement display can be switched between °C / °F

Temperature offset for analog input 2

A process value can be corrected by an offset in the range from —

20 … +20°C.

Temperature measurement range

-10 … +150°C or 14 … 302°F

Deviation from characteristic temperature curve

for Pt 100 / Pt 1000:

1.5 K

Outputs

Two analog outputs:

Freely configurable:

0(2) … 10V

10 … (2)0V

R load

R load

2 k

Ω

or

2 k

Ω

or else

66

0(4) … 20mA R load

20 … (4)0mA R load

400

Ω or

400

Ω galvanically isolated from inputs:

Δ

U

30V AC or

Δ

U

50 V DC

Scope of scaling at least 10% of the scope of the measurement range

deviation from the characteristic curve of the output signal

0.075% of the measurement range

Relay output:

Switching contact switching capacity: 8 A, 250 V AC or 8 A, 24 V DC with resistive load contact life: > 100,000 operations at normal load

Key general parameters

A/D converter

14-bit resolution

Sampling time

500ms = 2 measurements / second

Effect of ambient temperature

0.6% / 10 K

Measuring circuit monitoring

Input 1(main value): out-of-range

Input 2 (temperature): out-of-range, sensor short circuit, broken sensor.

In case of error, the outputs assume a defined

(configurable) state.

Data backup

EEPROM

67

Supply

DC 20 … 30 V, residual ripple<5%, power consumption

4 W, with reverse polarity protection.

Operates only on SELV or PELV circuits.

Electrical connection

Screw terminals up to 2.5 mm

2

Operating temperature range

0 … +50°C

Functional temperature range

-10 … +60°C

Storage temperature range

-20 … +75°C

Climatic rating

Rel. humidity

75% no condensation

Enclosure protection (complies with EN 60 529)

IP 20

Electrical safety

Complies with EN 61 010 air gaps and creep zones for

— Overvoltage category II

— Pollution degree 2

Electromagnetic compatibility

Complies with EN 61 326

Interference immunity:To industrial requirements

Interference emission:Class B

Enclosure

DIN rail enclosure made of PC (polycarbonate)

68

Mounting

On DIN rail 35mm x 7.5mm to DIN EN 60 715

Installation position

Any

Weight

Approx. 150g

15 Environment/disposal

Defective instruments can be sent to JUMO for proper disposal.

69

70

JUMO GmbH & Co. KG

Street address:

Moritz-Juchheim-Straße 1

36039 Fulda, Germany

Delivery address:

Mackenrodtstraße 14

36039 Fulda, Germany

Postal address:

36035 Fulda, Germany

Phone: +49 661 6003-0

Fax: +49 661 6003-607

Email: [email protected]

Internet: www.jumo.net

JUMO Instrument Co. Ltd.

JUMO House

Temple Bank, Riverway

Harlow, Essex CM 20 2DY, UK

Phone: +44 1279 63 55 33

Fax: +44 1279 62 50 29

Email: [email protected]

Internet: www.jumo.co.uk

JUMO Process Control, Inc.

6733 Myers Road

East Syracuse, NY 13057, USA

Phone: +1 315 437 5866

Fax: +1 315 437 5860

Email: [email protected]

Internet: www.jumousa.com

  • Возможность переключения с рН на мВ (редокс-потенциал)
  • Простое подключение сенсоров с помощью винтовых зажимов
  • Асимметричное и симметричное подключение рН-метрических цепей
  • Релейный переключающий контакт
  • Возможен контроль температуры среды

Описание

Прибор измеряет и регулирует – в зависимости от конфигурации – величину pH или редокс-потенциал в водных растворах. Типичные области применения – общая водоподготовка, измерения в питьевой, сточной и технологической воде, измерения в поверхностной и морской воде, бассейны, профессиональная аквариумистика, различные технологические процессы.

Измерительный преобразователь имеет два аналоговых входа. Первый аналоговый вход (основной вход для величины pH или редокс-потенциала) предусмотрен для подключения комбинированных или раздельных электродов. К прибору можно также подключать сурьмяные электроды. Ко второму аналоговому входу может подключаться термометр сопротивления Pt100 или Pt1000.

В распоряжении также имеются до двух аналоговых выходов и одно реле — переключающий контакт. Аналоговые выходы гальванически развязаны и поставлены в соответствие входам. Релейному контакту может быть поставлена в соответствие основная измеряемая величина (величина pH или редокс-потенциал) или температура.

Настройка и конфигурация прибора осуществляется с помощью кнопок и интегрированного ЖК-дисплея. Альтернативно настройку и конфигурацию можно очень удобно проводить с помощью ноутбука или ПК, подключенного к setup-интерфейсу прибора, и русифицированной setup-программы (опция). С помощью setup-программы можно также делать распечатку данных конфигурации; это облегчает процесс документирования.

Приборы поставляются с калибровочным сертификатом, в котором отражены информация о приборе и данные настройки.

Особенности

  • Возможность переключения с рН на мВ (редокс-потенциал)
  • Простое подключение сенсоров с помощью винтовых зажимов
  • асимметричное и симметричное подключение рН-метрических цепей
  • Два гальванически изолированных аналоговых выхода 0(4)… 20 мА / 0(2)… 10 В, произвольно конфигурируемых как выход действительного значения рН, редокс-потенциала или температуры
  • Релейный переключающий контакт
  • Возможен контроль температуры среды
  • Простая процедура калибровки, сопровождаемая подсказками
  • Тройная гальваническая развязка (вход, выход и напряжение питания гальванически изолированы друг от друга)
  • Для монтажа на DIN-рейку
  • Таймер калибровки
  • Калибровочный сертификат в стандартном объеме поставки

  • Аналогичные товары

Вся информация на сайте о товарах и ценах носит справочный характер и не является публичной офертой. Производитель оставляет за собой право изменять характеристики товара, его внешний вид и комплектность без предварительного уведомления продавца

Content

9

Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40

9.1

General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40

9.2

9.3

9.3.1 Zero point error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42

9.3.2 Slope error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42

9.3.5 Additional measures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43

9.4

9.5

9.6

9.7

9.7.1 General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48

10

Analog output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52

10.1

10.2

10.3

10.4

10.5

11

Relay output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54

11.1

Relay behavior . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54

11.2

11.3

11.4

11.5

11.6

11.7

Error detection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60

. . . . . . . . . . . . . . . . . . . 42

. . . . . . . . . . . . . . . . . . . . . . . . . 43

. . . . . . . . . . . . . . . . . . . . . . . 44

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53

Понравилась статья? Поделить с друзьями:
  • Jump starter jstar one инструкция на русском
  • Jumo logoscreen 600 инструкция на русском
  • Jsd 520 автомагнитола инструкция на русском
  • Jracdrive jac580b инструкция на русском
  • Jr t03s инструкция на русском