Summary of Contents for Sartorius MA150C-000115V1
-
Page 1
This document courtesy of: Data Weighing Systems, Inc. Contact Us For immediate assistance call 1-800-750-6842… -
Page 2
Operating Instructions Sartorius Moisture Analyzer Model MA45 Electronic Moisture Analyzer 98648-010-37… -
Page 3
Intended Use The MA45 Moisture Analyzer can be The following features make the Symbols used for quick and reliable determi- moisture analyzer ideal for use as an The following symbols are used in these nation of the moisture content of inspection, measuring and test instru- instructions: materials of liquid, pasty and solid sub-… -
Page 4: Table Of Contents
Contents Intended Use Error Codes Contents Care and Maintenance Service Warning and Safety Information Repairs Cleaning Getting Started Safety Inspection Instructions for Recycling Equipment Supplied Overview General View of the Moisture Analyzer Specifications Connecting the Moisture Analyzer Accessories (Options) C Marking to AC Power Leveling the Moisture Analyzer Index…
-
Page 5: Warning And Safety Information
Improper use or handling, however, Fire Explosion can result in damage and/or injury. – Connect only Sartorius accessories and options, as these are optimally designed – Flammable or explosive substances To prevent damage to the equipment, for use with your moisture analyzer read these operating instructions –…
-
Page 6: Getting Started
Sartorius office or dealer. 2 hours at room temperature, leaving it Storage and Shipping Conditions unplugged from AC power. Afterwards,…
-
Page 7
Designation Order no. for replacement Designation Order no. for replacement Hinged cover with heating ‘Down/Back’ key element ‘Up/Forward’ key Disposable sample pan 6965542 (80 pcs.) Print key Pan support 69MA0092 Leveling foot Pan draft shield 69MA0093 Interface port Display Power socket Keypad On/off key Not shown:… -
Page 8: Connecting The Moisture Analyzer
– 230 volts: MA45-…230.. – 115 volts: MA45-…115.. If they do not match: Do not operate your moisture analyzer; contact your local Sartorius office or dealer. Use only – Original power cords – Power cords approved by a certified electrician/Sartorius service technician –…
-
Page 9: Leveling The Moisture Analyzer
Leveling the Moisture Analyzer Purpose: – To compensate for unevenness at the place of installation – This is necessary especially for testing liquid samples that need to be at a uniform level in the disposable sample pan Always level the moisture analyzer again any time after it has been moved to a different location.
-
Page 10: Operating Design
Operating Design Operation of the moisture analyzer follows a standardized “philosophy” which is described below. Keys The keys have following functions: e On/off key y Down/Back Entering Letters, Special Characters Turns the moisture analyzer on/off. When the operating menu is active: and Numbers: The moisture analyzer remains in Shows the next menu item on…
-
Page 11: Operation: Analysis And Test Functions
Operation Bar Graph: Function Line: There are two fundamentally different types of display: The bar graph is shown during moisture This line indicates the functions that can analysis if the “Initial Weight” option in the be activated by pressing “Enter” (e.g., –…
-
Page 12: Menu Operation
Menu Operation Parameter Settings Data Output This display is divided into two sections. The MA45 is equipped with an interface The MA45 has an operating menu for port for connecting either Line for Operating Status finding and setting parameters. Each menu has several levels.
-
Page 13: Error Codes
Error Codes Saving Data If you press a key that has no function, or which is blocked at a certain point Storing Parameter Settings in an application program, the error is The most recent configurations of indicated as follows: Setup menu parameters and drying programs are active when you switch on –…
-
Page 14: Configuring The Moisture Analyzer
Configuring the Moisture Analyzer Purpose Setting the Language You can configure your moisture ana- You can choose from 5 languages for lyzer to meet individual requirements by the display: entering user data and setting selected menu parameters in the Setup menu. –…
-
Page 15: Setting The Device Parameters
Setting the Device Parameters Factory Settings Entering or Changing the Password (DEVICE) Parameters: The factory-set configura- – The password (up to 8 characters) tions are identified by an “ο” in the list protects the following functions: Purpose starting on page 17. –…
-
Page 16
Additional Functions Reset the device parameters to the Print out parameter settings: factory settings: § If “Device parameters” are displayed: § Select the Setup menu: Press the r key Use the x or y key to select SETUP in the Function line and press u to confirm >… -
Page 17
Example: Set the display contrast to “3” Step Key (or Instruction) Display/Printout x or y 1. Select SETUP in the Function line C Auto. 0092 g SETUP PROGRAM TARE 2. Confirm SETUP 3. Select Device parameters 4. Confirm Device parameters 2 ƒ… -
Page 18: Device Parameters (Overview)
Device Parameters (Overview) ο Factory setting √ User-defined setting Setup – Password: empty Enter 8 Device characters max. parameters Interface ο SBI Baud rate 150 baud 300 baud ο 1,200 baud 2,400 baud 4,800 baud 9,600 baud 19,200 baud ο 7 bit Number of data bits 8 bit Parity…
-
Page 19: Device Information
Device Information § Print device information: Press the r key Purpose Display of device information > Printout (Example) Display Device Information ——————— § Select the Setup menu: 13.08.2001 13:02 Use the x or y key to select Model MA45-000230V SETUP in the Function line and press Ser.
-
Page 20: Basics Of Moisture Analysis
Basics of Moisture Analysis Purpose Material When deductive methods are used, the The MA45 moisture analyzer can be The moisture of a material is often mis- moisture content is indirectly deter- used for quick and reliable deter- takenly equated with its water content. mined.
-
Page 21
The Sartorius Moisture Analyzer cannot be used. Applications Guide will provide you with important information on the use of your moisture analyzer. -
Page 22: Preparation
Preparation Adjustment to an Existing Measuring System Before drying a sample, you must carry A moisture analysis method often replaces another drying method (e.g., oven drying out the following preparations: method), because it is simple to use and requires shorter analysis time. In this case, you should adapt this method to that of the moisture analyzer in order to obtain values –…
-
Page 23: Preparing A Sample
$ Use an appropriately designed tool for shredding a sample. Using disposable sample pans § Use only Sartorius disposable sample pans (inner diameter = 92 mm). Reusing sample pans leads to poor reproducibility of results for several reasons: – after cleaning, sample residues can still remain on the pan –…
-
Page 24
Applying a Sample to the Sample Pan § Apply the sample to the sample pan in a thin, even layer (height: 2 to 5 mm, weight: 5 to 15 g); otherwise: – a sample applied unevenly will result in a non-uniform distribution of heat –… -
Page 25: Operating The Moisture Analyzer
Operating the Moisture Analyzer Setting the Drying Parameters Purpose Adapt the moisture analyzer to the special requirements of products. Parameters can be configured individually for each program. Drying Parameters (Overview) ο Factory setting √ User-defined setting Program ο P1 memory (PROGRAM) Heating program ο…
-
Page 26
Features Standby Temperature Fully Automatic Mode: – Temperature set to defined value when The analysis ends as soon as the weight Number of Drying Programs the sample chamber is closed loss per 24 s is below an automatically MA45: 3 programs detected threshold. -
Page 27
——————— Printout of the Analysis Parameters § When the desired program is displayed: 13.08.2001 13:06 Press the r key Model MA45-000230V Ser. no. 13206969 Ver. no. 01-43-01 > Parameters are printed (see example on the right) ——————— Lines with more than 20 characters PROGRAM are truncated. -
Page 28
End of analysis: automatic (factory setting) Part A: Configuring the Drying Parameters Step Key (or Instruction) Display/Printout 1. Turn on the moisture analyzer Sartorius logo is displayed Self-test runs y or x 2. Select the PROGRAM function C Auto. 0092 g SETUP… -
Page 29
Step Key (or Instruction) Display/Printout 8. Confirm the temperature 9. Select the Initial weight 2 ƒ y parameter 10. Confirm the Initial weight parameter 11. Select the On setting and confirm 3 ƒ x 12. Enter the target weight 2.0 13. -
Page 30
Part B: Performing the Analysis Step Key (or Instruction) Display/Printout 1. Turn on the moisture analyzer Sartorius logo is displayed Self-test runs C Auto. 0092 g SETUP PROGRAM TARE 2. Prepare the sample: not necessary for corn starch 3. Open the sample chamber and position a new sample pan 4. -
Page 31
Step Key (or Instruction) Display/Printout ——————— 13.08.2001 16:01 Model MA45-000230V Ser. no. 13206969 Ver. no. 01-43-01 ——————— Prg. 1 Heating STANDARD Fin. temp. 130 ’C Stdby temp. Start W/STABIL. AUTOMATIC IniWt + 2.036 g The current moisture loss is displayed C Auto. -
Page 32: Heater Adjustment
“isoTEST” Calibration/Adjustment Functions The following functions are available Heater Adjustment Weighing System Settings in the SETUP menu, under the isoTEST Using 1-point and 2-point temperature menu item: adjustment and the YTM03MA temper- Calibration, Adjustment ature adjustment set (see “Accessories”) Purpose –…
-
Page 33
External Calibration/ Adjustment with a Factory-Defined Weight Externally calibrate and adjust the weighing system using a 30-g calibration weight Step Key (or Instruction) Display/Printout x or y 1. Select SETUP in the Function line C Auto. 0092 g SETUP PROGRAM TARE 2. -
Page 34
——————— 13.08.2001 16:15 Model MA45-000230V *) Date and time shown only when Ser. no. 99992581 using the YDP02 or YDP03 printer Ver. no. 01-43-01 from Sartorius ——————— External calibration W-ID Nom. 30.000 g Diff. — 0.002 g ——————— 13.08.2001 16:16… -
Page 35
13.08.2001 16:15 Model MA45-000230V *) Date and time shown only when Ser. no. 13206969 using the YDP02 or YDP03 printer Ver. no. 01-43-01 from Sartorius ——————— External calibration W-ID Nom. 30.000 g Diff. — 0.002 g External adjustment completed Diff. -
Page 36: Hardware Tests
Hardware Tests The following device elements can be tested: Purpose Hardware tests are performed to check – SBI communication whether the system communication – Heater with internal and external devices functions properly. These tests are not elementary hardware tests. Activating Hardware Tests Step Key (or Instruction) Display/Printout…
-
Page 37
Testing SBI Communication Step Key (or Instruction) Display/Printout 1. Prepare the test connector for the RS-232 port Connect TƒD (pin 2) with (see “Pin Assignment Chart”) RƒD (pin 3) 2. Select “Hardware tests” see above 3. Select “Test interfaces” function and confirm 4. -
Page 38: Data Output
Data Output There are three options for data output: – Output to the moisture analyzer – Output to an external printer (YDP02-0CE or YDP03-0CE) – Output to a peripheral device (e.g., a computer) via the interface port Output to the Moisture Analyzer Display (Weights and Calculated Values) The display is divided into 9 sections.
-
Page 39
Busy Symbol The “Busy” symbol is shown here when the moisture analyzer is processing a function activated by pressing a key. Plus/Minus Sign Any plus or minus sign required for a displayed value is shown here. Measured Value/Result This line shows: 5234 –… -
Page 40: Interface Port
Factory Settings for the SBI Operating Mode Parameters: RS-232 cables purchased from other manufacturers often have incorrect pin Transmission rate: 1,200 baud assignments for use with Sartorius Number of data bits 7 bits weighing systems! Be sure to check the pin assignments against the chart below…
-
Page 41: Data Output Format
Data Output Format Output Format with 20 Characters + CR LF You can output the values, displayed in the corresponding line for moisture 9 10 11 12 13 14 15 16 17 18 19 20 21 22 analysis values, and the weight unit U CR LF with or without a data ID code.
-
Page 42
During data communication between The moisture analyzer SBI interface Once XOFF has been received, it pre- the moisture analyzer and an on-line (Sartorius Balance Interface) has vents further transmission of characters. device (computer), messages consisting transmit and receive buffers. You can… -
Page 43: Pin Assignment Chart
Pin Assignment Chart Female Interface Connector: Pin Assignment Chart: 25-position D-Submini (DB25S) with Pin 1: Signal ground screw lock hardware for cable gland Pin 2: Data output (TƒD) Pin 3: Data input (RƒD) Male Connector Used: Pin 4: Signal return (TƒD/RƒD) (please use connectors with the Pin 5: Clear to send (CTS) same specifications):…
-
Page 44: Cabling Diagram
Cabling Diagram – Diagram for interfacing a computer or other peripheral device to the moisture analyzer using the RS-232/V24 standard and cables up to 15 m (50 ft.) long Moisture analyzer Peripheral device 25-pin 9-pin – – Type of cable: AWG 24 specification…
-
Page 45
External device not ready to send Transmit XON, then CTS (interface handshake interrupted; XOFF, CTS) Print fct. blocked ERR 53 No temperature compensation Contact the Sartorius Service Center No meas. value ERR 101 Key is stuck Release key Key pressed when switching on… -
Page 46
– Increase the sample quantity Insufficient heat output – Clean the protective glass or temperature sensor Place of installation exposed to – Change the place of installation interfering factors (vibration, etc.) If any other errors occur, please contact your local Sartorius Service Center… -
Page 47
Care and Maintenance Service Cleaning the Heating Unit and Regular servicing by a Sartorius techni- Temperature Sensor § Open the hinged cover cian will extend the service life of your moisture analyzer and ensure its contin- ued weighing accuracy. Sartorius can… -
Page 48
In this case, notify your nearest Sartorius Service Center. Maintenance and repair work may be performed only by service technicians who are authorized by Sartorius and who: – have access to the required mainte- nance manuals – have attended the relevant service… -
Page 49
Overview Specifications Model MA45 Dryer functions: Heating element Ceramic IR heater Temperature range 40 – 230°C Temperature increments adjustable in 1°C increments Temperature adjustment with YTM03MA temperature adjustment set Weighing Functions: Weighing capacity 45 g Readability 1 mg, 0.01% moisture content Repeatability, average (%) sample weight 1–5 g: 0.2% sample weight 5 g: 0.05%… -
Page 50
Calibration weight, 30 g (± 0.3 mg) YSS43 Standard Operating Procedure (SOP)YSL02A For Working with a Moisture Analyzer (For Quality Assurance Documentation) Spare Parts Dust cover for keypad 6960MA02 Forceps 69MA0072 Other spare parts can be ordered directly through your local Sartorius Service. -
Page 51
Important Note: and laboratory use The operator shall be responsible for Part 1: any modifications to Sartorius equip- General requirements ment and must check and, if necessary, Part 2-010: correct these modifications. On request, Particular requirements for laboratory… -
Page 52
Index Page Page Accessories 2, 4, 31, 49 Keypad 2, 5 Acoustic signal 14, 17 Keys: description Adjustment to an existing measuring system Language: setting 8, 13 Adjustment 31, 44 Leveling the moisture analyzer Alphanumeric input MA30 mode 24, 48 Basics Maintenance C marking… -
Page 53
General Password Enter/Change Password § Select the password function: § Select the Setup menu: Press the u key Use the x or y key to select SETUP in the Function line and press > Password prompt is displayed, together u to confirm with current password §… -
Page 54
Brief Instructions Sartorius MA45 y Down/Back Overview of Key Functions e On/Off Key In menu: Go to next menu item on Turn moisture analyzer on/off; same level analyzer remains in standby mode when In alphanumeric input: Show next lowest switched off… -
Page 55
Moisture Analysis Add sample: Switch analyzer on: e key Close sample chamber: Select drying program: Start analysis: START function and u key Select PROGRAM: or directly after closing sample chamber y and u keys Drying in progress: wait Use the y and u keys to select program P1, P2 or P3 Read off result: Cancel password prompt, if necessary: c key… -
Page 56
Configuration (SETUP) Language Activate menu: Use the y and u keys to German English select the SETUP function; confirm U.S. mode Select desired menu level: y and x keys Français Italiano Exit menu level or exit SETUP: c key Español Select menu item and setting: Device parameters y, x and u keys… -
Page 57
Drying Parameters (PROGRAM) Heating program Activate menu: Use the y and u keys to Standard drying MA30 mode select the PROGRAM function; confirm Select program P1, P2 or P3 Standby temperature Exit menu level or exit PROGRAM: Initial weight c key Start analysis Select menu item and setting: With stability when Enter is pressed… -
Page 58
Sartorius AG Weender Landstrasse 94–108 37075 Goettingen, Germany Phone +49.551.308.0 Fax +49.551.308.3289 www.sartorius.com Copyright by Sartorius AG, Goettingen, Germany. All rights reserved. No part of this publication may be reprinted or translated in any form or by any means without the prior written permission of Sartorius AG.
Table of Contents for Sartorius MA150C-000115V1:
-
Moisture Analysis Switch analyzer on: e key Select drying program: Select PROGRAM: y and u keys Use the y and u keys to select program P1, P2 or P3 Cancel password prompt, if necessary: c key Exit program selection: c key twice Open sample chamber: Position sample pan: Tare sample pan: TARE function and u key Add sample: Close sample chamber: Start analysis: START function and u key or directly after closing sample chamber Drying in progress: wait Read off result: 1890 PROGRAMM P1 130 o C Auto. 128 o C 5.2
-
WMA6026-e01111 Brief Instructions Sartorius MA45 Overview of Key Functions eOn/Off Key Turn moisture analyzer on/off; analyzer remains in standby mode when switched off c Clear Function During analysis: Cancel function (e.g., taring) In menu: Delete input; exit current menu level u Enter During analysis: Activate selected function (e.g., taring, analysis, cancel) In menu: Store selected setting/input y Down/Back In menu: Go to next menu item on same level In alphanumeric input: Show next lowest number/letter x Up/Forward In menu: Go to previous menu ite
-
Sartorius AG Weender Landstrasse 94–108 37075 Goettingen, Germany Phone +49.551.308.0 Fax +49.551.308.32 89 www.sartorius.com Copyright by Sartorius AG, Goettingen, Germany. All rights reserved. No part of this publication may be reprinted or translated in any form or by any means without the prior written permission of Sartorius AG. The status of the information, specifications and illustrations in this ma
-
Service Regular servicing by a Sartorius techni- cian will extend the service life of your moisture analyzer and ensure its contin- ued weighing accuracy. Sartorius can offer you service contracts, with your choice of regular maintenance intervals ranging from 1 month to 2 years. The frequency of the maintenance intervals depends on operating condi- tions and user tolerance requirements. Repairs Repair work must be performed by trained service technicians. Any attempt by untrained persons to per- form repairs may le
-
49 Accessories (Options) Product Order No. External printer YDP03-0CE Printer expendables: – 5 paper rolls 69 06937 – ink ribbon 69 06918 Temperature adjustment set YTM03MA Exchangeable panels for drying unit (aluminum) YDS04MA Carrying case YDB05MA Software for recording moisture analysis values YMW02MA Expendables: – 80 disposable sample pans, ‡ 90 mm (aluminum) 69 65542 – 80 glass fiber filters (for liquid samples
-
No. Designation Order no. for replacement 1 Hinged cover with heating element 2 Disposable sample pan 6965542 (80 pcs.) 3 Pan support 69MA0092 4 Pan draft shield 69MA0093 5 Display 6 Keypad 7 On/off key 8 CF key (clear function) 9 ‘Enter’ key (confirm) No. Designation Order no. for replacement 10 ‘Down/Back’ key 11 ‘Up/Forward’ key 12 Print key 13 Leveling foot 14 Interface
-
Operating Instructions Sartorius Moisture Analyzer Model MA45 Electronic Moisture Analyzer 98648-010-37
-
Preparing a Sample Selecting a Sample § Select a representative part of the whole substance as a sample – a representative number of individual samples for quality control – samples which indicate a trend are sufficient for in-process control $ Homogenize the product before a sample is taken, if required, by: – mixing or stirring, – taking several samples from different areas of the product, or – taking sever
-
Safety Inspection If there is any indication that safe operation of the moisture analyzer is no longer warranted: § Turn off the power and disconnect the equipment from the wall outlet immediately > Lock the equipment in a secure place to ensure that it cannot be used for the time being Safe operation of the moisture analyzer is no longer ensured when: – there is visible damage to the moisture analyzer or power cable – the moisture analyzer no longer functions properly – the moisture analyzer has been stored for a relatively long period under unfavorabl
Questions, Opinions and Exploitation Impressions:
You can ask a question, express your opinion or share our experience of Sartorius MA150C-000115V1 device using right now.
Макс вес: 150 г. d: 0,001 г. Калибровка: внешняя. Дисплей: Жидкокристаллический. Габаритные размеры: 213×320×181 мм.Диапазон температуры сушки: 40-220 °C. Дискретность содержания влаги: 0,01 %. Шаг установки температуры: 1 °C.
(товар НДС не облагается)
- Описание
Описание
Анализатор влажности Sartorius MA-150 применяется при измерении влажности чувствительных к нагреву образцов, в частности, в пищевой промышленности. Температурная стабильность и вибростойкость достигается благодаря монолитной весовой ячейке, а использование керамического нагревателя гарантирует высокую скорость работы, стабильность и надежность работы в течение всего срока эксплуатации. Прибор имеет прочный корпус с камерой из нержавеющей стали. Энергозависимая память позволяет запоминать несколько программ с различными параметрами.
Особенности
- Внесен в Госреестр СИ РФ
- Сокращенное время анализа по сравнению с сушильным шкафом – 5-20 минут
- Три режима нагрева: стандартный – для большинства веществ, деликатный – для органических веществ, особо чувствительных к температуре и для сложных, композитных веществ – ступенчатый.
- Аналитическая монолитная весовая ячейка обеспечивает первоклассную температурную стабильность и вибростойкость
- Несколько режимов окончания сушки: полностью автоматический, по времени, полуавтоматический – для разнообразных веществ и всего множества задач, связанных с измерением влажности.
- 20 настраиваемых пользователем программ под различные вещества: достаточно выбрать заранее введенную программу, аккуратно разместить пробу, закрыть крышку и забыть об анализаторе влажности до полного завершения анализа
- Стекловолоконные подложки для подверженных быстрому спеканию веществ, одноразовые металлические кюветы.
Параметр | Значение |
Метод сушки | Инфракрасный нагрев/керамический нагреватель |
Максимальный весовой | 150 г |
диапазон | |
Цена деления | 1 г |
Воспроизводимость по | 0.01 % |
влажности | |
Вывод результатов анализа | % влажности, % сухого остатка, % RATIO, g residue |
Температурный диапазон, шаг установки | 40°C — 180°C, 1° |
Режим нагрева | Стандартное высушивание, деликатное высушивание, ступенчатое высушивание |
Режим анализа | Полностью автоматический, по времени 1-99 мин, |
полуавтоматический | |
Калибровочный вес | Внешний |
Тип нагревателя | Керамический или кварцевый |
Возможность сохранения программ в памяти | Да, 20 программ |
Сохранение данных в памяти | Сохранение результатов анализа до старта следующего |
Защита настроек паролем | Да |
Интерфейс | RS-232 |
Отчет в соответствии с требованиями GLP | Да |
Защитный чехол на клавиатуру | Да |
Вес | 5.5 кг |
Регистрационное удостоверение, инструкция по эксплуатации Анализатор влажности Sartorius MA-150
-
Page 1
This document courtesy of: Data Weighing Systems, Inc. Contact Us For immediate assistance call 1-800-750-6842… -
Page 2
Operating Instructions Sartorius Moisture Analyzer Model MA45 Electronic Moisture Analyzer 98648-010-37… -
Page 3
Intended Use The MA45 Moisture Analyzer can be The following features make the Symbols used for quick and reliable determi- moisture analyzer ideal for use as an The following symbols are used in these nation of the moisture content of inspection, measuring and test instru- instructions: materials of liquid, pasty and solid sub-… -
Page 4: Table Of Contents
Contents Intended Use Error Codes Contents Care and Maintenance Service Warning and Safety Information Repairs Cleaning Getting Started Safety Inspection Instructions for Recycling Equipment Supplied Overview General View of the Moisture Analyzer Specifications Connecting the Moisture Analyzer Accessories (Options) C Marking to AC Power Leveling the Moisture Analyzer Index…
-
Page 5: Warning And Safety Information
Improper use or handling, however, Fire Explosion can result in damage and/or injury. – Connect only Sartorius accessories and options, as these are optimally designed – Flammable or explosive substances To prevent damage to the equipment, for use with your moisture analyzer read these operating instructions –…
-
Page 6: Getting Started
Sartorius office or dealer. 2 hours at room temperature, leaving it Storage and Shipping Conditions unplugged from AC power. Afterwards,…
-
Page 7
Designation Order no. for replacement Designation Order no. for replacement Hinged cover with heating ‘Down/Back’ key element ‘Up/Forward’ key Disposable sample pan 6965542 (80 pcs.) Print key Pan support 69MA0092 Leveling foot Pan draft shield 69MA0093 Interface port Display Power socket Keypad On/off key Not shown:… -
Page 8: Connecting The Moisture Analyzer
– 230 volts: MA45-…230.. – 115 volts: MA45-…115.. If they do not match: Do not operate your moisture analyzer; contact your local Sartorius office or dealer. Use only – Original power cords – Power cords approved by a certified electrician/Sartorius service technician –…
-
Page 9: Leveling The Moisture Analyzer
Leveling the Moisture Analyzer Purpose: – To compensate for unevenness at the place of installation – This is necessary especially for testing liquid samples that need to be at a uniform level in the disposable sample pan Always level the moisture analyzer again any time after it has been moved to a different location.
-
Page 10: Operating Design
Operating Design Operation of the moisture analyzer follows a standardized “philosophy” which is described below. Keys The keys have following functions: e On/off key y Down/Back Entering Letters, Special Characters Turns the moisture analyzer on/off. When the operating menu is active: and Numbers: The moisture analyzer remains in Shows the next menu item on…
-
Page 11: Operation: Analysis And Test Functions
Operation Bar Graph: Function Line: There are two fundamentally different types of display: The bar graph is shown during moisture This line indicates the functions that can analysis if the “Initial Weight” option in the be activated by pressing “Enter” (e.g., –…
-
Page 12: Menu Operation
Menu Operation Parameter Settings Data Output This display is divided into two sections. The MA45 is equipped with an interface The MA45 has an operating menu for port for connecting either Line for Operating Status finding and setting parameters. Each menu has several levels.
-
Page 13: Error Codes
Error Codes Saving Data If you press a key that has no function, or which is blocked at a certain point Storing Parameter Settings in an application program, the error is The most recent configurations of indicated as follows: Setup menu parameters and drying programs are active when you switch on –…
-
Page 14: Configuring The Moisture Analyzer
Configuring the Moisture Analyzer Purpose Setting the Language You can configure your moisture ana- You can choose from 5 languages for lyzer to meet individual requirements by the display: entering user data and setting selected menu parameters in the Setup menu. –…
-
Page 15: Setting The Device Parameters
Setting the Device Parameters Factory Settings Entering or Changing the Password (DEVICE) Parameters: The factory-set configura- – The password (up to 8 characters) tions are identified by an “ο” in the list protects the following functions: Purpose starting on page 17. –…
-
Page 16
Additional Functions Reset the device parameters to the Print out parameter settings: factory settings: § If “Device parameters” are displayed: § Select the Setup menu: Press the r key Use the x or y key to select SETUP in the Function line and press u to confirm >… -
Page 17
Example: Set the display contrast to “3” Step Key (or Instruction) Display/Printout x or y 1. Select SETUP in the Function line C Auto. 0092 g SETUP PROGRAM TARE 2. Confirm SETUP 3. Select Device parameters 4. Confirm Device parameters 2 ƒ… -
Page 18: Device Parameters (Overview)
Device Parameters (Overview) ο Factory setting √ User-defined setting Setup – Password: empty Enter 8 Device characters max. parameters Interface ο SBI Baud rate 150 baud 300 baud ο 1,200 baud 2,400 baud 4,800 baud 9,600 baud 19,200 baud ο 7 bit Number of data bits 8 bit Parity…
-
Page 19: Device Information
Device Information § Print device information: Press the r key Purpose Display of device information > Printout (Example) Display Device Information ——————— § Select the Setup menu: 13.08.2001 13:02 Use the x or y key to select Model MA45-000230V SETUP in the Function line and press Ser.
-
Page 20: Basics Of Moisture Analysis
Basics of Moisture Analysis Purpose Material When deductive methods are used, the The MA45 moisture analyzer can be The moisture of a material is often mis- moisture content is indirectly deter- used for quick and reliable deter- takenly equated with its water content. mined.
-
Page 21
The Sartorius Moisture Analyzer cannot be used. Applications Guide will provide you with important information on the use of your moisture analyzer. -
Page 22: Preparation
Preparation Adjustment to an Existing Measuring System Before drying a sample, you must carry A moisture analysis method often replaces another drying method (e.g., oven drying out the following preparations: method), because it is simple to use and requires shorter analysis time. In this case, you should adapt this method to that of the moisture analyzer in order to obtain values –…
-
Page 23: Preparing A Sample
$ Use an appropriately designed tool for shredding a sample. Using disposable sample pans § Use only Sartorius disposable sample pans (inner diameter = 92 mm). Reusing sample pans leads to poor reproducibility of results for several reasons: – after cleaning, sample residues can still remain on the pan –…
-
Page 24
Applying a Sample to the Sample Pan § Apply the sample to the sample pan in a thin, even layer (height: 2 to 5 mm, weight: 5 to 15 g); otherwise: – a sample applied unevenly will result in a non-uniform distribution of heat –… -
Page 25: Operating The Moisture Analyzer
Operating the Moisture Analyzer Setting the Drying Parameters Purpose Adapt the moisture analyzer to the special requirements of products. Parameters can be configured individually for each program. Drying Parameters (Overview) ο Factory setting √ User-defined setting Program ο P1 memory (PROGRAM) Heating program ο…
-
Page 26
Features Standby Temperature Fully Automatic Mode: – Temperature set to defined value when The analysis ends as soon as the weight Number of Drying Programs the sample chamber is closed loss per 24 s is below an automatically MA45: 3 programs detected threshold. -
Page 27
——————— Printout of the Analysis Parameters § When the desired program is displayed: 13.08.2001 13:06 Press the r key Model MA45-000230V Ser. no. 13206969 Ver. no. 01-43-01 > Parameters are printed (see example on the right) ——————— Lines with more than 20 characters PROGRAM are truncated. -
Page 28
End of analysis: automatic (factory setting) Part A: Configuring the Drying Parameters Step Key (or Instruction) Display/Printout 1. Turn on the moisture analyzer Sartorius logo is displayed Self-test runs y or x 2. Select the PROGRAM function C Auto. 0092 g SETUP… -
Page 29
Step Key (or Instruction) Display/Printout 8. Confirm the temperature 9. Select the Initial weight 2 ƒ y parameter 10. Confirm the Initial weight parameter 11. Select the On setting and confirm 3 ƒ x 12. Enter the target weight 2.0 13. -
Page 30
Part B: Performing the Analysis Step Key (or Instruction) Display/Printout 1. Turn on the moisture analyzer Sartorius logo is displayed Self-test runs C Auto. 0092 g SETUP PROGRAM TARE 2. Prepare the sample: not necessary for corn starch 3. Open the sample chamber and position a new sample pan 4. -
Page 31
Step Key (or Instruction) Display/Printout ——————— 13.08.2001 16:01 Model MA45-000230V Ser. no. 13206969 Ver. no. 01-43-01 ——————— Prg. 1 Heating STANDARD Fin. temp. 130 ’C Stdby temp. Start W/STABIL. AUTOMATIC IniWt + 2.036 g The current moisture loss is displayed C Auto. -
Page 32: Heater Adjustment
“isoTEST” Calibration/Adjustment Functions The following functions are available Heater Adjustment Weighing System Settings in the SETUP menu, under the isoTEST Using 1-point and 2-point temperature menu item: adjustment and the YTM03MA temper- Calibration, Adjustment ature adjustment set (see “Accessories”) Purpose –…
-
Page 33
External Calibration/ Adjustment with a Factory-Defined Weight Externally calibrate and adjust the weighing system using a 30-g calibration weight Step Key (or Instruction) Display/Printout x or y 1. Select SETUP in the Function line C Auto. 0092 g SETUP PROGRAM TARE 2. -
Page 34
——————— 13.08.2001 16:15 Model MA45-000230V *) Date and time shown only when Ser. no. 99992581 using the YDP02 or YDP03 printer Ver. no. 01-43-01 from Sartorius ——————— External calibration W-ID Nom. 30.000 g Diff. — 0.002 g ——————— 13.08.2001 16:16… -
Page 35
13.08.2001 16:15 Model MA45-000230V *) Date and time shown only when Ser. no. 13206969 using the YDP02 or YDP03 printer Ver. no. 01-43-01 from Sartorius ——————— External calibration W-ID Nom. 30.000 g Diff. — 0.002 g External adjustment completed Diff. -
Page 36: Hardware Tests
Hardware Tests The following device elements can be tested: Purpose Hardware tests are performed to check – SBI communication whether the system communication – Heater with internal and external devices functions properly. These tests are not elementary hardware tests. Activating Hardware Tests Step Key (or Instruction) Display/Printout…
-
Page 37
Testing SBI Communication Step Key (or Instruction) Display/Printout 1. Prepare the test connector for the RS-232 port Connect TƒD (pin 2) with (see “Pin Assignment Chart”) RƒD (pin 3) 2. Select “Hardware tests” see above 3. Select “Test interfaces” function and confirm 4. -
Page 38: Data Output
Data Output There are three options for data output: – Output to the moisture analyzer – Output to an external printer (YDP02-0CE or YDP03-0CE) – Output to a peripheral device (e.g., a computer) via the interface port Output to the Moisture Analyzer Display (Weights and Calculated Values) The display is divided into 9 sections.
-
Page 39
Busy Symbol The “Busy” symbol is shown here when the moisture analyzer is processing a function activated by pressing a key. Plus/Minus Sign Any plus or minus sign required for a displayed value is shown here. Measured Value/Result This line shows: 5234 –… -
Page 40: Interface Port
Factory Settings for the SBI Operating Mode Parameters: RS-232 cables purchased from other manufacturers often have incorrect pin Transmission rate: 1,200 baud assignments for use with Sartorius Number of data bits 7 bits weighing systems! Be sure to check the pin assignments against the chart below…
-
Page 41: Data Output Format
Data Output Format Output Format with 20 Characters + CR LF You can output the values, displayed in the corresponding line for moisture 9 10 11 12 13 14 15 16 17 18 19 20 21 22 analysis values, and the weight unit U CR LF with or without a data ID code.
-
Page 42
During data communication between The moisture analyzer SBI interface Once XOFF has been received, it pre- the moisture analyzer and an on-line (Sartorius Balance Interface) has vents further transmission of characters. device (computer), messages consisting transmit and receive buffers. You can… -
Page 43: Pin Assignment Chart
Pin Assignment Chart Female Interface Connector: Pin Assignment Chart: 25-position D-Submini (DB25S) with Pin 1: Signal ground screw lock hardware for cable gland Pin 2: Data output (TƒD) Pin 3: Data input (RƒD) Male Connector Used: Pin 4: Signal return (TƒD/RƒD) (please use connectors with the Pin 5: Clear to send (CTS) same specifications):…
-
Page 44: Cabling Diagram
Cabling Diagram – Diagram for interfacing a computer or other peripheral device to the moisture analyzer using the RS-232/V24 standard and cables up to 15 m (50 ft.) long Moisture analyzer Peripheral device 25-pin 9-pin – – Type of cable: AWG 24 specification…
-
Page 45
External device not ready to send Transmit XON, then CTS (interface handshake interrupted; XOFF, CTS) Print fct. blocked ERR 53 No temperature compensation Contact the Sartorius Service Center No meas. value ERR 101 Key is stuck Release key Key pressed when switching on… -
Page 46
– Increase the sample quantity Insufficient heat output – Clean the protective glass or temperature sensor Place of installation exposed to – Change the place of installation interfering factors (vibration, etc.) If any other errors occur, please contact your local Sartorius Service Center… -
Page 47
Care and Maintenance Service Cleaning the Heating Unit and Regular servicing by a Sartorius techni- Temperature Sensor § Open the hinged cover cian will extend the service life of your moisture analyzer and ensure its contin- ued weighing accuracy. Sartorius can… -
Page 48
In this case, notify your nearest Sartorius Service Center. Maintenance and repair work may be performed only by service technicians who are authorized by Sartorius and who: – have access to the required mainte- nance manuals – have attended the relevant service… -
Page 49
Overview Specifications Model MA45 Dryer functions: Heating element Ceramic IR heater Temperature range 40 – 230°C Temperature increments adjustable in 1°C increments Temperature adjustment with YTM03MA temperature adjustment set Weighing Functions: Weighing capacity 45 g Readability 1 mg, 0.01% moisture content Repeatability, average (%) sample weight 1–5 g: 0.2% sample weight 5 g: 0.05%… -
Page 50
Calibration weight, 30 g (± 0.3 mg) YSS43 Standard Operating Procedure (SOP)YSL02A For Working with a Moisture Analyzer (For Quality Assurance Documentation) Spare Parts Dust cover for keypad 6960MA02 Forceps 69MA0072 Other spare parts can be ordered directly through your local Sartorius Service. -
Page 51
Important Note: and laboratory use The operator shall be responsible for Part 1: any modifications to Sartorius equip- General requirements ment and must check and, if necessary, Part 2-010: correct these modifications. On request, Particular requirements for laboratory… -
Page 52
Index Page Page Accessories 2, 4, 31, 49 Keypad 2, 5 Acoustic signal 14, 17 Keys: description Adjustment to an existing measuring system Language: setting 8, 13 Adjustment 31, 44 Leveling the moisture analyzer Alphanumeric input MA30 mode 24, 48 Basics Maintenance C marking… -
Page 53
General Password Enter/Change Password § Select the password function: § Select the Setup menu: Press the u key Use the x or y key to select SETUP in the Function line and press > Password prompt is displayed, together u to confirm with current password §… -
Page 54
Brief Instructions Sartorius MA45 y Down/Back Overview of Key Functions e On/Off Key In menu: Go to next menu item on Turn moisture analyzer on/off; same level analyzer remains in standby mode when In alphanumeric input: Show next lowest switched off… -
Page 55
Moisture Analysis Add sample: Switch analyzer on: e key Close sample chamber: Select drying program: Start analysis: START function and u key Select PROGRAM: or directly after closing sample chamber y and u keys Drying in progress: wait Use the y and u keys to select program P1, P2 or P3 Read off result: Cancel password prompt, if necessary: c key… -
Page 56
Configuration (SETUP) Language Activate menu: Use the y and u keys to German English select the SETUP function; confirm U.S. mode Select desired menu level: y and x keys Français Italiano Exit menu level or exit SETUP: c key Español Select menu item and setting: Device parameters y, x and u keys… -
Page 57
Drying Parameters (PROGRAM) Heating program Activate menu: Use the y and u keys to Standard drying MA30 mode select the PROGRAM function; confirm Select program P1, P2 or P3 Standby temperature Exit menu level or exit PROGRAM: Initial weight c key Start analysis Select menu item and setting: With stability when Enter is pressed… -
Page 58
Sartorius AG Weender Landstrasse 94–108 37075 Goettingen, Germany Phone +49.551.308.0 Fax +49.551.308.3289 www.sartorius.com Copyright by Sartorius AG, Goettingen, Germany. All rights reserved. No part of this publication may be reprinted or translated in any form or by any means without the prior written permission of Sartorius AG.
Sartorius MA150C-000115V1 Operating Instructions Manual
Download Operating instructions manual of Sartorius MA150 Measuring Instruments for Free or View it Online on All-Guides.com. This version of Sartorius MA150 Manual compatible with such list of devices, as: MA150C-000115V1, MA150C-000115V2, MA150C-000230V1, MA150C-000230V2, MA45
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58
For immediate assistance call
1-800-750-6842
This document courtesy of:
Data Weighing Systems, Inc.
Contact Us