Pt878 расходомер инструкция на русском

  • Contents

  • Table of Contents

  • Troubleshooting

  • Bookmarks

Quick Links

GE

Sensing & Inspection Technologies

TransPort™ PT878GC

Portable Gas Flowmeter

User’s Manual

loading

Related Manuals for GE TransPort PT878GC

Summary of Contents for GE TransPort PT878GC

  • Page 1
    Sensing & Inspection Technologies TransPort™ PT878GC Portable Gas Flowmeter User’s Manual…
  • Page 2
    Sensing & Inspection Technologies TransPort™ PT878GC Portable Gas Flowmeter User’s Manual 910-229B3 April 2009…
  • Page 3
    Liability under this warranty is limited to restoring the instrument to normal operation or replacing the instrument, at the sole discretion of GE Infrastructure Sensing, Inc. Fuses and batteries are specifically excluded from any liability. This warranty is effective from the date of delivery to the original purchaser.
  • Page 4
    (RAN), and shipping instructions for the return of the instrument to a service center will be provided. 2. If GE Infrastructure Sensing, Inc. instructs you to send your instrument to a service center, it must be shipped prepaid to the authorized repair station indicated in the shipping instructions.
  • Page 5: Table Of Contents

    April 2009 Table of Contents Chapter 1: Features and Capabilities Overview …………1-1 System Description .

  • Page 6
    April 2009 Table of Contents (cont.) Installing Dampening Material ……. . 3-25 Installing DMP-1 Dampening Material .
  • Page 7
    April 2009 Table of Contents (cont.) Chapter 5: Creating and Managing Sites The Site Manager ……….5-2 Creating a New Site .
  • Page 8
    April 2009 Table of Contents (cont.) Changing Date and Time Appearance (Locale) ….7-9 Adjusting the Contrast ……..7-12 Setting Backlight Timeout .
  • Page 9
    April 2009 Table of Contents (cont.) Setting up the Thickness Gauge ……. . 9-5 Measuring Pipe Wall Thickness .
  • Page 10
    April 2009 Table of Contents (cont.) Chapter 11: Specifications Operation and Performance……..11-2 Fluid Types .
  • Page 11
    April 2009 Table of Contents (cont.) Clamp-On Ultrasonic Flow Transducers ……11-6 Temperature Range(s) ……..11-6 Transducer Materials .
  • Page 12: Chapter 1: Features And Capabilities

    April 2009 Chapter 1 Features and Capabilities ® The TransPort Model PT878GC is a transit-time clamp-on gas flowmeter which combines all the features of a full-size gas flowmeter with the advantages of a portable instrument. This section describes the TransPort features and general system, and explains the theory of operation.

  • Page 13
    April 2009 Overview (cont.) The TransPort has the ability to store site data in files which can be accessed at a later time. Within the Main Menu, a set of forms (windows) asks you all the necessary setup information for a particular site.
  • Page 14: System Description

    April 2009 System Description The TransPort is only one part of the flowmeter system. The other part of the system is the flowcell, which consists of the pipe and the transducers. The Flowcell The flowcell is that part of the system that uses ultrasonic pulses to interrogate the flow.

  • Page 15: Electronics Package

    April 2009 Electronics Package The TransPort consists of circuits that generate, receive, and measure the travel time of the ultrasonic pulses. It also contains a microcomputer that controls operation and calculates flow measurement parameters. Specific circuits function as follows: A. TRANSMIT SIGNAL GENERATOR — The transmit signal generator, under control of the microcomputer and timing circuit, synthesizes the signal that drives the transmitter.

  • Page 16: Theory Of Operation

    April 2009 Theory of Operation The TransPort is a transit-time ultrasonic flowmeter. When ultrasonic pulses are transmitted through a moving liquid or gas, the pulses that travel in the same direction as the fluid flow (downstream) travel slightly faster than the pulses that travel against the flow (upstream).

  • Page 17: Chapter 2: Initial Setup

    April 2009 Chapter 2 Initial Setup Before making measurements, you must prepare the TransPort for operation. This includes the following procedures: • Making Electrical Connections • Charging and/or Replacing Batteries • Powering On and Off • Using the Screen and Keypad •…

  • Page 18: Making Electrical Connections

    April 2009 Making Electrical Connections Before making measurements with the TransPort, you must make all the necessary connections to the unit. This section describes how to connect the following: • Power • Transducer • Analog Input/Output • Infrared Interface Make all connections to the top of the TransPort unit as shown in Figure 2-2 below.

  • Page 19: Power Connections

    April 2009 Power Connections The PT878GC is powered by either a 100-120/200-260 VAC wall mount plug-in module, or by 5 internal C -size NiCad high- energy rechargeable batteries or by a pack of 3.0 Ahr NiMH batteries. (An optional power supplement, part #703-1283, uses 6 AA alkaline batteries.) When you receive the PT878GC, the batteries are not charged;…

  • Page 20: The Infrared Wireless Interface

    April 2009 Analog Input/Output Connections (cont.) Table 2-1: Cable Assembly for Analog Inputs/Outputs Pin Number Wire Color Description Black Analog Out 1 16 V (switched) White Input A (Temperature or Pressure) Yellow Input B (Temperature or Pressure) Green Analog Ground Orange Digital Output Blue…

  • Page 21: Charging And/Or Replacing Batteries

    April 2009 Charging and/or Replacing Batteries The PT878GC comes with self-contained, built-in rechargeable batteries to support portable operation. For optimum performance, these batteries require a minimum of maintenance. Charging the Batteries When you receive the PT878GC, you will need to initially charge the batteries.

  • Page 22: Replacing The Batteries

    (see Figure 2-3 below). Battery Location (Behind Panel) Figure 2-3: Rear View of PT878GC To further extend the battery power on the PT878GC, the GE Sensing Part #705-1283 power pack option uses 6 AA alkaline batteries. Initial Setup…

  • Page 23: Powering On And Off

    GE Sensing logo and the software version and emits a long beep. If the meter fails any of these tests, contact GE Sensing. Caution! If the meter fails the backup battery test, you must send the unit back to the factory for a battery replacement.

  • Page 24
    April 2009 Powering On and Off (cont.) After the meter conducts all the self checks, the screen then appears similar to the one shown in Figure 2-4 below. ABC.SIT 2004/11/30 09:53 AM Velocity, ft/s Signal, dB 0.00 Volume, l/s Delta-T, ns 0.10 E0: No Errors Figure 2-4: Screen After Powering On…
  • Page 25
    April 2009 Powering On and Off (cont.) To turn the PT878GC off, press the red key for 3 seconds. The screen now appears similar to Figure 2-5 below. Velocity, ft/s Signal, dB SHUTDOWN: Meter OFF SLEEP: Meter Idle CANCEL: Resume Operations Volume, l/s Delta-T, ns Resume…
  • Page 26: Using The Screen And Keypad

    April 2009 Using the Screen and Keypad The essential features for operating the TransPort are the screen and keypad. Although these features are common on portable instruments, the PT878GC design offers unique features to simplify and speed operation. Screen The primary function of the screen is to display information in order for you to accurately and easily take measurements.

  • Page 27
    April 2009 Screen (cont). The middle of the screen is the work area, which displays the measured parameters, numeric measurements, and both bar and line graphs. (When you enter a selection on the Main Menu discussed in Chapter 4, Programming Site Data, this area displays menu prompts.) A line at the bottom of the area also displays error code messages, which are described in more detail in Chapter 10, Diagnostics and Troubleshooting.
  • Page 28: Keypad

    April 2009 Keypad The PT878GC keypad has 25 keys. The functions for each key are as follows (see Figure 2-7 on the next page): • [F1] [F2] [F3] 3 function keys ( ) — enable you to select the special functions which appear at the bottom of the screen. •…

  • Page 29
    April 2009 Keypad (cont.) Figure 2-7: The TransPort PT878GC Keypad Initial Setup 2-13…
  • Page 30: Obtaining On-Line Help

    April 2009 Obtaining On-Line Help The TransPort offers on-line help screens that describe various features. You can access on-line help at any time by pressing the [?] key. The screen appears similar to Figure 2-8 below. Help Velocity, ft/s Signal, dB Table of Contents Use the arrows and the enter key to select from the links below.

  • Page 31: Chapter 3: Installing The Dampening Material, Transducers And Fixtures

    Since the PT878GC is specifically designed for gas measurement with clamp-on transducers, it requires the use of specially designed fixtures and dampening material to maintain the highest possible measurement accuracy. GE Sensing supplies the CFG series of fixtures: • The V1 clamping fixture for pipes with diameters between 0.75 and 1.25 in.

  • Page 32
    April 2009 CFG-V1 CFG-V1 CFG-V4 CFG-V8 CFG-V8 CFG-V12 Layout Wrap PI Fixture Dummy Block Chain Strap Figure 3-1: The V Series and PI Fixtures Choosing the transducer installation locations on the pipe is an important aspect of proper flow measurement. Specifications assume a fully developed flow profile typically requiring 20 diameters upstream and 10 diameters downstream of straight pipe run.
  • Page 33: Application Requirements

    April 2009 Application Requirements Before you begin to use the PT878GC, you should ensure that your meter can handle the frequencies required for your particular application. To check the capabilities of your particular PT878GC, refer to the back label of the PT878GC, shown in Figure 3-2 below.

  • Page 34
    April 2009 Application Requirements (cont.) • For air, nitrogen, oxygen or argon, refer to Table 3-1 on the next page. • For natural gas, refer to Table 3-2 on page 3-7. • For steam, refer to Table 3-3 on page 3-9. 1.
  • Page 35
    April 2009 Table 3-1: PT878GC Installation Requirements for Air, Nitrogen, Oxygen or Argon Maximum Velocity, ft/s (m/s) Transducer Min. Pressure Single Dual Triple Four Five Pipe Size ANSI (DIN) Pipe Wall Inches (mm) psig (bar) Traverse Traverse Traverse Traverse Traverse 3/4 (20) <0.07 (1.8) 60 (5.1)
  • Page 36
    April 2009 Table 3-1: PT878GC Installation Requirements for Air, Nitrogen, Oxygen or Argon (Continued) Maximum Velocity, ft/s (m/s) Transducer Min. Pressure Single Dual Triple Four Five Pipe Size ANSI (DIN) Pipe Wall Inches (mm) psig (bar) Traverse Traverse Traverse Traverse Traverse 12 (300) <0.38 (9.7)
  • Page 37
    April 2009 Table 3-2: PT878GC Installation Requirements for Natural Gas Maximum Velocity, ft/s (m/s) Pipe Size ANSI Pipe Wall Inches Transducer Min. Pressure (DIN) (mm) psig (bar) Single Traverse Dual Traverse Triple Traverse Four Traverse 2 (50) <0.16 (4.1) 200 (14.8) 110 (33.5) 88 (26.8) 66 (20.5)
  • Page 38
    April 2009 Table 3-2: PT878GC Installation Requirements for Natural Gas (Continued) Maximum Velocity, ft/s (m/s) Pipe Size ANSI Pipe Wall Inches Transducer Min. Pressure (DIN) (mm) psig (bar) Single Traverse Dual Traverse Triple Traverse Four Traverse 18 (450) <0.38 (9.7) 300 (21.7) 78 (23.8) 59 (18.0)
  • Page 39
    April 2009 Table 3-3: PT878GC Installation Requirements for Steam Maximum Flow Velocity Pipe Pipe Wall Min. ft/s (m/s) Size in. Inches Transducer Pressure Single (mm) (mm) (MHz) psig (bar) Traverse 3 (80) <0.22 (5.6) 110 (8.6) 120 (36.6) <0.3 (7.6) 200 (14.8) 4 (100) <0.24 (6.1)
  • Page 40: Preparing The Pipe

    April 2009 Preparing the Pipe 1. Locate a transducer measurement point with 20 diameters of upstream straight run pipe and 10 diameters of downstream straight run pipe. In addition, the point should be at least 10 diameters from any butt welds or flanges. Keep appropriate clearance on either side of the pipe for easy transducer installation: •…

  • Page 41: Performing A Pipe Survey

    April 2009 Performing a Pipe Survey Finding a location where the pipe is concentric is important for optimum accuracy and performance. If possible, perform a pipe survey with an ultrasonic thickness gauge to find the best location. 2. Clear rust or loose paint and measure the wall thickness at five points along the pipe axis at 25 mm (1-in.) intervals using an ultrasonic thickness gauge, as shown in Figure 3-4 below.

  • Page 42
    April 2009 Performing a Pipe Survey (cont.) 4. Now measure the OD and the wall thickness at eight points along the pipe circumference at 45° intervals (shown in Figure 3-6 below), three times per point, and record the mean values. Figure 3-6: Measuring the OD and Pipe Wall Thickness 5.
  • Page 43: Obtaining The Transducer Spacing

    April 2009 Obtaining the Transducer Spacing 1. Using the measured OD and the pipe wall thickness, program the PT878GC (discussed in Chapter 4, Programming Site Data) to determine the required transducer spacing. 2. To determine the PT878GC correction factor, calculate the mean inside pipe diameter (ID) and the pipe ID at the transducer locations.

  • Page 44: Installing The V Series Clamping Fixture And Transducers

    9 o’clock position. Figure 3-7 below shows the two mounted halves. Note: If you are using a V4 fixture with a 1.5 to 2-in. pipe, GE Sensing suggests removing the set screws and the mounting bracket, rotating them and the transducer yoke 180°, and replacing the bracket, yoke and screws for a…

  • Page 45
    April 2009 Installing the Fixture (cont.) The two fixture halves have measuring scales; ensure that the scales are on the same side of the fixture, so that both zeros start at the same origin, as shown in Figure 3-8 below. Figure 3-8: Fixture with Scale Origins Properly Aligned 3.
  • Page 46: Installing The Transducers

    April 2009 Installing the Transducers 1. Apply a bead of couplant 6 mm (0.25 in.) wide along the entire length of each transducer face, as shown in Figure 3-10 below. Figure 3-10: Couplant on Transducer Face Note: Do not slide the transducer with couplant along the surface of the pipe when mounting the transducer.

  • Page 47
    April 2009 Installing the Transducers (cont.) 3. Slide the second mounting block to the calculated spacing plus the initial scale number selected for the first mounting block. For example: a. Initial convenient number for the first mounting block = 5 cm or 2 in. b.
  • Page 48: Installing The Pi Fixture And Transducers

    The PI clamping fixture holds transducers on pipes from 8 to 24 in. in diameter. It comes with either a chain or strap, depending on the selection made with the initial order from GE Sensing. To install the fixture and transducers, complete the following steps: Surveying the Pipe 1.

  • Page 49
    April 2009 Surveying the Pipe (cont.) Figure 3-14: Marking the 3 o’Clock Position 4. To find the coinciding point on the opposite site of the pipe (180° away from each other), divide the measured circumference by 2 and measure this distance along the circumferential lines from the zero point, as shown in Figure 3-15 below.
  • Page 50
    April 2009 Surveying the Pipe (cont.) Figure 3-16: Measuring the 180° Point from Top and Bottom 3-20 Installing the Dampening Material, Transducers and Fixtures…
  • Page 51: Installing The First Bracket With A Chain Or Strap

    April 2009 Installing the First Bracket with a Chain or Strap The following steps describe how to install the PI fixture with a supplied chain or strap. 1. Carefully wrap the chain or strap around the pipe, taking care not to twist it. 2.

  • Page 52: Installing The Second Bracket With A Chain Or Strap

    April 2009 Installing the Second Bracket with a Chain or Strap Note: The following step requires the transducer spacing discussed on page 3-7. 1. Measure the spacing from the zero point (the point of circumferential origin). Mark the spacing point with a crosshair on the opposite side of the pipe, 180°…

  • Page 53: Installing The Transducers

    April 2009 Installing the Second Bracket (cont.) The spacing should now appear similar to that shown in Figure 3-21 below. Figure 3-21: The Installed CFG-PI Fixture, with Calculated Spacing Installing the Transducers 1. Check to be sure the second CFG-PI holder bracket is correctly positioned.

  • Page 54
    April 2009 Installing the Second Bracket (cont.) Note: Do not slide the transducer with couplant along the surface of the pipe when mounting the transducer. 3. With one hand, mount one transducer into the PI fixture. With the other hand, tighten the transducer hold down screw, gradually pushing the transducer down to the pipe surface.
  • Page 55: Installing Dampening Material

    The pipe is old, with a history of scaling or rust, • The pipe experiences condensation on the outside. Note: Consult a GE Sensing flowmeter applications engineer or sales engineer if you have any questions regarding dampening material. Installing the Dampening Material, Transducers and Fixtures…

  • Page 56: Installing Dmp-1 Dampening Material

    April 2009 Installing DMP-1 Dampening Material 1. Be sure the CFG-V clamping fixture is installed on the pipe with the transducers as described in the section Installing the V Series Clamping Fixture and Transducers on page 3-14. With a marker, mark scribe lines on the inside edges of the brackets onto the pipe.

  • Page 57
    April 2009 Installing DMP-1 Dampening Material with CFG-V Series Fixtures (cont.) 7. With a marker, trace around the transducer footprint, as shown in Figure 3-24 below. Figure 3-24: DMP-1 Material with Transducer Footprint Traced 8. Remove the fixture and transducers. Then use a utility knife to cut out the area under the transducer footprint, and peel the cut material off the pipe, as shown in Figure 3-25 below.
  • Page 58
    April 2009 Installing DMP-1 Dampening Material with CFG-V Series Fixtures (cont.) 11. Lay the second strip of dampening material flat. Then cut the strip into two pieces, each 4.5-in. (12 cm) wide. 12. Wrap each of these strips around the pipe on the outside edge of the clamping fixture, one upstream and one downstream.
  • Page 59: Installing Dmp-1 Dampening Material With Pi Fixture

    April 2009 Installing DMP-1 Dampening Material with PI Fixture 1. Be sure the PI clamping fixture is installed onto the pipe with the transducers as described in the section Installing the PI Fixture and Transducers on page 3-18. 2. Approximate the axial distance from one transducer face to the other transducer face, as shown in Figure 3-27 below.

  • Page 60
    April 2009 Installing DMP-1 Dampening Material with PI Fixture (cont.) Figure 3-28: Completed DMP-1 Installation 3-30 Installing the Dampening Material, Transducers and Fixtures…
  • Page 61: Installing Dmp-3 Compound With All Fixtures

    April 2009 Installing DMP-3 Compound with All Fixtures 1. Be sure the fixture and transducers are installed on the pipe as described in the previous sections of this chapter. 2. Remove the fixture and transducers, but be sure to mark the approximate area of installation.

  • Page 62
    April 2009 Installing DMP-3 Compound with All Fixtures (cont.) 6. Spread the DMP-3 material so that it covers the whole area under the fixture to a thickness of about 0.25 in. (6.4 mm), as shown in Figure 3-31a below. Figure 3-31: Spreading the DMP-3 (a) and Reinstalling the Fixture (b) 7.
  • Page 63: Installing The Pdj Dampening Jacket

    April 2009 Installing the PDJ Dampening Jacket If the pipe temperature is over 150°F, you must use the PDJ pipe dampening jacket with preapplied DMP-3. As the dampening material dries out over several hours after installation, its effectiveness increases. The jacket is available in standard pipe sizes from 4 to 12 in.

  • Page 64
    April 2009 Installing the PDJ Dampening Jacket (cont.) !WARNING! The pipe and the dripping fluid will cause severe burns upon contact with bare skin. Also, be sure not to inhale the fumes generated during the DMP-3 curing cycle. 4. Install the fixture over the jacket, adjusting the spacing to match the prestamped transducer holes and PT878GC spacing calculations.
  • Page 65: Chapter 4: Programming Site Data

    April 2009 Chapter 4 Programming Site Data On the PT878GC, a Program Menu (part of the Main Menu) enables you to enter information that is specific to each site: • Transducer types and paths • Pipe materials and linings • Fluid types •…

  • Page 66: Entering The Program Menu

    April 2009 Entering the Program Menu [MENU] To enter the Program Menu, press the key at the lower right of the PT878GC keypad. The Menu Bar replaces the Status Bar at the top of the screen. Press the [ ] arrow key once to scroll from the Site Menu to the Program Menu.

  • Page 67: Entering Transducer Parameters

    Use the [ ] and [ ] keys to scroll between the two types. b. Press [ENTER] [SEL] to confirm the choice. IMPORTANT: While wetted transducers are listed as a choice, the PT878GC is designed for use with Shear wave clamp-on transducers. Consult GE before selecting wetted transducers. Programming Site Data…

  • Page 68
    April 2009 Entering Transducer Parameters (cont.) Note: The choices made earlier in the Transducer and Pipe options determine the prompts available later. If the PT878GC does not scroll to a particular parameter, it is not necessary for that transducer or pipe type. 2.
  • Page 69: Parameters For Special Transducers

    From the Frequency prompt, press the [ ] key to reach the [ENTER] Tw prompt, and press b. Use the numeric keys to enter the GE Sensing-supplied [ENTER] time delay and press 5. When calculating the flow rate, the PT878GC must take into account the wedge angle, the angle of the ultrasonic transmission.

  • Page 70
    ] key to reach [ENTER] the Wedge SS prompt, and press b. Use the numeric keys to enter the GE Sensing-supplied [ENTER] wedge sound speed (in ft/sec or m/sec) and press Pressing the [ ] key returns the meter to the Transducer tab at the top.
  • Page 71: Entering Pipe Parameters

    April 2009 Entering Pipe Parameters To enter the Pipe option, scroll to the Pipe entry on the Program [ENTER] Menu and press . (From the Transducer window, you can scroll back up to the Transducer tab and press the [ ] arrow key [ENTER] to reach the Pipe window and press .) The screen appears…

  • Page 72
    April 2009 Entering Pipe Parameters (cont.) 1. The first prompt asks you to select the pipe material. a. Press [ENTER] to enter the material prompt. b. A drop-down list of materials opens. Table 4-1 below lists the available preprogrammed materials on the list. Press the [ ] or [ ] keys to scroll to the appropriate material,…
  • Page 73
    April 2009 Entering Pipe Parameters (cont.) 2. At this point, the program follows one of two paths: a. If you have selected a preprogrammed material, the PT878GC automatically supplies the correct sound speed, and proceeds to step 3. b. If you have selected “Other,” the meter prompts you to enter the sound speed.
  • Page 74
    April 2009 Entering Pipe Parameters (cont.) If you select a material that has the Schedule option: a. The prompt asks if you wish to apply DIN (the DIN [ENTER] schedule). Press to select (or deselect) the DIN box. (If you do not select the DIN option, the prompt moves to the OD text box, and you enter the parameters for the Diameter alternative as discussed on the previous page.)
  • Page 75: Entering Pipe Lining Parameters

    April 2009 Entering Pipe Lining Parameters To enter the Lining option, scroll to the Lining entry on the [ENTER] Program Menu and press . (From the Pipe window, you can scroll back up to the Pipe tab and press the [ ] arrow key to [ENTER] reach the Lining window, and press .) The screen appears…

  • Page 76
    April 2009 Entering Pipe Lining Parameters (cont.) Note: If your pipe lining is not on the drop-down list, consult GE Sensing for further information. 2. The program follows one of two paths: a. If you have selected a preprogrammed material, the PT878GC automatically supplies the correct sound speed, and you proceed to step 3.
  • Page 77: Entering Fluid Types And Speeds

    April 2009 Entering Fluid Types and Speeds To access the Fluid option, scroll to the Fluid entry on the [ENTER] Program Menu and press . (If you are already in the Transducer/Pipe form, press the [ ] arrow key to reach the Fluid [ENTER] window, and press .) The screen appears similar to Figure…

  • Page 78
    April 2009 Entering Fluid Types and Speeds (cont.) b. Scroll to the appropriate entry. If you do not see your fluid on the list, select “Other.” c. Press [ENTER] to confirm your selection. If you have selected a preprogrammed gas, the PT878GC supplies the sound speed.
  • Page 79: Entering The Signal Path Parameters

    April 2009 Entering the Signal Path Parameters To enter the Path option, scroll to the Path entry on the Program [ENTER] Menu and press . (From the Lining window, you can scroll back up to the Lining tab and press the [ ] arrow key to reach the [ENTER] Path window, and press .) The screen appears similar to…

  • Page 80: Path Parameters For Clamp-On Transducers

    If you enter an invalid entry, the PT878GC rejects the entry and displays an error message. Note: GE Sensing does not recommend using a spacing other than the one calculated by the PT878GC. After you enter the spacing, pressing the […

  • Page 81: Entering The Standard Volume

    April 2009 Entering the Standard Volume/ Mass Flow Parameters The Standard Volume/Mass Flow option enables you to correct standard volumetric units for pressure and temperature, as well as to disable or enable mass flow. To enter this option, scroll to the [ENTER] Std Vol/Mass Flow entry on the Program Menu and press The screen appears similar to Figure 4-7 below.

  • Page 82
    April 2009 Entering the Standard Volume Parameters (cont.) 2. The next prompt asks for the base pressure. a. Press [ENTER] to enter the text box. b. Use the numeric keys to enter the known pressure. c. Press [ENTER] to confirm the entry. 3.
  • Page 83: Entering Inputs In The Standard Volume Option

    April 2009 Entering Inputs in the Standard Volume Option To enter input parameters in the Standard Volume/Mass Flow option, return to the Settings tab at the top of the Standard [ENTER] Volume window. Press the [ ] arrow key and to open the Inputs window.

  • Page 84
    April 2009 Entering Inputs in the Standard Volume Option (cont.) 2. The next prompt appears if you have selected a fixed pressure. The PT878GC asks for the fixed pressure value. a. Press [ENTER] to open the text box. b. Use the numeric keys to enter the known pressure. c.
  • Page 85
    April 2009 Entering Inputs in the Standard Volume Option (cont.) Note: If you have switched both active inputs to #A (or to #B), the PT878GC automatically switches one input to the other letter. This change ensures that only one analog input is assigned to one measurement type.
  • Page 86: Entering Mass Flow Parameters

    April 2009 Entering Mass Flow Parameters To enter mass flow parameters in the Standard Volume/Mass Flow option, press the [ ] arrow key until you reach the Mass [ENTER] Flow tab, and then press to open the Mass Flow window. The screen appears similar to Figure 4-9 below.

  • Page 87
    April 2009 Entering Mass Flow Parameters (cont.) 2. If you have enabled Mass Flow, the next prompt asks for the calculation method. Three options are available: static density, saturated steam, and steam. a. Press [ENTER] to open the drop-down menu. b.
  • Page 88: Entering Skan/Measure Parameters

    Entering Skan/Measure Parameters IMPORTANT: Do not change the Skan/Measure parameters unless recommended by GE Sensing. If you select one of the Skan/Measure modes in the Measurement mode window in the Signal Setup option (see page 9-24), the PT878GC requires you to set up the appropriate parameters. You can enter this option in either of two ways: •…

  • Page 89
    M>S Switch value. The default value is 5.0 µsec and values from 0 to 250 µsec are acceptable. Note: Do not change this value without consulting GE Sensing. If you wish to use Skan only, set the M/S Switch to a value of 0.
  • Page 90
    April 2009 Entering Skan/Measure Parameters (cont.) Pressing the [ ] key highlights the Advanced prompt. To enter [ENTER] advanced parameters for the Skan/Measure modes, press The screen now appears similar to Figure 4-11 below. In each case, to enter a value: a.
  • Page 91
    April 2009 Entering Skan/Measure Parameters (cont.) 6. The next prompt enables you to reset the size of the transmit window. Normally, the PT878GC calculates the size of both transmit and receive windows based on pipe size and fluid sound speed. However, for special diagnostic purposes, it may sometimes be necessary to reset the window size.
  • Page 92: Entering Skan/Measure Integrate Parameters

    April 2009 Entering Skan/Measure Integrate Parameters To enter parameters for the Skan/Measure Integrate mode in the Skan/Measure option, press the [ ] arrow key once to reach the [ENTER] Integrate tab. Then press to open the Integrate window, which appears similar to Figure 4-12 below. Skan/Measure Integrate Count…

  • Page 93
    April 2009 Entering Skan/Integrate Parameters (cont.) 3. If you wish to use Measure Times only (instead of both Skan [ENTER] and Measure times), press at the Measure Only [ENTER] prompt. (You can also press to deselect the prompt.) Pressing the [ ] key returns the meter to the A Divisor entry, and pressing the [ ] key returns the meter to the Integrate tab.
  • Page 94: Entering Skan/Measure Count Parameters

    April 2009 Entering Skan/Measure Count Parameters To open the Skan/Measure Count window in the Skan/Measure option, press the [ ] arrow key twice (from the All tab) to reach [ENTER] the Count tab. Press , and the screen appears similar to Figure 4-13 below.

  • Page 95
    April 2009 Entering Skan/Measure Count Parameters (cont.) 3. The next prompt asks for the percentage of Noise Level. a. Press [ENTER] to enter the text box. b. Use the numeric keys to enter the recommended value. c. Press [ENTER] to confirm the entry. 4.
  • Page 96: Entering Skan/Correlation Parameters

    April 2009 Entering Skan/Correlation Parameters To open the Skan/Correlation window in the Skan/Measure option, press the [ ] arrow key three times (from the All tab) to [ENTER] reach the Correlate tab, and press . The screen appears similar to Figure 4-14 below. Skan/Measure Integrate Count…

  • Page 97
    April 2009 Entering Skan/Correlation Parameters (cont.) 2. Repeat step 1 to enter the maximum percentage of peak. 3. The next prompt asks for the Percent of Peak, which is used to calculate the transit times and Delta-T. a. Press [ENTER] to enter the text box.
  • Page 98: Entering Analog Inputs

    April 2009 Entering Analog Inputs The Analog Inputs option enables you to specify parameters for general purpose or standard volume inputs. To enter the Analog Inputs option, scroll to the Analog Inputs entry on the Program [ENTER] Menu and press .

  • Page 99
    April 2009 Entering General-Purpose Analog Inputs (cont.) 2. The next prompt asks you to create a label for the input. The label corresponds to the measurement type (i.e., velocity or temperature), while the units symbol corresponds to the measurement units (i.e., feet/sec or degrees F). a.
  • Page 100
    April 2009 Entering General-Purpose Analog Inputs (cont.) 3. The next prompt asks for the Units Symbol. a. Press [ENTER] to reopen the text creation window. b. Repeat the procedure used for the Label (on the previous page) to create the Units Symbol. c.
  • Page 101: Entering Analog Inputs In Standard Volume

    April 2009 Entering Analog Inputs in Standard Volume or Mass Flow Option If you have enabled the Standard Volume or Mass Flow options, the screen displays different options, as shown in Figure 4-17 below. Analog Input Input A Input B Pressure Function Label…

  • Page 102
    April 2009 Entering Analog Inputs in the Standard Volume or Mass Flow Option (cont.) You have completed entering parameters in the Analog Inputs option. • [F3] To confirm the entries and return to Operate mode, press (OK). The PT878GC returns to Operate Mode. •…
  • Page 103: Entering The Analog Output

    April 2009 Entering the Analog Output The Analog Output option enables you to enter information to set up output parameters. To enter the option, scroll to the Analog [ENTER] Output entry on the Program Menu and press . The screen appears similar to Figure 4-18 below.

  • Page 104
    April 2009 Entering the Analog Output (cont.) 2. The next prompt asks you to select the analog output type from a number of choices, as shown in Figure 4-19 below. Select Measurement Diagnostics Velocity Temperature Pressure Volume UserFunction Aux Input A Aux Input B Skan/Measure Mass Flow…
  • Page 105
    April 2009 Entering the Analog Output (cont.) 3. The next prompt asks you to enter the zero (base) value for the analog output. This value represents the 0/4 mA output (in flow units). a. Press [ENTER] to open the text box. b.
  • Page 106: Entering The Digital Output

    April 2009 Entering the Digital Output While resembling the Analog Output option, the Digital Output option enables you to set up parameters necessary for a digital output. To enter the option, scroll to the Digital Output entry on [ENTER] the Program Menu and press .

  • Page 107
    April 2009 Entering the Digital Output (cont.) a. Press [ENTER] to open the drop-down menu. b. Scroll to the desired function. c. Press [ENTER] to confirm your selection. Note: If you select Off, you will not be able to access any other parameters in this option.
  • Page 108
    April 2009 Entering the Digital Output (cont.) d. The prompt then moves to a list of unit types. (The available units depend on the selection made at the Data Source prompt.) Scroll to the desired output unit. e. Press [F3] (OK) to confirm your selection.
  • Page 109
    April 2009 Entering the Digital Output (cont.) If you entered Test Points: The prompt asks for the signal type from two choices: transmit or receive. a. Press [ENTER] to open the drop-down menu. b. Scroll to the required signal. c. Press [ENTER] to confirm your selection.
  • Page 110: Entering User Functions

    April 2009 Entering User Functions User functions enable you to perform mathematical equations on each measurement. You can also use any parameter in the meter to calculate a different parameter. To enter the User Functions option, scroll to the User Functions entry on the Program Menu [ENTER] and press .

  • Page 111
    April 2009 Entering User Functions (cont.) 2. The next prompt asks you to create a label for the function. The label corresponds to the measurement type (i.e., velocity or temperature), while the units symbol corresponds to the measurement units (i.e., feet/sec or degrees F). a.
  • Page 112
    April 2009 Entering User Functions (cont.) 3. The next prompt asks for the Units Symbol. a. Press [ENTER] to reopen the text creation window. b. Repeat the procedure used for the Label (on the previous page) to create the Units Symbol. c.
  • Page 113
    When you have completed entering the function, press [F3] (OK) to confirm the entry and return to the User Function window. 6. GE Sensing recommends pressing [F2] (Check) to test the validity of the function. The PT878GC displays either “OK”…
  • Page 114: Entering Correction Factors

    April 2009 Entering Correction Factors The final option in the Program Menu, Correction Factors, allows you to enter and modify three correction factors: Reynolds Correction, Kinematic Viscosity and Calibration Factor. To enter the Correction Factors option, scroll to the Correction Factors [ENTER] entry on the Program Menu and press .

  • Page 115
    April 2009 Entering Reynolds Correction (cont.) b. Press [ENTER] to confirm your selection. Note: If you disable Reynolds Correction, you will not be able to enter any other values in this window. 2. If you enable Reynolds Correction, you will also need to enter the Kinematic Viscosity of the gas (available in the brochure Sound Speeds and Pipe Size Data, 914-004).
  • Page 116
    April 2009 Entering Reynolds Correction (cont.) b. Enter the desired measurement and units as described on page 4-40. c. Press [F3] (OK) to confirm your selection. 4. The prompt then asks if you wish to edit the Reynolds Correction table. a.
  • Page 117: Entering A Calibration Factor

    April 2009 Entering Reynolds Correction (cont.) e. When you have completed entering values, press [F3] (OK) to confirm the table and return to the Correction Factors window. You have completed entering data for a Reynolds Correction table. • [F3] To confirm the entries, press (OK).

  • Page 118
    April 2009 Entering a Calibration Factor (cont.) 1. The first prompt asks if you wish to enable the calibration factor. a. Use the [ ] and [ ] keys to scroll to the appropriate radio button. b. Press [ENTER] to confirm your selection. 2.
  • Page 119
    April 2009 Entering a Calibration Factor (cont.) b. Enter the desired measurement and units as described on page 4-40. c. Press [F3] (OK) to confirm your selection. 4. The prompt then asks if you wish to edit the K factor table. a.
  • Page 120
    April 2009 Entering a Calibration Factor (cont.) You have completed entering data for correction factors. • [F3] To confirm the entries and return to Operate Mode, press (OK). • To leave the window without confirming the entries and return [F2] [ESC] to Operate mode, press (Cancel) or the…
  • Page 121: Chapter 5: Creating And Managing Sites

    April 2009 Chapter 5 Creating and Managing Sites The PT878GC can store site data in files for current and future access. (To learn how to program setup data, refer to Chapter 4, Programming Site Data.) After you answer the necessary questions, simply save the information to a site file.

  • Page 122: The Site Manager

    April 2009 The Site Manager [ENTER] From the Site Menu, press to open the Site Manager. The screen appears similar to Figure 5-2 below. Site Manager File Site Sort Info: DEFAULT:SIT DEFAULT New Site **CURRENT SITE** 03/04/04 14:51:10 5111 bytes 118272 bytes free Refresh Exit…

  • Page 123
    April 2009 The Site Manager (cont.) Site Manager File Site Sort Info: DEFAULT:SIT DEFAULT Open New Site Save Save As Refresh Rename 09/04/04 14:51:10 Delete 5111 bytes 118272 bytes free Refresh Exit Figure 5-3: The Site Manager Menu • To create a new site, go to page 5-4. •…
  • Page 124: Creating A New Site

    April 2009 Creating a New Site [MENU] To create a new site in the Site Manager, press in the Site Manager. Be sure the option New is highlighted, and then press [ENTER] . The screen appears similar to Figure 5-4 below. New Site Delete Cancel…

  • Page 125
    April 2009 Creating a New Site (cont.) New Site New Site Use the selected site ‘DEFAULT.SIT’ as a template? Figure 5-5: Template Confirmation Window The meter returns to Operate Mode, with the new site name displayed in the upper left corner of the screen. Creating and Managing Sites…
  • Page 126: Opening An Existing Site

    April 2009 Opening an Existing Site If you want to return to a second site that you have previously saved, first highlight the replacement site in the left window of [MENU] the Site Manager. Then press to open the File Menu. [ENTER] Scroll to the Open option, and press .

  • Page 127: Saving A Site

    April 2009 Saving a Site You can save the current site in one of two ways: • From the Site Menu, you can scroll down to the Save Now [ENTER] option and press , or • [MENU] From within the Site Manager, press , scroll to the Save [ENTER] option, and press…

  • Page 128: Saving A Site With A Different Name

    April 2009 Saving a Site with a Different Name If you want to save the current site with a different name, open [MENU] the Site Manager, press to open the File Menu, scroll to [ENTER] the Save As option, and press .

  • Page 129: Refreshing A Site

    April 2009 Refreshing a Site You can refresh a site (updating the display with the most current information) in one of two ways: • [F2] Press , Refresh, to refresh the highlighted file. • From the File Menu, scroll to the Refresh option and press [ENTER] The updated information on the highlighted file appears in the window on the right.

  • Page 130: Deleting A Site

    April 2009 Deleting a Site To delete a site in the Site Manager, first be sure you have highlighted that site in the left window of the Site Manager. Then [MENU] [ENTER] press , scroll to the Delete option, and press .

  • Page 131: Creating A Site Message

    April 2009 Creating a Site Message The Site Message option allows you to add an explanatory message (with up to 30 characters or spaces) for any given site. To create a site message: 1. Press [MENU] to enter the File Menu. Then press the [ ] arrow key once to scroll from the File Menu to the Site Menu.

  • Page 132
    April 2009 Creating a Site Message (cont.) “ ‘ < > Delete Cancel Figure 5-12: The Text Creation Window 3. Use the four arrow keys to scroll to the desired letter or [ENTER] symbol, and press to add the letter to the message. Note: Pressing [SEL] causes the screen to alternate between a…
  • Page 133: Printing A File

    April 2009 Printing a File [MENU] To print a file from the Site Manager, press to enter the File Menu, press the [ ] arrow key once, scroll to the Print [ENTER] option, and press . The PT878GC shows a message indicating that its infrared scanner is looking for a receiving device.

  • Page 134: Transferring A File As Text

    April 2009 Transferring a Site to a PC (cont.) Figure 5-13: The IR_Inbox Folder with Transferred Site File Transferring a File as Text If you want to review site programming data as text, the PT878GC can transfer a site file to a PC in a Unicode text format that can be opened by text editors such as Notepad or Word.

  • Page 135: Transferring A File From A Pc To The Pt878Gc

    April 2009 Transferring a File from a PC to the PT878GC Once you have stored site or meter files to a PC, you can then transfer them back to the PT878GC over the IR interface. The PT878GC only accepts files with a .sit (site) or .met (meter) extension.

  • Page 136
    April 2009 Transferring a File from a PC to the PT878GC (cont.) QuickBeam opens a window indicating that it is downloading the file. If you close and reopen the Site Manager, the file appears as one of the listed sites. Note: While sending or receiving files, the PT878GC continues to perform measurements, but at a slower rate than normal.
  • Page 137: Listing Files By Name

    April 2009 Listing Files by Name If you want to list your files alphabetically by site name within [MENU] the Site Manager, press to open the File Menu. Then press the [ ] arrow key twice to scroll from the File Menu to the Sort [ENTER] Menu, shown in Figure 5-15 below.

  • Page 138: Chapter 6: Displaying And Configuring Data

    April 2009 Chapter 6 Displaying and Configuring Data The PT878GC allows you to view from one to four different measurement parameters simultaneously. The screen can show these parameters not only in numeric format, but as line or bar graphs as well. You can configure any given measurement for your particular requirements.

  • Page 139: The Format Option

    April 2009 The Format Option The first option on the menu, Format, allows you to specify the type of notation and the number of decimal places for the [ENTER] measurement in that window. Press and the submenu shows two options, Decimal Place and Notation. Press the [ ] or [ENTER] ] arrow keys to select either entry, and press…

  • Page 140
    April 2009 The Format Option (cont.) 2. The second step asks you to choose the number of decimal places to be displayed. Available choices range from 0 to 4 places. a. Press [ENTER] to open the drop-down list. b. Scroll to the appropriate number. c.
  • Page 141: The View Option

    April 2009 The View Option The second option, View, allows you to select the presentation of a parameter in one of three formats: numeric, line graph or bar graph. 1. From the Display Menu, press the [ ] key once to reach the [ENTER] View option, and then press 2.

  • Page 142: The Limits Option

    April 2009 The Limits Option Once you have configured a parameter as a line or bar graph, you might need to change its presentation or values. The Limits option (replacing the Format option for line and bar graphs) enables you to program the minimum or maximum values displayed, the time interval and the display of the average value.

  • Page 143
    April 2009 The Limits Option (cont.) 4. The first prompt asks for the minimum value shown in the graph. a. Press [ENTER] to open the text box. b. Use the numeric keys to enter the desired value. c. Press [ENTER] to confirm the value.
  • Page 144: The Measurement Option

    April 2009 The Measurement Option On occasion, you might need to change the actual parameter measured in a given window.The Measurement option enables you to reconfigure the window with one of five categories of data source (velocity, volume, forward or reverse totalizer, or diagnostics) and appropriate English or metric measurement units.

  • Page 145
    April 2009 The Measurement Option (cont.) Select Measurement Programming Diagnostics Delta-T Delta-T Velocity Amplitude Up Amplitude Down Temperature T Up Pressure T Down Volume Gain Up Fwd Totalizer Gain Dn Rev Totalizer Signal Up StdFwdTotalizer Signal Down Threshold Up StdRevTotalizer Threshold Down User Function Norm Factor…
  • Page 146: Customizing The Display Screen

    April 2009 Customizing the Display Screen You might wish to display one or two parameters, or customize the soft keys to quickly access particular menus. The Site Menu enables you to make more comprehensive changes in your display screen. [MENU] To enter the Site Menu, press the key at the lower right of the PT878GC keypad.

  • Page 147: Specifying The Number Of Displayed Parameters

    April 2009 Specifying the Number of Displayed Parameters As mentioned earlier, the PT878GC can display one to four different measurement parameters simultaneously. However, sometimes you might wish to display only one or two parameters. To change the number of open display windows: 1.

  • Page 148: Customizing Softkeys

    April 2009 Customizing Softkeys When the screen is in Operate mode, you might wish to access a particular submenu frequently without the trouble of scrolling through menus. Customizing the softkeys (the function keys — [F1] [F2] [F3] ) allows you to access up to three submenus by pressing the associated softkey.

  • Page 149
    April 2009 Customizing Softkeys (cont.) Configure FKey Select a menu item (CLEAR) Site Manager Save Now 1 View 2 Views Cancel Figure 6-9: The Configure FKey Window 4. Press [ENTER] to open the menu drop-down list. Then use the ] or [ ] arrow keys to scroll to the desired submenu (for example, Contrast).
  • Page 150
    April 2009 Customizing Softkeys (cont.) 2004/3/30 09:53 AM ABC.SIT Velocity, m/s Gain, dB 5000 -1.5121 -5000 Volume, ft Delta-T, ns -1.65 E2: Soundspeed Contrast Figure 6-10: The Display Screen with Modified Softkey • To clear a softkey: 1. Press [MENU] 2.
  • Page 151: Managing Files — The Drive Manager

    April 2009 Managing Files — The Drive Manager On occasion, you might want to review, print or transfer some or all of the files in the PT878GC. The Drive Manager allows you to view all the files stored in the meter. To open the Drive Manager: 1.

  • Page 152
    April 2009 Managing Files — The Drive Manager (cont.) File Manager Sort File Transfer Info: SYSLOG.MET SYSLOG. Refresh DEFAULT. Delete GLOBAL. TABLE.MET FUNCTIO AAA.LOG 03/15/04 10:39:20 BBB.LOG 6402 bytes CCC.LOG 91136 bytes free Q.LOG Refresh Exit Figure 6-12: The File Manager Menu To transfer, refresh or delete a file, first be sure the file is [MENU] highlighted in the window on the left.
  • Page 153: Refreshing A File

    April 2009 Refreshing a File You can refresh a file (updating the display with the most current information) in one of two ways: • [F2] Press , Refresh, to refresh the highlighted file. • From the File Menu, scroll to the Refresh option and press [ENTER] The updated information on the highlighted file appears in the window on the right.

  • Page 154: Transferring A File From A Pc To The Pt878Gc

    April 2009 Transferring a File to a PC (cont.) Figure 6-13: The IR_Inbox Folder with Transferred File Transferring a File from a PC to the PT878GC Once you have stored site or meter files to a PC, you can then transfer them back to the PT878GC over the IR interface.

  • Page 155
    April 2009 Transferring a File from a PC to the PT878GC (cont.) Note: For Windows NT 4.0, check that the QuickBeam software (available through the path C:\Program Files\QuickBeam Suite) is running, and that the IR beam on the PT878GC has clear access to the IR sensor connected to the PC port.
  • Page 156: Deleting A File

    April 2009 Deleting a File To delete a file in the File Manager, first be sure you have highlighted that site in the left window of the File Manager. Then [MENU] [ENTER] press , scroll to the Delete option, and press .

  • Page 157: Listing Files By Name

    April 2009 Listing Files by Name If you want to list your files alphabetically by site name within [MENU] the File Manager, press to open the File Menu. Then press the [ ] arrow key to scroll from the File Menu to the Sort Menu, [ENTER] shown in Figure 6-16 below.

  • Page 158: Accessing Meter Data -The About Option

    To open the About window, scroll to the About option on the Site Menu and press [ENTER] . The screen appears similar to Figure 6-17 below. GE Industrial Sensing PT878GC Portable Flowmeter Copyright ©2002 General Electric Co. All rights reserved.

  • Page 159: Chapter 7: Programming Meter Settings

    April 2009 Chapter 7 Programming Meter Settings Along with display formats and site data, PT878GC users can program global settings for the meter that suit their individual preferences. The global settings include: • English or Metric measurement units • Battery power •…

  • Page 160: Entering The Meter Menu

    April 2009 Entering the Meter Menu [MENU] To enter the Meter Menu, press the key at the lower right of the PT878GC keypad. The Main Menu replaces the Status Bar at the top of the screen. Press the [ ] arrow key twice to scroll from the Site Menu to the Meter Menu.

  • Page 161: Selecting Measurement Units

    April 2009 Selecting Measurement Units The first option, Units, enables you to select either English or Metric units as global measurement units for the PT878GC. The selected units then become the default settings for every measurement that has the option for metric/English units. In addition, you can select the type of pressure units displayed.

  • Page 162
    April 2009 Selecting Measurement Units (cont.) a. Press [ENTER] to open the drop-down menu. b. Use the [ ] and [ ] arrow keys to scroll to the desired choice: • PSIA • • Bars • MMGH c. Press [ENTER] to confirm the entry.
  • Page 163: The Battery Charger

    April 2009 The Battery Charger The Battery option allows you to monitor the current run time and status of the internal rechargeable batteries, as well as to condition NiCad batteries to maintain the maximum life possible. Conditioning NiCad batteries (a process that can take up to 12 hours for a fully charged pack) discharges the pack completely and then performs a fast charge.

  • Page 164
    April 2009 The Battery Charger (cont.) 3. To condition the batteries, press the [ENTER] key. The “Condition Battery” window should now read “Stop.” The status line should change to “Discharging.” (Updating the status could take up to 30 seconds.) Note: It is advisable to condition NiCad batteries when their run time is reduced to 50% of their previous run time.
  • Page 165: Entering Date And Time

    April 2009 Entering Date and Time In Operate mode, the Status Bar displays the current date and time above the measurements in the upper right corner of the screen. The Date/Time option allows you to set the date or time, which are required for correct data logging operation.

  • Page 166
    April 2009 Entering Date and Time (cont.) To set the date: 1. From the date text box, press the [ ] key once to scroll to the time text box. This box displays the current meter time. 2. Press [ENTER] to enter the text box.
  • Page 167: Changing Date And Time Appearance (Locale)

    April 2009 Changing Date and Time Appearance (Locale) In addition to setting the correct date and time, you can also change its presentation to suit local preferences. You can select a time display of AM/PM or 24-hour time To alter the time and date display: 1.

  • Page 168
    April 2009 Changing Date and Time Appearance (cont.) 2. The first series of prompts asks you to select separator symbols for the date, time and decimal. You can choose from a dash, comma and slash (/) for the date; from a period and colon for the time: and from a period or comma for the decimal place.
  • Page 169
    April 2009 Changing Date and Time Appearance (cont.) A line at the bottom, the Date/Time, Decimal Formats, displays how the format and separator selections will appear on the screen. Pressing the [ ] arrow key returns the PT878GC to the Locale tab.
  • Page 170: Adjusting The Contrast

    April 2009 Adjusting the Contrast For more comfortable viewing in a particular environment, the PT878GC enables you to adjust the screen contrast. To adjust the screen contrast: 1. From the Meter menu, scroll to the Contrast entry and press [ENTER] .

  • Page 171
    April 2009 Adjusting the Contrast (cont.) • [F3] To confirm the entries and return to Operate mode, press (OK). • To leave the window without confirming the entries and return [F2] [ESC] to Operate mode, press (Cancel) or the key. Programming Meter Settings 7-13…
  • Page 172: Setting Backlight Timeout

    April 2009 Setting Backlight Timeout By using the Backlight Timeout option, you can set a specified time that the PT878GC backlight will remain on before turning itself off. Automatic turnoff enables the PT878GC to conserve battery power. To set the backlight timeout: 1.

  • Page 173: Changing Communications Parameters

    2. The first prompt asks for the node identification number, which can be any number from 1 to 240. IMPORTANT: Do NOT change the node ID unless instructed by GE Sensing. a. Press [ENTER] to open the text box. b. Use the numeric keys to enter the node ID.

  • Page 174
    April 2009 Changing Communications Settings (cont.) 3. The next prompt asks you to select the communications interface from two options: IrDA (Infrared Data Association) and the IR232 virtual interface. You must use IrDA when using an IR printer or transfer application such as QuickBeam. Use the [ ] and [ ] keys to scroll to the desired selection, [ENTER] and press…
  • Page 175
    April 2009 Changing Communications Settings (cont.) 6. The next prompt asks you to select either one or two stop bits.The default number is one. Use the [ ] and [ ] keys to [ENTER] scroll to the desired number, and press 7.
  • Page 176: Resetting Forward And Reverse Totals

    April 2009 Resetting Forward and Reverse Totals On occasion, it might be necessary to clear and reset the forward and reverse totals computed by the Forward and Reverse Totalizers. To reset the totals: 1. From the Meter menu, scroll to the Totals entry and press [ENTER] .

  • Page 177: Setting Up User Tables

    April 2009 Setting Up User Tables When you program user functions (see page 4-46), you can also support them with up to six user tables of non-linear or empirical data. To program one or more user tables: 1. From the Meter menu, scroll to the User Tables entry and [ENTER] press .

  • Page 178
    April 2009 Setting up User Tables (cont.) 3. The next step is to create a Table ID. a. Press [ENTER] to open the window. The screen now appears similar to Figure 7-9 below. UserTable1 “ ‘ < > Delete Cancel Figure 7-9: The Text Creation Window b.
  • Page 179
    April 2009 Setting up User Tables (cont.) 4. The final prompt asks you to enter or change data in the user table. a. Press [ENTER] to open the window. The screen appears similar to Figure 7-10 below. Edit User Table 1 Programming UserTable1 Cancell…
  • Page 180
    April 2009 Setting up User Tables (cont.) e. Repeat steps b, c and d on page 7-21 until you have completed entering data for the table. f. When you have finished, press [F3] (OK) to confirm the [F2] entries or (Cancel) to leave the window without confirming the table.
  • Page 181: Taking A Bitmap Capture Of A Current Screen

    April 2009 Taking a Bitmap Capture of a Current Screen The Snapshot option enables you to take a “screen capture” of the current screen in bitmap format (.bmp) for display or storage in a Windows-based PC. To take a “snapshot” of the screen: 1.

  • Page 182: Chapter 8: Logging Data

    April 2009 Chapter 8 Logging Data A powerful and flexible feature of the PT878GC is data logging. The meter enables you to choose up to 12 parameters to log. You can also select the start time and date, end time and date, and time interval.

  • Page 183: Entering The Logging Menu

    April 2009 Entering the Logging Menu [MENU] To enter the Logging Menu, press the key at the lower right of the PT878GC keypad. The Main Menu replaces the Status Bar at the top of the screen. Press the [ ] arrow key three times to scroll from the Site Menu to the Logging Menu.

  • Page 184: The Log Manager

    April 2009 The Log Manager The Log Manager offers users a way to check the status and memory size of all the logs currently pending, running or finished. To select Log Manager, scroll to the Manager entry on [ENTER] the Logging Menu and press .

  • Page 185: The File Menu

    April 2009 The File Menu The File Menu allows you, not only to create new logs, but also to copy, rename or delete logs, as well as to print them or transfer them to a PC. To open the File menu from the Log Manager, press [MENU] [ENTER] key and then…

  • Page 186: Setting Up A New Log

    April 2009 Setting up a New Log The New Log option enables you to create and set up parameters for a new log. You can access this option in two ways: • by scrolling to the New Log option in the Logging Menu (as [ENTER] shown in Figure 8-1 on page 8-2) and pressing , or…

  • Page 187
    April 2009 Setting up a New Log (cont.) Measurements General Log Name 10SEC.LOG Format Circular Linear Type Error Standard Start Date/Time 01:38:08 2004/11/01 09:38:08 End Date/Time 2004/11/01 Logging Interval secs Activate Cancel Figure 8-5: The General Log Format Window To step through each parameter, press the [ ] key. 1.
  • Page 188
    April 2009 Setting up a New Log (cont.) 3. The next prompt asks for the starting date and time. a. Press [ENTER] to enter the text box. The meter highlights the first number. Use the [ ] and [ ] keys to scroll to any number you wish to change, or the [ ] key to scroll to the time box.
  • Page 189
    April 2009 Setting up a New Log (cont.) General Measurements NO UNIT NO UNIT NO UNIT NO UNIT NO UNIT NO UNIT NO UNIT NO UNIT NO UNIT NO UNIT NO UNIT NO UNIT Activate Cancel Figure 8-6: The Log Measurements Window To step through each entry, press the [ ] key.
  • Page 190
    April 2009 Setting up a New Log (cont.) Select Measurement Pressure Diagnostics Velocity Temperature Volume Fwd Totalizer Rev Totalizer StdFwdTotalizer StdRevTotalizer Cancel No Unit Figure 8-7: The Select Measurement Window 2. Scroll to the desired output type. 3. Press [SEL] to confirm your selection.
  • Page 191: Copying (Cloning) A Selected Log

    April 2009 Copying (Cloning) a Selected Log In the Clone Selected Log option, you can copy the parameters of a particular log, modify the parameters, and start the copy. To copy a log: 1. First, select the log you wish to copy in the Log Manager (shown in Figure 8-3 on page 8-4).

  • Page 192: Renaming A Log

    April 2009 Renaming a Log To rename a log: 1. First, select the log you wish to rename in the Log Manager (shown in Figure 8-3 on page 8-4). 2. Then open the File menu, scroll to the Rename option, and [ENTER] press 3.

  • Page 193: Deleting All Logs

    April 2009 Deleting a Log (cont.) Log Manager Delete File Programming Info:DEFAULT DEFAULT Delete Log ‘555.LOG?’ GLOBAL 1 View Figure 8-8: The Delete Confirmation Window Deleting All Logs To clear the Log Manager and memory of all logs, open the File [ENTER] menu, scroll to the Delete All Logs option, and press [F2]…

  • Page 194: Printing A Log

    April 2009 Printing a Log To print a file from the Log Manager: 1. Highlight the desired log in the window on the left of the Log Manager. 2. Press [MENU] [ENTER] , scroll to the Print option, and press The PT878GC shows a message indicating that its infrared sensor is looking for a printer.

  • Page 195: The Log Menu

    April 2009 The Log Menu The Log Menu allows you to pause, restart or end any or all logs that are currently pending or running. (However, you cannot restart any finished logs, even if they were finished before the programmed end time.) To open the Log Menu in the Log [MENU] Manager, press .

  • Page 196: Stopping (Pausing) A Log

    April 2009 Stopping (Pausing) a Log To stop a log that is currently pending or running, 1. First, select the log you wish to pause in the Log Manager (shown in Figure 8-2 on page 8-4). 2. Then open the Log menu, scroll to the Pause option, and press [ENTER] The PT878GC returns to the Log Manager, which displays the highlighted log with “Paused”…

  • Page 197: Pausing All Logs

    April 2009 Pausing All Logs To pause all logs that are currently pending or running, open the [ENTER] Log menu, scroll to the Pause All Logs option, and press The PT878GC returns to the Log Manager, which displays all logs programmed to run now or in the future with a status of “Paused.”…

  • Page 198: The View Menu

    April 2009 The View Menu Through the View menu, you can view the data of individual logs in graphical or spreadsheet formats. To open the View menu from [MENU] the Log Manager, press . Scroll to the View menu, and [ENTER] press .

  • Page 199: Displaying Log Details

    April 2009 Displaying Log Details To view details of a given log, be sure the log is highlighted in the left window of the Log Manager. Then scroll to the View menu [ENTER] [ENTER] and press . Scroll to the Details option and press The screen now appears similar to Figure 8-11 below.

  • Page 200: Displaying Log Data In Graphical Form

    April 2009 Displaying Log Data in Graphical Form To view a log in graphical form, be sure the log is highlighted in the left window of the Log Manager. 1. Scroll to the View menu and press [ENTER] 2. Scroll to the Graph option and press [ENTER] 3.

  • Page 201
    April 2009 Displaying Log Data in Graphical Form (cont.) AAA.LOG (Dew Point, °C) Y-Axis -52.0 Limits Range Minimum 0.1234 Maximum 0.5678 2004/03/01 2004/03/01 13:53:41 PM 13:23:41 PM Cancel Figure 8-13: The Y-Axis Window The Y-Axis window allows you to specify whether the Y axis on the graph extends to the maximum value (Max), over the entire range (Range) or between certain specified values (Set).
  • Page 202
    April 2009 Displaying Log Data in Graphical Form (cont.) [F2] If you wish to alter the time scale, press (Time). The Enter Time window appears similar to Figure 8-14 below. AAA.LOG (P#Up) -52.0 Enter Time Start Date Start Time 2004/03/01 13:23:41 PM End Date End Time…
  • Page 203: Displaying Log Data In Spreadsheet Form

    April 2009 Displaying Log Data in Spreadsheet Form To view a log in spreadsheet form, be sure the log is highlighted in the left window of the Log Manager. Then scroll to the View [ENTER] menu and press . Scroll to the Spreadsheet option and [ENTER] press .

  • Page 204: The Sort Menu

    April 2009 The Sort Menu The Sort Menu within the Log Manager allows you to arrange your log list either alphabetically (By Name) or chronologically (By Date). Log Manager Sort File View By Name Info: AAA.LOG State: Finished By Date S:03/01/04 13:13:41 E:03/01/04 13:23:41 Interval: 10 Seconds…

  • Page 205: Chapter 9: Servicing The Pt878Gc

    April 2009 Chapter 9 Servicing the PT878GC For user convenience, the PT878GC offers a Service Menu. This menu enables users to perform a variety of functions that they might occasionally require: • print out reports • calibrate the PT878GC • run diagnostics •…

  • Page 206: Entering The Service Menu

    April 2009 Entering the Service Menu [MENU] To enter the Service Menu, press the key at the lower right of the PT878GC keypad. The Main Menu replaces the Status Bar at the top of the screen. Press the [ ] arrow key four times to scroll from the Site Menu to the Service Menu.

  • Page 207: Printing Reports

    April 2009 Printing Reports When used with an IR-compatible printer, the PT878GC can print out a variety of data (current site, logs, drive contents, and user functions, tables, and settings) in the Reports option. To enter the Reports option, scroll to the Reports entry on the Service Menu [ENTER] and press .

  • Page 208
    April 2009 Printing Reports (cont.) Figure 9-3: Printout of a Typical Drive Report Servicing the PT878GC…
  • Page 209: Setting Up The Thickness Gauge

    April 2009 Setting up the Thickness Gauge For greatest accuracy in flow applications, the PT878GC can measure pipe wall thickness using an optional thickness gauge transducer, instead of relying on the nominal pipe wall thickness. In Thickness Gauge mode, the PT878GC does not measure flow, but it can determine the thickness of most standard metal and plastic pipe materials over a range from 0.05 to 3 in.

  • Page 210
    April 2009 Setting up the Thickness Gauge (cont.) Calibrating the thickness gauge involves up to two additional steps: 4. Calibrating the thickness gauge itself (in the Zero window on page 9-12), and 5. Calibrating the velocity of sound in the pipe material (in the Velocity window on page 9-14), if possible and if the pipe material is not the material entered in the Zero window.
  • Page 211: Measuring Pipe Wall Thickness

    April 2009 Measuring Pipe Wall Thickness Entering the Material and Sound Speed IMPORTANT: GE Sensing recommends calibrating the thickness gauge periodically (as discussed on page 9-12) before measuring thickness. To enter the Thickness Gauge Display option, scroll to the [ENTER] T-Gauge Display entry on the Service Menu and press The screen appears similar to Figure 9-5 below.

  • Page 212
    April 2009 Entering the Material and Sound Speed (cont.) If you have selected a preprogrammed material, you have completed entering data in this window. Pressing the [ ] key returns the program to the Material tab. But if you selected “Other,”…
  • Page 213: Measuring Thickness In Numeric Format

    April 2009 Measuring Thickness in Numeric Format To measure the actual thickness of a pipe, apply couplant to the calibrated thickness gauge transducer and hold the transducer steady against the pipe. Then press the [ ] or [ ] arrow key [ENTER] until you reach the Display tab and press .

  • Page 214: Displaying The Receive Signal In Graphical Format

    April 2009 Displaying the Receive Signal in Graphical Format Used chiefly for diagnostic purposes, the Graph option helps to determine why the thickness gauge is not working if you suspect a problem. The graph shows an image of the acoustic signal. If the display does not show a signal image similar to Figure 9-7 below, you may have a problem with the transducer, couplant, or ®…

  • Page 215
    April 2009 Displaying the Receive Signal in Graphical Format (cont.) Thickness Gauge Measure Display Material Graph Velocity Zero Signal Display Figure 9-8: The Graph Window in Zoom Format • [F1] Press (In) to zoom in to magnify screen details. • [F2] Press (Out) to zoom out fully.
  • Page 216: Calibrating The Thickness Gauge Transducer

    See Appendix C for more information. 3. The next prompt asks you to enter the length of Block 1. (If you are using a GE Sensing-supplied test block, the length is [ENTER] printed on the block.) Press to open the text box.

  • Page 217
    IMPORTANT: Be sure to hold the transducer steady until the GE Sensing wait cursor disappears. 7. If you have chosen dual calibration, repeat Steps 3 through 6 above for Block 2. 8. The PT878GC asks for confirmation of the calculated and current values.
  • Page 218: Calculating Velocity (Pipe Material Sound Speed)

    April 2009 Calculating Velocity (Pipe Material Sound Speed) The nominal sound speed for your pipe material will give thickness measurements of reasonable accuracy. However, for greater accuracy, calibrate pipe material sound velocity. IMPORTANT: This step is necessary only if the actual pipe material sound speed differs from the sound speed of the material used in the calibration block.

  • Page 219
    April 2009 Calculating Velocity (Pipe Material Sound Speed) (cont.) 2. To enter the block length (here, the thickness of the sample as measured with a caliper or calibrated sample), press the [ [ENTER] arrow key and then press to open the Block Length text box.
  • Page 220
    April 2009 Calculating Velocity (Pipe Material Sound Speed) (cont.) 7. The “Calculated” box shows the thickness value measured. The PT878GC asks for confirmation of the calculated and [ENTER] current values. Press to commit (accept) the calculated value (in which case it becomes the current value), and to recalculate the value (which becomes the new “Calculated”…
  • Page 221: Programming The Thickness Gauge

    • Detection Threshold IMPORTANT: The thickness gauge programming settings are entered at the factory. You should not change them unless instructed by GE Sensing. The Programming window appears similar to Figure 9-12 below. Thickness Gauge Setup Programming Low Signal Thresh Transducer Delay µs…

  • Page 222
    April 2009 Programming the Thickness Gauge (cont.) 1. Press the [ ] arrow key to enter the window. 2. The first prompt asks for the low signal threshold. If the signal strength falls below this value, a “Low Signal” message appears on the screen.
  • Page 223
    April 2009 Programming the Thickness Gauge (cont.) 6. The final value, the detection threshold, represents the percent of peak the PT878GC uses to make measurements. It will consider anything above the entered percentage as part of the signal. To change this value: a.
  • Page 224: Displaying Diagnostic Parameters

    April 2009 Displaying Diagnostic Parameters The Diagnostics option enables you to view current diagnostic parameters without having to open a display window in Operate Mode. To enter the option, scroll to the Diagnostics entry on the [ENTER] Service Menu and press .

  • Page 225: Calibrating The Analog Output And Inputs

    April 2009 Calibrating the Analog Output and Inputs The Calibration option allows you to calibrate the analog output and inputs. To enter the option, scroll to the Calibrate entry on the [ENTER] Service Menu and press . The screen appears similar to Figure 9-14 below.

  • Page 226
    April 2009 Calibrating the Analog Output (cont.) 1. Connect a digital multimeter or ammeter to the analog output. 2. The next prompt asks for the setpoint (4 or 20 mA). Use the [ ] and [ ] arrow keys to move to the appropriate radio [ENTER] button and press 3.
  • Page 227: Calibrating Inputs

    April 2009 Calibrating Inputs To open the Inputs window, press the [ ] arrow key and press [ENTER] . The screen appears similar to Figure 9-15 below. Note: Calibrating the analog inputs requires use of a current source. Calibrate/Test I/O Analog Input Analog Output Input#…

  • Page 228
    April 2009 Calibrating Inputs (cont.) 4. Connect the current source to the input and set it to deliver the value entered in the previous step. 5. Press the [ ] arrow key to move to the Read box, and then [ENTER] press to read the low input.
  • Page 229: Setting Up Signal Parameters

    April 2009 Setting up Signal Parameters The Signal Setup option enables you to set parameters that affect the transducer signal: • Delta-T Offset • Transmit Sample Size • Zero Cutoff • Velocity Averaging • Errors Allowed • Peak Detection Method •…

  • Page 230
    2. The first prompt asks for the Delta-T offset. Delta-T is the difference between the upstream and downstream transit times of the transducers. The Delta-T offset should normally be set to zero. Note: Consult GE Sensing before performing this step. a. Press [ENTER] to open the text box.
  • Page 231
    April 2009 Setting up Signal Parameters (cont.) 5. The next prompt asks for velocity averaging, the number of velocity measurements to average together to smooth out noise in the system. a. Press [ENTER] to open the drop-down list. b. Use the [ ] or [ ] arrow keys to scroll to the desired number.
  • Page 232: Setting Up The Measurement Mode

    Note: Do not change the measurement mode or values unless recommended by GE Sensing. From the Signal Parameter tab, press the [ ] arrow key to move to the Meas Mode tab. The window appears similar to Figure 9-17 below.

  • Page 233
    April 2009 Setting up Measurement Mode (cont.) The first prompt asks for the measurement mode detection method. Use the [ ] and [ ] arrow keys to move to the [ENTER] appropriate radio button and press You must now enter the setup data covered on page 4-23, if you have not already done so.
  • Page 234: Setting Up Pulse/Code Parameters

    April 2009 Setting Up Pulse/Code Parameters When you select the Skan detection method, you must also choose what type of signal to send to the transducers: either a coded signal at 1, 2, 4, 11 or Twin bits, or a specific number of pulses to be sent.

  • Page 235
    April 2009 Setting Up Pulse/Code Parameters (cont.) 2. The next prompt asks if you want to send a specific number of [ENTER] pulses. Press to select or deselect the pulse option. Note: If you do not select this option, pressing the [ ] key returns the meter to the Transmit Code prompt.
  • Page 236: Setting Error Limits

    April 2009 Setting Error Limits The Error Limits option enables you to set limits for an incoming signal. When the signal falls outside the programmed limits, an error indication appears. To enter this option, scroll to the Error [ENTER] Limits entry on the Service Menu and press .

  • Page 237
    April 2009 Setting Error Limits (cont.) d. Press the [ ] arrow key to move to the maximum signal box, and repeat steps a, b and c. e. Press the [ ] arrow key to return to the minimum signal box, and the [ ] key to move to the next limit.
  • Page 238
    April 2009 Setting Error Limits (cont.) Note: For the velocity and acceleration boxes, the F1 softkey toggles between English and metric measurements. Thus, the key will always display the opposite measurement [F1] from that which is currently active. Press to display the measurement in the alternate format.
  • Page 239: The Test Option

    April 2009 The Test Option Within the Service Menu, the Test option includes seven tests to ensure that the PT878GC is performing properly: Test Screen, Test Keys, Watchdog Test, Impulse Response, Wave Snapshot, Simulate and Battery Test.To enter this option, scroll to the Test [ENTER] entry on the Service Menu and press .

  • Page 240: Testing The Screen

    Pressing the key through this sequence should return the PT878GC to Operate Mode. If the test does not proceed according to this sequence, please consult GE Sensing. 9-36 Servicing the PT878GC…

  • Page 241: Testing The Keys

    Pressing the key returns the meter to the Operate Mode. If any key does not appear on the screen or does not darken when pressed, contact GE Sensing. Note: The power key does not appear in this test. Servicing the PT878GC…

  • Page 242: Testing The Watchdog Timer Circuit

    (No) to cancel the test and return to the Menu screen, or [F3] press (Yes) to start the test. The PT878GC should go blank for a few seconds, and then restart. If it does not follow this sequence, consult GE Sensing. 9-38 Servicing the PT878GC…

  • Page 243: Setting Impulse Response

    April 2009 Setting Impulse Response The Impulse Response option enables you to force the meter to transmit in one direction only, without changing the AGC setting. You can then diagnose problems with transducer, pipe, or fluid configurations. To enter the option, scroll down to the Impulse [ENTER] Response entry in the Service Menu and press .

  • Page 244: Taking Wave Samples For Diagnosis

    April 2009 Setting Impulse Response (cont.) 2. The second prompt asks for the percent of gain. a. Press [ENTER] to open the text box. b. Use the numeric keys to enter the desired percentage. c. Press [ENTER] to confirm your entry. You have completed entering data for the Impulse Response option.

  • Page 245: Testing The Battery

    April 2009 Testing the Battery The Battery Test option provides additional battery information for service troubleshooting. To enter the option, scroll down to [ENTER] the Battery Test entry in the Service Menu and press . The screen appears similar to Figure 9-25 below. Service Battery Form Status: Fast Charge Run Time…

  • Page 246: Resetting To Factory Default Parameters

    April 2009 Resetting to Factory Default Parameters For various reasons, you might wish to return the PT878GC to its original settings. The Factory Defaults option enables you to return the meter to its preprogrammed default settings. To enter the option, scroll down to the Factory Defaults entry in the [ENTER] Service Menu and press .

  • Page 247: Updating Pt878Gc Software

    • IrCOMM, an infrared standard supported by certain Microsoft operating systems. GE Sensing recommends updating software via the IrOBEX standard; however, the IrCOMM standard is available for users who have problems with IrOBEX. This section covers procedures for both standards.

  • Page 248: Updating Software Via Irobex

    4. The meter asks for confirmation. Repeat the options shown in Step 1 above. After the PT878GC reboots, the screen appears similar to Figure 9-28 below. GE Sensing PCI Loader v3.0 2/26/04 [HW Rev3+] Backup Battery: OK Firmware CRC: Valid…

  • Page 249
    April 2009 Updating Software Via IrOBEX (cont.) 5. The meter now asks if you wish to reload the flash memory via the infrared communications port. • [F1] Press (Yes) to begin reloading via OBEX, or • [F2] Press (No) to stop the reloading procedure. Be sure the infrared sensor on the meter has a clear path to the infrared adapter on the PC.
  • Page 250: Updating Software Via Ircomm

    April 2009 Updating Software Via IrCOMM Note: While Windows 95/98/98SE and NT (with QuickBeam) support IrCOMM, Windows 2000 and XP do not. Refer to Appendix B. Before you install new coding, you must be sure that the PC has the correct protocols to transmit the software to the meter. Setting up the PC Note: The use of Hyperterminal is shown here as an example.

  • Page 251
    April 2009 Setting up the PC (cont.) Figure 9-29: Port Settings for COMM 4 Installing New Software To begin the loading procedure, scroll down to the Flash Update [ENTER] entry in the Service Menu and press . The screen appears similar to Figure 9-30 below.
  • Page 252
    2. The meter asks for confirmation. Repeat the options shown in Step 1 above. The screen on both the PC and the PT878GC now appears similar to Figure 9-31 below. GE Sensing PCI Loader v3.0 2/26/04 [HW Rev3+] Backup Battery: OK…
  • Page 253
    April 2009 Installing New Software (cont.) 1. In the Hyperterminal window, pull down the Transfer menu and click Send File. A window opens similar to Figure 9-32 below. Figure 9-32: The Send File Window in the Transfer Menu 2. Click on the replacement software (designated by a .cod extension) from the folder where it has been stored.
  • Page 254: Chapter 10: Diagnostics And Troubleshooting

    April 2009 Chapter 10 Diagnostics and Troubleshooting The TransPort PT878GC is a reliable instrument that is easy to maintain. It will provide accurate flow measurement readings as long as it is operated as described in this manual. If problems do…

  • Page 255: Error Code Messages

    Use Table 10-1 on page 10-3 to isolate and remedy the problem. If you are unable to remedy the problem, contact GE Sensing. Note: We suggest that, in order to isolate the indicated problem more easily, you obtain a test flowcell.

  • Page 256
    April 2009 Table 10-1: Error Messages Error Possible Message Problem Cause Action No error. Displays briefly None required. Mea- after the dis- surement is valid. play of another error message Low Signal — Broken cable. Check transducer Poor ultra- Flowcell prob- cable.
  • Page 257
    April 2009 Table 10-1: Error Messages (cont.) Error Possible Message Problem Cause Action Velocity Programming Make sure the maxi- Range — error, poor flow mum velocity range Velocity conditions, or is within the range exceeds pro- bad transducer listed in Table 3-1 on grammed spacing.
  • Page 258
    April 2009 Table 10-1: Error Messages (cont.) Error Possible Message Problem Cause Action Cycle skip, Poor flow con- Check transducer Acceleration ditions or bad spacing as pro- transducer spac- grammed and as set ing. on pipe.See Flowcell Problems on page 10- 9 and Transducer Problems on page 10- Analog Out…
  • Page 259: Displaying Diagnostic Parameters

    April 2009 Displaying Diagnostic Parameters As part of its measurement menu, the PT878GC offers a list of diagnostic parameters to aid in troubleshooting in the event of flowcell, transducer, or electrical problems. You can select any diagnostic parameter for display as a measurement as discussed in Chapter 6 on pages 6-7 and 6-8.

  • Page 260
    April 2009 Table 10-2: Diagnostic Parameters (cont.) Diagnostic Displays Good Parameter Soundspeed Displays the measured sound Check the sound speed (m/s or f/s) speed of the fluid. of fluid in Sound Speeds and Pipe Size Data. P#up Displays signal peaks for the 100- <100, >900 upstream transducer.
  • Page 261
    April 2009 Table 10-2: Diagnostic Parameters (cont.) Diagnostic Displays Good Parameter Qdown Displays the signal quality for ±300 or between the downstream transducer. higher ±100 Norm Factor Displays the normalization 0.85 — <0.85 factor. Threshold Up Displays the Percent of Peak for the upstream transducer.
  • Page 262: Flowcell Problems

    Most gas-related problems can be solved by proper design of the flowmeter installation and the proper choice of transducer frequency. Contact GE Sensing if you cannot solve a gas-related problem. If the physical installation of the system meets the recommended specifications, it is possible that the gas itself may be preventing accurate flow rate measurements.

  • Page 263: Pipe Problems

    4-13 and 4-14. Pipe Problems Improper pipe conditions and/or flowcell installation can cause problems with measurement of the gas flow. Contact GE Sensing if you cannot solve pipe-related problems. By far, the most common pipe problems are the following 1. THE COLLECTION OF MATERIAL AT THE TRANSDUCER LOCATION(S).

  • Page 264: Transducer Problems

    Note: Transducers must be replaced in pairs. Refer to Chapter 4, Programming Site Data, to program the replacement transducer data into the meter. Contact GE Sensing if you cannot solve a transducer-related problem. Diagnostics and Troubleshooting 10-11…

  • Page 265: Specifications

    April 2009 Chapter 11 Specifications This section contains specifications for the following: • Operations and Performance • Electronics • Clamp-on Transducers • Thickness Gauge Option • Additional Options Specifications 11-1…

  • Page 266: Operation And Performance

    April 2009 Operation and Performance Fluid Types Acoustically conductive gases with minimum density requirements. See Table 3-1 on page 3-5, Table 3-2 on page 3-7 and Table 3-3 on page 3-9. Pipe Sizes 0.75 to 12 in. (20 to 300 mm) with PT878GC-01 electronics 4 to 24 in.

  • Page 267: Pressure Requirements, Air, Nitrogen, Oxygen, Argon

    See Table 3-2 on page 3-7. Pressure Requirements, Steam See Table 3-3 on page 3-9. Pressure Requirements, Other Gases Dependent upon gas composition, and pipe size and material. Consult GE Sensing. Measurement Parameters Standard and actual volumetric flow, flow velocity and mass flow. Specifications…

  • Page 268: Electronics

    April 2009 Electronics Flow Measurement Patented Correlation Transit-Time mode Enclosures Submersible IP67 Dimensions Weight: 3 lb (1.36 kg) Size: (h x w x d) 9.4 x 5.5 x 1.5 in. (238 x 138 x 38 mm) Display 240- x 200-pixel backlit LCD graphic display Keypad 25-key rubberized tactile membrane keypad Internal Battery…

  • Page 269: Standard Inputs/Outputs

    April 2009 Standard Inputs/Outputs • One 0/4- to 20-mA current output • One user-selectable pulse (solid state, 5-V maximum) or frequency (5-V square wave; 100 to 10,000 Hz) • Two 4- to 20-mA analog inputs with switchable power supply for loop-powered temperature or pressure transmitters Digital Interfaces Infrared communication port for printer or PC interface Site Parameter Programming…

  • Page 270: Clamp-On Ultrasonic Flow Transducers

    April 2009 Clamp-On Ultrasonic Flow Transducers Temperature Range(s) Standard: –40° to 130°C (–40° to 266°F) Optional (overall range): –40° to 230°C (–40° to 446°F) Transducer Materials Stainless steel and plastic Fixture Materials Anodized aluminum blocks with stainless steel rod, chain or strap Mountings Portable gas-transducer clamping fixtures in polycarbonate carrying cases:…

  • Page 271: Transducer Cables

    April 2009 Transducer Cables • Standard: One pair of LEMO® coaxial transducer connectors with 25-ft (8-m) cables • Optional: 500-ft (152-m) extension cables available for most transducers Specifications 11-7…

  • Page 272: Thickness Gauge Option

    April 2009 Thickness Gauge Option Transducer GE Sensing dual element transducer Pipe Thickness Range 0.05 to 3 in. (1.3 to 76.2 mm) Pipe Materials Most standard metal and plastic pipe materials Accuracy ±1% typical or ±0.002 in. (±0.05 mm) Thermal Exposure Continuous operation to 37°C (100°F);…

  • Page 273
    April 2009 Appendix A Menu Maps Menu Service Program Meter Logging Site See Figure See Figure A-10 Figure Figure A-2 Figures to A-7 Figure A-1: The Main Menu Menu Maps…
  • Page 274: Appendix A: Menu Maps

    April 2009 Menu Site Manager Save Fkeys 1 View 2 Views 3 Views 4 Views Clear Clear Clear Menu Item File Site Sort Transfer Message Print Transfer Name Date as Text Drive Manager File Sort Save Save Refresh Rename Open Delete Name File Name File…

  • Page 275
    April 2009 Menu Program (See Figure A-5) (Clamp-on) Transducer Lining Pipe Fluid Path (See Figure A-4) Fluid (Wetted) (Clamp-on) Wetted Clamp-on Natural Steam Other Path Length Traverses Axial Length Spacing Sound Speed (Preprogrammed Special Units) Frequency Wedge Temperature Time Delay (Wetted) (Clamp-on) Material…
  • Page 276
    April 2009 Menu Program Pipe Material Steel Aluminum Brass Copper Glass Gold Inconel Iron Monel Nickel Plastic Tin Titanium Tungsten Zinc Other Sound Speed (For cast iron, steel, PVC, CPVC) Note: Press [F2] (Cancel) to cancel the entries and return to Menu. ANSI? Press [F3] (OK) to confirm the entries and return to Menu.
  • Page 277
    April 2009 Menu Program (See Figures A-3 and A-4) (See Figure A-6) Standard Volume Mass Flow Inputs Settings Pressure Disabled Enabled Fixed Active Disabled Enabled Base Pressure Pressure Input # Calculation Method Atmospheric Pressure Static Saturated Steam Base Temperature Density Steam Temperature Value…
  • Page 278
    April 2009 Menu Program (See Figure A-5) (See Figure A-7) Analog Output Analog Input Scan/Measure Function Input A Input B 0-20 mA 4-20 mA Correlate Integrate Count (If Standard Min Threshold (If Standard Cycle Number A Divisor % of Peak Volume or Mass Volume or Mass Data Source…
  • Page 279
    April 2009 Menu Program See Figure User Digital Functions Correction Factors Output Function Pulse Frequency Gate Input Test Points Function Calibration Factor Reynolds Correction Totalizer User1 User2 User4 User3 Reynolds Correction Calibration Factor Gate Active Data Source (Pulse Totalizer) (Frequency) Contact Contact Closed…
  • Page 280
    April 2009 Note: Press [F2] (Cancel) to cancel Menu the entries and return to Menu. Press [F3] (OK) to confirm the entries Meter and return to Menu. Battery Snapshot Date/Time Contrast Backlight Units Locale User Tables Totals Date Backlight Reset Time NiCd NiMH…
  • Page 281
    April 2009 Menu Note: Press [F2] (Cancel) to cancel Logging the entries and return to Menu. Press [F3] (OK) to confirm the entries and return to Menu. Log Manager New Log Name Format Sort File View Linear Circular Name Date Type Details Graph Spreadsheet Standard…
  • Page 282
    April 2009 Menu Note: Press [F2] (Cancel) to cancel entries and return to Menu. Press [F3] (OK) to confirm entries and return to Menu. Service T-Gauge T-Gauge Diagnostics Calibrate Factory Flash Reports Error Limits Test Signal Setup Setup Display Defaults Update Min Signal Output…
  • Page 283
    April 2009 Menu Meter Communications Node ID Comm Interface IrDA IR 232 Baud Rate 300 600 1200 2400 4800 9600 19.2K 38.4K 57.6K 115.2K Parity None Mark Space Even Stop Bits Data Bits Note: Press [F2] (Cancel) to cancel the entries and return to Menu. Press [F3] (OK) to confirm the entries and return to Menu.
  • Page 284: Appendix B: Establishing Ir Communications With The Pt878Gc

    IR port (available on most laptops) or an IR dongle (IR to RS232 adapter for PCs without a built-in IR port). If your PC does not have an IR dongle, GE Sensing recommends the ActiSys ACT-IR220L+ infrared to RS- 232 adapter, which has been tested for compatibility with the PT878GC.

  • Page 285: Windows 2000/Xp

    April 2009 Table B-1: IrDA Compatibility of Windows Operating Systems Operating System IrCOMM OBEX Windows XP Windows 2000 Windows NT 4.0 QuickBeam* QuickBeam* Windows Me* Windows 98 SE* Windows 98* Windows 95* Windows 3.1 DOS 6.22 *May require QuickBeam installation under some circumstances when Windows built-in support fails.

  • Page 286: Windows Nt4.0

    April 2009 Windows NT4.0 For IR use, Windows NT4.0 requires installation of QuickBeam software from Extended Systems (www.extendedsystems.com). Desktop PCs and laptops without built-in IR ports also require adding an IR dongle. While laptops with a built-in IR port normally do not require a dongle, its addition might be necessary in some cases.

  • Page 287: Appendix C: Ultrasonic Thickness Gauge Theory Of Operation

    April 2009 Appendix C Ultrasonic Thickness Gauge Theory of Operation All ultrasonic thickness gauging involves timing the round trip of a sound pulse in a test material. Because solid metal has an acoustic impedance that differs from that of gases, liquids, or corrosion products such as scale or rust, the sound pulse will reflect from the far surface of the remaining metal.

  • Page 288: Factors Affecting Performance And Accuracy

    April 2009 Factors Affecting Performance and Accuracy A. Surface Condition — Loose or flaking scale, rust, corrosion or dirt on the outside surface of a test piece will interfere with the coupling of sound energy from the transducer into the test material.

  • Page 289
    April 2009 Factors Affecting Performance and Accuracy (cont.) B. Transducer Positioning/Alignment — For proper sound coupling the transducer must be pressed firmly against the test surface. On small diameter cylindrical surfaces such as pipes, hold the transducer so that the sound barrier material visible on the probe face is aligned perpendicular to the center axis of the pipe.
  • Page 290
    April 2009 Factors Affecting Performance and Accuracy (cont.) Remember that an ultrasonic test measures thickness at only one point within the beam of the transducer, and that in corrosion situations wall thicknesses often vary considerably. Test procedures usually call for making a number of measurements within a defined area and establishing a minimum and/or average thickness.
  • Page 291
    April 2009 Factors Affecting Performance and Accuracy (cont.) D. Taper or Eccentricity — If the contact surface and the back surface are tapered or eccentric with respect to each other, the return echo again becomes distorted and the accuracy of measurement is diminished.
  • Page 292
    A variety of couplants are available for use with GE Sensing transducers. To obtain a material safety data sheet (MSDS) for any couplant GE Sensing supplies, go to the GE Sensing web site and enter the Technical Support portal. Material Safety Data Sheets for Couplants…
  • Page 293: Analog Outputs

    April 2009 Index About Option……… . . 6-21 Acceleration Limit.

  • Page 294
    April 2009 Index (cont.) Calibration Analog Input ……… 9-21 Analog Output .
  • Page 295
    April 2009 Index (cont.) Dampening Material ……..3-25 Data Acquisition .
  • Page 296: File Menu

    April 2009 Index (cont.) Electrical Connections ……… . 2-2 Electrical Connections .

  • Page 297
    April 2009 Index (cont.) Graph Display ……… . . 6-4 Help .
  • Page 298
    April 2009 Index (cont.) Limits Option……….6-5 Line Graph Display .
  • Page 299: Main Menu

    April 2009 Index (cont.) M>S Switch ……….4-25 Main Menu Menu Map.

  • Page 300
    April 2009 Index (cont.) On-Line Help ……… . . 2-14 Getting .
  • Page 301
    April 2009 Index (cont.) Parameters, Displayed ……..6-10 Path for Clamp-on .
  • Page 302
    April 2009 Index (cont.) Range ……….11-2 Rangeability .
  • Page 303
    April 2009 Index (cont.) Sampling Waves……..9-40 Schedule for Pipe Dimensions .
  • Page 304
    April 2009 Index (cont.) Sound Speed, Calculating ……9-14, 9-15 Span (Full Scale) Value for Analog Output .
  • Page 305
    April 2009 Index (cont.) Task Keys ……….2-12 Temperature Operating .
  • Page 306
    April 2009 Index (cont.) Entering Data for ……..4-3 Materials .
  • Page 307
    April 2009 Index (cont.) Watchdog Test ……… 9-38 Wave Sampling .
  • Page 308
    DECLARATION Sensing CONFORMITY Panametrics Limited Shannon Industrial Estate Shannon, County Clare Ireland declare under our sole responsibility that the ™ TransPort PT878 Portable Ultrasonic Flowmeter ™ TransPort PT878GC Clamp-On Portable Ultrasonic Flowmeter to which this declaration relates, are in conformity with the following standards: •…
  • Page 309
    DECLARATION Sensing CONFORMITE Nous, Panametrics Limited Shannon Industrial Estate Shannon, County Clare Ireland déclarons sous notre propre responsabilité que les ™ TransPort PT878 Portable Ultrasonic Flowmeter ™ TransPort PT878GC Clamp-On Portable Ultrasonic Flowmeter rélatif á cette déclaration, sont en conformité avec les documents suivants: •…
  • Page 310
    Sensing KONFORMITÄTS- ERKLÄRUNG Wir, Panametrics Limited Shannon Industrial Estate Shannon, County Clare Ireland erklären, in alleiniger Verantwortung, daß die Produkte ™ TransPort PT878 Portable Ultrasonic Flowmeter ™ TransPort PT878GC Clamp-On Portable Ultrasonic Flowmeter folgende Normen erfüllen: • EN 61326:1998, Class A, Annex C, Continuous Unmonitored Operation gemäß…
  • Page 311
    1100 Technology Park Drive Billerica, MA 01821-4111 Web: www.gesensing.com Ireland Sensing House Shannon Free Zone East Shannon, County Clare Ireland…

  • Page 1
    Sensing & Inspection Technologies TransPort™ PT878GC Portable Gas Flowmeter User’s Manual…
  • Page 2
    Sensing & Inspection Technologies TransPort™ PT878GC Portable Gas Flowmeter User’s Manual 910-229B3 April 2009…
  • Page 3
    Liability under this warranty is limited to restoring the instrument to normal operation or replacing the instrument, at the sole discretion of GE Infrastructure Sensing, Inc. Fuses and batteries are specifically excluded from any liability. This warranty is effective from the date of delivery to the original purchaser.
  • Page 4
    (RAN), and shipping instructions for the return of the instrument to a service center will be provided. 2. If GE Infrastructure Sensing, Inc. instructs you to send your instrument to a service center, it must be shipped prepaid to the authorized repair station indicated in the shipping instructions.
  • Page 5: Table Of Contents

    April 2009 Table of Contents Chapter 1: Features and Capabilities Overview …………1-1 System Description .

  • Page 6
    April 2009 Table of Contents (cont.) Installing Dampening Material ……. . 3-25 Installing DMP-1 Dampening Material .
  • Page 7
    April 2009 Table of Contents (cont.) Chapter 5: Creating and Managing Sites The Site Manager ……….5-2 Creating a New Site .
  • Page 8
    April 2009 Table of Contents (cont.) Changing Date and Time Appearance (Locale) ….7-9 Adjusting the Contrast ……..7-12 Setting Backlight Timeout .
  • Page 9
    April 2009 Table of Contents (cont.) Setting up the Thickness Gauge ……. . 9-5 Measuring Pipe Wall Thickness .
  • Page 10
    April 2009 Table of Contents (cont.) Chapter 11: Specifications Operation and Performance……..11-2 Fluid Types .
  • Page 11
    April 2009 Table of Contents (cont.) Clamp-On Ultrasonic Flow Transducers ……11-6 Temperature Range(s) ……..11-6 Transducer Materials .
  • Page 12: Chapter 1: Features And Capabilities

    April 2009 Chapter 1 Features and Capabilities ® The TransPort Model PT878GC is a transit-time clamp-on gas flowmeter which combines all the features of a full-size gas flowmeter with the advantages of a portable instrument. This section describes the TransPort features and general system, and explains the theory of operation.

  • Page 13
    April 2009 Overview (cont.) The TransPort has the ability to store site data in files which can be accessed at a later time. Within the Main Menu, a set of forms (windows) asks you all the necessary setup information for a particular site.
  • Page 14: System Description

    April 2009 System Description The TransPort is only one part of the flowmeter system. The other part of the system is the flowcell, which consists of the pipe and the transducers. The Flowcell The flowcell is that part of the system that uses ultrasonic pulses to interrogate the flow.

  • Page 15: Electronics Package

    April 2009 Electronics Package The TransPort consists of circuits that generate, receive, and measure the travel time of the ultrasonic pulses. It also contains a microcomputer that controls operation and calculates flow measurement parameters. Specific circuits function as follows: A. TRANSMIT SIGNAL GENERATOR — The transmit signal generator, under control of the microcomputer and timing circuit, synthesizes the signal that drives the transmitter.

  • Page 16: Theory Of Operation

    April 2009 Theory of Operation The TransPort is a transit-time ultrasonic flowmeter. When ultrasonic pulses are transmitted through a moving liquid or gas, the pulses that travel in the same direction as the fluid flow (downstream) travel slightly faster than the pulses that travel against the flow (upstream).

  • Page 17: Chapter 2: Initial Setup

    April 2009 Chapter 2 Initial Setup Before making measurements, you must prepare the TransPort for operation. This includes the following procedures: • Making Electrical Connections • Charging and/or Replacing Batteries • Powering On and Off • Using the Screen and Keypad •…

  • Page 18: Making Electrical Connections

    April 2009 Making Electrical Connections Before making measurements with the TransPort, you must make all the necessary connections to the unit. This section describes how to connect the following: • Power • Transducer • Analog Input/Output • Infrared Interface Make all connections to the top of the TransPort unit as shown in Figure 2-2 below.

  • Page 19: Power Connections

    April 2009 Power Connections The PT878GC is powered by either a 100-120/200-260 VAC wall mount plug-in module, or by 5 internal C -size NiCad high- energy rechargeable batteries or by a pack of 3.0 Ahr NiMH batteries. (An optional power supplement, part #703-1283, uses 6 AA alkaline batteries.) When you receive the PT878GC, the batteries are not charged;…

  • Page 20: The Infrared Wireless Interface

    April 2009 Analog Input/Output Connections (cont.) Table 2-1: Cable Assembly for Analog Inputs/Outputs Pin Number Wire Color Description Black Analog Out 1 16 V (switched) White Input A (Temperature or Pressure) Yellow Input B (Temperature or Pressure) Green Analog Ground Orange Digital Output Blue…

  • Page 21: Charging And/Or Replacing Batteries

    April 2009 Charging and/or Replacing Batteries The PT878GC comes with self-contained, built-in rechargeable batteries to support portable operation. For optimum performance, these batteries require a minimum of maintenance. Charging the Batteries When you receive the PT878GC, you will need to initially charge the batteries.

  • Page 22: Replacing The Batteries

    (see Figure 2-3 below). Battery Location (Behind Panel) Figure 2-3: Rear View of PT878GC To further extend the battery power on the PT878GC, the GE Sensing Part #705-1283 power pack option uses 6 AA alkaline batteries. Initial Setup…

  • Page 23: Powering On And Off

    GE Sensing logo and the software version and emits a long beep. If the meter fails any of these tests, contact GE Sensing. Caution! If the meter fails the backup battery test, you must send the unit back to the factory for a battery replacement.

  • Page 24
    April 2009 Powering On and Off (cont.) After the meter conducts all the self checks, the screen then appears similar to the one shown in Figure 2-4 below. ABC.SIT 2004/11/30 09:53 AM Velocity, ft/s Signal, dB 0.00 Volume, l/s Delta-T, ns 0.10 E0: No Errors Figure 2-4: Screen After Powering On…
  • Page 25
    April 2009 Powering On and Off (cont.) To turn the PT878GC off, press the red key for 3 seconds. The screen now appears similar to Figure 2-5 below. Velocity, ft/s Signal, dB SHUTDOWN: Meter OFF SLEEP: Meter Idle CANCEL: Resume Operations Volume, l/s Delta-T, ns Resume…
  • Page 26: Using The Screen And Keypad

    April 2009 Using the Screen and Keypad The essential features for operating the TransPort are the screen and keypad. Although these features are common on portable instruments, the PT878GC design offers unique features to simplify and speed operation. Screen The primary function of the screen is to display information in order for you to accurately and easily take measurements.

  • Page 27
    April 2009 Screen (cont). The middle of the screen is the work area, which displays the measured parameters, numeric measurements, and both bar and line graphs. (When you enter a selection on the Main Menu discussed in Chapter 4, Programming Site Data, this area displays menu prompts.) A line at the bottom of the area also displays error code messages, which are described in more detail in Chapter 10, Diagnostics and Troubleshooting.
  • Page 28: Keypad

    April 2009 Keypad The PT878GC keypad has 25 keys. The functions for each key are as follows (see Figure 2-7 on the next page): • [F1] [F2] [F3] 3 function keys ( ) — enable you to select the special functions which appear at the bottom of the screen. •…

  • Page 29
    April 2009 Keypad (cont.) Figure 2-7: The TransPort PT878GC Keypad Initial Setup 2-13…
  • Page 30: Obtaining On-Line Help

    April 2009 Obtaining On-Line Help The TransPort offers on-line help screens that describe various features. You can access on-line help at any time by pressing the [?] key. The screen appears similar to Figure 2-8 below. Help Velocity, ft/s Signal, dB Table of Contents Use the arrows and the enter key to select from the links below.

  • Page 31: Chapter 3: Installing The Dampening Material, Transducers And Fixtures

    Since the PT878GC is specifically designed for gas measurement with clamp-on transducers, it requires the use of specially designed fixtures and dampening material to maintain the highest possible measurement accuracy. GE Sensing supplies the CFG series of fixtures: • The V1 clamping fixture for pipes with diameters between 0.75 and 1.25 in.

  • Page 32
    April 2009 CFG-V1 CFG-V1 CFG-V4 CFG-V8 CFG-V8 CFG-V12 Layout Wrap PI Fixture Dummy Block Chain Strap Figure 3-1: The V Series and PI Fixtures Choosing the transducer installation locations on the pipe is an important aspect of proper flow measurement. Specifications assume a fully developed flow profile typically requiring 20 diameters upstream and 10 diameters downstream of straight pipe run.
  • Page 33: Application Requirements

    April 2009 Application Requirements Before you begin to use the PT878GC, you should ensure that your meter can handle the frequencies required for your particular application. To check the capabilities of your particular PT878GC, refer to the back label of the PT878GC, shown in Figure 3-2 below.

  • Page 34
    April 2009 Application Requirements (cont.) • For air, nitrogen, oxygen or argon, refer to Table 3-1 on the next page. • For natural gas, refer to Table 3-2 on page 3-7. • For steam, refer to Table 3-3 on page 3-9. 1.
  • Page 35
    April 2009 Table 3-1: PT878GC Installation Requirements for Air, Nitrogen, Oxygen or Argon Maximum Velocity, ft/s (m/s) Transducer Min. Pressure Single Dual Triple Four Five Pipe Size ANSI (DIN) Pipe Wall Inches (mm) psig (bar) Traverse Traverse Traverse Traverse Traverse 3/4 (20) <0.07 (1.8) 60 (5.1)
  • Page 36
    April 2009 Table 3-1: PT878GC Installation Requirements for Air, Nitrogen, Oxygen or Argon (Continued) Maximum Velocity, ft/s (m/s) Transducer Min. Pressure Single Dual Triple Four Five Pipe Size ANSI (DIN) Pipe Wall Inches (mm) psig (bar) Traverse Traverse Traverse Traverse Traverse 12 (300) <0.38 (9.7)
  • Page 37
    April 2009 Table 3-2: PT878GC Installation Requirements for Natural Gas Maximum Velocity, ft/s (m/s) Pipe Size ANSI Pipe Wall Inches Transducer Min. Pressure (DIN) (mm) psig (bar) Single Traverse Dual Traverse Triple Traverse Four Traverse 2 (50) <0.16 (4.1) 200 (14.8) 110 (33.5) 88 (26.8) 66 (20.5)
  • Page 38
    April 2009 Table 3-2: PT878GC Installation Requirements for Natural Gas (Continued) Maximum Velocity, ft/s (m/s) Pipe Size ANSI Pipe Wall Inches Transducer Min. Pressure (DIN) (mm) psig (bar) Single Traverse Dual Traverse Triple Traverse Four Traverse 18 (450) <0.38 (9.7) 300 (21.7) 78 (23.8) 59 (18.0)
  • Page 39
    April 2009 Table 3-3: PT878GC Installation Requirements for Steam Maximum Flow Velocity Pipe Pipe Wall Min. ft/s (m/s) Size in. Inches Transducer Pressure Single (mm) (mm) (MHz) psig (bar) Traverse 3 (80) <0.22 (5.6) 110 (8.6) 120 (36.6) <0.3 (7.6) 200 (14.8) 4 (100) <0.24 (6.1)
  • Page 40: Preparing The Pipe

    April 2009 Preparing the Pipe 1. Locate a transducer measurement point with 20 diameters of upstream straight run pipe and 10 diameters of downstream straight run pipe. In addition, the point should be at least 10 diameters from any butt welds or flanges. Keep appropriate clearance on either side of the pipe for easy transducer installation: •…

  • Page 41: Performing A Pipe Survey

    April 2009 Performing a Pipe Survey Finding a location where the pipe is concentric is important for optimum accuracy and performance. If possible, perform a pipe survey with an ultrasonic thickness gauge to find the best location. 2. Clear rust or loose paint and measure the wall thickness at five points along the pipe axis at 25 mm (1-in.) intervals using an ultrasonic thickness gauge, as shown in Figure 3-4 below.

  • Page 42
    April 2009 Performing a Pipe Survey (cont.) 4. Now measure the OD and the wall thickness at eight points along the pipe circumference at 45° intervals (shown in Figure 3-6 below), three times per point, and record the mean values. Figure 3-6: Measuring the OD and Pipe Wall Thickness 5.
  • Page 43: Obtaining The Transducer Spacing

    April 2009 Obtaining the Transducer Spacing 1. Using the measured OD and the pipe wall thickness, program the PT878GC (discussed in Chapter 4, Programming Site Data) to determine the required transducer spacing. 2. To determine the PT878GC correction factor, calculate the mean inside pipe diameter (ID) and the pipe ID at the transducer locations.

  • Page 44: Installing The V Series Clamping Fixture And Transducers

    9 o’clock position. Figure 3-7 below shows the two mounted halves. Note: If you are using a V4 fixture with a 1.5 to 2-in. pipe, GE Sensing suggests removing the set screws and the mounting bracket, rotating them and the transducer yoke 180°, and replacing the bracket, yoke and screws for a…

  • Page 45
    April 2009 Installing the Fixture (cont.) The two fixture halves have measuring scales; ensure that the scales are on the same side of the fixture, so that both zeros start at the same origin, as shown in Figure 3-8 below. Figure 3-8: Fixture with Scale Origins Properly Aligned 3.
  • Page 46: Installing The Transducers

    April 2009 Installing the Transducers 1. Apply a bead of couplant 6 mm (0.25 in.) wide along the entire length of each transducer face, as shown in Figure 3-10 below. Figure 3-10: Couplant on Transducer Face Note: Do not slide the transducer with couplant along the surface of the pipe when mounting the transducer.

  • Page 47
    April 2009 Installing the Transducers (cont.) 3. Slide the second mounting block to the calculated spacing plus the initial scale number selected for the first mounting block. For example: a. Initial convenient number for the first mounting block = 5 cm or 2 in. b.
  • Page 48: Installing The Pi Fixture And Transducers

    The PI clamping fixture holds transducers on pipes from 8 to 24 in. in diameter. It comes with either a chain or strap, depending on the selection made with the initial order from GE Sensing. To install the fixture and transducers, complete the following steps: Surveying the Pipe 1.

  • Page 49
    April 2009 Surveying the Pipe (cont.) Figure 3-14: Marking the 3 o’Clock Position 4. To find the coinciding point on the opposite site of the pipe (180° away from each other), divide the measured circumference by 2 and measure this distance along the circumferential lines from the zero point, as shown in Figure 3-15 below.
  • Page 50
    April 2009 Surveying the Pipe (cont.) Figure 3-16: Measuring the 180° Point from Top and Bottom 3-20 Installing the Dampening Material, Transducers and Fixtures…
  • Page 51: Installing The First Bracket With A Chain Or Strap

    April 2009 Installing the First Bracket with a Chain or Strap The following steps describe how to install the PI fixture with a supplied chain or strap. 1. Carefully wrap the chain or strap around the pipe, taking care not to twist it. 2.

  • Page 52: Installing The Second Bracket With A Chain Or Strap

    April 2009 Installing the Second Bracket with a Chain or Strap Note: The following step requires the transducer spacing discussed on page 3-7. 1. Measure the spacing from the zero point (the point of circumferential origin). Mark the spacing point with a crosshair on the opposite side of the pipe, 180°…

  • Page 53: Installing The Transducers

    April 2009 Installing the Second Bracket (cont.) The spacing should now appear similar to that shown in Figure 3-21 below. Figure 3-21: The Installed CFG-PI Fixture, with Calculated Spacing Installing the Transducers 1. Check to be sure the second CFG-PI holder bracket is correctly positioned.

  • Page 54
    April 2009 Installing the Second Bracket (cont.) Note: Do not slide the transducer with couplant along the surface of the pipe when mounting the transducer. 3. With one hand, mount one transducer into the PI fixture. With the other hand, tighten the transducer hold down screw, gradually pushing the transducer down to the pipe surface.
  • Page 55: Installing Dampening Material

    The pipe is old, with a history of scaling or rust, • The pipe experiences condensation on the outside. Note: Consult a GE Sensing flowmeter applications engineer or sales engineer if you have any questions regarding dampening material. Installing the Dampening Material, Transducers and Fixtures…

  • Page 56: Installing Dmp-1 Dampening Material

    April 2009 Installing DMP-1 Dampening Material 1. Be sure the CFG-V clamping fixture is installed on the pipe with the transducers as described in the section Installing the V Series Clamping Fixture and Transducers on page 3-14. With a marker, mark scribe lines on the inside edges of the brackets onto the pipe.

  • Page 57
    April 2009 Installing DMP-1 Dampening Material with CFG-V Series Fixtures (cont.) 7. With a marker, trace around the transducer footprint, as shown in Figure 3-24 below. Figure 3-24: DMP-1 Material with Transducer Footprint Traced 8. Remove the fixture and transducers. Then use a utility knife to cut out the area under the transducer footprint, and peel the cut material off the pipe, as shown in Figure 3-25 below.
  • Page 58
    April 2009 Installing DMP-1 Dampening Material with CFG-V Series Fixtures (cont.) 11. Lay the second strip of dampening material flat. Then cut the strip into two pieces, each 4.5-in. (12 cm) wide. 12. Wrap each of these strips around the pipe on the outside edge of the clamping fixture, one upstream and one downstream.
  • Page 59: Installing Dmp-1 Dampening Material With Pi Fixture

    April 2009 Installing DMP-1 Dampening Material with PI Fixture 1. Be sure the PI clamping fixture is installed onto the pipe with the transducers as described in the section Installing the PI Fixture and Transducers on page 3-18. 2. Approximate the axial distance from one transducer face to the other transducer face, as shown in Figure 3-27 below.

  • Page 60
    April 2009 Installing DMP-1 Dampening Material with PI Fixture (cont.) Figure 3-28: Completed DMP-1 Installation 3-30 Installing the Dampening Material, Transducers and Fixtures…
  • Page 61: Installing Dmp-3 Compound With All Fixtures

    April 2009 Installing DMP-3 Compound with All Fixtures 1. Be sure the fixture and transducers are installed on the pipe as described in the previous sections of this chapter. 2. Remove the fixture and transducers, but be sure to mark the approximate area of installation.

  • Page 62
    April 2009 Installing DMP-3 Compound with All Fixtures (cont.) 6. Spread the DMP-3 material so that it covers the whole area under the fixture to a thickness of about 0.25 in. (6.4 mm), as shown in Figure 3-31a below. Figure 3-31: Spreading the DMP-3 (a) and Reinstalling the Fixture (b) 7.
  • Page 63: Installing The Pdj Dampening Jacket

    April 2009 Installing the PDJ Dampening Jacket If the pipe temperature is over 150°F, you must use the PDJ pipe dampening jacket with preapplied DMP-3. As the dampening material dries out over several hours after installation, its effectiveness increases. The jacket is available in standard pipe sizes from 4 to 12 in.

  • Page 64
    April 2009 Installing the PDJ Dampening Jacket (cont.) !WARNING! The pipe and the dripping fluid will cause severe burns upon contact with bare skin. Also, be sure not to inhale the fumes generated during the DMP-3 curing cycle. 4. Install the fixture over the jacket, adjusting the spacing to match the prestamped transducer holes and PT878GC spacing calculations.
  • Page 65: Chapter 4: Programming Site Data

    April 2009 Chapter 4 Programming Site Data On the PT878GC, a Program Menu (part of the Main Menu) enables you to enter information that is specific to each site: • Transducer types and paths • Pipe materials and linings • Fluid types •…

  • Page 66: Entering The Program Menu

    April 2009 Entering the Program Menu [MENU] To enter the Program Menu, press the key at the lower right of the PT878GC keypad. The Menu Bar replaces the Status Bar at the top of the screen. Press the [ ] arrow key once to scroll from the Site Menu to the Program Menu.

  • Page 67: Entering Transducer Parameters

    Use the [ ] and [ ] keys to scroll between the two types. b. Press [ENTER] [SEL] to confirm the choice. IMPORTANT: While wetted transducers are listed as a choice, the PT878GC is designed for use with Shear wave clamp-on transducers. Consult GE before selecting wetted transducers. Programming Site Data…

  • Page 68
    April 2009 Entering Transducer Parameters (cont.) Note: The choices made earlier in the Transducer and Pipe options determine the prompts available later. If the PT878GC does not scroll to a particular parameter, it is not necessary for that transducer or pipe type. 2.
  • Page 69: Parameters For Special Transducers

    From the Frequency prompt, press the [ ] key to reach the [ENTER] Tw prompt, and press b. Use the numeric keys to enter the GE Sensing-supplied [ENTER] time delay and press 5. When calculating the flow rate, the PT878GC must take into account the wedge angle, the angle of the ultrasonic transmission.

  • Page 70
    ] key to reach [ENTER] the Wedge SS prompt, and press b. Use the numeric keys to enter the GE Sensing-supplied [ENTER] wedge sound speed (in ft/sec or m/sec) and press Pressing the [ ] key returns the meter to the Transducer tab at the top.
  • Page 71: Entering Pipe Parameters

    April 2009 Entering Pipe Parameters To enter the Pipe option, scroll to the Pipe entry on the Program [ENTER] Menu and press . (From the Transducer window, you can scroll back up to the Transducer tab and press the [ ] arrow key [ENTER] to reach the Pipe window and press .) The screen appears…

  • Page 72
    April 2009 Entering Pipe Parameters (cont.) 1. The first prompt asks you to select the pipe material. a. Press [ENTER] to enter the material prompt. b. A drop-down list of materials opens. Table 4-1 below lists the available preprogrammed materials on the list. Press the [ ] or [ ] keys to scroll to the appropriate material,…
  • Page 73
    April 2009 Entering Pipe Parameters (cont.) 2. At this point, the program follows one of two paths: a. If you have selected a preprogrammed material, the PT878GC automatically supplies the correct sound speed, and proceeds to step 3. b. If you have selected “Other,” the meter prompts you to enter the sound speed.
  • Page 74
    April 2009 Entering Pipe Parameters (cont.) If you select a material that has the Schedule option: a. The prompt asks if you wish to apply DIN (the DIN [ENTER] schedule). Press to select (or deselect) the DIN box. (If you do not select the DIN option, the prompt moves to the OD text box, and you enter the parameters for the Diameter alternative as discussed on the previous page.)
  • Page 75: Entering Pipe Lining Parameters

    April 2009 Entering Pipe Lining Parameters To enter the Lining option, scroll to the Lining entry on the [ENTER] Program Menu and press . (From the Pipe window, you can scroll back up to the Pipe tab and press the [ ] arrow key to [ENTER] reach the Lining window, and press .) The screen appears…

  • Page 76
    April 2009 Entering Pipe Lining Parameters (cont.) Note: If your pipe lining is not on the drop-down list, consult GE Sensing for further information. 2. The program follows one of two paths: a. If you have selected a preprogrammed material, the PT878GC automatically supplies the correct sound speed, and you proceed to step 3.
  • Page 77: Entering Fluid Types And Speeds

    April 2009 Entering Fluid Types and Speeds To access the Fluid option, scroll to the Fluid entry on the [ENTER] Program Menu and press . (If you are already in the Transducer/Pipe form, press the [ ] arrow key to reach the Fluid [ENTER] window, and press .) The screen appears similar to Figure…

  • Page 78
    April 2009 Entering Fluid Types and Speeds (cont.) b. Scroll to the appropriate entry. If you do not see your fluid on the list, select “Other.” c. Press [ENTER] to confirm your selection. If you have selected a preprogrammed gas, the PT878GC supplies the sound speed.
  • Page 79: Entering The Signal Path Parameters

    April 2009 Entering the Signal Path Parameters To enter the Path option, scroll to the Path entry on the Program [ENTER] Menu and press . (From the Lining window, you can scroll back up to the Lining tab and press the [ ] arrow key to reach the [ENTER] Path window, and press .) The screen appears similar to…

  • Page 80: Path Parameters For Clamp-On Transducers

    If you enter an invalid entry, the PT878GC rejects the entry and displays an error message. Note: GE Sensing does not recommend using a spacing other than the one calculated by the PT878GC. After you enter the spacing, pressing the […

  • Page 81: Entering The Standard Volume

    April 2009 Entering the Standard Volume/ Mass Flow Parameters The Standard Volume/Mass Flow option enables you to correct standard volumetric units for pressure and temperature, as well as to disable or enable mass flow. To enter this option, scroll to the [ENTER] Std Vol/Mass Flow entry on the Program Menu and press The screen appears similar to Figure 4-7 below.

  • Page 82
    April 2009 Entering the Standard Volume Parameters (cont.) 2. The next prompt asks for the base pressure. a. Press [ENTER] to enter the text box. b. Use the numeric keys to enter the known pressure. c. Press [ENTER] to confirm the entry. 3.
  • Page 83: Entering Inputs In The Standard Volume Option

    April 2009 Entering Inputs in the Standard Volume Option To enter input parameters in the Standard Volume/Mass Flow option, return to the Settings tab at the top of the Standard [ENTER] Volume window. Press the [ ] arrow key and to open the Inputs window.

  • Page 84
    April 2009 Entering Inputs in the Standard Volume Option (cont.) 2. The next prompt appears if you have selected a fixed pressure. The PT878GC asks for the fixed pressure value. a. Press [ENTER] to open the text box. b. Use the numeric keys to enter the known pressure. c.
  • Page 85
    April 2009 Entering Inputs in the Standard Volume Option (cont.) Note: If you have switched both active inputs to #A (or to #B), the PT878GC automatically switches one input to the other letter. This change ensures that only one analog input is assigned to one measurement type.
  • Page 86: Entering Mass Flow Parameters

    April 2009 Entering Mass Flow Parameters To enter mass flow parameters in the Standard Volume/Mass Flow option, press the [ ] arrow key until you reach the Mass [ENTER] Flow tab, and then press to open the Mass Flow window. The screen appears similar to Figure 4-9 below.

  • Page 87
    April 2009 Entering Mass Flow Parameters (cont.) 2. If you have enabled Mass Flow, the next prompt asks for the calculation method. Three options are available: static density, saturated steam, and steam. a. Press [ENTER] to open the drop-down menu. b.
  • Page 88: Entering Skan/Measure Parameters

    Entering Skan/Measure Parameters IMPORTANT: Do not change the Skan/Measure parameters unless recommended by GE Sensing. If you select one of the Skan/Measure modes in the Measurement mode window in the Signal Setup option (see page 9-24), the PT878GC requires you to set up the appropriate parameters. You can enter this option in either of two ways: •…

  • Page 89
    M>S Switch value. The default value is 5.0 µsec and values from 0 to 250 µsec are acceptable. Note: Do not change this value without consulting GE Sensing. If you wish to use Skan only, set the M/S Switch to a value of 0.
  • Page 90
    April 2009 Entering Skan/Measure Parameters (cont.) Pressing the [ ] key highlights the Advanced prompt. To enter [ENTER] advanced parameters for the Skan/Measure modes, press The screen now appears similar to Figure 4-11 below. In each case, to enter a value: a.
  • Page 91
    April 2009 Entering Skan/Measure Parameters (cont.) 6. The next prompt enables you to reset the size of the transmit window. Normally, the PT878GC calculates the size of both transmit and receive windows based on pipe size and fluid sound speed. However, for special diagnostic purposes, it may sometimes be necessary to reset the window size.
  • Page 92: Entering Skan/Measure Integrate Parameters

    April 2009 Entering Skan/Measure Integrate Parameters To enter parameters for the Skan/Measure Integrate mode in the Skan/Measure option, press the [ ] arrow key once to reach the [ENTER] Integrate tab. Then press to open the Integrate window, which appears similar to Figure 4-12 below. Skan/Measure Integrate Count…

  • Page 93
    April 2009 Entering Skan/Integrate Parameters (cont.) 3. If you wish to use Measure Times only (instead of both Skan [ENTER] and Measure times), press at the Measure Only [ENTER] prompt. (You can also press to deselect the prompt.) Pressing the [ ] key returns the meter to the A Divisor entry, and pressing the [ ] key returns the meter to the Integrate tab.
  • Page 94: Entering Skan/Measure Count Parameters

    April 2009 Entering Skan/Measure Count Parameters To open the Skan/Measure Count window in the Skan/Measure option, press the [ ] arrow key twice (from the All tab) to reach [ENTER] the Count tab. Press , and the screen appears similar to Figure 4-13 below.

  • Page 95
    April 2009 Entering Skan/Measure Count Parameters (cont.) 3. The next prompt asks for the percentage of Noise Level. a. Press [ENTER] to enter the text box. b. Use the numeric keys to enter the recommended value. c. Press [ENTER] to confirm the entry. 4.
  • Page 96: Entering Skan/Correlation Parameters

    April 2009 Entering Skan/Correlation Parameters To open the Skan/Correlation window in the Skan/Measure option, press the [ ] arrow key three times (from the All tab) to [ENTER] reach the Correlate tab, and press . The screen appears similar to Figure 4-14 below. Skan/Measure Integrate Count…

  • Page 97
    April 2009 Entering Skan/Correlation Parameters (cont.) 2. Repeat step 1 to enter the maximum percentage of peak. 3. The next prompt asks for the Percent of Peak, which is used to calculate the transit times and Delta-T. a. Press [ENTER] to enter the text box.
  • Page 98: Entering Analog Inputs

    April 2009 Entering Analog Inputs The Analog Inputs option enables you to specify parameters for general purpose or standard volume inputs. To enter the Analog Inputs option, scroll to the Analog Inputs entry on the Program [ENTER] Menu and press .

  • Page 99
    April 2009 Entering General-Purpose Analog Inputs (cont.) 2. The next prompt asks you to create a label for the input. The label corresponds to the measurement type (i.e., velocity or temperature), while the units symbol corresponds to the measurement units (i.e., feet/sec or degrees F). a.
  • Page 100
    April 2009 Entering General-Purpose Analog Inputs (cont.) 3. The next prompt asks for the Units Symbol. a. Press [ENTER] to reopen the text creation window. b. Repeat the procedure used for the Label (on the previous page) to create the Units Symbol. c.
  • Page 101: Entering Analog Inputs In Standard Volume

    April 2009 Entering Analog Inputs in Standard Volume or Mass Flow Option If you have enabled the Standard Volume or Mass Flow options, the screen displays different options, as shown in Figure 4-17 below. Analog Input Input A Input B Pressure Function Label…

  • Page 102
    April 2009 Entering Analog Inputs in the Standard Volume or Mass Flow Option (cont.) You have completed entering parameters in the Analog Inputs option. • [F3] To confirm the entries and return to Operate mode, press (OK). The PT878GC returns to Operate Mode. •…
  • Page 103: Entering The Analog Output

    April 2009 Entering the Analog Output The Analog Output option enables you to enter information to set up output parameters. To enter the option, scroll to the Analog [ENTER] Output entry on the Program Menu and press . The screen appears similar to Figure 4-18 below.

  • Page 104
    April 2009 Entering the Analog Output (cont.) 2. The next prompt asks you to select the analog output type from a number of choices, as shown in Figure 4-19 below. Select Measurement Diagnostics Velocity Temperature Pressure Volume UserFunction Aux Input A Aux Input B Skan/Measure Mass Flow…
  • Page 105
    April 2009 Entering the Analog Output (cont.) 3. The next prompt asks you to enter the zero (base) value for the analog output. This value represents the 0/4 mA output (in flow units). a. Press [ENTER] to open the text box. b.
  • Page 106: Entering The Digital Output

    April 2009 Entering the Digital Output While resembling the Analog Output option, the Digital Output option enables you to set up parameters necessary for a digital output. To enter the option, scroll to the Digital Output entry on [ENTER] the Program Menu and press .

  • Page 107
    April 2009 Entering the Digital Output (cont.) a. Press [ENTER] to open the drop-down menu. b. Scroll to the desired function. c. Press [ENTER] to confirm your selection. Note: If you select Off, you will not be able to access any other parameters in this option.
  • Page 108
    April 2009 Entering the Digital Output (cont.) d. The prompt then moves to a list of unit types. (The available units depend on the selection made at the Data Source prompt.) Scroll to the desired output unit. e. Press [F3] (OK) to confirm your selection.
  • Page 109
    April 2009 Entering the Digital Output (cont.) If you entered Test Points: The prompt asks for the signal type from two choices: transmit or receive. a. Press [ENTER] to open the drop-down menu. b. Scroll to the required signal. c. Press [ENTER] to confirm your selection.
  • Page 110: Entering User Functions

    April 2009 Entering User Functions User functions enable you to perform mathematical equations on each measurement. You can also use any parameter in the meter to calculate a different parameter. To enter the User Functions option, scroll to the User Functions entry on the Program Menu [ENTER] and press .

  • Page 111
    April 2009 Entering User Functions (cont.) 2. The next prompt asks you to create a label for the function. The label corresponds to the measurement type (i.e., velocity or temperature), while the units symbol corresponds to the measurement units (i.e., feet/sec or degrees F). a.
  • Page 112
    April 2009 Entering User Functions (cont.) 3. The next prompt asks for the Units Symbol. a. Press [ENTER] to reopen the text creation window. b. Repeat the procedure used for the Label (on the previous page) to create the Units Symbol. c.
  • Page 113
    When you have completed entering the function, press [F3] (OK) to confirm the entry and return to the User Function window. 6. GE Sensing recommends pressing [F2] (Check) to test the validity of the function. The PT878GC displays either “OK”…
  • Page 114: Entering Correction Factors

    April 2009 Entering Correction Factors The final option in the Program Menu, Correction Factors, allows you to enter and modify three correction factors: Reynolds Correction, Kinematic Viscosity and Calibration Factor. To enter the Correction Factors option, scroll to the Correction Factors [ENTER] entry on the Program Menu and press .

  • Page 115
    April 2009 Entering Reynolds Correction (cont.) b. Press [ENTER] to confirm your selection. Note: If you disable Reynolds Correction, you will not be able to enter any other values in this window. 2. If you enable Reynolds Correction, you will also need to enter the Kinematic Viscosity of the gas (available in the brochure Sound Speeds and Pipe Size Data, 914-004).
  • Page 116
    April 2009 Entering Reynolds Correction (cont.) b. Enter the desired measurement and units as described on page 4-40. c. Press [F3] (OK) to confirm your selection. 4. The prompt then asks if you wish to edit the Reynolds Correction table. a.
  • Page 117: Entering A Calibration Factor

    April 2009 Entering Reynolds Correction (cont.) e. When you have completed entering values, press [F3] (OK) to confirm the table and return to the Correction Factors window. You have completed entering data for a Reynolds Correction table. • [F3] To confirm the entries, press (OK).

  • Page 118
    April 2009 Entering a Calibration Factor (cont.) 1. The first prompt asks if you wish to enable the calibration factor. a. Use the [ ] and [ ] keys to scroll to the appropriate radio button. b. Press [ENTER] to confirm your selection. 2.
  • Page 119
    April 2009 Entering a Calibration Factor (cont.) b. Enter the desired measurement and units as described on page 4-40. c. Press [F3] (OK) to confirm your selection. 4. The prompt then asks if you wish to edit the K factor table. a.
  • Page 120
    April 2009 Entering a Calibration Factor (cont.) You have completed entering data for correction factors. • [F3] To confirm the entries and return to Operate Mode, press (OK). • To leave the window without confirming the entries and return [F2] [ESC] to Operate mode, press (Cancel) or the…
  • Page 121: Chapter 5: Creating And Managing Sites

    April 2009 Chapter 5 Creating and Managing Sites The PT878GC can store site data in files for current and future access. (To learn how to program setup data, refer to Chapter 4, Programming Site Data.) After you answer the necessary questions, simply save the information to a site file.

  • Page 122: The Site Manager

    April 2009 The Site Manager [ENTER] From the Site Menu, press to open the Site Manager. The screen appears similar to Figure 5-2 below. Site Manager File Site Sort Info: DEFAULT:SIT DEFAULT New Site **CURRENT SITE** 03/04/04 14:51:10 5111 bytes 118272 bytes free Refresh Exit…

  • Page 123
    April 2009 The Site Manager (cont.) Site Manager File Site Sort Info: DEFAULT:SIT DEFAULT Open New Site Save Save As Refresh Rename 09/04/04 14:51:10 Delete 5111 bytes 118272 bytes free Refresh Exit Figure 5-3: The Site Manager Menu • To create a new site, go to page 5-4. •…
  • Page 124: Creating A New Site

    April 2009 Creating a New Site [MENU] To create a new site in the Site Manager, press in the Site Manager. Be sure the option New is highlighted, and then press [ENTER] . The screen appears similar to Figure 5-4 below. New Site Delete Cancel…

  • Page 125
    April 2009 Creating a New Site (cont.) New Site New Site Use the selected site ‘DEFAULT.SIT’ as a template? Figure 5-5: Template Confirmation Window The meter returns to Operate Mode, with the new site name displayed in the upper left corner of the screen. Creating and Managing Sites…
  • Page 126: Opening An Existing Site

    April 2009 Opening an Existing Site If you want to return to a second site that you have previously saved, first highlight the replacement site in the left window of [MENU] the Site Manager. Then press to open the File Menu. [ENTER] Scroll to the Open option, and press .

  • Page 127: Saving A Site

    April 2009 Saving a Site You can save the current site in one of two ways: • From the Site Menu, you can scroll down to the Save Now [ENTER] option and press , or • [MENU] From within the Site Manager, press , scroll to the Save [ENTER] option, and press…

  • Page 128: Saving A Site With A Different Name

    April 2009 Saving a Site with a Different Name If you want to save the current site with a different name, open [MENU] the Site Manager, press to open the File Menu, scroll to [ENTER] the Save As option, and press .

  • Page 129: Refreshing A Site

    April 2009 Refreshing a Site You can refresh a site (updating the display with the most current information) in one of two ways: • [F2] Press , Refresh, to refresh the highlighted file. • From the File Menu, scroll to the Refresh option and press [ENTER] The updated information on the highlighted file appears in the window on the right.

  • Page 130: Deleting A Site

    April 2009 Deleting a Site To delete a site in the Site Manager, first be sure you have highlighted that site in the left window of the Site Manager. Then [MENU] [ENTER] press , scroll to the Delete option, and press .

  • Page 131: Creating A Site Message

    April 2009 Creating a Site Message The Site Message option allows you to add an explanatory message (with up to 30 characters or spaces) for any given site. To create a site message: 1. Press [MENU] to enter the File Menu. Then press the [ ] arrow key once to scroll from the File Menu to the Site Menu.

  • Page 132
    April 2009 Creating a Site Message (cont.) “ ‘ < > Delete Cancel Figure 5-12: The Text Creation Window 3. Use the four arrow keys to scroll to the desired letter or [ENTER] symbol, and press to add the letter to the message. Note: Pressing [SEL] causes the screen to alternate between a…
  • Page 133: Printing A File

    April 2009 Printing a File [MENU] To print a file from the Site Manager, press to enter the File Menu, press the [ ] arrow key once, scroll to the Print [ENTER] option, and press . The PT878GC shows a message indicating that its infrared scanner is looking for a receiving device.

  • Page 134: Transferring A File As Text

    April 2009 Transferring a Site to a PC (cont.) Figure 5-13: The IR_Inbox Folder with Transferred Site File Transferring a File as Text If you want to review site programming data as text, the PT878GC can transfer a site file to a PC in a Unicode text format that can be opened by text editors such as Notepad or Word.

  • Page 135: Transferring A File From A Pc To The Pt878Gc

    April 2009 Transferring a File from a PC to the PT878GC Once you have stored site or meter files to a PC, you can then transfer them back to the PT878GC over the IR interface. The PT878GC only accepts files with a .sit (site) or .met (meter) extension.

  • Page 136
    April 2009 Transferring a File from a PC to the PT878GC (cont.) QuickBeam opens a window indicating that it is downloading the file. If you close and reopen the Site Manager, the file appears as one of the listed sites. Note: While sending or receiving files, the PT878GC continues to perform measurements, but at a slower rate than normal.
  • Page 137: Listing Files By Name

    April 2009 Listing Files by Name If you want to list your files alphabetically by site name within [MENU] the Site Manager, press to open the File Menu. Then press the [ ] arrow key twice to scroll from the File Menu to the Sort [ENTER] Menu, shown in Figure 5-15 below.

  • Page 138: Chapter 6: Displaying And Configuring Data

    April 2009 Chapter 6 Displaying and Configuring Data The PT878GC allows you to view from one to four different measurement parameters simultaneously. The screen can show these parameters not only in numeric format, but as line or bar graphs as well. You can configure any given measurement for your particular requirements.

  • Page 139: The Format Option

    April 2009 The Format Option The first option on the menu, Format, allows you to specify the type of notation and the number of decimal places for the [ENTER] measurement in that window. Press and the submenu shows two options, Decimal Place and Notation. Press the [ ] or [ENTER] ] arrow keys to select either entry, and press…

  • Page 140
    April 2009 The Format Option (cont.) 2. The second step asks you to choose the number of decimal places to be displayed. Available choices range from 0 to 4 places. a. Press [ENTER] to open the drop-down list. b. Scroll to the appropriate number. c.
  • Page 141: The View Option

    April 2009 The View Option The second option, View, allows you to select the presentation of a parameter in one of three formats: numeric, line graph or bar graph. 1. From the Display Menu, press the [ ] key once to reach the [ENTER] View option, and then press 2.

  • Page 142: The Limits Option

    April 2009 The Limits Option Once you have configured a parameter as a line or bar graph, you might need to change its presentation or values. The Limits option (replacing the Format option for line and bar graphs) enables you to program the minimum or maximum values displayed, the time interval and the display of the average value.

  • Page 143
    April 2009 The Limits Option (cont.) 4. The first prompt asks for the minimum value shown in the graph. a. Press [ENTER] to open the text box. b. Use the numeric keys to enter the desired value. c. Press [ENTER] to confirm the value.
  • Page 144: The Measurement Option

    April 2009 The Measurement Option On occasion, you might need to change the actual parameter measured in a given window.The Measurement option enables you to reconfigure the window with one of five categories of data source (velocity, volume, forward or reverse totalizer, or diagnostics) and appropriate English or metric measurement units.

  • Page 145
    April 2009 The Measurement Option (cont.) Select Measurement Programming Diagnostics Delta-T Delta-T Velocity Amplitude Up Amplitude Down Temperature T Up Pressure T Down Volume Gain Up Fwd Totalizer Gain Dn Rev Totalizer Signal Up StdFwdTotalizer Signal Down Threshold Up StdRevTotalizer Threshold Down User Function Norm Factor…
  • Page 146: Customizing The Display Screen

    April 2009 Customizing the Display Screen You might wish to display one or two parameters, or customize the soft keys to quickly access particular menus. The Site Menu enables you to make more comprehensive changes in your display screen. [MENU] To enter the Site Menu, press the key at the lower right of the PT878GC keypad.

  • Page 147: Specifying The Number Of Displayed Parameters

    April 2009 Specifying the Number of Displayed Parameters As mentioned earlier, the PT878GC can display one to four different measurement parameters simultaneously. However, sometimes you might wish to display only one or two parameters. To change the number of open display windows: 1.

  • Page 148: Customizing Softkeys

    April 2009 Customizing Softkeys When the screen is in Operate mode, you might wish to access a particular submenu frequently without the trouble of scrolling through menus. Customizing the softkeys (the function keys — [F1] [F2] [F3] ) allows you to access up to three submenus by pressing the associated softkey.

  • Page 149
    April 2009 Customizing Softkeys (cont.) Configure FKey Select a menu item (CLEAR) Site Manager Save Now 1 View 2 Views Cancel Figure 6-9: The Configure FKey Window 4. Press [ENTER] to open the menu drop-down list. Then use the ] or [ ] arrow keys to scroll to the desired submenu (for example, Contrast).
  • Page 150
    April 2009 Customizing Softkeys (cont.) 2004/3/30 09:53 AM ABC.SIT Velocity, m/s Gain, dB 5000 -1.5121 -5000 Volume, ft Delta-T, ns -1.65 E2: Soundspeed Contrast Figure 6-10: The Display Screen with Modified Softkey • To clear a softkey: 1. Press [MENU] 2.
  • Page 151: Managing Files — The Drive Manager

    April 2009 Managing Files — The Drive Manager On occasion, you might want to review, print or transfer some or all of the files in the PT878GC. The Drive Manager allows you to view all the files stored in the meter. To open the Drive Manager: 1.

  • Page 152
    April 2009 Managing Files — The Drive Manager (cont.) File Manager Sort File Transfer Info: SYSLOG.MET SYSLOG. Refresh DEFAULT. Delete GLOBAL. TABLE.MET FUNCTIO AAA.LOG 03/15/04 10:39:20 BBB.LOG 6402 bytes CCC.LOG 91136 bytes free Q.LOG Refresh Exit Figure 6-12: The File Manager Menu To transfer, refresh or delete a file, first be sure the file is [MENU] highlighted in the window on the left.
  • Page 153: Refreshing A File

    April 2009 Refreshing a File You can refresh a file (updating the display with the most current information) in one of two ways: • [F2] Press , Refresh, to refresh the highlighted file. • From the File Menu, scroll to the Refresh option and press [ENTER] The updated information on the highlighted file appears in the window on the right.

  • Page 154: Transferring A File From A Pc To The Pt878Gc

    April 2009 Transferring a File to a PC (cont.) Figure 6-13: The IR_Inbox Folder with Transferred File Transferring a File from a PC to the PT878GC Once you have stored site or meter files to a PC, you can then transfer them back to the PT878GC over the IR interface.

  • Page 155
    April 2009 Transferring a File from a PC to the PT878GC (cont.) Note: For Windows NT 4.0, check that the QuickBeam software (available through the path C:Program FilesQuickBeam Suite) is running, and that the IR beam on the PT878GC has clear access to the IR sensor connected to the PC port.
  • Page 156: Deleting A File

    April 2009 Deleting a File To delete a file in the File Manager, first be sure you have highlighted that site in the left window of the File Manager. Then [MENU] [ENTER] press , scroll to the Delete option, and press .

  • Page 157: Listing Files By Name

    April 2009 Listing Files by Name If you want to list your files alphabetically by site name within [MENU] the File Manager, press to open the File Menu. Then press the [ ] arrow key to scroll from the File Menu to the Sort Menu, [ENTER] shown in Figure 6-16 below.

  • Page 158: Accessing Meter Data -The About Option

    To open the About window, scroll to the About option on the Site Menu and press [ENTER] . The screen appears similar to Figure 6-17 below. GE Industrial Sensing PT878GC Portable Flowmeter Copyright ©2002 General Electric Co. All rights reserved.

  • Page 159: Chapter 7: Programming Meter Settings

    April 2009 Chapter 7 Programming Meter Settings Along with display formats and site data, PT878GC users can program global settings for the meter that suit their individual preferences. The global settings include: • English or Metric measurement units • Battery power •…

  • Page 160: Entering The Meter Menu

    April 2009 Entering the Meter Menu [MENU] To enter the Meter Menu, press the key at the lower right of the PT878GC keypad. The Main Menu replaces the Status Bar at the top of the screen. Press the [ ] arrow key twice to scroll from the Site Menu to the Meter Menu.

  • Page 161: Selecting Measurement Units

    April 2009 Selecting Measurement Units The first option, Units, enables you to select either English or Metric units as global measurement units for the PT878GC. The selected units then become the default settings for every measurement that has the option for metric/English units. In addition, you can select the type of pressure units displayed.

  • Page 162
    April 2009 Selecting Measurement Units (cont.) a. Press [ENTER] to open the drop-down menu. b. Use the [ ] and [ ] arrow keys to scroll to the desired choice: • PSIA • • Bars • MMGH c. Press [ENTER] to confirm the entry.
  • Page 163: The Battery Charger

    April 2009 The Battery Charger The Battery option allows you to monitor the current run time and status of the internal rechargeable batteries, as well as to condition NiCad batteries to maintain the maximum life possible. Conditioning NiCad batteries (a process that can take up to 12 hours for a fully charged pack) discharges the pack completely and then performs a fast charge.

  • Page 164
    April 2009 The Battery Charger (cont.) 3. To condition the batteries, press the [ENTER] key. The “Condition Battery” window should now read “Stop.” The status line should change to “Discharging.” (Updating the status could take up to 30 seconds.) Note: It is advisable to condition NiCad batteries when their run time is reduced to 50% of their previous run time.
  • Page 165: Entering Date And Time

    April 2009 Entering Date and Time In Operate mode, the Status Bar displays the current date and time above the measurements in the upper right corner of the screen. The Date/Time option allows you to set the date or time, which are required for correct data logging operation.

  • Page 166
    April 2009 Entering Date and Time (cont.) To set the date: 1. From the date text box, press the [ ] key once to scroll to the time text box. This box displays the current meter time. 2. Press [ENTER] to enter the text box.
  • Page 167: Changing Date And Time Appearance (Locale)

    April 2009 Changing Date and Time Appearance (Locale) In addition to setting the correct date and time, you can also change its presentation to suit local preferences. You can select a time display of AM/PM or 24-hour time To alter the time and date display: 1.

  • Page 168
    April 2009 Changing Date and Time Appearance (cont.) 2. The first series of prompts asks you to select separator symbols for the date, time and decimal. You can choose from a dash, comma and slash (/) for the date; from a period and colon for the time: and from a period or comma for the decimal place.
  • Page 169
    April 2009 Changing Date and Time Appearance (cont.) A line at the bottom, the Date/Time, Decimal Formats, displays how the format and separator selections will appear on the screen. Pressing the [ ] arrow key returns the PT878GC to the Locale tab.
  • Page 170: Adjusting The Contrast

    April 2009 Adjusting the Contrast For more comfortable viewing in a particular environment, the PT878GC enables you to adjust the screen contrast. To adjust the screen contrast: 1. From the Meter menu, scroll to the Contrast entry and press [ENTER] .

  • Page 171
    April 2009 Adjusting the Contrast (cont.) • [F3] To confirm the entries and return to Operate mode, press (OK). • To leave the window without confirming the entries and return [F2] [ESC] to Operate mode, press (Cancel) or the key. Programming Meter Settings 7-13…
  • Page 172: Setting Backlight Timeout

    April 2009 Setting Backlight Timeout By using the Backlight Timeout option, you can set a specified time that the PT878GC backlight will remain on before turning itself off. Automatic turnoff enables the PT878GC to conserve battery power. To set the backlight timeout: 1.

  • Page 173: Changing Communications Parameters

    2. The first prompt asks for the node identification number, which can be any number from 1 to 240. IMPORTANT: Do NOT change the node ID unless instructed by GE Sensing. a. Press [ENTER] to open the text box. b. Use the numeric keys to enter the node ID.

  • Page 174
    April 2009 Changing Communications Settings (cont.) 3. The next prompt asks you to select the communications interface from two options: IrDA (Infrared Data Association) and the IR232 virtual interface. You must use IrDA when using an IR printer or transfer application such as QuickBeam. Use the [ ] and [ ] keys to scroll to the desired selection, [ENTER] and press…
  • Page 175
    April 2009 Changing Communications Settings (cont.) 6. The next prompt asks you to select either one or two stop bits.The default number is one. Use the [ ] and [ ] keys to [ENTER] scroll to the desired number, and press 7.
  • Page 176: Resetting Forward And Reverse Totals

    April 2009 Resetting Forward and Reverse Totals On occasion, it might be necessary to clear and reset the forward and reverse totals computed by the Forward and Reverse Totalizers. To reset the totals: 1. From the Meter menu, scroll to the Totals entry and press [ENTER] .

  • Page 177: Setting Up User Tables

    April 2009 Setting Up User Tables When you program user functions (see page 4-46), you can also support them with up to six user tables of non-linear or empirical data. To program one or more user tables: 1. From the Meter menu, scroll to the User Tables entry and [ENTER] press .

  • Page 178
    April 2009 Setting up User Tables (cont.) 3. The next step is to create a Table ID. a. Press [ENTER] to open the window. The screen now appears similar to Figure 7-9 below. UserTable1 “ ‘ < > Delete Cancel Figure 7-9: The Text Creation Window b.
  • Page 179
    April 2009 Setting up User Tables (cont.) 4. The final prompt asks you to enter or change data in the user table. a. Press [ENTER] to open the window. The screen appears similar to Figure 7-10 below. Edit User Table 1 Programming UserTable1 Cancell…
  • Page 180
    April 2009 Setting up User Tables (cont.) e. Repeat steps b, c and d on page 7-21 until you have completed entering data for the table. f. When you have finished, press [F3] (OK) to confirm the [F2] entries or (Cancel) to leave the window without confirming the table.
  • Page 181: Taking A Bitmap Capture Of A Current Screen

    April 2009 Taking a Bitmap Capture of a Current Screen The Snapshot option enables you to take a “screen capture” of the current screen in bitmap format (.bmp) for display or storage in a Windows-based PC. To take a “snapshot” of the screen: 1.

  • Page 182: Chapter 8: Logging Data

    April 2009 Chapter 8 Logging Data A powerful and flexible feature of the PT878GC is data logging. The meter enables you to choose up to 12 parameters to log. You can also select the start time and date, end time and date, and time interval.

  • Page 183: Entering The Logging Menu

    April 2009 Entering the Logging Menu [MENU] To enter the Logging Menu, press the key at the lower right of the PT878GC keypad. The Main Menu replaces the Status Bar at the top of the screen. Press the [ ] arrow key three times to scroll from the Site Menu to the Logging Menu.

  • Page 184: The Log Manager

    April 2009 The Log Manager The Log Manager offers users a way to check the status and memory size of all the logs currently pending, running or finished. To select Log Manager, scroll to the Manager entry on [ENTER] the Logging Menu and press .

  • Page 185: The File Menu

    April 2009 The File Menu The File Menu allows you, not only to create new logs, but also to copy, rename or delete logs, as well as to print them or transfer them to a PC. To open the File menu from the Log Manager, press [MENU] [ENTER] key and then…

  • Page 186: Setting Up A New Log

    April 2009 Setting up a New Log The New Log option enables you to create and set up parameters for a new log. You can access this option in two ways: • by scrolling to the New Log option in the Logging Menu (as [ENTER] shown in Figure 8-1 on page 8-2) and pressing , or…

  • Page 187
    April 2009 Setting up a New Log (cont.) Measurements General Log Name 10SEC.LOG Format Circular Linear Type Error Standard Start Date/Time 01:38:08 2004/11/01 09:38:08 End Date/Time 2004/11/01 Logging Interval secs Activate Cancel Figure 8-5: The General Log Format Window To step through each parameter, press the [ ] key. 1.
  • Page 188
    April 2009 Setting up a New Log (cont.) 3. The next prompt asks for the starting date and time. a. Press [ENTER] to enter the text box. The meter highlights the first number. Use the [ ] and [ ] keys to scroll to any number you wish to change, or the [ ] key to scroll to the time box.
  • Page 189
    April 2009 Setting up a New Log (cont.) General Measurements NO UNIT NO UNIT NO UNIT NO UNIT NO UNIT NO UNIT NO UNIT NO UNIT NO UNIT NO UNIT NO UNIT NO UNIT Activate Cancel Figure 8-6: The Log Measurements Window To step through each entry, press the [ ] key.
  • Page 190
    April 2009 Setting up a New Log (cont.) Select Measurement Pressure Diagnostics Velocity Temperature Volume Fwd Totalizer Rev Totalizer StdFwdTotalizer StdRevTotalizer Cancel No Unit Figure 8-7: The Select Measurement Window 2. Scroll to the desired output type. 3. Press [SEL] to confirm your selection.
  • Page 191: Copying (Cloning) A Selected Log

    April 2009 Copying (Cloning) a Selected Log In the Clone Selected Log option, you can copy the parameters of a particular log, modify the parameters, and start the copy. To copy a log: 1. First, select the log you wish to copy in the Log Manager (shown in Figure 8-3 on page 8-4).

  • Page 192: Renaming A Log

    April 2009 Renaming a Log To rename a log: 1. First, select the log you wish to rename in the Log Manager (shown in Figure 8-3 on page 8-4). 2. Then open the File menu, scroll to the Rename option, and [ENTER] press 3.

  • Page 193: Deleting All Logs

    April 2009 Deleting a Log (cont.) Log Manager Delete File Programming Info:DEFAULT DEFAULT Delete Log ‘555.LOG?’ GLOBAL 1 View Figure 8-8: The Delete Confirmation Window Deleting All Logs To clear the Log Manager and memory of all logs, open the File [ENTER] menu, scroll to the Delete All Logs option, and press [F2]…

  • Page 194: Printing A Log

    April 2009 Printing a Log To print a file from the Log Manager: 1. Highlight the desired log in the window on the left of the Log Manager. 2. Press [MENU] [ENTER] , scroll to the Print option, and press The PT878GC shows a message indicating that its infrared sensor is looking for a printer.

  • Page 195: The Log Menu

    April 2009 The Log Menu The Log Menu allows you to pause, restart or end any or all logs that are currently pending or running. (However, you cannot restart any finished logs, even if they were finished before the programmed end time.) To open the Log Menu in the Log [MENU] Manager, press .

  • Page 196: Stopping (Pausing) A Log

    April 2009 Stopping (Pausing) a Log To stop a log that is currently pending or running, 1. First, select the log you wish to pause in the Log Manager (shown in Figure 8-2 on page 8-4). 2. Then open the Log menu, scroll to the Pause option, and press [ENTER] The PT878GC returns to the Log Manager, which displays the highlighted log with “Paused”…

  • Page 197: Pausing All Logs

    April 2009 Pausing All Logs To pause all logs that are currently pending or running, open the [ENTER] Log menu, scroll to the Pause All Logs option, and press The PT878GC returns to the Log Manager, which displays all logs programmed to run now or in the future with a status of “Paused.”…

  • Page 198: The View Menu

    April 2009 The View Menu Through the View menu, you can view the data of individual logs in graphical or spreadsheet formats. To open the View menu from [MENU] the Log Manager, press . Scroll to the View menu, and [ENTER] press .

  • Page 199: Displaying Log Details

    April 2009 Displaying Log Details To view details of a given log, be sure the log is highlighted in the left window of the Log Manager. Then scroll to the View menu [ENTER] [ENTER] and press . Scroll to the Details option and press The screen now appears similar to Figure 8-11 below.

  • Page 200: Displaying Log Data In Graphical Form

    April 2009 Displaying Log Data in Graphical Form To view a log in graphical form, be sure the log is highlighted in the left window of the Log Manager. 1. Scroll to the View menu and press [ENTER] 2. Scroll to the Graph option and press [ENTER] 3.

  • Page 201
    April 2009 Displaying Log Data in Graphical Form (cont.) AAA.LOG (Dew Point, °C) Y-Axis -52.0 Limits Range Minimum 0.1234 Maximum 0.5678 2004/03/01 2004/03/01 13:53:41 PM 13:23:41 PM Cancel Figure 8-13: The Y-Axis Window The Y-Axis window allows you to specify whether the Y axis on the graph extends to the maximum value (Max), over the entire range (Range) or between certain specified values (Set).
  • Page 202
    April 2009 Displaying Log Data in Graphical Form (cont.) [F2] If you wish to alter the time scale, press (Time). The Enter Time window appears similar to Figure 8-14 below. AAA.LOG (P#Up) -52.0 Enter Time Start Date Start Time 2004/03/01 13:23:41 PM End Date End Time…
  • Page 203: Displaying Log Data In Spreadsheet Form

    April 2009 Displaying Log Data in Spreadsheet Form To view a log in spreadsheet form, be sure the log is highlighted in the left window of the Log Manager. Then scroll to the View [ENTER] menu and press . Scroll to the Spreadsheet option and [ENTER] press .

  • Page 204: The Sort Menu

    April 2009 The Sort Menu The Sort Menu within the Log Manager allows you to arrange your log list either alphabetically (By Name) or chronologically (By Date). Log Manager Sort File View By Name Info: AAA.LOG State: Finished By Date S:03/01/04 13:13:41 E:03/01/04 13:23:41 Interval: 10 Seconds…

  • Page 205: Chapter 9: Servicing The Pt878Gc

    April 2009 Chapter 9 Servicing the PT878GC For user convenience, the PT878GC offers a Service Menu. This menu enables users to perform a variety of functions that they might occasionally require: • print out reports • calibrate the PT878GC • run diagnostics •…

  • Page 206: Entering The Service Menu

    April 2009 Entering the Service Menu [MENU] To enter the Service Menu, press the key at the lower right of the PT878GC keypad. The Main Menu replaces the Status Bar at the top of the screen. Press the [ ] arrow key four times to scroll from the Site Menu to the Service Menu.

  • Page 207: Printing Reports

    April 2009 Printing Reports When used with an IR-compatible printer, the PT878GC can print out a variety of data (current site, logs, drive contents, and user functions, tables, and settings) in the Reports option. To enter the Reports option, scroll to the Reports entry on the Service Menu [ENTER] and press .

  • Page 208
    April 2009 Printing Reports (cont.) Figure 9-3: Printout of a Typical Drive Report Servicing the PT878GC…
  • Page 209: Setting Up The Thickness Gauge

    April 2009 Setting up the Thickness Gauge For greatest accuracy in flow applications, the PT878GC can measure pipe wall thickness using an optional thickness gauge transducer, instead of relying on the nominal pipe wall thickness. In Thickness Gauge mode, the PT878GC does not measure flow, but it can determine the thickness of most standard metal and plastic pipe materials over a range from 0.05 to 3 in.

  • Page 210
    April 2009 Setting up the Thickness Gauge (cont.) Calibrating the thickness gauge involves up to two additional steps: 4. Calibrating the thickness gauge itself (in the Zero window on page 9-12), and 5. Calibrating the velocity of sound in the pipe material (in the Velocity window on page 9-14), if possible and if the pipe material is not the material entered in the Zero window.
  • Page 211: Measuring Pipe Wall Thickness

    April 2009 Measuring Pipe Wall Thickness Entering the Material and Sound Speed IMPORTANT: GE Sensing recommends calibrating the thickness gauge periodically (as discussed on page 9-12) before measuring thickness. To enter the Thickness Gauge Display option, scroll to the [ENTER] T-Gauge Display entry on the Service Menu and press The screen appears similar to Figure 9-5 below.

  • Page 212
    April 2009 Entering the Material and Sound Speed (cont.) If you have selected a preprogrammed material, you have completed entering data in this window. Pressing the [ ] key returns the program to the Material tab. But if you selected “Other,”…
  • Page 213: Measuring Thickness In Numeric Format

    April 2009 Measuring Thickness in Numeric Format To measure the actual thickness of a pipe, apply couplant to the calibrated thickness gauge transducer and hold the transducer steady against the pipe. Then press the [ ] or [ ] arrow key [ENTER] until you reach the Display tab and press .

  • Page 214: Displaying The Receive Signal In Graphical Format

    April 2009 Displaying the Receive Signal in Graphical Format Used chiefly for diagnostic purposes, the Graph option helps to determine why the thickness gauge is not working if you suspect a problem. The graph shows an image of the acoustic signal. If the display does not show a signal image similar to Figure 9-7 below, you may have a problem with the transducer, couplant, or ®…

  • Page 215
    April 2009 Displaying the Receive Signal in Graphical Format (cont.) Thickness Gauge Measure Display Material Graph Velocity Zero Signal Display Figure 9-8: The Graph Window in Zoom Format • [F1] Press (In) to zoom in to magnify screen details. • [F2] Press (Out) to zoom out fully.
  • Page 216: Calibrating The Thickness Gauge Transducer

    See Appendix C for more information. 3. The next prompt asks you to enter the length of Block 1. (If you are using a GE Sensing-supplied test block, the length is [ENTER] printed on the block.) Press to open the text box.

  • Page 217
    IMPORTANT: Be sure to hold the transducer steady until the GE Sensing wait cursor disappears. 7. If you have chosen dual calibration, repeat Steps 3 through 6 above for Block 2. 8. The PT878GC asks for confirmation of the calculated and current values.
  • Page 218: Calculating Velocity (Pipe Material Sound Speed)

    April 2009 Calculating Velocity (Pipe Material Sound Speed) The nominal sound speed for your pipe material will give thickness measurements of reasonable accuracy. However, for greater accuracy, calibrate pipe material sound velocity. IMPORTANT: This step is necessary only if the actual pipe material sound speed differs from the sound speed of the material used in the calibration block.

  • Page 219
    April 2009 Calculating Velocity (Pipe Material Sound Speed) (cont.) 2. To enter the block length (here, the thickness of the sample as measured with a caliper or calibrated sample), press the [ [ENTER] arrow key and then press to open the Block Length text box.
  • Page 220
    April 2009 Calculating Velocity (Pipe Material Sound Speed) (cont.) 7. The “Calculated” box shows the thickness value measured. The PT878GC asks for confirmation of the calculated and [ENTER] current values. Press to commit (accept) the calculated value (in which case it becomes the current value), and to recalculate the value (which becomes the new “Calculated”…
  • Page 221: Programming The Thickness Gauge

    • Detection Threshold IMPORTANT: The thickness gauge programming settings are entered at the factory. You should not change them unless instructed by GE Sensing. The Programming window appears similar to Figure 9-12 below. Thickness Gauge Setup Programming Low Signal Thresh Transducer Delay µs…

  • Page 222
    April 2009 Programming the Thickness Gauge (cont.) 1. Press the [ ] arrow key to enter the window. 2. The first prompt asks for the low signal threshold. If the signal strength falls below this value, a “Low Signal” message appears on the screen.
  • Page 223
    April 2009 Programming the Thickness Gauge (cont.) 6. The final value, the detection threshold, represents the percent of peak the PT878GC uses to make measurements. It will consider anything above the entered percentage as part of the signal. To change this value: a.
  • Page 224: Displaying Diagnostic Parameters

    April 2009 Displaying Diagnostic Parameters The Diagnostics option enables you to view current diagnostic parameters without having to open a display window in Operate Mode. To enter the option, scroll to the Diagnostics entry on the [ENTER] Service Menu and press .

  • Page 225: Calibrating The Analog Output And Inputs

    April 2009 Calibrating the Analog Output and Inputs The Calibration option allows you to calibrate the analog output and inputs. To enter the option, scroll to the Calibrate entry on the [ENTER] Service Menu and press . The screen appears similar to Figure 9-14 below.

  • Page 226
    April 2009 Calibrating the Analog Output (cont.) 1. Connect a digital multimeter or ammeter to the analog output. 2. The next prompt asks for the setpoint (4 or 20 mA). Use the [ ] and [ ] arrow keys to move to the appropriate radio [ENTER] button and press 3.
  • Page 227: Calibrating Inputs

    April 2009 Calibrating Inputs To open the Inputs window, press the [ ] arrow key and press [ENTER] . The screen appears similar to Figure 9-15 below. Note: Calibrating the analog inputs requires use of a current source. Calibrate/Test I/O Analog Input Analog Output Input#…

  • Page 228
    April 2009 Calibrating Inputs (cont.) 4. Connect the current source to the input and set it to deliver the value entered in the previous step. 5. Press the [ ] arrow key to move to the Read box, and then [ENTER] press to read the low input.
  • Page 229: Setting Up Signal Parameters

    April 2009 Setting up Signal Parameters The Signal Setup option enables you to set parameters that affect the transducer signal: • Delta-T Offset • Transmit Sample Size • Zero Cutoff • Velocity Averaging • Errors Allowed • Peak Detection Method •…

  • Page 230
    2. The first prompt asks for the Delta-T offset. Delta-T is the difference between the upstream and downstream transit times of the transducers. The Delta-T offset should normally be set to zero. Note: Consult GE Sensing before performing this step. a. Press [ENTER] to open the text box.
  • Page 231
    April 2009 Setting up Signal Parameters (cont.) 5. The next prompt asks for velocity averaging, the number of velocity measurements to average together to smooth out noise in the system. a. Press [ENTER] to open the drop-down list. b. Use the [ ] or [ ] arrow keys to scroll to the desired number.
  • Page 232: Setting Up The Measurement Mode

    Note: Do not change the measurement mode or values unless recommended by GE Sensing. From the Signal Parameter tab, press the [ ] arrow key to move to the Meas Mode tab. The window appears similar to Figure 9-17 below.

  • Page 233
    April 2009 Setting up Measurement Mode (cont.) The first prompt asks for the measurement mode detection method. Use the [ ] and [ ] arrow keys to move to the [ENTER] appropriate radio button and press You must now enter the setup data covered on page 4-23, if you have not already done so.
  • Page 234: Setting Up Pulse/Code Parameters

    April 2009 Setting Up Pulse/Code Parameters When you select the Skan detection method, you must also choose what type of signal to send to the transducers: either a coded signal at 1, 2, 4, 11 or Twin bits, or a specific number of pulses to be sent.

  • Page 235
    April 2009 Setting Up Pulse/Code Parameters (cont.) 2. The next prompt asks if you want to send a specific number of [ENTER] pulses. Press to select or deselect the pulse option. Note: If you do not select this option, pressing the [ ] key returns the meter to the Transmit Code prompt.
  • Page 236: Setting Error Limits

    April 2009 Setting Error Limits The Error Limits option enables you to set limits for an incoming signal. When the signal falls outside the programmed limits, an error indication appears. To enter this option, scroll to the Error [ENTER] Limits entry on the Service Menu and press .

  • Page 237
    April 2009 Setting Error Limits (cont.) d. Press the [ ] arrow key to move to the maximum signal box, and repeat steps a, b and c. e. Press the [ ] arrow key to return to the minimum signal box, and the [ ] key to move to the next limit.
  • Page 238
    April 2009 Setting Error Limits (cont.) Note: For the velocity and acceleration boxes, the F1 softkey toggles between English and metric measurements. Thus, the key will always display the opposite measurement [F1] from that which is currently active. Press to display the measurement in the alternate format.
  • Page 239: The Test Option

    April 2009 The Test Option Within the Service Menu, the Test option includes seven tests to ensure that the PT878GC is performing properly: Test Screen, Test Keys, Watchdog Test, Impulse Response, Wave Snapshot, Simulate and Battery Test.To enter this option, scroll to the Test [ENTER] entry on the Service Menu and press .

  • Page 240: Testing The Screen

    Pressing the key through this sequence should return the PT878GC to Operate Mode. If the test does not proceed according to this sequence, please consult GE Sensing. 9-36 Servicing the PT878GC…

  • Page 241: Testing The Keys

    Pressing the key returns the meter to the Operate Mode. If any key does not appear on the screen or does not darken when pressed, contact GE Sensing. Note: The power key does not appear in this test. Servicing the PT878GC…

  • Page 242: Testing The Watchdog Timer Circuit

    (No) to cancel the test and return to the Menu screen, or [F3] press (Yes) to start the test. The PT878GC should go blank for a few seconds, and then restart. If it does not follow this sequence, consult GE Sensing. 9-38 Servicing the PT878GC…

  • Page 243: Setting Impulse Response

    April 2009 Setting Impulse Response The Impulse Response option enables you to force the meter to transmit in one direction only, without changing the AGC setting. You can then diagnose problems with transducer, pipe, or fluid configurations. To enter the option, scroll down to the Impulse [ENTER] Response entry in the Service Menu and press .

  • Page 244: Taking Wave Samples For Diagnosis

    April 2009 Setting Impulse Response (cont.) 2. The second prompt asks for the percent of gain. a. Press [ENTER] to open the text box. b. Use the numeric keys to enter the desired percentage. c. Press [ENTER] to confirm your entry. You have completed entering data for the Impulse Response option.

  • Page 245: Testing The Battery

    April 2009 Testing the Battery The Battery Test option provides additional battery information for service troubleshooting. To enter the option, scroll down to [ENTER] the Battery Test entry in the Service Menu and press . The screen appears similar to Figure 9-25 below. Service Battery Form Status: Fast Charge Run Time…

  • Page 246: Resetting To Factory Default Parameters

    April 2009 Resetting to Factory Default Parameters For various reasons, you might wish to return the PT878GC to its original settings. The Factory Defaults option enables you to return the meter to its preprogrammed default settings. To enter the option, scroll down to the Factory Defaults entry in the [ENTER] Service Menu and press .

  • Page 247: Updating Pt878Gc Software

    • IrCOMM, an infrared standard supported by certain Microsoft operating systems. GE Sensing recommends updating software via the IrOBEX standard; however, the IrCOMM standard is available for users who have problems with IrOBEX. This section covers procedures for both standards.

  • Page 248: Updating Software Via Irobex

    4. The meter asks for confirmation. Repeat the options shown in Step 1 above. After the PT878GC reboots, the screen appears similar to Figure 9-28 below. GE Sensing PCI Loader v3.0 2/26/04 [HW Rev3+] Backup Battery: OK Firmware CRC: Valid…

  • Page 249
    April 2009 Updating Software Via IrOBEX (cont.) 5. The meter now asks if you wish to reload the flash memory via the infrared communications port. • [F1] Press (Yes) to begin reloading via OBEX, or • [F2] Press (No) to stop the reloading procedure. Be sure the infrared sensor on the meter has a clear path to the infrared adapter on the PC.
  • Page 250: Updating Software Via Ircomm

    April 2009 Updating Software Via IrCOMM Note: While Windows 95/98/98SE and NT (with QuickBeam) support IrCOMM, Windows 2000 and XP do not. Refer to Appendix B. Before you install new coding, you must be sure that the PC has the correct protocols to transmit the software to the meter. Setting up the PC Note: The use of Hyperterminal is shown here as an example.

  • Page 251
    April 2009 Setting up the PC (cont.) Figure 9-29: Port Settings for COMM 4 Installing New Software To begin the loading procedure, scroll down to the Flash Update [ENTER] entry in the Service Menu and press . The screen appears similar to Figure 9-30 below.
  • Page 252
    2. The meter asks for confirmation. Repeat the options shown in Step 1 above. The screen on both the PC and the PT878GC now appears similar to Figure 9-31 below. GE Sensing PCI Loader v3.0 2/26/04 [HW Rev3+] Backup Battery: OK…
  • Page 253
    April 2009 Installing New Software (cont.) 1. In the Hyperterminal window, pull down the Transfer menu and click Send File. A window opens similar to Figure 9-32 below. Figure 9-32: The Send File Window in the Transfer Menu 2. Click on the replacement software (designated by a .cod extension) from the folder where it has been stored.
  • Page 254: Chapter 10: Diagnostics And Troubleshooting

    April 2009 Chapter 10 Diagnostics and Troubleshooting The TransPort PT878GC is a reliable instrument that is easy to maintain. It will provide accurate flow measurement readings as long as it is operated as described in this manual. If problems do…

  • Page 255: Error Code Messages

    Use Table 10-1 on page 10-3 to isolate and remedy the problem. If you are unable to remedy the problem, contact GE Sensing. Note: We suggest that, in order to isolate the indicated problem more easily, you obtain a test flowcell.

  • Page 256
    April 2009 Table 10-1: Error Messages Error Possible Message Problem Cause Action No error. Displays briefly None required. Mea- after the dis- surement is valid. play of another error message Low Signal — Broken cable. Check transducer Poor ultra- Flowcell prob- cable.
  • Page 257
    April 2009 Table 10-1: Error Messages (cont.) Error Possible Message Problem Cause Action Velocity Programming Make sure the maxi- Range — error, poor flow mum velocity range Velocity conditions, or is within the range exceeds pro- bad transducer listed in Table 3-1 on grammed spacing.
  • Page 258
    April 2009 Table 10-1: Error Messages (cont.) Error Possible Message Problem Cause Action Cycle skip, Poor flow con- Check transducer Acceleration ditions or bad spacing as pro- transducer spac- grammed and as set ing. on pipe.See Flowcell Problems on page 10- 9 and Transducer Problems on page 10- Analog Out…
  • Page 259: Displaying Diagnostic Parameters

    April 2009 Displaying Diagnostic Parameters As part of its measurement menu, the PT878GC offers a list of diagnostic parameters to aid in troubleshooting in the event of flowcell, transducer, or electrical problems. You can select any diagnostic parameter for display as a measurement as discussed in Chapter 6 on pages 6-7 and 6-8.

  • Page 260
    April 2009 Table 10-2: Diagnostic Parameters (cont.) Diagnostic Displays Good Parameter Soundspeed Displays the measured sound Check the sound speed (m/s or f/s) speed of the fluid. of fluid in Sound Speeds and Pipe Size Data. P#up Displays signal peaks for the 100- <100, >900 upstream transducer.
  • Page 261
    April 2009 Table 10-2: Diagnostic Parameters (cont.) Diagnostic Displays Good Parameter Qdown Displays the signal quality for ±300 or between the downstream transducer. higher ±100 Norm Factor Displays the normalization 0.85 — <0.85 factor. Threshold Up Displays the Percent of Peak for the upstream transducer.
  • Page 262: Flowcell Problems

    Most gas-related problems can be solved by proper design of the flowmeter installation and the proper choice of transducer frequency. Contact GE Sensing if you cannot solve a gas-related problem. If the physical installation of the system meets the recommended specifications, it is possible that the gas itself may be preventing accurate flow rate measurements.

  • Page 263: Pipe Problems

    4-13 and 4-14. Pipe Problems Improper pipe conditions and/or flowcell installation can cause problems with measurement of the gas flow. Contact GE Sensing if you cannot solve pipe-related problems. By far, the most common pipe problems are the following 1. THE COLLECTION OF MATERIAL AT THE TRANSDUCER LOCATION(S).

  • Page 264: Transducer Problems

    Note: Transducers must be replaced in pairs. Refer to Chapter 4, Programming Site Data, to program the replacement transducer data into the meter. Contact GE Sensing if you cannot solve a transducer-related problem. Diagnostics and Troubleshooting 10-11…

  • Page 265: Specifications

    April 2009 Chapter 11 Specifications This section contains specifications for the following: • Operations and Performance • Electronics • Clamp-on Transducers • Thickness Gauge Option • Additional Options Specifications 11-1…

  • Page 266: Operation And Performance

    April 2009 Operation and Performance Fluid Types Acoustically conductive gases with minimum density requirements. See Table 3-1 on page 3-5, Table 3-2 on page 3-7 and Table 3-3 on page 3-9. Pipe Sizes 0.75 to 12 in. (20 to 300 mm) with PT878GC-01 electronics 4 to 24 in.

  • Page 267: Pressure Requirements, Air, Nitrogen, Oxygen, Argon

    See Table 3-2 on page 3-7. Pressure Requirements, Steam See Table 3-3 on page 3-9. Pressure Requirements, Other Gases Dependent upon gas composition, and pipe size and material. Consult GE Sensing. Measurement Parameters Standard and actual volumetric flow, flow velocity and mass flow. Specifications…

  • Page 268: Electronics

    April 2009 Electronics Flow Measurement Patented Correlation Transit-Time mode Enclosures Submersible IP67 Dimensions Weight: 3 lb (1.36 kg) Size: (h x w x d) 9.4 x 5.5 x 1.5 in. (238 x 138 x 38 mm) Display 240- x 200-pixel backlit LCD graphic display Keypad 25-key rubberized tactile membrane keypad Internal Battery…

  • Page 269: Standard Inputs/Outputs

    April 2009 Standard Inputs/Outputs • One 0/4- to 20-mA current output • One user-selectable pulse (solid state, 5-V maximum) or frequency (5-V square wave; 100 to 10,000 Hz) • Two 4- to 20-mA analog inputs with switchable power supply for loop-powered temperature or pressure transmitters Digital Interfaces Infrared communication port for printer or PC interface Site Parameter Programming…

  • Page 270: Clamp-On Ultrasonic Flow Transducers

    April 2009 Clamp-On Ultrasonic Flow Transducers Temperature Range(s) Standard: –40° to 130°C (–40° to 266°F) Optional (overall range): –40° to 230°C (–40° to 446°F) Transducer Materials Stainless steel and plastic Fixture Materials Anodized aluminum blocks with stainless steel rod, chain or strap Mountings Portable gas-transducer clamping fixtures in polycarbonate carrying cases:…

  • Page 271: Transducer Cables

    April 2009 Transducer Cables • Standard: One pair of LEMO® coaxial transducer connectors with 25-ft (8-m) cables • Optional: 500-ft (152-m) extension cables available for most transducers Specifications 11-7…

  • Page 272: Thickness Gauge Option

    April 2009 Thickness Gauge Option Transducer GE Sensing dual element transducer Pipe Thickness Range 0.05 to 3 in. (1.3 to 76.2 mm) Pipe Materials Most standard metal and plastic pipe materials Accuracy ±1% typical or ±0.002 in. (±0.05 mm) Thermal Exposure Continuous operation to 37°C (100°F);…

  • Page 273
    April 2009 Appendix A Menu Maps Menu Service Program Meter Logging Site See Figure See Figure A-10 Figure Figure A-2 Figures to A-7 Figure A-1: The Main Menu Menu Maps…
  • Page 274: Appendix A: Menu Maps

    April 2009 Menu Site Manager Save Fkeys 1 View 2 Views 3 Views 4 Views Clear Clear Clear Menu Item File Site Sort Transfer Message Print Transfer Name Date as Text Drive Manager File Sort Save Save Refresh Rename Open Delete Name File Name File…

  • Page 275
    April 2009 Menu Program (See Figure A-5) (Clamp-on) Transducer Lining Pipe Fluid Path (See Figure A-4) Fluid (Wetted) (Clamp-on) Wetted Clamp-on Natural Steam Other Path Length Traverses Axial Length Spacing Sound Speed (Preprogrammed Special Units) Frequency Wedge Temperature Time Delay (Wetted) (Clamp-on) Material…
  • Page 276
    April 2009 Menu Program Pipe Material Steel Aluminum Brass Copper Glass Gold Inconel Iron Monel Nickel Plastic Tin Titanium Tungsten Zinc Other Sound Speed (For cast iron, steel, PVC, CPVC) Note: Press [F2] (Cancel) to cancel the entries and return to Menu. ANSI? Press [F3] (OK) to confirm the entries and return to Menu.
  • Page 277
    April 2009 Menu Program (See Figures A-3 and A-4) (See Figure A-6) Standard Volume Mass Flow Inputs Settings Pressure Disabled Enabled Fixed Active Disabled Enabled Base Pressure Pressure Input # Calculation Method Atmospheric Pressure Static Saturated Steam Base Temperature Density Steam Temperature Value…
  • Page 278
    April 2009 Menu Program (See Figure A-5) (See Figure A-7) Analog Output Analog Input Scan/Measure Function Input A Input B 0-20 mA 4-20 mA Correlate Integrate Count (If Standard Min Threshold (If Standard Cycle Number A Divisor % of Peak Volume or Mass Volume or Mass Data Source…
  • Page 279
    April 2009 Menu Program See Figure User Digital Functions Correction Factors Output Function Pulse Frequency Gate Input Test Points Function Calibration Factor Reynolds Correction Totalizer User1 User2 User4 User3 Reynolds Correction Calibration Factor Gate Active Data Source (Pulse Totalizer) (Frequency) Contact Contact Closed…
  • Page 280
    April 2009 Note: Press [F2] (Cancel) to cancel Menu the entries and return to Menu. Press [F3] (OK) to confirm the entries Meter and return to Menu. Battery Snapshot Date/Time Contrast Backlight Units Locale User Tables Totals Date Backlight Reset Time NiCd NiMH…
  • Page 281
    April 2009 Menu Note: Press [F2] (Cancel) to cancel Logging the entries and return to Menu. Press [F3] (OK) to confirm the entries and return to Menu. Log Manager New Log Name Format Sort File View Linear Circular Name Date Type Details Graph Spreadsheet Standard…
  • Page 282
    April 2009 Menu Note: Press [F2] (Cancel) to cancel entries and return to Menu. Press [F3] (OK) to confirm entries and return to Menu. Service T-Gauge T-Gauge Diagnostics Calibrate Factory Flash Reports Error Limits Test Signal Setup Setup Display Defaults Update Min Signal Output…
  • Page 283
    April 2009 Menu Meter Communications Node ID Comm Interface IrDA IR 232 Baud Rate 300 600 1200 2400 4800 9600 19.2K 38.4K 57.6K 115.2K Parity None Mark Space Even Stop Bits Data Bits Note: Press [F2] (Cancel) to cancel the entries and return to Menu. Press [F3] (OK) to confirm the entries and return to Menu.
  • Page 284: Appendix B: Establishing Ir Communications With The Pt878Gc

    IR port (available on most laptops) or an IR dongle (IR to RS232 adapter for PCs without a built-in IR port). If your PC does not have an IR dongle, GE Sensing recommends the ActiSys ACT-IR220L+ infrared to RS- 232 adapter, which has been tested for compatibility with the PT878GC.

  • Page 285: Windows 2000/Xp

    April 2009 Table B-1: IrDA Compatibility of Windows Operating Systems Operating System IrCOMM OBEX Windows XP Windows 2000 Windows NT 4.0 QuickBeam* QuickBeam* Windows Me* Windows 98 SE* Windows 98* Windows 95* Windows 3.1 DOS 6.22 *May require QuickBeam installation under some circumstances when Windows built-in support fails.

  • Page 286: Windows Nt4.0

    April 2009 Windows NT4.0 For IR use, Windows NT4.0 requires installation of QuickBeam software from Extended Systems (www.extendedsystems.com). Desktop PCs and laptops without built-in IR ports also require adding an IR dongle. While laptops with a built-in IR port normally do not require a dongle, its addition might be necessary in some cases.

  • Page 287: Appendix C: Ultrasonic Thickness Gauge Theory Of Operation

    April 2009 Appendix C Ultrasonic Thickness Gauge Theory of Operation All ultrasonic thickness gauging involves timing the round trip of a sound pulse in a test material. Because solid metal has an acoustic impedance that differs from that of gases, liquids, or corrosion products such as scale or rust, the sound pulse will reflect from the far surface of the remaining metal.

  • Page 288: Factors Affecting Performance And Accuracy

    April 2009 Factors Affecting Performance and Accuracy A. Surface Condition — Loose or flaking scale, rust, corrosion or dirt on the outside surface of a test piece will interfere with the coupling of sound energy from the transducer into the test material.

  • Page 289
    April 2009 Factors Affecting Performance and Accuracy (cont.) B. Transducer Positioning/Alignment — For proper sound coupling the transducer must be pressed firmly against the test surface. On small diameter cylindrical surfaces such as pipes, hold the transducer so that the sound barrier material visible on the probe face is aligned perpendicular to the center axis of the pipe.
  • Page 290
    April 2009 Factors Affecting Performance and Accuracy (cont.) Remember that an ultrasonic test measures thickness at only one point within the beam of the transducer, and that in corrosion situations wall thicknesses often vary considerably. Test procedures usually call for making a number of measurements within a defined area and establishing a minimum and/or average thickness.
  • Page 291
    April 2009 Factors Affecting Performance and Accuracy (cont.) D. Taper or Eccentricity — If the contact surface and the back surface are tapered or eccentric with respect to each other, the return echo again becomes distorted and the accuracy of measurement is diminished.
  • Page 292
    A variety of couplants are available for use with GE Sensing transducers. To obtain a material safety data sheet (MSDS) for any couplant GE Sensing supplies, go to the GE Sensing web site and enter the Technical Support portal. Material Safety Data Sheets for Couplants…
  • Page 293: Analog Outputs

    April 2009 Index About Option……… . . 6-21 Acceleration Limit.

  • Page 294
    April 2009 Index (cont.) Calibration Analog Input ……… 9-21 Analog Output .
  • Page 295
    April 2009 Index (cont.) Dampening Material ……..3-25 Data Acquisition .
  • Page 296: File Menu

    April 2009 Index (cont.) Electrical Connections ……… . 2-2 Electrical Connections .

  • Page 297
    April 2009 Index (cont.) Graph Display ……… . . 6-4 Help .
  • Page 298
    April 2009 Index (cont.) Limits Option……….6-5 Line Graph Display .
  • Page 299: Main Menu

    April 2009 Index (cont.) M>S Switch ……….4-25 Main Menu Menu Map.

  • Page 300
    April 2009 Index (cont.) On-Line Help ……… . . 2-14 Getting .
  • Page 301
    April 2009 Index (cont.) Parameters, Displayed ……..6-10 Path for Clamp-on .
  • Page 302
    April 2009 Index (cont.) Range ……….11-2 Rangeability .
  • Page 303
    April 2009 Index (cont.) Sampling Waves……..9-40 Schedule for Pipe Dimensions .
  • Page 304
    April 2009 Index (cont.) Sound Speed, Calculating ……9-14, 9-15 Span (Full Scale) Value for Analog Output .
  • Page 305
    April 2009 Index (cont.) Task Keys ……….2-12 Temperature Operating .
  • Page 306
    April 2009 Index (cont.) Entering Data for ……..4-3 Materials .
  • Page 307
    April 2009 Index (cont.) Watchdog Test ……… 9-38 Wave Sampling .
  • Page 308
    DECLARATION Sensing CONFORMITY Panametrics Limited Shannon Industrial Estate Shannon, County Clare Ireland declare under our sole responsibility that the ™ TransPort PT878 Portable Ultrasonic Flowmeter ™ TransPort PT878GC Clamp-On Portable Ultrasonic Flowmeter to which this declaration relates, are in conformity with the following standards: •…
  • Page 309
    DECLARATION Sensing CONFORMITE Nous, Panametrics Limited Shannon Industrial Estate Shannon, County Clare Ireland déclarons sous notre propre responsabilité que les ™ TransPort PT878 Portable Ultrasonic Flowmeter ™ TransPort PT878GC Clamp-On Portable Ultrasonic Flowmeter rélatif á cette déclaration, sont en conformité avec les documents suivants: •…
  • Page 310
    Sensing KONFORMITÄTS- ERKLÄRUNG Wir, Panametrics Limited Shannon Industrial Estate Shannon, County Clare Ireland erklären, in alleiniger Verantwortung, daß die Produkte ™ TransPort PT878 Portable Ultrasonic Flowmeter ™ TransPort PT878GC Clamp-On Portable Ultrasonic Flowmeter folgende Normen erfüllen: • EN 61326:1998, Class A, Annex C, Continuous Unmonitored Operation gemäß…
  • Page 311
    1100 Technology Park Drive Billerica, MA 01821-4111 Web: www.gesensing.com Ireland Sensing House Shannon Free Zone East Shannon, County Clare Ireland…

Расходомер с накладными датчиками

РТ878 — это времяимпульсный ультразвуковой расходомер, который объединяет в себе характеристики стационарного расходомера с преимуществами портативного прибора. Расходомер измеряет акустически проводящие однофазные жидкости, в которых возможно наличие второй фазы. К ним относятся большинство чистых жидкостей, сточные воды, некоторые суспензии и водонефтяные смеси, жидкости с небольшим содержанием растворенных газовых пузырьков. Прибор имеет два входа для подключения температурных датчиков с целью измерения энергии теплового потока, переносимого жидкостями. Энергия теплового потока может быть вычислена для воды, гликоля и водно-гликолевой смеси.

Области применения

Расходомер TransPort PT878 – это полностью укомплектованная портативная ультразвуковая система измерения расхода жидкостей:

  • Питьевая вода.
  • Сточные воды.
  • Вода систем охлаждения и нагрева.
  • Сверхчистая вода и жидкости.
  • Растворы вода/гликоль.
  • Сырая нефть.
  • Очищенные углеводороды.
  • Дизельное топливо.
  • Смазочные масла.
  • Химикаты.
  • Напитки.
  • Другие жидкости.

Основные характеристики

  • Легкий, небольшого размера и простой в использовании расходомер.
  • Бесконтактное измерение расхода с помощью накладных ультразвуковых (УЗ) преобразователей
  • Определение скорости, объемного расхода и энергии.
  • Данные по суммарному расходу и трендам.
  • Большой ЖК-дисплей с подсветкой.
  • Алфавитно-цифровой и графический форматы представления данных.
  • Многоязычный интерфейс пользователя.
  • Блок подзаряжаемых батарей.
  • Запись до 100000 точек данных по расходу.
  • Исполнение, допускающее погружение в воду.
  • Сохранение данных о 32 объектах измерения.
  • Дополнительно – функция измерения толщины стенки трубы.
  • Дополнительно – функция определения энергии теплового потока.
  • Расходомер применим для большинства размеров и материалов труб, включая трубы с покрытием.

Универсальный портативный расходомер

Расходомер TransPort PT878 – это универсальная, автономная, портативная система, основанная на времяимпульсном методе измерения и имеющая все необходимые опции и дополнительные принадлежности для измерения расхода жидкостей. Небольшие раз-меры и вес, встроенный блок аккумуляторов и универсальное зарядное устройство обеспечивают высокую гибкость применения этого расходомера, практически, в любых задачах измерения расхода жидких сред.

Высокая точность измерения расхода двухфазных и чистых жидкостей

Запатентованная корреляционная времяимпульсная технология измерения расхода Correlation Transit-Time™ и цифровая обработка сигналов, реализованные в TransPort PT878, значительно увеличивают отношение сигнал-шум, обеспечивая точное, без дрейфа нуля измерение расхода жидкостей, которые содержат вторую фазу – захваченные потоком частицы или газовые пузырьки. Расходомер TransPort надежно работает в этих и других сложных условиях, где традиционные расходомеры не действуют. Расходомер TransPort PT878 также позволяет точно измерять расход в идеально чистых жидкостях, не содержащих «отражатели», где Допплеровские расходомеры работать не могут. Расходомер TransPort пригоден для использования, как в случае обычных задач применения времяимпульсного метода, так и в большинстве других задач, где традиционные время-импульсные расходомеры «встречаются» со значительными трудностями.

Простота и легкость использования

Возможность выполнения первых измерений расхода в пределах нескольких минут после открытия упаковки говорит о легкости использования расходомера TransPort. Просто введите параметры объекта, прикрепите УЗ преобразователи на трубу, установите расстояние между ними, и «процесс пошел». При этом нет необходимости в дополнительном оборудовании, а также во врезке в трубопровод. Опытный пользователь может выполнить множество различных измерений за один рабочий день. Расходомер TransPort PT878 – это идеальное средство для всех видов обследований, связанных с измерением расхода.

PC совместимый программный интерфейс

Обмен информацией Transport РТ878 с РС осуществляется с помощью USB-интерфейса и операционной системы Windows. Обращайтесь к руководству по эксплуатации, где приведена подробная информация о введении данных об объектах, выполнении записей и других операциях с PC.

УЗ преобразователи и зажимные приспособления

С помощью накладных УЗ преобразователей расходомер TransPort PT878 позволяет измерять расход через металлические, пластиковые или даже покрытые цементом трубы без врезки в стенку трубы. Расходомер TransPort PT878 обеспечивает стабильную точность измерения расхода без загрязнения и утечек измеряемой среды. Расходомер TransPort PT878 не имеет движущихся деталей, которые подвержены износу, или сужений, приводящих к засорению трубы, и не требует регулярного технического обслуживания. Для удовлетворения жестким требованиям промышленного применения доступен широкий спектр УЗ преобразователей с различными рабочими частота-ми, материалами конструкции, рабочими температурами и размерами. Для обеспечения контакта накладных УЗ преобразователей с трубой доступны различные зажимные приспособления для разных труб и размеров УЗ преобразователей. Эти зажимные приспособления реализуют различные способы крепления к трубе, включая цепи, металлические ленты, ленты Velcro® и магнитные зажимы.

Алфавитно-цифровой и графический ЖК-дисплей завершает картину

Большой многофункциональный ЖК-дисплей позволяет представлять измеренные данные, как в алфавитно-цифровом, так и в графическом формате. Кроме того, он помогает легко выполнять программирование, представляя меню программы, которое обеспечивает ввод данных и выбор функций. Стандартные функции представления данных в буквенно-цифровом формате включают в себя отображение скорости потока, объемного расхода и энергии теплового потока, а также суммарного количества в английских или метрических единицах измерения. В графическом режиме ЖК-дисплей показывает, как данные в реальном масштабе времени, так и записанные данные. При этом справа на дисплее воспроизводится график, который полезен для просмотра данных и наблюдения за трендами на объекте.

Прочный корпус, допускающий погружение в воду

Ваши инвестиции в этот расходомер надежно защищены от ежедневной тяжелой его эксплуатации в жестких промышленных условиях. Расходомер TransPort PT878 имеет резиновый чехол, который обеспечивает защиту от вибраций и ударов. Полностью герметизированный корпус и порты соответствуют требованиям IP67, что позволяет погружать прибор в воду на глубину до 1 м в течение ограниченного периода времени. Он будет продолжать безопасно функционировать, даже если упадет в воду. Полностью укомплектованная система измерения расхода TransPort PT878 размещается в компактном кейсе для переноски.

Дополнительная возможность измерения энергии теплового потока

Расходомер TransPort PT878 соединяет в себе возможности измерения расхода ультразвуковым методом и точные измерения температуры термометром сопротивления для определения энергии теплового потока в системах нагрева и охлаждения жидкостью. При наличии этой опции, расходомер TransPort поставляется с встроенными датчиками температуры (термометры сопротивления) с питанием по токовой петле, а также со всеми схемами и программным обеспечением для выполнения измерения энергии. В комплект опции измерения энергии входят два равноценных платиновых термометра сопротивления градуировки Pt1000 для поверхностного монтажа и кабель длиной 9 м для подключения к расходомеру PT878.

Дополнительный УЗ преобразователь для измерения толщины стенки трубы

Толщина стенки трубы является очень важным параметром, используемым в расходомере TransPort при измерении расхода накладными УЗ преобразователями. Эта опция позволяет точно измерить толщину стенки снаружи трубы.

USB-порт

Расходомер TransPort PT878 имеет USB-порт для связи с PC.

Экономически эффективное решение

Для того чтобы быть действительно полезным, портативный расходомер должен быть экономически выгодным для владельца и работать в любых полевых условиях. Расходомер TransPort PT878 разработан так, чтобы оставаться в эксплуатации в течение многих лет. Это устройство, полностью изготовленное из твердотельных элементов, практически не изнашивается или не нуждается в сервисе, обеспечивая минимальное время простоя и низкие затраты на техническое обслуживание.

Расходомер TransPort PT878 реализует времяимпульсный метод измерения расхода

При использовании этого метода, два УЗ преобразователя служат по очереди как генераторы и приемники УЗ сигналов. После установки их на трубе, они акустически связаны друг с другом, т.е. один из УЗ преобразователей может принимать УЗ сигналы, переданные другим УЗ преобразователем, и наоборот. Во время работы каждый из УЗ преобразователей функционирует как генератор, генерируя определенное число акустических импульсов, а затем как приемник такого же числа импульсов. Интервал времени между подачей и приемом УЗ сигналов измеряется в обоих направлениях. Когда жидкость в трубе неподвижна, время распространения этих сигналов по потоку равно времени распространения сигналов против потока. Когда жидкость течет по трубе, время распространения сигналов по потоку меньше времени распространения сигналов против потока. Разность этих времен распространения пропорциональна скорости потока жидкости, а ее знак указывает направление потока.

  • Page 1
    Flow TransPort™ PT878GC Portable Gas Flowmeter Instruction Manual 910-229 B3 panametrics.com August 2021…
  • Page 3
    User’s Manual 910-229 B3 August 2021 panametrics.com Copyright 2021 Baker Hughes company. This material contains one or more registered trademarks of Baker Hughes Company and its subsidiaries in one or more countries. All third-party product and company names are trademarks of their respective…
  • Page 4
    [no content intended for this page]…
  • Page 5: Table Of Contents

    Contents Features and Capabilities ………………1 1.1.1 Overview .

  • Page 6
    Contents 4.1.9.1 Entering General-Purpose Analog Inputs …………61 4.1.9.2 Entering Analog Inputs in Standard Volume or Mass Flow Option .
  • Page 7
    Contents Logging Data ………………..121 8.1.1 Entering the Logging Menu .
  • Page 8
    Contents 9.1.10.7 Testing the Battery …………….167 9.1.11 Resetting to Factory Default Parameters .
  • Page 9
    Contents 11.1.4.3 Pipe Materials ………………185 11.1.4.4 Accuracy.
  • Page 10
    Contents viii TransPort™ PT878GC User’s Manual…
  • Page 11: Features And Capabilities

    Chapter 1. Chapter 1. Features and Capabilities ® The TransPort Model PT878GC is a transit-time clamp-on gas flowmeter which combines all the features of a full-size gas flowmeter with the advantages of a portable instrument. This section describes the TransPort features and general system, and explains the theory of operation. 1.1.1 Overview The PT878GC is available in two models.

  • Page 12: System Description

    Chapter 1. 1.1.2 System Description The TransPort is only one part of the flowmeter system. The other part of the system is the flowcell, which consists of the pipe and the transducers. 1.1.2.1 The Flowcell The flowcell is that part of the system that uses ultrasonic pulses to interrogate the flow. It consists of the flowcell pipe and the transducers.

  • Page 13: Electronics Package

    Chapter 1. 1.1.2.2 Electronics Package The TransPort consists of circuits that generate, receive, and measure the travel time of the ultrasonic pulses. It also contains a microcomputer that controls operation and calculates flow measurement parameters. Specific circuits function as follows: TRANSMIT SIGNAL GENERATOR — The transmit signal generator, under control of the microcomputer and timing circuit, synthesizes the signal that drives the transmitter.

  • Page 14: Theory Of Operation

    Chapter 1. 1.1.3 Theory of Operation The TransPort is a transit-time ultrasonic flowmeter. When ultrasonic pulses are transmitted through a moving liquid or gas, the pulses that travel in the same direction as the fluid flow (downstream) travel slightly faster than the pulses that travel against the flow (upstream).

  • Page 15: Initial Setup

    Chapter 2. Chapter 2. Initial Setup Before making measurements, you must prepare the TransPort for operation. This includes the following procedures: • Making Electrical Connections • Charging and/or Replacing Batteries • Powering On and Off • Using the Screen and Keypad •…

  • Page 16
    Chapter 2. Figure 3: Connection Locations TransPort™ PT878GC User’s Manual…
  • Page 17: Power Connections

    Chapter 2. 2.1.1.1 Power Connections The PT878GC is powered by either a 100-120/200-260 VAC wall mount plug-in module, or by 5 internal C -size NiCad high-energy rechargeable batteries or by a pack of 3.0 Ahr NiMH batteries. (An optional power supplement, part #703-1283, uses 6 AA alkaline batteries.) When you receive the PT878GC, the batteries are not charged;…

  • Page 18: Charging And/Or Replacing Batteries

    -size NiCad high energy rechargeable batteries (Panametrics Sensing Part Number 200-058) or 3.0 Ahr NiMH batteries (part number 200-081). To replace the batteries, remove the rubber boot, open the panel located on the back of the PT878GC unit, disconnect the batteries, and replace with new ones (see Figure 4 below).

  • Page 19: Powering On And Off

    PT878GC emits a short beep and displays a “PCI Loader” message. It then validates the instrument programming, and then displays the Panametrics Sensing logo and the software version and emits a long beep. If the meter fails any of these tests, contact Panametrics Sensing.

  • Page 20
    Chapter 2. ABC.SIT 2004/11/30 09:53 AM Velocity, ft/s Signal, dB 0.00 Volume, l/s Delta-T, ns 0.10 E0: No Errors Figure 5: Screen After Powering On To turn the PT878GC off, press the red key for 3 seconds. The screen now appears similar to Figure 6 below. Velocity, ft/s Signal, dB SHUTDOWN: Meter OFF…
  • Page 21: Using The Screen And Keypad

    Chapter 2. 2.1.4 Using the Screen and Keypad The essential features for operating the TransPort are the screen and keypad. Although these features are common on portable instruments, the PT878GC design offers unique features to simplify and speed operation. 2.1.4.1 Screen The primary function of the screen is to display information in order for you to accurately and easily take measurements.

  • Page 22
    Chapter 2. Table 3: Icons in the System Tray Icon Function Meaning Stopwatch Calibration Gate Operation: Watch is stopped when the gate is closed, or runs when it is open. (See page 3-46.) Snapshot Indicates that the Snapshot (To file) function has been activated, so users can take screen (To Printer)
  • Page 23: Keypad

    [ESC] — enables you to exit menus or menu options at any time; cancels numeric entry. • Red key [ ] turns the power on or off, and toggles the backlight on or off. Figure 8: The TransPort PT878GC Keypad TransPort™ PT878GC User’s Manual…

  • Page 24: Obtaining On-Line Help

    Chapter 2. 2.1.5 Obtaining On-Line Help The TransPort offers on-line help screens that describe various features. You can access on-line help at any time by pressing the [?] key. The screen appears similar to Figure 9 below. Help Velocity, ft/s Signal, dB Table of Contents Use the arrows and the enter key to…

  • Page 25: Installing The Dampening Material, Transducers And Fixtures

    Since the PT878GC is specifically designed for gas measurement with clamp-on transducers, it requires the use of specially designed fixtures and dampening material to maintain the highest possible measurement accuracy. Panametrics Sensing supplies the CFG series of fixtures: • The V1 clamping fixture for pipes with diameters between 0.75 and 1.25 in.

  • Page 26
    Chapter 3. CFG-V1 CFG-V8 Layout Wrap PI Fixture Dummy Block Chain Strap Figure 10: The V Series and PI Fixtures Choosing the transducer installation locations on the pipe is an important aspect of proper flow measurement. Specifications assume a fully developed flow profile typically requiring 20 diameters upstream and 10 diameters downstream of straight pipe run.
  • Page 27: Application Requirements

    Chapter 3. 3.1.1 Application Requirements Before you begin to use the PT878GC, you should ensure that your meter can handle the frequencies required for your particular application. To check the capabilities of your particular PT878GC, refer to the back label of the PT878GC, shown in Figure 11 below.

  • Page 28
    Chapter 3. Table 4: PT878GC Installation Requirements for Air, Nitrogen, Oxygen or Argon Maximum Velocity, ft/s (m/s) Min. Single Dual Four Pipe Size ANSI Pipe Wall Inches Transducer Pressure Travers Travers Triple Travers Five (DIN) (mm) psig (bar) Traverse Traverse 3/4 (20) <0.07 (1.8) 60 (5.1)
  • Page 29
    Chapter 3. Table 4: PT878GC Installation Requirements for Air, Nitrogen, Oxygen or Argon Maximum Velocity, ft/s (m/s) Min. Single Dual Four Pipe Size ANSI Pipe Wall Inches Transducer Pressure Travers Travers Triple Travers Five (DIN) (mm) psig (bar) Traverse Traverse 10 (250) <0.37 (9.4) 60 (5.1)
  • Page 30
    Chapter 3. Table 5: PT878GC Installation Requirements for Natural Gas Maximum Velocity, ft/s (m/s) Min. Pipe Size ANSI Pipe Wall Transduce Pressure Dual Triple Four (DIN) Inches (mm) r MHz psig (bar) Single Traverse Traverse Traverse Traverse 2 (50) <0.16 (4.1) 200 (14.8) 110 (33.5) 88 (26.8)
  • Page 31
    Chapter 3. Table 6: PT878GC Installation Requirements for Steam Maximum Flow Velocity Pipe Wall Min. ft/s (m/s) Pipe Size Inches Transducer Pressure Single in. (mm) (mm) (MHz) psig (bar) Traverse 3 (80) <0.22 (5.6) 110 (8.6) 120 (36.6) <0.3 (7.6) 200 (14.8) 4 (100) <0.24 (6.1)
  • Page 32: Preparing The Pipe

    Chapter 3. 3.1.2 Preparing the Pipe Locate a transducer measurement point with 20 diameters of upstream straight run pipe and 10 diameters of downstream straight run pipe. In addition, the point should be at least 10 diameters from any butt welds or flanges.

  • Page 33: Performing A Pipe Survey

    Chapter 3. 3.1.2.1 Performing a Pipe Survey Finding a location where the pipe is concentric is important for optimum accuracy and performance. If possible, perform a pipe survey with an ultrasonic thickness gauge to find the best location. Clear rust or loose paint and measure the wall thickness at five points along the pipe axis at 25 mm (1-in.) intervals using an ultrasonic thickness gauge, as shown in Figure 13 below.

  • Page 34: Obtaining The Transducer Spacing

    Chapter 3. 3.1.2.2 Obtaining the Transducer Spacing Using the measured OD and the pipe wall thickness, program the PT878GC (discussed in Chapter 4, Programming Site Data) to determine the required transducer spacing. To determine the PT878GC correction factor, calculate the mean inside pipe diameter (ID) and the pipe ID at the transducer locations.

  • Page 35: Installing The V Series Clamping Fixture And Transducers

    Note: If you are using a V4 fixture with a 1.5 to 2-in. pipe, Panametrics Sensing suggests removing the set screws and the mounting bracket, rotating them and the transducer yoke 180°, and replacing the bracket, yoke and screws for a more secure fit.

  • Page 36
    Chapter 3. The two fixture halves have measuring scales; ensure that the scales are on the same side of the fixture, so that both zeros start at the same origin, as shown in Figure 17 below. Figure 17: Fixture with Scale Origins Properly Aligned Install the four nuts onto the threaded rods with the convex side of the nut facing the fixture.
  • Page 37: Installing The Transducers

    Chapter 3. 3.1.3.2 Installing the Transducers Apply a bead of couplant 6 mm (0.25 in.) wide along the entire length of each transducer face, as shown in Figure 19 below. Figure 19: Couplant on Transducer Face Note: Do not slide the transducer with couplant along the surface of the pipe when mounting the transducer. Set the first mounting block (either left edge or right edge) at a convenient number on the scale, such as 1 in.

  • Page 38: Installing The Pi Fixture And Transducers

    The PI clamping fixture holds transducers on pipes from 8 to 24 in. in diameter. It comes with either a chain or strap, depending on the selection made with the initial order from Panametrics Sensing. To install the fixture and transducers, complete the following steps: 3.1.4.1…

  • Page 39
    Chapter 3. Figure 23: Marking the 3 o’Clock Position To find the coinciding point on the opposite site of the pipe (180° away from each other), divide the measured circumference by 2 and measure this distance along the circumferential lines from the zero point, as shown in Figure 24 below.
  • Page 40: Installing The First Bracket With A Chain Or Strap

    Chapter 3. 3.1.4.2 Installing the First Bracket with a Chain or Strap The following steps describe how to install the PI fixture with a supplied chain or strap. Carefully wrap the chain or strap around the pipe, taking care not to twist it. Loosen the wing nuts up to the end of the J-hooks.

  • Page 41: Installing The Transducers

    Chapter 3. Line up the other edge of the CFG-PI holder bracket with the scribe line and tighten the chain or strap, as shown in Figure 29 below. Mark on Opposite Scribe Line Rear Edge of Second Block Figure 29: Lining up the Rear Edge of the Bracket with the 180° Scribe Line The spacing should now appear similar to that shown in Figure 30 below.

  • Page 42
    Chapter 3. With one hand, mount one transducer into the PI fixture. With the other hand, tighten the transducer hold down screw, gradually pushing the transducer down to the pipe surface. Use a wrench to tighten the backing nut to prevent loosening due to vibration and thermal expansion.
  • Page 43: Installing Dampening Material

    • The DMP-3 is a clay-like compound for all temperature applications. If the temperature is over 150°F, the PDJ pipe dampening jacket (available from Panametrics Sensing with preapplied DMP-3) must be used with the material. At a minimum, you should consider applying dampening material if you have any of the following conditions: •…

  • Page 44: Installing Dmp-1 Dampening Material

    Chapter 3. 3.1.5.1 Installing DMP-1 Dampening Material Be sure the CFG-V clamping fixture is installed on the pipe with the transducers as described in the section Installing the V Series Clamping Fixture and Transducers on page 25. With a marker, mark scribe lines on the inside edges of the brackets onto the pipe.

  • Page 45
    Chapter 3. Figure 34: DMP-1 Material with Transducer Footprint Cut Out To remove any residual adhesive, clean the cut-out area with a rag or a piece of sandpaper, or use varnish remover. 10. Reinstall the fixture and transducers on the pipe. Lay the second strip of dampening material flat.
  • Page 46: Installing Dmp-1 Dampening Material With Pi Fixture

    Chapter 3. 3.1.5.2 Installing DMP-1 Dampening Material with PI Fixture Be sure the PI clamping fixture is installed onto the pipe with the transducers as described in the section Installing the PI Fixture and Transducers on page 28. Approximate the axial distance from one transducer face to the other transducer face, as shown in Figure 36 below.

  • Page 47: Installing Dmp-3 Compound With All Fixtures

    Chapter 3. 3.1.5.3 Installing DMP-3 Compound with All Fixtures Be sure the fixture and transducers are installed on the pipe as described in the previous sections of this chapter. Remove the fixture and transducers, but be sure to mark the approximate area of installation. Remove any loose paint or rust with a file or emery cloth, as shown in Figure 38a below.

  • Page 48: Installing The Pdj Dampening Jacket

    Chapter 3. Figure 41: Clearing DMP-3 from the Transducer Location (a) and Installing the Transducers (b) Apply the couplant to the transducers, and install the transducers onto the pipe. Note: If the measurement point is near a flange or weld, apply DMP-3 between that structure and the fixture as well.

  • Page 49
    Chapter 3. WARNING! The pipe and the dripping fluid will cause severe burns upon contact with bare skin. Also, be sure not to inhale the fumes generated during the DMP-3 curing cycle. Install the fixture over the jacket, adjusting the spacing to match the prestamped transducer holes and PT878GC spacing calculations.
  • Page 50
    Chapter 3. TransPort™ PT878GC User’s Manual…
  • Page 51: Programming Site Data

    Chapter 4. Chapter 4. Programming Site Data On the PT878GC, a Program Menu (part of the Main Menu) enables you to enter information that is specific to each site: • Transducer types and paths • Pipe materials and linings • Fluid types •…

  • Page 52: Entering Transducer Parameters

    Press [ENTER] or [SEL] to confirm the choice. IMPORTANT: While wetted transducers are listed as a choice, the PT878GC is designed for use with Shear wave clamp-on transducers. Consult Panametrics before selecting wetted transducers. Note: The choices made earlier in the Transducer and Pipe options determine the prompts available later. If the PT878GC does not scroll to a particular parameter, it is not necessary for that transducer or pipe type.

  • Page 53: Parameters For Special Transducers

    From the Wedge Temp prompt, press the [ ] key to reach the Wedge SS prompt, and press Use the numeric keys to enter the Panametrics Sensing-supplied wedge sound speed (in ft/sec or m/sec) and press [ENTER]. Pressing the [] key returns the meter to the Transducer tab at the top.

  • Page 54: Entering Pipe Parameters

    Chapter 4. 4.1.3 Entering Pipe Parameters To enter the Pipe option, scroll to the Pipe entry on the Program Menu and press [ENTER]. (From the Transducer  window, you can scroll back up to the Transducer tab and press the [ ] arrow key to reach the Pipe window and …

  • Page 55
    Chapter 4. The first prompt asks you to select the pipe material. [ENTER] Press to enter the material prompt. A drop-down list of materials opens. Table 7 below lists the available preprogrammed materials on the list.   Press the [ ] or [ ] keys to scroll to the appropriate material, or scroll to “Other”…
  • Page 56
    Chapter 4. Note: The measurement units shown depend on the choices you have made in the English/Metric window of the Meter menu. If you select a material that has the Schedule option: [ENTER] The prompt asks if you wish to apply DIN (the DIN schedule). Press to select (or deselect) the DIN box.
  • Page 57: Entering Pipe Lining Parameters

    Press [ENTER] to confirm your choice. Note: If your pipe lining is not on the drop-down list, consult Panametrics Sensing for further information. The program follows one of two paths: If you have selected a preprogrammed material, the PT878GC automatically supplies the correct sound speed, and you proceed to step 3.

  • Page 58: Entering Fluid Types And Speeds

    Chapter 4. 4.1.5 Entering Fluid Types and Speeds To access the Fluid option, scroll to the Fluid entry on the Program Menu and press [ENTER]. (If you are already in the Transducer/Pipe form, press the [] arrow key to reach the Fluid window, and press [ENTER].) The screen appears similar to Figure 48 below.

  • Page 59: Entering The Signal Path Parameters

    Chapter 4. 4.1.6 Entering the Signal Path Parameters To enter the Path option, scroll to the Path entry on the Program Menu and press [ENTER]. (From the Lining window, you can scroll back up to the Lining tab and press the [] arrow key to reach the Path window, and press [ENTER].) The screen appears similar to Figure 49 below.

  • Page 60: Path Parameters For Clamp-On Transducers

    [ENTER] to confirm the entry. If you enter an invalid entry, the PT878GC rejects the entry and displays an error message. Note: Panametrics Sensing does not recommend using a spacing other than the one calculated by the PT878GC.  After you enter the spacing, pressing the [ ] key returns the prompt to the Traverses box, and then to the Path tab at the top of the screen.

  • Page 61
    Chapter 4. The first prompt asks if you want to disable or enable the Standard Volume option.   Use the [ ] and [ ] keys to scroll to the appropriate radio button. Press [ENTER] to confirm your selection. Note: If you select “Disabled,”…
  • Page 62: Entering Inputs In The Standard Volume Option

    Chapter 4. 4.1.7.1 Entering Inputs in the Standard Volume Option To enter input parameters in the Standard Volume/Mass Flow option, return to the Settings tab at the top of the Standard Volume window. Press the [] arrow key and [ENTER] to open the Inputs window. The screen appears similar to Figure 51 below.

  • Page 63
    Chapter 4. The next prompt appears if you have selected a fixed pressure. The PT878GC asks for the fixed pressure value. [ENTER] Press to open the text box. Use the numeric keys to enter the known pressure. Press [ENTER] to confirm the entry. Whether you have selected a fixed or an active supply, the PT878GC asks for the desired input.
  • Page 64: Entering Mass Flow Parameters

    Chapter 4. 4.1.7.2 Entering Mass Flow Parameters To enter mass flow parameters in the Standard Volume/Mass Flow option, press the [] arrow key until you reach the Mass Flow tab, and then press [ENTER] to open the Mass Flow window. The screen appears similar to Figure 52 below. To step through each parameter, press the [] key.

  • Page 65: Entering Skan/Measure Parameters

    M>S Switch value. The default value is 5.0 µsec and values from 0 to 250 µsec are acceptable. Note: Do not change this value without consulting Panametrics Sensing. If you wish to use Skan only, set the M/S Switch to a value of 0.

  • Page 66
    Chapter 4. The program now asks if you wish to Limit the Measurement Delta. (The default is unchecked or off.) [ENTER] Press to enter the text box. Use the numeric keys to enter the recommended value in µsec. Press [ENTER] to confirm the entry. Pressing the [] key highlights the Advanced prompt.
  • Page 67: Entering Skan/Measure Integrate Parameters

    Chapter 4. 4.1.8.1 Entering Skan/Measure Integrate Parameters To enter parameters for the Skan/Measure Integrate mode in the Skan/Measure option, press the [] arrow key once to reach the Integrate tab. Then press [ENTER] to open the Integrate window, which appears similar to Figure 55 below.

  • Page 68: Entering Skan/Measure Count Parameters

    Chapter 4. 4.1.8.2 Entering Skan/Measure Count Parameters To open the Skan/Measure Count window in the Skan/Measure option, press the [] arrow key twice (from the All tab) to reach the Count tab. Press [ENTER], and the screen appears similar to Figure 56 below. Skan/Measure Integrate Count…

  • Page 69: Entering Skan/Correlation Parameters

    Chapter 4. 4.1.8.3 Entering Skan/Correlation Parameters To open the Skan/Correlation window in the Skan/Measure option, press the [] arrow key three times (from the All tab) to reach the Correlate tab, and press [ENTER]. The screen appears similar to Figure 57 below. Skan/Measure Integrate Count…

  • Page 70
    Chapter 4. Repeat step 1 to enter the maximum percentage of peak. The next prompt asks for the Percent of Peak, which is used to calculate the transit times and Delta-T. [ENTER] Press to enter the text box. Use the numeric keys to enter the desired value. Press [ENTER] to confirm the entry.
  • Page 71: Entering Analog Inputs

    Chapter 4. 4.1.9 Entering Analog Inputs The Analog Inputs option enables you to specify parameters for general purpose or standard volume inputs. To enter the Analog Inputs option, scroll to the Analog Inputs entry on the Program Menu and press [ENTER]. If you have not enabled the Standard Volume Option (see page 4-18), the screen appears similar to Figure 58 below.

  • Page 72
    Chapter 4. Use the four arrow keys to scroll to the desired letter or symbol, and press [ENTER] to add the letter to the label. Note: Pressing [SEL] causes the screen to alternate between a set of upper-case (capital) letters, a set of lower-case letters, and a set of symbols.
  • Page 73: Entering Analog Inputs In Standard Volume

    Chapter 4. 4.1.9.2 Entering Analog Inputs in Standard Volume or Mass Flow Option If you have enabled the Standard Volume or Mass Flow options, the screen displays different options, as shown in Figure 60 below. Analog Input Input A Input B Pressure Function Label…

  • Page 74: Entering The Analog Output

    Chapter 4. 4.1.10 Entering the Analog Output The Analog Output option enables you to enter information to set up output parameters. To enter the option, scroll to the Analog Output entry on the Program Menu and press [ENTER]. The screen appears similar to Figure 61 below. To step through each parameter, press the [] key.

  • Page 75
    Chapter 4. [ENTER] Press to open the window. Scroll to the desired output type. Press [SEL] to confirm your selection. The prompt then moves to a list of unit types. (The available units depend on the selection made at the Data Source prompt.) Scroll to the desired output unit.
  • Page 76: Entering The Digital Output

    Chapter 4. 4.1.11 Entering the Digital Output While resembling the Analog Output option, the Digital Output option enables you to set up parameters necessary for a digital output. To enter the option, scroll to the Digital Output entry on the Program Menu and press [ENTER]. The screen appears similar to Figure 63 below.

  • Page 77
    Chapter 4. • Temperature • Volume • User Function • Skan/Measure • Mass Flow [ENTER] Press to open the data source selection window, shown in Figure 62. Scroll to the desired source. Press [SEL] to confirm your selection. The prompt then moves to a list of unit types. (The available units depend on the selection made at the Data Source prompt.) Scroll to the desired output unit.
  • Page 78: Entering User Functions

    Chapter 4. [ENTER] Press to open the drop-down menu. Scroll to the required mode. Press [ENTER] to confirm your selection. Stopwatch Totalizer Through the Gate Input option, you can implement the Stopwatch Totalizer function to measure totals manually. To set up the Totalizer function: At the Gate Active prompt, select Contact Closed.

  • Page 79
    Chapter 4. “ ‘ < > Cancel Delete Figure 65: The Text Creation Window Use the four arrow keys to scroll to the desired letter or symbol, and press [ENTER] to add the letter to the label. Note: Pressing [SEL] causes the screen to alternate between a set of upper-case (capital) letters, a set of lower-case letters, and a set of symbols.
  • Page 80: Entering Correction Factors

    When you have completed entering the function, press [F3] (OK) to confirm the entry and return to the User Function window. Panametrics Sensing recommends pressing [F2] (Check) to test the validity of the function. The PT878GC displays either “OK” or a message such as “Syntax Error.”…

  • Page 81: Entering Reynolds Correction

    Chapter 4. Correction Factors Calibration Factor Reynolds Correction Inputs Re Correction K/V Factors Single Table K/V x 10 Data Source Edit Table Cancel Figure 67: Reynolds Correction Window 4.1.13.1 Entering Reynolds Correction The first prompt asks if you want to enable the Reynolds Correction factor, a number based on the Kinematic Viscosity and flow rate of the gas.

  • Page 82: Entering A Calibration Factor

    Chapter 4. The prompt then asks if you wish to edit the Reynolds Correction table. [ENTER] Press to open the table, which appears similar to Figure 68 below. Edit Reynolds Table Inputs Data Source Cancel Figure 68: Window for Reynolds Correction Table Use the numeric keys to enter the first value for the data source, and press [ENTER] to confirm the entry.

  • Page 83
    Chapter 4. Correction Factors Reynolds Correction Calibration Factor Inputs Calibration Factor K Factor Single Table Meter K-Factor 1.0000 Data Source Edit Table Edit Table Cancel Figure 69: The Calibration Factor Window The first prompt asks if you wish to enable the calibration factor. …
  • Page 84
    Chapter 4. Edit KFactor Table Inputs KFactor Data Source Cancel Figure 70: The KFactor Table Use the numeric keys to enter the first value for the data source, and press [ENTER] to confirm the entry.  Press the [ ] key to move to the KFactor column. Use the numeric keys to enter the corresponding value, and press [ENTER] to confirm the entry.
  • Page 85: Creating And Managing Sites

    Chapter 5. Chapter 5. Creating and Managing Sites The PT878GC can store site data in files for current and future access. (To learn how to program setup data, refer to Chapter 4, Programming Site Data.) After you answer the necessary questions, simply save the information to a site file.

  • Page 86
    Chapter 5. Site Manager File Site Sort Info: DEFAULT:SIT DEFAULT New Site **CURRENT SITE** 03/04/04 14:51:10 5111 bytes 118272 bytes free Refresh Exit Figure 72: The Site Manager Window Note: Each PT878GC comes preprogrammed with a basic site, Default, which serves as a basis for saving data and creating other sites.
  • Page 87: Creating A New Site

    Chapter 5. • To open an existing site (thus replacing the current site), go to page 79. • To save a current site, go to page 79. • To refresh site information, go to page 81. • To rename a site, go to page 81. •…

  • Page 88
    Chapter 5. New Site New Site Use the selected site ‘DEFAULT.SIT’ as a template? Figure 75: Template Confirmation Window The meter returns to Operate Mode, with the new site name displayed in the upper left corner of the screen. TransPort™ PT878GC User’s Manual…
  • Page 89: Opening An Existing Site

    Chapter 5. 5.1.1.2 Opening an Existing Site If you want to return to a second site that you have previously saved, first highlight the replacement site in the left window of the Site Manager. Then press [MENU] to open the File Menu. Scroll to the Open option, and press [ENTER]. The screen appears similar to Figure 76 below.

  • Page 90
    Chapter 5. Save Site Save the current site (DEFAULT.SIT)? Figure 77: The Save Current Site Window • Press [F2] (No) to cancel saving the site, or • Press [F3] (Yes) to save the site. The PT878GC remains in the current window (Operate Mode or Site Manager), with the current site saved. TransPort™…
  • Page 91: Saving A Site With A Different Name

    Chapter 5. 5.1.1.4 Saving a Site with a Different Name If you want to save the current site with a different name, open the Site Manager, press [MENU] to open the File Menu, scroll to the Save As option, and press [ENTER]. The screen appears similar to Figure 78 below. Save Current Site Delete Cancel…

  • Page 92: Deleting A Site

    Chapter 5. Rename Site DEFAULT Delete Cancel Figure 79: The Rename Site Window Use the four arrow keys to scroll to the desired letter or number, and press [ENTER]. Repeat this procedure until you have created the desired site name. (Press [F1], Delete, to remove any unwanted letters or numbers.) When you have finished, •…

  • Page 93: Creating A Site Message

    Chapter 5. 5.1.1.8 Creating a Site Message The Site Message option allows you to add an explanatory message (with up to 30 characters or spaces) for any given site. To create a site message: Press [MENU] to enter the File Menu. Then press the [] arrow key once to scroll from the File Menu to the Site Menu.

  • Page 94
    Chapter 5. Use the four arrow keys to scroll to the desired letter or symbol, and press [ENTER] to add the letter to the message. Note: Pressing [SEL] causes the screen to alternate between a set of upper-case (capital) letters, a set of lower-case letters, and a set of symbols.
  • Page 95: Printing A File

    Chapter 5. 5.1.1.9 Printing a File To print a file from the Site Manager, press [MENU] to enter the File Menu, press the [] arrow key once, scroll to the Print option, and press [ENTER]. The PT878GC shows a message indicating that its infrared scanner is looking for a receiving device.

  • Page 96: Transferring A File From A Pc To The Pt878Gc

    Chapter 5. 5.1.1.12 Transferring a File from a PC to the PT878GC Once you have stored site or meter files to a PC, you can then transfer them back to the PT878GC over the IR interface. The PT878GC only accepts files with a .sit (site) or .met (meter) extension. If you rename another type of file with one of these extensions and transfer it, it will be transferred, but will not function if you open it.

  • Page 97: Listing Files By Name

    Chapter 5. 5.1.1.13 Listing Files by Name If you want to list your files alphabetically by site name within the Site Manager, press [MENU] to open the File Menu. Then press the [] arrow key twice to scroll from the File Menu to the Sort Menu, shown in Figure 85 below. Press [ENTER].

  • Page 98
    Chapter 5. TransPort™ PT878GC User’s Manual…
  • Page 99: Displaying And Configuring Data

    Chapter 6. Chapter 6. Displaying and Configuring Data The PT878GC allows you to view from one to four different measurement parameters simultaneously. The screen can show these parameters not only in numeric format, but as line or bar graphs as well. You can configure any given measurement for your particular requirements.

  • Page 100
    Chapter 6. Number Format Programming Fixed Decimal Format Decimal Places Cancel Figure 87: The Number Format Window The first entry, Format, asks you to select the numeric format from three choices: fixed decimal, default and scientific. Default provides the default resolution, while Fixed Decimal allows users to override the standard resolution.
  • Page 101: The View Option

    Chapter 6. 6.1.2 The View Option The second option, View, allows you to select the presentation of a parameter in one of three formats: numeric, line graph or bar graph. From the Display Menu, press the [] key once to reach the View option, and then press [ENTER]. A drop-down menu shows the three formats.

  • Page 102: The Limits Option

    Chapter 6. 6.1.3 The Limits Option Once you have configured a parameter as a line or bar graph, you might need to change its presentation or values. The Limits option (replacing the Format option for line and bar graphs) enables you to program the minimum or maximum values displayed, the time interval and the display of the average value.

  • Page 103: The Measurement Option

    Chapter 6. 6.1.4 The Measurement Option On occasion, you might need to change the actual parameter measured in a given window.The Measurement option enables you to reconfigure the window with one of five categories of data source (velocity, volume, forward or reverse totalizer, or diagnostics) and appropriate English or metric measurement units.

  • Page 104: Customizing The Display Screen

    Chapter 6. 6.1.5 Customizing the Display Screen You might wish to display one or two parameters, or customize the soft keys to quickly access particular menus. The Site Menu enables you to make more comprehensive changes in your display screen. To enter the Site Menu, press the [MENU] key at the lower right of the PT878GC keypad.

  • Page 105: Specifying The Number Of Displayed Parameters

    Chapter 6. 6.1.5.1 Specifying the Number of Displayed Parameters As mentioned earlier, the PT878GC can display one to four different measurement parameters simultaneously. However, sometimes you might wish to display only one or two parameters. To change the number of open display windows: Press [MENU].

  • Page 106: Customizing Softkeys

    Chapter 6. 6.1.5.2 Customizing Softkeys When the screen is in Operate mode, you might wish to access a particular submenu frequently without the trouble of scrolling through menus. Customizing the softkeys (the function keys —[F1], [F2] and [F3]) allows you to access up to three submenus by pressing the associated softkey.

  • Page 107
    Chapter 6. Press [ENTER] to confirm the entry, and press [F3] (OK) to confirm the entry and close the window. (Press [F2] (Cancel) to close the window without changing the key.) The screen now appears similar to Figure 6-10 on the next page, with the [F1] window displaying “Contrast.” Pressing [F1] opens the Contrast window.
  • Page 108: Managing Files — The Drive Manager

    Chapter 6. 6.1.6 Managing Files — The Drive Manager On occasion, you might want to review, print or transfer some or all of the files in the PT878GC. The Drive Manager allows you to view all the files stored in the meter. To open the Drive Manager: Press [MENU].

  • Page 109
    Chapter 6. The window on the left lists all the meter, site and log files in the PT878GC, while the window on the right displays information on the file highlighted in the left window. File Manager File Sort Transfer Info: SYSLOG.MET SYSLOG.
  • Page 110: Refreshing A File

    Chapter 6. 6.1.6.1 Refreshing a File You can refresh a file (updating the display with the most current information) in one of two ways: • Press [F2], Refresh, to refresh the highlighted file. • From the File Menu, scroll to the Refresh option and press [ENTER]. The updated information on the highlighted file appears in the window on the right.

  • Page 111
    Chapter 6. Note: For Windows NT 4.0, check that the QuickBeam software (available through the path C:Program FilesQuickBeam Suite) is running, and that the IR beam on the PT878GC has clear access to the IR sensor connected to the PC port. You can send the file in one of two ways: •…
  • Page 112: Deleting A File

    Chapter 6. 6.1.6.4 Deleting a File To delete a file in the File Manager, first be sure you have highlighted that site in the left window of the File Manager. Then press [MENU], scroll to the Delete option, and press [ENTER]. The screen appears similar to Figure 100 below. File Manager Delete Confirmation File…

  • Page 113: Listing Files By Name

    Chapter 6. 6.1.6.5 Listing Files by Name If you want to list your files alphabetically by site name within the File Manager, press [MENU] to open the File Menu. Then press the [] arrow key to scroll from the File Menu to the Sort Menu, shown in Figure 101 below. Press [ENTER]. The File Manager screen refreshes, with the sites listed in alphabetical order.

  • Page 114: Accessing Meter Data -The About Option

    While the window normally appears briefly at startup, users might want to access the information for a longer period. To open the About window, scroll to the About option on the Site Menu and press [ENTER]. The screen appears similar to Figure 102 below. Panametrics Sensing PT878GC Portable Flowmeter Copyright ©2002 Baker Hughes All rights reserved.

  • Page 115: Programming Meter Settings

    Chapter 7. Chapter 7. Programming Meter Settings Along with display formats and site data, PT878GC users can program global settings for the meter that suit their individual preferences. The global settings include: • English or Metric measurement units • Battery power •…

  • Page 116: Entering The Meter Menu

    Chapter 7. 7.1.1 Entering the Meter Menu To enter the Meter Menu, press the [MENU] key at the lower right of the PT878GC keypad. The Main Menu replaces the  Status Bar at the top of the screen. Press the [ ] arrow key twice to scroll from the Site Menu to the Meter Menu.

  • Page 117
    Chapter 7. Meter Settings Units English Metric Pressure Units: Date: Bars 2000/11/1 Time: 10:08:53 AM Cancel Figure 104: The Meter Settings Window   At the Units prompt, use the [ ] and [ ] keys to scroll between English and Metric units. Press [ENTER] to confirm the choice.
  • Page 118: The Battery Charger

    Chapter 7. 7.1.3 The Battery Charger The Battery option allows you to monitor the current run time and status of the internal rechargeable batteries, as well as to condition NiCad batteries to maintain the maximum life possible. Conditioning NiCad batteries (a process that can take up to 12 hours for a fully charged pack) discharges the pack completely and then performs a fast charge.

  • Page 119: Entering Date And Time

    Chapter 7. 7.1.4 Entering Date and Time In Operate mode, the Status Bar displays the current date and time above the measurements in the upper right corner of the screen. The Date/Time option allows you to set the date or time, which are required for correct data logging operation.

  • Page 120: Changing Date And Time Appearance (Locale)

    Chapter 7. 7.1.5 Changing Date and Time Appearance (Locale) In addition to setting the correct date and time, you can also change its presentation to suit local preferences. You can select a time display of AM/PM or 24-hour time To alter the time and date display: From the Meter menu, scroll to the Locale entry and press [ENTER].

  • Page 121
    Chapter 7. The PT878GC now asks you to select whether you want the time presented in a 12-hour format (for example, 11:53:23 PM) or in a 24-hour format (23:53:23). [ENTER] Press to open the drop-down menu.   Use the [ ] or [ ] arrow keys to scroll to the 12-hour or 24-hour entry.
  • Page 122: Adjusting The Contrast

    Chapter 7. 7.1.6 Adjusting the Contrast For more comfortable viewing in a particular environment, the PT878GC enables you to adjust the screen contrast. To adjust the screen contrast: From the Meter menu, scroll to the Contrast entry and press [ENTER]. The Display Options window opens on the …

  • Page 123: Setting Backlight Timeout

    Chapter 7. 7.1.7 Setting Backlight Timeout By using the Backlight Timeout option, you can set a specified time that the PT878GC backlight will remain on before turning itself off. Automatic turnoff enables the PT878GC to conserve battery power. To set the backlight timeout: From the Meter menu, scroll to the Backlight entry and press [ENTER].

  • Page 124: Changing Communications Parameters

    Cancel Figure 108: The Communications Window The first prompt asks for the node identification number, which can be any number from 1 to 240. IMPORTANT: Do NOT change the node ID unless instructed by Panametrics Sensing. [ENTER] Press to open the text box.

  • Page 125
    Chapter 7. [ENTER] Press to open the drop-down menu.   Use the [ ] or [ ] arrow keys to scroll to the desired parity. Press [ENTER] to confirm the entry.   The next prompt asks you to select either one or two stop bits.The default number is one. Use the [ ] and [ keys to scroll to the desired number, and press [ENTER].
  • Page 126: Resetting Forward And Reverse Totals

    Chapter 7. 7.1.9 Resetting Forward and Reverse Totals On occasion, it might be necessary to clear and reset the forward and reverse totals computed by the Forward and Reverse Totalizers. To reset the totals: From the Meter menu, scroll to the Totals entry and press [ENTER]. The window now appears similar to Figure 109 below.

  • Page 127: Setting Up User Tables

    Chapter 7. 7.1.10 Setting Up User Tables When you program user functions (see page 68), you can also support them with up to six user tables of non-linear or empirical data. To program one or more user tables: From the Meter menu, scroll to the User Tables entry and press [ENTER]. The window now appears similar to Figure 110 below.

  • Page 128
    Chapter 7. UserTable1 “ ‘ < > Delete Cancel Figure 111: The Text Creation Window Use the four arrow keys to scroll to the desired letter or symbol, and press [ENTER] to add the letter to the label. Note: Pressing [SEL] causes the screen to alternate between a set of upper-case (capital) letters and a set of symbols.
  • Page 129
    Chapter 7. Use the four arrow keys to move to the desired entry in the table. Press [ENTER]. Then use the numeric keys to enter the desired data, which appears in the right corner of the window above the table. Press [ENTER] to confirm the data, which then appears in the appropriate slot in the table.
  • Page 130: Taking A Bitmap Capture Of A Current Screen

    Chapter 7. 7.1.11 Taking a Bitmap Capture of a Current Screen The Snapshot option enables you to take a “screen capture” of the current screen in bitmap format (.bmp) for display or storage in a Windows-based PC. To take a “snapshot” of the screen: From the Meter menu, scroll to the Snapshot entry and press [ENTER].

  • Page 131: Logging Data

    Chapter 8. Chapter 8. Logging Data A powerful and flexible feature of the PT878GC is data logging. The meter enables you to choose up to 12 parameters to log. You can also select the start time and date, end time and date, and time interval. Logs can run one at a time or simultaneously.

  • Page 132: Entering The Logging Menu

    Chapter 8. 8.1.1 Entering the Logging Menu To enter the Logging Menu, press the [MENU] key at the lower right of the PT878GC keypad. The Main Menu replaces  the Status Bar at the top of the screen. Press the [ ] arrow key three times to scroll from the Site Menu to the Logging Menu.

  • Page 133
    Chapter 8. Log Manager File View Sort Info: AAA.LOG State: Finished S:03/01/04 13:13:41 E:03/01/04 13:23:41 Interval: 10 Seconds Records:51 03/01/04 13:23:21 1216 bytes 109568 bytes free Refresh Exit Figure 114: The Log Manager Window To access the menu for the Log Manager, press the [MENU] key. The cursor highlights the File Menu in the upper left …
  • Page 134: The File Menu

    Chapter 8. 8.1.3 The File Menu The File Menu allows you, not only to create new logs, but also to copy, rename or delete logs, as well as to print them or transfer them to a PC. To open the File menu from the Log Manager, press the [MENU] key and then [ENTER]. The screen appears similar to Figure 115 below.

  • Page 135: Setting Up A New Log

    Chapter 8. 8.1.3.1 Setting up a New Log The New Log option enables you to create and set up parameters for a new log. You can access this option in two ways: • by scrolling to the New Log option in the Logging Menu (as shown in Figure 8-1 on page 8-2) and pressing [ENTER], or •…

  • Page 136
    Chapter 8. General Measurements Log Name 10SEC.LOG Format Circular Linear Type Error Standard Start Date/Time 01:38:08 2004/11/01 End Date/Time 09:38:08 2004/11/01 Logging Interval secs Activate Cancel Figure 117: The General Log Format Window To step through each parameter, press the [] key. The first prompt asks you to choose between a linear or circular format for the log.
  • Page 137
    Chapter 8.  Finally, press the [ ] key to move to the Measurements tab, and press [ENTER]. The Measurements window appears similar to Figure 118. General Measurements NO UNIT NO UNIT NO UNIT NO UNIT NO UNIT NO UNIT NO UNIT NO UNIT NO UNIT…
  • Page 138: Copying (Cloning) A Selected Log

    Chapter 8. The prompt then moves to a list of unit types. (The available units depend on the selection made at the Data Source prompt.) Scroll to the desired output unit. Press [F3] (OK) to confirm your selection. You can repeat this procedure for up to 12 different parameters. When you have finished, press [F2] (Cancel) to cancel the entries, or [F3] (Activate) to confirm the entries and start the log.

  • Page 139: Deleting All Logs

    Chapter 8. Log Manager File Delete Programming Info:DEFAULT DEFAULT Delete Log ‘555.LOG?’ GLOBAL 1 View Figure 120: The Delete Confirmation Window 8.1.3.5 Deleting All Logs To clear the Log Manager and memory of all logs, open the File menu, scroll to the Delete All Logs option, and press [ENTER].

  • Page 140: The Log Menu

    Chapter 8. message indicating that it is uploading the log. (The meter also displays a message if the upload fails.) When the upload is complete, the meter returns to the Log Manager. The PC holds the transferred log in a CIr_Inbox folder (C:Ir_Inbox for Windows NT, C:My Received Files for Windows 98, or Desktop for Windows 2000).

  • Page 141: Pausing All Logs

    Chapter 8. The PT878GC returns to the Log Manager, which displays the highlighted log with a status of “Finished.” The space not used by the finished log is freed for reuse. Note: You cannot restart a finished log. You must create a new log with the same parameters. 8.1.4.4 Pausing All Logs To pause all logs that are currently pending or running, open the Log menu, scroll to the Pause All Logs option, and…

  • Page 142: The View Menu

    Chapter 8. 8.1.5 The View Menu Through the View menu, you can view the data of individual logs in graphical or spreadsheet formats. To open the View menu from the Log Manager, press [MENU]. Scroll to the View menu, and press [ENTER]. The screen appears similar to Figure 122 below.

  • Page 143: Displaying Log Details

    Chapter 8. 8.1.5.1 Displaying Log Details To view details of a given log, be sure the log is highlighted in the left window of the Log Manager. Then scroll to the View menu and press [ENTER]. Scroll to the Details option and press [ENTER]. The screen now appears similar to Figure 123 below.

  • Page 144: Displaying Log Data In Graphical Form

    Chapter 8. 8.1.5.2 Displaying Log Data in Graphical Form To view a log in graphical form, be sure the log is highlighted in the left window of the Log Manager. Scroll to the View menu and press [ENTER]. Scroll to the Graph option and press [ENTER]. The “Select Measurement”…

  • Page 145
    Chapter 8. The Y-Axis window allows you to specify whether the Y axis on the graph extends to the maximum value (Max), over the entire range (Range) or between certain specified values (Set).   Use the [ ] and [ ] keys to scroll to the desired limit type.
  • Page 146: Displaying Log Data In Spreadsheet Form

    Chapter 8. 8.1.5.3 Displaying Log Data in Spreadsheet Form To view a log in spreadsheet form, be sure the log is highlighted in the left window of the Log Manager. Then scroll to the View menu and press [ENTER]. Scroll to the Spreadsheet option and press [ENTER]. The screen now appears similar to Figure 127 below.

  • Page 147: The Sort Menu

    Chapter 8. 8.1.6 The Sort Menu The Sort Menu within the Log Manager allows you to arrange your log list either alphabetically (By Name) or chronologically (By Date). Log Manager Sort File View By Name  Info: AAA.LOG State: Finished By Date S:03/01/04 13:13:41 E:03/01/04 13:23:41…

  • Page 148
    Chapter 8. TransPort™ PT878GC User’s Manual…
  • Page 149: Servicing The Pt878Gc

    Chapter 9. Chapter 9. Servicing the PT878GC For user convenience, the PT878GC offers a Service Menu. This menu enables users to perform a variety of functions that they might occasionally require: • print out reports • calibrate the PT878GC • run diagnostics •…

  • Page 150: Entering The Service Menu

    Chapter 9. 9.1.1 Entering the Service Menu To enter the Service Menu, press the [MENU] key at the lower right of the PT878GC keypad. The Main Menu replaces  the Status Bar at the top of the screen. Press the [ ] arrow key four times to scroll from the Site Menu to the Service Menu.

  • Page 151
    Chapter 9. Reports Reports Drive Contents Current Site Global Settings User Functions User Tables All User Settings Menu Commands All Reports Exit Print Figure 130: The Reports Window   Press [ENTER] to open the drop-down list of available reports, as shown in Figure 130 above. Press the [ ]or [ arrow keys to scroll to the desired reports, and press [ENTER].
  • Page 152
    Chapter 9. Figure 131: Printout of a Typical Drive Report TransPort™ PT878GC User’s Manual…
  • Page 153: Setting Up The Thickness Gauge

    Chapter 9. 9.1.3 Setting up the Thickness Gauge For greatest accuracy in flow applications, the PT878GC can measure pipe wall thickness using an optional thickness gauge transducer, instead of relying on the nominal pipe wall thickness. In Thickness Gauge mode, the PT878GC does not measure flow, but it can determine the thickness of most standard metal and plastic pipe materials over a range from 0.05 to 3 in.

  • Page 154: Measuring Pipe Wall Thickness

    Measuring Pipe Wall Thickness 9.1.4.1 Entering the Material and Sound Speed IMPORTANT: Panametrics Sensing recommends calibrating the thickness gauge periodically (as discussed on page 148) before measuring thickness. To enter the Thickness Gauge Display option, scroll to the T-Gauge Display entry on the Service Menu and press [ENTER]. The screen appears similar to Figure 133 below.

  • Page 155: Measuring Thickness In Numeric Format

    Chapter 9. 9.1.4.2 Measuring Thickness in Numeric Format To measure the actual thickness of a pipe, apply couplant to the calibrated thickness gauge transducer and hold the   transducer steady against the pipe. Then press the [ ] or [ ] arrow key until you reach the Display tab and press [ENTER].

  • Page 156: Displaying The Receive Signal In Graphical Format

    Chapter 9. 9.1.4.3 Displaying the Receive Signal in Graphical Format Used chiefly for diagnostic purposes, the Graph option helps to determine why the thickness gauge is not working if you suspect a problem. The graph shows an image of the acoustic signal. If the display does not show a signal image similar to Figure 9-7 below, you may have a problem with the transducer, couplant, or the programmed ®…

  • Page 157
    Chapter 9. Thickness Gauge Measure Display Material Velocity Graph Zero Signal Display Figure 136: The Graph Window in Zoom Format • Press [F1] (In) to zoom in to magnify screen details. • Press [F2] (Out) to zoom out fully. • Press [F3] to toggle between the left and right cursors. …
  • Page 158: Calibrating The Thickness Gauge Transducer

    See Appendix C for more information. The next prompt asks you to enter the length of Block 1. (If you are using a Panametrics Sensing-supplied test block, the length is printed on the block.) Press [ENTER] to open the text box.

  • Page 159: Calculating Velocity (Pipe Material Sound Speed)

    Chapter 9.   • Press the [ ] or [ ] arrow key to move to another tab to take readings or perform a velocity calibration. • Press [F2] (Cancel) to return to Operate Mode without confirming the sound speed value. •…

  • Page 160
    Chapter 9. Thickness Gauge Measure Display Velocity Graph Zero Material Block Length Press Set to Commit Value Current Calculated 22129. 74733. Cancel Figure 139: The Velocity Window, Displaying the Calculated and Current Values The “Calculated” box shows the thickness value measured. The PT878GC asks for confirmation of the calculated and current values.
  • Page 161: Programming The Thickness Gauge

    Noise Threshold • Detection Threshold IMPORTANT: The thickness gauge programming settings are entered at the factory. You should not change them unless instructed by Panametrics Sensing. The Programming window appears similar to Figure 140 below. Thickness Gauge Setup Programming Low Signal Thresh Transducer Delay µs…

  • Page 162
    Chapter 9.   Signal inversion should be on, unless you have received other instructions. To change its status, use the [ ] and [ arrow keys to move to the appropriate radio button and press [ENTER]. To enter the noise threshold: [ENTER] Press to open the text box.
  • Page 163: Displaying Diagnostic Parameters

    Chapter 9. 9.1.6 Displaying Diagnostic Parameters The Diagnostics option enables you to view current diagnostic parameters without having to open a display window in Operate Mode. To enter the option, scroll to the Diagnostics entry on the Service Menu and press [ENTER]. The screen appears similar to Figure 141 below.

  • Page 164: Calibrating The Analog Output And Inputs

    Chapter 9. 9.1.7 Calibrating the Analog Output and Inputs The Calibration option allows you to calibrate the analog output and inputs. To enter the option, scroll to the Calibrate entry on the Service Menu and press [ENTER]. The screen appears similar to Figure 142 below. Calibrate/Test I/O Analog Input Analog Output…

  • Page 165: Calibrating Inputs

    Chapter 9. 9.1.7.2 Calibrating Inputs  To open the Inputs window, press the [ ] arrow key and press [ENTER]. The screen appears similar to Figure 143 below. Note: Calibrating the analog inputs requires use of a current source. Calibrate/Test I/O Analog Input Analog Output Input#…

  • Page 166: Setting Up Signal Parameters

    The first prompt asks for the Delta-T offset. Delta-T is the difference between the upstream and downstream transit times of the transducers. The Delta-T offset should normally be set to zero. Note: Consult Panametrics Sensing before performing this step. [ENTER] Press to open the text box.

  • Page 167
    The next prompt, the transmitter sample size, is the number of pulses each transmitter (upstream and downstream) emits. It is set to 8 by default. Note: Consult Panametrics Sensing before performing this step. [ENTER] Press to open the drop-down list.
  • Page 168: Setting Up The Measurement Mode

    Skan/Correlation mode, is used primarily for liquid flow measurement. Note: Do not change the measurement mode or values unless recommended by Panametrics Sensing. From the Signal Parameter tab, press the [] arrow key to move to the Meas Mode tab. The window appears similar to Figure 145 below.

  • Page 169: Setting Up Pulse/Code Parameters

    Chapter 9. 9.1.8.2 Setting Up Pulse/Code Parameters When you select the Skan detection method, you must also choose what type of signal to send to the transducers: either a coded signal at 1, 2, 4, 11 or Twin bits, or a specific number of pulses to be sent. From the Meas Mode tab, press …

  • Page 170: Setting Error Limits

    Chapter 9. 9.1.9 Setting Error Limits The Error Limits option enables you to set limits for an incoming signal. When the signal falls outside the programmed limits, an error indication appears. To enter this option, scroll to the Error Limits entry on the Service Menu and press [ENTER].

  • Page 171
    Chapter 9. Use the numeric keys to enter the desired speed. Press [ENTER] to confirm your entry. Note: For the velocity and acceleration boxes, the F1 softkey toggles between English and metric measurements. Thus, the key will always display the opposite measurement from that which is currently active. Press [F1] to display the measurement in the alternate format.
  • Page 172: The Test Option

    Chapter 9. 9.1.10 The Test Option Within the Service Menu, the Test option includes seven tests to ensure that the PT878GC is performing properly: Test Screen, Test Keys, Watchdog Test, Impulse Response, Wave Snapshot, Simulate and Battery Test.To enter this option, scroll to the Test entry on the Service Menu and press [ENTER].

  • Page 173: Testing The Screen

    Pressing a key two more times should result in two more checkerboard patterns, followed by a series of dark and light screens. Pressing the key through this sequence should return the PT878GC to Operate Mode. If the test does not proceed according to this sequence, please consult Panametrics Sensing. TransPort™ PT878GC User’s Manual…

  • Page 174: Testing The Keys

    Figure 150: The Test Keys Window Pressing the [F3] key returns the meter to the Operate Mode. If any key does not appear on the screen or does not darken when pressed, contact Panametrics Sensing. Note: The power key does not appear in this test.

  • Page 175: Testing The Watchdog Timer Circuit

    Press [F2] (No) to cancel the test and return to the Menu screen, or press [F3] (Yes) to start the test. The PT878GC should go blank for a few seconds, and then restart. If it does not follow this sequence, consult Panametrics Sensing.

  • Page 176: Setting Impulse Response

    Chapter 9. 9.1.10.4 Setting Impulse Response The Impulse Response option enables you to force the meter to transmit in one direction only, without changing the AGC setting. You can then diagnose problems with transducer, pipe, or fluid configurations. To enter the option, scroll down to the Impulse Response entry in the Service Menu and press [ENTER].

  • Page 177: Applying A Stored Signal For Diagnosis

    Chapter 9. 9.1.10.6 Applying a Stored Signal for Diagnosis The Simulate option places the PT878GC in a mode in which it uses a stored signal (instead of the live signal from the transducers) to make flow calculations for diagnostic purposes. On the PC, you must rename a Wavexx.met file as Wave.met and send it back to the PT878GC over the IR link.

  • Page 178
    • IrCOMM, an infrared standard supported by certain Microsoft operating systems. Panametrics Sensing recommends updating software via the IrOBEX standard; however, the IrCOMM standard is available for users who have problems with IrOBEX. This section covers procedures for both standards.
  • Page 179
    • Press [F3] (OK) to confirm that you wish to erase the program. The meter asks for confirmation. Repeat the options shown in Step 1 above. After the PT878GC reboots, the screen appears similar to Figure 156 below. Panametrics Sensing PCI Loader v3.0 2/26/04 [HW Rev3+]…
  • Page 180
    Chapter 9. 9.1.12.2 Updating Software Via IrCOMM Note: While Windows 95/98/98SE and NT (with QuickBeam) support IrCOMM, Windows 2000 and XP do not. Refer to Appendix B. Before you install new coding, you must be sure that the PC has the correct protocols to transmit the software to the meter.
  • Page 181
    • Press [F3] (OK) to confirm that you wish to erase the program. The meter asks for confirmation. Repeat the options shown in Step 1 above. The screen on both the PC and the PT878GC now appears similar to Figure 159 below. Panametrics Sensing PCI Loader v3.0 2/26/04 [HW Rev3+]…
  • Page 182
    Chapter 9. In the Hyperterminal window, pull down the Transfer menu and click Send File. A window opens similar to Figure 160 below. Figure 160: The Send File Window in the Transfer Menu Click on the replacement software (designated by a .cod extension) from the folder where it has been stored. From the Protocol drop-down menu, select Xmodem.
  • Page 183
    Diagnostics and Troubleshooting The TransPort PT878GC is a reliable instrument that is easy to maintain. It will provide accurate flow measurement readings as long as it is operated as described in this manual. If problems do arise with the electronics, transducers or the flowcell, the TransPort displays an error message specifying the possible problem.
  • Page 184
    “Error.”The Error Code messages are only general descriptions of the possible problems. Use Table 8 to isolate and remedy the problem. If you are unable to remedy the problem, contact Panametrics Sensing. Note: We suggest that, in order to isolate the indicated problem more easily, you obtain a test flowcell.
  • Page 185
    Chapter 10. Table 8: Error Messages Error Message Problem Possible Cause Action No error. Displays briefly after the None required. Measurement is valid. display of another error message Low Signal — Poor ultrasonic Broken cable. Flowcell problem. Check transducer cable. See Flowcell signal strength.
  • Page 186
    Chapter 10. 10.1.2 Displaying Diagnostic Parameters As part of its measurement menu, the PT878GC offers a list of diagnostic parameters to aid in troubleshooting in the event of flowcell, transducer, or electrical problems. You can select any diagnostic parameter for display as a measurement as discussed in The Measurement Option.
  • Page 187
    Chapter 10. Table 9: Diagnostic Parameters Diagnostic Parameter Displays Good Signal up Displays the signal strength for the upstream 50-80 <50 transducer. Signal dn Displays the signal strength for the 50-80 <50 downstream transducer. Displays the signal quality for the upstream ±300 or higher between ±100 transducer.
  • Page 188
    Pipe Problems Improper pipe conditions and/or flowcell installation can cause problems with measurement of the gas flow. Contact Panametrics Sensing if you cannot solve pipe-related problems. By far, the most common pipe problems are the following THE COLLECTION OF MATERIAL AT THE TRANSDUCER LOCATION(S). Accumulated debris at the transducer location(s) will interfere with transmission of the ultrasound signals.
  • Page 189
    If corrosion is suspected, remove the transducer from the flowcell and carefully inspect the electrical connector and the transducer surface for roughness and/or pitting. Any transducer damaged in this manner must be replaced. Contact Panametrics Sensing for information on transducers in materials suitable for the application.
  • Page 190
    Chapter 10. TransPort™ PT878GC User’s Manual…
  • Page 191
    Chapter 11. Chapter 11. 11.1 Specifications This section contains specifications for the following: • Operations and Performance • Electronics • Clamp-on Transducers • Thickness Gauge Option • Additional Options TransPort™ PT878GC User’s Manual…
  • Page 192
    Pressure Requirements, Steam See Table 6. 11.1.1.12 Pressure Requirements, Other Gases Dependent upon gas composition, and pipe size and material. Consult Panametrics Sensing. 11.1.1.13 Measurement Parameters Standard and actual volumetric flow, flow velocity and mass flow. TransPort™ PT878GC User’s Manual…
  • Page 193
    Chapter 11. 11.1.2 Electronics 11.1.2.1 Flow Measurement Patented Correlation Transit-Time mode 11.1.2.2 Enclosures Submersible IP67 11.1.2.3 Dimensions Weight: 3 lb (1.36 kg) Size: (h x w x d) 9.4 x 5.5 x 1.5 in. (238 x 138 x 38 mm) 11.1.2.4 Display 240- x 200-pixel backlit LCD graphic display…
  • Page 194
    Chapter 11. 11.1.2.16 European Compliance Battery-powered system complies with EMC Directive 89/336/EEC, and transducers comply with PED 97/23/EC for DN<25. 11.1.3 Clamp-On Ultrasonic Flow Transducers 11.1.3.1 Temperature Range(s) Standard: –40° to 130°C (–40° to 266°F) Optional (overall range): –40° to 230°C (–40° to 446°F) 11.1.3.2 Transducer Materials Stainless steel and plastic…
  • Page 195: Pipe Materials

    Chapter 11. 11.1.4 Thickness Gauge Option 11.1.4.1 Transducer Panametrics Sensing dual element transducer 11.1.4.2 Pipe Thickness Range 0.05 to 3 in. (1.3 to 76.2 mm) 11.1.4.3 Pipe Materials Most standard metal and plastic pipe materials 11.1.4.4 Accuracy ±1% typical or ±0.002 in. (±0.05 mm) 11.1.4.5…

  • Page 196
    Chapter 11. TransPort™ PT878GC User’s Manual…
  • Page 197: Menu Maps

    Appendix A. Appendix A. Menu Maps Figure 162: The Main Menu TransPort™ PT878GC User’s Manual…

  • Page 198
    Appendix A. TransPort™ PT878GC User’s Manual…
  • Page 199
    Appendix A TransPort™ PT878GC User’s Manual…
  • Page 200
    Appendix A TransPort™ PT878GC User’s Manual…
  • Page 201
    Appendix A TransPort™ PT878GC User’s Manual…
  • Page 202
    Appendix A TransPort™ PT878GC User’s Manual…
  • Page 203
    Appendix A TransPort™ PT878GC User’s Manual…
  • Page 204
    Appendix A TransPort™ PT878GC User’s Manual…
  • Page 205
    Appendix A TransPort™ PT878GC User’s Manual…
  • Page 206
    Appendix A Figure -169: The Logging Menu Figure -170: The Service Menu TransPort™ PT878GC User’s Manual…
  • Page 207: Establishing Ir Communications With The Pt878Gc

    IrDA compatibility and either a built-in IR port (available on most laptops) or an IR dongle (IR to RS232 adapter for PCs without a built-in IR port). If your PC does not have an IR dongle, Panametrics Sensing recommends the ActiSys ACT-IR220L+ infrared to RS-232 adapter, which has been tested for compatibility with the PT878GC.

  • Page 208: Windows Nt4.0

    Appendix B. B.1.0.2 Windows NT4.0 For IR use, Windows NT4.0 requires installation of QuickBeam software from Extended Systems (www.extendedsystems.com). Desktop PCs and laptops without built-in IR ports also require adding an IR dongle. While laptops with a built-in IR port normally do not require a dongle, its addition might be necessary in some cases. B.1.0.3 Windows Me/98SE/98/95 The Windows 98SE and Me operating systems include IR drivers.

  • Page 209: Ultrasonic Thickness Gauge Theory Of Operation

    Appendix C. Appendix C. Ultrasonic Thickness Gauge Theory of Operation All ultrasonic thickness gauging involves timing the round trip of a sound pulse in a test material. Because solid metal has an acoustic impedance that differs from that of gases, liquids, or corrosion products such as scale or rust, the sound pulse will reflect from the far surface of the remaining metal.

  • Page 210
    Appendix C. Transducer Figure 171: Proper Alignment of Transducers for Cylindrical Surfaces While firm hand pressure on the transducer is necessary for good readings, the probe should never be scraped along or twisted against a rough metal surface. This will scratch the face of the transducer and eventually degrade performance.
  • Page 211
    Material Safety Data Sheets for Couplants A variety of couplants are available for use with Panametrics Sensing transducers. To obtain a material safety data sheet (MSDS) for any couplant Panametrics Sensing supplies, go to the Panametrics Sensing web site and enter the Technical Support portal.
  • Page 212
    Appendix D. TransPort™ PT878GC User’s Manual…
  • Page 213
    Index Connections ……….5 About Option .
  • Page 214
    Index Electronics ……….183 Input/Output Connections.
  • Page 215
    Index Pipe Sizes ……….182 M>S Switch.
  • Page 216
    Index Site Steam Deleting a ……….82 Installation Requirements for .
  • Page 217
    Index Transducers ……….2 Area Classifications .
  • Page 218
    Index TransPort™ PT878GC User’s Manual…
  • Page 219
    Panametrics Limited Shannon Industrial Estate Shannon, County Clare Ireland declare under our sole responsibility that the  TransPort PT878 Portable Ultrasonic Flowmeter  TransPort PT878GC Clamp-On Portable Ultrasonic Flowmeter to which this declaration relates, are in conformity with the following standards: •…
  • Page 220
    <Doc Tittle> <Manual Type>…
  • Page 221
    Notify Panametrics Sensing, giving full details of the problem, and provide the model number and serial number of the instrument. If the nature of the problem indicates the need for factory service, Panametrics will issue a RETURN AUTHORIZATION NUMBER (RAN), and shipping instructions for the return of the instrument to a service center will be provided.
  • Page 222
    Warranty [no content intended for this page] TransPort™ PT878GC User’s Manual…
  • Page 224
    Customer Support Centers U.S.A. The Boston Center 1100 Technology Park Drive Billerica, MA 01821 U.S.A. Tel: 800 833 9438 (toll-free) 978 437 1000 E-mail: mstechsupport@bakerhughes.com Ireland Sensing House Shannon Free Zone East Shannon, County Clare Ireland Tel: +353 (0)61 470291 E-mail: mstechsupport@bakerhughes.com Copyright 2021 Baker Hughes company.

Предложите, как улучшить StudyLib

(Для жалоб на нарушения авторских прав, используйте

другую форму
)

Ваш е-мэйл

Заполните, если хотите получить ответ