Measurement and Calibration

For better accuracy, frequent calibration of the meter is recommended. In

addition, the instrument must be recalibrated:

• When high accuracy is required.

• At least once a month.

Measurement

Press the ON/OFF button to turn the meter ON. Place the probe to the

desired solution. The conductivity value, automatically compensated for

temperature, will be shown on the LCD.

The measured temperature will be shown on the secondary LCD.

Note: Before taking any measurement make sure the meter has been

calibrated.

Calibration Procedure

• Press CAL button to enter calibration mode from measurement mode.

• The meter will enter the calibration mode, displaying «mS/cm 12.88

USE» with CAL tag blinking.

1. Pour 2″ (5 cm) of standard solution into a container.

2. Place the probe in calibration solution. The probe tip should be centered

in the solution and submersed 1.18″ (3 cm). The meter will automatically

recognize the solution.

3. If the solution is not recognized or is out of the accepted range «—- Err»

is displayed.

4. If the calibration solution is recognized «REC» is displayed until the

reading is stable and the calibration is accepted.

• After acceptance, the «Stor» message is displayed and meter returns to

measurement mode.

Clear Calibration

Place meter in calibration mode. Press ON/OFF and «Clr» is displayed. The

meter will now be at default calibration.

Error messages

During user calibration, if the reading is out of the accepted range, the meter

will display «—- Err».i.e. 20.00 mS/cm blinking means the limit of the

measured value is out of range.

If the measured temperature is higher than 50.0 ºC or lower than 0.0 ºC, the

50.0 ºC or 0.0 ºC temperature value will blink on the LCD in measure mode.

Care and Maintenance

To obtain the highest accuracy for measurement it is important to follow

these tips:

• Calibration is only as good as the solution being used. The calibration

solution values change over time. Fresh solution should be used for each

calibration.

• The probe should be rinsed with purified water each time before placing

in calibration solution or sample to be tested.

Battery Replacement

The meter features a low battery indicator. When the battery is running low

(under 10%), the battery indicator will blink on the LCD. When the battery

is discharged «dEAd bAtt» will be displayed on the LCD for 2 seconds and

the meter will turn off.

To change the CR2032 Li-ion battery, turn the battery cover located on the

back of the meter counterclockwise to unlock. Remove cover and replace with

new battery + side facing up.

Note: Batteries should only be replaced in a safe area using the

battery type specified in this instruction manual. Old batteries should

be disposed in accordance with local regulations.

Recommendations for Users

Before using Hanna Instruments products, make sure that they are entirely

suitable for your specific application and for the environment in which they

are used. Any variation introduced by the user to the supplied equipment

may degrade the meter’ s performance. Avoid touching the electrode area.

To avoid damages or burns, do not put the meter in microwave oven. For

yours and the meter’s safety do not use or store the meter in hazardous

environments.

Accessories

Solutions

Code

Description

12.88 mS/cm calibration solution,

HI70030P

20 mL sachets (25 pcs.)

12.88 mS/cm calibration solution,

HI7030M

230 mL bottle

12.88 mS/cm calibration solution,

HI7030L

500 mL bottle

Warranty

These meters are warranted for a period of one year against defects in

workmanship and materials when used for their intended purpose and

maintained according to instructions. Probes are warranted for a period

of six months. This warranty is limited to repair or replacement free of

charge. Damage due to accidents, misuse, tampering or lack of prescribed

maintenance is not covered. If service is required, contact your local Hanna

Instruments Office. If under warranty, report the model number, date of

purchase, serial number and the nature of the problem. If the repair is not

covered by the warranty, you will be notified of the charges incurred. If the

instrument is to be returned to Hanna Instruments, first obtain a Returned

Goods Authorization (RGA) number from the Technical Service department

and then send it with shipping costs prepaid. When shipping any instrument,

make sure it is properly packaged for complete protection.

IST98304N 06/18

INSTRUCTION MANUAL

DiST 4

HI98304

EC Tester

Тестеры серии DiST представляют собой прочные и надежные карманные тестеры для получения быстрых и точных показаний электрической проводимости. Серия тестеров DiST® широко используется для мониторинга электрической проводимости питьевой воды, в процессах водоподготовки, обратном осмосе, градирнях, сточных водах, лабораториях, сельском хозяйстве, аквакультуре и аквариумах, гидропонике и полиграфической промышленности. Эти тестеры имеют неокисляемый амперометрический графитовый электрод, обеспечивающий лучшую воспроизводимость при измерениях. Амперометрическое измерение электрической проводимости основано на применении закона Ома, I = V / R, где R зависит от расстояния между двумя электродами и величины их поверхности. Окисление электродов изменяют и расстояние, и величину поверхности, что влияет на точность измерений. Неокисляемые графитовые электроды тестеров DiSTs обеспечивают оптимальную поверхность для получения точных и надежных результатов. Калибровка автоматическая.

OPERATION:

• Remove the protective cap.

• Turn the DiST

DiST

DiST

DiST

DiST on.

• Immerse it in the solution to be tested without

exceeding the maximum immersion level.

• Stir gently and wait for the reading

to stabilize.
DiST

DiST

DiST

DiST

DiST automatically compensates for
temperature variations.

• The reading on display is directly expressed in:

ppm (mg/L) unit for DiST1

DiST1

DiST1

DiST1

DiST1 (

HI98300 

HI98300 

HI98300 

HI98300 

HI98300 & HI98301

 HI98301

 HI98301

 HI98301

 HI98301

),

ppt (g/L) unit for DiST2

DiST2

DiST2

DiST2

DiST2 (HI98302

HI98302

HI98302

HI98302

HI98302),

µS/cm for DiST3

DiST3

DiST3

DiST3

DiST3 (HI98303

HI98303

HI98303

HI98303

HI98303) and

mS/cm for DiST4

DiST4

DiST4

DiST4

DiST4 (HI98304

HI98304

HI98304

HI98304

HI98304).

CALIBRATION:

• Immerse the DiST 

DiST 

DiST 

DiST 

DiST in the indicated calibration

solution (see Specifications), without exceeding
the maximum immersion level.

• Allow the reading to stabilize and with a small

screwdriver turn the calibration trimmer until the
display shows the required solution value.

E.g. for calibrating HI98301

HI98301

HI98301

HI98301

HI98301 the correct calibra-

tion solution is HI70032

HI70032

HI70032

HI70032

HI70032, and the trimmer should

be turned until the display shows «

1382

«.

BEPS AND BATTERY REPLACEMENT:

DiST 

DiST 

DiST 

DiST 

DiST is supplied with BEPS (Battery Error Preventing
System) which avoids any erroneous reading due to
low batteries level, by switching the meter off.
To replace the batteries, slide off the battery com-
partment cover and replace all four 1.5V batteries
while paying attention to their polarity.

Batteries should only be replaced in a nonhazardous
area using the battery type specified in this instruction
sheet.

DiST

HI98300/1/2/3/4

Conductivity/TDS Meters

with Automatic Temperature
Compensation

Manufacturers since 1978

Visit our Internet Home Page:

http://www.hannainst.com

ACCESSORIES:

HI 70030P

12880 µS/cm calibration kit
(25 x 20 mL sachets)

HI 70031P

1413 µS/cm calibration kit
(25 x 20 mL sachets)

HI 70032P

1382 ppm (mg/L) calibration kit
(25 x 20 mL sachets)

HI 70038P

6.44 ppt (g/L) calibration kit
(25 x 20 mL sachets)

HI 70039P

5000 µS/cm calibration kit
(25 x 20 mL sachets)

HI 70442P

1500 ppm calibration kit
(25 x 20 mL sachets)

HI 73301

Spare electrode for DiST1

HI 73302

Spare electrode for DiST2

HI 73303

Spare electrode for DiST3

HI 73304

Spare electrode for DiST4

HI 731326

Calibration screwdriver (20 pcs)


3.0

Rated 3 out of 5

3 out of 5 stars (based on 1 review)

Your overall rating

HANNA INSTRUMENTS DIST 4 HI98304 (01) PDF MANUAL

Click here to download HANNA INSTRUMENTS DIST 4 HI98304 (01) PDF MANUAL


HANNA INSTRUMENTS DIST 4 HI98304 (01) PDF MANUAL

FREE ENGLISH PDF

OPERATING INSTRUCTIONS

USER GUIDE – USER MANUAL

OWNER GUIDE – OWNER MANUAL

REFERENCE GUIDE – REFERENCE MANUAL

INSTRUCTION GUIDE – INSTRUCTION MANUAL


Your overall rating

  • Twitter
  • YouTube
  • Reddit
  • Facebook
  • Instagram
  • Pinterest

HANNA INSTRUMENTS DIST 4 HI98304 (01) PDF MANUAL

HANNA INSTRUMENTS DIST 4 HI98304 (01) PDF MANUAL

Thank You

Thank you for choosing a Hanna Instruments product. Please read

this instruction manual carefully before using the instrument.

For more information about Hanna Instruments and our products,

visit www.hannainst.com or e-mail us at sales@hannainst.com.

For technical support, contact your local Hanna Instruments

Office or e-mail us at tech@hannainst.com

Find your local Hanna Instruments Office on www.hannainst.com

Preliminary Examination

Remove the meter from the packing material and examine it carefully to

make sure that no damage has occurred during shipment. If noticeable

damage is evident, contact your local Hanna Instruments Office.

Each meter is supplied with:

• CR2032 battery (1 pc.)

• Storage / Protection sleeve

• 12.88 mS/cm calibration solution sachet (4 pcs.)

• Instruction manual

• Quality Certificate

Note: Save all packing material until you are sure that the instrument

functions correctly. All defective items must be returned in the original

packaging with the supplied accessories.

Preparation

The probe is shipped dry. Before using the electrode, remove the protective

cap. Then follow the calibration procedure.

• Turn the tester on by pressing ON/OFF button.

• Remove the protective cap and immerse the tip of the probe in the

sample to be tested.

• Stir gently and wait for a stability tag to disappear.

• The electrode automatically compensates for temperature variations.

• The reading on display is directly expressed in: mS/cm.

• For best results, recalibrate periodically.

• After use rinse the probe with water.

• Always replace the protective cap after each use.

Operation

mS/cm

Liquid Crystal

Display

CAL

CAL button

ON/OFF button

DiST 4

Recommended

Immersion

Level

Probe

Temperature

Sensor (inside)

All rights are reserved. Reproduction in whole or in part is prohibited

without the written consent of the copyright owner.

Specifications

0.00 to 20.00 mS/cm

Range

0.0 to 50.0 ˚C (32.0 to 122.0 ˚F)

0.01 mS/cm

Resolution

0.1 ˚C (0.1 ˚F)

Accuracy

±2% F.S.

(@25 °C/77 °F)

±0.5 °C (±1.0 °F)

Calibration

HI70030

Solutions

(12.88 mS/cm)

Calibration

automatic, single point (12.88 mS/cm)

Temperature Compensation automatic, 0.0 to 50.0 ºC (32.0 to 122.0 ºF)

Battery type

CR2032 Li-Ion 3V (1 pc.)

Battery life

approximately 250 hours of continuous use

0 to 50 ºC (32 to 122 ºF);

Environment

RH 100% max

Weight

68 g (2.4 oz.)

(without battery)

Hanna Instruments reserves the right to modify the design,

construction, or appearance of its products without advance notice.

Operational Guide

To turn the meter ON and check the battery status

Press the ON/OFF button to turn the meter on. At start-up, all the LCD

segments are displayed for 1 second, then the percent indication of the

remaining battery life is displayed for another second. The meter then enters

the normal measuring mode.

Note: Keeping the ON/OFF button pressed while turning the meter on

will display all LCD segments as long as the button is pressed.

To enter calibration mode

Press the CAL button. «CAL» message is displayed.

To enter setup mode

Remove the battery cover and press the Setup button located on the side

of the battery.

Meter Setup

While in measurement mode, remove the battery cover. Press

the Setup button located on the side of the battery in the battery

compartment. The meter will enter in setup mode. Press the

ON/OFF button to move through setup parameters. Change option by

pressing CAL button. The default settings are: «Set t» measure unit — °C,

«AOFF» — 8 min.

Setup

Button

To select the temperature unit (°C/°F)

To select the temperature unit when «SET t» is displayed press the CAL

button to change between °C or °F.

To select the Auto-Off time

To select the AUTO OFF when «AOFF» is displayed press the CAL button to

change between 8 min, 60 min or — (disabled).

To return to measurement mode

Press the ON/OFF button.