Schiller at 2 plus инструкция по эксплуатации

Schiller AT-2plus Manual

  • Contents

  • Table of Contents

  • Bookmarks

Related Manuals for Schiller AT-2plus

Summary of Contents for Schiller AT-2plus

  • Page 1
    SCHILLER AT-2plus 6-Channel ECG Unit AT-2plus 6-Channel ECG unit SCHILLER AG Altgasse 68 CH-6341 Baar, Switzerland Phone: + 41 41 766 42 42 Fax: + 41 41 761 08 80 Home page: http://www.schiller.ch/…
  • Page 2
    SERVICE HANDBOOK Issue d May 2002 What’s in this book The service philosophy for the AT-2plus Memory is fault finding to module level. The purpose of this book is to provide all the information necessary to enable the service engineer to efficiently locate and replace a faulty module.
  • Page 3
    The full technical specification of the AT-2plus Memory is given in this chapter. Chapter 8 — Glossary This chapter explains all the acronyms and signal titles used in this book and in the AT-2plus Memory circuit diagrams. Circuit Diagrams & Board layouts The circuit diagrams and component layouts are provided for all boards.
  • Page 4: Table Of Contents

    SCHILLER AT-2plus 6-Channel ECG Unit Chapter 1 Operating Elements SERVICE HANDBOOK Issue d May 2002 Chapter 1 Operating Elements Contents Introduction List of Symbols Location & Power Switching On and Off The Keyboard LCD Screen AT-2plus Short Form Operating Instructions…

  • Page 5: Introduction

    Operating Elements Introduction The CARDIOVIT AT-2plus is a 6-channel ECG recorder with all (12) ECG signals simultaneously processed to provide instant ECG recordings. Two automatic recording modes can be individually preset to enable one button ECG recording of preferred print formats.

  • Page 6: Location & Power

    Potential Equalisation If the AT-2plus is used in conjunction with other patient connected equipment, we recommend that the potential equalisation stud on the rear of the unit is connected to the hospital/ building common ground with the yellow/green ground cable (Part-no.

  • Page 7: The Keyboard

    SCHILLER AT-2plus 6-Channel ECG Unit Chapter 1 SERVICE HANDBOOK Issue d May 2002 Operating Elements The Keyboard Page 1.4…

  • Page 8
    In the `Born` (date of birth field), only the patients year of birth need be entered if desired. When two digits are entered (patients year of birth), the AT-2plus calculates the age of the patient according to the year entered. When the full DOB is entered, the age is calculated precisely.
  • Page 9: Lcd Screen

    SCHILLER AT-2plus 6-Channel ECG Unit Chapter 1 SERVICE HANDBOOK Issue d May 2002 Operating Elements LCD Screen 1. Current Heart Rate (averaged over 4 beats and refreshed every 2 seconds). The HR is also given on a manual printout. Note that with an auto mode printout the HR is averaged over the full 10 seconds of the recording.

  • Page 10: At-2Plus Short Form Operating Instructions

    SCHILLER AT-2plus 6-Channel ECG Unit Chapter 1 Operating Elements SERVICE HANDBOOK Issue d May 2002 AT-2plus Short Form Operating Instructions Automatic ECG Recording • Prepare skin, hook up patient. • Switch unit on, press ON • Press and enter patient data.

  • Page 11: Modes Of Operation

    SCHILLER AT-2plus 6-Channel ECG Unit Chapter 1 SERVICE HANDBOOK Issue d May 2002 Operating Elements Modes of Operation Automatic Mode Automatic Mode provides a printout giving 10 seconds of ECG recording of all 12 leads with a choice of 2 different formats.

  • Page 12: Automatic Mode

    SCHILLER AT-2plus 6-Channel ECG Unit Chapter 1 Operating Elements SERVICE HANDBOOK Issue d May 2002 Automatic Mode In automatic mode, a full 12-lead ECG is printed in one of two predefined formats with a sensitivity of 10 mm/mV. These two formats are selected by the user to suit his specific needs and requirements.

  • Page 13: Manual Mode

    SCHILLER AT-2plus 6-Channel ECG Unit Chapter 1 SERVICE HANDBOOK Issue d May 2002 Operating Elements Manual Mode Manual mode provides a direct printout of the real-time ECG with full control of parameter selection. To start the manual recording of a real-time ECG, press the MANUAL Printout key…

  • Page 14
    SCHILLER AT-2plus 6-Channel ECG Unit Chapter 1 Operating Elements SERVICE HANDBOOK Issue d May 2002 Manual Mode (cont.) Chart Speed Select speed 5, 10, 25 or 50mm/s by means of the SPEED keys: 5/10 Notes: Key 7 is a toggle key -press once and 5 is selected, press a second time and 10mm/s is selected.
  • Page 15: Settings

    Example If you want to reset your AT-2plus to the basic default settings, the key sequence given on page 14 is ALT; 0; 6; 6. STOP. On the following pages the programmable parameters and the programming sequences are described in detail.

  • Page 16
    SCHILLER AT-2plus 6-Channel ECG Unit Chapter 1 Operating Elements SERVICE HANDBOOK Issue d May 2002 Settings (cont.) The defined formats and settings for your unit can be checked as follows: ALT — 0 — 1 — 1 A printout of the defined settings will be produced and gives the following information, depending…
  • Page 17
    SCHILLER AT-2plus 6-Channel ECG Unit Chapter 1 SERVICE HANDBOOK Issue d May 2002 Operating Elements Settings (cont.) Interpretation settings: N/A:+/- ‘normal/abnormal’ is written (+) or suppressed (-) U:+/- ‘unconfirmed report’ is written (+) or suppressed (-) A30:+/- patient age is assumed to be < 30 (-) or >30 (+)
  • Page 18
    SCHILLER AT-2plus 6-Channel ECG Unit Chapter 1 Operating Elements SERVICE HANDBOOK Issue d May 2002 Settings (cont.) Language The language can only be set with the SWUP program. User Identification The user identification is printed on all recordings. The user ID can be the department, doctor or hospital etc.
  • Page 19
    SCHILLER AT-2plus 6-Channel ECG Unit Chapter 1 SERVICE HANDBOOK Issue d May 2002 Operating Elements Settings (cont.) Filters There are three different filters which can be set individually as follows: • Baseline filter • Mains filter • Myogram filter Baseline Filter The digital Baseline filter suppresses excessive baseline drifts.
  • Page 20
    SCHILLER AT-2plus 6-Channel ECG Unit Chapter 1 Operating Elements SERVICE HANDBOOK Issue d May 2002 Settings (cont.) Myogram Filter The Myogram filter suppresses disturbances caused by strong muscle tremor. The set value will be the new upper limit of the frequency range as soon as the FILTER key is pressed on or programmed as default when the unit is switched on.
  • Page 21
    All ECG traces are centred dynamically for optimal use of paper width. Auto-Centering OFF ECG traces are set to a fixed baseline position and may possibly overlap. The Standard and Cabrera lead groups available for the AT-2plus are: Lead Groups Standard Cabrera -aVR…
  • Page 22
    SCHILLER AT-2plus 6-Channel ECG Unit Chapter 1 Operating Elements SERVICE HANDBOOK Issue d May 2002 Settings (cont.) Acoustic QRS Indication The acoustic QRS beep can be switched on or off at any time by pressing the QRS key Time / Date…
  • Page 23: Automatic Mode (Ecg) Settings

    Operating Elements Automatic Mode (ECG) Settings Two separate Auto formats can be defined for the AT-2plus. When defining auto format 1 the key sequence ALT `1` precedes the setting. When defining auto format 2 the key sequence ALT `2` precedes the setting.

  • Page 24
    SCHILLER AT-2plus 6-Channel ECG Unit Chapter 1 Operating Elements SERVICE HANDBOOK Issue d May 2002 Automatic Mode (ECG) Settings (cont.) Average Cycles The Average cycles are defined as follows: Note: Lead selection for the rhythm lead(s) are defined on page 25.
  • Page 25
    Confirm the selection by pressing the STOP key STOP Full details of the interpretation option are given in the SCHILLER ECG Measurement and Interpretation booklet (Art. No. 2.510 179). Interpretation Settings (C version only) The interpretation settings enable the user to determine whether or not certain comments will be added to the interpretation statements on the ECG printout.
  • Page 26
    SCHILLER AT-2plus 6-Channel ECG Unit Chapter 1 Operating Elements SERVICE HANDBOOK Issue d May 2002 Automatic Mode (ECG) Settings (cont.) Selecting Rhythm Leads The rhythm leads are printed out as defined. Two separate rhythm leads can be selected. The following formats can be set:…
  • Page 27: Memory And Data Transmission Option

    SCHILLER AT-2plus 6-Channel ECG Unit Chapter 1 SERVICE HANDBOOK Issue d May 2002 Operating Elements Memory and Data Transmission Option WARNINGS & CAUTIONS WHEN NON-MEDICAL DEVICES ARE CONNECTED TO THE RS-232 INTERFACE ENSURE THAT BOTH UNITS ARE SECURELY CONNECTED TO THE SAME EARTH POTENTIAL.

  • Page 28: Erasing Memory Files

    LINE transmission; sending over the telephone system requires a modem and this form of sending is termed MODEM. Transmission Settings The speed settings options for the AT-2plus are as follows: Serial Communication Interface Transmission Entry Key Sequence…

  • Page 29: Modem Transmission

    Check all settings in the SEMACOMM program (baud rate; parity — none; stop bit — 2; time between blocks, records — 100ms). • Check that the transmission speed is the same in both the AT-2plus and the SEMACOMM program. •…

  • Page 30: Line Transmission

    Check all settings in the SEMACOMM program (baud rate; parity — none; stop bit — 2; time between blocks, records — 100ms). • Check that the transmission speed is the same in both the AT-2plus and the SEMACOMM program. •…

  • Page 31: Software Updating Via Rs-232

    SWUP Programme for Win 95 ( 2 disks or Zip-file from Mailbox) Procedure 1. Connect the AT-2plus Memory with the RS-232 cable to any free COM port on your PC 2. Switch the PC ON and go to Win 95.

  • Page 32: Care & Maintenance

    CLEANING LIQUID OR STERILIZE WITH HOT WATER, STEAM, OR AIR. The casing of the AT-2plus can be cleaned with a soft damp cloth on the surface only. Where necessary a domestic non-caustic cleaner can be used for grease and finger marks.

  • Page 33
    SCHILLER AT-2plus 6-Channel ECG Unit Chapter 1 SERVICE HANDBOOK Issue d May 2002 Operating Elements Care & Maintenance (cont.) Cleaning the Patient Cable Align the leads in such a way as to prevent anyone stumbling over them or any damage caused by the wheels of instrument trolleys.
  • Page 34: Replacing The Recording Paper

    Return the paper tray cover in position and press firmly until secure. • Press the STOP key to transport the paper to the start position. Note: SCHILLER can only guarantee perfect printouts when SCHILLER original chart paper or chart paper of the same quality is used. Page 1.31…

  • Page 35: Thermal Paper Handling

    Operating Elements Thermal Paper Handling The thermal paper used in the AT-2plus requires slightly different handling to normal paper as it can react with chemicals and to heat. However, when the following points are remembered, the paper will give reliable results: The following points apply to both storage, and when archiving the results.

  • Page 36
    SCHILLER AT-2plus 6-Channel ECG Unit Chapter 2 SERVICE HANDBOOK Issue d May 2002 Functional Overview Chapter 2 Functional Overview Contents Introduction MK 14 — 10 Main Board Power Supply Program and ECG Memory Serial EEPROM Thermal Print Head Controller Printer Timing…
  • Page 37: Introduction

    SERVICE HANDBOOK Issue d May 2002 Introduction This chapter provides a functional overview of the AT-2plus electronics. The aim of this overview is to enable the service engineer to identify processing paths in order to help identify possible faulty modules. A functional block diagram supports the text.

  • Page 38
    SCHILLER AT-2plus 6-Channel ECG Unit Chapter 2 SERVICE HANDBOOK Issue d May 2002 Functional Overview Page 2.3…
  • Page 39: Mk 14 — 10 Main Board

    Chapter 2 SCHILLER AT-2plus 6-Channel ECG Unit Functional Overview SERVICE HANDBOOK Issue d May 2002 MK 14 — 10 Main Board Power Supply The mains supply is full wave rectified to produce an unregulated dc supply of approximately 30 V (+U). This voltage is used by a switched voltage generator to produce +UD (13.5V). +UD charges the battery when mains is connected.

  • Page 40: Power On Reset

    SCHILLER AT-2plus 6-Channel ECG Unit Chapter 2 SERVICE HANDBOOK Issue d May 2002 Functional Overview MK 14-10 Main Board (cont.) Power On Reset The Power on reset circuit controls the master reset of the CPU. This circuit has two functions as follows: •…

  • Page 41: Mk 14-11 Rs-232 Interface Board

    / serial and serial / parallell conversion for the transmission and reception of data and provide signal level compatibility with RS-232 standard. External Modem An external modem can be connected to the RS-232 output from the AT-2plus and be used for transmitting memory contents over a telephone line. Page 2.6…

  • Page 42: Top Assembly

    SCHILLER AT-2plus 6-Channel ECG Unit Chapter 2 SERVICE HANDBOOK Issue d May 2002 Functional Overview Top Assembly LCD Screen The LCD power supply produces the high voltage for the LCD backlight and the contrast voltage. The ECG signals are stored in a video RAM and the LCD controller converts the data to the proper form for the LCD screen.

  • Page 43
    Chapter 2 SCHILLER AT-2plus 6-Channel ECG Unit Functional Overview SERVICE HANDBOOK Issue d May 2002 Page 2.8…
  • Page 44
    SCHILLER AT-2plus 6-Channel ECG Unit Chapter 3 SERVICE HANDBOOK Issue d May 2002 Fault Diagnosis Chapter 3 Fault Diagnosis Contents Introduction General Check Procedures Printer Check Print Head Alignment and Print Head Tension RS-232 Test with Test Plug Memory and Data Transmission Check…
  • Page 45: Introduction

    SERVICE HANDBOOK Issue d May 2002 Introduction The AT-2plus is designed to be simple to use and simple to service: the service philosophy of the AT-2plus is module replacement and not board repair. The purpose of this chapter is to provide fault-finding procedures that will quickly and efficiently identify a fault to a specific module.

  • Page 46: General Check Procedures

    AT-2plus functional areas. The instructions given here are guides to the basic functions. If more operating information is required (general settings, comprehensive menu guides etc.) please refer to Chapter 1 in this publication or the relevant User Manual for the software version applicable.

  • Page 47: Printer Check

    Chapter 3 SCHILLER AT-2plus 6-Channel ECG Unit Fault Diagnosis SERVICE HANDBOOK Issue d May 2002 Printer Check To check the printer and to ensure that every pixel is operational, a built-in printer test is provided. To carry out the printer check press: ALT — MAN START A test printout is given.

  • Page 48: Test With Test Plug

    To carry out this test proceed as follows: Note: If the SCHILLER test plug is not available a test plug can easily be fabricated from a standard 9-pin D-type connector (female). Shortcircuit the following pins: •…

  • Page 49: Memory And Data Transmission Check

    • An RS-232 cable assembly, Art.No. 2.310 159 for connecting the RS-232 interface on the AT-2plus with the COM port of the PC. • A patient simulator attached to the patient connector on the AT-2plus. Test Procedure Command Screen Display…

  • Page 50
    SCHILLER AT-2plus 6-Channel ECG Unit Chapter 4 SERVICE HANDBOOK Issue d May 2002 Module Removal and Replacement Chapter 4 Module Removal and Replacement Contents Introduction Safety Notices Physical Overview Exploded View Base Assembly and Printer Exploded View Paper Feed Opening and Closing the Case Printer Tray Assembly and Thermal printer 4.10…
  • Page 51
    Chapter 4 SCHILLER AT-2plus 6-Channel ECG Unit Module Removal and Replacement SERVICE HANDBOOK Issue d May 2002 Remove top assembly In Base assembly access to : Mains trans- Paper tray Board Battery former and assembly MK 14-10 pack connector and motor…
  • Page 52: Introduction

    SCHILLER AT-2plus 6-Channel ECG Unit Chapter 4 SERVICE HANDBOOK Issue d May 2002 Module Removal and Replacement Introduction This Chapter provides an overview of the procedures to remove and replace the modules that are spared at service level. The instructions given in this chapter are autonomous, with each module containing the following: •…

  • Page 53: Safety Notices

    TOP ASSEMBLY REMOVED AND WITH MAINS CONNECTED. WHEN CARRYING OUT THESE PROCEDURES BEWARE THAT POTENTIALLY LETHAL VOLTAGES ARE PRESENT. CAUTIONS THE AT-2PLUS CONTAINS STATIC SENSITIVE CMOS COMPONENTS; OBSERVE ANTISTATIC PRECAUTIONS: WHEN CARRYING OUT ANY MAINTENANCE PROCEDURES ALWAYS PLACE THE UNIT ON AN EARTHED ANTISTATIC MAT.

  • Page 54: Physical Overview

    Module Removal and Replacement Physical Overview The AT-2plus unit is enclosed in a two part, medical standard, moulded plastic case. The top part contains the keyboard and the LCD screen with the base section containing all the electronics of the unit, the RS-232 interface, the thermal printer, the paper tray, the battery and mains transformer.

  • Page 55: Exploded View Base Assembly And Printer

    Chapter 4 SCHILLER AT-2plus 6-Channel ECG Unit Module Removal and Replacement SERVICE HANDBOOK Issue d May 2002 Exploded View Base Assembly and Printer Page 4.6…

  • Page 56: Exploded View Paper Feed

    SCHILLER AT-2plus 6-Channel ECG Unit Chapter 4 SERVICE HANDBOOK Issue d May 2002 Module Removal and Replacement Exploded View Paper Feed Page 4.7…

  • Page 57: Opening And Closing The Case

    Prerequisites • The unit must be placed on an antistatic mat and antistatic precautions observed when any maintenance is carried out on the AT-2plus. The room temperature should be between 18 and 28 degrees. • THE WARNINGS AND CAUTIONS AT THE BEGINNING OF THE CHAPTER MUST BE OBSERVED.

  • Page 58
    SCHILLER AT-2plus 6-Channel ECG Unit Chapter 4 SERVICE HANDBOOK Issue d May 2002 Module Removal and Replacement Opening and Closing the Case Top Assembly Replacement To replace the Top Assembly proceed as follows: 1. Check that all boards and components are firmly secured. Check for loose screws. Ensure that no screws or foreign bodies are loose in the bottom of the case.
  • Page 59: Printer Tray Assembly And Thermal Printer

    Chapter 4 SCHILLER AT-2plus 6-Channel ECG Unit Module Removal and Replacement SERVICE HANDBOOK Issue d May 2002 Printer Tray Assembly and Thermal printer Prerequisite • The Warnings and Cautions at the beginning of the Chapter must be observed. • The Top Assembly must be removed as detailed previously. All external cable assemblies must be disconnected.

  • Page 60: Thermal Printer Replacement

    SCHILLER AT-2plus 6-Channel ECG Unit Chapter 4 SERVICE HANDBOOK Issue d May 2002 Module Removal and Replacement Printer Tray Assembly and Thermal printer Thermal Printer Replacement. To replace the Thermal Printer proceed as follows: 1. Position the Printer in the paper tray/print assembly so that the printer mounting plate lips slot into the dedicated cutouts in the assembly;…

  • Page 61: Main Board Mk

    Chapter 4 SCHILLER AT-2plus 6-Channel ECG Unit Module Removal and Replacement SERVICE HANDBOOK Issue d May 2002 Main Board MK 14 — 10 Prerequisite • The Warnings and Cautions at the beginning of the chapter must be observed. • The top assembly must be removed and all external cable assemblies disconnected.

  • Page 62: Battery Pack

    SCHILLER AT-2plus 6-Channel ECG Unit Chapter 4 SERVICE HANDBOOK Issue d May 2002 Module Removal and Replacement Battery Pack Prerequisite • The Warnings and Cautions at the beginning of the Chapter must be observed. • The Top Assembly must be removed and all external cable assemblies disconnected.

  • Page 63: Keyboard

    Chapter 4 SCHILLER AT-2plus 6-Channel ECG Unit Module Removal and Replacement SERVICE HANDBOOK Issue d May 2002 Keyboard The keyboard comes as a complete assembly with the top casing. The part number of the keyboard is given in Chapter 6.

  • Page 64: Lcd Screen Board

    SCHILLER AT-2plus 6-Channel ECG Unit Chapter 4 SERVICE HANDBOOK Issue d May 2002 Module Removal and Replacement LCD screen board Prerequisite • The Warnings and Cautions at the beginning of the Chapter must be observed. • The Top Assembly must be removed as detailed previously. All external cable assemblies must be disconnected.

  • Page 65: Rs-232 Serial Interface Board Mk

    Chapter 4 SCHILLER AT-2plus 6-Channel ECG Unit Module Removal and Replacement SERVICE HANDBOOK Issue d May 2002 RS-232 Serial Interface Board MK 14 — 11 Prerequisite • The Warnings and Cautions at the beginning of the Chapter must be observed.

  • Page 66
    SCHILLER AT-2plus 6-Channel ECG Unit Chapter 5 SERVICE HANDBOOK Issue d May 2002 Adjustments Chapter 5 Adjustments Contents Introduction Safety Notices and Conditions Conditions Test Equipment Proprietary Test Equipment/Tools Main Board MK 14-10 Adjustment locations Battery Charge Voltage Precautions and Requirements…
  • Page 67: Introduction

    Specific warnings and cautions are given in the text where applicable. The part numbers for all replaceable modules are given in Chapter 6. The AT-2plus has the following adjustments: • Battery Charge Voltage (VR1 on the main board MK 14-10) •…

  • Page 68: Safety Notices And Conditions

    USE FORCE. NEVER STRAIN THE CABLE ASSEMBLIES. THE PROCEDURAL STEPS GIVEN FOR EACH MODULE MUST BE FOLLOWED IN THE ORDER GIVEN. THE OUTER SURFACES OF THE AT-2PLUS ARE SUSCEPTIBLE TO ABRASION DAMAGE. TO PREVENT SCRATCHING, ALWAYS PLACE ON A SOFT, NON-ABRASIVE CLOTH.

  • Page 69: Test Equipment

    ECG Simulator, e.g. Phantom 320 • Oscilloscope • Digital Multimeter • Standard tool kit with a selection of cross-bladed, flat-bladed and posi-drive screwdrivers, pliers and general tools • SCHILLER 10 lead patient cable Number 2.400070 (2. 400071 for USA) Page 5.4…

  • Page 70: Main Board Mk 14-10 Adjustment Locations

    SCHILLER AT-2plus 6-Channel ECG Unit Chapter 5 SERVICE HANDBOOK Issue d May 2002 Adjustments Main Board MK 14-10 Adjustment locations Page 5.5…

  • Page 71: Battery Charge Voltage

    Battery Charge Voltage Precautions and Requirements The unit must be placed on an antistatic mat and antistatic precautions observed when any maintenance is carried out on the AT-2plus. The room temperature should be between 18 and 28 degrees. Tools and Equipment •…

  • Page 72: Paper Mark Detector Check

    SCHILLER AT-2plus 6-Channel ECG Unit Chapter 5 SERVICE HANDBOOK Issue d May 2002 Adjustments Paper mark Detector Check Tools, Equipment and Material • Digital voltmeter • Small flat bladed screwdriver • 70% alcohol solution and clean lint free cloth Procedure To adjust the paper-mark Schmitt trigger sensitivity proceed as follows: •…

  • Page 73: Ecg Amplifier +2V, -2V And Pwm Ramp Time Adjustment

    • DISCONNECT THE MAINS SUPPLY CAUTIONS THE AT-2PLUS CONTAINS STATIC SENSITIVE CMOS COMPONENTS; OBSERVE ANTISTATIC PRECAUTIONS . PLACE THE UNIT ON AN EARTHED ANTISTATIC MAT. PERSONNEL MUST BE EARTHED WHEN HANDLING THE UNIT. EXERCISE CARE WHEN REMOVING AND REPLACING CONNECTORS. NEVER USE FORCE.

  • Page 74: Ecg Amplifier Reference Voltage

    SCHILLER AT-2plus 6-Channel ECG Unit Chapter 5 SERVICE HANDBOOK Issue d May 2002 Adjustments ECG Amplifier reference voltage • Switch the unit on and measure the voltage difference between the +2V reference and the — 2V reference on pins 1 and 7 of operational amplifier U5. Adjust trimmer VR2 to achieve a voltage difference of 4000 mV ±2mV.

  • Page 75: Service Screen

    Chapter 5 SCHILLER AT-2plus 6-Channel ECG Unit Adjustments SERVICE HANDBOOK Issue d May 2002 Service Screen The service screen provides information of the patient cable and electrodes and gives the value of certain reference voltages and important internal offset values. These values are for information only.

  • Page 76
    SCHILLER AT-2plus 6-Channel ECG Unit Chapter 5 SERVICE HANDBOOK Issue d May 2002 Adjustments Service Screen (cont.) Electrode dc offset This gives the voltage drop in the patient cable and can indicate any faults in the patient cable or patient electrode.
  • Page 77: Software Options / Updates

    Acceptance of the code is indicated by a series of beeps. CAUTION MORE THAN 10 ATTEMPTS TO ENTER THE INCORRECT CODE BLOCKS THE UNIT Installing Software Updates To update the operating software in the AT-2plus see instructions in Chapter 1 of this book. Page 5.12…

  • Page 78
    SCHILLER AT-2plus 6-Channel ECG Unit Chapter 6 SERVICE HANDBOOK Issue d May 2002 Spare Parts Chapter 6 Spare Parts Contents Ordering Information Service Department Spare Parts Page 6.1…
  • Page 79
    Your local representative stocks all the disposables and accessories available for the AT-2plus. In case of difficulty or to obtain the address of your local dealer, please contact the head office. Our staff will be pleased to help process your order or to provide any details for all SCHILLER products.
  • Page 80
    SCHILLER AT-2plus 6-Channel ECG Unit Chapter 6 SERVICE HANDBOOK Issue d May 2002 Spare Parts Spare Parts Description Part Number Mikroprocessor and power supply board MK 14-10 3.248 3BC Keyboard MK 14-3B 3.248 2BA RS-232 Interface board MK 14-11 3.248 5BA LCD Module 4.600 062…
  • Page 81
    Chapter 6 SCHILLER AT-2plus 6-Channel ECG Unit Spare Parts SERVICE HANDBOOK Issue d May 2002 Page 6.4…
  • Page 82
    SCHILLER AT-2plus 6-Channel ECG Unit Chapter 7 SERVICE HANDBOOK Issue d May 2002 Technical Data Chapter 7 Technical Data Contents Technical Data Available Configurations Page 7.1…
  • Page 83
    Chapter 7 SCHILLER AT-2plus 6-Channel ECG Unit Technical Data SERVICE HANDBOOK Issue d May 2002 Technical Data Technical data subject to change without notice. Dimensions: 400 x 330 x 100 mm Weight: 5.0 kg ( 5.35 kg with full paper tray)
  • Page 84
    25 to 95% (non condensing) Atmospheric pressure: 700 to 1060 hPa Control Panel: Rubber keys Available Configurations The CARDIOVIT AT-2plus is available in several versions: Standard Version: Unit with ECG recording and printout capabilities. Version C: Unit with additional ECG Interpretation program (including measurements).
  • Page 85
    Chapter 7 SCHILLER AT-2plus 6-Channel ECG Unit Technical Data SERVICE HANDBOOK Issue d May 2002 Page 7.4…
  • Page 86
    SCHILLER AT-2plus 6-Channel ECG Unit Chapter 8 SERVICE HANDBOOK Issue d May 2002 Glossary Chapter 8 Glossary Contents Introduction Acronyms Page 8.1…
  • Page 87
    The following list provides a glossary of the important signals and acronyms used in the circuit diagrams for the SCHILLER instruments. They will not all apply to the AT-2plus Memory. Only abbreviations that are specific to SCHILLER equipment are included here. General electrical and electronic abbreviations are not included.
  • Page 88
    SCHILLER AT-2plus 6-Channel ECG Unit Chapter 8 SERVICE HANDBOOK Issue d May 2002 Glossary Acronyms ..OS Offset signal (on the ECG amplifier). A(1…n) CPU Address Bus ALBEEP Alarm beeper signal to the audio amplifier. The frequency of this signal is about 1000 Hz.
  • Page 89
    Chapter 8 SCHILLER AT-2plus 6-Channel ECG Unit Glossary SERVICE HANDBOOK Issue d May 2002 Acronyms DTACK Transfer data acknowledge. Bus signal to acknowledge transfer of data. Outgoing serial data, turns modem ON. ECGI ECG in — serial ECG data to the CPU sent over the optical interface.
  • Page 90
    Motor ON — printer motor enable signal. NWTZ Mains supply. Non-maskable interrupt — interrupt for U47 (Schiller gate array) activated by the reset button. Off signal from the OFF key to switch off the power supply. Control signal derived from FLM (unity waveform 1/2 FLM frequency).
  • Page 91
    Upper column address strobe (for dynamic RAM). UD1, UD2 Upper data strobe — used for generating UOE and UWE. Upper data strobe — used on the Schiller gate array. UOE, USRAM Upper output enable — for static RAM. Page 8.6…
  • Page 92
    SCHILLER AT-2plus 6-Channel ECG Unit Chapter 8 SERVICE HANDBOOK Issue d May 2002 Glossary Acronyms Voltage rectified from the mains input and regulated to approximately +15 V. Upper write enable. Unregulated dc supply from mains (approximately 30V). +UBU Back-up voltage for the real time clock and static RAM.
  • Page 93
    Chapter 8 SCHILLER AT-2plus 6-Channel ECG Unit Glossary SERVICE HANDBOOK Issue d May 2002 Page 8.8…

Schiller AT 2 Plus

Schiller Cardiovit AT 2 Plus 6 канальный электрокардиограф, осуществляющий одновременную регистрацию всех 12 отведений, и таким образом обеспечивающий немедленную регистрацию ЭКГ.

Schiller Cardiovit AT 2 Plus

Schiller Cardiovit AT 2 Plus имеет следующие преимущества:

  • Небольшой вес и компактные размеры.
  • Распечатка формата А4 со встроенного термопринтера.
  • Встроенный аккумулятор для независимой работы от сети в течение 4 часов обычного использования или 300 распечаток на один заряд.
  • Большой жидкокристаллический экран для просмотра ЭКГ перед печатью.
  • Действие совершается простым нажатием клавиш.
  • Автоматический и ручной режимы регистрации.
  • Возможность выбора форматов печати.
  • Память ЭКГ с возможностью копирования ЭКГ.
  • Программа интерпретации (включая измерения) для взрослых и детей.
  • Буквенно-цифровая клавиатура для ввода данных пациента и замечаний врача.

Производитель Schiller Cardiovit AT 2 Plus

Производитель электрокардиографа Schiller Cardiovit AT 2 Plus – Фирма «SCHILLER AG». Страна производитель Швейцария.

Технические характеристики Schiller Cardiovit AT 2 Plus

Ниже указаны технические характеристики на аппарат ЭКГ Schiller Cardiovit AT 2 Plus.

Технические характеристики AT 2 Plus
Технические характеристики
Технические характеристики
Технические характеристики

Режимы работы Schiller Cardiovit AT 2 Plus

Режимы работы Schiller Cardiovit AT 2 Plus

Автоматический режим

Автоматический режим допускает печать десяти секунд ЭКГ, зарегистрированной по всем 12 отведениям, в двух различных форматах.

До регистрации в каждом из форматов можно запрограммировать следующее:

  • Формат отведения.
  • Скорость подачи бумаги.
  • При инсталлированной программе интерпретации возможен выбор таблицы измерений , усредненных комплексов с отметками и интерпретационными сообщениями для печати.

Ручной режим

Ручной режим допускает печать шести отведений, выбираемых и индицируемых на экране, в режиме реального времени.

В течение регистрации возможен следующий выбор:

  • Группы отведений.
  • Скорости подачи бумаги.
  • Чувствительности.
  • Миографического фильтра.

Подсоединение кабеля пациента

Набор принадлежностей электрокардиографа включает 10-проводный кабель пациента. Этот кабель подсоединяется к гнезду кабеля пациента на правой боковой панели прибора и закрепляется двумя винтами.

Электрокардиограф относится к классу СF. Это означает, что отведения и наложенные на пациента электроды для ЭКГ полностью изолированы и защищены от дефибрилляционного напряжения. Защита от дефибрилляционного напряжения гарантирована только в случае использования кабеля пациента производства фирмы SCHILLER (катал. номер 2.400070). Для записи электрокардиограммы нужно убедиться в том, что ни пациент, ни токопроводящие части наложенных на него электродов (включая нейтральный электрод), не контактируют с другим человеком или другими токопроводящими предметами, (даже если они заземлены).

Качество ЭКГ зависит от подготовки области наложения электрода и величины сопротивления между кожей и электродом. Для гарантии хорошего качества ЭКГ и минимизации сопротивления кожа/
электрод необходимо помнить о следующем:

  • Убедитесь, что в помещении тепло и пациент расслаблен.
  • Побрейте область наложения электрода до того, как очистить ее.
  • Очистите область наложения электрода спиртовым тампоном.
  • Первым прикрепите электрод С4 в пятом межреберье по среднеключичной линии.

Расположение электродов:

  • C1 в четвертом межреберье по правому краю грудины.
  • C2 в четвертом межреберье по левому краю грудины.
  • C3 между С4 и С2 на равном расстоянии.
  • C6 по левой среднеподмышечной линии на уровне С4
  • C5 между С4 и С6 на равном расстоянии.

Подсоединение кабеля ЭКГ

Подсоединение кабеля ЭКГ

Клавиатура

Ниже представлено описание функций клавиш клавиатуры Schiller Cardiovit AT 2 Plus.

Клавиатура Schiller Cardiovit AT 2 Plus
  1. Печать дополнительной копии  если текущая регистрация произведена в автоматическом режиме (из памяти). Для получения дополнительной распечатки в формате 2 , сначала нажмите клавишу АЛЬТ.
  2. Визуализация / ввод данных пациента. При выведенных на экран данных пациента нажмите клавишу снова для возвращения к экрану ЭКГ.
  3. Включение / выключение миографического фильтра. Отсекающая частота устанавливается в “Установках”.
  4. Эти клавиши служат для изменения групп отведений, представленных на экране.
  5. Установки автоматической чувствительности для печати (только в автоматическом режиме) для подбора оптимальной величины сигнала (5мм/мВ или 10мм/мВ).
  6. Верхняя часть клавиш имеет обозначения 5, 10 и 20 для выбора чувствительности ЭКГ на экране и на (ручной) печати. Возможные установки чувствительности 5, 10, 20 мм/мВ.
  7. Верхняя часть клавиш имеет обозначения 5, 10, 25 и 50, которые являются установками скорости ЭКГ на экране и (ручной) печати. Скорость экрана может быть установлена на 25 или 50 мм/с. Скорость распечатки в ручном режиме может быть 5, 10, 25 или 50 мм/с. Установками скорости 5 и 10 мм/с обладает одна клавиша.
  8. Верхний символ отвечает за включение / выключение бипера QRS.
  9. Удаление последнего действия / символа.
  10. Выключение прибора.
  11. Включение прибора.
  12. В ручном режиме  запуск непрерывной печати ЭКГ – до нажатия клавиши СТОП.
  13. Регистрации ЭКГ в автоматическом режиме (формат 1). Нажмите АЛЬТ перед клавишей АВТО для получения формата 2.
  14. Остановка печати / подтверждение (новых) установок.
  15. Клавиша АЛЬТ  инициирует установку и выбирает второй формат для печати и регистрации автоматического режима.
  16. В режиме ЭКГ используйте клавиши вверх / вниз для настройки контрастности экрана. При вводе данных пациента используйте клавиши направо / налево для передвижения курсора в поле.
  17. Индикатор сети  горит при работе от сети.

Экран Schiller Cardiovit AT 2 Plus

Экран Schiller Cardiovit AT 2 Plus
  1. Текущая ЧСС (усредняются каждые 4 сокращения и обновляется каждые 2 секунды). ЧСС приводится и на распечатке в ручном режиме. Учтите, в автоматическом режиме ЧСС усредняется за каждые 10 секунд регистрации.
  2. Верхняя строка – текущие день недели, дата и время. Нижняя строка – системные сообщения.
  3. Верхняя строка  источник питания  сеть или аккумулятор. Когда мощность аккумулятора снижается, символ мигает Нижняя строка – АЛЬТ в этом поле означает, что клавиша АЛЬТ – нажата.
  4. Подсоединение электродов  если отведение мигает, сопротивление электрода слишком высоко. Электрод должен быть переустановлен.
  5. Индикатор отведений (текущих отведений, демонстрируемых экраном). Изменить группу отведений можно клавишами (1) и (2).
  6. Индикатор миографического фильтра: “ФИЛЬТР”  фильтр включен, нет индикации  фильтр выключен.
  7. (А) В этом поле свидетельствует о том, что выбрана установка автоматической чувствительности (активна только при авто режиме). Включение / выключение авто чувствительности клавишей (3).
  8. Чувствительность  5, 10, 20 мм/мВ. Изменение чувствительности с помощью клавиш (4), (5) и (6).
  9. Скорость  25 или 50 мм/с. Изменение скорости с помощью клавиш (8) и (9).

Установки Schiller Cardiovit AT 2 Plus

Все параметры устанавливаются с помощью кода. Этот код состоит из комбинации клавиш, которая всегда начинается с клавиши АЛЬТ, затем следует последовательность цифровых клавиш, и в заключении ввод подтверждается клавишей СТОП. Как только нажата клавиша ALT, клавиатура работает на функцию программирования.

При нажатии клавиши АЛЬТ символ ‘АЛЬТ’ появляется на экране. Функция АЛЬТ активна только 4 секунды. Если клавиши для программирования не нажаты в течении 4 секунд прибор возвращается в стандартный режим. Для активации режима программирования клавиша АЛЬТ должна быть нажата заново.

Установки запоминаются после нажатия клавиши STOP, и клавиатура освобождается для выполнения других функций. Однажды подтвержденные установки сохраняются даже тогда, когда прибор выключен.

Более подробные установки можно посмотреть в инструкции на аппарат. Скачать руководство пользователя можно в конце статьи.

Установка языка Schiller Cardiovit AT 2 Plus

Язык выбирается следующим образом:

Установка языка Schiller Cardiovit AT 2 Plus

Подтвердите выбор нажатием клавиши СТОП.

Установка даты и времени Schiller Cardiovit AT 2 Plus

Установка даты и времени Schiller Cardiovit AT 2 Plus осуществляется следующим образом:

Установка даты и времени

Техническое обслуживание Schiller Cardiovit AT 2 Plus

Прибор должен проходить технический контроль каждые 12 месяцев. Во время проверки должно быть выполнено и документировано следующее:

  • Визуальный осмотр прибора и кабеля пациента.
  • Тесты электрической безопасности по IEC 6011 и IEC 601225.
  • Тесты функционального состояния прибора согласно сервисному руководству.

Результаты тестов должны быть документированы.

Самотест

Самотестирование Schiller Cardiovit AT 2 Plus инициируется следующим образом:

Самотест

Очистка кабеля пациента

Кабель пациента не должен подвергаться излишним механическим нагрузкам. При отсоединении отведений, держитесь за штеккер, а не за кабель. Обеспечьте безопасные условия хранения и транспортировки прибора и кабеля.

Кабель можно мыть мыльной водой. Стерилизация, если требуется, должна проводиться только газом и без пара. Для дезинфекции протрите кабель стандартными средствами дезинфекции, применяющимися в Вашем учреждении.

Очистка термопечатающеего устройства

При частом использовании термоголовка может загрязняться, что может привести к ухудшению качества печати. Для предотвращения этого рекомендуется ежемесячно очищать термоголовку принтера.

Снимите крышку отсека бумаги. Термоголовка находится за планкой фиксации крышки отсека для бумаги. Мягкой тканью, слегка смоченной спиртом, осторожно протрите термоголовку.

Замена бумаги

Красная полоска по нижнему краю регистрирующей бумаги свидетельствует о том, что бумага заканчивается. Появление этой полоски указывает на то, что осталось около 8 страниц. Рекомендуется заменить бумагу сразу после появления полоски. В случае окончания бумаги начнет мигать индикаторная лампа, и процесс печати будет прерван.

Для замены бумаги проделайте следующее:

Замена бумаги Schiller Cardiovit AT 2 Plus
  • Захватите пальцами задвижку, крепящую отсек для бумаги, и потяните вверх.
  • В подтверждение правильности выполненных действий отсек, покрывающий бумагу, слегка отскочит влево.
  • Удалите отсек, затем оставшуюся бумагу.
  • Расположите в отделении новую упаковку бумаги. Проверьте, чтобы печатаемая (разлинованная) сторона располагалась «лицом» вверх, бумага должна выходить на переднюю панель прибора, огибая черный ролик.
  • Установите крышку отсека для бумаги на место для щелчка.
  • Для перемещения бумаги в начальную позицию нажмите клавишу СТОП.

Ошибки и неисправности Schiller Cardiovit AT 2 Plus

Ниже указаны неисправности аппарата ЭКГ Schiller Cardiovit AT 2 Plus и способы их устранения.

Прибор не включается, экран не функционирует.

Включен ли индикатор сети? Нет? Проверьте источник питания. Да? Проверьте контрастность экрана с помощью курсоров вверх / вниз. Если сеть в порядке, а экран так и не функционирует: Нажмите клавишу ВЫКЛ. Подождите несколько секунд и включите прибор снова. Если экран так и не реагирует, свяжитесь с представителем Schiller медицинское оборудование.

Наложение комплексов QRS.

Убедитесь, что автоматическое уменьшение чувствительности не отключено. Нажмите клавишу 1 мВ для переустановки изолинии. Проверьте контакт электродов.

«Шумные» кривые.

Проверьте контакт электродов. Наложите электроды заново. Убедитесь, что пациент расслаблен и ему тепло. Проверьте все установки фильтров. Включите миографический фильтр – проверьте частоту. Убедитесь, что установка сетевого фильтра соответствует напряжению сети.

После автоматического запуска нет распечатки.

Убедитесь в наличии бумаги. Проверьте установки – убедитесь, что, в авто формате ЭКГ выбрана, по крайней мере, одна позиция для печати. Запустите печать в ручном режиме, если принтер все еще бездействует  обратитесь к представителю SCHILLER.

Распечатка постепенно исчезает или «нечистая».

Убедитесь, что вставлена новая бумага от SCHILLER. Помните, что термочувствительная бумага для AT4 чувствительна к теплу и свету. Если бумага хранится не в оригинальной упаковке, при высокой температуре или она просто старая, то качество печати может ухудшиться. Убедитесь, что бумага вставлена правильно с отметкой в верху. Проверьте, защелкнут ли отсек для бумаги. При истечении некоторого времени чернила с печатной стороны бумаги могут сохраняться на головке принтера. Очистите головку принтера.

Нет распечатки интерпретационных сообщений или измерений.

Проверьте, доступны ли для печати опции интерпретация и измерения. Проверьте, чтобы выбранный формат включал установки измерения и интерпретации.

Клавиши не функционируют, экран завис.

Выключите прибор, подождите несколько секунд и включите снова.

Принадлежности и расходные материалы

Принадлежности и расходные материалы

Аккумулятор Schiller Cardiovit AT 2 Plus

Аккумулятор 12V-2,2Ah для ЭКГ аппарата Schiller Cardiovit AT 2 Plus.

Аккумулятор Schiller Cardiovit AT 2 Plus

Технические характеристики аккумулятора

  • Напряжение (В): 12;
  • Емкость (А.ч): 2,2;
  • Габариты (ДхШхВ): 178x34x60 мм;
  • Вес: 0.93 кг;
  • Срок службы: при t = 20 C — 5 лет;
  • Тип клем: Faston 187;
  • Применение: в циклическом или буферном режиме.

Скачать инструкцию на Schiller Cardiovit AT 2 Plus

Скачать инструкцию и другую документацию на электрокардиограф Schiller Cardiovit AT 2 Plus можно здесь.

Руководство пользователя ( user manual ) на русском языке Schiller Cardiovit AT 2 Plus скачать.

Регистрационное удостоверение Schiller Cardiovit AT 2 Plus скачать.

Сертификат соответствия Schiller Cardiovit AT 2 Plus скачать.

Свидетельство об утверждении типа средств измерений Schiller Cardiovit AT 2 Plus скачать.

Так же смотрите Schiller AT-101 12-канальный электрокардиограф, сфигмоманометр Schiller BR-102 plus.

SCHILLER AT-2plus 6-Channel ECG Unit

SERVICE HANDBOOK Issue d May 2002

AT-2plus

Memory

6-Channel ECG unit

Service Handbook

SCHILLER AG

Altgasse 68

CH-6341 Baar, Switzerland

Phone: + 41 41 766 42 42

Fax: + 41 41 761 08 80

Home page: http://www.schiller.ch/

May 2002

Article Number: 2.540014

i

SCHILLER AT-2plus 6-Channel ECG Unit

SERVICE HANDBOOK Issue d May 2002

AT-2plus Memory Service Handbook

Article Number 2. 540014d

Issue 1 :

June 1998

Issue 2 :

December 1999

Schematics: 2480BC, 2482CA, 2483BF, 2485BA

Issue c :

September 2001

Schematics:

2480BC, 2482CA, 2483DA, 2485BA

Issue d :

May 2002

Schematics:

2480BC, 2482CA, 2483DA, 2485BA

Associated Documents

Guide to the SCHILLER Interpretation

and Measurement Program E/ D/ F

Article No.

2. 510179

SCHILLER AT-2plus User Guide — English / German / French

Article No.

2. 510220

SCHILLER AT-2plus User Guide — Italian / Spanish / Portuguese

Article No.

2. 510221

SCHILLER AT-2plus User Guide Swedish

Article No.

2.510321

SCHILLER AT-2plus Short Form User Guide Swedish

Article No.

2.510245

© SCHILLERAG 2001

Windows™ is a trademark of Microsoft Corporation.

ii

SCHILLER AT-2plus 6-Channel ECG Unit

SERVICE HANDBOOK Issue d May 2002

Where to Obtain Service

WELCH ALLYN SCHILLER Inc., 7420 Carroll Road, San Diego, CA , US-92121-2334

USA

USA / Canada

Tel.: +1 858 635 6023

Fax : +1 858 635 6611

Home Page : www.welchallyn.com

SCHILLER Asia Pacific, 10 Jalan SS 3/33, Taman Universiti, 47300 Petaling Jaya, Selangor,

Asia Pacific

Malaysia

Tel.: + 603 7877 5336

Fax : + 603 7877 5744

Austria

SCHILLER HmbH, Kampmüllerweg 24, A-4044 Linz, Austria

Tel.: + 43 732 709 90

Fax : + 43 732 757 000

SCHILLER Medical S.A, BP 50, 19, Avenue de la Gare, F-67162 Wissembourg / Cedex,

France

France

Tel.: +33 3 88 63 36 00

Fax : +33 3 88 94 12 82

Germany (EU

SCHILLER Medizintechnik GmbH, Rudolf-Diesel Strasse 14, D-85521 Ottobrunn, Germany

authorized

Tel.: + 4989 629 981 0

Fax : + 4989 609 509 0

representative)

SCHILLER Healthcare India Pvt. Ltd.,D.C.Silk Mills Compound, ‘A’ Wing, 1st floor, 5,

India

Chunawala Estate, Kondivitta Lane, Andheri — Kurla Road, Andheri (E, Mumbai — 400 059,

India

Tel.: + 9122 826 3520

Fax : + 9122 826 3525

Italy

ESAOTE Spa (SCHILLER), Via di Caciolle 15, I-50127 Firenze, Italy

Tel.: + 39055 4229 201

Fax : + 39055 4229 208

Switzerland

SCHILLER Reomed AG, Riedstrasse 14, CH-8953 Dietikon, Switzerland

Tel.: +411 744 3000

Fax : + 411 740 3710

SCHILLER AG, Altgasse 68, CH-6341 Baar, Switzerland

All other

Tel.: + 4141 766 4242

Fax : + 4141 761 0880

countries

Home Page : www.schiller.ch

iii

SCHILLER AT-2plus 6-Channel ECG Unit

SERVICE HANDBOOK Issue d May 2002

Warranty

Terms of Warranty

The SCHILLER AT-2plus Memory is warranted against defects in material and manufacture for the duration of one year (as from date of purchase). Excluded from this guarantee is damage caused by an accident or as a result of improper handling. The warranty entitles free replacement of the defective part. Any liability for subsequent damage is excluded. The warranty is void if unauthorized or unqualified persons attempt to make repairs.

In case of a defect, contact your dealer or the manufacturer.

The manufacturer can only be held responsible for the safety, reliability, and performance of the apparatus if:

*assembly operations, extensions, readjustments, modifications, or repairs are carried out by persons authorized by him, and

*the AT-2plus Memory and approved attached equipment are used in accordance with the manufacturers instructions.

THERE ARE NO EXPRESS OR IMPLIED WARRANTIES WHICH EXTEND BEYOND THE WARRANTIES HEREINABOVE SET FORTH. SCHILLER MAKES NO WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE WITH RESPECT TO THE PRODUCT OR PARTS THEREOF.

iv

SCHILLER AT-2plus 6-Channel ECG Unit

SERVICE HANDBOOK Issue d May 2002

Safety Notices

WARNINGS & CAUTIONS

TO PREVENT ELECTRIC SHOCK DO NOT DISASSEMBLE THE UNIT. NO SERVICEABLE PARTS INSIDE. REFER SERVICING TO QUALIFIED PERSONNEL ONLY.

DO NOT USE THIS UNIT IN AREAS WHERE THERE IS ANY DANGER OF EXPLOSION OR THE PRESENCE OF FLAMMABLE GASES SUCH AS ANAESTHETIC AGENTS.

IF THE DISPLAY IS DAMAGED, A LEAKAGE OF FLUID MAY OCCUR. DO NOT INHALE THE VAPOUR FROM THIS FLUID AND AVOID CONTACT WITH MOUTH AND SKIN. IF CONTACT IS MADE, CLEAN CONTAMINATED AREA IMMEDIATELY WITH FRESH WATER.

THIS PRODUCT IS NOT DESIGNED FOR STERILE USE.

SWITCH THE UNIT OFF BEFORE CLEANING AND DISCONNECT FROM THE MAINS.

DO NOT, UNDER ANY CIRCUMSTANCES, IMMERSE THE UNIT OR CABLE ASSEMBLIES IN LIQUID.

DO NOT OPERATE THE UNIT IF THE EARTH CONNECTION IS SUSPECT OR IF THE MAINS LEAD IS DAMAGED OR SUSPECTED OF BEING DAMAGED.

DO NOT USE HIGH TEMPERATURE STERILISATION PROCESSES (SUCH AS AUTOCLAVING). DO NOT USE E-BEAM OR GAMMA RADIATION STERILISATION.

DO NOT USE SOLVENT CLEANERS

USE ONLY ACCESSORIES AND OTHER PARTS RECOMMENDED OR SUPPLIED BY SCHILLER AG. USE OF OTHER THAN RECOMMENDED OR SUPPLIED PARTS MAY RESULT IN INJURY INACCURATE INFORMATION AND/ OR DAMAGE TO THE UNIT.

THIS UNIT COMPLIES WITH EMC REGULATIONS FOR MEDICAL PRODUCTS WHICH AFFORDS PROTECTION AGAINST EMISSIONS AND ELECTRICAL INTERFERENCE. HOWEVER SPECIAL CARE MUST BE EXERCISED WHEN THIS UNIT IS USED WITH HIGH FREQUENCY EQUIPMENT.

IT MUST BE ENSURED THAT NEITHER THE PATIENT NOR THE ELECTRODES (INCLUDING THE NEUTRAL ELECTRODE) COME INTO CONTACT WITH OTHER PERSONS OR CONDUCTING OBJECTS (EVEN IF THESE ARE EARTHED).

THERE IS NO DANGER WHEN USING THE ECG UNIT FOR A PACEMAKER PATIENT OR WITH SIMULTANEOUS USE OF OTHER ELECTRICAL STIMULATION EQUIPMENT. HOWEVER, THE STIMULATION UNITS SHOULD ONLY BE USED AT A SUFFICIENT DISTANCE FROM THE ELECTRODES. IN CASE OF DOUBT, THE PATIENT SHOULD BE DISCONNECTED FROM THE RECORDER.

v

SCHILLER AT-2plus 6-Channel ECG Unit

SERVICE HANDBOOK Issue d May 2002

Safety Notices

WARNINGS & CAUTIONS

THIS UNIT IS CF CLASSIFIED ACCORDING TO IEC 601-1. THIS MEANS THAT THE PATIENT CONNECTION IS FULLY ISOLATED AND DEFIBRILLATION PROTECTED. SCHILLER CAN ONLY GUARANTEE PROTECTION AGAINST DEFIBRILLATION VOLTAGE, HOWEVER, WHEN THE ORIGINAL SCHILLER PATIENT CABLE IS USED.

WHEN NON-MEDICAL DEVICES ARE CONNECTED TO THE RS-232 INTERFACE ENSURE THAT BOTH UNITS ARE SECURELY CONNECTED TO THE SAME EARTH POTENTIAL.

WHEN OPERATING THE UNIT ON BATTERY AND SIMULTANEOUSLY USING NONMEDICAL DEVICES, THE RS-232 INTERFACE MUST BE FULLY ISOLATED.

AN EXTERNAL DEVICE MUST ONLY BE CONNECTED USING THE ORIGINAL SCHILLER INTERFACE CABLE ASSEMBLY.

BEFORE USING THE UNIT, ENSURE THAT AN INTRODUCTION REGARDING THE UNIT FUNCTIONS AND THE SAFETY PRECAUTIONS HAS BEEN PROVIDED BY A SCHILLER REPRESENTATIVE.

THE GUIDELINES FOR PATIENT ELECTRODE PLACEMENT ARE PROVIDED AS ON OVERVIEW ONLY. THEY ARE NOT A SUBSTITUTE FOR MEDICAL EXPERTISE.

THIS UNIT IS PROVIDED FOR THE EXCLUSIVE USE OF QUALIFIED PHYSICIANS OR PERSONNEL UNDER THEIR DIRECT SUPERVISION. THE NUMERICAL AND GRAPHICAL RESULTS AND ANY INTERPRETATION DERIVED FROM A RECORDING MUST BE EXAMINED WITH RESPECT TO THE PATIENTS OVERALL CLINICAL CONDITION. THE RECORDING PREPARATION QUALITY AND THE GENERAL RECORDED DATA QUALITY, WHICH COULD EFFECT THE REPORT DATA ACCURACY, MUST ALSO BE TAKEN INTO ACCOUNT.

IT IS THE PHYSICIANS RESPONSIBILITY TO MAKE THE DIAGNOSIS OR TO OBTAIN EXPERT OPINION ON THE RESULTS, AND TO INSTITUTE CORRECT TREATMENT IF INDICATED.

vi

SCHILLER AT-2plus 6-Channel ECG Unit

SERVICE HANDBOOK Issue d May 2002

What’s in this book

The service philosophy for the AT-2plus Memory is fault finding to module level. The purpose of this book is to provide all the information necessary to enable the service engineer to efficiently locate and replace a faulty module. This book assumes no detailed knowledge of the AT-2plus Memory but does require that the service engineer is familiar with standard workshop practices. The book is divided into the following chapters:

Chapter 1 — Operating Elements

The purpose of this chapter is to provide an easy reference for all the main operator functions and to give a basic introduction to the AT-2plus Memory. This chapter gives details of the operator controls with the operation and function of each key briefly explained. The information in this chapter provides a background to the operating functions only. Complete operating information is provided in the SCHILLER AT-2plus User Guide.

Chapter 2 — Functional Overview

This chapter provides a functional overview of the AT-2plus Memory. The description is supported by functional block diagrams.

Chapter 3 — Fault Diagnosis

This chapter provides a guide to locate a fault to module level. The diagnostics are presented in a logical sequence of fault finding algorithms and procedures. Illustrations are provided to support the text where needed.

Chapter 4 — Module Removal and Replacement

This chapter gives an overview of the physical construction of the AT-2plus Memory with the main physical attributes of the unit briefly described. The physical description is supported by illustrations showing the internal location of all modules. Removal and replacement instructions for all removable modules are also provided in this chapter. Each procedure is autonomous with details of tools, jumper settings, adjustments and settings or special requirements that are required before and after replacement. Functional checks that must be carried out after replacing a module are also provided.

Chapter 5 — Adjustments

This chapter provides all adjustments and settings. Also detailed in this chapter are basic functional test procedures that can be performed to check the functioning of the unit.

Chapter 6 — Spare Parts

This chapter provides the part numbers and reordering information for all replaceable modules. Also included in this chapter are details of any special test equipment or special tools required for adjustment or fault finding procedures.

vii

SCHILLER AT-2plus 6-Channel ECG Unit

SERVICE HANDBOOK Issue d May 2002

What’s in this book

Chapter 7 — Technical Data

The full technical specification of the AT-2plus Memory is given in this chapter.

Chapter 8 — Glossary

This chapter explains all the acronyms and signal titles used in this book and in the AT-2plus Memory circuit diagrams.

Circuit Diagrams & Board layouts

The circuit diagrams and component layouts are provided for all boards. These details are provided for information only.

viii

SCHILLER AT-2plus 6-Channel ECG Unit

Chapter 1

SERVICE HANDBOOK Issue d May 2002

Operating Elements

Chapter 1

Operating Elements

Contents

Introduction

1.2

List of Symbols

1.2

Location & Power

1.3

Switching On and Off

1.3

The Keyboard

1.4

LCD Screen

1.6

AT-2plus Short Form Operating Instructions

1.7

Modes of Operation

1.8

Automatic Mode

1.9

Manual Mode

1.10

Settings

1.12

Automatic Mode (ECG) Settings

1.20

Memory and Data Transmission Option

1.24

Erasing Memory Files

1.25

Transmitting Stored Files

1.25

Transmission Settings

1.25

Modem Transmission

1.26

Line Transmission

1.27

Software Updating via RS-232

1.28

Care & Maintenance

1.29

Replacing the Recording Paper

1.31

Thermal Paper Handling

1.32

Page 1.1

Chapter 1

SCHILLER AT-2plus 6-Channel ECG Unit

Operating Elements

SERVICE HANDBOOK Issue d May 2002

Introduction

The CARDIOVIT AT-2plus is a 6-channel ECG recorder with all (12) ECG signals simultaneously processed to provide instant ECG recordings. Two automatic recording modes can be individually preset to enable one button ECG recording of preferred print formats.

The AT-2plus includes the following features:

Low weight and compact dimensions

Large A4 size printout from integrated quality thermal printer

Built-in rechargeable battery for mains-independent use — 4hrs normal use or 300 printouts on one battery charge

Large, clear LCD for ECG preview prior to printing

Simple one key operation for main functions

Automatic or manual recording modes

Selectable printing formats

ECG memory for easy copying

Interpretation program option (including measurements) for children and adults

Alphanumeric keyboard for patient data entry and clinical comments

List of Symbols

Page 1.2

SCHILLER AT-2plus 6-Channel ECG Unit

Chapter 1

SERVICE HANDBOOK Issue d May 2002

Operating Elements

Location & Power

Location

Do not keep or operate the apparatus in a wet, moist, or dusty environment. Also, avoid exposure to direct sunlight or heat from other sources. Do not allow the unit to come into contact with acidic vapours or liquids, as such contact may cause irreparable damage. The unit should not be placed near X-ray or diathermy units, large transformers or motors. The unit must be placed on a flat surface and must not be operated in areas where there is any danger of explosion.

Power Supply

The mains connection is on the rear of the unit. The mains indicator lamp on the keyboard is always lit when the unit is connected to the mains supply. The unit can either be operated from the mains supply or from the built-in rechargeable battery. The power source isa indicted on the top line of the LCD.

Power Indication

HR: 76/min

Wed 20-AUG-96 18:20:21

R L F C1 C2

C3 C4 C5 C6

When mains is connected a mains symbol is displayed (as shown above). When the unit is running on battery power a battery symbol is displayed:

When battery capacity is limited, the battery symbol flashes on and off.

To recharge the battery, connect the apparatus to the mains supply by means of the supplied power cable. A totally discharged battery needs less than 15 hours to be fully recharged (60% in less than 3 hours, 90% in less than 7 hours). A fully charged battery gives approximately 4 hours of normal use. The unit can remain connected to the mains supply without any danger of damage to either the battery or the unit.

Switching On and Off

The CARDIOVIT AT-2plus is switched on with the green ON key

and off by means of the red OFF key

The unit is automatically switched off after 5 minutes (30 seconds if battery capacity is limited) if no key is pressed and the patient cable is not connected.

Potential Equalisation

If the AT-2plus is used in conjunction with other patient connected equipment, we recommend that the potential equalisation stud on the rear of the unit is connected to the hospital/ building common ground with the yellow/green ground cable (Part-no. 2.310005). When working from an emergency vehicle, the vehicle common ground can be used.

Page 1.3

Schiller AT-2+ Service Handbook

Chapter 1

SCHILLER AT-2plus 6-Channel ECG Unit

Operating Elements

SERVICE HANDBOOK Issue d May 2002

The Keyboard

Page 1.4

SCHILLER AT-2plus 6-Channel ECG Unit

Chapter 1

SERVICE HANDBOOK Issue d May 2002

Operating Elements

The Keyboard (cont.)

1Print extra copy — of Auto mode recording currently in memory. Press the ALT key first followed by this key to obtain a copy in Auto format 2.

2Display/enter patient data. When the patient data is displayed, pressing this key again returns to the ECG. Use the up/down arrows to go to the next data entry field.

In the `Born` (date of birth field), only the patients year of birth need be entered if desired. When two digits are entered (patients year of birth), the AT-2plus calculates the age of the patient according to the year entered. When the full DOB is entered, the age is calculated precisely.

3Myogram filter ON / OFF. The cutoff frequency can be defined and is detailed in `Settings`.

4.The top figures on the number keys designated > and < changes the lead group displayed on the screen.

5.Auto sensitivity key — automatically sets the ECG printout sensitivity ( in AUTO mode only) to the best setting for the signal strength (5mm/mV or 10mm/mV)

6.The top figures on the number keys designated 5, 10, and 20 set the sensitivity of the ECG both on the screen and on the (manual) printout. The sensitivity is 5, 10 or 20 mm / mV.

7.The top figures on the number keys designated 5/10, 25, and 50 set the speed of the ECG both on the screen and on the (manual) printout. The speed on the screen can only be set to 25 or 50 mm / s. The speed of the manual printout can be 5, 10, 25 or 50 mm/s. The 5 and 10 mm/s settings are both on the same key which toggles the two speeds.

8.The top character `QRS` toggles the QRS beeper ON/ OFF

9.Delete last typed character.

10.Switch the unit OFF.

11.Switch the unit ON.

12.Manual mode recording — start continuous printout of ECG — until STOP key pressed

13.Auto Mode recording (in Auto mode 1). Press ALT followed by the AUTO key for auto mode 2.

14.STOP printout / confirm (new) setting

15.ALT key — key for initiation of setups and selection of second format for printout and auto mode recording

16.In ECG mode use the UP/DOWN arrows to adjust screen contrast. When entering patient data use the LEFT/RIGHT arrow keys to move cursor in data field. Use the UP/DOWN arrow keys to go up/down to the next data entry

17.Mains Indicator — lit when mains connected.

Second letters on the keyboard — è, é, ç, ø are reached by holding the ALT key pressed before the letter key. Accents on a letter e.g. ô, ñ etc. are reached by pressing <SHIFT> and then the letter. In addition the following special characters are available (AT-2plus Memory only !):

Key combination: SHIFT +

1

2

3

4

5

6

7

8

9

0

Character

!

@

#

$

%

&

/

*

«

=

Page 1.5

Chapter 1

SCHILLER AT-2plus 6-Channel ECG Unit

Operating Elements

SERVICE HANDBOOK Issue d May 2002

LCD Screen

1.Current Heart Rate (averaged over 4 beats and refreshed every 2 seconds). The HR is also given on a manual printout. Note that with an auto mode printout the HR is averaged over the full 10 seconds of the recording.

2.Top line — Current Day, Date and Time Bottom Line — System messages

3.Top Line — Current power source — mains or battery. When battery capacity is limited the battery symbol flashes.

Bottom line — `ALT` in this box indicates that the ALT key has been pressed.

4.Electrode connections — when a lead flashes it indicates that the electrode resistance is too high. The electrode must be reapplied

5.Lead indication (leads currently displayed on the screen). Change the lead group with the keys `1` and `2`.

6.Myogram Filter indication — `Filter` = filter ON; no indication = filter OFF. Switch the filter on or off with the Filter key.

7.An `A` in this box indicates that automatic sensitivity is selected (auto mode printout only). Switch automatic sensitivity on or off with key `3`.

8.Sensitivity — 5, 10 or 20 mm/mV. Change the sensitivity with the keys `4`, `5`, and `6`.

Page 1.6

SCHILLER AT-2plus 6-Channel ECG Unit

Chapter 1

SERVICE HANDBOOK Issue d May 2002

Operating Elements

AT-2plus Short Form Operating Instructions

Automatic ECG Recording

Prepare skin, hook up patient.

Switch unit on, press ON .

Press and enter patient data.

Press again and wait for at least 10 seconds until a clear and stable trace is displayed.

Press AUTO to record and print.

Press COPY for additional copies.

Manual ECG Recording (Rhythm Strip)

·Prepare skin, hook up patient.

Switch unit on, press ON .

Press MAN START .

Change lead group with

1

and

2

.

Press STOP to stop the printout.

Electrode hook-up check

·

Press

ALT

1mV

AUTO

AUTO

for electrode check.

0

3

3

Best results are obtained when the electrode voltage readings (right column) are between ±50mV.

Filter On/Off

Press to switch the (Myogram) filter On / Off.

System Configuration

Press

ALT

1mV

to print system settings.

0

1

1

Page 1.7

Chapter 1

SCHILLER AT-2plus 6-Channel ECG Unit

Operating Elements

SERVICE HANDBOOK Issue d May 2002

Modes of Operation

Automatic Mode

Automatic Mode provides a printout giving 10 seconds of ECG recording of all 12 leads with a choice of 2 different formats.

Lead format and chart speed can be programmed freely for each of the 2 formats before recording.

With the optional interpretation program installed it is also possible to select the measurement table, average cycles with optional markings and interpretation statements for the printout.

For further information see paragraph `Settings for Automatic Mode`.

Manual Mode

Manual Mode provides a real time printout of 6 leads that are selected and indicated on the screen.

The following can be freely selected before or during recording:

Lead Group

Chart Speed

Sensitivity

Myogram Filter

For further information see paragraph `ECG Recording in Manual Mode` following.

Page 1.8

SCHILLER AT-2plus 6-Channel ECG Unit

Chapter 1

SERVICE HANDBOOK Issue d May 2002

Operating Elements

Automatic Mode

In automatic mode, a full 12-lead ECG is printed in one of two predefined formats with a sensitivity of 10 mm/mV. These two formats are selected by the user to suit his specific needs and requirements.

AUTO

When the AUTO SENSITIVITY key 3 is pressed before recording in automatic mode, the unit detects very large waveform amplitudes and sets the sensitivity for the extremity and/or precordial leads to 5 mm/mV to reduce the overlapping of traces. An `A` on the bottom line of the LCD indicates that Auto sensitivity is set.

To start the automatic ECG recording in Format 1, press the AUTO key:

To start the automatic recording in the second format, press the ALT key followed by the AUTO key:

ALT +

The printout gives the following:

ECG recording of all leads in either Standard or Cabrera format according to selection

Sensitivity

Heart Rate

Speed

Filter Settings

Time and Date

Interpretation statements

Average Cycles

Intervals

Axis

Sokolow Index (ECG index for hypertrophy)

Detailed Measurement Table

To obtain an extra printout of the ECG recording in Format 1, simply press the COPY key:

COPY

To obtain an extra printout of the second format, press the ALT key followed by the COPY key:

ALT — COPY

The Auto mode settings for the two formats are detailed in the paragraph entitled `Settings for Automatic Mode` later in this book

Page 1.9

Chapter 1

SCHILLER AT-2plus 6-Channel ECG Unit

Operating Elements

SERVICE HANDBOOK Issue d May 2002

Manual Mode

Manual mode provides a direct printout of the real-time ECG with full control of parameter selection.

To start the manual recording of a real-time ECG, press the MANUAL Printout key

To stop the manual recording (printout) press the STOP key

The printout provides you with the following:

Six (selected) leads with lead identification.

On the lower edge, the chart speed, user identification and filter settings (if on).

At the top, the heart rate as current average of 4 beats, trace sensitivity, and the time and date

The following can be freely chosen during or before the recording:

Lead Group

by means of the LEAD FORWARD and LEAD BACKWARD

key

1

2

The following lead groups are selectable:

I, II, III

aVR, aVL, aVF

(Cabrera: aVL, I, -aVR / II, aVF, III)

V1, V2, V3

/

V4, V5, V6

II, aVF, III

/

V2, V4, V5

V4, V5, V6

/

V7, V8, V9

Note:

The LCD only displays three leads at one time. When the lead forward or lead backward

key is pressed, the following /preceding three lead group is displayed

Leads

For software versions higher than 4.10 it is also possible to chose further leads. The desired leads are activated as shown in the table.

Select leads

Entry Key Sequence

Lead group

Confirm

0

Rhythm

ON

II, avF, III, V2, V4, V5

1

Rhythm

OFF

II, avF, III, V2, V4, V5

2

Left posterior

ON

V4, V5, V6, V7, V8, V9

3

Left posterior

OFF

V4, V5, V6, V7, V8, V9

4

Right precordial up to V5r

ON

ALT

7

9

V1, V2, V3, V3r, V4r, V5r

Press STOP

5

Right precordial up to V5r

OFF

key

V1, V2, V3, V3r, V4r, V5r

6

Right precordial up to V6r

ON

V1, V2, V3r, V4r, V5r, V6r

7

Right precordial up to V6r

OFF

V1, V2, V3r, V4r, V5r, V6r

8

NEHB

ON

D, A, J

9

NEHB

OFF

D, A, J

Page 1.10

SCHILLER AT-2plus 6-Channel ECG Unit

Chapter 1

SERVICE HANDBOOK Issue d May 2002

Operating Elements

Manual Mode (cont.)

Chart Speed

Select speed 5, 10, 25 or 50mm/s by means of the SPEED keys:

5/10

25

50

7

8

9

Notes:

Key 7 is a toggle key -press once and 5 is selected, press a second

time and 10mm/s is selected.

When the 25 or 50mm/s key is pressed, the same speed is set on

both the screen and the (manual) printout. When 5 or 10 mm/s is

selected, this affects the manual printout speed only.

Sensitivity

Select 5, 10 or 20 mm/mV by means of the SENSITIVITY keys:

5

10

20

4

5

6

Myogram Filter

Switch the filter ON or OFF with the FILTER key:

`FILTER` is displayed on the bottom line of the LCD when the

filter is switched on.

Recentering

To re-centre the ECG traces, press the 1mV key

1mV

0

WARNING

AFTER HEAVY ARTEFACTS OR LEAD OFF, THE INDICATION OF THE HEART RATE MAY NOT BE RELIABLE.

Page 1.11

Chapter 1

SCHILLER AT-2plus 6-Channel ECG Unit

Operating Elements

SERVICE HANDBOOK Issue d May 2002

Settings

Each parameter is set by means of a code. This code comprises a combination of keys starting with the ALT key followed by two or three numbers. The setting is confirmed with the STOP key. As soon as the ALT key is pressed, the keyboard is dedicated to the programming function.

Note:

When the ALT key is pressed `ALT` appears on the LCD (see previous page)

Note:

The Alternative (ALT) function is only active for 4 seconds. If a programming key is

not pressed within 4 seconds, the unit reverts to standard mode. The ALT key must

again be pressed to activate the programming mode

The setting is remembered and the keyboard released for other functions when the STOP key is pressed. Once a setting has been confirmed, it is stored in the memory even when the unit is switched off.

Example

If you want to reset your AT-2plus to the basic default settings, the key sequence given on page 14 is ALT; 0; 6; 6. STOP.

On the following pages the programmable parameters and the programming sequences are described in detail.

Page 1.12

SCHILLER AT-2plus 6-Channel ECG Unit

Chapter 1

SERVICE HANDBOOK Issue d May 2002

Operating Elements

Settings (cont.)

The defined formats and settings for your unit can be checked as follows:

ALT — 0 — 1 — 1

A printout of the defined settings will be produced and gives the following information, depending on the installed software:

Unit designation

Software version, Software option installed (C = Interpretation)

and interpretation version

Serial number

Serial number of the unit

Leads

Standard (S) or Cabrera (C)

ECG Format

Long (ooo), Short (o) or Suppressed (-)

MECG

Average cycles as defined in auto ECG recording setup (e.g. 4 *

3 (25 mm/s) + 2)

Measurements

Enabled (+) or Suppressed (-)

Marks

Enabled (+) or Suppressed (-)

Interpretation

Enabled (+) or Suppressed (-)

Selected Rhythm leads

Leads selected for R1, R2 resp.

Automatic Centering

Enabled (+) or Suppressed (-)

Printout of signals

Sequential or Simultaneous

Baseline Filter

0.05, 0.15 or 0.30 Hz

Mains Filter

50, 60 Hz or OFF (-)

Myogram Filter

25 or 35 Hz, ON (+) or OFF (-)

Page 1.13

Chapter 1

SCHILLER AT-2plus 6-Channel ECG Unit

Operating Elements

SERVICE HANDBOOK Issue d May 2002

Settings (cont.)

Interpretation settings:

N/A:+/- ‘normal/abnormal’ is written (+) or suppressed (-)

U:+/- ‘unconfirmed report’ is written (+) or suppressed (-)

A30:+/- patient age is assumed to be < 30 (-) or >30 (+)

S: +/- low (-) or high (+) sensitivity

For AT-2plus Memory only:

Automatic Save

Enabled (+) or Suppressed (-)

Automatic Erase

Enabled (+) or Suppressed (-)

Baud

Data transmission speed (9600 — 115200 bit/s, see page 1.25)

Com-Type

Communication mode, Line or Modem (see page 1.25)

Default Settings

To reset the unit to the basic default settings, proceed as follows:

ALT — 0 — 6 — 6

SETTINGS

STANDARD

WITH INTERPRETATION

LANGUAGE

AS SET

AS SET

LEADS

STANDARD (S)

STANDARD (S)

ECG : 25MM/S, SHORT (O)

MECG: 2*6 (50MM/S + 1)

AUTO FORMAT 1

ECG: 25MM/S, SHORT (O)

MEASUREMENTS:

SUPRESSED (-)

INTERPRETATION:

ENABLED (+)

MARKS: ENABLED (+)

ECG : 25MM/S, LONG

(OOO)

MECG: NONE

AUTO FORMAT 2

ECG: 25MM/S, LONG (OOO)

MEASUREMENTS:

SUPRESSED (-)

INTERPRETATION:

DISABLED (-)

MARKS: ENABLED (+)

RHYTHM LEADS

V1

V1, II

AUTOM.

ENABLED (+)

ENABLED (+)

CENTERING

PRINTOUT OF

SEQUENTIAL

SEQUENTIAL

SIGNALS

BASELINE FILTER

0.05HZ

0.05HZ

SETTING

MAINS FILTER

50HZ (60HZ)

50HZ (60HZ)

SETTINGS

MYOGRAM

35HZ, OFF

35HZ, OFF

FILTER SETTING

MEMORY AND

BAUD RATE 115200 BPS

BAUD RATE 115200 BPS

SERIAL

AUTO STORAGE ON (+)

AUTO STORAGE ON (+)

COMMUNICATIO-

AUTO DELETETION OFF (-)

AUTO DELETETION OFF (-)

N INTERFACE

OPTION

TRANS. MODE: LINE

TRANS. MODE: LINE

MEMORY

AUTO SAVE ENABLED (+)

AUTO SAVE ENABLED (+)

AUTO ERASE DISABLED (-)

AUTO ERASE DISABLED (-)

N/A: SUPRESSED (-)

INTERPRETATION

U: ENABLED (+)

SETTINGS

A30: UNDER THIRTY (-)

S: LOW (-)

Page 1.14

SCHILLER AT-2plus 6-Channel ECG Unit

Chapter 1

SERVICE HANDBOOK Issue d May 2002

Operating Elements

Settings (cont.)

Language

The language can only be set with the SWUP program.

User Identification

The user identification is printed on all recordings. The user ID can be the department, doctor or hospital etc. Enter the user ID as follows:

Press the ALT key followed by key 9, 3, 3

ALT — 9 — 3 — 3

The user entry field is displayed on the LCD. Enter up to 30 characters via the keyboard.

Confirm the new user ID by pressing the ENTER key.

Page 1.15

Chapter 1

SCHILLER AT-2plus 6-Channel ECG Unit

Operating Elements

SERVICE HANDBOOK Issue d May 2002

Settings (cont.)

Filters

There are three different filters which can be set individually as follows:

Baseline filter

Mains filter

Myogram filter

Baseline Filter

The digital Baseline filter suppresses excessive baseline drifts. The setting options are as follows:

Baseline Filter

Entry Key

Filter Setting

Confirm

Sequence

0

0.05 Hz

Press

(default)

ALT

5

STOP

1

0.15 Hz

key

3

0.30 Hz

Confirm the selection by pressing

STOP

Note:

The set value is the lower limit of the frequency range and is normally set to 0.05 Hz.

The settings 0.15 and 0.30 Hz should only be used when absolutely necessary, as the

possibility exists that they could affect the original ECG signal, especially the ST

segments.

Mains Filter

The Mains filter is an adaptive digital interference filter designed to suppress AC interference without attenuating or distorting the ECG.

Set the mains filter in accordance with the frequency of your local mains supply as follows:

Mains Filter

Entry Key

Filter Setting

Confirm

Sequence

5

Mains Filter 50 Hz

Press

ALT

8

6

Mains Filter 60 Hz

STOP

key

9

Mains Filter Off

Page 1.16

SCHILLER AT-2plus 6-Channel ECG Unit

Chapter 1

SERVICE HANDBOOK Issue d May 2002

Operating Elements

Settings (cont.)

Myogram Filter

The Myogram filter suppresses disturbances caused by strong muscle tremor. The set value will be the new upper limit of the frequency range as soon as the FILTER key is pressed on or programmed as default when the unit is switched on. When the Myogram filter is on `Filter` is displayed on the bottom line of the LCD.

Myogram Filter

Entry Key

Setting

Confirm

Sequence

2

Myogram Filter 25 Hz

3

Myogram Filter 35 Hz

Myogram Filter active

Press

1

when the unit is first

ALT

8

STOP

switched on (marked

key

on printout with +)

Myogram Filter off

8

when the unit is first

switched on (marked

on printout with -)

Confirm the selection by pressing the STOP key

STOP

The myogram filter is switched on and off manually with the FILTER KEY

Note:

An ECG recorded in auto mode is stored unfiltered. It is therefore possible to print the

stored ECG either with or without passing the myogram filter. Filter ON is indicated

in the bottom information line of the LCD. When the FILTER key is pressed again, the

filter is switched off and the `Filter` indication on the bottom information line of the

LCD is removed. The cutoff frequency of the myogram filter is set to either 25 or 35

Hz.

Page 1.17

Chapter 1

SCHILLER AT-2plus 6-Channel ECG Unit

Operating Elements

SERVICE HANDBOOK Issue d May 2002

Settings (cont.)

Defining Lead Sequence & Printout

The required settings can be selected as follows:

Sequences, Print & Auto-centering

Entry Key

Definition

Confirm

Sequence

1

Standard Lead

Sequence

2

Cabrera Lead

Sequence

Press

ALT

7

3

Simultaneous Print

STOP

key

4

Sequential Print

5

Auto-centering ON

6

Auto-centering OFF

Confirm the selection by pressing the STOP key

STOP

The selectable printout forms are:

Simultaneous

All ECG leads are printed in the same time segment (in automatic

mode only).

Sequential

Each group is a contiguous time segment of approximately 2.5 or

5 seconds (in automatic mode only).

Auto-Centering ON

All ECG traces are centred dynamically for optimal use of paper

width.

Auto-Centering OFF

ECG traces are set to a fixed baseline position and may possibly

overlap.

The Standard and Cabrera lead groups available for the AT-2plus are:

Lead Groups

Standard

Cabrera

I

V1

II

V4

aVL

V1

II

V4

II

V2

aVF

V5

I

V2

aVF

V5

III

V3

III

V6

-aVR

V3

III

V6

aVR

V4

V2

V7

II

V4

V2

V7

aVL

V5

V4

V8

aVF

V5

V4

V8

aVF

V6

V5

V9

III

V6

V5

V9

Note:

If the unit is reset to the default settings (see previous page), the user identification

must be re-entered

Page 1.18

SCHILLER AT-2plus 6-Channel ECG Unit

Chapter 1

SERVICE HANDBOOK Issue d May 2002

Operating Elements

Settings (cont.)

Acoustic QRS Indication

The acoustic QRS beep can be switched on or off at any time by pressing the QRS key

Time / Date

The required settings can be selected as follows:

Setting the Time and Date

Key Sequence

Enter Data

Confirmation

Time

ALT

9

1

1

HHMMSS

beep

Date

ALT

9

2

2

DDMMYY

beep

Seasonal Time Variation

Key Sequence

Wintertime to

ALT

9

4

4

Summertime (+1Hr)

Summertime to

ALT

9

5

5

Wintertime (-1Hr)

Page 1.19

Chapter 1

SCHILLER AT-2plus 6-Channel ECG Unit

Operating Elements

SERVICE HANDBOOK Issue d May 2002

Automatic Mode (ECG) Settings

Two separate Auto formats can be defined for the AT-2plus. When defining auto format 1 the key sequence ALT `1` precedes the setting. When defining auto format 2 the key sequence ALT `2` precedes the setting.

Automatic ECG Format

Entry Key

Setup Format

Sequence

1

Commence Setup for

Auto format 1

ALT

2

Commence Setup for

Auto format 2

The automatic mode formats are detailed on the following pages. The ECG format is set as follows:

ECG Format

Entry Key Sequence

Printout

Confirm

1

1page x 12 leads at 25mm/s

One page with the first 8 leads

2

printed for 5s and the last 4

leads printed for 10s

5

No leads printed

6

Leads are printed in short form

(1 sheet)

Press

ALT

1 or 2

1

7

Leads are printed in long form

(2 sheets)

STOP key

8

Chart Speed 25mm/s

9

Chart Speed 50mm/s

Leads are printed in format 4

0

* 3(25mm/s) + 1

rhythm(25mm/s)

Page 1.20

SCHILLER AT-2plus 6-Channel ECG Unit

SERVICE HANDBOOK Issue d May 2002

Automatic Mode (ECG) Settings (cont.)

Average Cycles

The Average cycles are defined as follows:

Note:

Lead selection for the rhythm lead(s) are defined on page 25.

Average Cycles (interpretation option only)

Entry Key Sequence

Printout

Confirm

5

No average lead

cycles are printed

4 x 3 (25 mm/s) + 2

rhythm leads

(25mm/s). The

6

average complexes

are printed in 4

groups of three leads

at a chart speed of

25mm/s

4 x 3 (50 mm/s) + 2

Press

1 or

rhythm leads

ALT

2

(25mm/s). The

STOP

2

7

average complexes

key

are printed in 4

groups of three leads

at a chart speed of

50mm/s

2 x 6 (50 mm/s) + 2

rhythm leads

(25mm/s). The

8

average complexes

are printed in 2

groups of six leads at

a chart speed of

50mm/s

Measurements and Markings (C version only)

To define the measurements and markings proceed as follows:

Measurements (Interpretation Option Only)

Entry Key Sequence

Printout

Confirm

Detailed table of

measurement results

omitted — however,

5

the values of electrical

axes, intervals, and

heart rate are not

suppressed.

1 or

Detailed table of

Press

6

measurement results is

ALT

3

STOP

2

printed

key

7

Referenece markings

are omitted

Reference markings

(beginning and end of

8

P wave and QRS,

and end of T wave)

are added to the ECG

average cycles

Chapter 1

Operating Elements

Page 1.21

Chapter 1

SCHILLER AT-2plus 6-Channel ECG Unit

Operating Elements

SERVICE HANDBOOK Issue d May 2002

Automatic Mode (ECG) Settings (cont.)

Interpretation (C version only)

To print or suppress interpretation statements on the printout proceed as follows:

Interpretation (Interpretation Option Only)

Entry Key Sequence

Printout

Confirm

5

Interpretation is

Press

1 or

omitted

ALT

4

STOP

2

Interpretation is

6

key

printed

Confirm the selection by pressing the STOP key

STOP

Full details of the interpretation option are given in the SCHILLER ECG Measurement and Interpretation booklet (Art. No. 2.510 179).

Interpretation Settings (C version only)

The interpretation settings enable the user to determine whether or not certain comments will be added to the interpretation statements on the ECG printout. Furthermore, the patient’s age can be defined (<30 or >30) and if low or high sensitivity should be applied. Low sensitivity will suppress certain nonspecific ECG diagnosis; this may be advisable when carrying out ECGs for screening.

Interpretation Settings

Entry Key

Setting

Confirm

Sequence

1

«Normal» / «Abnormal» is not

printed

2

«Normal» / «Abnormal» is printed

3

«Unconfirmed report» is not printed

Press

4

«Unconfirmed report» is printed

ALT

6

STOP

key

5

Patient age assummed to be < 30

6

Patient age assummed to be > 30

7

Low sensitivity

8

High sensitivity

Note:

The `Patient age assumed to be..` setting is only applicable when patient data has

not been entered.

Page 1.22

Loading…

SCHILLER AT-2plus 6-Channel ECG Unit

SERVICE HANDBOOK Issue d May 2002

What’s in this book

The service philosophy for the AT-2plus Memory is fault finding to module level. The purpose of

this book is to provide all the information necessary to enable the service engineer to efficiently

locate and replace a faulty module. This book assumes no detailed knowledge of the AT-2plus

Memory but does require that the service engineer is familiar with standard workshop practices.

The book is divided into the following chapters:

Chapter 1 — Operating Elements

The purpose of this chapter is to provide an easy reference for all the main operator functions and

to give a basic introduction to the AT-2plus Memory. This chapter gives details of the operator

controls with the operation and function of each key briefly explained. The information in this

chapter provides a background to the operating functions only. Complete operating information

is provided in the SCHILLER AT-2plus User Guide.

Chapter 2 — Functional Overview

This chapter provides a functional overview of the AT-2plus Memory. The description is supported

by functional block diagrams.

Chapter 3 — Fault Diagnosis

This chapter provides a guide to locate a fault to module level. The diagnostics are presented in

a logical sequence of fault finding algorithms and procedures. Illustrations are provided to support

the text where needed.

Chapter 4 — Module Removal and Replacement

This chapter gives an overview of the physical construction of the AT-2plus Memory with the

main physical attributes of the unit briefly described. The physical description is supported by

illustrations showing the internal location of all modules. Removal and replacement instructions

for all removable modules are also provided in this chapter. Each procedure is autonomous with

details of tools, jumper settings, adjustments and settings or special requirements that are required

before and after replacement. Functional checks that must be carried out after replacing a module

are also provided.

Chapter 5 — Adjustments

This chapter provides all adjustments and settings. Also detailed in this chapter are basic functional

test procedures that can be performed to check the functioning of the unit.

Chapter 6 — Spare Parts

This chapter provides the part numbers and reordering information for all replaceable modules.

Also included in this chapter are details of any special test equipment or special tools required for

adjustment or fault finding procedures.

vii

Schiller Supplies

Schiller Products

Categories

Nav Menu 5

Nav Menu 6

Nav Menu 7

Nav Menu 8

Loading…

SCHILLER AT-2plus 6-Channel ECG Unit

SERVICE HANDBOOK Issue d May 2002

What’s in this book

The service philosophy for the AT-2plus Memory is fault finding to module level. The purpose of

this book is to provide all the information necessary to enable the service engineer to efficiently

locate and replace a faulty module. This book assumes no detailed knowledge of the AT-2plus

Memory but does require that the service engineer is familiar with standard workshop practices.

The book is divided into the following chapters:

Chapter 1 — Operating Elements

The purpose of this chapter is to provide an easy reference for all the main operator functions and

to give a basic introduction to the AT-2plus Memory. This chapter gives details of the operator

controls with the operation and function of each key briefly explained. The information in this

chapter provides a background to the operating functions only. Complete operating information

is provided in the SCHILLER AT-2plus User Guide.

Chapter 2 — Functional Overview

This chapter provides a functional overview of the AT-2plus Memory. The description is supported

by functional block diagrams.

Chapter 3 — Fault Diagnosis

This chapter provides a guide to locate a fault to module level. The diagnostics are presented in

a logical sequence of fault finding algorithms and procedures. Illustrations are provided to support

the text where needed.

Chapter 4 — Module Removal and Replacement

This chapter gives an overview of the physical construction of the AT-2plus Memory with the

main physical attributes of the unit briefly described. The physical description is supported by

illustrations showing the internal location of all modules. Removal and replacement instructions

for all removable modules are also provided in this chapter. Each procedure is autonomous with

details of tools, jumper settings, adjustments and settings or special requirements that are required

before and after replacement. Functional checks that must be carried out after replacing a module

are also provided.

Chapter 5 — Adjustments

This chapter provides all adjustments and settings. Also detailed in this chapter are basic functional

test procedures that can be performed to check the functioning of the unit.

Chapter 6 — Spare Parts

This chapter provides the part numbers and reordering information for all replaceable modules.

Also included in this chapter are details of any special test equipment or special tools required for

adjustment or fault finding procedures.

vii

Schiller Supplies

Schiller Products

Categories

Nav Menu 5

Nav Menu 6

Nav Menu 7

Nav Menu 8


 
Home >
SHOP > ECG / EKG >




User Manual for AT-2 Plus ECG & Spirometry Combo (Electronic Edition)

User Manual for AT-2 Plus ECG & Spirometry Combo (Electronic Edition)

 

Alternative Views:



Price: $19.95

Availability:: Available for download now

Product Code: 2.510220I

Qty:


Description


Technical Specs

 


Electronic Edition — Can be downloaded now.  
User Manual for Schiller AT-2 Plus ECG & Spirometry Combo devices
Includes: Quick Reference Guide (Part No. 2.510255c)
               Spirometry User Guide (Part No. 2.510362e)

Features

  • User Manual for Schiller AT-2 Plus ECG & Spirometry Combo devices
  • Includes: Quick Reference Guide (Part No. 2.510255c)
  • Spirometry User Guide (Part No. 2.510362e)

Related Products…

EXEC Stress Upgrade — Software for AT-10 Plus SCHILLER Exercise ECG Analysis Program

Case of Recording Paper for CS-200, All AT-2 Series

Blue Sensor Electrodes for Stress ECG

10-wire patient cable, snap type, AHA for FT-1

Price: $430.00

Add

Price: $189.00

Add

Price: $35.00

Add

Price: $260.00

Add

EXEC Stress Upgrade - Software for AT-10 Plus SCHILLER Exercise ECG Analysis Program

Case of Recording Paper for CS-200, All AT-2 Series

Blue Sensor Electrodes for Stress ECG

10-wire patient cable, snap type, AHA for FT-1

Chart Paper for CARDIOVIT MS-2015

SNAP-CLIP Adapter for Resting ECG.

Case of Recording Paper for AT-101, LCM, DG 5000 and DG 6002

Case of paper for FT-1

Price: $20.00

Add

Price: $54.00

Add

Price: $169.50

Add

Price: $66.00

Add

Chart Paper for CARDIOVIT MS-2015

SNAP-CLIP Adapter for Resting ECG.

Case of Recording Paper for AT-101, LCM, DG 5000 and DG 6002

Case of paper for FT-1

Share your knowledge of this product with other customers… Be the first to write a review

Browse for more products in the same category as this item:

SHOP > ECG / EKG
SHOP > ECG / EKG > Parts
SHOP

Понравилась статья? Поделить с друзьями:
  • Scarlett grand barista pro инструкция
  • Schiller at 1 инструкция по эксплуатации на русском
  • Scarlett sc 143 пароварка инструкция
  • Scheriproct n инструкция по применению на русском языке
  • Scarlett england массажер инструкция по применению