Summary of Contents for Enotec OXITEC 5000 Series

  • Page 1
    Installation and Operation Manual Analyzer System ® ® OXITEC 5000 / OXITEC 5000+ Version 33 for Software Version: 4.13 Doc.-ID: OXI-11022020-EN…
  • Page 2
    ENOTEC. This leads to a considerably increased service life compared to «glued or cemented» sensors that have a tendency to leak or crack during operation. All ENOTEC instruments are thoroughly tested in the factory and are subject to a strict ISO 9001 and internal quality assurance ®…
  • Page 3: Table Of Contents

    Automatic Calibration (ACAL) ……24   Warranty     4.2.9   Automatic Calibration Settings ……24       4.2.10 ENOTEC REMOTE ……….. 25 Declaration of Conformity         4.2.11 Measuring units ……….25 Index   4.2.12  …

  • Page 4: System Description

    System Description ® Installation and Operation Manual OXITEC 5000 System Description System overview Info Depending on the system configuration ordered by the customer, the electronic unit and / or probe may vary. Refer to chapters B and B.2 for different versions of electronic units and probes. ®…

  • Page 5
    System Description ® Installation and Operation Manual OXITEC 5000 ® Figure 2 — OXITEC 5000 (probe with cooling tube for flue gas temperatures of up to 1600°C) Safe Area — max. ambient temperature: Electronic unit SME-5 / IP66 -20 °C to +55 °C (-4 °F to +131 °F) Safe Area — max.
  • Page 6: Measuring Principle

    Storage instructions ENOTEC equipment and spares are to be stored in a dry and ventilated environment at temperatures between -40 °C to +80 °C (-40 °F to 176 °F). Paint fumes, silicone sprays, etc. must be avoided in the storage environment.

  • Page 7: Name Plates

    System Description ® Installation and Operation Manual OXITEC 5000 Name plates The name plate contains information about the line voltage, the nominated current, frequency, protection class, year of manufacture, serial number, order number and system order code. The system order code refers to information which is detailed in the system test report and supplied with the system. SME-53 Electronic Unit Probe…

  • Page 8: Installation

    Installation ® Installation and Operation Manual OXITEC 5000 Installation Warning The system is not equipped with an external power-off switch. The line voltage switch/fuse/breaker must be installed and be in accordance with local technical standards and should be near to the electronic unit and must be clearly marked as such.

  • Page 9: Installation Of Probe Signal Cable Fep-0001

    Installation ® Installation and Operation Manual OXITEC 5000 Installation of probe signal cable FEP-0001 Abide by the maximum cable length (max. 150m) Note the minimum bending radius. FEP-0001  R = 96 mm +50 °C +90 °C Max. Max. Temperature during Temperature during operation -5 °C -40 °C…

  • Page 10: Access To The Terminals

    Installation ® Installation and Operation Manual OXITEC 5000 Access to the terminals Warning Before removing the terminal covers, switch off the mains voltage to the system. Switch the mains voltage on only after attaching the terminal cover. After the installation has been completed, live parts may no longer be accessible.

  • Page 11: Wiring Diagram Of The Electronic Unit

    Installation ® Installation and Operation Manual OXITEC 5000 Wiring diagram of the electronic unit Figure 7 — Wiring diagram of the Electronic Unit Ferrite sleeves (enclosed) Analogue output cable (customer) Power supply cable (customer) Status signal cable (customer) Probe signal cable Pressure transmitter analogue input cable (optional, customer) Power supply…

  • Page 12: Oxitec® 5000 Wiring Diagram

    Installation ® Installation and Operation Manual OXITEC 5000 OXITEC® 5000 Wiring diagram Figure 8 — Wiring diagram of the OXITEC 5000 Analyzer system Doc.-ID: OXI-11022020-EN…

  • Page 13: Installation Of The Probe

    For probe lengths exceeding 2000 mm, a support must be mounted inside the duct (every 2 m) to prevent the probe and mounting tube from flexing or bending. ENOTEC recommends installing the probe horizontally (- 1 ° to — 3 °) for the fastest possible response time.

  • Page 14: Mounting Of The Probe

    Installation ® Installation and Operation Manual OXITEC 5000 Mounting of the probe Figure 9 — Mounting of the O probe – for dimensions, see section B Counter flange which has been welded gas tight at correct Probe connection box angle to duct Flange gasket Duct wall Protection tube…

  • Page 15: Mounting A Probe With Cooling Protection Tube

    Installation ® Installation and Operation Manual OXITEC 5000 Mounting a probe with cooling protection tube Info If a cooling tube is used, the part of the protection tube projecting from the duct wall must be insulated or heated if necessary, to prevent its temperature from dropping below the dew point. Make sure that the gas outlet is not blocked.

  • Page 16: Electrical Connections Of The Probe

    Installation ® Installation and Operation Manual OXITEC 5000 2.11 Electrical connections of the probe Info The shield of the probe cable must only be connected at the electronic housing at the PE terminal. Under no circumstance should the shield also be connected at the probe. 1: — mV O2 sensor (white-brown) 2: + mV O2 sensor (brown) 3: + mV thermocouple 1 (green)

  • Page 17: Preparation Of The Pneumatic Cable

    Installation ® Installation and Operation Manual OXITEC 5000 2.13 Preparation of the pneumatic cable Pneumatic cable FEP-0002 Clamp ring Support sleeve Both, the pneumatic tubing for the reference air (blue) and the test gas (green) have to be prepared with support sleeves , clamp rings and nuts…

  • Page 18: Pneumatic Connections Of The Probe

    Installation ® Installation and Operation Manual OXITEC 5000 Electronics with external pneumatics — SME-50000080 Choke non return valve A — Reference air B — Test air Solenoid valve 3/2 ways Flow meter Instrument air in Test air in Test gas in Test gas out Reference air out Figure 15 — SME-50000080 gas plan for external pneumatics.

  • Page 19: Pneumatic Connections Of Electronic Unit

    Installation ® Installation and Operation Manual OXITEC 5000 2.16 Pneumatic connections of electronic Unit Tube Pump version Instrument air version 1/4“ Testgas in Testgas in 1/4“ Testgas out Testgas out 1/4“ Reference air in Instrument air in 1/4“ Reference air out Reference air out 1/4“…

  • Page 20: Initial Operation

    Switch on the line voltage to the system. After a short power up information, the user is prompted to Select language, set the System date, System time, enter a TAG number and ENOTEC REMOTE code (only if option ENOTEC REMOTE is factory activated).

  • Page 21: Display — Measuring Mode

    Initial Operation ® Installation and Operation Manual OXITEC 5000 Display — measuring mode TAG number Measured value Active measuring range Blinking indicators showing under or over measuring range Last access Measured component Limit alarm indicators min-Alarm / max-Alarm * Only if the O2 limit alarms are switched on and the limits are Softkey titles within the set measuring range.

  • Page 22: Software Overview And Explanations

    Software Overview and Explanations ® Installation and Operation Manual OXITEC 5000 Software Overview and Explanations Menu overview — SYS-MENU SYS-MENU System Information Actual measured values measured value (% O ) {may be ppm} -mA output 17A/B (mA) sensor input (mV) Flow rate reference air (l/h) probe temperature (°C / °F) probe heater power (%)

  • Page 23
    Software Overview and Explanations ® Installation and Operation Manual OXITEC 5000 Software version and options Software OXITEC 5000 ~~ Options ~~ ♦ ..SYS-MENU System configuration Measuring ranges Meas.range 1 from (% O , ppm O Meas.range 1 to (% O , ppm O Meas.range 2 from (% O , ppm O…
  • Page 24
    Location altitude (m) System clock/TAG number System date (jjjj-mm-tt) System time (hh:mm:ss) ENOTEC REMOTE settings (optional) } Visible when ENOTEC REMOTE interface is activated ENOTEC REMOTE (ON/OFF) ENOTEC REMOTE code (8 digit code) } Visible when ENOTEC REMOTE is ON…
  • Page 25: Software Explanations — Sys-Menu

    Software Overview and Explanations ® Installation and Operation Manual OXITEC 5000 Software Explanations — SYS-MENU 4.2.1 Measuring ranges (Scaling) The O measuring range is linearly scaled and converted to a linear current output (0/ 4-20 mA). The parameter „O measuring range from“ is the start value the O range, leading to an analogue output of 4,00 mA.

  • Page 26: O 2 Sensor Calibration Values

    Software Overview and Explanations ® Installation and Operation Manual OXITEC 5000 4.2.5 Sensor calibration values Info The sensor calibration values can be altered through a 1 or 2 point calibration. Manual entry of values is only necessary after replacing the O sensor.(The sensor calibration values, «cell constant»…

  • Page 27: Enotec Remote

    Figure 25 — ENOTEC REMOTE connection active Info A maximum of 16 users (smartphones /tablets) can connect to the ENOTEC REMOTE module of an ENOTEC analyzer. Should additional users attempt a connection, the connection will fail. In this case, manually switch off the ENOTEC REMOTE and switch it back on again SYS MENU =>…

  • Page 28: Service (Factory Service Settings)

    Software Overview and Explanations ® Installation and Operation Manual OXITEC 5000 4.2.15 Service (factory service settings) The service functions are password protected and are only accessible by trained service personnel. These functions are protected with a code, different to the system code. Calibration menu CAL MENU 1 point calibration, O…

  • Page 29: 1-Point Calibration (Manual)

    Software Overview and Explanations ® Installation and Operation Manual OXITEC 5000 4.3.2 1-point calibration (manual) During the 1-point calibration of the sensor, the calibration offset is determined. Test air is hereby applied to the sensor. In systems without an integrated pneumatic unit, test gas needs to be applied manually and the flow control also may need to be checked and adjusted if necessary.

  • Page 30: System Checks

    Software Overview and Explanations ® Installation and Operation Manual OXITEC 5000 System checks Sensor checks Source: Test air sensor .. mV = .. % Flow rate .. l/h Source: Test gas sensor .. mV = .. % Flow rate (3 bar max) ..

  • Page 31: Service And Maintenance

    Service and Maintenance ® Installation and Operation Manual OXITEC 5000 Service and Maintenance Exchange fuses Warning De-energize the system first! Abb. 27 Position of the fuses fuse system voltage ampere nominal voltage characteristic Size 115 / 230 V AC 0.5 A 250 V AC T / L 5×20 mm…

  • Page 32: Adjusting Flow Rate (Integrated Pneumatics)

    Service and Maintenance ® Installation and Operation Manual OXITEC 5000 Adjusting Flow Rate (integrated Pneumatics) In the instrument air version in the Safe Area housing with an integrated pneumatic unit, it is possible to adjust the reference and test air at the electronic unit. In this respect, the pump and instrument air versions are different: …

  • Page 33: Position Of The Adjustment Valves

    Service and Maintenance ® Installation and Operation Manual OXITEC 5000 Position of the adjustment valves Figure 30 — Terminal cover of the SME-53 electronic unit showing the position of the reference air and test air valves below. Adjusting Flow Rate (SME-54) With systems in 19“…

  • Page 34: Replacing The Filter

    Figure 33 — Cement new filter in Note The cement dries at room temperature within 24 hours. Using the cement supplied by ENOTEC, all types of ceramic, basalt and sintered metal filters supplied by can be inserted and fixated. Doc.-ID: OXI-11022020-EN…

  • Page 35: Replacing The Probe

    Service and Maintenance ® Installation and Operation Manual OXITEC 5000 Replacing the probe Warning Hot Surface The probe may only be removed with heat-insulated gloves. Before removing the probe, always switch off the supply voltage to the electronic system.  Disconnect the wires at the probe connection box.

  • Page 36
    Service and Maintenance ® Installation and Operation Manual OXITEC 5000 Clean the flange at the probe tube with fine sandpaper. Tighten the new measuring cell with a new metal gasket sealing and four new screws at the measuring cell flange of the probe tube. Insert the probe inner parts and ensure that the inner parts do not bind in the probe tube.
  • Page 37: Exchange Of Probe Inner Parts

    Service and Maintenance ® Installation and Operation Manual OXITEC 5000 Exchange of Probe Inner Parts Proceed as described in chapter 5.8 to dismantle the probe inner parts assembly. The exchange of the thermocouple element is an exception, because this can be changed without dismounting the probe inner part as a whole.

  • Page 38: Relay Outputs / Functions And Correlation

    Service and Maintenance ® Installation and Operation Manual OXITEC 5000 5.10 Relay Outputs / Functions and Correlation The relay contacts are designed for 24V and 1A ~, 1A = (Exception: probe valve) Relay Contact Function Terminal System error* Normally closed Signals operation-critical errors X5 (19A/B) Maintenance Normally open…

  • Page 39: Digital Inputs

    Service and Maintenance ® Installation and Operation Manual OXITEC 5000 5.11 Digital Inputs The digital inputs are designed for a direct voltage of 12 V — 24 V DC for logical “High”. Logical “Low” corresponds to a voltage less than 1 V. Digital input Function Calibration release…

  • Page 40: Status Messages

    Status Messages ® Installation and Operation Manual OXITEC 5000 Status Messages Error Messages Error Relay signal Description Message contact output System 2.00 mA, The error can occur at any time and signalizes a failure of one of the Hardware error, when not set electronic components.

  • Page 41: Alarm Messages

    Status Messages ® Installation and Operation Manual OXITEC 5000 Error Messages (cont.) Error Relay signal Description Message contact output The error can only occur during an O sensor calibration. An error reset can sensor System be carried out by the user. A successful re-calibration also resets the error. offset too error, open Possible causes: defective false test gas (not applicable with test air), test…

  • Page 42: Maintenance Messages

    Status Messages ® Installation and Operation Manual OXITEC 5000 Maintenance Messages Maintenance Relay Description message contact Measured value(s) Maint. When measured values held is set to on, the determined O -mA output remains held closed saved for a duration before a calibration. Doc.-ID: OXI-11022020-EN…

  • Page 43: Troubleshooting

    Troubleshooting ® Installation and Operation Manual OXITEC 5000 Troubleshooting Unsteady, widely varying measuring value (O Possible reasons Procedure Intermittent contact caused by wire breakage Eliminate bad/loose contact Intermittent contact inside the probe — internal mV connection Broken filter element Wrongly installed V-shield Visual inspection by dismounting the probe Probe has been installed without filter head display remains at the end of the measuring range or is higher than expected…

  • Page 44: Technical Data

    Technical Data ® Installation and Operation Manual OXITEC 5000 Technical Data Technical Specifications — Electronic Unit Housing: sheet steel powder coated; RAL6029 (GRP version optional) (SME-54, 19” rack optional) Safe Area Housing: IP66 IP Code: GRP cabinet: IP66 19“ housing: IP20 Display: LC Dot Matrix 240 x 64 — LED backlit…

  • Page 45: Technical Specifications — Probe

    Technical Data ® Installation and Operation Manual OXITEC 5000 Technical Specifications — Probe KES-132x: up to 400°C ( 752 °F) Process gas temperature: KES-200x: up to 600°C (1112 °F) KIS-200x: up to 800°C (1472 °F) KES-500x: up to 600°C (1112 °F) up to 1600°C (2912 °F) with cooling tube Insertion depth: KES-1321:…

  • Page 46: Gas Supply — Systems With Instrument Air

    Technical Data ® Installation and Operation Manual OXITEC 5000 Gas Supply — Systems with Instrument Air The analyzer system uses the connected instrument air continuously for the supply of reference air, and during calibration and system test respectively, for the supply of test air (test gas 1). Instrument air supply for reference air / test air Specification: According to ISO 8573-1 class 2…

  • Page 47: Dimensional Drawings

    Dimensional drawings ® Installation and Operation Manual OXITEC 5000 Dimensional drawings Dimensions of Electronic Unit Housing versions Figure 38 – Dimensions in mm of the SME-53 Figure 39 — Dimensions in mm of the SME-54 Figure 40 — GRP protective housing (optional) — Dimensions in mm Doc.-ID: OXI-11022020-EN…

  • Page 48: Probe Dimensions

    Dimensional drawings ® Installation and Operation Manual OXITEC 5000 Probe Dimensions Figure 41 — Dimensions of the KES-132X probes Figure 42 – Dimensions in mm of the KES-132X probes with cooling protection tube Doc.-ID: OXI-11022020-EN…

  • Page 49
    Dimensional drawings ® Installation and Operation Manual OXITEC 5000 Figure 43 — Dimensions of the KES-200X probes Figure 44 — Dimensions in mm of the KES-200X probes with cooling protection tube Doc.-ID: OXI-11022020-EN…
  • Page 50
    Dimensional drawings ® Installation and Operation Manual OXITEC 5000 Figure 45 — Dimensions in mm of the KES-500X probes Figure 46 — Dimensions of the KES-500X probes with cooling protection tube Doc.-ID: OXI-11022020-EN…
  • Page 51: Counter Flange Dimensions

    Dimensional drawings ® Installation and Operation Manual OXITEC 5000 Counter Flange Dimensions Figure 47 — Dimensions of counter flange ASR-1320 for KES 132X Figure 48 — Dimensions of counter flange ADP-2000 Doc.-ID: OXI-11022020-EN…

  • Page 52
    Dimensional drawings ® Installation and Operation Manual OXITEC 5000 Figure 49 — Dimensions of counter flange ADP-2000-S02 Doc.-ID: OXI-11022020-EN…
  • Page 53: Dimensions Of Protection Tube Flanges

    Dimensional drawings ® Installation and Operation Manual OXITEC 5000 Dimensions of protection tube flanges Figure 50 — Dimensions of protection tube flanges Doc.-ID: OXI-11022020-EN…

  • Page 54: Spare Parts

    Spare Parts ® Installation and Operation Manual OXITEC 5000 Spare Parts Probe Components — KES-132X Figure 51 — KES-132X Probe Components Part Part No. Connection Box ASK-1001 Measuring probe tube MSR-132X Probe flange gasket FLD-132X Probe inner part SIK-132X Oxygen sensor ZO2-00(0/1)1 Measuring sensor gasket MZD-0001…

  • Page 55: Probe Components — Kes-200X

    Spare Parts ® Installation and Operation Manual OXITEC 5000 Probe Components — KES-200X Figure 52 — KES-200X Probe Components Part Part No. Connection Box ASK-1001 Measuring probe tube MSR-200X Probe flange gasket FLD-200X Spacer 0-R-000476 Probe inner part SIK-200X Oxygen measuring cell ZO2-00(0/1)1 Measuring cell gasket MZD-0001…

  • Page 56: Probe Components — Kes-500X

    Spare Parts ® Installation and Operation Manual OXITEC 5000 Probe Components — KES-500X Figure 53 — KES-500X Probe Components Part Part No. Connection Box ASK-1001 Measuring probe tube MSR-200X Probe flange gasket FLD-200X Probe inner part SIK-200X Oxygen sensor cell ZO2-00(0/1)1 Measuring sensor gasket MZD-0001…

  • Page 57: Probe Inner Parts Assembly

    Spare Parts ® Installation and Operation Manual OXITEC 5000 Probe inner parts assembly Figure 54 — Probe inner parts assembly Figure 55 – Reinforced probe inner part assembly (KES/KIS-1323/200x/500x) Article Number (depends on probe type) Description KES-1323 / KES-1321-2 KES-200X / KES-500X KES-2001 Heater support tube HZH-00002…

  • Page 58: Mounting Plates Of The Electronic Unit

    Spare Parts ® Installation and Operation Manual OXITEC 5000 Mounting Plates of the Electronic Unit Parts List Test gas and reference air unit with internal pumps Transformer prim. 2*115V, TRA-0017 sec. 115V PLU-0028 Test air pump for SME5 720 l/h Test gas solenoid valve P01 for PGM-0001 SME-53 — with pneumatic unit…

  • Page 59: Display Board

    Spare Parts ® Installation and Operation Manual OXITEC 5000 Parts List Mounting plate 2 0-L-000047 Optional mA-output board. 0-L-000180 SME5 power board Filter for reference air and test air PLF-0003 — for version with internal pumps Figure 58 — Mounting plate 2 Figure 59 — Mounting plate 2 Display Board Parts List…

  • Page 60: Warranty

    E N O T E C W A R R A N T Y ENOTEC warrants that systems manufactured and sold by it will, upon shipment, be free of defects in workmanship or material. Should any failure to conform to this warranty become apparent during the relevant warranty periods, ENOTEC shall, upon prompt written notice from the purchaser, correct such nonconformity by repair or replacement of the defective part or parts.

  • Page 61: Declaration Of Conformity

    EC-directives as follows. Any unauthorized changes to the products, renders this declaration invalid. The manufacturer carries the sole responsibility for issuing this declaration. Manufacturer: ENOTEC GmbH Höher Birken 6 51709 Marienheide — Germany Tel.: +49 2264 45 78-0…

  • Page 62: Index

    Index ® Installation and Operation Manual OXITEC 5000 Index N  Access to the Terminals ……..8 Name Plate …………5 Adjusting the Flow Rate ……..30 Name Plates …………5 Adjusting the V-shield ……..13 O  C  OXITEC® 5000 Wiring diagram …… 10 Calibration …………

Быстрый.
Точный.
OXITEC® 5000

Информация

ENOTEC производит стационарные газоанализаторы кислорода для быстрого и точного контроля горения. Газоанализаторы кислорода  OXITEC®  разработаны и произведены в Германии в соответствии с ISO 9001.

Мы создаем сенсора на основе оксида циркония высочайшего качества, которые показывают себя наилучшим образом в самых сложных условиях процесса, а также в газах с агрессивными и коррозионными компонентами. Дополнительное преимущество циркониевого датчика  ENOTEC это его устойчивость к колебаниям условий процесса, что делает его долговечным при среднем сроке службы 7-8  лет, в зависимости от применения.

Стационарные газоанализаторы кислорода OXITEC® очень вариабельны и могут быть сконфигурированы под конкретный процесс: длина зонда, тип фильтра, интерфейс и даже материал зонда могут быть заданы индивидуально.

Превосходство газоанализаторов кислорода OXITEC® 5000 гарантируется благодаря использованию только качественных материалов на каждом этапе производства в дополнение к технологиям ноу-хау от наших экспертов. ENOTEC  адаптирует газоанализаторы под наиболее востребованные процессы горения, поэтому требования к обслуживанию минимальны.  

Технические параметры

Длина погружения зонда

до 3689 mm

диапазоны O2

2x 0 — 25 % O2

(другое по запросу)

Точность измерения

< 0,5 % от измеренного значения или
0,02 Вольт % O2
(более высокое значение действует)

Время отклика

< 1 сек. (на тестовом газе)

Температура процесса

максимум.. 800 °C
максимум. 1400 °C (с защитной трубкой охлаждения)

Температура окружающей среды

-40 °C до 80 °C (зонд)
-20 °C до 55 °C (электронный блок)

Интерфейс

MODBUS RTU via RS232/RS485, HART, FIELDBUS, ENOTEC REMOTE

IP код

IP65 (зонд)
IP66 (электронный блок)
IP20 (Электроника, стойка 19 дюймов)

Применение
  • Оптимизация потребления топлива
  • Измерение выбросов
  • Энергетика
  • Сжигание отходов
  • Контроль горения
Особенности
  • Высокая точность
  • Обзор истории калибровки
  • Прогноз срока службы
  • Герметичная конструкция сенсора
  • Самодиагностика системы
  • Быстрое InSitu измерение
  • Долгий срок службы

ENOTEC REMOTE

Одно приложение.
Все анализаторы. Безграничные возможности.

bg-img

Анализатор кислорода в дымовых газах OXITEC 5000

  • Производитель: Enotec GmbH
  • Измерение: O2

Анализатор кислорода OXITEC 5000 предназначен для измерения концентрации кислорода в дымовом газе, образующемся при процессах сгорания, а также в других негорючих газах. Анализатор выполнен в стандартном и во взрывозащищенном исполнении.

Основные сферы применения

  • технологические печи на нефтехимических, нефтеперерабатывающих, газоперерабатывающих, газодобывающих, и химических предприятиях.
  • энергетическая промышленность, металлургическая промышленность, тяжелое машиностроение, котельные.

Описание

Конструктивно анализатор OXITEC 5000 состоит из измерительного зонда, регистрирующего концентрацию кислорода непосредственно в трубе (газоходе), блока электроники, и пневматического блока. Принцип измерения О2 основан на уравнении Нернста с использованием оксида циркония в качестве измерительной ячейки.

Характеристики

Параметр Значение
Принцип измерения Твердотельный сенсор на основе оксида циркония (ZrO2)
Диапазон измерения от 0-2 до 0-25%
Предел определения 1 ppm O2
Погрешность ±0,3%
Время отклика 0,5 с при скорости газа >10 м/с
Темп. компенсация электронная
Температура газа Стандартно: до +500°С
Опционально: до +1400 °С
Давление газа от -50 до +50 мбар
Скорость газа от 0 до 50 м/с
Выходной сигнал Стандартно: 4-20 mA, ENOTEC REMOTE (Bluetooth);
Опционально: RS 232, RS 485, Fieldbus, HART
Экран графический дисплей, светодиодная подсветка
Температура окружающей среды — для блока электроники: -20…+60°С
— для зонда: -40…+80 °С
— опционально применяются обогреваемые шкафы, чехлы
Степень безопасности SIL2
Степень защиты Зонд — IP65
Электронный блок- IP66
Пневматический блок — IP65
Маркировка взрывозащиты Зонд — 1Ex d IIC T3 Gb X
Блок электроники — 1Ex d IIC T5 Gb
Соленоидные клапана — 1Ex e mb T4…T6 Gb X
Напряжение 230В/50-60Гц ±10%
Потребляемая мощность 400Вт при нагреве; 100-200Вт при работе
Предохранитель 10А
Срок эксплуатации более 10 лет, в зависимости от применения

Enotec OXITEC 5000 Series Installation And Operation Manual

  • Contents

  • Table of Contents

  • Troubleshooting

  • Bookmarks

Quick Links

Installation and Operation Manual

O

Analyzer System

2

®

®

OXITEC

5000 / OXITEC

5000+

Version 33

for Software Version: 4.13

Doc.-ID: OXI-11022020-EN

Related Manuals for Enotec OXITEC 5000 Series

Summary of Contents for Enotec OXITEC 5000 Series

  • Page 1
    Installation and Operation Manual Analyzer System ® ® OXITEC 5000 / OXITEC 5000+ Version 33 for Software Version: 4.13 Doc.-ID: OXI-11022020-EN…
  • Page 2
    ENOTEC. This leads to a considerably increased service life compared to «glued or cemented» sensors that have a tendency to leak or crack during operation. All ENOTEC instruments are thoroughly tested in the factory and are subject to a strict ISO 9001 and internal quality assurance ®…
  • Page 3: Table Of Contents

    Automatic Calibration (ACAL) ……24   Warranty     4.2.9   Automatic Calibration Settings ……24       4.2.10 ENOTEC REMOTE ……….. 25 Declaration of Conformity         4.2.11 Measuring units ……….25 Index   4.2.12  …

  • Page 4: System Description

    System Description ® Installation and Operation Manual OXITEC 5000 System Description System overview Info Depending on the system configuration ordered by the customer, the electronic unit and / or probe may vary. Refer to chapters B and B.2 for different versions of electronic units and probes. ®…

  • Page 5
    System Description ® Installation and Operation Manual OXITEC 5000 ® Figure 2 — OXITEC 5000 (probe with cooling tube for flue gas temperatures of up to 1600°C) Safe Area — max. ambient temperature: Electronic unit SME-5 / IP66 -20 °C to +55 °C (-4 °F to +131 °F) Safe Area — max.
  • Page 6: Measuring Principle

    Storage instructions ENOTEC equipment and spares are to be stored in a dry and ventilated environment at temperatures between -40 °C to +80 °C (-40 °F to 176 °F). Paint fumes, silicone sprays, etc. must be avoided in the storage environment.

  • Page 7: Name Plates

    System Description ® Installation and Operation Manual OXITEC 5000 Name plates The name plate contains information about the line voltage, the nominated current, frequency, protection class, year of manufacture, serial number, order number and system order code. The system order code refers to information which is detailed in the system test report and supplied with the system. SME-53 Electronic Unit Probe…

  • Page 8: Installation

    Installation ® Installation and Operation Manual OXITEC 5000 Installation Warning The system is not equipped with an external power-off switch. The line voltage switch/fuse/breaker must be installed and be in accordance with local technical standards and should be near to the electronic unit and must be clearly marked as such.

  • Page 9: Installation Of Probe Signal Cable Fep-0001

    Installation ® Installation and Operation Manual OXITEC 5000 Installation of probe signal cable FEP-0001 Abide by the maximum cable length (max. 150m) Note the minimum bending radius. FEP-0001  R = 96 mm +50 °C +90 °C Max. Max. Temperature during Temperature during operation -5 °C -40 °C…

  • Page 10: Access To The Terminals

    Installation ® Installation and Operation Manual OXITEC 5000 Access to the terminals Warning Before removing the terminal covers, switch off the mains voltage to the system. Switch the mains voltage on only after attaching the terminal cover. After the installation has been completed, live parts may no longer be accessible.

  • Page 11: Wiring Diagram Of The Electronic Unit

    Installation ® Installation and Operation Manual OXITEC 5000 Wiring diagram of the electronic unit Figure 7 — Wiring diagram of the Electronic Unit Ferrite sleeves (enclosed) Analogue output cable (customer) Power supply cable (customer) Status signal cable (customer) Probe signal cable Pressure transmitter analogue input cable (optional, customer) Power supply…

  • Page 12: Oxitec® 5000 Wiring Diagram

    Installation ® Installation and Operation Manual OXITEC 5000 OXITEC® 5000 Wiring diagram Figure 8 — Wiring diagram of the OXITEC 5000 Analyzer system Doc.-ID: OXI-11022020-EN…

  • Page 13: Installation Of The Probe

    For probe lengths exceeding 2000 mm, a support must be mounted inside the duct (every 2 m) to prevent the probe and mounting tube from flexing or bending. ENOTEC recommends installing the probe horizontally (- 1 ° to — 3 °) for the fastest possible response time.

  • Page 14: Mounting Of The Probe

    Installation ® Installation and Operation Manual OXITEC 5000 Mounting of the probe Figure 9 — Mounting of the O probe – for dimensions, see section B Counter flange which has been welded gas tight at correct Probe connection box angle to duct Flange gasket Duct wall Protection tube…

  • Page 15: Mounting A Probe With Cooling Protection Tube

    Installation ® Installation and Operation Manual OXITEC 5000 Mounting a probe with cooling protection tube Info If a cooling tube is used, the part of the protection tube projecting from the duct wall must be insulated or heated if necessary, to prevent its temperature from dropping below the dew point. Make sure that the gas outlet is not blocked.

  • Page 16: Electrical Connections Of The Probe

    Installation ® Installation and Operation Manual OXITEC 5000 2.11 Electrical connections of the probe Info The shield of the probe cable must only be connected at the electronic housing at the PE terminal. Under no circumstance should the shield also be connected at the probe. 1: — mV O2 sensor (white-brown) 2: + mV O2 sensor (brown) 3: + mV thermocouple 1 (green)

  • Page 17: Preparation Of The Pneumatic Cable

    Installation ® Installation and Operation Manual OXITEC 5000 2.13 Preparation of the pneumatic cable Pneumatic cable FEP-0002 Clamp ring Support sleeve Both, the pneumatic tubing for the reference air (blue) and the test gas (green) have to be prepared with support sleeves , clamp rings and nuts…

  • Page 18: Pneumatic Connections Of The Probe

    Installation ® Installation and Operation Manual OXITEC 5000 Electronics with external pneumatics — SME-50000080 Choke non return valve A — Reference air B — Test air Solenoid valve 3/2 ways Flow meter Instrument air in Test air in Test gas in Test gas out Reference air out Figure 15 — SME-50000080 gas plan for external pneumatics.

  • Page 19: Pneumatic Connections Of Electronic Unit

    Installation ® Installation and Operation Manual OXITEC 5000 2.16 Pneumatic connections of electronic Unit Tube Pump version Instrument air version 1/4“ Testgas in Testgas in 1/4“ Testgas out Testgas out 1/4“ Reference air in Instrument air in 1/4“ Reference air out Reference air out 1/4“…

  • Page 20: Initial Operation

    Switch on the line voltage to the system. After a short power up information, the user is prompted to Select language, set the System date, System time, enter a TAG number and ENOTEC REMOTE code (only if option ENOTEC REMOTE is factory activated).

  • Page 21: Display — Measuring Mode

    Initial Operation ® Installation and Operation Manual OXITEC 5000 Display — measuring mode TAG number Measured value Active measuring range Blinking indicators showing under or over measuring range Last access Measured component Limit alarm indicators min-Alarm / max-Alarm * Only if the O2 limit alarms are switched on and the limits are Softkey titles within the set measuring range.

  • Page 22: Software Overview And Explanations

    Software Overview and Explanations ® Installation and Operation Manual OXITEC 5000 Software Overview and Explanations Menu overview — SYS-MENU SYS-MENU System Information Actual measured values measured value (% O ) {may be ppm} -mA output 17A/B (mA) sensor input (mV) Flow rate reference air (l/h) probe temperature (°C / °F) probe heater power (%)

  • Page 23
    Software Overview and Explanations ® Installation and Operation Manual OXITEC 5000 Software version and options Software OXITEC 5000 ~~ Options ~~ ♦ ..SYS-MENU System configuration Measuring ranges Meas.range 1 from (% O , ppm O Meas.range 1 to (% O , ppm O Meas.range 2 from (% O , ppm O…
  • Page 24
    Location altitude (m) System clock/TAG number System date (jjjj-mm-tt) System time (hh:mm:ss) ENOTEC REMOTE settings (optional) } Visible when ENOTEC REMOTE interface is activated ENOTEC REMOTE (ON/OFF) ENOTEC REMOTE code (8 digit code) } Visible when ENOTEC REMOTE is ON…
  • Page 25: Software Explanations — Sys-Menu

    Software Overview and Explanations ® Installation and Operation Manual OXITEC 5000 Software Explanations — SYS-MENU 4.2.1 Measuring ranges (Scaling) The O measuring range is linearly scaled and converted to a linear current output (0/ 4-20 mA). The parameter „O measuring range from“ is the start value the O range, leading to an analogue output of 4,00 mA.

  • Page 26: O 2 Sensor Calibration Values

    Software Overview and Explanations ® Installation and Operation Manual OXITEC 5000 4.2.5 Sensor calibration values Info The sensor calibration values can be altered through a 1 or 2 point calibration. Manual entry of values is only necessary after replacing the O sensor.(The sensor calibration values, «cell constant»…

  • Page 27: Enotec Remote

    Figure 25 — ENOTEC REMOTE connection active Info A maximum of 16 users (smartphones /tablets) can connect to the ENOTEC REMOTE module of an ENOTEC analyzer. Should additional users attempt a connection, the connection will fail. In this case, manually switch off the ENOTEC REMOTE and switch it back on again SYS MENU =>…

  • Page 28: Service (Factory Service Settings)

    Software Overview and Explanations ® Installation and Operation Manual OXITEC 5000 4.2.15 Service (factory service settings) The service functions are password protected and are only accessible by trained service personnel. These functions are protected with a code, different to the system code. Calibration menu CAL MENU 1 point calibration, O…

  • Page 29: 1-Point Calibration (Manual)

    Software Overview and Explanations ® Installation and Operation Manual OXITEC 5000 4.3.2 1-point calibration (manual) During the 1-point calibration of the sensor, the calibration offset is determined. Test air is hereby applied to the sensor. In systems without an integrated pneumatic unit, test gas needs to be applied manually and the flow control also may need to be checked and adjusted if necessary.

  • Page 30: System Checks

    Software Overview and Explanations ® Installation and Operation Manual OXITEC 5000 System checks Sensor checks Source: Test air sensor .. mV = .. % Flow rate .. l/h Source: Test gas sensor .. mV = .. % Flow rate (3 bar max) ..

  • Page 31: Service And Maintenance

    Service and Maintenance ® Installation and Operation Manual OXITEC 5000 Service and Maintenance Exchange fuses Warning De-energize the system first! Abb. 27 Position of the fuses fuse system voltage ampere nominal voltage characteristic Size 115 / 230 V AC 0.5 A 250 V AC T / L 5×20 mm…

  • Page 32: Adjusting Flow Rate (Integrated Pneumatics)

    Service and Maintenance ® Installation and Operation Manual OXITEC 5000 Adjusting Flow Rate (integrated Pneumatics) In the instrument air version in the Safe Area housing with an integrated pneumatic unit, it is possible to adjust the reference and test air at the electronic unit. In this respect, the pump and instrument air versions are different: …

  • Page 33: Position Of The Adjustment Valves

    Service and Maintenance ® Installation and Operation Manual OXITEC 5000 Position of the adjustment valves Figure 30 — Terminal cover of the SME-53 electronic unit showing the position of the reference air and test air valves below. Adjusting Flow Rate (SME-54) With systems in 19“…

  • Page 34: Replacing The Filter

    Figure 33 — Cement new filter in Note The cement dries at room temperature within 24 hours. Using the cement supplied by ENOTEC, all types of ceramic, basalt and sintered metal filters supplied by can be inserted and fixated. Doc.-ID: OXI-11022020-EN…

  • Page 35: Replacing The Probe

    Service and Maintenance ® Installation and Operation Manual OXITEC 5000 Replacing the probe Warning Hot Surface The probe may only be removed with heat-insulated gloves. Before removing the probe, always switch off the supply voltage to the electronic system.  Disconnect the wires at the probe connection box.

  • Page 36
    Service and Maintenance ® Installation and Operation Manual OXITEC 5000 Clean the flange at the probe tube with fine sandpaper. Tighten the new measuring cell with a new metal gasket sealing and four new screws at the measuring cell flange of the probe tube. Insert the probe inner parts and ensure that the inner parts do not bind in the probe tube.
  • Page 37: Exchange Of Probe Inner Parts

    Service and Maintenance ® Installation and Operation Manual OXITEC 5000 Exchange of Probe Inner Parts Proceed as described in chapter 5.8 to dismantle the probe inner parts assembly. The exchange of the thermocouple element is an exception, because this can be changed without dismounting the probe inner part as a whole.

  • Page 38: Relay Outputs / Functions And Correlation

    Service and Maintenance ® Installation and Operation Manual OXITEC 5000 5.10 Relay Outputs / Functions and Correlation The relay contacts are designed for 24V and 1A ~, 1A = (Exception: probe valve) Relay Contact Function Terminal System error* Normally closed Signals operation-critical errors X5 (19A/B) Maintenance Normally open…

  • Page 39: Digital Inputs

    Service and Maintenance ® Installation and Operation Manual OXITEC 5000 5.11 Digital Inputs The digital inputs are designed for a direct voltage of 12 V — 24 V DC for logical “High”. Logical “Low” corresponds to a voltage less than 1 V. Digital input Function Calibration release…

  • Page 40: Status Messages

    Status Messages ® Installation and Operation Manual OXITEC 5000 Status Messages Error Messages Error Relay signal Description Message contact output System 2.00 mA, The error can occur at any time and signalizes a failure of one of the Hardware error, when not set electronic components.

  • Page 41: Alarm Messages

    Status Messages ® Installation and Operation Manual OXITEC 5000 Error Messages (cont.) Error Relay signal Description Message contact output The error can only occur during an O sensor calibration. An error reset can sensor System be carried out by the user. A successful re-calibration also resets the error. offset too error, open Possible causes: defective false test gas (not applicable with test air), test…

  • Page 42: Maintenance Messages

    Status Messages ® Installation and Operation Manual OXITEC 5000 Maintenance Messages Maintenance Relay Description message contact Measured value(s) Maint. When measured values held is set to on, the determined O -mA output remains held closed saved for a duration before a calibration. Doc.-ID: OXI-11022020-EN…

  • Page 43: Troubleshooting

    Troubleshooting ® Installation and Operation Manual OXITEC 5000 Troubleshooting Unsteady, widely varying measuring value (O Possible reasons Procedure Intermittent contact caused by wire breakage Eliminate bad/loose contact Intermittent contact inside the probe — internal mV connection Broken filter element Wrongly installed V-shield Visual inspection by dismounting the probe Probe has been installed without filter head display remains at the end of the measuring range or is higher than expected…

  • Page 44: Technical Data

    Technical Data ® Installation and Operation Manual OXITEC 5000 Technical Data Technical Specifications — Electronic Unit Housing: sheet steel powder coated; RAL6029 (GRP version optional) (SME-54, 19” rack optional) Safe Area Housing: IP66 IP Code: GRP cabinet: IP66 19“ housing: IP20 Display: LC Dot Matrix 240 x 64 — LED backlit…

  • Page 45: Technical Specifications — Probe

    Technical Data ® Installation and Operation Manual OXITEC 5000 Technical Specifications — Probe KES-132x: up to 400°C ( 752 °F) Process gas temperature: KES-200x: up to 600°C (1112 °F) KIS-200x: up to 800°C (1472 °F) KES-500x: up to 600°C (1112 °F) up to 1600°C (2912 °F) with cooling tube Insertion depth: KES-1321:…

  • Page 46: Gas Supply — Systems With Instrument Air

    Technical Data ® Installation and Operation Manual OXITEC 5000 Gas Supply — Systems with Instrument Air The analyzer system uses the connected instrument air continuously for the supply of reference air, and during calibration and system test respectively, for the supply of test air (test gas 1). Instrument air supply for reference air / test air Specification: According to ISO 8573-1 class 2…

  • Page 47: Dimensional Drawings

    Dimensional drawings ® Installation and Operation Manual OXITEC 5000 Dimensional drawings Dimensions of Electronic Unit Housing versions Figure 38 – Dimensions in mm of the SME-53 Figure 39 — Dimensions in mm of the SME-54 Figure 40 — GRP protective housing (optional) — Dimensions in mm Doc.-ID: OXI-11022020-EN…

  • Page 48: Probe Dimensions

    Dimensional drawings ® Installation and Operation Manual OXITEC 5000 Probe Dimensions Figure 41 — Dimensions of the KES-132X probes Figure 42 – Dimensions in mm of the KES-132X probes with cooling protection tube Doc.-ID: OXI-11022020-EN…

  • Page 49
    Dimensional drawings ® Installation and Operation Manual OXITEC 5000 Figure 43 — Dimensions of the KES-200X probes Figure 44 — Dimensions in mm of the KES-200X probes with cooling protection tube Doc.-ID: OXI-11022020-EN…
  • Page 50
    Dimensional drawings ® Installation and Operation Manual OXITEC 5000 Figure 45 — Dimensions in mm of the KES-500X probes Figure 46 — Dimensions of the KES-500X probes with cooling protection tube Doc.-ID: OXI-11022020-EN…
  • Page 51: Counter Flange Dimensions

    Dimensional drawings ® Installation and Operation Manual OXITEC 5000 Counter Flange Dimensions Figure 47 — Dimensions of counter flange ASR-1320 for KES 132X Figure 48 — Dimensions of counter flange ADP-2000 Doc.-ID: OXI-11022020-EN…

  • Page 52
    Dimensional drawings ® Installation and Operation Manual OXITEC 5000 Figure 49 — Dimensions of counter flange ADP-2000-S02 Doc.-ID: OXI-11022020-EN…
  • Page 53: Dimensions Of Protection Tube Flanges

    Dimensional drawings ® Installation and Operation Manual OXITEC 5000 Dimensions of protection tube flanges Figure 50 — Dimensions of protection tube flanges Doc.-ID: OXI-11022020-EN…

  • Page 54: Spare Parts

    Spare Parts ® Installation and Operation Manual OXITEC 5000 Spare Parts Probe Components — KES-132X Figure 51 — KES-132X Probe Components Part Part No. Connection Box ASK-1001 Measuring probe tube MSR-132X Probe flange gasket FLD-132X Probe inner part SIK-132X Oxygen sensor ZO2-00(0/1)1 Measuring sensor gasket MZD-0001…

  • Page 55: Probe Components — Kes-200X

    Spare Parts ® Installation and Operation Manual OXITEC 5000 Probe Components — KES-200X Figure 52 — KES-200X Probe Components Part Part No. Connection Box ASK-1001 Measuring probe tube MSR-200X Probe flange gasket FLD-200X Spacer 0-R-000476 Probe inner part SIK-200X Oxygen measuring cell ZO2-00(0/1)1 Measuring cell gasket MZD-0001…

  • Page 56: Probe Components — Kes-500X

    Spare Parts ® Installation and Operation Manual OXITEC 5000 Probe Components — KES-500X Figure 53 — KES-500X Probe Components Part Part No. Connection Box ASK-1001 Measuring probe tube MSR-200X Probe flange gasket FLD-200X Probe inner part SIK-200X Oxygen sensor cell ZO2-00(0/1)1 Measuring sensor gasket MZD-0001…

  • Page 57: Probe Inner Parts Assembly

    Spare Parts ® Installation and Operation Manual OXITEC 5000 Probe inner parts assembly Figure 54 — Probe inner parts assembly Figure 55 – Reinforced probe inner part assembly (KES/KIS-1323/200x/500x) Article Number (depends on probe type) Description KES-1323 / KES-1321-2 KES-200X / KES-500X KES-2001 Heater support tube HZH-00002…

  • Page 58: Mounting Plates Of The Electronic Unit

    Spare Parts ® Installation and Operation Manual OXITEC 5000 Mounting Plates of the Electronic Unit Parts List Test gas and reference air unit with internal pumps Transformer prim. 2*115V, TRA-0017 sec. 115V PLU-0028 Test air pump for SME5 720 l/h Test gas solenoid valve P01 for PGM-0001 SME-53 — with pneumatic unit…

  • Page 59: Display Board

    Spare Parts ® Installation and Operation Manual OXITEC 5000 Parts List Mounting plate 2 0-L-000047 Optional mA-output board. 0-L-000180 SME5 power board Filter for reference air and test air PLF-0003 — for version with internal pumps Figure 58 — Mounting plate 2 Figure 59 — Mounting plate 2 Display Board Parts List…

  • Page 60: Warranty

    E N O T E C W A R R A N T Y ENOTEC warrants that systems manufactured and sold by it will, upon shipment, be free of defects in workmanship or material. Should any failure to conform to this warranty become apparent during the relevant warranty periods, ENOTEC shall, upon prompt written notice from the purchaser, correct such nonconformity by repair or replacement of the defective part or parts.

  • Page 61: Declaration Of Conformity

    EC-directives as follows. Any unauthorized changes to the products, renders this declaration invalid. The manufacturer carries the sole responsibility for issuing this declaration. Manufacturer: ENOTEC GmbH Höher Birken 6 51709 Marienheide — Germany Tel.: +49 2264 45 78-0…

  • Page 62: Index

    Index ® Installation and Operation Manual OXITEC 5000 Index N  Access to the Terminals ……..8 Name Plate …………5 Adjusting the Flow Rate ……..30 Name Plates …………5 Adjusting the V-shield ……..13 O  C  OXITEC® 5000 Wiring diagram …… 10 Calibration …………

Быстрый.
Точный.
OXITEC® 5000
Информация

ENOTEC производит стационарные газоанализаторы кислорода для быстрого и точного контроля горения. Газоанализаторы кислорода  OXITEC®  разработаны и произведены в Германии в соответствии с ISO 9001.

Мы создаем сенсора на основе оксида циркония высочайшего качества, которые показывают себя наилучшим образом в самых сложных условиях процесса, а также в газах с агрессивными и коррозионными компонентами. Дополнительное преимущество циркониевого датчика  ENOTEC это его устойчивость к колебаниям условий процесса, что делает его долговечным при среднем сроке службы 7-8  лет, в зависимости от применения.

Стационарные газоанализаторы кислорода OXITEC® очень вариабельны и могут быть сконфигурированы под конкретный процесс: длина зонда, тип фильтра, интерфейс и даже материал зонда могут быть заданы индивидуально.

Превосходство газоанализаторов кислорода OXITEC® 5000 гарантируется благодаря использованию только качественных материалов на каждом этапе производства в дополнение к технологиям ноу-хау от наших экспертов. ENOTEC  адаптирует газоанализаторы под наиболее востребованные процессы горения, поэтому требования к обслуживанию минимальны.  

Технические параметры

Длина погружения зонда

до 3689 mm

диапазоны O2

2x 0 — 25 % O2

(другое по запросу)

Точность измерения

< 0,5 % от измеренного значения или
0,02 Вольт % O2
(более высокое значение действует)

Время отклика

< 1 сек. (на тестовом газе)

Температура процесса

максимум.. 800 °C
максимум. 1400 °C (с защитной трубкой охлаждения)

Температура окружающей среды

-40 °C до 80 °C (зонд)
-20 °C до 55 °C (электронный блок)

Интерфейс

MODBUS RTU via RS232/RS485, HART, FIELDBUS, ENOTEC REMOTE

IP код

IP65 (зонд)
IP66 (электронный блок)
IP20 (Электроника, стойка 19 дюймов)

Применение
  • Оптимизация потребления топлива
  • Измерение выбросов
  • Энергетика
  • Сжигание отходов
  • Контроль горения
Особенности
  • Высокая точность
  • Обзор истории калибровки
  • Прогноз срока службы
  • Герметичная конструкция сенсора
  • Самодиагностика системы
  • Быстрое InSitu измерение
  • Долгий срок службы

ENOTEC REMOTE

Одно приложение.
Все анализаторы. Безграничные возможности.

Описание
Программное обеспечение
Технические характеристики
Знак утверждения типа
Комплектность
Поверка
Сведения о методах измерений
Нормативные документы
Рекомендации к применению

Г азоанализаторы кислорода OXITEC Назначение средства измерений

Г азоанализаторы кислорода OXITEC предназначены для измерения объемной доли кислорода в отходящих газах топливосжигающих установок и в других негорючих газах.

Описание

Газоанализаторы кислорода OXITEC (далее — газоанализаторы) представляют собой стационарные многоблочные приборы непрерывного действия.

Конструктивно газоанализаторы выполнены многоблочными и состоят из измерительного зонда, блока электронной обработки информации (блока управления) и блока управления контрольным и эталонным газом (по заказу).

Измерительный зонд газоанализатора предназначен для монтажа непосредственно в объект, внутри которого необходимо проводить контроль концентрации кислорода (топливосжигающая установка, труба для отвода отходящих газов и т.п.).

Блок управления предназначен для приема, обработки и отображения измерительной информации, поступающей от измерительного зонда.

Блок управления имеет выходные сигналы:

—    показания жидкокристаллического цифрового дисплея;

—    выходной унифицированный аналоговый токовый сигнал (0-20) мА или (4-20) мА;

—    цифровой выход RS-232, RS-485, HART, FIELDBUS (по дополнительному заказу);

—    релейные выходы типа «сухой контакт» (по дополнительному заказу).

Газоанализаторы выпускаются в двух исполнениях: OXITEC 5000 и

OXITEC ECONOMY.

В состав газоанализатора в исполнении OXITEC 5000 входят:

1)    блок управления модели SME5x (исполнение для настенного монтажа или для установки в стандартную стойку 19″)

или модели SME 5D (взрывозащищенная версия, исполнение для настенного монтажа)

2)    измерительный зонд (модели KES-132x, KES-200x, KIS-200x, KES500x или KEX500x — где «x» условное обозначение типоразмера зонда).

Также в состав газоанализатора может входить блок управления контрольным и эталонным газом.

Блок управления, пневматический блок и прочие дополнительные устройства газоанализатора взрывозащищенного исполнения могут быть смонтированы в специальном шкафу.

В состав газоанализатора в исполнении OXITEC ECONOMY входят:

1)    блок управления модели SME5х, исполнение для настенного монтажа

2)    измерительный зонд модели ЕСО1000.

На передней панели блока управления любой модели расположены 4 программирующих клавиши управления, а также десятичная клавиатура для ввода численных параметров и жидкокристаллический дисплей, на котором отображается информация об измеряемой величине, состоянии газоанализатора и параметрах при его тестировании.

Принцип действия газоанализаторов — электрохимический метод с использованием циркониевого чувствительного элемента.

Способ отбора пробы — диффузионный.

Маркировка взрывозащиты комплектующих газоанализаторов:

—    измерительных зондов KEX500x    1 Ex d IIC T3 X

—    блока управления SME 5D    1 Ex d IIC T5 X

—    блок управления контрольным и эталонным газом    2 Ex emb II T4/T5

Программное обеспечение

Газоанализаторы имеют встроенное программное обеспечение, разработанное изготовителем специально для решения задач измерения объемной доли кислорода в отходящих газах топливосжигающих установок.

Идентификационные данные встроенного программного обеспечения приведены в таблице 1.

Таблица 1 — Идентификационные данные встроенного программного обеспечения

Наименование

программного

обеспечения

Идентификационное наименование программного обеспечения

Номер версии (идентификационный номер) программного обеспечения

Цифровой идентификатор программного обеспечения (контрольная сумма исполняемого кода)

Алгоритм вычисления цифрового идентификатора программного обеспечения

OXITEC 5000

OXITEC 5000

4.04

cdd9576879973e9162

e0cd9bc6ab20ea

MD5

Примечание — в газоанализаторах OXITEC ECONOMY используется то же программное обеспечение.

Влияние встроенного программного обеспечения газоанализаторов учтено при нормировании метрологических характеристик.

Газоанализаторы имеют защиту встроенного программного обеспечения от преднамеренных или непреднамеренных изменений, реализованную изготовителем на этапе производства путем установки системы защиты микроконтроллера от чтения и записи. Уровень защиты «C» по МИ 3286-2010.

Технические характеристики

1)    Диапазоны измерений объемной доли кислорода, %    от 0 до 2, от 0 до 5,

от 0 до 6, от 0 до 10 от 0 до 25

Примечание — диапазон измерений по требованию Заказчика.

2)    Пределы допускаемой основной абсолютной погрешности газоанализатора, объемная доля кислорода, %    ± 0,3

3)    Пределы допускаемой вариации выходного сигнала газоанализатора, в долях от пределов допускаемой основной погрешности    0,2

4)    Пределы допускаемой дополнительной погрешности газоанализатора от влияния изменения температуры окружающей среды в диапазоне от минус 20 0С до плюс 60 0С, в долях от пределов допускаемой основной погрешности    0,5

5)    Время прогрева газоанализатора, мин, не более    60

6)    Номинальное время установления показаний, Т0,9 ном, с    30

7)    Напряжение питания переменным током частотой 50/60 Гц, В 230 ± 10 % или 115 ± 10 %

8)    Потребляемая электрическая мощность В-А, не более:

—    в режиме прогрева    400

—    в рабочем режиме    200

9)    Габаритные размеры и масса элементов газоанализатора приведены в таблице 2.

Наименование элемента газоанализатора

Г абаритные размеры, мм, не более

Масса, кг

высота

ширина

длина

Блок управления SME5x для настенного монтажа

240

300

400

5

Блок управления SME5D для настенного монтажа

356

655

250

37

Блок управления SME5 исполнение для монтажа в стойку 19″

177

483

360

22

Измерительный зонд ЕСО1000 для OXITEC ECONOMY

160

160

До 1500 1)

Х2)

Измерительный зонд KES-132x, KES-200x, KIS-200x, KES500x или KEX500x

200

200

До 4500 1)

Х2)

Блок управления контрольным и эталонным газом SME-500000x

440

300

218

4

Примечания:

1)    * — по заказу может быть осуществлена поставка зондов различной длины;

2)    ** — масса измерительного зонда в зависимости от типоразмера.

10)    Срок службы циркониевого датчика, лет, не менее    5

11)    Средняя наработка на отказ, ч    20 000

Рабочие условия эксплуатации

—    диапазон температуры окружающей и контролируемой сред, С

блок управления    от минус 20 до 60

измерительный зонд    до 500

до 1400 (с охлаждающей трубкой)

—    диапазон атмосферного давления (для блока управления), кПа    от 84 до 106,7

—    диапазон относительной влажности при температуре 25°С, %    до 95

Знак утверждения типа

Знак утверждения типа наносится типографским способом на титульный лист руководства по эксплуатации и на корпус газоанализатора в виде наклейки.

Комплектность

Таблица 3 — Комплектность

Обозначение

Наименование

Количество, шт.

SME5x или SME5D

Блок управления

1 (модель по заказу)

KES-132x, или KES-200x, или KIS-200x, или KES500x, или KEX500x

Измерительный зонд OXITEC 5000

1 (модель по заказу)

ЕСО1000

Измерительный зонд для OXITEC ECONOMY

1

SME-500000x

Блок управления контрольным и эталонным газом

1 (по заказу)

Комплект электрических и пневматических кабелей

1 компл.

Руководство по эксплуатации

1 экз.

МП-242-1207-2011

Методика поверки

1 экз.

Поверка

осуществляется по документу МП-242-1207-2011 «Газоанализаторы кислорода OXITEC. Методика поверки», утвержденному ГЦИ СИ ФГУП «ВНИИМ им. Д.И. Менделеева» «17» августа 2011 г.

Основные средства поверки: ГСО-ПГС кислород — азот (номер по Госреестру 3720-87, 3722-87, 3724-87, 3726-87) в баллонах под давлением, выпускаемые по ТУ 6-16-2956-92 с изм. 1…7.

Сведения о методах измерений

Методика измерений приведена в документе «Газоанализаторы кислорода OXITEC. Руководство по эксплуатации», 2011 г.

Нормативные документы

1    ГОСТ 13320-81 Газоанализаторы промышленные автоматические. Общие технические условия.

2    ГОСТ Р 52931-2008 Приборы контроля и регулирования технологических процессов. Общие технические условия.

3    ГОСТ 8.578-2008 ГСИ. Государственная поверочная схема для средств измерений содержания компонентов в газовых средах.

Рекомендации к применению

обеспечение безопасных условий и охраны труда.

Preface

Dear Customer,

Thank you for selecting the OXITEC

®

OXITEC

5000 as well as the OXITEC

series.

Since 1980 our analyzer systems have been operating in numerous applications with tens of thousands of units being produced

and shipped worldwide. ENOTEC is committed to absolute quality and performance and over time we have continuously

enhanced our products to integrate various features and functions.

In this package, the electronic unit uses the very latest Microprocessor Technology, making the SME-5 electronic unit one of the

most advanced and up-to-date monitoring units, permitting you to reduce your maintenance & fuel costs and to achieve

increased measuring accuracy with operational reliability.

ENOTEC probes include the zirconium oxide sensor which has a gas-tight seal made by using a special process and technique

developed by ENOTEC. This leads to a considerably increased service life compared to «glued or cemented» sensors that have

a tendency to leak or crack during operation.

All ENOTEC instruments are thoroughly tested in the factory and are subject to a strict ISO 9001 and internal quality assurance

procedure. Therefore, with correct installation, the operation of the OXITEC

user friendly and will provide you with many years of trouble free operation with perfect measuring results.

Symbols used in this Manual

All symbols listed beneath, attached to the analyzer or noted in this manual show important information as well as safety

instructions for installation, operation and maintenance, to protect the personnel and the equipment.

Warning

Follow all instructions of this manual

Warning hot Surface

Warns of danger of burns which could occur from

hot system parts

Caution

Warns of risks of destroying the system or its

components or its functionality

The contents of this manual are protected by copyright. Alterations and errors reserved.

Safety Instructions

This system is operated with line voltage. If covers are removed during which line voltage is still connected, an electric shock

hazard will occur.

Only well trained and authorized personnel are allowed to conduct work on this system. Personnel have to understand all

precautions, safety instructions, installation and maintenance instructions in this manual. The trouble free and safe operation of

this system requires safe transportation, professional storage, installation, operation and maintenance.

Furthermore, all local safety requirements have to be considered.

This system may not be used in the vicinity of combustible gases as parts of the system may cause a risk of explosion.

The device may only be put into operation if the enclosed instructions have been fully understood.

This manual is also available in other languages on request.

®

5000 series as your InSitu flue gas oxygen analyzer system. This series includes the

®

5000+ special configuration. This manual uses OXITEC

®

5000 as umbrella term for the

®

5000 InSitu analyzer system is straightforward and

Consider Information

Points out important information which must be

considered before execution

Note

Contains further detailed information

Ground earth electrical protection

Doc.-ID: OXI-11022020-EN

Анализатор кислорода (O2) в дымовых газах OXITEC 5000

Анализатор кислорода (O2) в дымовых газах OXITEC 5000

Анализаторы кислорода  OXITEC  предназначены для измерения концентрации кислорода в дымовом газе, образующемся при процессах сгорания, а также в других негорючих газах. Система состоит из измерительного зонда, регистрирующего концентрацию кислорода непосредственно в трубе дымовых газов, электрического соединительного кабеля, пневматического соединительного кабеля, и электронного блока. Зонды и электронные блоки имеют разное исполнение и могут применяться в стандартных и во взрывоопасных областях. Принцип действия анализаторов кислорода  OXITEC  основан на уравнении Нернста с использованием оксида циркония в качестве измерительной ячейки для змерения кислорода в технологических и дымовых газах.

          

Характеристики:

Принцип измерения:

оксид циркония

Материал сенсора

V4A (1.4571/316SS)

Темп. компенсация:

электронная

Диапазон полной шкалы:

от 0-5 до 0-9000 ppm

Время отклика:

<200 мс (0,5с при скорости газа >10м/с)

Время установления показаний Т90:

5 с (при скорости газа >10 м/с)

Точность:

±0,2%

Температура газа:

до 1400 °С

Давление газа:

от -50 до +50 мбар

Скорость газа:

от 0 до 50 м/с

Выходной сигнал:

стандартно — 4-20 mA; опционально — RS 232, RS 485, Fieldbus, HART

Экран:

графический дисплей, светодиодная подсветка

Температура окружающей среды:

— для блока электроники: -20…+60°С

— для зонда: -40…+80 °С

— возможность комплектации обогреваемыми чехлами, шкафами

Степень защиты (терминальная часть):

IP65

Предел определения:

менее 1 ppm O2

Потребляемое напряжение:

через блок электроники

Напряжение:

230В/50-60Гц ±10%

Потребляемая мощность:

400VA при нагреве 100-200VA при работе

Предохранитель:

10А

Срок эксплуатации:

более 10 лет, в зависимости от применения

Анализатор кислорода (O2) в дымовых газах OXITEC 5000

Анализатор кислорода (O2) в дымовых газах OXITEC 5000

Анализаторы кислорода  OXITEC  предназначены для измерения концентрации кислорода в дымовом газе, образующемся при процессах сгорания, а также в других негорючих газах. Система состоит из измерительного зонда, регистрирующего концентрацию кислорода непосредственно в трубе дымовых газов, электрического соединительного кабеля, пневматического соединительного кабеля, и электронного блока. Зонды и электронные блоки имеют разное исполнение и могут применяться в стандартных и во взрывоопасных областях. Принцип действия анализаторов кислорода  OXITEC  основан на уравнении Нернста с использованием оксида циркония в качестве измерительной ячейки для змерения кислорода в технологических и дымовых газах.

          

Характеристики:

Принцип измерения:

оксид циркония

Материал сенсора

V4A (1.4571/316SS)

Темп. компенсация:

электронная

Диапазон полной шкалы:

от 0-5 до 0-9000 ppm

Время отклика:

<200 мс (0,5с при скорости газа >10м/с)

Время установления показаний Т90:

5 с (при скорости газа >10 м/с)

Точность:

±0,2%

Температура газа:

до 1400 °С

Давление газа:

от -50 до +50 мбар

Скорость газа:

от 0 до 50 м/с

Выходной сигнал:

стандартно — 4-20 mA; опционально — RS 232, RS 485, Fieldbus, HART

Экран:

графический дисплей, светодиодная подсветка

Температура окружающей среды:

— для блока электроники: -20…+60°С

— для зонда: -40…+80 °С

— возможность комплектации обогреваемыми чехлами, шкафами

Степень защиты (терминальная часть):

IP65

Предел определения:

менее 1 ppm O2

Потребляемое напряжение:

через блок электроники

Напряжение:

230В/50-60Гц ±10%

Потребляемая мощность:

400VA при нагреве 100-200VA при работе

Предохранитель:

10А

Срок эксплуатации:

более 10 лет, в зависимости от применения

Like this post? Please share to your friends:
  • Osteocare original инструкция на русском сироп
  • Otex кондиционер инструкция пульт управления
  • Oximeter c101a2 инструкция на русском языке
  • Osteocare calcium инструкция на русском
  • Oxibd nice инструкция на русском